Notes on Laser Resonators

Similar documents
ECE 185 HELIUM-NEON LASER

OPTI 511L Fall (Part 1 of 2)

SA210-Series Scanning Fabry Perot Interferometer

visibility values: 1) V1=0.5 2) V2=0.9 3) V3=0.99 b) In the three cases considered, what are the values of FSR (Free Spectral Range) and

CHAPTER 5 FINE-TUNING OF AN ECDL WITH AN INTRACAVITY LIQUID CRYSTAL ELEMENT

Ph 77 ADVANCED PHYSICS LABORATORY ATOMICANDOPTICALPHYSICS

MASSACHUSETTS INSTITUTE OF TECHNOLOGY Department of Electrical Engineering and Computer Science

EE119 Introduction to Optical Engineering Fall 2009 Final Exam. Name:

Diffraction. Interference with more than 2 beams. Diffraction gratings. Diffraction by an aperture. Diffraction of a laser beam

R. J. Jones Optical Sciences OPTI 511L Fall 2017

Constructing a Confocal Fabry-Perot Interferometer

Fabry Perot Resonator (CA-1140)

A novel tunable diode laser using volume holographic gratings

EE119 Introduction to Optical Engineering Spring 2003 Final Exam. Name:

R. J. Jones College of Optical Sciences OPTI 511L Fall 2017

EE119 Introduction to Optical Engineering Spring 2002 Final Exam. Name:

DPSS 266nm Deep UV Laser Module

R.B.V.R.R. WOMEN S COLLEGE (AUTONOMOUS) Narayanaguda, Hyderabad.

Experimental Physics. Experiment C & D: Pulsed Laser & Dye Laser. Course: FY12. Project: The Pulsed Laser. Done by: Wael Al-Assadi & Irvin Mangwiza

Principles of Optics for Engineers

Advanced Features of InfraTec Pyroelectric Detectors

Department of Electrical Engineering and Computer Science

ECEN. Spectroscopy. Lab 8. copy. constituents HOMEWORK PR. Figure. 1. Layout of. of the

OPTICS AND LASER PHYSICS LABORATORY #10 INSIDE A LASER CAVITY -- EXPLORING STABILITY, POLARIZATION, AND MODES with Mark Chawla and Chris Baird

Optics and Lasers. Matt Young. Including Fibers and Optical Waveguides

DESIGN OF COMPACT PULSED 4 MIRROR LASER WIRE SYSTEM FOR QUICK MEASUREMENT OF ELECTRON BEAM PROFILE

Optical Communications and Networking 朱祖勍. Sept. 25, 2017

Characteristics of point-focus Simultaneous Spatial and temporal Focusing (SSTF) as a two-photon excited fluorescence microscopy

Fabry-Perot Interferometer

B. Cavity-Enhanced Absorption Spectroscopy (CEAS)

The electric field for the wave sketched in Fig. 3-1 can be written as

Optical System Design

ADVANCED OPTICS LAB -ECEN Basic Skills Lab

Tutorial Zemax 9: Physical optical modelling I

LOS 1 LASER OPTICS SET

Fabry-Perot Interferometer

Lecture 04: Solar Imaging Instruments

Module 4 : Third order nonlinear optical processes. Lecture 24 : Kerr lens modelocking: An application of self focusing

Optical Components for Laser Applications. Günter Toesko - Laserseminar BLZ im Dezember

Will contain image distance after raytrace Will contain image height after raytrace

Laser Beam Analysis Using Image Processing

Lecture 2: Geometrical Optics. Geometrical Approximation. Lenses. Mirrors. Optical Systems. Images and Pupils. Aberrations.

Fabry-Perot Cavity FP1-A INSTRUCTOR S MANUAL

Laser stabilization and frequency modulation for trapped-ion experiments

Mirrors and Lenses. Images can be formed by reflection from mirrors. Images can be formed by refraction through lenses.

Chapter Ray and Wave Optics

Physics 431 Final Exam Examples (3:00-5:00 pm 12/16/2009) TIME ALLOTTED: 120 MINUTES Name: Signature:

Application Note (A11)

Lecture 3: Geometrical Optics 1. Spherical Waves. From Waves to Rays. Lenses. Chromatic Aberrations. Mirrors. Outline

Experiment 1: Fraunhofer Diffraction of Light by a Single Slit

ECE 185 ELECTRO-OPTIC MODULATION OF LIGHT

Properties of Structured Light

PHYS2090 OPTICAL PHYSICS Laboratory Microwaves

KULLIYYAH OF ENGINEERING

Lecture 2: Geometrical Optics. Geometrical Approximation. Lenses. Mirrors. Optical Systems. Images and Pupils. Aberrations.

