Enhancing Gray-to-White Matter Contrast in 3T T1 Spin-Echo Brain Scans by Optimizing Flip Angle

Similar documents
Simultaneous Multi-Slice (Slice Accelerated) Diffusion EPI

Inherent Insensitivity to RF Inhomogeneity in FLASH Imaging

3T Unlimited. ipat on MAGNETOM Allegra The Importance of ipat at 3T. medical

The SENSE Ghost: Field-of-View Restrictions for SENSE Imaging

HETERONUCLEAR IMAGING. Topics to be Discussed:

Half-Pulse Excitation Pulse Design and the Artifact Evaluation

Pulse Sequence Design and Image Procedures

MRI Metal Artifact Reduction

Supplementary Figure 1

Optimizing the Magnetization-Prepared Rapid Gradient-Echo (MP-RAGE) Sequence

Gradient Spoiling. Average balanced SSFP magnetization Reduce sensitivity to off-resonance. FFE, FISP, GRASS, GRE, FAST, Field Echo

The development of the RF-pulse for the low level SAR used by the MRI.

ISSN X CODEN (USA): PCHHAX. The role of dual spin echo in increasing resolution in diffusion weighted imaging of brain

H 2 O and fat imaging

Advanced MSK MRI Protocols at 3.0T. Garry E. Gold, M.D. Associate Professor Department of Radiology Stanford University

Background (~EE369B)

Applications Guide. Spectral Editing with SVS. (Works-in-Progress) MAGNETOM TaTs and Verio Systems (3T)

Pulse Sequence Design Made Easier

High Field MRI: Technology, Applications, Safety, and Limitations

Experience in implementing continuous arterial spin labeling on a commercial MR scanner

Cardiac MR. Dr John Ridgway. Leeds Teaching Hospitals NHS Trust, UK

MRI Summer Course Lab 2: Gradient Echo T1 & T2* Curves

IR/SR TrueFISP. Works-in-Progress package Version 1.2. For the SIEMENS Magnetom. Installation and User s Guide NUMARIS/4VA21B.

Clear delineation of optic radiation and very small vessels using phase difference enhanced imaging (PADRE)

Imaging the brain at ultra-high resolution using 3D FatNavs

MRI physics for SPM users

1 Introduction. 2 The basic principles of NMR

MR Basics: Module 8 Image Quality

NEMA Standards Publication MS (R2014) Determination of Signal-to-Noise Ratio (SNR) in Diagnostic Magnetic Resonance Imaging

10. Phase Cycling and Pulsed Field Gradients Introduction to Phase Cycling - Quadrature images

Works-in-Progress package Version 1.0. For the SIEMENS Magnetom. Installation and User s Guide NUMARIS/4VA21B. January 22, 2003

Pulse Sequences: Rapid Gradient Echo

Image Quality/Artifacts Frequency (MHz)

SIGNA Explorer Lift revives our MR

PET Performance Evaluation of MADPET4: A Small Animal PET Insert for a 7-T MRI Scanner

TimTX TrueShape. The parallel transmit architecture of the future. Answers for life.

RAD 229: MRI Signals and Sequences

High-Field Surface-Coil MR Imaging of Localized Anatomy

Field Simulation Software to Improve Magnetic Resonance Imaging

(N)MR Imaging. Lab Course Script. FMP PhD Autumn School. Location: C81, MRI Lab B0.03 (basement) Instructor: Leif Schröder. Date: November 3rd, 2010

MRI SYSTEM COMPONENTS Module One

Liver imaging beyond expectations with Ingenia

MR Advance Techniques. Flow Phenomena. Class II

M R I Physics Course. Jerry Allison Ph.D., Chris Wright B.S., Tom Lavin B.S., Nathan Yanasak Ph.D. Department of Radiology Medical College of Georgia

Improve Image Quality of Transversal Relaxation Time PROPELLER and FLAIR on Magnetic Resonance Imaging

SIEMENS MAGNETOM Skyra syngo MR D13

Numerical Evaluation of an 8-element Phased Array Torso Coil for Magnetic Resonance Imaging

