PIECAL Model 541 Frequency Calibrator with Totalizer Operating Instructions

Similar documents
PIECAL 211 Automated Universal RTD Calibrator Operating Instructions. Product Description. Practical Instrument Electronics

PIECAL 311 Automated Universal RTD Calibrator Operating Instructions. Product Description. Practical Instrument Electronics

PIECAL 211 Automated RTD Calibrator Operating Instructions. Product Description. Practical Instrument Electronics

PIECAL 311 Automated Universal RTD Calibrator Operating Instructions. Product Description

Frequency Calibrator with Totalizer

Frequency Calibrator with Totalizer

Model 532 ma/v Loop Calibrator with Loop Diagnostic Operating Instructions

Dual T/C & RTD Calibrator With Auto Stepping

Meriam Model M ma Loop Calibrator Operating Instructions

User s Guide CL535. Frequency Calibrator with Totalizer. Shop online at. omega.com For latest product manuals: omegamanual.

PIECAL 820 Multifunction Process Calibrator

PIE 525B. Automated Thermocouple & RTD Calibrator Operating Instructions. Practical Instrument Electronics

FREQUENCY CALIBRATOR WITH TOTALIZER MODEL 942 GENERAL DESCRIPTION ALTEK INDUSTRIES CORP A TRANSMATION COMPANY

Practical Instrument Electronics PIECAL 820

Automated Thermocouple & RTD Calibrator

Practical Instrument Electronics

Process Calibrator. TechChek 820

Model 422 Diagnostic Thermocouple & Milliamp Calibrator Operating Instructions

Practical Instrument Electronics. PIE 311 Diagnostic RTD & Milliamp Calibrator. Best RTD & Milliamp Calibrator on the market!

Milliamp Calibrator. Model 434. General description. Calibrate with laboratory accuracy. All 4 to 20 ma loop functions

PIE 525. Diagnostic Thermocouple RTD & Milliamp Calibrator Operating Instructions. Practical Instrument Electronics

941/942 Field Calibration C Procedure Created by: D.NEFF Date: 3JUN94 Sheet 1 of 5

CALIBRATORS. Calibrate and Troubleshoot Control Loops and Process Instruments. Multifunction Calibrators Find hidden problems in loops & transmitters

PIECAL 820 FIELD CALIBRATION PROCEDURE

Practical Instrument Electronics

User s Guide CL510A. RTD Simulator. Shop online at. omega.com. For latest product manuals: omegamanual.info CL510A-1 CL510A-10

PIE 820-ELITE Multifunction Diagnostic Process Calibrator

Model BPG-1P Biphasic Pulse Generator

34134A AC/DC DMM Current Probe. User s Guide. Publication number April 2009

True RMS AC / DC Power Clamp Meter Model

Resonance indicator LEDº. » Transducer connection. Button B º. » LCD display. Power/fine frequency º. frequency set

Model 7000 Low Noise Differential Preamplifier

Property of Ramsey Electronics, Inc. Do not reproduce or distribute.

INSTRUMENTS, INC. Model 2960AX Disciplined Quartz Frequency Standard 2960AX. Section Page Contents

Sound Meter Type

AC/DC Current Oscilloscope Probe Model SL261

PM 6669 High-Precision Frequency Counter Specifications

JTY2 PULSE DURATION RECEIVER BEFORE USE... COMPONENT IDENTIFICATION

User s Guide. Combination Photo-Tachometer Stroboscope Model

OPERATION MANUAL LDM-1000

DIGITAL SOUND LEVEL DSM8930 METER USER S MANUAL. Please read this manual carefully and thoroughly before using this product.

GE Infrastructure Sensing. Druck DPI 841/842. Frequency calibrator and Frequency loop calibrator User manual - K395

Model Hz to 10MHz Precision Phasemeter. Operating Manual

2001A. 200KHz Function Generator Instruction Manual. 99 Washington Street Melrose, MA Phone Toll Free

Electrical Multimeter

Introduction. Specifications. Features. Controls. Model 103

AC/DC Current Probes Model MR411 Model MR521

User s Guide CL514-PLUS. Automated Universal RTD Calibrator. Shop online at

Multi-Band Wireless Power/ VSWR Meter. PM-6000 User s Manual

User's Manual. Model f/1f Power Clamp-on Meter

OPERATION & SERVICE MANUAL FOR FC 110 AC POWER SOURCE

INSTRUCTION MANUAL. Model Autoranging DMM ProbeMeter TM. Measures voltage, resistance, frequency, capacitance, temperature, and duty cycle.