6545(Print), ISSN (Online) Volume 4, Issue 2, March April (2013), IAEME & TECHNOLOGY (IJEET)

APPLICATION NOTE

PHYS 3153 Methods of Experimental Physics II O2. Applications of Interferometry

06SurfaceQuality.nb Optics James C. Wyant (2012) 1

DIODE LASER SPECTROSCOPY (160309)

Heisenberg) relation applied to space and transverse wavevector

1. Evolution Of Fiber Optic Systems

Observational Astronomy

Lab 12 Microwave Optics.

GEOMETRICAL OPTICS Practical 1. Part I. BASIC ELEMENTS AND METHODS FOR CHARACTERIZATION OF OPTICAL SYSTEMS

Exercise 1-3. Radar Antennas EXERCISE OBJECTIVE DISCUSSION OUTLINE DISCUSSION OF FUNDAMENTALS. Antenna types

Lecture 5: Introduction to Lasers

First characterizations of a Ring Cavity for a colt atom gravimeter

Improving efficiency of CO 2

Wave Front Detection for Virgo

MASSACHUSETTS INSTITUTE OF TECHNOLOGY Department of Electrical Engineering and Computer Science

Spectroscopy Lab 2. Reading Your text books. Look under spectra, spectrometer, diffraction.

SELECTION GUIDE MULTIPLE-ORDER QUARTZ WAVEPLATES ZERO-ORDER QUARTZ WAVEPLATES DUAL-WAVELENGTH WAVEPLATES... 85

EUV Plasma Source with IR Power Recycling

FFP-TF2 Fiber Fabry-Perot Tunable Filter Technical Reference

101 W of average green beam from diode-side-pumped Nd:YAG/LBO-based system in a relay imaged cavity

Physics 476LW. Advanced Physics Laboratory - Microwave Optics

Fiber Optic Communications Communication Systems

Big League Cryogenics and Vacuum The LHC at CERN

Collimation Tester Instructions

Study on Imaging Quality of Water Ball Lens

Radial Polarization Converter With LC Driver USER MANUAL

NEW LASER ULTRASONIC INTERFEROMETER FOR INDUSTRIAL APPLICATIONS B.Pouet and S.Breugnot Bossa Nova Technologies; Venice, CA, USA

Mode analysis of Oxide-Confined VCSELs using near-far field approaches

Be aware that there is no universal notation for the various quantities.

Faraday Rotators and Isolators

ADVANCED OPTICS LAB -ECEN 5606

GIST OF THE UNIT BASED ON DIFFERENT CONCEPTS IN THE UNIT (BRIEFLY AS POINT WISE). RAY OPTICS

A Possible Design of Large Angle Beamstrahlung Detector for CESR

880 Quantum Electronics Optional Lab Construct A Pulsed Dye Laser

Use of Computer Generated Holograms for Testing Aspheric Optics

Single-photon excitation of morphology dependent resonance

Installation and Characterization of the Advanced LIGO 200 Watt PSL

Absorption: in an OF, the loss of Optical power, resulting from conversion of that power into heat.

The Beam Characteristics of High Power Diode Laser Stack

Introduction to the operating principles of the HyperFine spectrometer

Computer Generated Holograms for Optical Testing

2. Pulsed Acoustic Microscopy and Picosecond Ultrasonics

Swept Wavelength Testing:

Fibre optics (Item No.: P )

Transcription:

Notes on Laser Resonators 1 He-Ne Resonator Modes The mirrors that make up the laser cavity essentially form a reflecting waveguide. A stability diagram that will be covered in lecture is shown in Figure 1. Planar Symmetrical Confocal 0 < (1 + dr1 ) (1 + dr2 ) 1 R 1 =R 2 Figure 1: Stability Diagram for standing-wave modes. The solutions to the Maxwell equations, given a combination of spherical mirrors, results in several modes. The TEM modes are the transverse electric and transverse magnetic modes that resonate in the cavity. The lowest order mode approximates a Gaussian profile. A plot of these modes is shown in Figure 2 1

