USING LEICA AS LASER MICRODISSECTION (LMD6000) MICROSCOPE Written By Jungim Hur

Similar documents
Zeiss AxioImager.Z2 Brightfield Protocol

SHORT GUIDE TO LASER MICRODISSECTION USING THE PALM COMBI SYSTEM

User Manual. Digital Compound Binocular LED Microscope. MicroscopeNet.com

Zeiss Deconvolution Microscope: A Quick Guide

Contents STARTUP MICROSCOPE CONTROLS CAMERA CONTROLS SOFTWARE CONTROLS EXPOSURE AND CONTRAST MONOCHROME IMAGE HANDLING

SHORT GUIDE TO LASER MICRODISSECTION USING THE PALM COMBI SYSTEM

Zeiss Axio Imager.A1 manual

Guide to Confocal 5. Starting session

TABLE OF CONTENTS. Safety notes i. Care and Maintenance. ii. 1. Components Illustration Installation of Components.. 4

The invention of the microscope made it possible for scientists to view and study cells. Cells the basic units of all living organisms.

Nikon E800 Operating Instructions.

User Operation of JEOL 1200 EX II

Characterization Microscope Nikon LV150

STRUCTURE OF THE MICROSCOPE

Zeiss LSM 880 Protocol

Nikon Eclipse Ti2-E Widefield/Spinning Disk Confocal Microscope Standard Operation Protocol

UNIVERSITY OF WATERLOO Physics 360/460 Experiment #2 ATOMIC FORCE MICROSCOPY

JEOL 6500 User Manual

Artisan Technology Group is your source for quality new and certified-used/pre-owned equipment

User instructions Compound laboratory microscope

Components of the Microscope

Renishaw InVia Raman microscope

User Manual. Trinocular Metallurgical Microscope. MicroscopeNet.com

Things to check before start-up.

EXC500p-- PATHOLOGY MICROSCOPE. EXC500hd -- HD DIGITAL PATHOLOGY MICROSCOPE. EXC500r -- RESEARCH MICROSCOPE EXC500-LABORATORY SCOPE

Nikon Ti-E Microscope Manual. Rightmire Hall Ohio State University. Director: Tony Brown Rightmire

OMM300. Inverted Metallurgical Microscope

SHORT INSTRUCTIONS FOR OPERATING LSM1/2 (Zeiss LSM510) AT CIAN Version 1.4, September 2014

The ideal K-12 science microscope solution. User Guide. for use with the Nova5000

Olympus IX71 Microscope and DP71 Camera Instructions

CAPTURING IMAGES ON THE HIGH-MAGNIFICATION MICROSCOPE

1. Preliminary sample preparation

LEICA TCS SP5 AOBS TANDEM USER MANUAL

Simplified Instructions: Zeiss Brightfield Microscope S1000

Biology 29 Cell Structure and Function Spring, 2009 Springer LABORATORY 1: THE LIGHT MICROSCOPE

AxioVision 4.5 Brightfield Image Capture Procedure

Leica SP8 TCS Users Manual

Arcturus XT Laser Capture Microdissection System AutoScanXT Software Module. User Manual

JEOL 6700 User Manual 05/18/2009

Motorized Axio Observer Start-up instructions

Leica DMi8A Quick Guide

Zeiss LSM 780 Protocol

SWIFT SERIES M2252DGL MICROSCOPE

Manual for BMS E1 eplan series, compound microscope

BX-61: Brightfield Instruction /Continue to scroll for Fluorescent Instuctions

Leica DB LB Research microscope and Studo Lite Imaging software

DIC Imaging using Laser Scanning Microscopes (LSMs) on Axio Imager Stands

Operating Checklist for using the Laser Scanning Confocal Microscope. Leica TCS SP5.