LINCO MEASUREMENT MODEL CP-2B MASTER METER PROVER COUNTER INSTRUCTION MANUAL

Series 500. Owner s Manual. Analog Transmitters by Data Industrial. Data Industrial. Data Industrial 2/95 PN 72806

User s Manual. MiniTec TM Series. Model MN26 (Model MN26T includes temperature probe) Mini Autoranging MultiMeter

TRMS LEAKAGE CURRENT CLAMP-ON METER 565

Model Digital Light Meter

9200 Series, 300 MHz Programmable Pulse Generator

Power Meter. Measurement Guide. for Anritsu RF and Microwave Handheld Instruments BTS Master Site Master Spectrum Master Cell Master

DS1000B Series Digital Oscilloscopes

CIRCUIT-TEST ELECTRONICS

Dual Channel Function/Arbitrary Waveform Generators 4050 Series

Digi-Stop. Installation & Operation

HHMA2 DC / TRUE RMS AC NON-CONTACT MILLIAMMETER

99 Washington Street Melrose, MA Fax TestEquipmentDepot.com # # AAC Clamp Meter. Instruction Manual

TRANSDUCER IN-LINE AMPLIFIER

RIGOL Data Sheet. DG1000 Series Dual-Channel Function/Arbitrary Waveform Generator. Product Overview. Main Features. Applications. Easy to Use Design

12B/18 MultiMeter Instruction Sheet

Dual Channel Function/Arbitrary Waveform Generators 4050B Series

User s Guide. Digital AC/DC Clamp Meter Model 38394

Digi-Stop. User Manual for: Digi-Stop. Digital Readout Firmware version d & Higher

User s Guide. 400A AC/DC Clamp Meter. Model MA220

PIE mA/10-50mA Process Calibrator. Practical Instrument Electronics

Model OI-6940 Notis Quad 4-Gas Sensor Assembly

Dual Channel Function/Arbitrary Waveform Generators 4050 Series

SAFE USE &OPERATIONS MANUAL

CT-2 and CT-3 Channel Taggers OPERATION MANUAL

User s Manual. Ground Resistance Clamp On Tester MODEL

ProScale. Compact LCD Readout. Installation & Operation

User's Manual. Clamp-on Power Datalogger. Model

ProScale Compact LCD Readout. Operation. Firmware version C 2.xxx and higher

INSTRUMENTS, INC. Models 2960AR and 2965AR Disciplined Rubidium Frequency Standards. Section Page Contents

Signal Forge 2500M Frequency Expansion Module. 1.5 GHz to 2.6 GHz. User Manual

LUDLUM MODEL MODEL AND MODEL GAMMA SCINTILLATORS. June 2017

Non-contact Forehead IR Thermometer

Precision Frequency To Voltage Converter Card

TT-2Dx USER'S MANUAL

DDS Function Generator

User Guide. Digital AC/DC Clamp Meter Model 38394

Analog Arts SF990 SF880 SF830 Product Specifications

Arbitrary/Function Waveform Generators 4075B Series

POWER SERIES Plus. Volt / Amp / Hertz. Digital Switchboard Meter. User s Manual IM2493VAF-3

2013 MFJ ENTERPRISES, INC.

SMART TWEEZERS Model ST-1 User s Manual Version 1.0

TRANSDUCER IN-LINE AMPLIFIER

User s Guide. MultiView Series Digital MultiMeters Models: MV110 MV120 MV130

200Amp AC Clamp Meter + NCV Model MA250

DM-46 Instruction Manual

SA-027HQ 365-Day Annual Timer

Analog Arts SF900 SF650 SF610 Product Specifications

Transcription:

PIECAL Model 541 Frequency Calibrator with Totalizer Operating Instructions Easy to use With the Model 541 you can check & calibrate all your frequency instruments and measure flow sensors. Take it into the shop, plant or field Carry it without worry - it comes with a carrying case and fits into your toolbox. Accurate to ±0.005% of range. Calibrate in Hz, khz, CPM and CPH 0.001 to 20.000 khz 0.1 to 2000.0 Hz 0.01 Hz to 200.00 Hz 0.1 to 2000.0 CPM (Counts-per-Minute) 1 to 20000 CPH (Counts-per-Hour) Quickly output the frequency you need Simulate vibration pickups and variable speed drives with sine wave outputs and flow meters and magnetic pickups with square wave outputs. If you calibrate positive displacement flow meters, Watt-hour meters or slow rated integrators select frequencies as slow as 1 CPH (0.0002777 Hz). Easily set any value quickly to within 0.01 Hz with the two speed adjustable EZ-Dial plus store any three frequencies for instant recall with the EZ-Check switch. Verify optical pickups with the GATE TIME LED that is synchronized with the output pulses. Match your instrument signal levels Choose zero based or zero crossing square or sine waves output from 0.1 V to 12 V peak-to-peak. Measure from 0.1 V to 120 V peak with X1 or X10 attenuation. Turn the dial to adjust the output amplitude to the level required. Read signals from flow meters Measure flow meter, vibration, parts counter and other process frequency signals from 0.1 V to 120 V. GATE TIME LED flashes with received pulses indicating when the proper input adjustment level is achieved. Calibrate totalizers without a stopwatch Automatically output from 1 to 99999 pulses in 1 to 100 minutes. Easier and more accurate than waiting around with the stopwatch to stop generating pulses at the correct time. Count input pulses from 1 to 100 minutes to verify totalizer input signals. Copyright 2011 All rights reserved 541-9002 - Rev D 13 July 2011 Page 1