Figure 2: Hermite-Gaussian modes. 1.1 Modes Many applications of lasers require a minimum spot size. This requires that (a) the wavefront of the output has the same phase across its entire surface, and that (b) the edge of the wavefront falls off slowly, rather than abruptly as if it has passed through an aperture. TEMoo TEM01 Figure 3: Modes of a He-Ne The TEM 00 mode in Figure 3 satisfies the two conditions mentioned above, while the mode has two lobes and thus does not satisfy requirement (a). The TEM 01 mode is a superposition of two degenerate TEM 01 modes - one with a sine dependence on the azithuthal angle and one with a cosine dependence. The sum of these two mode produces a donut mode with no field in the center of the mode. This mode has a spot size that is larger than that of the TEM 00 mode (see Reference 2). 2

1.2 Brewster Windows Many laser lines including the 0.6328μm for an He-Ne line have very low gain. As a result the laser cavity must be tailored to minimize losses. Figure 4: Configuration of Brewster Windows. Brewster windows (see Figure 4) are utilized at both ends of the laser tube to minimize reflection losses. As drawn in the figure, the Brewster windows will transmit vertically polarized light with no surface reflections. Consequently, the laser output will be polarized. For a He-Ne laser, other laser lines (1.15μm and 3.39μm) can be used by constructing cavities with appropriately coated mirrors, and different Brewster window material, as windows made of glass or quartz absorb infrared radiation and will thus increase the lasing threshold for these wavelengths. 1.3 Types of Resonators For laser resonators, the only way for a uniphase wavefront to occur is to force all modes except for the lowest order mode TEM 00 to have high losses, so that they operate below threshold. To achieve this condition, the width of the lowest order mode should ideally equal the radius of the laser cavity mirrors. However, dust particles located on the mirror can produce different modal diffraction losses and can suppress the lowest order mode operation. It is therefore essential to design the laser with an adjustable mode waist W 0 to laser mirror aperture ratio. One way is accomplish this is to use an adjustable mirror separation. (see Figure 5) The beam waist W(z) as a function of z and the radius of curvature of the field R(z) are derived in the book and in ECE 181. They are ( )] λz 1/2 W(z) = W 0 [1 + πw0 2 (1) [ ) 2 ] R(z) = z 1 + ( πw 2 0 λz (2) 3

Figure 5: Resonator Configurations. The value b in the figure is the mirror radius R. Resontor Configurations 1. Plane-parallel resonator This was used in early designs of gas lasers. However, due to diffractions around the edges, the output has a diverging wavefront. Also, the ratio of TEM 01 diffraction loss to that of TEM 00 is a factor of two, which is much smaller than that of other resonator configurations. This small difference can easily lead to higher-order mode operation if other imperfections exist. Furthermore, a plane-parallel resonator is extremely susceptible to high losses resulting from mirror misalignment, mechanical vibrations, and thermal effects. It is also sensitive to the optical quality of the mirrors and the Brewster windows if they are used. For these reasons, plane parallel resonators are rarely used. 2. Large radius resonator For large R, we can neglect the factor of one in Eq.(2) and set z = d/2 so that the origin in halfway between the mirrors, we then have so that R 2π2 W 4 0 λ 2 d W 4 0 ( ) λ 2 R d π 2 Thus the waist W 0 is a very slowly varying function of R. Resonators with large radius mirrors can maximize the use of excited atoms in the cavity and thus potentially give high output power. Also, mirror alignment is not as critical as that in a plane-parallel resonator. We use large radius mirrors in the lab. 3. Confocal resonators When R = d, the resonator becomes confocal. From Eq.(2), W 0 is given by (3) Rλ W 0 = (4) 2π Thus for a given distance between the mirrors d or size of mirror R, this corresponds to the smallest W 0. As a result, confocal resonators are used whenever the smallest plasma tube diameter is desired, e.g. for R = 300 cm, λ = 0.6328µm, 2W 0 1.5 mm. When d is very close to R, W 0 varies very slowly with d. One disadvantage for confocal resonators 4

