Motion Solutions for Digital Pathology. White Paper

Similar documents
Motion Solutions for Digital Pathology

Fiber Optic Device Manufacturing

Mini-MAG Positioning Products

OPTICS IN MOTION. Introduction: Competing Technologies: 1 of 6 3/18/2012 6:27 PM.

P-611.Z Piezo Z-Stage

Compact Nanopositioning System Family with Long Travel Ranges

PIMag Precision Linear Stage

nano Motion Technology ANT130XY Series Two-Axis XY Direct-Drive Nanopositioning Stages ANT130XY Series NANO Technology Introduction

10 Things to Consider when Acquiring a Nanopositioning System

Synchronization Control Scheme for Hybrid Linear Actuator Based on One Common Position Sensor with Long Travel Range and Nanometer Resolution

PicoMaster 100. Unprecedented finesse in creating 3D micro structures. UV direct laser writer for maskless lithography

M-041 M-044 Tip/Tilt Stage

Advanced Motion Control Optimizes Laser Micro-Drilling

PICO MASTER 200. UV direct laser writer for maskless lithography

ATS50 Series. Mechanical Bearing, Lead-Screw Stage. Low profile, small footprint. Ultra-fine resolution

High accurate metrology on large surface areas with low reflectivity

Screw Driven automation tables

62xxH Series Galvanometer Scanners

The Air Bearing Throughput Edge By Kevin McCarthy, Chief Technology Officer

Machine Tools with an Enhanced Ball Screw Drive in Vertical Axis for Shaping of Micro Textures

An Introduction To Plug-and- Play Motion Subsystems

profile Using intelligent servo drives to filter mechanical resonance and improve machine accuracy in printing and converting machinery

Q-Motion Miniature Linear Stage

Press release. New 450 mm positioning system is twice as fast! Revolutionary Twin Gantry from Schneeberger

Active Vibration Isolation of an Unbalanced Machine Tool Spindle

WHO WE ARE MISSION STATEMENT

OPTOFORM 40 ENGINEERING SPECIFICATIONS

hurryscan, hurryscan II

PIglide AT3 Linear Stage with Air Bearings

Advanced Motion Control Optimizes Mechanical Micro-Drilling

KRONOS S. Key data. Precision for small workpieces. A member of the UNITED GRINDING Group

LEAD SCREW LINEAR ACTUATORS Profile Rail Linear Actuators

Using Autofocus in NIS-Elements

Linear vs. PWM/ Digital Drives

HLD117 Linear Motor Stage Series

V2018 SPINSTAND AND NEW SERVO-8 SYSTEM

ProScan DC Linear Servo Stage Technology

Upgrading from Stepper to Servo

EXC500p-- PATHOLOGY MICROSCOPE. EXC500hd -- HD DIGITAL PATHOLOGY MICROSCOPE. EXC500r -- RESEARCH MICROSCOPE EXC500-LABORATORY SCOPE

SGL Series. Single Guide Linear Motor Stage. Zero cogging and backlash ironless linear motor actuator. High speed and high acceleration

Computer Numeric Control

nanovea.com PROFILOMETERS 3D Non Contact Metrology

Considerations: Evaluating Three Identification Technologies

P-810 P-830 Piezo Actuators

Engineering Reference

All Servos are NOT Created Equal

Bring the Challenge. We ll Build the Solution.

Key data. Precision for small workpieces. A member of the United Grinding Group

INDIAN INSTITUTE OF TECHNOLOGY BOMBAY

P-736 PInano Z Microscope Scanner for Microtiter Plates

Why Nanopositioning is More than Just Nanometers or How to Find a State-of-the-Art System

attocube systems Probe Stations for Extreme Environments CRYOGENIC PROBE STATION fundamentals principles of cryogenic probe stations

Nmark AGV-HPO. High Accuracy, Open Frame, Thermally Stable Galvo Scanner. Highest accuracy scanner available attains singledigit,

