Computed Radiography

Similar documents
Film Replacement in Radiographic Weld Inspection The New ISO Standard

Minimum Requirements for Digital Radiography Equipment and Measurement Procedures by Different Industries and Standard Organizations

ISO INTERNATIONAL STANDARD

Dürr NDT GmbH & Co. KG Höpfigheimer Straße 22 D Bietigheim-Bissingen Germany. Contract No. BAM ZBA Dürr

Dürr NDT GmbH & Co. KG Höpfigheimer Straße 22 D Bietigheim-Bissingen Germany. Contract No. BAM ZBA Dürr

Essential Parameters for the Visibility of IQIs and Small Indications in Digital Radiography

» There are many possible techniques for making good radiographs. » Following a defined procedure will simplify technique

A COMPARATIVE STUDY ON THE PERFORMANCE OF DIGITAL DETECTOR SYSTEMS FOR HIGH ENERGY APPLICATIONS

Moving from film to digital: A study of digital x-ray benefits, challenges and best practices

Pre Conference Workshop on Advanced Digital imaging

Experiences of users in Digital Radiography

Strategies for Film Replacement in Radiography - a comparative study -

The HOIS recommended practice for in-service computed radiography of pipes. Dr Stephen F Burch, WCNDT 2012, Durban, SA. April 2012

Minimum Requirements for Digital Radiography Equipment and Measurement Procedures by Different Industries and Standard Organizations

Essential Parameters and Conditions for Optimum Image Quality in Digital Radiology

Digital Radiography : Flat Panel

ON THE WAY TO DIGITAL RADIOGRAPHY

APPLICATION OF THE DIGITAL RADIOGRAPHY IN WELD INSPECTION OF GAS AND OIL PIPELINES

Investigation of the film Kodak MX 125 in Agfa NDT S eco processing

ISO INTERNATIONAL STANDARD. Non-destructive testing of welds Radiographic testing Part 2: X- and gamma-ray techniques with digital detectors

Performance evaluation of a photon counting detector for high energy NDT applications

Determination of Optimum X-Ray Tube Output Parameters kv and ma for Digital Radiography Testing of Welded Tubes

Digital Radiography for the Inspection of Small Defects

Migrating from traditional to Digital Radiography in Aerospace

Equivalent Penetrameter Sensitivity (EPS) for Performance Evaluation of Computed Radiography Systems Muzibur Khan * and Mike Brothers

The HOIS recommended practice for in-service computed radiography of pipes

5th Pan American Conference for NDT 2-6 October 2011, Cancun, Mexico

Fig.2: Scanner VistaScan for image plates

Preliminary Modulation Transfer Function Study on Amorphous Silicon Flat Panel System for Industrial Digital Radiography

IQI-Sensitivity and Applications of Flat Panel Detectors and X-Ray Image Intensifiers A Comparison

DOSE REDUCTION BY USE OF DIGITAL X-RAY DETECTORS. Bernhard REDMER

Investigation of the film FUJIFILM IX 80 in TETENAL machine processing

Projection Radiography. application on stream

PERFORMANCE CHARACTERIZATION OF AMORPHOUS SILICON DIGITAL DETECTOR ARRAYS FOR GAMMA RADIOGRAPHY

THE CR SPECIALISTS HD-CR CR 35 NDT

Digital Radiographic Inspection replacing traditional RT and 3D RT Development

ScienceDirect. Evaluation of Pipe Wall Thickness Based on Contrast Measurement using Computed Radiography (CR)

Digital Radiology with Photon Counting Detectors

ISO INTERNATIONAL STANDARD. Non-destructive testing of welds Radiographic testing Part 1: X- and gamma-ray techniques with film

Amorphous Selenium Direct Radiography for Industrial Imaging

Recommended Training Curriculum For Digital Radiography Personnel (Level II)

Gamex CR 2.0 Program description and operating manual

COMPUTED RADIOGRAPHY THE STATE OF THE ART

PRACTICAL CONSIDERATIONS AND EFFECTS OF METALLIC SCREEN FLUORESCENCE AND BACKSCATTER CONTROL IN GAMMA COMPUTED RADIOGRAPHY

Digital Detector Array Image Quality for Various GOS Scintillators

Digital Radiography Systems Techniques and Performance Evaluation for Space Applications

Conversion to Digital Radiography from Film Radiography

DIGITAL RADIOGRAPHY. Digital radiography is a film-less technology used to record radiographic images.

