CHAPTER 6 QC Test For Fluoroscopic Equipment. Prepared by:- Kamarul Amin bin Abu Bakar School of Medical Imaging KLMUC

Similar documents
Nuclear Associates

I. PERFORMANCE OF X-RAY PRODUCTION COMPONENTS FLUOROSCOPIC ACCEPTANCE TESTING: TEST PROCEDURES & PERFORMANCE CRITERIA

Nuclear Associates

Nuclear Associates

TESTING FLAT-PANEL IMAGING SYSTEMS: What the Medical Physicist Needs to Know. JAMES A. TOMLINSON, M.S., D.A.B.R. Diagnostic Radiological Physicist

C-ARM FLUOROSCOPIC AND SPOT-FILM SYSTEMS

Nuclear Associates

Joint ICTP/IAEA Advanced School on Dosimetry in Diagnostic Radiology and its Clinical Implementation May 2009

LECTURE 1 The Radiographic Image

ABOVETABLE X-RAY SOURCE FLUOROSCOPIC AND SPOT-FILM SYSTEMS

Y11-DR Digital Radiography (DR) Image Quality

Introduction. Chapter 16 Diagnostic Radiology. Primary radiological image. Primary radiological image

NEMA. Cardiology Phantom Nuclear Associates Model Diagnostic Imaging. Introduction. Features. Applications

DENTAL RADIOGRAPHY KAMARUL AMIN BIN ABU BAKAR

NJDEP Medical Physicist s Radiographic QC Survey Registration Number:

Minnesota Rules, Chapter 4732 X-ray Revision

Test Equipment for Radiology and CT Quality Control Contents

FOUR CATEGORIES OF SAFETY

X-RAYS - NO UNAUTHORISED ENTRY

1-1. GENERAL 1-2. DISCOVERY OF X-RAYS

Beam-Restricting Devices

INTRODUCTION TO FLEXIBLE BRONCHOSCOPY. Fluoroscopy Synopsis HENRI G COLT MD SECOND EDITION THE BRONCHOSCOPY EDUCATION PROJECT SERIES

Nuclear Associates

DIAGNOSTIC ACCREDITATION PROGRAM. Radiology and CT Quality Control Procedures Workbook

QC Testing for Computed Tomography (CT) Scanner

Nuclear Associates

Fluoroscopy - Chapter 9

SPRINGFIELD TECHNICAL COMMUNITY COLLEGE ACADEMIC AFFAIRS

Acceptance Testing of a Digital Breast Tomosynthesis Unit

RULES OF TENNESSEE DEPARTMENT OF ENVIRONMENT AND CONSERVATION DIVISION OF RADIOLOGICAL HEALTH CHAPTER USE OF X-RAY APPARATUS

MAN Revision 002. InSight Mini C-arm Imaging System Technical Reference Manual

Unfors EDD-30 Radiation Protection in Fluoroscopy

STEREOTACTIC BREAST BIOPSY EQUIPMENT SURVEYS

Essentials of Digital Imaging

GE AMX 4+ Portable X-Ray

Ludlum Medical Physics

Collimation Assessment Using GAFCHROMIC XR-M2

Overview of Safety Code 35

10/3/2012. Study Harder

Visualization of sources of scattered radiation from x-ray equipment used for interventional radiology

10/26/2015. Study Harder

COMPUTED RADIOGRAPHY CHAPTER 4 EFFECTIVE USE OF CR

NATIONWIDE EVALUATION OF X-RAY TRENDS (NEXT) SUMMARY OF 2003 FLUOROSCOPY SURVEY

HIGH-RESOLUTION CORE FLUOROSCOPY, AN IMPORTANT TOOL FOR CORE ANALYSIS

Multiple Choice Identify the letter of the choice that best completes the statement or answers the question.

Investigation of the line-pair pattern method for evaluating mammographic focal spot performance

BASICS OF FLUOROSCOPY

MXHF-1500RF is controlled by Digital key panel console that displays KV, ma and mas with APR menu programmed.

Quality Control for Stereotactic Breast Biopsy. Robert J. Pizzutiello, Jr., F.A.C.M.P. Upstate Medical Physics, Inc

Ch. 223 VETERINARY MEDICINE CHAPTER 223. VETERINARY MEDICINE GENERAL PROVISIONS X-RAYS RADIOACTIVE MATERIAL. Authority

Sarah Hughes, MS, DABR Radiation Safety Officer

Surveying and QC of Stereotactic Breast Biopsy Units for ACR Accreditation

Diagnostic x-ray equipment compliance and facility survey

Nuclear Associates EZ CR-DIN Phantoms

CHAPTER 2 COMMISSIONING OF KILO-VOLTAGE CONE BEAM COMPUTED TOMOGRAPHY FOR IMAGE-GUIDED RADIOTHERAPY

Fluke199XRAY. Users Manual. Medical ScopeMeter

Dedicated Veterinary Imaging Solutions Digital, CR and Analog Imaging Solutions for any size patient and any size budget.

