DEVELOPMENT OF AN AUTOMATED DOBSON CONTROL SYSTEM FOR UNATTENDED OPERATION ABSTRACT

Similar documents
Preliminary investigation: The effect of field of view restrictions on retrieved

Brewer Ozone Spectrophotometer Mechanical Design Principles #2. C.T. McElroy York University Canada

COMPARISON OF STRATOSPHERIC OZONE COLUMN FROM OZONESONDES, RADIOMETER, MIPAS AND GOMOS INSTRUMENTS

GEO-SolarSIM-D2 and SunTracker-2000/3000

Spectrophotometer. An instrument used to make absorbance, transmittance or emission measurements is known as a spectrophotometer :

MUSKY: Multispectral UV Sky camera. Valentina Caricato, Andrea Egidi, Marco Pisani and Massimo Zucco, INRIM

Instrumental and Methodological Developments in UV Research

Oriel Cornerstone 130 1/8 m Monochromator

CHAPTER 7. Components of Optical Instruments

metcon meteorologieconsultgmbh, Instruments for Atmospheric Research W1aa_Feb_2017_1.doc 1 -

APPLICATION NOTE. Computer Controlled Variable Attenuator for Tunable Lasers. Technology and Applications Center Newport Corporation

IRRADIATION MEASUREMENTS ON GROUND

Broadband detectors Mario Blumthaler

Program for UV Intercomparison 2014 in Davos:

Space Weather and the Ionosphere

"Internet Telescope" Performance Requirements

MULTI-FILTER ROTATING SHADOW BAND RADIOMETER MODEL MFR-7 BULLETIN MFRSR-7

THE spectral response (SR) measurement of a solar cell is

USER S GUIDE MIDDLETON SOLAR. AST-02 and AST-03 ACTIVE SOLAR TRACKING SYSTEM. Version: 2.4

Chemistry Instrumental Analysis Lecture 10. Chem 4631

Calibration of ARM Spectral Shortwave Radiometers

LI-192 Underwater Quantum Sensor

Lecture Notes Prepared by Prof. J. Francis Spring Remote Sensing Instruments

RECOMMENDATION ITU-R S.733-1* (Question ITU-R 42/4 (1990))**

A Multiwavelength Interferometer for Geodetic Lengths

Chapter 5 Nadir looking UV measurement.

SUPPLEMENTARY INFORMATION

APPLICATION NOTE. Computer Controlled Variable Attenuator for Lasers. Technology and Applications Center Newport Corporation

Observational Astronomy

1/8 m GRATING MONOCHROMATOR

, PMOD-WRC IDEAS+ WP TD3370 Status Pandonia updates Some EPIC info

Microwave Radiometry Laboratory Experiment

CONFIGURING. Your Spectroscopy System For PEAK PERFORMANCE. A guide to selecting the best Spectrometers, Sources, and Detectors for your application

UVscope status report

FREQUENCY DECLARATION FOR THE ARGOS-4 SYSTEM. NOAA-WP-40 presents a summary of frequency declarations for the Argos-4 system.

INTERNATIONAL INTERCOMPARISON OF WAVELENGTH SCALE AND PHOTOMETRIC SCALE OF SPECTROPHOTOMETRY LABORATORIES CENAM - NRC - INMETRO - NIST.

Measuring photometric accuracy using the double aperture method

Will contain image distance after raytrace Will contain image height after raytrace

Imaging Photometer and Colorimeter

RADIOMETRIC CALIBRATION

CONTENTS. Page. Foreword Introduction The Use of Specialized Test Devices 5

The TSIS Spectral Irradiance Monitor: Prism Optical Degradation Studies

Brewer MkIII. Spectrophotometer. Operators manual

Measurements of Infrared Sources with the Missile Defense Transfer Radiometer

Chapter 7 HF Propagation. Ionosphere Solar Effects Scatter and NVIS

The equipment used share any common features regardless of the! being measured. Electronic detection was not always available.

USER S GUIDE. for MIDDLETON SOLAR SECONDARY STANDARD PYRANOMETER WITH INTEGRATING CAVITY DETECTOR

Linear Motion Servo Plants: IP01 or IP02. Linear Experiment #0: Integration with WinCon. IP01 and IP02. Student Handout

PRECISE MEASUREMENTS OF SOLAR BEAM IRRADIANCE THROUGH IMPROVED SENSOR CALIBRATION

Reading 28 PROPAGATION THE IONOSPHERE

SpectroMaster. High Precision Automatic Spectrometer-Goniometer

PRC Krochmann Sky Scanner Characterisation (parts 1 & 2)