10/26/2015. Study Harder

NON-DESTRUCTIVE EVALUATION UTILIZING IMAGING PLATES FOR FIELD RADIOGRAPHY APPLICATIONS

Introduction. Chapter 16 Diagnostic Radiology. Primary radiological image. Primary radiological image

10/3/2012. Study Harder

AUTOMATED AND QUANTITATIVE METHOD FOR QUALITY ASSURANCE OF DIGITAL RADIOGRAPHY IMAGING SYSTEMS

Acquisition, Processing and Display

Photon Counting and Energy Discriminating X-Ray Detectors - Benefits and Applications

A Modified Detectability Criterion for Conventional Radiography Simulation

Unit thickness. Unit area. σ = NΔX = ΔI / I 0

COMPUTED RADIOGRAPHY CHAPTER 4 EFFECTIVE USE OF CR

17th World Conference on Nondestructive Testing, Oct 2008, Shanghai, China

Spokane Industries Computed Radiography Implementation Experience. David Jolin Technical Sales

NEW POSSIBILITIES OF RADIATION CONTROL OF QUALITY OF WELDED JOINTS

A True Innovation in Non-Destructive Testing System FUJIFILM COMPUTED RADIOGRAPHY. Series 4 CR

COMPARATIVE STUDY BETWEEN COMPUTED RADIOGRAPHY AND CONVENTIONAL RADIOGRAPHY

Seminar 8. Radiology S8 1

Current technology in digital image production (CR/DR and other modalities) Jaroonroj Wongnil 25 Mar 2016

This document is a preview generated by EVS

Radiographic sensitivity improved by optimized high resolution X -ray detector design.

Ray Detection Digital Image Quality and Influential Factors

Ayaz Jhanorwala, Rishikesh Kumar, Satish Tilva. L & T MHPS Boilers Pvt Ltd., Hazira, Surat , India

Guidelines on Training, Examination and Certification in Digital Industrial Radiology Testing (RT-D)

DACON INSPECTION SERVICES. Computed RADIOGRAPHY

Research Article Digital Radiography Using Digital Detector Arrays Fulfills Critical Applications for Offshore Pipelines

Modelling Computed Radiography Detectors with a Cascaded Linear System Model

I. PERFORMANCE OF X-RAY PRODUCTION COMPONENTS FLUOROSCOPIC ACCEPTANCE TESTING: TEST PROCEDURES & PERFORMANCE CRITERIA

Part 3: Image quality classes

NAVSEA Requirements for Computed Radiography DRAFT

Nuclear Associates EZ CR-DIN Phantoms

TESTING FLAT-PANEL IMAGING SYSTEMS: What the Medical Physicist Needs to Know. JAMES A. TOMLINSON, M.S., D.A.B.R. Diagnostic Radiological Physicist

Hardware for High Energy Applications 30 October 2009

Tomographic 3D-Radiometry for the Visualisation and Measurement of the Defects of Girth Seams

Real Time Linear Array Imaging. Brian Caccamise

Examination of Pipe Welds by Image Plate Based Computed Radiography System

Radiographic testing: Increased detection sensitivity using optimum source to object distance

In-service inspections for primary coolant circuit components of light water reactors Part 6: Radiographic testing

A TUTORIAL FOR PERFORMING A RADIOGRAPHIC EXAMINATION

Setting up digital imaging department!

Mammography: Physics of Imaging

NDT Supply.com 7952 Nieman Road Lenexa, KS USA

Quality Control of Full Field Digital Mammography Units

HPX1-Plus. For Non-Destructive Testing. THE BENCHMARK IN COMPUTED RADIOGRAPHY.