is that for two mirrors which have slightly different radii, say R 1 < R 2, then the resonator is stable only if d < R 1 or d > R 2. Thus in practice, d has to be adjustable. 4. Spherical resonator When 2R = d, it is called a spherical resonator. From Eq.(2), W(z) is very large and it is focused down to a diffraction limited point at the center of the sphere. 1.4 Alignment of Resonators The sensitivity to alignment variations of resonators can be deduced from the fact that the uniphase wavefront mode must lie symmetrically along the line joining the center of curvature of the two mirrors. R 1 +R 2 -d R 1 φ R 2 d x θ y R 1 θ Figure 6: Mirror alignment parameters. The parameter b is the mirror radius R. In Figure 6, mirror M 2 is misaligned with respect to mirror M 1 by an angle θ measured from the center of M 1 with respect to the optical axis. The misalignment can also be measured from the center C 2 of the mirror M 2 with an angle φ. Using the distances shown in Figure 6, we can then write R 1 θ = (R 1 + R 2 d)φ. (5) Defining x, as the transverse displacement of the beam spot at M 1, and y as transverse displacement of the beam spot at M 2, we have x = xφ = (R 2 d)r 1 θ R 1 + R 2 d y = yφ = R 2 R 1 θ R 1 + R 2 d where x and y are measured from the left side of the figure. The alignment sensitivity is then defined as the smaller of the two angles θ = W xθ x or θ = W yθ y. (8) where the waist W depends on the type of resonator. Small values of θ are more sensitive. With Eqs. (5), (6) and (7), we have the following summary: 1. Large radius (R d): x y = Rθ/2 and θ 2W/R. The alignment is not as critical as in (1). 2. Confocal x = 0, and y = Rθ and θ y = W/R. The alignment is least sensitive among all the various configurations. 3. Spherical θ = Wδ/R 2, where δ = R 1 + R 2 d is small. The alignment is critical because δ is small. 5 (6) (7)

2 Resonator Based Interferometers Building a resonator and controlling the distance between the mirrors, one can vary the resonance frequency and thus construct a scaning interferometer. Scanning spherical-mirror interferometers (SSMI) are common tools for high resolution analysis of laser mode structures. They can be used in conjunction with a radio frequency (RF) spectrum analyzer which displays the frequency difference generated by the mixing laser modes when a square-law sensor is used. There are two types of laser modes. Longitudinal modes are associated with different modes of oscillation of the laser and are characterized by their oscillation frequency. Transverse modes are characterized by the field intensity distribution in a plane perpendicular to the direction of propagation. Several longitudinal modes can have the same form of the transverse field distribution. The SSMI are useful in examining the laser mode structure. Among the various SSMI, the mode degenerate interferometers, especially the confocal ones, are comparatively easy to use. 2.1 Scanning Fabry-Perot Interferometer SSMI belong to the general class of Fabry-Perot interferometers, which consist of two mirrors placed parallel to each another with a separation d. The resonance condition for such an interferometer is when the optical path between the mirrors is equal to an integral number (m) of half wavelengths of the incident light. For normal incidence, the resonance condition is f m = m c 2d where m is an integer and c/2d is the resonators free-spectral range. Let φ = kd = 2π f d c be the phase shift the field experiences where k = 2π λ = 2π f c is the wavenumber. The transmittance of the Fabry-Perot is specified by the Airy function I out I in = 1 1 + 4R (1 R) 2 sin 2 (φ) (9) where I in and I out are the input and output intensities, R and T are the reflection and transmission coefficient of the mirrors used in the resonator respectively. When R 1, then the resonance features become very sharp. The Airy function near one of the resonance features φ m = 2π f md c may then be written as I out I in = [ 1 + 4 (1 R) 1 ] 2 (10) (φ φm ) 2 which has the form of a Lorentzian with a full-width half-maximum width (FWHM) given by f = c(1 R) 2πd This width specifies the instrument resolution. The ratio of free-spectral range to f is called the finesse (11) F = c 2d c(1 R) 2πd = π (1 R) An ordinary Fabry-Perot is not useful in analyzing laser modes. A typical gas laser transition has a Doppler linewidth of a few gigahertz, and a mode spacing of a few tens of Mhz for 6