AUTOMATION OF 3D MEASUREMENTS FOR THE FINAL ASSEMBLY STEPS OF THE LHC DIPOLE MAGNETS

Specifications. Offers the best spatial resolution for multi-stripe spectroscopy. Provides the user the choice of either high accuracy slit mechanism

AGRON / E E / MTEOR 518 Laboratory

SCCH 4: 211: 2015 SCCH

The FTNIR Myths... Misinformation or Truth

The Standard for over 40 Years

Multispectral. imaging device. ADVANCED LIGHT ANALYSIS by. Most accurate homogeneity MeasureMent of spectral radiance. UMasterMS1 & UMasterMS2

PLANET SURFACE REFLECTANCE PRODUCT

RECOMMENDATION ITU-R SM * Measuring of low-level emissions from space stations at monitoring earth stations using noise reduction techniques

Spectroradiometer characterisations for traceable solar radiation measurements

Using Frequency Diversity to Improve Measurement Speed Roger Dygert MI Technologies, 1125 Satellite Blvd., Suite 100 Suwanee, GA 30024

Tunable KiloArc. Tunable Broadband Light Source.

SEMS-2000/3000 SOLAR ENERGY MEASUREMENT SYSTEM AND COMPLETE METEOROLOGICAL STATION WITH REALLY UNIQUE FEATURES

Components of Optical Instruments

PCS-150 / PCI-200 High Speed Boxcar Modules

ROTATING SHADOWBAND SPECTRORADIOMETER MODEL RSS-1024/UVRSS-1024 BULLETIN RSS/UVRSS-1024

Experimental Analysis of Luminescence in Printed Materials

Low Cost Earth Sensor based on Oxygen Airglow

The Photoelectric Effect

Spectral responsivity uncertainty of silicon photodiodes due to calibration spectral bandwidth

ECEN. Spectroscopy. Lab 8. copy. constituents HOMEWORK PR. Figure. 1. Layout of. of the

SUSIM UARS Calibration and Measurements of Solar UV Spectral Irradiance Variation ( )

AN INTRODUCTION TO MICROCARB, FIRST EUROPEAN PROGRAM FOR CO2 MONITORING.

LINEARPYROMETER LP4. Technical Documentation KE November TN

Cerro Tololo Inter-American Observatory. CHIRON manual. A. Tokovinin Version 2. May 25, 2011 (manual.pdf)

PoS(PhotoDet 2012)058

MICRO SPECTRAL SCANNER

SPECTRAL SCANNER. Recycling

Imaging Beyond the Basics: Optimizing Settings on the Leica SP8 Confocal

A Software Implementation of Data Acquisition Control and Management for Czerny Turner Monochromator

CAL/VAL ACTIVITIES IN ROSHYDROMET. GSICS Executive Panel 14, Tokyo, 15 July. 2013

Multi Sensor Reanalysis (MSR) of total ozone and ozone profiles

SpectroMaster. High Precision Automatic Spectrometer-Goniometer

Automated Double Aperture Accessory

AIXUV's Tools for EUV-Reflectometry Rainer Lebert, Christian Wies AIXUV GmbH, Steinbachstrasse 15, D Aachen, Germany

Picosecond Time Analyzer Applications in...

Radial Polarization Converter With LC Driver USER MANUAL

Basic Components of Spectroscopic. Instrumentation

Disclosure to Promote the Right To Information

RECOMMENDATION ITU-R F.1404*

MICROSCOPE PHOTOMETRY

Antennas and Propagation. Chapter 5

Applications of Steady-state Multichannel Spectroscopy in the Visible and NIR Spectral Region

The RGB code. Part 1: Cracking the RGB code (from light to XYZ)

BGS400 Lamp & Luminaire Goniophotometer (C-γ, B-ß Plane)

Chemistry 524--"Hour Exam"--Keiderling Mar. 19, pm SES

UNIVERSITY OF HAWAII Institute for Astronomy. f/31 High Angular Resolution Imaging Spectrograph HARIS USER MANUAL update June 10, 1997

ARCoptix. Radial Polarization Converter. Arcoptix S.A Ch. Trois-portes Neuchâtel Switzerland Mail: Tel:

Transcription:

DEVELOPMENT OF AN AUTOMATED DOBSON CONTROL SYSTEM FOR UNATTENDED OPERATION R. Stübi 1, W. Siegrist 1, H. Schill 1, S. Brönnimann 1 P.-A. Probst 2, D. Ruffieux 1, B. Calpini 1 1 Federal Office of Meteorology and Climatology, MeteoSwiss, Payerne, Switzerland. 2 Scientific Consulting Sàrl, Saint-Sulpice, Switzerland. ABSTRACT Dobson and Brewer sun spectrophotometers are considered as the reference instruments for total ozone column measurements. If the fully automatic Brewer has replaced the manual Dobson in many countries, the primary reason is rather the lower operational costs than the data quality. Arosa is one of the few stations continuing to measure the total ozone column with multiple colocated Brewer and Dobson instruments. The analysis of these series show that there is still a systematic difference between these two instrument types which is presently not fully understood. Therefore, continuing parallel measurements at different stations still make sense until Dobson and Brewer data agree within their respective uncertainty. A new project has been initiated at MeteoSwiss to develop an automated Dobson in order to continue the measurements at optimized costs and possibly with a more homogeneous and at higher data sampling. First results of the current development indicate a good agreement between the manually operated (D101) and the automatic (D062) Dobson units respectively, with a larger dispersion of the data for the manual instrument. Furthermore the automated Dobson demonstrates similar data dispersion with respect to collocated Brewer B040 (MKII, single monochromator) and Brewer B156 (MKIII, double monochromator). INTRODUCTION Dobson spectrophotometers have been developed at the beginning of the last century by G. M. B. Dobson to measure the ozone total column during sunny days. Figure 1 show the longest worldwide time series of the Arosa station. The long term ozone column has been fairly constant until about 1970 then it starts to decline due to the anthropogenic emission of CFC compounds. The Montreal protocol signed 25 years ago and the subsequent amendments have been successful enough to slow down this decline and stabilise the ozone layer above the mid-latitude. The ozone hole is still observed every year above the South Pole reminding that the problem is far from being solved and that the monitoring of the ozone layer is still required. Brewer spectrophotometers are automated instruments developed in Canada in the eighties. Both Dobson and Brewer are measuring the absorption of the solar UV radiation in the 305 340 nm range. The Brewer are measuring ozone by combining the absolute intensities at four different wavelengths while the Dobson use relative intensities of three different wavelength pairs. Figure 2 shows the distribution of the difference of the monthly means of pairs of collocated instruments. For a pair of Dobson (Fig 2a), respectively a pair of Brewer (Fig 2b), the centres of the density distribution (= systematic bias) are less than 0.5% and their width are within 1-2%. However, the pair comparison of a Dobson and a Brewer (fig 2c) shows a systematic bias of about 2% and variable widths of the distributions depending on the time periods considered. The distribution for the period 1990-1995 (fig 2c, green line) is quite different from the other periods and shows a -1% negative bias which illustrates problems linked to the calibration procedures of the Dobson and the Brewer networks which have evolved over the past 20 years.

Figure 1: Arosa station total ozone time series of the annual means reconstructed from the data of three successive Dobson (D002 D015 D101). Figure 2: Density distribution of the difference [%] of monthly mean values between different instruments and for different time periods as mentioned in the legend: 2a) D062 - D101, 2b) B072 - B040, 2c) B040 - D101 and 2d) D101 - OMI (red line) respectively B040 - OMI (blue line).

Such long term change shows up as an artificial trend linked to instrumental effects which have to be corrected as far as possible. Figure 2d shows the results of the comparison of the overpass data from the OMI satellite instrument with Dobson (red line) and Brewer (blue line) measurements. This figure 2d) reveals a significant difference between the two distributions and the difficulty encountered to validate satellites data against ground based stations data as long as the two instrument types do not agree within their uncertainties. AUTOMATED DOBSON CONTROL SYSTEM Considering that manual operation of the Dobson instrument is nowadays not the optimal solution, a project has been initiated at MeteoSwiss to develop an automated version of this instrument. The goals of the project can be summarised as: Preserving and continuing the world longest total ozone and Umkehr time series Achieving measurements quality equivalent or better than the manual observation Increasing the frequency of Dobson observation Carrying out a systematic comparison between manual and automatic Dobson The measurements principle of the Dobson spectrophotometer is illustrated in figure 3. The sun light beam entering through slit S1 is dispersed by the prism P1 in the right side of the instrument. In the middle, the slits S2 and S3 select two wavelength 1 and 2 in the UV range 305 340 nm. The shorter wavelength 1 in the range 305-320 nm has been significantly attenuated in the atmosphere by the ozone molecules while the longer wavelength 2 around 340 nm has crossed the atmosphere without attenuation. The wedges before slit S3 act as a calibrated attenuator which allows to artificially decrease the intensity of 2 to match the intensity of 1. The wedges are driven externally by the so-called R-dial graduated disk. P2 (focusing prism) P1 (dispersion prism) Figure 3: Illustration of the Dobson instruments light path and measurement principle. The rotating selector wheel after the slits S2 / S3 alternatively let the beam 1 or 2 pass to the left side of the instrument as illustrated by the dashed blue lines and alternatively illuminates the photomultiplier tube (PM) after passing the focusing prism. The PM output is an AC signal at the frequency generated by the rotating wheel which reads zero when the intensities of 1 and 2 are equal. The wavelengths selection is made by the rotation of the quartz plates Q1 (right side) and Q2 (left side) externally controlled by the so-called Q-levers. In the usual measurement mode, three such pairs of wavelengths 1-2 are measured within ~2-3 minutes. The combination of