Key words: fluoroscopy, dose-area-product, kerma-area-product, calibration of KAP meters, patient exposure

RAD 150 RADIOLOGIC EXPOSURE TECHNIQUE II

Visibility of Detail

Safelight Fog does what to contrast and density on film?

Evaluation of a quality control phantom for digital chest radiography

Assessment of Beam Alignment, Collimation and Half Value Layer of Some Selected X-Ray Machines in Plateau State, Nigeria

X-RAY FLUOROSCOPY IMAGING SYSTEMS. Dr Slavik Tabakov. Luminescence: Dept. Medical Eng. & Physics King s College London

Veterinary Science Preparatory Training for the Veterinary Assistant. Floron C. Faries, Jr., DVM, MS

Digital Imaging started in the 1972 with Digital subtraction angiography Clinical digital imaging was employed from the 1980 ~ 37 years ago Amount of

Overview. Professor Roentgen was a Physicist!!! The Physics of Radiation Oncology X-ray Imaging

QUALITY CONTROL TESTS IN SOME DIAGNOSTIC X-RAY UNITS IN BANGLADESH

Essentials of Digital Imaging

X-ray Tube and Generator Basic principles and construction

SPECIFICATION. Kilovoltage X-ray calibration system for protection and diagnostic level dosimetry. Prepared by

Half value layer and AEC receptor dose compliance survey in Estonia

Dose Reduction and Image Preservation After the Introduction of a 0.1 mm Cu Filter into the LODOX Statscan unit above 110 kvp

Ansur TNT Users Manual. Plug-In

Title: A COMPARISON OF Cs-137 AND X-RAY SOURCES AS CALIBRATION REFERENCES FOR THERMOLUMINESCENT DOSIMETER CHIPS

REQUIREMENTS FOR LICENCE HOLDERS WITH RESPECT TO QUALITY CONTROL TESTS FOR DIAGNOSTIC X-RAY IMAGING SYSTEMS

Notice of Rulemaking Hearing

Nuclear Associates &

IChapter Number,ChapterTiiie I0400: _OJ TJse of X-Ray Apparatus! Rule Number Rule Title

Exposure System Selection

Teaching Digital Radiography and Fluoroscopic Radiation Protection

SYLLABUS. TITLE: Equipment Operation I. DEPARTMENT: Radiologic Technology

The X-ray circuit: part II

Learning Objectives: What s my motivation? (unknown screen actor) Workshop Overview

Features and Weaknesses of Phantoms for CR/DR System Testing

X-ray Equipment in Medical Diagnosis Part A: Recommended Safety Procedures for Installation and Use

ProX Intraoral X-ray. PLANMECA is proud to introduce a new intraoral X-ray unit to its comprehensive collection of imaging products- the ProX.

(CR) PORTABLE CHEST EXAM Baseline Definition Worksheet

Nuclear Associates , , CT Head and Body Dose Phantom

CHAPTER 40 - PROFESSIONAL LICENSING AND FACILITY REGULATION. PART 4 - Diagnostic X-Rays and Associated Imaging Systems in the Healing Arts

Version 1.0. TechnicVR. Student Guide

Quality control of Gamma Camera. By Dr/ Ibrahim Elsayed Saad 242 NMT

Clinical Experiences with a Patient Skin Dose Monitoring and Tracking Program

Do you have any other questions? Please call us at (Toll Free) or , or

Luminos RF Classic. Where value meets performance.

X-ray Imaging. PHYS Lecture. Carlos Vinhais. Departamento de Física Instituto Superior de Engenharia do Porto

Interventional Radiological Equipment selection and installation

Breast Tomosynthesis. Bob Liu, Ph.D. Department of Radiology Massachusetts General Hospital And Harvard Medical School

AN ABSTRACT OF THE THESIS OF. W. Scott Helms for the degree of Master of Science in Radiation Health Physics

2012 :15th SESSION of ESMP

Transcription:

CHAPTER 6 QC Test For Fluoroscopic Equipment Prepared by:- Kamarul Amin bin Abdullah @ Abu Bakar School of Medical Imaging KLMUC

Lesson Outcomes Describe the objectives of each QC test done. Identify QC tools and equipments involved in each QC test. Discuss the analysis of result of QC.

