Easy Kohler Illumination Method

Similar documents
Biology 29 Cell Structure and Function Spring, 2009 Springer LABORATORY 1: THE LIGHT MICROSCOPE

A BRIEF INTRODUCTION TO MICROSCOPY The two key properties of a microscope that allow you to see microbes are resolution and magnification.

Marine Invertebrate Zoology Microscope Introduction

Manual for BMS E1 eplan series, compound microscope

Instruction Manual T Binocular Acromat Research Scope T Trinocular Acromat Research Scope

OMM300. Inverted Metallurgical Microscope

Optics Day 3 Kohler Illumination (Philbert Tsai July 2004) Goal : To build an bright-field microscope with a Kohler illumination pathway

Microscope Tutorial. How to use a compound microscope

User Manual. Trinocular Infinity Compound LED Microscope. MicroscopeNet.com

TEKSCOPE MICROSCOPE. Models N2 Series USER S MANUAL

Laboratory Introduction

Introduction. Instructional Objectives. Materials. Procedure. I. Microscope Parts and Function. Honors Biology

User Manual. Digital Compound Binocular LED Microscope. MicroscopeNet.com

Simplified Instructions: Zeiss Brightfield Microscope S1000

Chapter 2 Alignment C. Robert Bagnell, Jr., Ph.D., 2012

Education in Microscopy and Digital Imaging

User Manual. Trinocular Metallurgical Microscope. MicroscopeNet.com

DIC Imaging using Laser Scanning Microscopes (LSMs) on Axio Imager Stands

Therefore, all descriptions and illustrations in this instruction manual, including all specifications are subject to change without notice.

Figure 3.4 Approximate size of various types of cells. ~10 um. Red Blood Cells = mm 1500 um. Width of penny Pearson Education, Inc.

Using a Compound Light Microscope

Match the microscope structures given in the left column with the statements in the right column that identify or describe them.

The microscope is useful in making observations and collecting data in scientific experiments. Microscopy involves three basic concepts:

DIC Imaging using Laser Scanning Microscopes (LSM) on Inverted Stands


Artisan Technology Group is your source for quality new and certified-used/pre-owned equipment

MICROSCOPE LAB. Resolving Power How well specimen detail is preserved during the magnifying process.

SWIFT SERIES M2252DGL MICROSCOPE

TABLE OF CONTENTS. Safety notes i. Care and Maintenance. ii. 1. Components Illustration Installation of Components.. 4

CALIBRATION OF MICROSCOPE EYEPIECE GRATICULE

User Manual. Digital Compound Binocular LED Microscope. MicroscopeNet.com

The invention of the microscope made it possible for scientists to view and study cells. Cells the basic units of all living organisms.

Name: Period: Week of: January 21st-25th Root Words In-Class Homework. Picture: -Microscope Notes -Lesson on Focusing the Microscope

Eyepieces KHW10X. Diopter Adjustment Ring. Binocular Body Inclined 30. Binocular Clamp Screw. Analyzer control Lever. Reflected Light Illuminator

Basics of Light Microscopy and Metallography

Ocular Lenses. Head. Arm. Objective Lenses. Slide Holder Stage. On / Off Switch. Condenser with Iris Diaphragm. Light Intensity Control

Components of the Microscope

Introduction. Laboratory Equipment & Supplies. Model 1333PHi Shown (Phase Contrast) (2) Eyepieces (Eyecups installed) Diopter Adjustment Mechanism

Microscopy. Danil Hammoudi.MD

What is it? Study the mystery photos and try to identify each one! Have access to a computer?

OM FL400. Reflected Light Fluorescence Microscope. Instruction Manual. Please read instructions carefully before using microscope.

Microscope Review. 1. A compound light microscope is represented in the diagram below.

User instructions Metallurgical inverted microscope

Basic Microscopy. OBJECTIVES After completing this exercise, you should be able to do the following:

History of microscopy

MICROSCOPY MICROSCOPE TERMINOLOGY

STRUCTURE OF THE MICROSCOPE

Nikon Ti-E Microscope Manual. Rightmire Hall Ohio State University. Director: Tony Brown Rightmire

CAPTURING IMAGES ON THE HIGH-MAGNIFICATION MICROSCOPE

User instructions Metallurgical microscope

User instructions Biological inverted microscope

STEINDORFF METALLURGICAL MICROSCOPE. NYMCS-620 Instruction Manual

BX-61: Brightfield Instruction /Continue to scroll for Fluorescent Instuctions

ML7520 ML7530 DIOPTER ADJUSTMENT RING BINOCULAR BODY, INCLINED 30. (a) Field Iris Control Lever. (c) Filter Slots EYEPIECES, KHW10X

Using a Compound Light Microscope Lab Pre-Lab Assignment

2/4/15. Brightfield Microscopy! It s all about Magnification..! or is it?!

How Microscopes Work By Cindy Grigg

Care and Use of the Compound Light Microscope

Systems Biology. Optical Train, Köhler Illumination

User instructions Transmitted light laboratory microscope

Microscopy. Matti Hotokka Department of Physical Chemistry Åbo Akademi University

User instructions Transmitted light phase contrast microscope

User instructions Transmitted light laboratory microscope (digital)

STEINDORFF NYMC Polarizing Microscope

used for low power magnification of a sample image is 3 dimensional

Name: Date Completed: Class: Lab Minutes: Teacher:

Swift M10 Series Microscope Use and Care Manual

Introduction to Microscopes

User instructions Compound laboratory microscope

Biological Microscope Manual

Lab: Using a Compound Light Microscope

User instructions Polarisation microscope

Microscope & Measuring

Instructions for Making On-Line Reservations for Microscopes in NB11-204

Basic Microscopy for Plant Biology

Visual Anatomy ansd Physiology Lab Manual Pig Version 2nd Edition Sarikas TEST BANK

Objectives Be able to identify dependent and independent variables in experimental designs.

