Quick Guide for Zeiss 710 Laser Scanning Confocal MGH Cancer Center

Similar documents
Quick Guide for Zeiss 710 Laser Scanning Confocal MGH Cancer Center

LSM 710 Confocal Microscope Standard Operation Protocol

LSM 780 Confocal Microscope Standard Operation Protocol

LSM 800 Confocal Microscope Standard Operation Protocol

Zeiss LSM 880 Protocol

Zeiss 880 Training Notes Zen 2.3

Zeiss 780 Training Notes

MIF ZEISS VIOLET CONFOCAL ZEN 2009 PROTOCOL

Operating Instructions for Zeiss LSM 510

ZEISS LSM 710 CONFOCAL MICROSCOPE USER MANUAL

Quick Guide. LSM 5 MP, LSM 510 and LSM 510 META. Laser Scanning Microscopes. We make it visible. M i c r o s c o p y f r o m C a r l Z e i s s

Microscopy from Carl Zeiss

Things to check before start-up.

Training Guide for Carl Zeiss AxioZoom V16 Stereo Microscope

LSM 510 Meta Training Notes

Zeiss LSM 510 Confocor III Training Notes. Center for Cell Analysis & Modeling

ZEISS LSM 710 NLO Multiphoton microscope Manual/Quick guide

SHORT INSTRUCTIONS FOR OPERATING LSM1/2 (Zeiss LSM510) AT CIAN Version 1.4, September 2014

TRAINING MANUAL. Olympus FV1000

LSM 510 Training Notes

LEICA TCS SP5 AOBS TANDEM USER MANUAL

Training Guide for Carl Zeiss LSM 510 META Confocal Microscope

Zeiss LSM 780 Protocol

Leica SP8 TCS Users Manual

Zeiss Axiovert 135 Fluorescence Microscope Quick Guide / Operations Manual (v. 1.0 February 09)

Title: Leica SP5 Confocal User Manual

Guide to Confocal 5. Starting session

Training Guide for Carl Zeiss LSM 880 with AiryScan FAST

MIF ZEISS LSM510 CONFOCAL USER PROTOCOL

REMEMBER: You have 5GB of disk space on this microscope. Check before you start if you have room for your experiment. If not delete your old data.

Quick Start Guide. Leica SP5 X

Zeiss LSM880 Operating Instructions. UTMB Optical Microscopy Core Jan. 16, 2018

Training Guide for Carl Zeiss LSM 5 LIVE Confocal Microscope

Leica SP8 TCS Users Manual

OPERATING INSTRUCTIONS

Contents. Introduction

Cell Biology and Bioimaging Core

ZEISS LSM510META confocal manual

Operating Checklist for using the Laser Scanning Confocal Microscope. Leica TCS SP5.

CONFOCAL MICROSCOPE (Zeiss LSM 510 META v4.2)

Olympus Fluoview 1000S Spectral Confocal Microscope Introduction to the NRI-MCDB Microscopy Facility Spectral Confocal Microscope

Training Guide for Carl Zeiss LSM 7 MP Multiphoton Microscope

Leica Sp5 II Confocal User Guide

Nikon SIM-E & A1-R System

Nikon Eclipse Ti A1-A Confocal Operating Manual. Start-up. Microscope

Nikon Eclipse Ti2-E Widefield/Spinning Disk Confocal Microscope Standard Operation Protocol

QUICKSTART GUIDE: WIDEFIELD HWF1 Zeiss Cell Observer Live Cell Imaging System (HAMMERSMITH, L BLOCK, ROOM 314) Imperial College London

Topics. - How to calibrate the LSM scanner. - How to clean the microscope. - How to adjust the pinhole alignment. - How to adjust the Collimator

Nikon AZ100. Laser Scanning Macro Confocal Microscope. Jordan Briscoe Adam Fries Kyle Marchuk Kaitlin Corbin. May 2017.

