A Novel Low Noise High Gain CMOS Instrumentation Amplifier for Biomedical Applications

Similar documents
Nizamuddin M., International Journal of Advance Research, Ideas and Innovations in Technology.

CMOS Instrumentation Amplifier with Offset Cancellation Circuitry for Biomedical Application

ANALYSIS AND DESIGN OF HIGH CMRR INSTRUMENTATION AMPLIFIER FOR ECG SIGNAL ACQUISITION SYSTEM USING 180nm CMOS TECHNOLOGY

A Novel Design of Low Voltage,Wilson Current Mirror based Wideband Operational Transconductance Amplifier

DESIGN OF A NOVEL CURRENT MIRROR BASED DIFFERENTIAL AMPLIFIER DESIGN WITH LATCH NETWORK. Thota Keerthi* 1, Ch. Anil Kumar 2

Low Power Low Noise CMOS Chopper Amplifier

CHAPTER 3. Instrumentation Amplifier (IA) Background. 3.1 Introduction. 3.2 Instrumentation Amplifier Architecture and Configurations

Design of High Gain Two stage Op-Amp using 90nm Technology

Pankaj Naik Electronic and Instrumentation Deptt. SGSITS, Indore, India. Priyanka Sharma Electronic and. SGSITS, Indore, India

G m /I D based Three stage Operational Amplifier Design

Design of Low Voltage High Speed Operational Amplifier for Pipelined ADC in 90 nm Standard CMOS Process

Performance Evaluation of Different Types of CMOS Operational Transconductance Amplifier

A New Design Technique of CMOS Current Feed Back Operational Amplifier (CFOA)

Chapter 5. Operational Amplifiers and Source Followers. 5.1 Operational Amplifier

DESIGNING OF CURRENT MODE INSTRUMENTATION AMPLIFIER FOR BIO-SIGNAL USING 180NM CMOS TECHNOLOGY

Sensors & Transducers Published by IFSA Publishing, S. L.,

An Improved Recycling Folded Cascode OTA with positive feedback

Design and Implementation of High Gain, High Bandwidth CMOS Folded cascode Operational Transconductance Amplifier

Design and Analysis of High Gain Differential Amplifier Using Various Topologies

Design of Low Voltage Low Power CMOS OP-AMP

Design Analysis and Performance Comparison of Low Power High Gain 2nd Stage Differential Amplifier Along with 1st Stage

A Complete Analog Front-End IC Design for ECG Signal Acquisition

High Speed CMOS Comparator Design with 5mV Resolution

ECEN 474/704 Lab 6: Differential Pairs

Design and Analysis of Double Gate MOSFET Operational Amplifier in 45nm CMOS Technology

Design and Simulation of Low Voltage Operational Amplifier

Low Power High Speed Differential Current Comparator

Design and Simulation of Low Dropout Regulator

Analysis of Instrumentation Amplifier at 180nm technology

Enhancing the Slew rate and Gain Bandwidth of Single ended CMOS Operational Transconductance Amplifier using LCMFB Technique

Analysis of Two Stage Folded Cascode Operational Amplifier in 90nm Technology

DAT175: Topics in Electronic System Design

Design and Analysis of Low Power Two Stage CMOS Op- Amp with 50nm Technology

Design of Gain Enhanced and Power Efficient Op- Amp for ADC/DAC and Medical Applications

IJSRD - International Journal for Scientific Research & Development Vol. 4, Issue 03, 2016 ISSN (online):

A High Gain and Improved Linearity 5.7GHz CMOS LNA with Inductive Source Degeneration Topology

LOW VOLTAGE ANALOG IC DESIGN PROJECT 1. CONSTANT Gm RAIL TO RAIL INPUT STAGE DESIGN. Prof. Dr. Ali ZEKĐ. Umut YILMAZER

Design of CMOS Instrumentation Amplifier

Gain Boosted Telescopic OTA with 110db Gain and 1.8GHz. UGF

Low Power and Fast Transient High Swing CMOS Telescopic Operational Amplifier

Analog CMOS Interface Circuits for UMSI Chip of Environmental Monitoring Microsystem

Design and Performance Analysis of Low Power RF Operational Amplifier using CMOS and BiCMOS Technology