X-ray Imaging. PHYS Lecture. Carlos Vinhais. Departamento de Física Instituto Superior de Engenharia do Porto

UNDERSTANDING THE CURRENT CAPABILITIES AND LIMITATIONS OF DIGITAL INDUSTRIAL RADIOGRAPHY

Modeling computed radiography with imaging plates

Gamex Lite 3.0 Program description and operating manual

Dose Reduction and Image Preservation After the Introduction of a 0.1 mm Cu Filter into the LODOX Statscan unit above 110 kvp

Digital Imaging Considerations Computed Radiography

Features and Weaknesses of Phantoms for CR/DR System Testing

Advancements In Digital Radiography: CR, DR, and DICONDE

Gas scintillation Glass GEM detector for high-resolution X-ray imaging and CT

SINCE2011 Singapore International NDT Conference & Exhibition, 3-4 November 2011

Transcription:

BAM Berlin Computed Radiography --INDE 2007, Kalpakkam, India -- Uwe Zscherpel, Uwe Ewert BAM Berlin, Division VIII.3 Requests Requests and and information information to: to: Dr. Dr. U. U. Zscherpel Zscherpel // BAM BAM e-mail e-mail uwez@bam.de uwez@bam.de 1

New Technologies and Film Replacement Imaging Plates Scanner for Imaging Plates CR: Mobile Orex Scanner Radiography with Imaging Plates (CR) Agfa Hard Copy Grayscale Printer CR:FujiFilm XG1 NDT 2

Exposure Imaging Plate Cycle Computed Radiography Re-usage Erasure Read-out Lumisys ACR2000 Laser high sensitivity medium resolution high dynamic range re-usable 3

Storage Principle of of Luminescence Imaging Plates BaFCl:Eu scintillator BaFBr:Eu storage phosphor 4

5

IP BAS 2000 Comparison of of step wedge exposures a) a) Imaging plate plate system system with with log. log. amplifier. The The grey grey level level is is proportional to to the the wall wall thickness. b) b) Digitized film. film. The The grey grey level level is is proportional to to the the light light intensity behind behind the the film. film. X-ray tube Comparison: a) Imaging plates (IP) and b) Film Film 6

Image Quality Parameters Radiographic Image Quality is the basis for all comparisons Digital Detector Subject Contrast Detector Contrast Noise Detector Unsharpness Projected Unsharpness CNR CNR - Contrast/Noise Ratio SNR - Signal/Noise Ratio SNR CNR = SNR(I Tot ) Δw µ/(1 + k) k = I Scatter / I Primary BSR Detector Contrast: Signal for linear detectors Basic Spatial Resolution (effektive pixel size) Detectors can be qualified and classified by its SNR and BSR values! 7

D-D 0 Signal/Noise Ratio Definition of Signal/Noise Ratio and Contrast/Noise Ratio Signal proportional to radiation dose! 2*Noise Signal 0 Contrast 8

Standards, Regulations Welding Casting Comparison of Systems Digital Industrial Radiology Film Replacement Standards EN 14784 for CR were developed to be comparable with EN 444, EN 584-1, EN 462-5! All film based standards require: Minimum optical Density (e.g. > 2.0) Maximum film system class (e.g. ASTM special) Maximum unsharpness (> 0.1 mm, FFD/FOD) 9

Film replacement: What is the Optical Density? Film reading from light box, D=2: Light source Film Sensor Light intensity measurement 6000 Luxmeter I o Optical Density = log (I o / I 1 ) The opt. density is a measure of light absorption by film No digital X-ray detector yields an optical density 60 Luxmeter I 1 There exist software on the market which provides wrong equivalent values for the optical density of CR or DR systems! 10

NDT film as as linear radiation detector (D-D 00 )) K: K: steel, 160 kv 11

Film system classes have different SNR values at at opt. density = 2 for different dose values (EN 584-1, ASTM E 1815) all NDT film systems of all vendors along one line! AGFA films 2002-2004 In mixed systems SNR film film (D-D 00 )) SNR Max Max (class C x x,, D<4.5) square root (dose / mgy) 12

Film System Classes are the Basis for CR Classification and all Digital Detectors Tab. 1: Overview about the film system classes in different standards and the corresponding SNR values and G 2 /σ D values. SNR = log(e) (G 2 /σ D ) for linear detectors only System class Minimum gradientnoise ratio at Signal to Noise Ratio World ISO 11699-1 T1 T2 Europe CEN 584-1 USA ASTM E1815-01 Japan K7627-97 D=2 above D 0 D=2 above D 0 G 2 /σ D SNR C1 Special 300 130 C2 T1 270 230! 117 100! C3 I 180 78 C4 T2 150 65 T3 C5 II T3 120 52 T4 C6 III T4 100 43 W-A W-A 135 W-B W-B 110 W-C W-C 80 EN 14784 Part 1: Classification of Systems 13