Multi-channel SQUID-based Ultra-Low Field Magnetic Resonance Imaging in Unshielded Environment

Supplementary Material

Slice profile optimization in arterial spin labeling using presaturation and optimized RF pulses

Principles of MRI EE225E / BIO265. Lecture 21. Instructor: Miki Lustig UC Berkeley, EECS. M. Lustig, EECS UC Berkeley

Functional MRI with variable echo time acquisition

TITLE: Prostate Cancer Detection Using High-Spatial Resolution MRI at 7.0 Tesla: Correlation with Histopathologic Findings at Radical Prostatectomy

Fast Field-Cycling Magnetic Resonance Imaging (FFC-MRI)

Hardware. MRI System. MRI system Multicoil Microstrip. Part1

Personal Information. Contact information. Education / Training. Position

12/21/2016. Siemens Medical Systems Research Agreement Philips Healthcare Research Agreement AAN and ASN Committees

ACRIN 6686 / RTOG 0825

MAGNETIC RESONANCE IMAGING

Attenuation Correction in Hybrid MR-BrainPET Imaging

Downloaded from by on 02/07/18 from IP address Copyright ARRS. For personal use only; all rights reserved

2015 Spin echoes and projection imaging

Encoding of inductively measured k-space trajectories in MR raw data

2014 M.S. Cohen all rights reserved

Fundamental and Clinical Studies for Effectiveness of Zero-filling Interpolation on k-space for Improvement of Sharpness in Magnetic Resonance Imaging

Optimization of Gradient-Echo MR for Calcium Detection

Image quality evaluation of turbo-spin echo diffusion weighted image (TSE-DWI) : A phantom study

Iterative Reconstruction in Image Space. Answers for life.

Correction of the local intensity nonuniformity artifact in high field MRI

Alae Tracker: Tracking of the Nasal Walls in MR-Imaging

Module 2. Artefacts and Imaging Optimisation for single shot methods. Content: Introduction. Phase error. Phase bandwidth. Chemical shift review

Fast cine MRI of the orbit

Standards for Imaging Endpoints in Clinical Trials: Standardization and Optimization of Image Acquisitions: Magnetic Resonance

radiography detector

MRI Phase Mismapping Image Artifact Correction

Tissue classification based on relaxation environments

Coil Overlook Coil in MRI system TEM Coil Coil Overlook

BOLD fmri: signal source, data acquisition, and interpretation

RF pulse design and the Small Tip Angle Approximation

Application Guide & Release Notes

Functional Connectivity Mapping for Correlated Resting State Image Volumes

Analysis of spatial dependence of acoustic noise transfer function in magnetic resonance imaging

Virtual monochromatic imaging in dual-source and dual-energy CT for visualization of acute ischemic stroke

DESIGN PRINCIPLES FOR INSULATED INTERNAL LOOPLESS MRI RECEIVERS

Human Subjects in fmri Research

Optimization of Axial Resolution in Ultrasound Elastography

Keith S Cover, PhD. Department of Physics and Medical Technology, VU University Medical Center, Amsterdam

Besides long acquisition times and restricted spaces, acoustic noise

MR Basics: Module 6 Pulse Sequences

Magnetic Resonance Imaging Principles, Methods, and Techniques

Experimental Investigation of Crack Detection in Cantilever Beam Using Natural Frequency as Basic Criterion

A. SPECIFIC AIMS: phase graph (EPG) algorithms to cover a wide range of MRI

Receive Arrays and Circuitry

Magnetic Resonance Imaging

Diff Ganzkörper ep2d_diff_stir_b50_800_tra

Evaluation of MWT Materials Accusorb MRI Shield. Dr. E. Kanal, Department of Radiology, UPMC. Wednesday, April 21, 2010

Medical Imaging. X-rays, CT/CAT scans, Ultrasound, Magnetic Resonance Imaging

Steady-state sequences: Spoiled and balanced methods

Introduction to MR Hardware. RF Coils C M L C T. = g * B 0. Rotating magnetization produces alternating magnetic field