Rail-To-Rail Output Op-Amp Design with Negative Miller Capacitance Compensation

A Low Power Low-Noise Low-Pass Filter for Portable ECG Detection System

Common Mode Feedback for Fully Differential Amplifier in ami06 micron CMOS process

Radivoje Đurić, 2015, Analogna Integrisana Kola 1

Effect of Current Feedback Operational Amplifiers using BJT and CMOS

DESIGN OF A FULLY DIFFERENTIAL HIGH-SPEED HIGH-PRECISION AMPLIFIER

Design of Operational Amplifier in 45nm Technology

DESIGN AND SIMULATION OF CURRENT FEEDBACK OPERATIONAL AMPLIFIER IN 180nm AND 90nm CMOS PROCESSES

Transconductance Amplifier Structures With Very Small Transconductances: A Comparative Design Approach

Design of High gain and Low Offset CMOS Current Mode Front End Operational Amplifier

ECE 415/515 ANALOG INTEGRATED CIRCUIT DESIGN

A PSEUDO-CLASS-AB TELESCOPIC-CASCODE OPERATIONAL AMPLIFIER

Design and implementation of two stage operational amplifier

Design of Low Voltage, Low Power Rail to Rail Operational Transconductance Amplifier with enhanced Gain and Gain Bandwidth Product

IJSRD - International Journal for Scientific Research & Development Vol. 4, Issue 02, 2016 ISSN (online):

Design and Implementation of less quiescent current, less dropout LDO Regulator in 90nm Technology Madhukumar A S #1, M.

International Journal of Advance Engineering and Research Development

Analog Electronics. Lecture Pearson Education. Upper Saddle River, NJ, All rights reserved.

Comparative Analysis of Compensation Techniques for improving PSRR of an OPAMP

EE320L Electronics I. Laboratory. Laboratory Exercise #2. Basic Op-Amp Circuits. Angsuman Roy. Department of Electrical and Computer Engineering

A Switched-Capacitor Band-Pass Biquad Filter Using a Simple Quasi-unity Gain Amplifier

Performance Analysis of Low Power, High Gain Operational Amplifier Using CMOS VLSI Design

Design of a Folded Cascode Operational Amplifier in a 1.2 Micron Silicon-Carbide CMOS Process

Concepts to be Reviewed

Design of High Gain Low Voltage CMOS Comparator

DESIGN AND PERFORMANCE VERIFICATION OF CURRENT CONVEYOR BASED PIPELINE A/D CONVERTER USING 180 NM TECHNOLOGY

A Low Power Low Voltage High Performance CMOS Current Mirror

Differential Amplifier : input. resistance. Differential amplifiers are widely used in engineering instrumentation

Design of A Low Voltage Low Power CMOS Current Mirror with Enhanced Dynamic Range

AN-1106 Custom Instrumentation Amplifier Design Author: Craig Cary Date: January 16, 2017

Design and Analysis of Two-Stage Op-Amp in 0.25µm CMOS Technology

DESIGN HIGH SPEED, LOW NOISE, LOW POWER TWO STAGE CMOS OPERATIONAL AMPLIFIER. Himanshu Shekhar* 1, Amit Rajput 1

Instrumentation Amplifier design: Comparison of CMOS-memristive to CMOS design

SALLEN-KEY FILTERS USING OPERATIONAL TRANSCONDUCTANCE AMPLIFIER

Design of Low Power Preamplifier Latch Based Comparator

EE Analog and Non-linear Integrated Circuit Design

IMPLEMENTATION OF A LOW-KICKBACK-NOISE LATCHED COMPARATOR FOR HIGH-SPEED ANALOG-TO-DIGITAL DESIGNS IN 0.18

[Kumar, 2(9): September, 2013] ISSN: Impact Factor: 1.852

Lecture 300 Low Voltage Op Amps (3/28/10) Page 300-1

ALow Voltage Wide-Input-Range Bulk-Input CMOS OTA

Operational Amplifier with Two-Stage Gain-Boost

International Journal of Pure and Applied Mathematics

Topology Selection: Input

A new class AB folded-cascode operational amplifier

Design and Layout of Two Stage High Bandwidth Operational Amplifier

Designing CMOS folded-cascode operational amplifier with flicker noise minimisation