Classification by - SNR and - Spatial resolution e.g.: IP-3/200 CR System classes Table 1: IP-scanner system classes in dependence on the minimum SNR IP System classes System class CEN System class ISO System class ASTM IP 1 IP-AS Special 130 IP 2 IP-T 1 117 IP 3 78 IP 4 IP-T2 IP-AS 1 65 IP 5 IP-T3 IP-AS 2 52 IP 6 IP-T4 IP-AS 3 43 Minimum Signal-noise ratio New table 1 with 6 classes! NOTE:The SNR-values of table 1 are equivalent to EN 584-1, ISO 11699-1, ASTM E 1815-96 They are calculated by: SNR = log (e) (Gradient/Granularity) of table 1 of the corresponding standards and up-rounded. The calculation of measured SNR values shall be performed from dose proportional data. Part 1: Classification of Systems 14

FUJI XG-1 CR System Classification (BSR = 130 µm) CR problem: SNR limited by structure noise of imaging plates! SNR SNR 0.1mm Pb f&b versus SQRT(Dose/mGy) 180,00 160,00 140,00 y = -0,7114x 4 + 11,315x 3-63,635x 2 + 159x R 2 = 0,9974 120,00 100,00 80,00 60,00 40,00 20,00 01Pb f&b 130µm Polynomisch (01Pb f&b 130µm) 0,00 0,0 0,2 0,4 0,6 0,8 1,0 1,2 1,4 1,6 1,8 Wurzel (Dosis) 15

BAM 5 test weld, 8 mm Steel, 120 kv, different exposure time: Exp. Time 1x Exp. Time 4.3x Exp. Time 43x magnification CR-System: FujiFilm DynamiX (XG 1) 16

Basic differences Film Imaging Plate: Film systems are characterized by sensitivity (speed) and the SNR at D=2. The spatial resolution is high, rarely used and typically limited by geometrical unsharpness of exposure set-up (shortest exposure time!). Digital detectors have a higher dynamic range, the SNR depends on exposure. The spatial resolution is limited (digital file size!). 17

Status and Situation for CR: - 2005: Computed Radiology standards were completed and published in USA and Europe! CR Procedure CEN EN EN 14784-2 ASTM E 2033 - The CR procedure is different from the film radiography procedure - The optical impression of digital radiographic images is not different from film images in its structure (if not digitally processed, except brightness and contrast control) - RT-trained personal can interpret digital images in analogy to film - Digital images need a PC for presentation and may be altered by specialized image processing - A basic training in image processing is essential to avoid miss interpretation - Quantitative assessment of flaw sizes is improved by digital processing but the results may differ from those ones of film interpretation - Electronic reference catalogues may support correct image assessment 18

19

2005 20

Achievable Contrast Sensitivity and Exposure Time - The contrast sensitivity improves with the increasing exposure time (better SNR, weaker details becomes visible!) - The structural noise of the IPs limits the max. achievable SNR - Step hole, plate hole and wire image quality indicators measure the increase of CNR and improvement of contrast sensitivity - Unsharp imaging plates achieve a higher max. SNR and need less exposure time - The normalized SNR N (with BSR) is independent on the unsharpness, but depends on the efficiency and plate homogeneity - Only unsharp IPs for radiography with higher energy (about > 250 kev) yield a considerable reduction of exposure time in comparison to film. 21

Required Image Quality Indicators for CR See also ASTM E747-97 but different to EN 462 Description of Wire -IQI For For contrast only only!! Both IQI s are are required for for CR! For unsharpness only!! Wire IQI EN 462-1 Step-Hole IQI EN 462-2 Lead marks Plastic Duplex wire IQI / EN 462-5 / ASTM E 2002 22

CR System Selection European Standard EN 14784-2 23

EN 14784-2 24

FujiFilm NDT Correct IP-Type X-ray X-ray tube: tube: Seifert Seifert Isovolt Isovolt320 320 100 100 kv, kv, 1 ma, ma, 1 min, min, small small focus focus SDD SDD = 1000 1000 mm mm Agfa NDT Agfa Prototype (blue) 8 9 10 11 12 13 digitised grey values scanning: Agfa DPS, pixel distance: 28.2 µm (900 dpi) 8 9 10 11 12 13 double wire 0.32 0.25 0.20 0.16 0.13 0.10 mm unsharpness 25

