1D Transient NOE on the Bruker DRX-500 and DRX-600

Similar documents
User manual Bruker DPX200 NMR spectrometer

Your first NMR measurement

H Micro-Imaging. Tuning and Matching. i. Open any 1H data set and type wobb.

HMBC 17. Goto. Introduction AVANCE User s Guide Bruker 185

SUPPORTING INFORMATION

PULSED/CW NUCLEAR MAGNETIC RESONANCE

NMR Basics. Lecture 2

10. Phase Cycling and Pulsed Field Gradients Introduction to Phase Cycling - Quadrature images

400 MHz spectrometer user manual

Two Dimensional Homonuclear Correlation Spectroscopy

(N)MR Imaging. Lab Course Script. FMP PhD Autumn School. Location: C81, MRI Lab B0.03 (basement) Instructor: Leif Schröder. Date: November 3rd, 2010

Gradients. Effects of B0 gradients on transverse magnetisation Similar to figure 10 of Sattler review Progr. NMR 34 (1999), 93

PINMRF. Varian 300 MHz NMR Spectrometers User Guide for Advanced 1D and Basic 2D NMR Experiments

Step by step procedure for NMR data acquisition

NMR Spectrometer Operation: xwinnmr

Fast Methods for Small Molecules

Lab 8 6.S02 Spring 2013 MRI Projection Imaging

Student Name: Date Completed: Supervisor:

In a typical biological sample the concentration of the solute is 1 mm or less. In many situations,

PHY3902 PHY3904. Nuclear magnetic resonance Laboratory Protocol

Instruction for Operating the Bruker Avance III 800 MHz NMR Spectrometers in UTMB

Two Dimensional Homonuclear Correlation Spectroscopy

NMR spectrometer usage at the BioNMR facility ETH Zürich

8 COSY. 8.1 Introduction. 8.2 Magnitude COSY

Instructions for 1 H-, 13 C-, 19 F-, and 31 P-Spectra on the Varian Mercury-Vx-300

An NMR Caveman s Guide to Quickly Acquiring Spectroscopic Data By Brian Sparling

2015 Spin echoes and projection imaging

6.S02 MRI Lab Acquire MR signals. 2.1 Free Induction decay (FID)

KJM D-SELECTIVE NMR Experiments on the AVIIIHD-800. Version 1.0. Topspin 3.5 Windows 7

A Conceptual Tour of Pulsed NMR*

KJM D-SELECTIVE NMR Experiments on the AVI-600 and AVII-600. Version 1.0. Topspin 3.5 Windows 7 Topspin 1.3 Windows XP

If the magnetic field is larger, more energy is required to excite a given nucleus.

Chapter 11 Coherence Editing: Pulse-field Gradients and Phase Cycling

KJM Version 1.0. Topspin 3.5 Windows 7 Topspin 1.3 Windows XP

Optical Pumping Control Unit

Intermediate and Advanced Labs PHY3802L/PHY4822L

Implementation of parallel search algorithms using spatial encoding by nuclear magnetic resonance

Chem 203. Organic Spectroscopy. Midterm Examination, Part II (60 points total) Problem 4 of 4 (three out of four required, 20 points)

Chem 203. Organic Spectroscopy. Midterm Examination, Part II (60 points total) Problem 1 of 4 (three out of four required, 20 points)

Evaluation of Adiabatic Frequency-Modulated Schemes for Broadband Decoupling in Isotropic Liquids

Ultrahigh-resolution Total Correlation NMR Spectroscopy

The Agilent OneNMR Probe

Principios Básicos de RMN en sólidos destinado a usuarios. Gustavo Monti. Fa.M.A.F. Universidad Nacional de Córdoba Argentina

Arrayed Acquisition in VNMR and on the Gemini 1

Chem 203. Organic Spectroscopy. Midterm Examination, Part II (60 points total) Problem 4 of 4 (three out of four required, 20 points)

Background (~EE369B)

1 Introduction. 2 The basic principles of NMR

Combinational logic: Breadboard adders

Open acqi window if the button has been lost. autolocking routine, alock= y for autolocking, alock= n for typical manual locking

NMR Hardware 06/06/2017. Outline. Instrumentation: Magnet. Increasing magnetic field increases Sensitivity, by power of 3/2 Dispersion, linearly

EXP 9 ESR (Electron Spin Resonance)

Exercise 2-6. Target Bearing Estimation EXERCISE OBJECTIVE

PULSED NUCLEAR MAGNETIC RESONANCE. Advanced Laboratory, Physics 407 University of Wisconsin Madison, Wisconsin 53706

Transient Data Acquisition System, TAS 4-40 Potential-free measurement of fast rise pulses:

Two Dimensional Heteronuclear Correlation Spectroscopy

Implementing ultrafast 2D NMR experiments on a Bruker Avance Spectrometer

Free vibration of cantilever beam FREE VIBRATION OF CANTILEVER BEAM PROCEDURE

Gentec-EO USA. T-RAD-USB Users Manual. T-Rad-USB Operating Instructions /15/2010 Page 1 of 24

Efficacy of Wavelet Transform Techniques for. Denoising Polarized Target NMR Signals

EE320L Electronics I. Laboratory. Laboratory Exercise #2. Basic Op-Amp Circuits. Angsuman Roy. Department of Electrical and Computer Engineering

Qualion NMR Process\Lab NMR (Nuclei Magnetic Resonance) Analyzer Model MASH, Style D

4: EXPERIMENTS WITH SOUND PULSES

A TECHNIQUE TO EVALUATE THE IMPACT OF FLEX CABLE PHASE INSTABILITY ON mm-wave PLANAR NEAR-FIELD MEASUREMENT ACCURACIES

ECE 532 Hspice Tutorial

FLASH rf gun. beam generated within the (1.3 GHz) RF gun by a laser. filling time: typical 55 μs. flat top time: up to 800 μs

Chapter 5. Clock Offset Due to Antenna Rotation

Recording EPR Spectra using ER 4102ST Resonator

Notes on Experiment #1

Instruction manual for T3DS software. Tool for THz Time-Domain Spectroscopy. Release 4.0

Localization of microscale devices in vivo using addressable transmitters operated as magnetic spins

Supplementary Figure 1. Scanning Electron Microscopy images of the pristine electrodes. (a) negative electrode and (b) positive electrode.

AC/DC Current Probe CT6844/CT6845/CT6846

Advanced Lab LAB 6: Signal Acquisition & Spectrum Analysis Using VirtualBench DSA Equipment: Objectives:

7 Equipments. Spectrometers

DIODE LASER SPECTROSCOPY (160309)

EPR2010 Puerto Rico. Rapid Scan EPR. Mark Tseitlin, Deborah G. Mitchell, Joshua A. Biller, Richard W. Quine, George A. Rinard, Sandra S.

Standard Operating Procedure

Notes on OR Data Math Function

BLADE AND SHAFT CRACK DETECTION USING TORSIONAL VIBRATION MEASUREMENTS PART 2: RESAMPLING TO IMPROVE EFFECTIVE DYNAMIC RANGE

3D Distortion Measurement (DIS)

TopSpin Guide Book. Basic NMR Experiments User Manual. Innovation with Integrity. Version 002 NMR

Exercise 2: Hodgkin and Huxley model

DC and AC Circuits. Objective. Theory. 1. Direct Current (DC) R-C Circuit

An Alternative to Pyrotechnic Testing For Shock Identification

Setting up a Multi sine impedance measurement

Many applications. Mismatched Load Characterization for High-Power RF Amplifiers PA CHARACTERIZATION. This article discusses the

Experiment 6: Franck Hertz Experiment v1.3

THE BENEFITS OF DSP LOCK-IN AMPLIFIERS

Electronics Interview Questions

Development of a new Q-meter module

M R I Physics Course. Jerry Allison Ph.D., Chris Wright B.S., Tom Lavin B.S., Nathan Yanasak Ph.D. Department of Radiology Medical College of Georgia

Seized Drugs Operational Guidelines for the Thermo FTIR Comparative and Analytical Division

Exercise 6. Range and Angle Tracking Performance (Radar-Dependent Errors) EXERCISE OBJECTIVE

Photon Counters SR430 5 ns multichannel scaler/averager

S1. Current-induced switching in the magnetic tunnel junction.

Procidia Control Solutions Dead Time Compensation

The Oscilloscope. Vision is the art of seeing things invisible. J. Swift ( ) OBJECTIVE To learn to operate a digital oscilloscope.

Vibrating Wire R&D for Alignment of Multipole Magnets in NSLS-II

Class #8: Experiment Diodes Part I

Engineering 3821 Fall Pspice TUTORIAL 1. Prepared by: J. Tobin (Class of 2005) B. Jeyasurya E. Gill

Transcription:

1D Transient NOE on the Bruker DRX-500 and DRX-600 Reference: Stott, K., Stonehouse, J., Keeler, T.L. and Shaka, A.J., J. Amer. Chem. Soc. 1995, 117 (14), pp. 4199-4200. At thermal equilibrium in a strong magnetic field there is a slight excess of population of nuclei in the lower energy (aligned with the magnetic field) state and a slight depletion of nuclei in the higher energy state (opposed to the magnetic field). If this equilibrium is perturbed for one group of nuclei (corresponding to a peak in the 1 H spectrum), this perturbation is propapaged to nearby nuclei in the molecule due to the Nuclear Overhauser Effect (NOE). Because the intensity of a peak in an NMR spectrum is directly proportional to this population difference, the perturbation can be measured by simply recording a spectrum. The traditional 1D NOE experiment involve irradiating with low power radio frequency at the resonant frequency of one peak in the 1 H in order to equalize the populations of the two states ("saturation"). This saturated state is maintained by continued irradiation until the perturbation populations of nearby nuclei in the molecule reaches a steady-state and does not change any further. Then a 90 o pulse is applied and an FID is recorded to measure the amount of perturbation on the nearby nuclei. Since the enhancement of signals is quite small (a few percent), it is necessary to record a control spectrum with irradiation away from any peaks in the spectrum, and then subtract the control spectrum from the NOE spectrum. There are a number of disadvantages to this approach: 1. In any difference spectrum the conditions (temperature, RF power, sensitivity, magnetic field, vibration) must be identical in the two experiments in order to get perfect subtraction of the signals which are not affected. This subtraction is always imperfect since the two spectra are recorded at different times, so there are always big subtraction artefacts in the difference spectrum. 2. The magnitude of the NOE is proportional to inverse sixth power of the distance between two nuclei only for very short times between the perturbation and the measurement of the effect on other nuclei. The magnitude of the steady-state NOE is dependent on many other competing relaxation processes so it cannot be used as an accurate measure of distance. To accurately measure distances you need to measure the NOE with a number of different times between perturbation and measurement ("mixing times") and measure the initial slope of the curve as the effect increases with time. 3. The selectivity of irradiation is limited, and in crowded regions of the spectrum nearby peaks are also affected. This sometimes makes the results ambiguous. With shaped (selective) pulses we can apply a 180 o pulse specifically to a single peak in the spectrum. This is the most dramatic pertubation you can create, since the excess population in the lower energy level is now in the higher energy level and the depleted population is now in the lower energy level. If we then wait a short time for this 21

perturbation to propagate to nearby nuclei, a 90 o pulse will "read" the effect on the other nuclei in the form of a spectrum with enhanced peak areas. To avoid having to subtract two spectra, we can also use gradients to kill any magnetization on other nuclei at the same time that we invert the desired peak. Thus the only thing that will be detected with the 90 o "read" pulse is the perturbation due to the NOE. In principle, an NOE spectrum similar to an NOE difference spectrum could be obtained with one scan, but for improved signal-to-noise ratio we will average the result of many scans. This method relies only on the transient NOE and does not involve the subtraction of spectra. The results are excellent and there are practically no artefacts. The pulse sequence is as follows: A non-selective 90 o pulse rotates all of the sample magnetization onto the y axis. Then a gradient "twists" the magnetization into a helix. The selective (shaped) 180 o pulse is applied to invert the magnetization of the peak of interest, so that its "twist" is now in the reverse direction. A second gradient of equal intensity and duration to the first now unwinds the twist for the peak of interest. But all the other peaks in the spectrum are just twisted twice as far, since their magnetization helix was not reversed by the selective 180 o pulse. This destroys this magnetization and leaves only one thing in the sample: the peak of interest with its magnetization aligned along the -y axis. A second non-selective 90 o pulse is now applied to rotate this magnetization from the -y axis to the -z axis. Thus we have accomplished two things: the peak of interest has been inverted (population inversion) and the rest of the peaks have been destroyed. This approach of destroying all of the unwanted magnetization to "reveal" the desired magnetization was developed by A. J. Shaka at the University of California, Irvine, who called it "excitation sculpting" by analogy to a sculptor revealing a work of art by chipping away pieces of stone. 22

Setting Up for the Selective 1D NOE Experiment Insert your sample, lock and shim. Do not spin the sample for this experiment. Make sure the temperature is stable at the regulated value (temon) and re-shim if necessary after the temperature has stabilized. Get parameters for a simple 1D 1H experiment (rpar) and record a 1 H spectrum. Transform and phase, and set SR (spectral reference) to the same value as O1. This makes the scale read directly in units of absolute audio frequency in Hz relative to the center of the spectral window. Click the left mouse button to get the vertical arrow cursor, and move the cursor to the position of the peak you wish to select. Read off the cursor frequency from the small window above the spectrum and write down this precise value. This is the offset frequency (SPOFF1) for the shaped pulse. If this value were set to zero you would get selective excitation at the center of the spectral window (that is, at the O1 frequency). By setting an offset frequency you can shift the frequency of excitation to any desired peak in the spectrum. For example, let's say you enter O1 and find that the O1 value is 2858.20 Hz. Then you enter SR and enter 2858.20 for SR. Place the vertical arrow cursor on the peak of interest in your spectrum and read -1706.13 Hz for the cursor frequency. This is the value you need for the shaped pulse frequency offset (SPOFF1). Write this down and enter it when you run ased. (Do not use this exact number, of course, yours will be different!!!!) For the cholesterol sample, you will see that the two proton signal at 2.25 which represents the h4 ax and h4 eq is beginning to be resolved into two signals at 500 MHz. This is the signal which gave NOEs to m19 (the 19-Me group) and h6 (the olefinic proton) in the steady-state NOE difference experiment on the Unity-300. Looking at the structure of cholesterol, it seems likely that the NOE to the angular methyl group is primarily from h4 ax, and the NOE to h6 is primarily from h4 eq. With shaped pulses (and at higher field) we may be able to separate these two effects. In the expanded plot below, 23