Basic Operation HI 20000 SOURCE CPH 1-20000 r MENU Key Press and release the menu key to enable the configuration menu. This allows you to choose the settings for AUTO OFF, X1/X10, Zero Based/ Zero Crossing, EZ-Check HI/LO Readings, SINE/SQUARE Wave output and reset to BASIC CONFIGURATION. t READ/GATE TIME Key Press and release READ/GATE TIME key to step through the read ranges. Read CPH 1-20000 Read CPM 0.1 2000.0 Read HZ 0.01 200.00 Read HZ 0.1 2000.0 Read KHZ 0.001 20.000 Totalizer q EZ-Check Switch/EZ-Step Push-button For Source - Slide the switch to select from three user stored values for the desired calibration points. The HI, DIAL and LO values can easily be changed to suit the calibration requirements. Press the push-button like a stopwatch to start or stop the totalizer output. For Read - Slide the switch to the DIAL position. The Model 541 will begin making readings. Slide the switch to high and you will get the highest point read and then slide the switch to the low position and you will get the lowest range. Press the push-button like a stopwatch to start or stop the totalizer counting pulses. Press and hold the READ/GATE TIME key to enable time adjustment for Source & Read Totalize mode. Turn the EZ-Dial knob u to adjust the totalize time from 1 to 100 minutes. Press and release the SOURCE/AMPLITUDE or STORE/CLEAR button y STORE/CLEAR Key For Source - press the STORE/CLEAR key to save the calibration values. The display will flash STORED to confirm. For Read - press STORE/CLEAR to clear the values saved in the EZ-Check HI and LO positions. The display will flash CLEARED to confirm. w SOURCE/AMPLITUDE Key Press and release the SOURCE/AMPLITUDE Key to step through the source ranges. Source CPH 1-20000 Source CPM 0.1 2000.0 Source HZ 0.01 200.00 Source HZ 0.1 2000.0 Source KHZ 0.001 20.000 Totalizer Press and hold SOURCE/AMPLITUDE key to enable amplitude adjustment. Turn the EZ-Dial knob u to adjust the amplitude voltage from 0.1-12.0 Vp. Press and release the SOURCE/AMPLITUDE or STORE/CLEAR button e ON / OFF Key Press the ON/OFF key to turn the Model 541 on or off. u EZ-Dial Knob For Source - turn the EZ-Dial knob to change the output frequency in increments of one least significant digit. Push down and turn for faster dialing with increments of 100 digits. For Read - turn the EZ-Dial knob to adjust the trigger level. CHANGING BATTERY Low battery is indicated by BAT on the display. Approximately one to four hours of typical operation remain before the 541 will automatically turn off. To change the battery remove the battery cover from the back of the unit. Replace with a single 9V battery. Note: Alkaline battery is supplied and recommended for maximum battery life & performance. Page 2

