Digital radiography (DR) post processing techniques for pediatric radiology

Similar documents
Exposure Indices and Target Values in Radiography: What Are They and How Can You Use Them?

COMPUTED RADIOGRAPHY CHAPTER 4 EFFECTIVE USE OF CR

Disclosures. Outline 7/31/2017. Current Implementation Status of IEC Standard : Exposure Index (EI) for Digital Radiography

Outline. Digital Radiography. Understanding Digital Modalities: Image Quality and Dose. Image Quality. Dose Control

Digital Imaging started in the 1972 with Digital subtraction angiography Clinical digital imaging was employed from the 1980 ~ 37 years ago Amount of

Image Display and Perception

2217 US Highway 70 East Garner, NC Main: Fax:

Practical Medical Physics Session: TG-151 Dose Monitoring. August 5, 2013 Katie Hulme, M.S.

3/31/2011. Objectives. Emory University. Historical Development. Historical Development. Historical Development

Digital Imaging Considerations Computed Radiography

Acquisition, Processing and Display

of sufficient quality and quantity

Visibility of Detail

NOTES Teaching Digital Processing and Display Quinn B Carroll, MEd, RT WCEC Conference, 2015

A Practical Overview of the Clinical and Operational Impact of Computed Radiography(CR) Implementations. Shirley Weddle, RT(R)(M), CIIP, BBA

Teaching Digital Radiography and Fluoroscopic Radiation Protection

Features and Weaknesses of Phantoms for CR/DR System Testing

RAD 150 RADIOLOGIC EXPOSURE TECHNIQUE II

Beam-Restricting Devices

New Exposure Indicators for Digital Radiography Simplified for Radiologists and Technologists

An Exposure Indicator for Digital Radiography. Report of AAPM Task Group 116

COMPUTED RADIOGRAPHY (CR)

The effect of compensating filter on image quality in lateral projection of thoraco lumbar radiography

THE ART OF THE IMAGE: IDENTIFICATION AND REMEDIATION OF IMAGE ARTIFACTS IN MAMMOGRAPHY

DELWORKS DR MEDICAL. take the next step

SYLLABUS. TITLE: Equipment Operation I. DEPARTMENT: Radiologic Technology

The Evaluation of Collimator Alignment of Diagnostic X-ray Tube Using Computed Radiography System

Automated dose control in multi-slice CT. Nicholas Keat Formerly ImPACT, St George's Hospital, London

Essentials of Digital Imaging

While digital techniques have the potential to reduce patient doses, they also have the potential to significantly increase them.

DISC QC/QA Program for Digital Imaging Systems using the DR Radchex Plus Meter

SECTION I - CHAPTER 1 DIGITAL RADIOGRAPHY: AN OVERVIEW OF THE TEXT. Exam Content Specifications 8/22/2012 RADT 3463 COMPUTERIZED IMAGING

Advanced digital image processing for clinical excellence in fluoroscopy

Acceptance Testing of a Digital Breast Tomosynthesis Unit

Studies on reduction of exposure dose using digital scattered X-ray removal processing

Studies on reduction of exposure dose using digital scattered X-ray removal processing

Outline ASRT Changes Impact on current curriculum Potential new courses WECM Changes Last update Resources and needs

STEREOTACTIC BREAST BIOPSY EQUIPMENT SURVEYS

- KiloVoltage. Technique 101: Getting Back to Basics

CR35-SP+ Computed Radiography (CR) System. Innovations in Digital Imaging

Ask EuroSafe Imaging Tips & Tricks. Paediatric Imaging Working Group. Dose Management in Digital Radiography

Get more from your images with Symphony Image Processing

Redefining Ergonomics

Essentials of Digital Imaging

Y11-DR Digital Radiography (DR) Image Quality

COST EFFECTIVE FLAT PANEL DIGITAL RADIOGRAPHY UPGRADE SOLUTIONS

MUSICA Nerve Center. Artificial Intelligence. Intelligent tools for your Digital Radiography workflow. Fluoroscopy. Workflow Optimization

radiography detector

Learning Objectives. Outline. Getting Started with CR. Converting the Radiology Department from Film-Screen to Digital: Making the Transition

Do you have any other questions? Please call us at (Toll Free) or , or

Multiple Choice Identify the letter of the choice that best completes the statement or answers the question.