Acoustic noise reduction of MRI systems by means of magnetic shielding

Transcription:

AJNR Am J Neuroradiol 26:2000 2004, September 2005 Enhancing Gray-to-White Matter Contrast in 3T T1 Spin-Echo Brain Scans by Optimizing Flip Angle Bernd L. Schmitz, Georg Grön, Florian Brausewetter, Martin H.K. Hoffmann, and Andrik J. Aschoff BACKGROUND AND PURPOSE: Compared with MR imaging at 1.5T T1-weighted spin-echo imaging at 3T shows up with reduced gray-to-white matter contrast. The purpose of the present study was to show the effects of alterations of different flip angles as an easily accessible parameter to increase gray-to-white matter contrast. METHODS: Spin-echo T1 sequences of 6 healthy volunteers were acquired in a 3T head scanner with 5 different flip angles. Observer-independent contrast-to-noise ratios for gray versus white matter from different flip angles, as well as subjective ratings of image quality from 2 blinded neuroradiologists, were compared statistically. RESULTS: Gray-to-white matter contrast increased significantly with decreasing flip angle. No artifacts were introduced by decreasing flip angles, and T1 contrast characteristics were robust and stable at lowered flip angles. Also, specific absorption ratios significantly decreased with decreasing flip angles. CONCLUSION: Using a flip angle of 50 significantly increases gray-to-white matter contrast in T1 spin-echo brain scans at 3T B0 field strength. To an increasing extent, 3T high-field-strength scanners have entered clinical routine (1, 2). In addition to the theoretical doubling in signal intensity, there are some limitations (3) associated with going to higher field strengths that, in turn, are mainly due to the doubling of spin resonance frequencies when compared with 1.5T. One of the major changes of brain imaging at higher magnetic fields is the rather low gray-to-white matter contrast in spin-echo T1 sequences. These sequences are widely used in clinical imaging, and neuroradiologists diagnostic decisions frequently rely heavily on this usually robust technique; however, the contrast reduction in spin-echo T1 sequences has already led to doubts on the usefulness of these sequences at higher field strengths in routine clinical brain imaging (4). Moreover, increasing resonance frequency at higher fields often causes problems with limits in individual specific absorption rates (SARs). In contrast to the linear increase in signal intensity with Received September 22, 2004; accepted after revision February 4, 2005. From the Departments of Radiology (B.L.S., F.B., M.H.K.H., A.J.A.) and Psychiatry (G.G.), University Hospitals, Ulm, Germany. Address correspondence to Bernd L. Schmitz, MD, Department of Radiology, University Hospitals Ulm, Steinhoevelstr. 9, D-89075 Ulm, Germany. American Society of Neuroradiology increasing B0 field strength, SAR increases with the square of the B0 field strength, or resonance frequency (5). The frequently experienced problems with SAR limitations in T1 spin-echo sequences, as well as the contrast reduction in T1 imaging, motivated the present study. From a theoretical perspective (see Material and Methods, below), we hypothesized that altering the flip angle as a single and easily accessible parameter has the potential to target both increasing contrastto-noise ratios (CNRs) and decreasing SARs (5 8). To test empirically the effects of varying flip angles on CNRs from T1-weighted spin-echo sequences upon 3T head imaging, an observer-independent method to measure gray-to-white matter contrasts reliably and objectively was introduced. Subjective ratings by 2 experienced neuroradiologists were used to test the practical gain of the above approach. Finally, we empirically investigated the relationship between SAR and different flip angles to determine whether decreasing flip angles might represent a possible and easily accessible parameter to compensate for SAR problems at higher field strengths. Materials and Methods Theoretical Considerations With T1 and T2 relaxation times for gray and white matter known (9), signal intensity (I) for gray and white matter can be 2000