LSM 780 Confocal Microscope Standard Operation Protocol

OPT3: Operating Procedure for Horiba Jobin Yvon LabRam Aramis Raman/PL System See LabSpec_6_2 General User Quick Start Guide on the computer desktop

Operation Guide for the Leica SP2 Confocal Microscope Bio-Imaging Facility Hunter College October 2009

Training Guide for Carl Zeiss AxioZoom V16 Stereo Microscope

OM FL400. Reflected Light Fluorescence Microscope. Instruction Manual. Please read instructions carefully before using microscope.

Nikon E800 Operating Instructions.

DIC Imaging using Laser Scanning Microscopes (LSM) on Inverted Stands

Nikon E800 Microscope. Operating Instructions

A BRIEF INTRODUCTION TO MICROSCOPY The two key properties of a microscope that allow you to see microbes are resolution and magnification.

User instructions Compound laboratory microscope

Zoom Stereo Microscope NYMCS-360 Instruction Manual

Using the Nikon TE2000 Inverted Microscope

Operating Checklist for using the Scanning Electron. Microscope, JEOL JSM 6400.

Basic Users Manual for Tecnai-F20 TEM

User Manual. Digital Compound Binocular LED Microscope. MicroscopeNet.com

Jeol JEM Responsible personell: Endy ( ) Online booking is compulsory!

User Manual. Trinocular Infinity Compound LED Microscope. MicroscopeNet.com

Therefore, all descriptions and illustrations in this instruction manual, including all specifications are subject to change without notice.

ISONIC PA AUT Spiral Scan Inspection of Tubular Parts Operating Manual and Inspection Procedure Rev 1.00 Sonotron NDT

WITec Alpha 300R Quick Operation Summary October 2018

Using the AmScope Microscope Cameras

START-UP PROCEDURE 1 THE MICROSCOPE STAND 3 OBJECTIVES 5 STARTING WITH LAS (SOFTWARE) AND SETTING UP THE MICROSCOPE STAND 7

Key Points Refer to How to Use the Compound Light Microscope :

Microbiology Laboratory 2

MEASUREMENT CAMERA USER GUIDE

Leica TCS SP8 Quick Start Guide

STEINDORFF NYMC Polarizing Microscope

Digital Microscope. User Manual

Horiba LabRAM ARAMIS Raman Spectrometer Revision /28/2016 Page 1 of 11. Horiba Jobin-Yvon LabRAM Aramis - Raman Spectrometer

Confocal Raman Microscopy (WITec Alpha 300R)

Basic Microscopy. OBJECTIVES After completing this exercise, you should be able to do the following:

User instructions Metallurgical microscope

Brief manual how to start and close the Leica sp2 Confocal. (TCS SP2 AOBS system mounted on a DM IRE2)

User instructions Transmitted light phase contrast microscope

ML7520 ML7530 DIOPTER ADJUSTMENT RING BINOCULAR BODY, INCLINED 30. (a) Field Iris Control Lever. (c) Filter Slots EYEPIECES, KHW10X

CALIBRATION OF MICROSCOPE EYEPIECE GRATICULE

LSM 800 Confocal Microscope Standard Operation Protocol

Swift M3600, M3700 Series Digital & Compound Microscope

3. are adherent cells (ie. cells in suspension are too far away from the coverslip)

Brightfield Microscopy and Image Acquisition on Spotcam1. by Ryan Taylor/Nancy Kleene Last modified 10/02/05 by Birgit Ehmer

Quick Guide for Zeiss 710 Laser Scanning Confocal MGH Cancer Center

Using the Hitachi 3400-N VP-SEM

Biological Microscope Manual

Operating Checklist for using the Scanning Electron Microscope, JEOL JSM 6400.

User instructions Transmitted light laboratory microscope

King Saud University Dept. of Bot. & Microbiology. General Microbiology 140 MIC

MIF ZEISS LSM510 CONFOCAL USER PROTOCOL

1 Placing particles on the slide

Leica SP8 TCS Users Manual

Operating the Hitachi 7100 Transmission Electron Microscope Electron Microscopy Core, University of Utah

SOP: Polyvar Met Light Microscope

Microscope Skills. Scientific Skills the Microscope!

