Pitfalls and Remedies of MDCT Scanners as Quantitative Instruments

Similar documents
12/21/2016. Siemens Medical Systems Research Agreement Philips Healthcare Research Agreement AAN and ASN Committees

1. Patient size AEC. Large Patient High ma. Small Patient Low ma

Software and Hardware in CCTA. Elly Castellano PhD

Automated dose control in multi-slice CT. Nicholas Keat Formerly ImPACT, St George's Hospital, London

TOPICS: CT Protocol Optimization over the Range of Patient Age & Size and for Different CT Scanner Types: Recommendations & Misconceptions

Research Support. Dual-Source CT: What is it and How Do I Test it? Cynthia H. McCollough, Ph.D.

Wide-Detector CT for TAVR Planning:

diagnostic examination

HISTORY. CT Physics with an Emphasis on Application in Thoracic and Cardiac Imaging SUNDAY. Shawn D. Teague, MD

Suppression of metal artifacts using image-based monoenergetic DECT imaging

Image Quality and Dose. Image Quality and Dose. Image Quality and Dose Issues in MSCT. Scanner parameters affecting IQ and Dose

PET/CT Instrumentation Basics

Breast Tomosynthesis. Bob Liu, Ph.D. Department of Radiology Massachusetts General Hospital And Harvard Medical School

Yinsheng Li 1, Peter Bannas 2, M.D., Perry Pickhardt M.D. 2, Meghan Lubner M.D. 2, Ke Li Ph.D. 1,2, and Guang-Hong Chen Ph.D. 1,2

QC Testing for Computed Tomography (CT) Scanner

PD233: Design of Biomedical Devices and Systems

TORNIER BLUEPRINT. 3D Planning + PSI SCAN PROTOCOL

COMPUTED TOMOGRAPHY 1

Introduction. Chapter 16 Diagnostic Radiology. Primary radiological image. Primary radiological image

PET: New Technologies & Applications, Including Oncology

CT Basics: Image Quality Module 6

CHAPTER 2 COMMISSIONING OF KILO-VOLTAGE CONE BEAM COMPUTED TOMOGRAPHY FOR IMAGE-GUIDED RADIOTHERAPY

Detector technology in simultaneous spectral imaging

CT Basics: Data Acquisition Module 3

Authors: Cabral, Ricardo 1 ; Carvoeiras, Pedro 2 ; Fatana, João 2, ; Alves, Rita 1. 1 Centro Hospitalar Lisboa Norte - Hospital de Santa Maria; 2

160-slice CT SCANNER / New Standard for the Future

Dose Reduction and Image Preservation After the Introduction of a 0.1 mm Cu Filter into the LODOX Statscan unit above 110 kvp

Clinical Experience Using the Open Bore Multislice CT System Supria (16 slice CT) MEDIX VOL. 61 P.8 P.11

Maximum Performance, Minimum Space

Electronic Noise in CT Detectors: Impact on Image Noise and Artifacts

Metal Artifact Reduction for Orthopedic Implants (O-MAR)

Influence of different iteration levels in fourth generation iterative reconstruction technique on image noise in CT examinations of the neck

abc MHRA Philips Mx8000 IDT CT scanner technical evaluation September 2004 Best choice best practice nww.medical-devices.nhs.

NeuViz 16 Computed Tomography. Elevating routine imaging for exceptional results

Model Based Iterative Reconstructions represent a paradigm shift - Imaging with almost no noise

Redefining Ergonomics

CT Basics: Equipment and Instrumentation Module 2

Beam-Restricting Devices

Translating Protocols Between Scanner Manufacturer and Model

Radionuclide Imaging MII Single Photon Emission Computed Tomography (SPECT)

Features and Weaknesses of Phantoms for CR/DR System Testing

Iterative Reconstruction in Image Space. Answers for life.

This document is published in:

Diffraction-enhanced X-ray Imaging (DEXI) Medical Solutions. More information using less radiation

Abdominal Radiology ISSN X. Abdom Radiol DOI /s

COMPUTED RADIOGRAPHY CHAPTER 4 EFFECTIVE USE OF CR

QC by the MPE in Belgium

Dose Reduction in Helical CT: Dynamically Adjustable z-axis X-Ray Beam Collimation

DOUBLE gated in vivo small animal cone beam micro

The future of nuclear imaging is clear

Veterinary Science Preparatory Training for the Veterinary Assistant. Floron C. Faries, Jr., DVM, MS

MC SIMULATION OF SCATTER INTENSITIES IN A CONE-BEAM CT SYSTEM EMPLOYING A 450 kv X-RAY TUBE

Improvement of CT image quality with iterative reconstruction idose4

X-ray Imaging. PHYS Lecture. Carlos Vinhais. Departamento de Física Instituto Superior de Engenharia do Porto

FOREWORD. Acknowledgements

7/24/2014. Image Quality for the Radiation Oncology Physicist: Review of the Fundamentals and Implementation. Disclosures. Outline

Lunar Technology Advantages

Iterative Reconstruction

Mammography: Physics of Imaging

K-edge subtraction X-ray imaging with a pixellated spectroscopic detector

Multislice CT: Current Technology and Future Developments

January 2012 February 1, Release Date: Expiration Date:

Clinical Importance on CT

Chiara Secco. PET Performance measurements of the new LSO-Based Whole Body PET/CT. Scanner biograph 16 HI-REZ using the NEMA NU Standard.