AJNR: 26, September 2005 3T T1 SPIN-ECHO BRAIN SCANS 2001 predicted for given TRs and TEs and different flip angles according to the following formula (5): I N H sin exp TE/T2 1 exp TR/T1 1 cos exp TR/T1 2 exp TE/2 TR /T1 Here, N(H) is the hydrogen attenuation, denotes the flip angle, T1 and T2 are the relaxation times of the gray and white matter, and TR and TE are repetition time and echo time, respectively. Solving the equation above, signal intensities for gray and white matter at different flip angels can be calculated and gray-to-white matter contrast can be predicted in terms of differences between signal intensities of both types of tissue. Theoretically, from the relationships it can be shown that a lowering of flip angles with TR and TE kept constant leads to a reduction of signal intensities while there is an increase of gray-to-white matter contrast. This relationship, however, is of practical use only until a certain lower bound of flip angles is reached, beyond which gray and white matter no longer yield reasonable amounts of absolute signal intensity. Within this framework, we empirically tested the effects of different flip angles on gray-to-white matter contrast. It is noteworthy, however, that technical restrictions on the commercially available software did not permit us to test flip angles 50 empirically. With respect to the relationship between flip angle and the SAR it is of relevance that deposition of radio-frequency (RF) power is related to the square of excitation pulse at 90 ( /90 ) 2 (5). Again, by computing SAR for different flip angles, the exponential increase of depositing RF power with increasing flip angles can be demonstrated. Subjects Six healthy volunteers (1 woman and 5 men; mean age, 28.8 years; SD, 1.7) gave written informed consent before the study. The subjects heads were fixed by means of foam pads within the head coil, and the subjects were instructed not to move during MR acquisition. Imaging Protocol MR brain scanning was performed by use of 3T dedicated head scanner (Magnetom Allegra; Siemens, Erlangen, Germany) with the following parameters: TR, 700 ms; TE, 10 ms; matrix, 256 256; field of view, 220 220; section thickness, 5 mm; 19 transversal sections oriented parallel to the hypophysis-fastigium line. Within each subject, this sequence was repeated with flip angles of 50, 70, 90, 110, and 130 ; all other imaging parameters were kept constant. In addition, for purposes of coregistration, a necessary prerequisite to compute fully automated CNR measurements described below, each subject was scanned with a 3D-MPRAGE (magnetization-prepared rapid acquisition of gradient echo) sequence (10). Section orientation was equivalent to the T1 spin-echo sequence, while covering the whole brain. Calculation of Gray-to-White Matter Contrast Gray-to-white matter contrasts were calculated in an observer-independent, objective manner. All images were transferred to a workstation running SPM 2 (Statistical Parametric Mapping; Wellcome Department of Cognitive Neurology, London) under Matlab 6 (Mathworks, Natick, MA). The MPRAGE volume was segmented into gray and white matter probability maps (11, 12) by using SPM 2. To correct for head movements between the acquisition of the spin-echo T1 sequences with different flip angles and the MPRAGE sequence, the MPRAGE volume image was coregistered to each of the T1 spin-echo sequences by using a rigid body algorithm, and the coregistration was applied to the gray and white matter segments. To ascertain exclusive masking of gray and white matter in further computations, probability maps were thresholded at a probability of 95%, to create gray and white matter masks. These masks were applied to the spin-echo images, which resulted in matrixes containing the individual gray and white matter intensity values. Means and SDs of these matrices were calculated. To account for differences in general scanner signal intensity, values were scaled to the overall mean signal intensity of each individual sequence. CNRs were calculated as differences between intensity means of gray and white matter divided by the mean of their SDs (13). In addition, signal intensity tonoise ratios (SNR) were calculated in the same manner as intensity means and divided by the means of their SDs. Other Dependent Variables In addition to the method described above, image quality was rated independently, on a scale of 1 10, by 2 experienced neuroradiologists blinded to the imaging parameters. For each of the different flip angles, observer-related scoring differences were tested by means of 5 separate unpaired t tests. Because none of the tests revealed any significant difference between observers (all P values.15), scores from both observers were averaged for further analysis. Individual SARs, in units of watt per kilogram body weight, and estimated for the whole body according to the manufacturer s algorithm were tabulated for each measured sequence. Statistical Analysis For each of the various dependent variables (CNR, mean observer scores, SAR, and SNR) differences were tested on significance by means of separate multivariate analyses of variance (MANOVA) for repeated measures with the within-subjects factor flip-angle (5 levels). In cases of a significant main effect of this factor, post hoc comparisons (Fisher LSD, P.05) were calculated to further locate differences between factor levels. Results All acquired images were free of artifacts and were suitable for statistical analysis. Visual inspection of the calculated gray and white matter masks, as well as visual control of the coregistration, showed optimal fitting of the different volumes. Observer-Independent Parameters Comparing CNR for gray versus white matter revealed a highly significant effect of the different flip angles (F [4, 2] 207.86; P.0048). Post hoc comparisons showed significant differences between all CNR values for the different flip angles (Fig 1A). CNR was greatest for a flip angle of 50. Comparing SNR values, a MANOVA for repeated measures demonstrated a significant effect of the factor flip angle (F [4, 2] 104.75; P.009). Post hoc tests revealed significant differences between all SNR values for the different flip angles. A MANOVA for repeated measures on individual SAR values demonstrated highly significant differences (F [4, 2] 2290.04; P.00044). Here, post hoc