if you ignore the small couplings, you can see a strongly skewed downfield doublet (h4 eq ) and a skewed upfield triplet (h4 ax ). They are leaning toward each other because their chemical shifts are so close. The triplet structure of h4 ax is due to the fact that it has two large couplings: one to h4 eq (geminal) and one to h3 (antiperiplanar). The h4 ax signal has only one large coupling, the geminal coupling to h4 eq. The two smaller couplings observed in the upfield (h4 eq ) signal are due to h3 (vicinal axial-equatorial) and h2 eq (Wcoupling). Get the offset values for the center of the h4 eq signal and for the center of the h4 ax signal. These two values will be used for two separate 1D NOE experiments. Note that there is no need to record a "control" frequency since this is a direct NOE measurement and not a difference spectrum. This offset value and the mixing time are the only parameters you need to edit! Everything else is set for you in the standard parameter set for this experiment. "Clone" your 1D proton experiment to a new area so you don't overwrite your proton spectrum: enter edc change experiment number to a new experiment (e.g., 2) click on SAVE You should now be in a new 1D experiment with no FID data, but with standard 1 H parameters. Now get the standard parameters for this experiment using rpar. Scroll down to the bottom of the list where the parameter sets starting with "std" are found, and select std-1dnoe-n3. Select Copy All to load the parameters. You must have the Nalorac Triple-Resonance Gradient Probe installed to run this experiment!!! To check the parameters, enter eda and make sure the O1 and SW_h values are set the same as they were for your 1D 1 H experiment. Click on SAVE and enter ased. Scroll to the bottom half of the list and enter the value you calculated for the shaped pulse frequency offset (SPOFF1). Check the value of the delay time for NOE transfer (the mixing time) which is d8 in this experiment. A value of 0.35 ms should be optimal for a molecule the size of cholesterol (larger molecules need shorter mixing times and smaller ones need longer). Click on SAVE. To adjust the receiver gain, go to the acquisition window (enter a or acqu) and enter gs. You should see the lock signal periodically fall and then recover as the gradients fire. Adjust the vertical scale (using the *2, /2, *8 and /8 buttons) so that the 100,000 value in absolute units is near the top of the scale. If the vertical scale is in centimeters, you need to exit the acquisition window (click Return) and select Utilities and YU, then return to the acquisition window (a). If the FID at any point exceeds the limits from -100,000 to +100,000 on this absolute scale, you need to reduce the RG value. You can do so by entering RG at any time, even 24

while gs is running. If the FID is significantly smaller than +/-100,000 you should increase the value of RG. When you are satisfied with the receiver gain, enter halt (DO NOT click on STOP - this can lock up the gradient amplifier and the lock display!!!). You are now ready to start the experiment. For this concentrated sample (43 mg cholesterol) you can set NS to 16 and DS to 4 scans. Start the experiment with zg. Set up the second experiment in the next consecutive experiment number (use edc to clone) and edit the value of SPOFF1. When the first NOE experiment is done you can start the second one. If you are really being efficient you can set up both experiments and queue them using the command multizg. Data Processing The selective 1D NOE experiment is just a 1D 1 H spectrum so you can process it normally. Phase correct so that the selected peak is upside-down (negative) and you will see NOE peaks as normal (positive) peaks. To compare to the normal 1D spectrum, click on Display / Dual Display. Select the second spectrum to display by editing the experiment number and setting the processing number (PROCNO) to 1. If you don't see the second spectrum, its color may be the same as the background. Enter setres and select Second Spectrum, using the sliders to change the color to 100% red (or any color that stands out). Click on Apply and then exit the User Interface Settings window. Again select Display / Dual Display and you should see the second spectrum displayed in red. In this mode you can change the vertical scale of the second spectrum (*2, /2, etc.) or shift it horizontally relative to the first spectrum. This makes it easy to compare two 1D spectra. When finished, click on return to exit the Dual Display mode. Compare the two 1D NOE spectra and verify that the NOEs are very different (and make sense in terms of distances) for h4 ax and h4 eq. 25