LOW CURRENT REFERENCES WITH SUPPLY INSENSITIVE BIASING

Ultra Low Power Multistandard G m -C Filter for Biomedical Applications

CMOS Operational Amplifier

V d = "1" if V in > V m. Fig 2: Frequency analysis of the PDM signal. Fig 1: PDM signal generation

High Voltage Operational Amplifiers in SOI Technology

DESIGN OF LOW POWER AND HIGH GAIN BOOSTED OTA FOR HIGH FREQUENCY RADIO MODULATIONS AND TELECOMMUNICATION SYSTEMS

LOW VOLTAGE / LOW POWER RAIL-TO-RAIL CMOS OPERATIONAL AMPLIFIER FOR PORTABLE ECG

Cascode Bulk Driven Operational Amplifier with Improved Gain

I1 19u 5V R11 1MEG IDC Q7 Q2N3904 Q2N3904. Figure 3.1 A scaled down 741 op amp used in this lab

DESIGN OF 2.4 GHZ LOW POWER CMOS TRANSMITTER FRONT END

Design of Rail-to-Rail Op-Amp in 90nm Technology

A High-Driving Class-AB Buffer Amplifier with a New Pseudo Source Follower

Transcription:

International Journal of Electrical and Computer Engineering (IJECE) Vol. 3, No. 4, August 2013, pp. 516~523 ISSN: 2088-8708 516 A Novel Low Noise High Gain CMOS Instrumentation Amplifier for Biomedical Applications Akshay Goel, Gurmohan Singh Centre for Development of Advanced Computing (C-DAC),Mohali, India Article Info Article history: Received May 10, 2013 Revised Jun 29, 2013 Accepted Jul 12, 2013 Keyword: Instrumentation Amplifier Gain Input and Output Referred Noise CMRR Folded Cascode Amplifier ABSTRACT This work describes a novel technique of designing a low noise high gain CMOS instrumentation amplifier for biomedical applications. A three opamp instrumentation amplifier have been designed by employing two simple op-amps at the two inputs and a folded cascode op-amp at the output. Both input and output stage op-amps are 2-stage. Most of the earlier designed opamp in literature uses same type of op-amp at the input and output stages of instrumentation amplifier. By using folded cascode two stage op-amp at the output, we have achieved significant improvement in gain and CMRR. Transistors sizing plays a vital role in achieving high gain and CMRR. To achieve a desirable gain, common mode rejection ratio and other performance metrics, selection of most appropriate op-amp circuit topologies & optimum transistor sizing was the main criteria for design of instrumentation amplifier for biomedical applications. The instrumentation amplifier is simulated using Cadence Spectre tool and layout is designed in Cadence Layout editor at 0.18µm CMOS technology. Each of the input opamp provides a gain and CMRR of 45dB and 72dB respectively. The output stage folded cascode amplifier provides a gain of 82dB and a CMRR of 92dB. The design achieves an overall gain and CMRR of 67dB and 92db respectively. The designed instrumentation amplifier consumes only 263µW of power suitable for biomedical signal processing applications. Copyright 2013 Institute of Advanced Engineering and Science. All rights reserved. Corresponding Author: Akshay Goel, Centre for development of advanced computing Mohali, Punjab Email: akshayece0753@gmail.com 1. INTRODUCTION Biomedical signal processing aims at fetching useful information from biomedical signals. Earlier the primary focus was on processing the biomedical signal to extract the original signal and remove noise. Detecting biomedical signals is a very challenging task as these signals normally consists of very weak amplitude in the order of few mv with almost equivalent noise signal levels. These signals also have a very low frequency range usually below 1KHZ. With the growth of microelectronics and embedded design more and more applications require very weak amplitude signal detection and measurement. These weak amplitude biomedical signals need to be suitably amplified along with rejection of noise. To overcome this problem instrumentation amplifiers are used to suppress any unwanted noise or the common mode signals that affects the original signal and also provide proper amplification to the desired signal. Due to this property of rejecting common mode noise and providing amplification, instrumentation amplifier is widely used in various applications like transducers, micro-electro mechanical systems and biomedical applications. Unlike op-amp, instrumentation amplifier is a closed loop gain block that has a differential block with two opposite inputs and a single ended output. The impedances of the two input terminals are balanced and have high values typically 10 9 ohm or greater. The major achievements of designed instrumentation amplifier as compared to the previous similar works in literature are increment of 22.7db gain, and low noise due to the Journal homepage: http://iaesjournal.com/online/index.php/ijece