Configure the Calibrator Press and release the ON/OFF Key w to turn the calibrator on then press and release the MENU Key r to enable the configuration menu. Turn the EZ-Dial Knob to scroll through the list of configuration options. Press the EZ-Dial Knob to change configuration items. >AUTO OFF ON X1/X10 X10 0 XING/BASED BASED EZ-CHECK ON SINE/SQ SQ BASIC CONFIGURATION Default Settings Shown AUTO OFF - pressing the knob will toggle between ON and OFF. If AUTO OFF is ON, the unit will turn off after 30 minutes of inactivity to save battery life. If AUTO OFF is OFF the unit will stay on until the ON/OFF key is pressed. Set If you are using the Totalizer functions for over 30 minutes you should set AUTO OFF to OFF. X1/X10 - pressing the knob will toggle between X1 and X10 attenuation while reading frequency. Choose X1 when you are measuring signals with amplitudes between 0.1 and 12 V peak and X10 when measuring between signals with amplitudes between 12 and 120 V peak. Configuration 0 XING/BASED - pressing the knob will toggle between zero crossing and zero based output waveforms. Select zero crossing when the instruments you are calibrating require a waveform that has both positive and negative peaks that are centered around zero. Select zero based when the instruments you are calibrating require only positive signals. Note: To simulate negative only signals select zero based and connect the red lead to the ground and the black lead to the input connection of the instrument being calibrated. EZ-CHECK - pressing the knob will toggle between ON and OFF. Select ON to have the highest and lowest measured frequencies recalled by moving the EZ-Check switch on the side of the calibrator to HI or LO. Select OFF and the currently measured freqency will be displayed regardless of the position of the EZ-Check switch. SINE/SQ - pressing the knob will toggle between SINE and SQ. Select SINE when the instruments you are calibrating require a sine wave and SQ when the instruments you are calibrating require a square wave or pulse input. BASIC CONFIGURATION - pressing the knob will restore the 541 to the basic configuration with factory defaults as shown in the menu diagram. Output values stored in the EZ-Check HI and LO positions are not changed. Note: All settings are remembered even with the power off. Removing the batteries resets the configuration to factory defaults (Basic Configuration) and resets the EZ-Check HI and LO outputs to factory values. Page 3

Calibrating Frequency Instruments SOURCE Choose this function to provide a frequency signal into flow meters or any input devices that measure frequency signals. HI 20000 SOURCE CPH 1-20000 1) Disconnect the input signal from the device to be calibrated. 2) Press and release the SOURCE/ AMPLITUDE Key until the desired frequency range is displayed w. 3) Press and hold SOURCE/AMPLITUDE key to enable amplitude adjustment. Turn the EZ-Dial knob u to adjust the amplitude voltage from 0.1-12.0 Vp. Press and release the SOURCE/AMPLITUDE or STORE/CLEAR button 4) Connect the 541 to the input of the device to be calibrated. Turn the EZ-Dial knob u to change the output frequency in increments of one least significant digit. Push down and turn for faster dialing with increments of 100 digits. Move the EZ-Check slide switch q on the side of the calibrator up to recall the HI output value and down to select the LO output value. STORING HI and LO EZ-Check Outputs Storing EZ-Check Outputs 1) Store your high (SPAN) output frequency by moving the EZ-Check switch q to the HI position and turn the EZ-Dial knob u until the desired frequency is on the display. Press the STORE/ CLEAR Key and STORED appears on the display. 2) Store your low (ZERO) output frequency by moving the EZ-Check switch q to the LO position and turn the EZ-Dial knob u until the desired frequency is on the display. Press the STORE/ CLEAR Key and STORED appears on the display. 3) Instantly output your SPAN and ZERO frequency outputs by moving the EZ-Check switch q between HI and LO. You may also select any third frequency output (such as mid-range) using the middle position on the EZ-Check switch. Calibrating Optical Pickups Position the optical pickup to be verified over the green GATE TIME LED. The LED is synchronized to the pulse output of the 541. Follow the Calibrating Frequency Instruments section to adjust the output frequency. Page 4

Calibrating Totalizers Choose this function to provide a number of pulses for a set time into parts counters or totalizers. 1) Disconnect the input signal from the device to be calibrated. 2) Press and release the SOURCE/AMPLITUDE Key until TOTALIZER appears on the display. 3) Press and hold the READ/GATE TIME key to enable time adjustment. Turn the EZ-Dial knob u to adjust the totalize time from 1 to 100 minutes. Press and release the READ/GATE TIME or STORE/CLEAR button 4) Turn the EZ-Dial knob u to select the number of pulses you would like to output with increments of 1 count. Push down and turn for faster dialing with increments of 100 counts. 5) Connect the 541 to the input of the device to be calibrated. 5) Press the EZ-Step push-button q like a stop watch to start outputting pulses. The word RUN will appear on the display and the 541 display increments with each pulse while the totalizer is running. When the selected number of seconds has elapsed the word STOP appears on the display and the total number of pulses sourced during the time appears on the display. Press the EZ-Step push-button q like a stop watch to stop or re-start totalizing. Note: If you are using the Totalizer functions for over 30 minutes you should set AUTO OFF to OFF (see Configure the Calibrator on page 3). CPM/CPH CONVERSIONS To Convert From: To: Divide By: To Convert From: To: Multiply By: CPM Hz 60 Hz CPM 60 CPH Hz 3600 Hz CPH 3600 Page 5