QUICKSTART GUIDE: WIDEFIELD WF3 Zeiss Cell Observer Live Cell Imaging System (SAF, ROOM 409) Imperial College London

Zeiss AxioImager.Z2 Brightfield Protocol

Leica TCS SP8 Quick Start Guide

Title: Nikon A1R Confocal User Manual

Zeiss AxioObserver with ApoTome

Operation Guide for the Leica SP2 Confocal Microscope Bio-Imaging Facility Hunter College October 2009

Zeiss LSM 510 Multiphoton Confocal Microscope

Bi/BE 227 Winter Assignment #3. Adding the third dimension: 3D Confocal Imaging

Zeiss Axio Imager.A1 manual

SPINNING DISK CSU-X1 USER MANUAL

Widefield-NikonEclipseTE200-ORCA Nikon Eclipse TE200 Inverted Microscope with Hamamatsu 1394 Orca-ER Cooled CCD Camera and Micromanager Software

Zeiss Deconvolution Microscope: A Quick Guide

Zeiss LSM 510 META Guide

Olympus Digital Microscope Camera (DP70) checklist

Practical work no. 3: Confocal Live Cell Microscopy

Leica SPEII confocal microscope. Short Manual

Zeiss Axioskop II. The AIF's "routine" light microscope. (Installed 8/24/04)AxioCam installed July 11th 2005

Zeiss LSM 510 Multiphoton Confocal Microscope

START-UP PROCEDURE 1 THE MICROSCOPE STAND 3 OBJECTIVES 5 STARTING WITH LAS (SOFTWARE) AND SETTING UP THE MICROSCOPE STAND 7

Training Guide for Leica SP8 Confocal/Multiphoton Microscope

Internal Medicine Imaging Core Emory University Department of Medicine

Everest System / Slidebook Operating Procedures

Leica TCS SP8 Quick Start Guide

BX-61: Brightfield Instruction /Continue to scroll for Fluorescent Instuctions

Nikon A1R. Multi-Photon & Laser Scanning Confocal Microscope. Kyle Marchuk Adam Fries Jordan Briscoe Kaitlin Corbin. April 2017.

Microscope Confocal LSM510 META

Focus Strategies with ZEN2. Sven Terclavers ZEISS Embedded 3D Imaging Specialist Harvard University,

DIC Imaging using Laser Scanning Microscopes (LSM) on Inverted Stands

Diskovery Spinning Disk Guide

Confocal Application Notes Vol. 5 July 2010

Nikon A1Rsi Confocal Start-Up Sequence

The Zeiss AiryScan System, Confocal Four.

Widefield 1. Switching on

Using the Nikon TE2000 Inverted Microscope

Leica SP8 Resonant Confocal. Quick-Start Guide

b. Turn the power switch and key to on position for blue laser.

Nikon C1si Spectral Laser Scanning Confocal Microscope. User Guide

DIC Imaging using Laser Scanning Microscopes (LSMs) on Axio Imager Stands

Confocal imaging on the Leica TCS SP8. 1) Turn the system on. 2) Use TCS user account. 3) Start LAS X software:

Renishaw InVia Raman microscope

Before you start, make sure that you have a properly calibrated system to obtain high-quality images.

Leica TCS SL Confocal Training. Neuroscience Imaging Core Staff. Core Director. Facility Manager

Leica DMi8A Quick Guide

Brief manual how to start and close the Leica sp2 Confocal. (TCS SP2 AOBS system mounted on a DM IRE2)

Simplified Instructions: Olympus Widefield Microscope S1230

Boulevard du Temple Daguerrotype (Paris,1838) a busy street? Nyquist sampling for movement

Zeiss Axioplan 2 imaging microscope and Axiovision software

ZEN 2012 SP5 black edition Hotfix 12

WITec Alpha 300R Quick Operation Summary October 2018

MAKE SURE YOUR SLIDES ARE CLEAN (TOP & BOTTOM) BEFORE LOADING DO NOT LOAD SLIDES DURING SOFTWARE INITIALIZATION

Nikon E800 Operating Instructions.