517 ISSN: 2088-8708 addition of folded cascode stage at the output. Section II explains the proposed CMOS instrumentation amplifier design. Section III discusses the simulation results and comparison while Section IV describes the conclusion. 2. PROPOSED INSTRUMENTATION AMPLIFIER CIRCUIT DESIGN One of the main concerns while designing an instrumentation amplifier is the interferences generated in the form of common mode voltages [5]. To reduce these interferences and to improve the gain, we have designed a conventional three stage instrumentation amplifier by employing two simple op-amps at the input stage and a folded cascode op-amp at the output stage. Figure 1. shows the proposed design with symbols 1 and 2 showing the simple two stage op-amps and symbol 3 showing two stage folded cascode op-amp. Figure 1. Instrumentation Amplifier The advantages of three op-amp type of configuration are high gain, High Common Mode Rejection Ratio and easy gain control by changing value of only single resistance. A major drawback of this instrumentation amplifier is its larger chip area due to the use of different resistances. So, we have drawn layout of instrumentation amplifier with both on-chip and off-chip resistors. The op-amp blocks 1 & 2 shown in Fig.1 are connected in non-inverting configuration. The complete circuit of simple two stage op-amp [4] is shown in figure 2. Input stage comprises of a differential pair MN1 & MN2 and a current mirror load MP1 &MP2. The output stage comprises of transistors MP3 & MN5. Transistors MN3 & MN4 form the current sink for proper biasing of the circuit. Figure 2. Two Stage CMOS OP-AMP IJECE Vol. 3, No. 4, August 2013 : 516 523

IJECE ISSN: 2088-8708 518 For calculating the aspect ratio of various transistors methodology [3] is widely used for transistor sizing of the analog circuits. Apart from calculation of transistor sizing, methodology also helps in identifying device operating region. The mode of operation can be found according to the following equation (In / ) = (1) The ratio decreases as the operating point moves toward strong inversion. The main focus is on transistor sizing of differential pair MN1 & MN2 which senses the difference between the two input voltages. Biasing current to the differential pair is calculated by the transistors MN3 and MN4. Output current can be calculated by using the relation 2. /W/L (2) Where represents the total output current, is the drain current and W/L is the aspect ratio of the transistor. The Table 1 shows optimum calculated aspect ratios of various transistor of a two stage CMOS opamps used in blocks 1 & 2 respectively. Table 1. Input stage op-amp 1 & 2 transistor sizing TRANSISTORS ASPECT RATIO (W/L) MN1, MN2 3µ/1µ MN3 9µ/1µ MN5 10µ/1µ MN4 4.5µ/1µ MP1, MP2 15µ/1µ MP3 30µ/1µ To enhance the gain of instrumentation amplifier, we have added the folded cascode stage [9] at the output as block 3 shown in Figure 3. The folded cascode op-amp circuit is shown in Figure 3. Figure 3. Two Stage Folded Cascode CMOS OP-AMP A Novel Low Noise High Gain CMOS Instrumentation Amplifier for Biomedical Applications (Akshay Goel)