2002 SCHMITZ AJNR: 26, September 2005 FIG 1. A, Black bar charts represent measured mean gray-to-white matter CNR at increasing flip angles. (Values were scaled to their grand mean to scale with predicted signal intensity.) Error is in SD. Striped bar charts represent predicted gray-to-white matter contrast. B, Black bar charts represent measured SNR at increasing flip angles. (Values were scaled to their grand mean to scale with predicted signal intensity.) Error is in SD. Striped bar charts represent predicted mean signal intensity for gray and white matter. comparisons were significant when SAR values from the 50 flip angle were compared to SAR values from all the other flip angles. In addition, SAR values from a 70 flip angle were significantly lower than from a 90 flip angle, whereas the latter was still significantly lower than the SAR value for a flip angle of 130. Observer-Dependent Parameters Subjective image reading by the 2 neuroradiologists also showed a significant effect of the different flip angles (F [4, 2] 30.04; P.032). Again, post hoc comparisons showed significant differences between all observer scores for the different flip angles (Fig 1B). Although low flip angles resulted in a robust and stable even improved T1 contrast characteristic, flip angles 90 introduced additional artifacts that even inverted the contrast of the basal ganglia region (Fig 2). FIG 2. Bar charts represent mean subjective observer scores on image quality at increasing flip angles. Error is in SD. Observers were 2 experienced neuroradiologists. Discussion Brain imaging at 3T high field strength is prone to a substantial reduction of gray-to-white matter contrast in T1 spin-echo sequences (4). The present study showed a highly significant effect of different flip

AJNR: 26, September 2005 3T T1 SPIN-ECHO BRAIN SCANS 2003 FIG 3. Transversal sections of a prototypical subject scanned at different flip angles during a spin-echo T1 sequence at 3T B0 field strength to demonstrate the obvious decrease of gray-to-white matter contrast with increasing flip angle. Also noteworthy is the inverted T1 contrast in the basal ganglia region at a flip angle of 130. angles on gray-to-white matter contrast in 3T imaging, with the best CNR at the lowest flip angle of 50. The variance in contrast was more pronounced than theoretically predicted, which indicates that other factors such as, for example, magnetization transfer or shielding effects contribute to the contrast reduction with higher flip angles. SNR was lowest for a flip angle of 50, but this effect was less pronounced than theoretically predicted, probably because of more uniform signal intensity distribution at lower flip angles. Moreover, flip angle reduction led to a significant decrease of SAR. CNR measurements in MR data sets to compare gray and white matter most frequently involve an expert defining regions in which to find exclusively gray or white matter (9, 14); however, the small cortical band is especially prone to partial volume effects, which ultimately might bias the computation toward erroneously higher gray-to-white matter contrasts because every expert must define gray and white matter areas on the images to judge. Therefore, gray and white matter intensities are already implied in defining the regions, leading to a bias toward higher grayto-white matter contrast. In our present method, gray and white matter of the images could be defined independently by automatically segmenting an additional MPRAGE sequence with high resolution into gray and white matter probability maps (11, 12, 15). The MPRAGE sequence with high isotropic resolution was aligned to the lower resolved T1 images, to avoid introduction of partial volume effects from rotation or translation of these anisotropic images. Moreover, thresholds for defining gray and white matter were set at a high level, 95%, to further increase robustness of voxel selection. As a result, and in contrast to the standard region of interest-based analysis, the present method is able to include gray and white matter intensities from the entirely scanned brain and is no longer confined to subjectively selected brain areas. Certainly, image quality is not solely affected by CNR. To account for other factors that cannot be objectively measured, subjective observer ratings based on clinical expertise were included. These subjective ratings by 2 independent neuroradiologists blinded to the sequence parameters confirmed the results of the automated process, and the high overlap of the results of both methods was seen as indicative for the robustness and reliability of the automated, and observer-independent, method.