Near-field Optical Microscopy

R30D RVDTs DC-Operated Rotary Variable Differential Transformers

Nmark AGV-HP. High Accuracy, Thermally Stable Galvo Scanner

This is how PI Does Measuring - Part I

Absolute exposed linear encoder For high accuracy and large measuring length

Introducing the New DMC-42x0 Ethernet Controller

Diagnosis and compensation of motion errors in NC machine tools by arbitrary shape contouring error measurement

Innovations in PWM Drive Technology for Nanometer Level Motion Control Applications. May 14, 2014 Jason Goerges

Magnetic Direct Drives and Air Bearing Technology PRECISION ENGINEERING AND MOTION CONTROL EXPERTISE

Nmark AGV-HP. High Accuracy, Thermally Stable Galvo Scanner

TECHNICAL DATA OPTIV CLASSIC 432

OTC18 OVERHEAD TUBE CRANE SYSTEM

Key data. Maximum performance for large workpieces. A member of the United Grinding Group

Key data. Flexibility for medium-sized workpieces. A member of the United Grinding Group

Introduction to Servo Control & PID Tuning

MPI TS300-SE 300 mm Manual Probe System with ShielDEnvironment TM For accurate and reliable DC/CV, RF and mmw measurements

Measurement of Microscopic Three-dimensional Profiles with High Accuracy and Simple Operation

OPTIV CLASSIC 321 GL TECHNICAL DATA

GUZIK V2002 Spinstand with XY-Positioning For Head, Headstack and Disk Testing

High Performance Low Voltage Servo Drives

ATS 200 & ATS 400 Alignment Turning Stations with Ful y Integrated Measurement Technology

Fast Optical Form Measurements of Rough Cylindrical and Conical Surfaces in Diesel Fuel Injection Components

PiezoMike Linear Actuator

SCANLAB Perfect Scanning

Compact and Reliable Speckle Reduction

Q-Motion Miniature Linear Stage

The Air Bearing Advantage

PICO MASTER. UV direct laser writer for maskless lithography

Outline: Introduction: What is SPM, history STM AFM Image treatment Advanced SPM techniques Applications in semiconductor research and industry

Technical Explanation for Displacement Sensors and Measurement Sensors

When the machine makes a movement based on the Absolute Coordinates or Machine Coordinates, instead of movements based on work offsets.

Mechatronics. Bring the challenge. We ll build the solution.

ATX115SL/SLE Series Mechanical-Bearing, Screw-Driven Linear Stage

FemtoFAB. Femtosecond laser micromachining system. tel fax Konstitucijos ave. 23C LT Vilnius, Lithuania

production competitive Quality Vision International Defines Accuracy Townsend Machine Transforms Their Business New Makino a51nx and

Size 23 Single Stack Size 23 Double Stack. 30-in (760 mm) 225 lbs (1,000 N) lbs-ft (30.5 Nm) lbs-ft (26.25 Nm) lbs-ft (30.

Be aware that there is no universal notation for the various quantities.

High Precision Gantry Motion Systems

Optimizing Performance Using Slotless Motors. Mark Holcomb, Celera Motion

Unit-25 Scanning Tunneling Microscope (STM)

Prepare Sample 3.1. Place Sample in Stage. Replace Probe (optional) Align Laser 3.2. Probe Approach 3.3. Optimize Feedback 3.4. Scan Sample 3.

the Art of turning

TECHNICAL DATA. OPTIV CLASSIC 322 Version 3/2013

Compact Photonics Control Solutions

6-Axis Nanopositioning Systems

Cutting-edge Atomic Force Microscopy techniques for large and multiple samples

Electronic Design of a Semi-Automated Micromanipulator Cell Injection System

Transcription:

Motion Solutions for Digital Pathology White Paper

Design Considerations for Digital Pathology Instruments With an ever increasing demand on throughput, pathology scanning applications are some of the toughest to solve. Digital pathology is a rapidly growing segment of in vitro diagnostics with the potential to streamline the overall pathology process for the benefit of both doctors and patients. This relatively new market segment requires motion systems that support both high-quality and high-speed imaging. In addition, the scanners are dispersed globally, which drives the demand for extremely high reliability. Lastly, as with all diagnostic instruments, laboratory space is at a premium, so minimizing the overall footprint is critical. Automated pathology scanners require unique positioning specifications that are not found in off-the-shelf components and will not fit into the instrument required footprint. Because of the packaging challenges, the motion system is often integrated into structural elements of the instrument. Product consistency is required for high-volume manufacturing at a globally competitive cost. Whether you are designing your own pathology scanning stage or working with a motion control partner, this paper covers key considerations for creating a successful design. Figure 1: In addition to requiring high throughput, digital pathology electromechanical systems must have excellent straightness and flatness and produce extremely smooth motion. Jim Monnich Jim is the Engineering Manager in Parker s Life Science Automation group. He has over 30 years of experience in the electrical motion and control industry, specializing in mechanical and electrical design. His primary expertise is in servo and stepper systems and position feedback technologies, as well as design and development of motion mechanics focusing on the areas of submicron positioning and ultra-precision velocity control for scanning. Brian Handerhan Brian is a Business Development Manager focusing on Parker s Life Science Automation group. He has over 20 years of experience in the implementation of automation across a broad range of industries. His primary expertise has been as a process improvement leader, change agent and P&L owner. He now focuses that broad experience working with industry OEM s to develop lasting business relationships built on both operational and technical value. Digital Pathology Motion Overview Pathology scanners present OEMs with some of the most demanding motion control requirements in life science equipment. These scanners rely on the quality and performance of positioning systems to allow for the capture of the highest definition images possible with their imaging components. The advent of digital whole slide imaging has made these devices some of the most promising tools in diagnostic medicine today, producing better quality, faster and less costly prediction, diagnosis and prognosis of diseases such as cancer. In a pathology scanner, several slides are loaded into a cassette. The cassette is placed on an XY stage, which is mounted to the

scanner frame. A vertical Z stage is mounted to the scanner frame above the XY stage, carrying the focusing optics. During the scanning process, the XY stages move the slides under the camera at a typical rate of 5 mm/s. The camera takes digital images within its field of view, which is on the order of 1 mm 2. While the XY stages are moving, to cover the entire slide area, which is typically on the order of 50 mm x 25 mm, the Z-axis focuses the camera and holds a stable position for a fraction of a second, while snapshots are being taken by the camera. Figure 2: An example of a high precision, monolithic pathology scanner All of this must be accomplished, of course, within a defined space. The market demands a minimized footprint, which factors in the given camera s shape and size, illumination requirements and special slide holder requirements. The space constraints required limit the choice of XYZ stage configurations, often requiring low profiles, built-in apertures and highly customized axes with miniaturized motion control components. Scanning Stages The scanning XY stages typically move the slides beneath a fixed microscope or camera system so that images can be captured at a high rate of speed. To improve throughput, images are often captured on-the-fly. The key motion characteristics are: Flatness scanner flatness must be tightly controlled to reduce the dynamic of the Z-axis to autofocus. Straightness scanner straightness is required to minimize overlap between scanning passes and increase process throughput. Smoothness of motion smoothness of motion is required to assure the highest possible image resolution Flatness Flatness of motion is determined by several factors, including flatness of the slide and base surfaces of the stage and flatness of the linear guidance system. Typical flatness requirements for pathological scanners are on the order of submicron per mm. For example, consider an X stage which scans at 5 mm/s and has a flatness of +/- 1 µm/mm. This flatness and scanning rate represents a position disturbance to the autofocusing stage with an amplitude of 1 µm and a frequency of 5 Hz. This in turn implies that the required bandwidth of the Z stage should be at least three times higher than the disturbance frequency. It also implies that the natural frequency will be three times higher than the bandwidth. The flatness of Parker scanners, as an example, is on the order of submicron levels. This flatness performance is achieved by using linear bearing technologies that are specifically selected for this type of application. Straightness Straightness is determined primarily by the rails and their assembly procedures. These systems require a straightness spec, using high-precision rails and special assembly fixtures, that are smaller than +/- 0.25 µm/mm. Smoothness of Motion As shown in Figure 3, this system is using miniature linear motors to drive the XY stages with linear encoders as feedback. Miniature linear motors have the advantage of being compact with low-profile size and are completely noncontact. They therefore have an advantage in applications with limited space, such as pathology scanners. Linear motors are virtually jitter- and friction-free positioning technology, resulting in high smoothness of motion and long life. The linear encoders used in Parker scanners have resolutions on the order of 10 nanometers. The mounting design of feedback and other elements critical to the operation are designed to achieve very high system response and reduce vibrations, which may otherwise limit the optical resolution. Figure 3: An example of a Parker single linear motor stage (upper right) and an XYZ stage (lower left).