Transcription:

USING LEICA AS LASER MICRODISSECTION (LMD6000) MICROSCOPE Written By Jungim Hur Digital Video Camera Eyepieces Laser module Laser safety UV shield Specimen holder Smart move control LEICA CTR6500 electronics box 1. Mains switch 2. Pilot indicator Front side of the laser box 1. Key-operated switch 2. Switch for external trigger 3. Status LED: Operating status of laser 4. Push-button: Laser on/off 5. Status LED: Laser active/inactive 6. Service plug 1

Leica LMD6000 Operation Protocol 1) Clean the LMD station including slide holder, PCR tube tray sub-stage holder, the 3-D Satellite controls and keyboard with Kimwipes damped with DEPC treated H 2 O (500 µl DEPC solution in 500 ml H 2 O). a 2 2) Turn on power to: a. CTR-MIC control box (microscope will go through self diagnostic tests, stage will move; requires about 30 seconds). b. CPU/monitor the usual way by the on button on the computer. c. Laser (with key). During warm-up, the laser power box will display one green light (left). When the second green light (right) comes on, the laser is ready. Touch the red button between the two green lights on the laser power supply to activate the laser. Now there will be one green light (left) and one red light (right). The laser is ready to fire. Requires about 5 minutes. If you want to take a break more than 5 minute, then press red button and the laser box will display two green lights. It will be in the stand-by mode and will be reactivated by pressing the red button between the two green lights. 3) Allow the microscope to fully initialize the stage before starting the LMD application software. 4) Check to be sure the Laser Safety Shield is properly in place. Laser will not fire without the laser safety shield in place. 5) Activate the LMD software from the Desktop by double clicking on the LMD icon. Click on the LOAD slide (left) button in the software menu. Slide holder will move forward; pull the slide holder forward and out of the microscope stage. Insert the special foil membrane slides with mounted tissue sections into the slide holder (be sure to have the proper orientation of the specimen facing down in the microscope). Return the slide holder back into the microscope stage by sliding the holder along the slide rails until it snaps into place. Click the Continue button on the dialog box to take the slide holder into the microscope. Create Specimen Overview images by clicking Options/Specimen Overview. Use lower magnification (4X-10X) and define scanning are. Then, save scanned overview image. 2

6) Click the Unload Collector (right) button in the menu bar to unload the collection device (moves forward), Load PCR tubes with caps into the collection device. Inserting the cap into the holder 1. Lid of the cap 2. Holder For RNA isolation from LCM samples, place Extraction buffer (20-30 µl) from PicoPure RNA Isolation kit into the caps (Avoid bubbles when you place buffer into the caps). Return the collector to the sub-stage holder and click OK to return the collector into the sub-stage of the microscope. Select a cap to collect samples by clicking the corresponding circular red marking at the bottom left of the work window. The selected cap turns green. While collecting the cell types that you are interested, the buffer added into the tube cap might be dried. You may add 10 µl of H 2 O into the tube cap after 1-1.5 hour of cutting if the light is getting dimmer because the buffer is dried and should not collect samples more than 2.5 hours into sample cap. 7) The 3-D Satellite controls all basic remote functions of the microscope. a. These include motorized X and Y stage movements and Z microscope focus functions. b. The buttons on the right side switch between coarse or fine focus and stage motion. c. The buttons on the left side control the up and down magnification by changing the objectives on the nosepiece, one position at a time. Smartmove control 1. Movement in x direction 2. Movement in y direction 3. Setting the focus 4. Individual knob height position 3