Slide 1. Slide 2. Slide 3 ACR CT Accreditation. Multi-Slice CT Artifacts and Quality Control. What are the rules or recommendations for CT QC?

SOMATOM Esprit A Bundle of Energy

SAFIRE. Sinogram Affirmed Iterative Reconstruction. Answers for life.

Y11-DR Digital Radiography (DR) Image Quality

Wide beam CT dosimetry. Elly Castellano

Computerized Medical Imaging and Graphics

IBEX TECHNOLOGY APPLIED TO DIGITAL RADIOGRAPHY

2010 Philips BrightView XCT SPECT/CT

SPECT Reconstruction & Filtering

PHYSICS QUESTIONNAIRE FORM

Diagnostic X-Ray Shielding

of sufficient quality and quantity

Advanced MSK MRI Protocols at 3.0T. Garry E. Gold, M.D. Associate Professor Department of Radiology Stanford University

Advanced Noise Reduction Processing for X-ray CT System with Iterative Processing. Koichi Hirokawa MEDIX VOL. 56 P.43 P.46

A Basic Overview of Veterinary Computed Tomography

Exposure Indices and Target Values in Radiography: What Are They and How Can You Use Them?

Optimized CT metal artifact reduction using the Metal Deletion Technique (MDT)

Ludlum Medical Physics

Efficacy of Beam-Hardening Correction for Vascular Morphometry Using 3D Conebeam Micro-CT

Radiology Physics Lectures: Digital Radiography. Digital Radiography. D. J. Hall, Ph.D. x20893

Acceptance Testing of a Digital Breast Tomosynthesis Unit

T h e P h a n t o m L a b o r a t o r y

Overview of Safety Code 35

Quantitation of clinical feedback on image quality differences between two CT scanner models

A positioning QA procedure for 2D/2D (kv/mv) and 3D/3D (CT/CBCT) image matching for radiotherapy patient setup

Digital radiography (DR) post processing techniques for pediatric radiology

2012 :15th SESSION of ESMP

Inside Biograph mct.

X-Ray-Based Attenuation Correction for Positron Emission Tomography/Computed Tomography Scanners

Phantoms in Medical Physics (RT) U. Oelfke. Division of Radiotherapy & Imaging

The disclaimer on page 1 is an integral part of this document. Copyright November 30, 2017 by AAPM. All rights reserved.

The influence of bowtie filtration on cone-beam CT image quality

Enhanced Functionality of High-Speed Image Processing Engine SUREengine PRO. Sharpness (spatial resolution) Graininess (noise intensity)

INNOVATION BY DESIGN. Toshiba A History of Leadership REMOTE CONTROL R/F SYSTEM

An Activity in Computed Tomography

Noise Characteristics of the FORE+OSEM(DB) Reconstruction Method for the MiCES PET Scanner

Transcription:

intensity m(e) m (/cm) 000 00 0 0. 0 50 0 50 Pitfalls and Remedies of MDCT Scanners as Jiang Hsieh, PhD GE Healthcare Technology University of Wisconsin-Madison Root-Causes of CT Number Inaccuracies Nature of the X-ray Physics Beam Hardening Scatter Aliasing New Technology Helical Cone Beam Incomplete Sampling Patient Motion Photon Starvation Operator Protocols (scan thin, recon thick) Partial Volume scanner patient operator 2 Nature of Physics: Beam Hardening Bone Beam Hardening measured channel ideal poly-energetic x-ray photon energy-dependent m(e) measured projection error polynomial correction N p n ' n p n energy, E The attenuation characteristics of bone is significantly different from that of soft tissue. Similar observation can be made for contrast agents. bone water energy (kev) 0 cm water 5 cm water 30 cm water no correction with correction 3 Water BH Error Bone BH 4

m (/cm) Dual Energy Correction Dual energy takes advantage of the two types of x-ray interaction with matter: photoelectric and Compton. Use of the additional information can effectively remove BH. Impact of Iodine Contrast Iodine contrast Cardiac phantom experiments Dual energy solution Conventional 20kVp Dual Energy 70keV 0000 000 00 0 Iodine 80 kvp 40 kvp Projection-space DECT Soft-tissue Bone 0. 0 50 0 50 energy (kev) Cardiac Phantom ww=300 5 6 Regular 40 kvp Perfusion Perfusion is important to test the viability of a tissue Perfusion requires accurate quantitation of CT numbers BH is one of the major sources of error DE Mono 70 kev Metal-induced Error CT is sensitive to the presence of high-density and highfrequency objects. Special care must be taken to avoid placing foreign objects inside the scanner FOV. 0.625mm thickness,.0mm apart 40kVp Images courtesy of Drs. T. Lee and A. So of St. Mary s Hospital, London, OT, Canada metal reinforcement 7 8 2

