X-Ray Medical Imaging and Pixel detectors

Similar documents
Radiology Physics Lectures: Digital Radiography. Digital Radiography. D. J. Hall, Ph.D. x20893

Introduction. Chapter 16 Diagnostic Radiology. Primary radiological image. Primary radiological image

Amorphous Selenium Direct Radiography for Industrial Imaging

Current technology in digital image production (CR/DR and other modalities) Jaroonroj Wongnil 25 Mar 2016

Setting up digital imaging department!

X-ray detectors in healthcare and their applications

10/26/2015. Study Harder

DIGITAL IMAGE PROCESSING IN X-RAY IMAGING

10/3/2012. Study Harder

BASICS OF FLUOROSCOPY

Unit thickness. Unit area. σ = NΔX = ΔI / I 0

X-ray Tube and Generator Basic principles and construction

Seminar 8. Radiology S8 1

DALLA LUCE VISIBILE AI RAGGI X: NUOVI RIVELATORI DI IMMAGINI PER RAGGI X A DISCRIMINAZIONE IN ENERGIA ED APPLICAZIONI

Basis of Computed Radiography & PACS

Chromatic X-Ray imaging with a fine pitch CdTe sensor coupled to a large area photon counting pixel ASIC

COMPUTED TOMOGRAPHY 1

LARGE FORMATTED AND HIGH RESOLUTION CMOS FLAT PANEL SENSORS FOR X-RAY

X-RAY FLUOROSCOPY IMAGING SYSTEMS. Dr Slavik Tabakov. Luminescence: Dept. Medical Eng. & Physics King s College London

Radiographic sensitivity improved by optimized high resolution X -ray detector design.

Photon Counting and Energy Discriminating X-Ray Detectors - Benefits and Applications

X-rays in medical diagnostics

Flat Panel Detectors in Industrial Radiography

Digital Detector Array Image Quality for Various GOS Scintillators

High Performance Amorphous Silicon Image Sensor for X-ray Diagnostic Medical Imaging Applications

Mammography: Physics of Imaging

Strategies to improve the signal and noise performance of active matrix, flat-panel imagers for diagnostic x-ray applications

Small Animal Radiographic Techniques and Positioning COPYRIGHTED MATERIAL

STATUS AND PROSPECTS OF DIGITAL DETECTOR TECHNOLOGY FOR CR AND DR Ulrich Neitzel Philips Medical Systems, Röntgenstrasse 24, D Hamburg, Germany

DELWORKS DR MEDICAL. take the next step

Medical Imaging: A Look inside. Medical Imaging. Medical Imaging. Visible Human Project

Some Essential Physics of Flat-Panel Detectors

Features and Weaknesses of Phantoms for CR/DR System Testing

Advances in X-Ray Scintillator Technology Roger D. Durst Bruker AXS Inc.

PERFORMANCE CHARACTERIZATION OF AMORPHOUS SILICON DIGITAL DETECTOR ARRAYS FOR GAMMA RADIOGRAPHY

10/15/2012 SECTION III - CHAPTER 6 DIGITAL FLUOROSCOPY RADT 3463 COMPUTERIZED IMAGING

X-ray light valve (XLV): a novel detectors technology for digital mammography*

Basis of Computed Radiography & PACS

Gas scintillation Glass GEM detector for high-resolution X-ray imaging and CT

Shad-o-Box HS Product Family

Do you have any other questions? Please call us at (Toll Free) or , or

SECTION I - CHAPTER 1 DIGITAL RADIOGRAPHY: AN OVERVIEW OF THE TEXT. Exam Content Specifications 8/22/2012 RADT 3463 COMPUTERIZED IMAGING

Dosepix Detector as kvp-meter in Radiology and Mammography: First steps

DEVELOPMENT OF A FLAT PANEL DETECTOR WITH AVALANCHE GAIN FOR INTERVENTIONAL RADIOLOGY MATTHEW M. WRONSKI

PD233: Design of Biomedical Devices and Systems

SmartRAD. Advanced Digital Radiography System

Digital radiography: Practical advantages of Digital Radiography. Practical Advantages in image quality

K-edge subtraction X-ray imaging with a pixellated spectroscopic detector

Digital Image Management: the Basics

SYLLABUS. TITLE: Equipment Operation I. DEPARTMENT: Radiologic Technology

Mammography is a radiographic procedure specially designed for detecting breast pathology Approximately 1 woman in 8 will develop breast cancer over