Transcription:

Quick Guide for Zeiss 710 Laser Scanning Confocal MGH Cancer Center For any questions or concerns, please contact: Linda Nieman lnieman@mgh.harvard.edu Office: (617) 643-9684 Cell: (512) 565-8076 Chenyue Lu clu0@mgh.harvard.edu Cell: (413) 561-3316 Power Up Turn on equipment in the following order: 1. Main System Power Make sure Key on right hand side in in the horizontal position. 1A. Flip MAIN SWTICH to ON position 1B. Flip SYSTEM/PC switch to ON position i. Wait for the Definite Focus module to display OFF 1C. Flip COMPONENTS switch to ON position i. Wait for components to initialize 2. Argon Laser Black Box Power Supply Turn the key to the right (clockwise) 3. Argon Laser Remote Controller Flip switch to RUN position 4. X-Cite widefield fluorescence light source Flip switch to ON position Write down the bulb hours on the sign in sheet 5. Turn on PC Login as LSM USER Wait for ~3 min for RTC to establish connection with computer 6. Zen software Double click ZEN black icon in center of desktop (Not Zen blue) When startup screen appears, click on the BOOT STATUS drop down menu Click START SYSTEM i. If the screen stalls at 2% or an error about the Real Time Controllers pops up. Press the RTC reset button on the box under the confocal for 5 seconds. Wait 30 seconds. If it fails to continue, close the software, press the button again, and then restart the ZEN software. If that fails, close Zen software, then power off the RTC. Wait 30 seconds. Power the RTC back on. Wait 1-2 minutes. Re-launch Zen. When startup screen appears, click on BOOT STATUS then START SYSTEM 1

Loading your sample 1. Push microscope head back. Do NOT push on gray HAL100 light box. 2. On Touchpad, choose objective. View using 10X first, then progress to higher magnification 3. Tap LOAD POSITION. (Drops down stage) 4. Load slide onto stage coverslip down (inverted microscope) 5. Bring microscope head back into position 6. On Touchpad, tap SET WORK POSITION Finding your sample 1. In Zen, go to the LOCATE tab 2. Click Transmitted Light ON if using BF, Reflected Light ON if using fluorescence 3. Choose configuration: a. BRIGHTFIELD b. DIC II For use with dry objectives i. 10x, 20x, 63x dry ii. Requires polarizers c. DIC III for use with wet objectives i. 40x oil, 40x water, 63x oil ii. Requires polarizers d. DAPI blue filter e. GFP green filter f. DsRED red filter 4. Focus on sample Too bright? Lower bright field transmission in ocular tab or using large black dial on XCite light box. 5. Adjust Kohler illumination, if needed: a. Close top aperture b. Focus with condenser head until a clear hexagon is formed c. Re-open top aperture 2

Setting confocal configuration 1. Click ACQUISITION tab to enter confocal mode NOTE: There are three different ways to setup a scan: a. Load saved configuration (These can be changed, so check before use.) b. Load template image (File / Open), then click REUSE c. Smart Setup (see instructions below) a. Load Configuration b. Re-use template image settings c. Smart Setup Smart setup 1. In the EXPERIMENT MANGER panel, choose SMART SETUP a. Configure your experiment. Choose dye / fluorophore and image display color b. Add additional dyes / fluorophores using + button. You can also delete dyes using the buttons 2. Choose imaging option a. FASTEST: i. Advantage: faster image acquisition ii. Disadvantage: potential cross-talk (spectral overlap) between channels. b. BEST SIGNAL: i. Advantage: Only one detector and one laser are switched on at any one time. This reduces crosstalk ii. Disadvantage: slower image acquisition (each color track imaged separately) c. BEST COMPROMISE - between both speed and best signal 3. Click APPLY 3

Image capture settings 1. In the ACQUISITION MODE panel, set the following parameters: a. SCAN MODE = FRAME b. FRAME SIZE (i.e. pixel density) = 1024x1024 c. SCAN SPEED should be set between 7-9 i. NOTE: faster scanning gives smaller time interval between images, but lower signal-tonoise ratio. Slower scanning gives longer time intervals, but higher signal-to-noise ratio d. AVERAGING = 1 i. NOTE: if you have a dim sample with white noise, you may want to consider averaging at least 4 frames. Note that adding frames adds to the total scan time. e. BIT DEPTH = 12 for publication images (or 16 bit for high dynamic range images) i. NOTE: increasing bit depth increases number of gray scale levels (2 n ) 2. Set PINHOLE to desired spatial resolution NOTE: For best resolution, click 1 AU NOTE: There is an inherent tradeoff between spatial resolution and image brightness. If your sample is dim and high resolution is not critical, you may want to set pinhole larger 4