S PR SPR CT Number (HU) Nature of Physics: Scatter Scatter Artifacts X-ray source Scatter-to-Primary Ratio (SPR) increases almost linearly with the -coverage. SPR decreases with the increase in kvp SPR increases with the object sie 0.5 Measured SPR in VC T /H D at 40mm coverage scatter Scatter-to-Primary Ratio (SPR) increases almost linearly with the -coverage. SPR decreases with the increase in kvp SPR increases with the object sie ghost image scatter 0.4 0.3 35 30 20kVp 48cm Poly 0.2 8 0 kvp 25 20 0. 0 0 kvp 5 2 0 kvp 0 5 25 35 45 55 w ater diameter (cm ) 5 0 0 50 0 50 200 Beam Width (mm) 9 32mm Detector shading 60mm Detector Lung CT Number Accuracy Inspiration Level Beam Hardening Patient centering Variation of object sie Presence of non-water materials Scatter Increased volume coverage Insufficient post-patient collimation JP Sieren, PF Judy, DA Lynch, JD Newell, HO Coxson, and EA Hoffman, QIBA Quantitative CT: Towards routine quantitative CT in obstructive lung disease, QIBA COPD/Asthma Subcommittee Lung Volume CT Number Air-Lung Accuracy Air CT number can be impacted by Increased cone beam effect Increased scatter Axial Sagittal view -980-985 32mm Helical -990-995 -00-05 - -5-20 60mm Axial -25-30 5 5 20 2 3

Scatter Correction Increased collimator aspect ratio: H/W 2D collimation Algorithmic correction possible Phantom calibration Truncation Issue For x-ray radiography, truncation does not cause artifact inside FOV. Truncated projection will cause significant artifacts for CT W COPDGene Phantom (CTP657) Chest X-ray H x Full FOV Truncated original corrected 3 4 Sie Matters CT is sensitive to both high density object and objects extended outside the FOV. Patient arms need to be positioned outside the FOV. Beer s law indicate that the amount of attenuation increases exponentially with path length. I e I 0 At low signal level, the noise in the projection is no longer dominated by the x-ray photon. Convolution filtering operation will further amplify the noise and streak artifacts will result. ml Photon Starvation 50cm FOV Scout Image CT Image 5 6 4

m (/cm) Pitfalls and Remedies of MDCT Scanners as Protocol optimiation Scan speed Helical pitch kvp selection Slice thickness Reconstruction kernel ma modulation Protocol Optimiation 0 soft-tissue 0. Bone 0 30 60 90 20 50 energy (kev) Algorithmic Correction Adaptive filtering for streak reduction original adaptively filtered 80kVp, 6.4mGy 40kVp, 6.3mGy 7 Sinogram 8 Iterative Reconstruction Dose report Cardiac Motion 6-slice Examples FBP 20 kv, ma, 0.5 rot time 0.625mm slice thickness.78msv* MBIR Complexity of cardiac motion Artifacts from several factors In-plane vs. cross-plane temporal resolution *Obtained by EUR-6262 EN, using an abdomen factor of 0.05*DLP and a pelvis factor of 0.09*DLP Images courtesy of Dr Barrau, CCN, France sub-optimal gating In-Plane sub-optimal table speed Cross-Plane phase mis-registration 20 5

Artifact Correction In-Plane Temporal Resolution Improvement: Phase optimiation Faster scan speed Multiple source-detector Advanced reconstruction Advanced Technology-induced Artifact Half-scan: p + fan angle acquisition. Dual source reduces the acquisition by 40-45% Two detectors have different sie Object outside small FOV may cause artifacts. Compensation algorithm can reduce the impact centered phantom off-centered phantom half-scan smaller detector FOV sub-optimal phase optimal phase 25 g at 0.35 s 8X safety margin 200 g 76 g at 0.2 s 8X safety margin 62 g 2 22 Cross-Plane Temporal Resolution: Cone Beam CT Increase -coverage will reduce phase mis-registration artifacts. Entire heart covered by a single rotation. Step-and-shoot mode artifacts are due to: Missing frequencies Mis-handled frequencies Z-truncation desired ROI Axial Truncation detector Cone Beam Axial Exact cone-beam reconstruction does not exist. Large cone angle can lead to cone-beam artifact. ww=600 Small Detector Large Detector source trajectory phase mis-registration Helical Mode Step-and-shoot 23 24 6

Advanced Conventional Cone Beam Axial Exact cone-beam reconstruction does not exist for step-and-shoot mode. Large cone angle can lead to cone-beam artifact. Active research in on-going Axial, Phantom Temporal Resolution Improvement: Advanced Reconstruction Algorithms Characterie cardiac motion Compensate for motion Axial, Phantom Coronal, Simulation 25 Conventional Rest/Low HR Conventional Stress/High HR Advanced Stress/High HR High HR + motion correction 26 Conclusion There are many sources of error for quantitative CT Nature of Physics New Technologies Patient Operator Many approaches are taken to reduce or eliminate the error. It takes a combined efforts from manufactures, CT operators, and patient corporation to reduce the impact of these pitfalls to a minimum. 27 28 7