these three pairs of wavelengths allows improving the final estimate of the atmospheric ozone column measured in Dobson unit [DU]. The first step toward the automation consists in a detailed analysis of the manual operation that has been partially unconsciously developed by the observers. In the second step, the strategy for the implementation of the automatic procedures is defined and the hardware assembly redesigned. The third step is the work at the mechanical workshop and it is a clear milestone in the project since it requires high precision machining and simultaneously avoiding any new friction in the control of the different precision movements (in particular Q-levers and R-dial). In figure 4 below, the details of the realized motor / encoder assemblies are illustrated in the left panel for the R-dial control and in the right panel for the Q-levers wavelength selectors. DC Motor Q2 lever Encoder Q1 lever R-dial driving the attenuator DC Motor Encoder Figure 4: Left panel shows details of the R-dial automatic control: the DC motor at the top drives the axis which goes through the encoder. Right panel: illustration of R-dial, Q1- and Q2-levers assemblies. The control hardware is built around the National Instrument PXI series and consists in: a chassis PXI-1036 a windows PC controller PXI-8101 a data acquisition board M-series PXI-6221 a 4-axis Stepper/Servo Motion Controller PXI-7344 The control software is written in Labview language and the PM output signal measurement is done with a Digital Lock-in Amplifier (DLA). The measurements sequence consists in: pointing the instruments to the sun (azimuth and elevation) setting the chosen wavelength pair (Q1 and Q2 levers) selecting a position of the R-dial (attenuator) close to the zero signal performing a set of three fixed positions around the zero to interpolate the zero position getting to this position and having a PID controller tracking this zero position. recording the R-dial position for typically 20 second.

FIRST RESULTS An example of the R-dial position time series recently measured by the automated Dobson is illustrated in figure 5 for the three wavelength pairs called C, D and A. The first part of the record until 15h50 is a direct sun total ozone column observation followed at 15h55 by a zenith sky Umkehr observation until the sunset. The rupture of the R-dial position around 15h50 reflects the change of the light intensity when passing from a direct sun (high intensity) to a zenith observation (low diffuse intensity) which is compensated by a significant increase of the PM high voltage. One C-D-A cycle is recorded every 2 minutes, half of that time is devoted to the change of the positions of Q1, Q2 and R. The open symbols (right axis of figure 5) are the standard deviations of the R-dial positions measured during the 20 seconds averaging time. The values are of the order of 0.1 deg and increase to 0.2 deg at the end of the Umkehr measurements. This illustrated the very good overall stability of the control system even at low light intensity and high PM voltage. Figure 5: Time series of the raw data for direct sun total ozone column measurements (14h00-15h50) followed by zenith sky Umkehr measurements (~15:55-19:30). The open symbols with the corresponding right axis are the standard deviation of the R-dial positions over the 20 seconds averaging period. An example of the total ozone time series measured by four different spectrophotometers collocated in Arosa is illustrated in figure 6. The automated Dobson (black symbols) are compared to the manual operated Dobson (red symbols) and to the automatic Brewer instruments (blue and green symbols). The systematic difference of 3-8 DU between Dobson and Brewer instruments is well known as already illustrated in figure 2c. The precision of the automated Dobson is comparable to that of B040 instrument (green) and is a little better to the precision of B156 (blue) for that particular day. Figure 6 also clearly shows the good agreement between the two Dobson with a larger dispersion of the data for the manually operated instrument.

Figure 6: Time series of the total ozone column measured by 4 different instruments collocated at Arosa station, Switzerland. Presently, the automated data acquisition is operational and a systematic comparison with the manual measurements has started under the control of the operator. The next step will be the development of a new shelter and the automatic instruments orientation. CONCLUSION Recent developments towards an automated version of the Dobson sun spectrophotometer have been presented. The first results of the comparison with a manually operated Dobson as well as with collocated Brewer instruments show that the data are in good agreement. The precision of the automated Dobson is as good as the manual one and it is also comparable to the Brewer instruments. The frequency of the data acquisition for the automated Dobson is comparable to the one of the Brewer. This first stage of the project gives very promising results and the development will be continued towards a 100 % automatic version of the Dobson. In parallel, systematic series of comparison will be started to assure the equivalence of the manual and automated versions of the Dobson.