2004 SCHMITZ AJNR: 26, September 2005 Conclusion Within the context of clinical routine diagnostics, the data presented here strongly suggest low flip angles of 50 for T1 spin-echo brain imaging at a B0 field strength of 3T to increase gray-to-white matter contrast. As a side product, considerable reduction in SAR can also be gained by means of this approach. References 1. Schmitt F, Grosu D, Mohr C, et al. 3 tesla MRI: successful results with higher field strengths. Radiologe 2004;44:31 47 2. Baudendistel KT, Heverhagen JT, Knopp MV. Clinical MR at 3 tesla: current status. Radiologe 2004;44:11 18 3. Heidemann RM, Griswold MA, Muller M, et al. Feasibilities and limitations of high field parallel MRI. Radiologe 2004;44:49 55 4. Ross JS. The high-field-strength curmudgeon. AJNR Am J Neuroradiol 2004;25:168 169 5. Mills TC, Ortendahl DA, Hylton NM, et al. Partial flip angle MR imaging. Radiology 1987;162:531 539 6. Bonny JM, Foucat L, Laurent W, Renou JP. Optimization of signal intensity and T1-dependent contrast with nonstandard flip angles in spin-echo and inversion-recovery MR imaging. J Magn Reson 1998;130:51 57 7. Sugimura K, Kawamitsu H, Yoshikawa K, et al. Low flip angle spin-echo MR imaging to obtain better Gd-DTPA enhanced imaging with ECG gating. Nippon Igaku Hoshasen Gakkai Zasshi 1992;52:338 343 8. Winkler ML, Ortendahl DA, Mills TC, et al. Characteristics of partial flip angle and gradient reversal MR imaging. Radiology 1988;166:17 26 9. Wansapura JP, Holland SK, Dunn RS, Ball WS Jr. NMR relaxation times in the human brain at 3.0 tesla. J Magn Reson Imaging 1999;9:531 538 10. Mugler JP 3rd, Brookeman JR. Three-dimensional magnetizationprepared rapid gradient-echo imaging (3D MP RAGE). Magn Reson Med 1990;15:152 157 11. Ashburner J, Friston K. Multimodal image coregistration and partitioning: a unified framework. Neuroimage 1997;6:209 217 12. Ashburner J, Friston KJ. Voxel-based morphometry: the methods. Neuroimage 2000;11:805 821 13. Constable RT, Henkelman RM. Contrast, resolution, and detectability in MR imaging. J Comput Assist Tomogr 1991;15:297 303 14. Whittall KP, MacKay AL, Graeb DA, et al. In vivo measurement of T2 distributions and water contents in normal human brain. Magn Reson Med 1997;37:34 43 15. Woermann FG, Free SL, Koepp MJ, et al. Voxel-by-voxel comparison of automatically segmented cerebral gray matter: a raterindependent comparison of structural MRI in patients with epilepsy. Neuroimage 1999;10:373 384