519 ISSN: 2088-8708 The Table 2 shows optimum calculated aspect ratios of various transistor of a two stage folded cascode CMOS op-amp used in block 3. Table 2. Output Folded Cascode Stage op-amp Transistor Sizing TRANSISTORS ASPECT RATIO(W/L) MN1,NM2 MN3,MN4 MN5,MN6 MP1,MP2 MP3.PM4 MP5,MP6 MN8 MN7 5µ/200n 3µ/200n 20µ/200n 20µ/200n 20µ/200n 50µ/200n 5µ/200n 10µ/200n As we know the input signal of a common-gate stage may be a current and also that in the commonsource arrangement a transistor converts a voltage signal to a current signal. The cascade of a CS stage and a CG stage is called a cascode topology, which provides many useful properties. The main advantage of using the folded cascode at the output is that the gain can be further increased without affecting the output swing. As shown in Fig 3 MN3 and MN4 act as a differential pair used to sense the input voltage difference. Special care has been exercised to operate the input differential pair in the saturation region not in the triode region. The operation in triode region causes non-linerity in the behavior of the folded cascode op-amp as non linear and also results in poor DC gain. In this proposed instrumentation amplifier we have set the values of the resistances according to gain requirement. For a good common mode rejection ratio and gain proper resistance matching should be done such that R3=R4 and R5=R6. The output voltage of the instrumentation amplifier is given by relation 3. The two input op amps 1 & 2 are used in buffer configuration so as to provide high input impedance followed by the third folded cascode two stage op-amps is used in differential configuration. The above configuration is capable of amplifying low amplitude signals in the order of few mv to operational range for further processing in several volts. With the use of active impedances not only temperature dependency but also power consumption is greatly reduced. To check the effects of on chip & off-chip resistors on the proposed instrumentation amplifier performance & area, the layouts of the proposed circuit have been generated with both on-chip and off-chip resistors. If on chip resistors are used than a much larger area is consumed by the instrumentation amplifier as very high value resistors are required and the gain is also fixed as the value of the resistors cannot be changed. But with off chip resistors a significant reduction in the area is achieved with benefit of variable gain as it can be adjusted by changing the value of the single resistor R1. The values of the resistors used in instrumentation amplifier are given in Table 3. These values can be adjusted according to the need of the amplification. (3) Table 3. Resistance Value for IA Resistors Resistance Value (Ω) R1, R2 100, 1K R3,R4 32K R5,R6 1K 3. SIMULATION RESULTS To achieve a desirable gain, common mode rejection ratio and other performance metrics, selection of most appropriate op-amp circuit topologies & optimum transistor sizing was the main criteria for design of low noise high gain CMOS instrumentation amplifier for biomedical applications. Layouts of the proposed instrumentation amplifier with on-chip & off-chip resistors have been generated. The schematic of proposed CMOS instrumentation amplifier is generated in Virtuoso Schematic Editor. Cadence Spectre is used for IJECE Vol. 3, No. 4, August 2013 : 516 523

IJECE ISSN: 2088-8708 520 circuit simulator using BSIM3v3 models in 180nm CMOS technology. The layouts of proposed CMOS instrumentation amplifier with on-chip resistor and without on-chip resistors are generated in Virtuoso Layout Editor. Figure 4 and 5 shows gain versus frequency plot in both magnitude and db of the proposed instrumentation amplifier design. Here we have applied a 5mV amplitude signal at the input. The output AC response has been plotted with frequency range from 10Hz to 100MHz.The fig. 5 shows the obtained gain of 67.7dB, 3-dB bandwidth of 1.1MHz and unity gain frequency of 91.33MHz. Figure 4. Gain Curve in Magnitude Figure 5. Gain Curve in Decibels To perform noise analysis, a 50Ω resistor is connected at the output. The input/output referred noise graphs have been plotted in Cadence Spectre simulator. The output referred noise plot in Figure 6 gives a peak value of 1.75 pv/ Hz. Noise analysis has been performed with a 5mV signal applied at the input. The input referred noise plot in Figure 7 gives a peak value of 89nV/ Hz. A Novel Low Noise High Gain CMOS Instrumentation Amplifier for Biomedical Applications (Akshay Goel)

521 ISSN: 2088-8708 Figure 6. Output Referred Noise Voltage Figure 7. Input Referred Noise Voltage Figure 8 shows layout of designed instrumentation amplifier with on-chip resistors. The layout of area of the chip is 1.49x10 6 μm 2. The area consumed by the on-chip resistors R1, R2, R3, R4, R5, and R6 is 1.48x10 6 μm 2 of the total area. Figure 8. Layout with On-Chip Resistors IJECE Vol. 3, No. 4, August 2013 : 516 523