User instructions Compound laboratory microscope

The light microscope

Light Microscopy. Upon completion of this lecture, the student should be able to:

Person s Optics Test KEY SSSS

Microscope. Dr. Leena Barhate Department of Microbiology M.J.College, Jalgaon

Biology Lab #1: Using Microscopes to Observe and Measure Cells

MANUAL EXC-350 MICROSCOPE SERIES

Lab: The Compound Microscope

Things to check before start-up.

Therefore, all descriptions and illustrations in this instruction manual, including all specifications are subject to change without notice.

Observing Living Things

STEINDORFF Digital LCD Biological Microscope

Ⅰ. Application. Ⅱ. Main Technical Specification. AS1 Biological microscope,which is widely used in medical and

The Compound Microscope. Brightfield: Köhler Illumination

Microscopy Primer. Fig A compound light microscope with important parts labeled.

VISUAL PHYSICS ONLINE DEPTH STUDY: ELECTRON MICROSCOPES

H S P. User Manual. Cat.-No

REVIEW FOR TEST ON MONDAY

Zoom Stereo Microscope NYMCS-360 Instruction Manual

General Physics Experiment 5 Optical Instruments: Simple Magnifier, Microscope, and Newtonian Telescope

MANUAL MICROSCOPE SERIES. 73 Mall Drive, Commack, NY (P) (F)

ECEN 4606, UNDERGRADUATE OPTICS LAB

Microscopes. A guide to use, general Maintenance, and repair tailored to the Olympus CX-21 microscope

Transcription:

Easy Kohler Illumination Method ACADEMIC SKILLS CENTRE (ASC) A. Silverberg Completion of a Kohler illumination method is required before a microscope can be used efficiently. The Kohler method is designed to achieve maximum resolution and light focussing when observing an image using a microscope. To understand how an image is generated from a microscope, it is essential to grasp how a light path is generated. A path of light is created by a small halogen lamp located at the base of a microscope. The light is directed upwards by mirrors to the condenser. The condenser filters light to remove long wavelengths of light, thereby, permitting only the shorter wavelengths to proceed throughout the system and hence, improve resolution. When the condenser is positioned correctly, the light focuses onto the specimen held by the specimen holder. The correct positioning of light is achieved using a Kohler

illumination method. Furthermore, the shorter wavelength of light is passed through the objective lens, which produces a magnified image of the specimen. The eyepiece lenses further magnify the image which is projected onto the eye s retina. Kohler Illumination Procedure: Detailed Steps 1) Place a specimen slide onto the microscope stage and secure between the specimen holders. Refer to Appendix A for a breakdown of operational parts. 2) Ensure the microscope is turned on and the light source is adjusted to half of the possible brightness, otherwise the user will experience visual fatigue and discomfort. 3) Using the eyepiece portion, set the interpupillary distance to ensure comfort. 4) Place the 10X objective lens into position and focus the specimen using the coarse adjustment knob. 5) Each eyepiece requires adjustment according to an individual s sight requirements. To achieve an eyepiece adjustment: a. Close the right eye. Use the fine adjustment knob to sharpen the image. b. Close the left eye. Focus is obtained using the Diopter ring (Figure 1). The Diopter ring is normally located on the right eyepiece and moves both in the counter clockwise and clockwise directions to achieve the necessary focus requirements. Slowly, turn the Diopter ring to focus the right eye with the left eye closed. Figure 1: Location and Movement of the Diopter Ring 6) Close the field diaphragm ring and the condenser aperture diaphragm ring. A small circle of light will be visible. The circle of light needs to be centred in the field of view. If not, adjust the condenser screws as necessary to centre the circle of light. Recommendation: adjust the condenser screws slowly! 7) If you do not see a circle of light, this indicates the microscope is extremely out of Kohler. To correct the misalignment, open the field diaphragm ring to the point where the circle of light is just visible. Adjust the condenser screws as necessary to centre the circle of light. Close the field diaphragm ring again and proceed to step 8. 8) Adjust the condenser focus knob until the image of the circle of light is sharp.

NOTE: Completion of steps 6, 7 and 8 are important for a) centring the light path and b) focussing light onto the specimen. 9) Take out the removable eyepiece (Figure 2). Look down the eyepiece cylinder. Observe a circle of light located in the centre of a dark field. Figure 2: Removable left eyepiece Open the condenser aperture diaphragm ring slowly until the circle of light fills ¾ of the field. Replace the removable eyepiece. Record the setting for the condenser aperture diaphragm for the 10X objective (Figure 3). Figure 3: Settings for the condenser aperture diaphragm 10) The Kohler illumination procedure is now complete. Note: a Kohler illumination is required on a daily basis and every time the objective lens is changed. Once the condenser aperture diaphragm setting is established for each objective lens for the day, it is easy to dial the setting to switch between the objectives rather than repeat the entire procedure when an higher or lower magnification is necessary.

References Leica Microsystems [Online]. Richmond Hill, Ontario. Cited September 15, 2010. Available online: http://www.leica-microsystems.com/ Misseri, A., 2008. Microscopy. Kingston, Ontario: St. Lawrence College. Appendix A: Operational Parts of a Microscope

The Academic Skills Centre www.trentu.ca/academicskills acdskills@trentu.ca 705-748-1720