SYLLABUS. 1. Identification of Subject:

Fluoroscopy - Chapter 9

COPYRIGHT 2002 by Srinivasan Vedantham ALL RIGHTS RESERVED. Use or inclusion of any portion of this document in another work intended for

Detectors for microscopy - CCDs, APDs and PMTs. Antonia Göhler. Nov 2014

SECTION I - CHAPTER 2 DIGITAL IMAGING PROCESSING CONCEPTS

Determination of the detective quantum efficiency of a prototype, megavoltage indirect detection, active matrix flat-panel imager

A ghost story: Spatio-temporal response characteristics of an indirect-detection flat-panel imager

Digital Radiography : Flat Panel

Scintillation Counters

GE Healthcare. Senographe 2000D Full-field digital mammography system

Digital Industrial Radiography

Diffraction-enhanced X-ray Imaging (DEXI) Medical Solutions. More information using less radiation

X-ray phase-contrast imaging

2 nd generation TOMOSYNTHESIS

High Energy Digital Radiography & 3D-CT for Industrial Systems

radiography detector

Preliminary Modulation Transfer Function Study on Amorphous Silicon Flat Panel System for Industrial Digital Radiography

X-ray Imaging. PHYS Lecture. Carlos Vinhais. Departamento de Física Instituto Superior de Engenharia do Porto

Performance of Image Intensifiers in Radiographic Systems

CXDI-70C WIRELESS SPECIFICATIONS

Marten Bosma 1, Alex Fauler 2, Michael Fiederle 2 en Jan Visser Nikhef, Amsterdam, The Netherlands 2. FMF, Freiburg, Germany

Digital Images & Image Quality

Introduction to X-ray Detectors for Synchrotron Radiation Applications

Examination, TEN1, in courses SK2500/SK2501, Physics of Biomedical Microscopy,

ARCO Rk.5 Digital Mobile C-Arm

Veterinary Science Preparatory Training for the Veterinary Assistant. Floron C. Faries, Jr., DVM, MS

Estimation of signal transfer property for wireless digital detector in different measurement schemes

Predicted image quality of a CMOS APS X-ray detector across a range of mammographic beam qualities

Modelling Computed Radiography Detectors with a Cascaded Linear System Model

HIGH SPEED, HIGH RESOLUTION AND LOW COST DIGITAL RADIOGRAPHY

Acquisition, Processing and Display

X-ray Tube and Generator Basic principles and construction

A Comprehensive Review of Image Production

Pixel hybrid photon detectors

We are IntechOpen, the first native scientific publisher of Open Access books. International authors and editors. Our authors are among the TOP 1%

Breast Tomosynthesis. Bob Liu, Ph.D. Department of Radiology Massachusetts General Hospital And Harvard Medical School

I. PERFORMANCE OF X-RAY PRODUCTION COMPONENTS FLUOROSCOPIC ACCEPTANCE TESTING: TEST PROCEDURES & PERFORMANCE CRITERIA

Digital Radiographic Inspection replacing traditional RT and 3D RT Development

Moving from film to digital: A study of digital x-ray benefits, challenges and best practices

Simulation of High Resistivity (CMOS) Pixels

3/31/2011. Objectives. Emory University. Historical Development. Historical Development. Historical Development

IBEX TECHNOLOGY APPLIED TO DIGITAL RADIOGRAPHY

Production of HPDs for the LHCb RICH Detectors

3D light microscopy techniques

Lecture 12 OPTICAL DETECTORS

Phase Contrast Imaging with X-ray tube

ProX Intraoral X-ray. PLANMECA is proud to introduce a new intraoral X-ray unit to its comprehensive collection of imaging products- the ProX.

brief history of photography foveon X3 imager technology description

Experiences of users in Digital Radiography

Transcription:

X-Ray Medical Imaging and Pixel detectors PIXEL 2000 Genova, June 5-8 th 2000 J.P.Moy, TRI XELL, Moirans, France 1

OUTLINE - X-ray medical imaging. The requirements, some particular features - Present detectors. - The new X-ray Flat detectors scintillator and photoconductor approach - How can pixel detectors help medical imaging? The detecting material, the readout circuit CONCLUSIONS 2