the size of the resonators used in the lab. This implies that the interferometer must have a FSR larger than the Doppler linewidth, and a f smaller than the mode spacing. For gas lasers, this typically means a finesse of at least 100. However, the finesse of an ordinary Fabry- Perot is limited by the flatness and the apertures of the plane mirrors, and for small apertures, diffraction loss becomes significant. A plot of a single resonance of a Fabry-Perot for three values of the finesse is shown in Figure 7. 1.0 Fabry Perot 0.8 Trans. 0.6 0.4 F=20 0.2 F=200 F=100 0.0 0.5 0.0 0.5 phase Figure 7: Plot of a single resonance feature of a Fabry-Perot for three values of the finesse. 2.2 Spherical Mirror Resonators The diffraction effects of the Fabry-Perot etalon can be eliminated by using spherical, instead of plane, mirrors in the interferometer. However, the radius of curvature R must be greater than d/2. The spherical mirror also alleviates the constraint on surface figure, because only a small area of the mirror is used. One requirement of a general spherical-mirror interferometer is that it must be illuminated with a narrow, diffraction limited beam, i.e. the beam must be mode-matched with the interferometer for proper operation. The various resonant modes of such a cavity is given by f qmn = c 2d [q + 1π ( (1 + m + n) cos 1 1 d )] R where q is an integer denoting longitudinal mode number, m and n are integers denoting transverse mode numbers. From Eq.(12), it can be seen that in order to have a large enough FSR, m and n must be fixed such that FSR = c/2d. In practice, the only transverse mode of the interferometer, which is convenient to excite is the TEM 00 mode. This single transverse mode requirement limits the use of a general spherical-mirror interferometer, as the laser itself may not operate in the TEM 00 mode. Moreover, an optical isolator must be used to avoid feedback of light from a well aligned interferometer into the laser. 2.3 Mode Degenerate Interferometers In contrast to the general spherical-mirror interferometer, the mode-degenerate interferometer does not need to be mode-matched to the incident laser beam. This results in several simplifications: 1. The laser does not have to operate in a single transverse mode. 2. There is no spurious resonances due to mode-mismatching. 7 (12)

3. The interferometer need not be accurately aligned along the axis of the incident laser beam. This also allows the operation without an optical isolator as the light does not reflect back into the laser. A mode-degenerate interferometer is a spherical-mirror interferometer whose transverse modes are degenerate in frequency. When the condition cos 1 (1 d/r) = π l (13) (where l is an integer) is satisfied, then the resonance condition in Eq.(12) becomes f qmn = c [ q + 1 2d π (1 + m + n) π ] l = c [lq + (1 + m + n)] 2ld By increasing m f n by l, and decreasing q by 1, f qmn remains unchanged. Thus the interferometer will have l "sets" of degenerate transverse modes with equal mode spacing. It has an FSR of c/2ld and a finesse of π/l(1 R). The best known mode-degenerate interferometer is the confocal one, with l = 2. It has "even symmetric" modes corresponding to (m + n) even, and "odd symmetric" modes corresponding to (m + n) odd. Mode degenerate interferometers can also be analyzed in terms of geometric optics. For instance, Eq.(13) is equivalent to the condition that a ray launched in the cavity retraces its path after l complete travels of the cavity. A typical ray path for the confocal interferometer is shown in Figure 8. For this resonator, l = 2 and the ray retraces its path after two complete round trips in the resonator regardless of the initial launch angle. This independence of round-trip on launch angle is the ray optics interpretation of mode degeneracy. (14) Figure 8: Ray paths in a confocal resonator. The performance of mode-degenerate interferometers is limited by the mirror reflectivity, mirror surface figure, and spherical aberration. The reflectivity limitation is not serious as high reflectance (R>0.998) has been achieved in practice giving a finesse greater than 750. In order to minimize the effect of surface figure of the mirrors, the laser beam diameter should be as small as possible (approximately that of the TEM 00 mode of the interferometer). To avoid aberration, the beam should be close to the interferometer axis, so that the paraxial optics approximation applies. 2.4 Practical Mode Degenerate Interferometers Figure 9 shows a typical scanning confocal interferometer. It consists of two spherical mirrors, separated by a distance equal to their radius of curvature. The back surfaces of the mirrors are made such that the mirrors are self-collimating, so that a plane wavefront incident on the interferometer is transformed into a spherical wavefront with a radius of curvature that is 8

matched to the transverse modes of the interferometer. The concave surfaces are coated with high reflectance dielectric films; the convex ones are coated with anti-reflection films to eliminate spurious resonances associated with the back surfaces. The mirrors are mounted in a cell whose spacing can be controlled (to within a few wavelengths) by a voltage applied to the piezoelectric spacer. For confocal interferometers, a scan of λ/3 in spacing corresponds to one FSR. An aperture is placed outside the entrance mirror to limit the diameter of the incident beam and hence reduce spherical aberration. The transmitted light is detected by a photodetector whose electrical output is viewed on an oscilloscope, as a function of the voltage applied to the piezoelectric spacer. Piezoelectric Spacer Figure 9: Confocal Interferometer with Piezoelectric actuator for scanning. It is important that the mirror separation be quite close to confocal. The tolerance on the mirror spacing depends on the length of the interferometer and its finesse. For very short, high finesse interferometers, the mirror spacing should be set to within a few wavelengths. In practice, the adjustment of length is quite easy to make. The mirror spacing should first be set so that it is approximately confocal; then by observing the mode structure of a laser on an oscilloscope, the final adjustment of the mirror spacing can be made to maximize the finesse of the interferometer. Once the separation of the interferometer mirrors is set, there is no need for further adjustment. 2.4.1 Use of Scanning Confocal Interferometers Confocal resonators are useful for a variety of laser diagnostics. By focusing the laser beam into the scanning confocal interferometer, different transverse laser modes can be monitored, and the laser can be adjusted to operate in a single transverse mode. Fig. 4 shows the mode spectra for a He-Ne laser in double transverse mode, and single transverse mode operation. 9