Contents of QC Fluoroscopic A. Exposure rate B. Beam alignment C. Source-to-skin distance limits D. Intensifier viewing system high contrast E. Intensifier viewing system low contrast F. TV monitors and recorders G. Automatic brightness control

EXPOSURE RATE

Objective and Equipments To determine the variation in reproducibility over a number of exposures at the same generator setting. To determine the variation in average kv over a number of exposures at the same generator setting. To evaluate the linearity of generator for commonly used exposure settings. Phantom, dosimeter, stopwatch, digital kv meter, energized fluoroscopic unit

Reproducibility (analysis)

kv Accuracy (Analysis) To calculate the kilovolt variance, we used this equation : kvp variance : (kvpmax -kvpmin) (kvpmax+ kvpmin) The kv variance should be less than @ equal to 0.05 The measure kv and the value indicated on the control should coincide within + 5%.

Milliampere Linearity (Analysis) To calculate the ma linearity,we must used this equation: Linearity variance: (mr/masmax - mr/masmin 2) (mr/masaverage) The linearity variance should within 0.1 (10%)

BEAM ALIGNMENT

Objective and Equipments To evaluate the x-ray beam alignment of fluoroscopy unit in ensuring it is not larger than normal collimation. Fluoroscopic beam alignment device, x-ray film, protection shielding, energized fluoro unit.

Analysis Theedgesofthebeamfieldarenotinalign,thereforeb eamfieldandimagereceptoraremisaligned. The center of the field is indicated by the central dot and any misalignment of the beam can be checked by counting the number of dots visible in each channel. Supposedly, the dot at the intersection of the two lines should be at the center of the image There should be an equal number of dots visible on each of the channels when counted from the centre.

SOURCE TO SKIN DISTANCE

Objective and Equipments To reduce the skin exposure to the patient as much as possible while still assuring that sufficient x-ray output is obtainable to perform the diagnostic procedure. Energized fluoro unit, measuring tape.

Analysis Standard result: the source to skin distance must be not less than 15 inches (38cm) for stationary fluoroscopy for mobile fluoroscopy must be not less than 12 inches (30cm)

INTENSIFIER VIEWING SYSTEM HIGH CONTRAST

Objective and Equipments

Analysis The minimum number of lp/mm that must be resolved by the system is given by the equation: 2 lp/mm x (6 inches (15cm) / size of FOV used) = minimum number of lp/mm.

INTENSIFIER VIEWING SYSTEM LOW CONTRAST

Objective and Equipments To resolve relatively large objects that differ slightly in radiolucency from the surrounding area. Energized fluoroscopic unit, low contrast resolution test tool

Analysis When the plates are sandwiched together and imaged on a fluoroscopic system, low contrast evaluation can be made based on the shallowest visible hole. With the low contrast resolution test tool, the contrast between the holes and the surrounding area is 2%. with CDRH phantom, the two deepest holes should be visible clearly. Better system are able to visualize the smaller holes on the test tools or the shallower holes on the CDRH phantom.

TV MONITORS AND RECORDERS

Objective and Equipments To assure that the TV monitor and recorder are functioning properly. Energized fluoro unit, phantom, test tool

For kvp and ma changes: Analysis If the indicated kvp differs from the baseline value by more than 5% or the ma by more than 10%, recheck the setup of the phantom and fluoro system. If the setup is correct and the changes persists, then contact service personnel or the medical physicist. Decrease in number of meshes seen: If one mesh group is lost and the kvp value is the same then try adjusting the monitor brightness if possible to optimize visualization. If no improvement is obtained then contact service or medical physicist to perform a fluoroscopic resolution test and system focusing adjustment. Decrease number of holes seen: If one hole is lost and the kvp value is the same then try adjusting the monitor brightness if possible to optimize visualization. Ensure that you are viewing the monitor from a distance of 4 feet. If no improvement is obtained then contact service or medical physicist to perform a fluoroscopic resolution test and system focusing adjustment.

TEST TOOL

AUTOMATIC BRIGHTNESS CONTROL

Objective and Equipments To assure that the automatic brightness control are functioning properly. Energized fluoro unit, lead apron, phantom

Analysis As the thickness of structure increases, the brightness or contrast remain the same. ABC adjusts the contrast automatically. Allows the fluoroscopic unit to automatically maintain the brightness level of the image (variations of thickness and attenuation)