X-ray Imaging in Medicine : Radiography, Fluoroscopy, Computed Tomography (1) The oldest medical imaging technique : projection radiography discovered by Röntgen in 1895 : About 200 000 systems in the world: best for bones, but also widely used for soft tissues, often with contrats agents, such as barium sulfate for gastro-intestinal imaging. Mammography is a particular case, as it concerns soft tissues and aims at the detection of very fine calcifications. 3

X-ray Imaging in Medicine : Radiography, Fluoroscopy, Computed Tomography (2) Fluoroscopy : Originally visual observation of the fluorescent screen. Now with electronic image converters : XRII Angiography is a particularly important application of fluoroscopy : imaging blood vessels after injection of an iodine compound in an artery to detect stenosis or other pathologies. Computed tomography is a 3D imaging technique based on the reconstruction of the object from many linear projections. At present, it does not rely on imaging detectors 4

X-ray Imaging in Medicine : competition with Ultra-Sound, Magnetic Resonance Imaging? Imaging techniques without ionizing radiation will certainly develop at the expense of X-rays : - US is easy to use and cheaper than other modalities. - MRI provides invaluable information on soft tissues, and is becoming fast enough to adress cardiac imaging, but will remain expensive. X-rays will definitely remain for many years the most practical and cost effective imaging technique for bones, joints, and mammography. 5

Physical limitations Poisson statistics imply a trade off between size-dose-contrast. For instance, a 100 µm detail with 10 % contrast will be detected with a 10:1 Signal to Noise ratio only if the photon flux exceeds 10 6 photons /mm² (with an ideal detector) 100 10 2 absorbed photons /mm² S/N = 3 10 1 contrast (%) 0,1 Absorbed photon flux = 10 6 photons/mm² 10 6 absorbed photons/mm² for S/N = 30 S/N = 10 S/N = 3 0,01 After M.Arques, JRI 97 1 10 100 1000 10000 Object size (µm) 6

X-ray image sampling The image from a digital detector is spatially sampled, and therefore must comply with the laws of sampling : 1 2.sampling pitch Neither signal nor noise spectra should exceed (Nyquist) Failure to comply with this law results in aliasing. A spatial response of the converter layer smaller than the pixel is deceptive : the noise spectrum extends well beyond the Nyquist limit, so that it piles up 1 in the [0- ] range. 2.sampling pitch When the spatial response stops at the Nyquist limit, signal and quantum noise are filtered by the same MTF, and the input S/N is preserved as long as the other noises remain small. 7

Simulated images : photoconductor and scintillator based detectors Photoconductor, PSF = Pixel 500 µm CsI, measured PSF 6.8 x 10 5 120 6.8 x 10 5 120 6.6 100 6.6 100 intensité en e- 6.4 6.2 6 5.8 5.6 80 60 40 20 inte ns ité e n e - 6.4 6.2 6 5.8 5.6 5.4 80 60 40 20 5.4 LUT LUT 20 40 60 80 100 120 20 40 60 80 100 120 8

DETECTIVE QUANTUM EFFICIENCY : A measure of how well X-rays are used MTF Readout noise DQE Quantum Noise X-ray absorption Dose X-ray Energy 9

Compared requirements for RADIOGRAPHY and FLUOROSCOPY General radiography Mammography Fluoroscopy Size > 40 x 40 cm >18 x 24 cm >30 x 30 cm Pixel size ~ 150 µm 60-100 µm 200-400 µm Typical nb of incid.x/pel ~1000 ~5000 ~10 Corresponding dose 2.5 µgy 100µGy 25 ngy Energy range 30-120 kev ~20 kev 30-120 kev Input equiv. noise < 5 X quanta < 5 X quanta < 1 X quantum Dynamic range 12 bit 12 bit 12 bit Readout time 1-5 s 1-5 s ~30 ms (30fps) 10

The present detectors in Radiography (1) Film At present, the most widely used detection scheme is the screen-film. A light sensitive silver halide film is sandwiched between two radioluminescent screens, usually made of Gd 2 O 2 S:Tb powder in a binding agent. The sensitivity vs resolution trade-off results from : - the thickness of the absorbing screen, - the light absorption or reflection of the backing layer, - the size of the grains in the screen. 11