(a) (b) Figure 10: (a) Output spectrum of a helium-neon laser operating simultaneously in two transverse modes, (b) Output spectrum of a helium-neon laser operating in a single transverse mode. The vertical sensitivity and horizontal dispersion are identical for both (a) and (b). 2.5 Laboratory Scanning Interferometer The Tropel model 240 used in the laboratory has relatively high reflectance dielectric mirrors. These mirror provide a good compromise between high spectral resolution and high instrumental transmission at the center of the bandpass. The exact separation of the two mirrors has been made insensitive to reasonable room temperature variations by the use of thermal compensation in the interferometer assembly. Parameter Free spectral range Instrumental bandwidth Spectral Resolving Power Finesse Peak instrumental transmission Symbol FSR f Q F Value 1500MHz 75MHz 8 107 200 20-30% Table 1: Specification for Tropel 240 Interferometer. We will used the Tropel 240 in the scanning mode shown in Figure 11. Figure 11: Set-up for scanning interferometer. 10

2.5.1 Arrangement for scanning. When used in the scanning mode, a time-varying voltage is applied to the "scanning voltage" terminal. This causes the mirror separation to change through the action of a piezoelectric transducer, and this in turn varies the resonant frequency of the interferometer. The model 240 will scan over one free spectral range, or 1500MHz, with the application of 30-40 volts. If the time-varying voltage is made periodic, the resultant repetitive display of the spectrum, with a 1500MHz scan period, provides a convenient self-calibration of the frequency scale. Note that an increase in applied scan voltage produces a decrease in the resonant frequency of the interferometer. Figure 12 illustrates a typical oscilloscope trace obtained when the model 240 is used in the scanning mode. sinusoidal scanning voltage input spectrum analyzer detector output Figure 12: Typical laser spectra. There are a few points to consider when using the model 240: 1. If the spectrum analyzer and laser beam are exactly co-linear, the spectrum of the laser may become; an erratic function of time due to optical feedback from the spectrum analyzer into the laser. This can be eliminated by a slight adjustment of the alignment, or by the insertion of a circular polarizer (linear polarizer plus 1/4-wave plate) between the laser and the spectrum analyzer. A relatively slow temporal variation of the spectrum of the laser is typical of many commercially available lasers and is due to thermal variations in the laser cavity length. 2. Once a high resolution display is obtained on the scope, the user should experiment to find out the angle through which the alignment can be varied without seriously affecting the instrumental bandwidth. In general, the final alignment is accomplished by observing the displayed spectrum while touching up the alignment adjustment screws to maximize the amplitude of the display, and to minimize the linewidth of each observed longitudinal mode. 3. Whenever possible, use a collimated beam with a small beam diameter (on the order of 1-2mm). This will provide the maximum signal and minimum instrumental bandwidth, while at the same time will allow maximum alignment tolerance. 11

4. Scan rates greater than a 200Hz should be avoided because the scan becomes nonlinear at high frequencies. Moreover, the silicon photo diode cannot resolve pulses shorter than a few microseconds, since zero reverse bias voltage is applied to the detector in the model 240. For continuous scans, a sinusoidal scanning voltage should be used to prevent damage to the piezoelectric transducers. In practice, a scanning voltage of around 35Vp-p at 100 Hz is recommended for obtaining an initial display on the scope. sinusoidal scanning voltage input 3 References These notes are abstracted from: 1. Spectra Physics Laser Technical Bulletin No. 2 by A.L. Bloom, June 1963. 2. W.W. Rigrod, "Isolation of axi-symmetrical-optical-resonator modes," App. Phy. Lett., Vol. 2, pp. 51-53, Feb. 1, 1963. 3. G.D. Boyd and J. P. Gordon, "Confocal multimode resonator for millimeter through optical wavelength masers," Bell System Tech. /., Vol. 40, pp. 489-508, March 1961. 4. G.D. Boyd and H. Kogelnik, "Generalized confocal resonator theory," Bell System Tech.J., Vol. 41, pp. 1347-1369, July 1962. 12