Get more from your images with Symphony Image Processing

Exposure System Selection

Digital radiography: Practical advantages of Digital Radiography. Practical Advantages in image quality

I. PERFORMANCE OF X-RAY PRODUCTION COMPONENTS FLUOROSCOPIC ACCEPTANCE TESTING: TEST PROCEDURES & PERFORMANCE CRITERIA

CR Basics and FAQ. Overview. Historical Perspective

Quality assurance: a comparison study of radiographic exposure for neonatal chest radiographs at 4 academic hospitals

Essentials of Digital Imaging

DICOM3.0 Conformance Statement

SECTION I - CHAPTER 2 DIGITAL IMAGING PROCESSING CONCEPTS

Mammography is a radiographic procedure specially designed for detecting breast pathology Approximately 1 woman in 8 will develop breast cancer over

Patient-Assisted Compression Impact on Image Quality and Workflow

Mammography: Physics of Imaging

X-RAYS - NO UNAUTHORISED ENTRY

DigiMam Conformance Statement for DICOM V3.0

Digital Radiography X-Ray System. X Twin with X Mobil Roesys GmbH [rshsmi] 1/11

Digital radiography. bucky table and wall stand as a dual detector or wireless system. Amadeo DR Systems

TOPICS: CT Protocol Optimization over the Range of Patient Age & Size and for Different CT Scanner Types: Recommendations & Misconceptions

Ludlum Medical Physics

LECTURE 1 The Radiographic Image

Introduction. Sam R. Kottamasu Lawrence R. Kuhns

X-RAY IMAGING EE 472 F2017. Prof. Yasser Mostafa Kadah

Radiology Physics Lectures: Digital Radiography. Digital Radiography. D. J. Hall, Ph.D. x20893

Image Quality Artifacts in Digital Imaging

4. Contrast is the. There must The function of contrast is to:. The types of contrast are.

TESTING FLAT-PANEL IMAGING SYSTEMS: What the Medical Physicist Needs to Know. JAMES A. TOMLINSON, M.S., D.A.B.R. Diagnostic Radiological Physicist

Open. the Digitized world. Fuji Computed Radiography

2 COMPENSATING FILTERS PETER J. BARGER

Overview. Professor Roentgen was a Physicist!!! The Physics of Radiation Oncology X-ray Imaging

Technical Paper CONSISTENT PRESENTATION OF MEDICAL IMAGES

FMT18 FLOOR MOUNTED SYSTEM

White Paper. The Advantages of Hybrid Gamma PXL. 1 Introduction Outline of Hybrid Gamma PXL Study Results...

GE Healthcare. Senographe 2000D Full-field digital mammography system

Practical Aspects of Medical Physics Surveys of Mammography Equipment and Facilities

Introduction. Chapter 16 Diagnostic Radiology. Primary radiological image. Primary radiological image

REGIUS CONSOLE CS-3. DICOM 3.0 Conformance Statement CODE NO Manufacturer: 1 Sakura-machi, Hino-shi Tokyo , Japan

Test Equipment for Radiology and CT Quality Control Contents

Quality Control for Stereotactic Breast Biopsy. Robert J. Pizzutiello, Jr., F.A.C.M.P. Upstate Medical Physics, Inc

X-ray Imaging. PHYS Lecture. Carlos Vinhais. Departamento de Física Instituto Superior de Engenharia do Porto

CONTRASTING VIEWS DIGITAL VS CONVENTIONAL RADIOGRAPHY

Digital Image Management: the Basics

Surveying and QC of Stereotactic Breast Biopsy Units for ACR Accreditation

SUSPENSION CRITERIA FOR IMAGE MONITORS AND VIEWING BOXES.