8) Focus the specimen with a low power objective and adjust for proper illumination levels and binocular phototube prism setting for an on screen image as necessary. Move the stage to bring the area or cells of interest into the center of the field of view. Switch to a higher power objective by touching your selection on the front touch screen of the microscope, for better detail and identification of the cells of interest for cutting, and adjust focus and light level if necessary. If obtained sample image is not focused or it is hard to define the cell type that you are interested, you may align condenser adjust. To align condenser, close the field diaphragm all the way (control the field diaphragm on the microscope control dialogue window) and rotate the knob on the condenser until the circle of light is crisp and in focus. Then open up closed field diaphragm and adjust the aperture diaphragm. Refer to Condenser Alignment Protocol provided by Leica microsystem. 9) Perform Laser calibration on set-up before cutting the region of interest. 10) Test different laser settings (power and speed) to obtain optimized condition for the cell types that you are interested. Then, record the setting and proceed. On the right side of the software interface, select the Draw and Cut mode. Use the optional touch screen pen to drawn the region of interest (ROI) directly on the monitor or move the mouse cursor to the image area, and click and hold the left mouse button to draw a line or shape on the monitor around the cell or cells of interest. If the drawing is not desirable, remove the line from the shape list. iclick on the Start Cut button in the software, or alternately right click and select Cut. 4

It is recommended to save the image with drawn the region of interest before and after cut. 11) If the laser does not cut properly through the specimen, optimize laser settings in Laser Control by adjusting Power, and/or Speed. Sometimes, laser loses power after few hours of use, then increase laser power and re-draw the line and cut it again. (Refer to suggested laser settings for a normal histological tissue sections by clicking Factory Settings in the Laser Control window. Also, you can refer to the settings previously used to cut the cells of interest.) 12) Cutting options include: free hand figures with close line option, circle and rectangle figures. 13) Cutting options also include: Draw and Cut or Move and Cut. If region of interest is not released after cut, re-draw the line and cut again or increase laser power and click Move and Cut button to release cut. You can inspect the samples in the cap of collection tube by pressing the "Collector" key in the toolbar. Pressing the "Specimen" key restores the previous cutting position. 14) The Close Line feature can simply be checked on or off, but area measurements are only functional with the Close Line activated. 15) Laser Alignment and laser controls settings are specific to each objective and can be saved under File and Save Configuration File As. By naming this file and saving it to the folder Profiles, one is able to recall all laser settings saved during a cutting session. 16) To recall a user specific configuration files, go to File and Restore Configuration File, find and activate your specific file. 17) When you finish collecting samples, take out the tube from the collection device by unloading collector. The samples on the tube cap should be collected into the tube by brief spin and stored at -80 C until use. Also, unload slide by clicking on the LOAD slide (left) button in the software menu. Pull the slide holder out of the microscope stage and remove slide and store in the slide box. Unload collector device Unload slide holder 5

18) Make sure to record the laser setting and the information of the cell types you have captured, the objective used to capture, the volume of extraction buffer and H 2 O, the collection tube name and the date of capture. 19) Close LMD Software 20) Shut down Windows (computer) 21) Turn power off to: a. CTR MIC controller box b. Laser (turn key to off position) c. Fluorescence Power Supply (if utilized during your session) d. Monitor Adjustment of CORR objective settings for the L20x, L40x and L63x objective 1) Depending on whether you use the Glass PEN foil slide or Metal PET Frame foil slides or Metal POL Frame foil slides, one must properly set the correction collar of all CORR objectives in order to achieve the best image possible. The wrong correction collar setting will result in a poor or hazy image and a sub-optimal laser cutting. The correction collar settings are dramatically different between the glass PEN foil slides and the metal frame foil slides. Using the settings of one for the other type of slide will not give optimal imaging results. 2) For Glass PEN foil slides, the correction collar setting should be approximately 0.9 while for any Metal frame foil slide, the correction collar setting should be approximately 0.1. However, proper setting should be obtained by observing the image and turning the correction collar until optimum contrast and clarity in the image is achieved. NOTE: Refer to Leica LMD6000 V6.3.1. Quick (Basic Operation) Start Up and comprehensive online Help files from the LMD software or consult the LMD printed instruction manual provided with the system. 6