The present detectors in Radiography (2) Screen - Optics - CCD Based on existing elements. Possible extension to dynamic imaging. The basic obstacle is to get more than 1 el. /X-ray in the CCD ("Quantum sink" situation ) - According to the laws of optics the collection of light decreases as 1/demagnification². Coupling a 20 cm screen to a 2 cm CCD results in a very poor light collection - Fiber optics are the best way to couple a screen to a CCD (but the most expensive...) Some optical gain is necessary : X-ray Image Intensifiers 12

The present detectors in Radiography (3) Storage Phosphors Electrons created by the absorption of X-rays are stored as a latent image in a screen. It is then read by laser scanning Provides a digital image with a very broad dynamic range. Handled like film : Thin, identical formats and read time, disposable if damaged. Image quality and resolution comparable to that of sreen-films. Single reading station for several units. Not suitable for fluoroscopy 13

The present detectors in fluoroscopy X-ray Image Intensifiers are widely used. They offer an unequaled range of performance : X-ray detection efficiency close to the theoretical limits, Excellent S/N, even for very low X-ray flux Large size, up to Ø 400 mm Dynamic imaging capability, Zooming Mature technology, affordable However, they are bulky, especially for large diameters, and suffer from strong geometrical and magnetic distortion. 14

Operation of an X-ray image intensifier G1 Metal vacuum bottle X-ray G2 Output window Lens G3 Anode Camera photocathode CsI input screen Gain : input screen = 200 el.. / X-photon P20 output screen Gain : output screen = 1000 vis. photons / el. Aluminum input window Total gain = 200.000 vis. photons / X-photon 15

X ray Image Intensifiers from TTE 16

X-ray Flat Detectors, the emerging technology Two approaches : - The scintillator/visible image sensor - The photoconductor/charge sensor Both have led to commercial systems. So far, only amorphous silicon can be obtained in the required sizes. Image sensors as well as charge detection arrays can be built with a technology derived from that of LCD active matrices An assembly of standard single crystal Si circuits is also possible, but such tiling results in challenging technical obstacles. 17

Readout Architecture bias bias Line drivers PC Line drivers PD Charge amplifiers Multiplex, coding Photoconductor scintillator / Photodiode 18

The Photoconductor based pixel -HV bias Se h e - TFT Data column a-si gate 19

Cross-section of a scintillator-photodiode-tft pixel Photodiode CsI:Tl Bias column TFT Data column a-si gate 20

Photodiode quantum efficiency and CsI:Tl fluorescence spectrum 100% 80% Photodiode quantum efficiency CsI:Tl emission (nb photons) 60% 40% 20% 0% 350 400 450 500 550 600 650 700 750 wavelength (nm) 21

Energy absorption of different materials (standard DN spectra, escape taken into account) X-ray absorption (% energy). 100 90 80 70 60 50 40 30 20 10 0 500 µm CsI, 75 % Pack.fr. 800 µm Se Lanex regular (67 mg/cm²) 1 2 3 4 5 6 7 8 9 10 DN # 22

Compared Performance of scintillator based detectors Colbeth et al. 1 Jung et al. 2 Weisfield et al. 3 Kameshima et al. 4 Chaussat et al. 5 Granfors 6 Structure Gd 2 O 2 S:Tb or CsI:Tl/ TFT CsI:Tl/TFT Gd 2 O 2 S:Tb/TFT Powd.phos./MIS CsI:Tl/DD CsI:Tl/TFT Overall active size (cm) 19.5 x 24.4 20 x 20 40.6 x29.3 43 x 43 43 x 43 41 x 41 Number of pixels 1536 x 1920 1024 x 1024 2304 x 3200 2752 x 2752 3120 x 3120 2048 x 2048 Pixel size 127 µm 200 µm 127 µm 160 µm 143 µm 200 µm X-ray abs. @RQA5 ~40%(Gd screen) ~80 % ~40% (Gd screen) N.A. ~80% ~75% Presamp.MTF @ 2 lp/mm 20% 20% 40% 40% 35% N.A. Read noise (equ. X phot.) / acq.time 4-5X / 35ms ~1X / 35ms 3-4 X / 5s N.A./ 1s 4-5 X / 1.5s N.A. /<5s ( ~1X / 35ms for 20 x 20cm.) Dynamic range N.A. N.A. 4000:1 6000:1 4000:1 N.A. N.A.= not available. 1 Varian 99, 2 Philips 98, 3 dpix 98, 4 Canon 98, 5 Trixell 98, 6 General Electric 2000 RQA5 is a standard for X-ray quality : 70 kv DC on the X-ray tube, 23 mm of Al filtration to simulate the patient. 23