DICOM Correction Item

Breast Tomosynthesis. Bob Liu, Ph.D. Department of Radiology Massachusetts General Hospital And Harvard Medical School

Small Animal Radiographic Techniques and Positioning COPYRIGHTED MATERIAL

Artefacts found in computed radiography

1. Patient size AEC. Large Patient High ma. Small Patient Low ma

RADIOGRAPHIC EXPOSURE

Maximizing clinical outcomes

Transcription:

Digital radiography (DR) post processing techniques for pediatric radiology St Jude Children s Research Hospital Samuel Brady, MS PhD DABR samuel.brady@stjude.org

Purpose Review common issues and solutions facing pediatric digital radiography (DR) Exposure techniques Post processing Review physics of DR post processing

Default technique set up A properly acquired DR image starts with proper technologist training/ongoing education Post processing should never replace proper technique Patient alignment (may be difficult in pediatrics) Center the patient (especially when using AEC) Under 4 years old AEC not effective Collimation Techs will widen collimation for squirmy/noncompliant children Poor collimation leads to More scatter and less subject contrast More direct detector exposure» May throw off post processing algorithms

Default technique set up Technologist should review technique before acquiring image kvp, mas No grid under 1 yr old (extremities, and small parts in general) Filtration Adjust for patient size Correct anatomical protocol is selected e.g., skeletal survey Techs will move the detector with out changing the protocol Technologists should review captured image For adequate quality EI/DI values Physicist/technologist QA Review image data for outlier examination Identify examination techniques/post processing problems

Default technique set up Review results in a measurable way

Default technique set up Age/size- adjusted protocols RIS system interaction Auto select technique factors based on body habitus (e.g., age) e.g., Philips uses patient age Based on bone density (S) 0-12 mon (M) 12-48 mon (L) 48-96 mon Adult equivalent > 96 mon Post processing applied to the image Manual vs. automatic

Default technique set up Review default age/size-adjusted technique factors May not be defined for peds if you are an adult-centric site kvp, mas, grid options, etc. Work with your vendor s apps specialists AAPM TG 252/ACR Methodology to set up ped technique charts Philips GE

Default technique set up Review default age/size post processing Left image (adult protocol-general NA chest) Right image [(S), (M), (L)-NAP chest] Adult D1 = 1.5 D2 = 2.0 Pediatric D1 = 1.8 D2 = 2.1 Density (D); D1 = VOI (lung field); D2 = soft tissue

Image post processing The image is created by taking the raw image through a series of steps before it is ready for presentation 1. Image correction Raw data 2. For processing Log compression 3. Histogram formation/ comparison 4. Segmentation Detector exposure determination (K IND ) 5. Image processing 6. For presentation AAPM Report 116, Figure 2

Caveat Calibrate DR viewing monitor if possible If not possible, verify processed images on a diagnostic quality monitor

Image post processing

Image post processing Pixel values within for-processing image are organized into a histogram Histogram x-axis represents exposure amount Histogram y-axis displays number of pixels for each exposure Low kvp = wide histogram High kvp = narrow histogram Underexposed histogram is skewed left Overexposed histogram is skewed right Shape of the histogram stays fairly constant for each exposed body part White values Black values

Image post processing Third, for-processing pixel values (Q K ) are modified to produce an image with values suitable for display (Q P ) Computer auto detects collimator boundaries Shutters remove white borders that cause veiling glare in eye Shutters should not be used to compensate for poor collimation! Exclude pixel values outside collimator boundaries Images Courtesy of Stefan Specht, Philips

Image post processing Fourth, relevant anatomical areas of the radiograph are segmented These points are called the values of interest (VOI) e.g., lung parenchyma in chest x-ray; mediastinum tissue in thoracic spine, etc. D B C D C B A A

Image post processing Auto ranging function determines relevant min and max pixel intensity values from the histogram Sets VOI look up table (VOI LUT) with respect to histogram VOI LUT maps digital densities to P-values for display Generally, histogram shapes are similar for similar anatomy Important: select correct body part view before imaging patient

Image post processing Different VOI LUT may be optional and will affect image appearance FB FC LIN SCL SWL Philips

Image post processing Incorrect segmenting of histogram changes the look of the image VOI LUT provides good contrast only for tissue of interest Chest VOI Soft Tissue VOI Images Courtesy of Stefan Specht, Philips