IJECE ISSN: 2088-8708 522 Figure 9 shows layout of designed instrumentation amplifier without on-chip resistors. The layout of area of the chip is 1.28x10 4 μm 2. Figure 9. Layout without On-Chip Resistors Layout of with and without on chip resistors instrumentation amplifier shows that the area consumed by without on-chip resistor is much lesser than with on-chip resistors. Moreover, with off-chip resistors the gain can be easily controlled externally. Table 4 shows a comparison of the proposed instrumentation amplifier with some related existing instrumentation amplifier designs in literature. The critical performance metrics taken into consideration for comparison are gain, power dissipation, unity gain frequency, input referred noise, output referred noise and common mode rejection ratio for proposed design and other related references. The comparison explicitly shows an effective increment of 22.7 db in gain, 17 db in CMRR and 3.88 (nv/ Hz) in output referred noise. Table 4. Comparison With Previous Designs PARAMETER THIS WORK [5] [1] [2] [3] Technology (μm) 0.18 0.5 0.5 0.18 0.8 Gain (db) 67.7 45 19.9 19.6 40 CMRR(dB) 92 75 >11 92 N/A Unity Gain Frequency 91.33M N/A N/A N/A N/A GB(Hz) 1.75 1.1M 20K 100M N/A Power Dissipation (µw) Output Referred Noise (pv/ Hz) Input Referred Noise (V/ Hz) 263 283 N/A N/A 122 1.75 5.63 N/A N/A N/A 89n 22n 175n 22.8 m N/A 4. CONCLUSION A novel high gain and low noise CMOS instrumentation amplifier design for biomedical applications has been presented. The post and the pre layout simulations show that the design achieves comparatively high gain and low noise as compared to the previous designs existing in literature. The design also ensures low output referred noise one of main requirement of biomedical signal processing. REFERENCES [1] Chih-Jen Yen, Wen-Yaw Chung and Mely Chen Chi. Micro-Power Low Offset Instrumentation Amplifier IC Design For Bio-Medical System Applications. IEEE Transactions On Circuits And Systems-I: Regular Papers. 2004; 51(4): 691-699. [2] Yasin FM, Yap MT, and Reaz MBI. CMOS Instrumentation Amplifier with Offset Cancellation Circuitry for Biomedical Applications. In Proc. of 5th WSEAS. 2006, Spain, pp 168-171. A Novel Low Noise High Gain CMOS Instrumentation Amplifier for Biomedical Applications (Akshay Goel)

523 ISSN: 2088-8708 [3] Silveira F, Flandre D, Jespers PGA. A gm/id based methodology for the design of CMOS analog circuit and its application to the synthesis of a silicon-ion- insulator micropower OTA. IEEE Journal of Solid State Circuits. 1996; 31: 1314-1319. [4] Manish Goswami. and Smriti Khanna. DC Suppressed High Gain Active CMOS Instrumentation Amplifier for Biomedical Application. In Proc. IEEE Inter. Symp. on Circuits and Systems (ISCAS 04). vol 4, pp. 9-12. [5] MSJ Steyaert, WMC Sansen and C Zhongyuan. A micropower low-noise monolithic instrumentation amplifier for medical purposes. IEEE Journal of Solid State Circuits. 22(6), 1987. [6] MA Smither, DR Pugh and LM Woolard. C.M.R.R. analysis of the 3-op-amp instrumentation amplifier. Electronic Letters. 1997; 13(20): 586. [7] Ramakant A Gayakward. Op-Amps and Linear Integrated Circuits: Pearson Education. [8] Behzad Razavi, Design of Analog CMOS Integrated Circuits: Tata McGraw Hill Education. [9] Ananth RS and Lee EK. Design of a low power implantable electromyogram amplifier. In Proc. IEEE Inter. Symp. on Circuits and Systems (ISCAS 04). vol 4, pp. 9-12. BIOGRAPHIES OF AUTHORS Akshay Goel has done his bachelor of technology degree in Electronics and Communication Engineering from Punjab Technical University Jalandhar in year 2011 and currently pursuing his Master of Technology degree in VLSI Design from CDAC, Mohali. His areas of interests are Analog and Digital VLSI Design, Embedded technology and low power CMOS circuit design. His email-id is akshayece0753@gmail.com Gurmohan Singh has obtained his Bachelor of Technology degree in Electronics & Communication Engineering from Giani Zail Singh College of Engineering & Technology, Bathinda and Master of Technology degree in Microelectronics from Panjab University, Chandigarh in 2001 and 2005 respectively. He is working as a Senior Engineer in Digital Electronics & Comm. Division at C-DAC, Mohali. He is involved in many technological research areas in the field of VLSI Design. He has more than 7 years of experience. He has also worked in Bharat Sanchar Nigam Limited for 3 years. He was involved in installation, configuration & Testing of latest communication equipments like DWDM, GPON OLTEs, LAN Switches and MADMs etc. His major research interests are analog integrated circuits design, low power CMOS circuit design techniques and VLSI Testing. His email-id is gurmohan@cdac.in IJECE Vol. 3, No. 4, August 2013 : 516 523