Setting the detector gain 1. Click SET EXPOSURE a. (System will find optimal gain setting for each color channel) 2. Click LIVE a. (next to SET EXPOSURE) b. NOTE: live continuously scans sample at low resolution to allow quick adjustments. The SNAP image will be captured with Image capture settings above 3. Adjust focus, if needed 4. Click STOP a. (next to SET EXPOSURE) 5. Click TRACK1 a. Desired track will be highlighted in light gray. Other color channels can be unchecked 6. Click LIVE a. (next to SET EXPOSURE) 7. Enable SINGLE CHANNEL a. (Active channel will be highlighted blue) 8. Enable RANGE INDICATOR a. Red shows pixels that are saturated b. Black are pixels that are underexposed c. Blue are pixels that have optimal exposure d. Aim for an image that is has few red pixels 5

9. Adjust GAIN (MASTER) a. A good initial setting is ~650 b. The maximum gain is 1200 *** Aim for an image that is has few red pixels *** 10. Click STOP in EXPERIMENT MANAGER a. (next to SET EXPOSURE) 11. To view another track (i.e. color channel), click on TRACK2 in the CHANNELS panel. 12. Adjust GAIN (MASTER) for TRACK2 13. Repeat until all color channels give good images 14. Click STOP in EXPERIMENT MANAGER a. (next to SET EXPOSURE) 6

Acquiring a z-stack 1. Check the Z-STACK box 2. Click LIVE 3. Focus (see the blue plane on the screen) manually on the plane closest to the objective, then click SET FIRST 4. Focus (see the blue plane on the screen) manually on the plane furthest to the objective, then click SET LAST 5. Click STOP to stop scanning 6. Select either SMALLEST slice interval or select the number of SLICES desired 7. Click START EXPERIMENT to begin acquiring z- stack 7

Tile scan 1. Check the TILE SCAN box 2. Find the TILE SCAN panel 3. Modify number of tiles 4. Choose overlap percentage (10%) 5. Enable ONLINE STITCHING Crop 1. Click on CROP 2. Adjust the zoom area on your image 3. Go LIVE and then STOP 4. In ACQUISITION MODE, and click on OPTIMAL Frame size to get best resolution 8

Graphics / Overlays 1. Add in: a. Text b. Scale bar c. Lines d. Arrows e. Boxes f. Circles g. Free draw Saving your images 1. Click disk icon to save 2. Save to Data / LAB PI NAME / YOUR NAME if saving locally 3. Please do not store data permanently on computer. It is not backed up and data folders are periodically deleted 4. Images can be opened using FIJI (ImageJ) 5. Images can be exported as TIFF from the FILE / EXPORT menu Finishing up Turn off equipment in following order: 6. Close ZEN software 5. Shut down PC 4. Turn off X-Cite light source 3. Argon Laser Remote controller: Flip switch to OFF Position 2. Argon Laser Black Box Power Supply: Turn key to left (counter-clockwise) 1. Main system power: 1C. Flip COMPONENTS switch to OFF position 1B. Flip SYSTEM / PC switch to OFF position *** WAIT 5 MINUTES TO LET LASER COOL *** (you will hear fan shut off when laser is cool enough) 1A. Flip MAIN SWITCH to OFF position 1. Clean work area 2. Clean any immersion objectives that you used with lens paper DO NOT USE KIMWIPES! 3. Sign out 4. Put dust cover over microscope 9

ZEN lite for Windows Download ZEN lite, your free copy of the powerful ZEN imaging software. Use this free microscope software to try out key features of the ZEN core package or simply install it as a viewer for your CZI files. Use ZEN lite to: Control Axiocam microscope cameras Create microscope images and simple video sequences Measure distances interactively Manage images/videos and export them into various formats View meta information relating to CZI image files Use simple report functions Expand ZEN lite with selected modules: create impressive multichannel and time lapse images of your specimens or use extended measurement functions. To download go to: http://www.zeiss.com/microscopy/en_us/products/microscope-software/zen-lite.html FIJI (is just imagej) ImageJ is an open source Java image processing program inspired by NIH Image. It runs on any computer with a Java 1.8 or later virtual machine. Downloadable distributions are available for Windows, Mac OS X and Linux. ImageJ has a strong, established user base, with thousands of plugins and macros for performing a wide variety of tasks. To download go to: https://imagej.net/fiji/downloads 10