Image post processing Internal metal hardware, external fixators, restraining devices, Pb shilelds confuse segmentation function Set up separate processing algorithms for internal hardware Close collimation (external hardware) reprocess then open up collimation Original knee Contrast Adjust 132% 147% Brightness Adjust 140% 144% Tissue Contrast 0.14 0.05 Edge 7 7 Knee w/ metal Under Penetration Area 20% 20% Under Penetration Strength 20% 20% Over Penetration Area 50% 60% Original knee GE Over Penetration Strength 80% 30%

Image post processing Fifth, Exposure index or air kerma index (K IND ) is determined from segmented anatomical areas (VOI) K IND relates the median exposure to these areas NOT to whole image Does not indicate an actual patient dose IEC 62424-1/AAPM TG116 describe universal methodology Deviation Index (DI) Patient dose vs. image post processing Traditional screen-film served as type of exposure indicator Digital radiography is forgiving of poor exposure If exposure is < 50% ideal exposure, quantum mottle results If exposure is > 200% ideal exposure, contrast loss results dose creep occurs technologists increase exposure over time for better IQ

Image post processing: example Post processing order of operation Segmentation Make sure the correct anatomical protocol is selected When possible, make sure the VOI is properly segmented Apply the VOI LUT curve

Image post processing: example Start by setting brightness (window level) and contrast (window width) of image VOI LUT

Image post processing: example Brightness (window level)

Image post processing: example Contrast (window width)

Image post processing: example Balancing contrast goals: Enhances details in the white range If image is too digital increase contrast Increased contrast = film like

Image post processing: Recap Brightness (window level) For this adjustment, only look at the image in the area of VOI Brightness (window level) is generally set a little higher for pediatrics than adults Contrast (window width) Increase or decrease until it meets the radiologists expectations (i.e., visibility of soft tissue against bones) Ped patients have poor internal contrast (no fat)» small adjustments make a big difference Avoid clipping in all areas of the image Once you have a good base image then proceed with detail enhancement Zoom image De-magnified image does not show sub band 1 and presents as smoother than in reality

Image post processing: example Detail edge enhancement Affects bony definition & noise Pediatric radiologists generally accept more noise for more bony enhancement Lower bone density for pediatric < 9 yrs old No detail enhancement

Normalized for-processing pixel values (Q K ) are modified to produce an image with values suitable for display (Q P ) by first: 1. Segmenting image anatomy and establishing VOIs 2. Calculating the log transform of the raw image 3. De-noising image 4. Enhancing small detail structures 5. Converting pixel values to grayscale P-values 61% 21% 11% 3% 3% 1. 2. 3. 4. 5.

Answer 1. Segmenting image anatomy and establishing VOIs Not all pixel values in an image are associated with objects that are of interest to the viewer for the purposes of diagnosis. Those that are of interest are referred to as the values of interest (or VOI). The pixels that are associated with the VOI are typically identified based on their physical location and their relative signal strength characteristics in the image histogram. This identification process is referred to as segmentation. Reference: AAPM TG-116, P.6

A critical first step for digital radiographic post processing is: 1. Correcting the raw image 38% for dark noise, gain, and 35% bad pixels 2. Calibrating the monitor(s) 3. Establishing anatomical specific VOIs 4. Establishing a good quality low pass image for 13% detail enhancement modification 8% 5. Calibrating entrance 5% exposure to detector for accurate exposure index (EI) calculation 1. 2. 3. 4. 5.

Answer 2. Calibrating the monitor(s) Arguably this is my opinion, but if the DICOM calibration of the display monitor you are using to develop DR post processing techniques does not match the display from which your radiologist is reading, much of your work will be in vain (the both of you are seeing different things).

Selecting the proper anatomical acquisition protocol before acquiring the x-ray image is important because: 1. The proper kvp and mas are anatomical specific 2. Post processing segmentation is based on histogram analysis that is anatomical specific 3. To set the proper filtration 4. To calculate exposure index (EI) accurately 5. Because post processing algorithms are age specific 1. 2. 3. 4. 5. 50% 48% 0% 0% 2%

Answer 2. Post processing segmentation is based on histogram analysis that is anatomical specific X-ray images are binned into a histogram whose general shape is anatomical specific. The overall location and magnitude of the histogram peaks are used to segment the image and apply the VOI LUT. Reference: AAPM TG-116, P.12

Thank You!