READ Measuring Frequencies Choose this function to measure frequency from flow, vibration sensors or frequency outputs. 1) Disconnect the frequency sensor from any other device. 2) Press and release READ/GATE TIME Key t to step through the read ranges. 3) Connect the 541 to the output of the device to be calibrated. 4) Turn the EZ-Dial knob u to adjust the trigger level of the 541. The LED will light when the 541 locks onto the input signal indicating that it is properly detecting the signal. HI 20000 SOURCE CPH 1-20000 Totalize (Count Pulses) Choose this function to verify that parts counters or totalizers are working correctly. 1) Disconnect the frequency sensor from any other device. 2) Press and release READ/GATE TIME key t to step through the read ranges until TOTALIZER appears on the display. 3) Press and hold the READ/GATE TIME key to enable time adjustment. Turn the EZ-Dial knob u to adjust the totalize time from 1 to 100 minutes. Press and release the READ/GATE TIME or STORE/CLEAR button 3) Connect the 541 to the output of the device to be calibrated. 4) Turn the EZ-Dial knob u to adjust the trigger level of the 541. The LED will light when the 541 locks onto the input signal indicating that it is properly detecting the signal. 5) Press the EZ-Step push-button q like a stop watch to start totalizing. The word RUN will appear on the display and the 541 displays increments with each pulse while the totalizer is running. When the selected number of seconds has elapsed the word STOP appears on the display and the total number of pulses counted during the time appears on the display. Press the EZ-Step push-button q like a stop watch to stop or re-start totalizing. Note: If you are using the Totalizer functions for over 30 minutes you should set AUTO OFF to OFF (see Configure the Calibrator on page 3). Page 6

General Accuracy Ranges Model 541 Specifications (Unless otherwise indicated all specifications are rated from a nominal 23 C, 70 % RH for 1 year from calibration) Operating Temperature Range Temperature Drift Relative Humidity Range Size Weight Batteries & Battery Life Low Battery Protection against mis-connection Display Normal Mode/Common Mode Rejection ±0.005% of Range 0.001 to 20.000 khz 0.1 to 2000.0 Hz 0.01 to 200.00 Hz 0.1 to 2000 CPM (Counts-per-Minute) 1 to 20000 CPH (Counts-per-Hour -25 to 60 C (-10 to 140 F) ± 0.01% of span outside of 23 C ±10 C (73 C ±18 F) 10 % RH 90 % (0 to 35 C), Non-condensing 10 % RH 70 % (35 to 60 C), Non-condensing 7.00 X 3.30 X 2.21 inches (177.8 X 83.8 X 56.1 mm 12.0 oz (340 grams) (including battery) One Alkaline 9V (6LR61), 45 Hours Low battery indication with nominal 1 hour of life left Over-voltage protection to 120 Vrms rated for 30 seconds or 240 Vrms for 15 seconds High contrast graphic liquid crystal display. 50/60 Hz, 50 db/ 50/60 Hz, 90 db Read Frequency Input Impedance > 1 Megohms + 100pF Signal Attenuation Adjustment Range Adjustable trigger level with X1 (0.1 V peak to 12.0 V peak) and X10 (1 V Peak to 120 V peak) Note on reading slow sine waves High signal noise and low slew rate (Volts-per-second) will affect reading uncertainty Source Frequency Output Impedance Source Current Amplitude Adjustment Range & Accuracy Square Wave (Pulse) Rise/Fall Time Frequency Jitter Duty Cycle Sine Wave Offset and Zero Crossing Symmetry < 25 Ohms > 6 ma at 12 V peak-to-peak, 20 khz 0.1 to 12.0 V peak ±10% of setting Selectable Zero Based or Zero Crossing <0.0001% of output V peak per second <0.5 Least Significant Digit of full scale value 50% ±2% <±10% of V peak output amplitude setting Page 7

Accessories Included: 9V Alkaline battery, Certificate of Calibration Large Carrying Case with PIE Logo Part No. 020-0204 Additional Information This product is calibrated on equipment traceable to NIST and includes a Certificate of Calibration. Test Data is available for an additional charge. recommends a calibration interval of one year. Contact your local representative for recalibration and repair services. Warranty Our equipment is warranted against defective material and workmanship (excluding batteries) for a period of three years from the date of shipment. Claims under warranty can be made by returning the equipment prepaid to our factory. The equipment will be repaired, replaced or adjusted at our option. The liability of (PIE) is restricted to that given under our warranty. No responsibility is accepted for damage, loss or other expense incurred through sale or use of our equipment. Under no condition shall, Inc. be liable for any special, incidental or consequential damage. Page 8