High Definition CR Dürr Germany -- Important Advantages: - flexible imaging plates (IP) in direct contact with inner canister wall - arbitrary formats possible (length not limited, width max. 14 (35cm) - could even be tried with existing setup (colimator, rotating canister) - small Laser spot (8 µm) allows small BSR for high resolution plates Disadvantages: - No automated inspection - Manual handling for IP read-out - IP based on BaFBr have limited absorption efficiency scan direction Flexible IP (14 x17 ) on mobile CR scanner (Dürr) 50 k High resolution mode (BSR=40 µm with suitable Imaging plate) available, with thicker IP high Sensitivity achievable (BSR 100 200 µm) 26

High Definition CR Systems Systems available down to 12 µm pixel pitch and 40 µm basic spatial resolution weld quality HD CR 35 NDT: of Dürr, Germany HD-IP, light blue 27

BAM Certification 2006 28

Under Discussion: - Intermediate filters - Pre filters - Max tube voltage Improvement of CNR Lead screens may damage the IPs! EN 14784-2 29

Improvement of CNR Under Discussion: - Intermediate filters - Pre filters Draft - Max tube voltage (lower than for film!) Designation: E 2033 xx TABLE 1 Lead Screen and Filter Thickness Lead Thickness A KV Range Front Screen Maximum, in. Back Screen Minimum, in. Filter at Tube port > 40kV None. None. <2mm Be Tube Port 40 to 55kV None. None. 0.04 Alum. (1mm) 55 to 90kV None 0.001 (25µm) None - 0.001 (25µm) 0.04 Alum. (1mm) 90 to 120 kv 0.002 (50µm) 0.002 (50µm) 0.08 Alum. (2mm) 120 to 160kV 0.005 (125µm) 0.005 (125µm) 0.01 Copper (250µm) 160 to 200 kv 0.01 (0.25 mm) 0.01 (0.25 mm) 0.04 Copper (1mm) 200 to 320kV 0.02 (0.5mm) 0.02 (0.5mm) 0.118 Copper (3mm) 320kV-420kV 0.04 (1mm) 0.04 (1mm) 0.236 Copper (6mm) 1MeV to <? 0.04 (1mm)?0.04 (1mm) None Se 135 0.01 (0.25mm) 0.01 (0.25mm) None Ir 192 0.02 (0.5mm) 0.02 (0.5mm) None Co 60 0.04 (1mm) 0.04 (1mm) None Lead screens may damage the IPs! 30

CR Phantom - long term stability tests - EN 14784-1, ASTM E 2445 Evaluation of Phantom: spatial resolution (by duplex-wire method, optional converging line pairs) contrast (recognized contrast percentage of the material to examine) slipping (yes/no) jitter (yes/no) MTF (European standard only) shading (percentage at selected distance) SNR N (Intensity) near BAM-snail 20x400 window recommended 31

EN 14784-1 Appendix B: 32

Photograph: 35x43 cm² Protoype Available from: - Fluke Biomedical, Part/No. 07-605-2435, 14 x17 www.flukebiomedical.com/rms - CIT (8 x10 ) in U.K. 33

CR radiograph 90 90 kv, 2m, 4mAmin 34

Duplex wire IQI Pair 9 = 260 µm BAM-snail for test of source position 20% dip EN 462-5: unsharpness 0.26 mm BSR = 0.13 mm 35

BAM-snail for test of source position Converging line pair IQI 3.6 lp/mm Unsharpness = 280 µm Basic Spatial Resolution = 140 µm Problems arise arise for for high high resolution systems; aliasing aliasing effects effects 36

Linearity test - Check rulers - Slipping IP was detected by a line structure in the image. Scanner needs repair! 37

Profile for MTF - Jitter Test: no problem - MTF calculation from pofile: Zero Frequency drop indicates internal scatter MTF MTF 20 = 1.6 lp/mm Unsharpness = 0.62 mm (too high!) 38

SNR N (Intensity) near BAM-snail 20x400 window recommended 39

Documentation of Long Term Stability Tests 40

Summary: The practice for CR (EN 14784-2) is based on EN 444 and ISO 5579 for NDT film. Film can be substituted by IPs The exposure conditions have to be selected to achieve the required SNR and CNR 2 IQIs are required for CNR and BSR Tab. 4 in EN 14784-2 regulates the device selection, especially the required basic spatial resolution (BSR) The max tube voltage should be reduced in comparison to film application Long term stability can be checked by CR Phantom 41

END 42