Autofocus Z Stage Options The Z-axis stage, as shown in Figure 4, supports the optical head, including lenses and mounting bracket for focal adjustment. The operation of the autofocus stage is critical to the performance of the scanner and to the image clarity of the scanned objects. It must have extremely high-resolution (in the nanometer range), fast response, short settling time, stable holding position and repeatable operation. It also requires a precision controller with very consistent performance from one stage to the next. The following are alternative options for this stage configuration: Stacked Piezo Stage Stacked piezo stages have very fast responses and excellent position holding stability. Their limitations are in positioning hysteresis, which makes the servo control nonlinear and more difficult to tune. They also have extremely short travel, on the order of a few hundred µm, which limits other tasks needed by the Z stage, such as mapping the entire XY range for Z height, or adapting to several chuck sizes or optics with different heights. Hybrid Piezo Stage Its disadvantages are its larger size, additional servo axis to control and difficult tuning of both stages. Walking Piezo Stage These stages have the advantages of miniature size, holding stability and long travel. The disadvantage is in their change of performance over time. Most OEMs require that the system will be tuned with a consistent set of parameters for all similar stages. They expect all stages of the same type to act the same way over the lifetime of their systems. Voice Coil or Linear Motor Stages These stages have the advantages of offering long travel, miniature size, frictionless operation, low maintenance and consistent performance over the lifetime of the stage. Their disadvantage is in the need for a counterbalance in vertical applications. Parker has developed unique solutions to solve this limitation, including, among others, a non-contact magnetic counterbalance with a uniform force constant throughout the travel range of the state, with a repeatability of 0.1 µm. Recommendations for Additional Design Considerations For high smoothness of motion and short settling time, there are several additional design considerations OEMs should consider for any positioning systems being developed: Cables that are part of the design must be managed properly to minimize rattling effects. Wires need to be shielded and grounded to a common ground to reduce EMI and RFI noise. The equipment frame should be isolated from ground disturbances. Structural damping should be optimized to attenuate high-frequency noise amplification. Stage structure and machine frame natural frequency should be designed to meet the required move and settle frequency response. A controller s capability should be matched to the required frequency response. Proper servo filtering will attenuate high frequency noise. Hybrid piezo stages consist of a combination of a ball screw stage and a stacked piezo stage mounted on top. The ball screw stage provides long travel at low accuracy. The precision stacked piezo stage provides holding stability and precise positioning. The advantage of this hybrid configuration is in increased travel as a result of the two positioning technologies used. Figure 4: An example of a magnetically counterbalanced Parker linear motor stage used for autofocusing.

Manufacturing the Solution When looking for positioning solutions for automated pathology scanners, the key is understanding the importance and role of every component in the required design. These products have very specific and unique positioning motion specification requirements that off-the-shelf products often are not designed to meet. Positioning solutions can meet these requirements, but producing the best possible custom-designed solutions requires a technology partner with specialized knowledge and experience in the design and manufacturing of custom positioning systems as well as their components, with access to testing environments specific to the life science industry. Following a careful design process with experienced partners will lead to the best design solution that can also be affordably produced and still offer the reliability and demanding performance required for life science equipment. 2018 Parker Hannifin Corporation Motion Solutions for Digital Pathology 01/18 Parker Hannifin Corporation Electromechanical & Drives Division 1140 Sandy Hill Road Irwin, PA 15642-3049 724-861-8200 www.parker.com/emdusa