Compared performance of photoconductor based detectors A.Tsukamoto et al. 1 G.Shaber et al. 2 J.Rowlands et al. 3 Structure 500 µm a-se/tft 500 µm a-se/tft 300 µm a-se/tft Overall active size (cm) 23 x 23 35.6 x 43 5 x 7.5 Number of pixels 1536 x 1536 2560 x 3072 360 x 480 Pixel size 150µm 139µm 160µm X-ray abs. @RQA5 70% 52% 37% Presampl.MTF @ 2 lp/mm 80% 85% 80% Read noise (equ. X phot.)/ acquisition time N.A./ 35 ms 12-15X/a few sec N.A. Dynamic range N.A. 4000:1 N.A. N.A.= not available. 1 Toshiba 99, 2 Sterling 98, 3 University of Toronto 98. RQA5 is a standard for X-ray quality : 70 kv DC on the X-ray tube, 23 mm of Al filtration to simulate the patient. 24

Commercial devices Clinical tests have been performed for many years, and several manufacturers are now starting the production : Scintillator screen / a-si array : TRIXELL, GEMS, CANON Selenium : KodaK-HOLOGIC (formerly STERLING) According to the manufacturers, various applications are (or will soon be) covered : General and chest radiography, mammography, cardiac angiography 25

Pixium 4600 and radiographic table 26

Thorax image with a pixium TRI XELL 27

The benefits of the new X-ray Flat detectors Improved conditions : Immediate readout. The patient no longer waits in painful positions for the development of the film, and a new shot. The clinician has easier access to the patient during intervention. Reduced running cost (increased throughput of radiology rooms, no film, no chemicals, no waste processing, cheaper storage of data). Less dose: depending on the device, the required dose is 100 to 40% of the film dose (for a given S/N in the image). All the advantages of a digital image : processing, transfer, archiving, access to Computer Aided Diagnosis,... 28

The weak points of the new X-ray Flat detectors High investment costs (detector + image display & process.), because the detector relies on specific techniques (a-si photodiodes, converter material,...) and requires the assembly of many expensive components. Difficult image corrections : offset and gain correction accuracy limited by small non-linearities and drifts. At very low dose (fluoroscopy), obtaining a S/N comparable to that of XRII requires extreme care (costly!) 29

What next? Development time extremely long : The work on the present generation started in the mid eighties => It is time to prepare the next generation! Which improvements are worth a new development? A spectacular improvement in resolution or dose is unlikely. Reduce manufacturing costs without compromising on performance : make it simpler! Increase the S/N in fluoroscopy Open new modalities dual energy, tomosynthesis,... 30

The detecting material Obviously the cornerstone of future devices. Should combine : Strong X-ray absorption from 20 to 150 kev, large area deposition technique, Chemical, thermal compatibility with Si, high resistivity, high µτe, preferably with a low E, low e-h creation energy (50 ev in Se, 5 ev desirable), environmentally acceptable (HgI 2...?) At present, there is no consensus on a potential workhorse. 31

Can pixel detectors meet the requirements of medical imaging (1)? Single crystal silicon technology will soon reach the point where elaborate functions can be implemented in a 100-200 µm pixel, with a realistic yield over a large area. A suitable converter material is still to be found : CdTe, PbI 2, HgI 2, PbO,... Discrimination and counting in a pixel would open new possibilities : suppression of offset correction, dual energy, better linearity, no longer escape noise... However, it should be borne in mind that the counting rate will be huge : in the worst (but common) case where the patient does not cover the whole detector, ~10 7 photons/s hit each pixel. 32

Can pixel detectors meet the requirements of medical imaging (2)? Tiling will still be required for general radiography. Integration of driving and readout circuits will help to reduce manufacturing costs, but redundancy will be mandatory in order to reach reasonable yields. The better linearity should alleviate the task of matching the different tiles to avoid the checkerboard effects. Realistic assembly techniques are still to be found COST will most likely be the driving force, more than performance, as the pressure on health budgets will undoubtedly increase. 33

Conclusions X-ray imaging may benefit from the development of pixel detectors : - Simpler devices thanks to higher integration. - Rely on the standard Si technology. - Better S/N at very low doses. Besides the work on Si circuits, the need for a good X-ray converter is a prerequisite. 34