Quick Check of EIS System Performance

Similar documents
Measuring Batteries using the Right Setup: Dual-cell CR2032 and Battery Holder

EIS Measurement of a Very Low Impedance Lithium Ion Battery

Testing Electrochemical Capacitors Part 3: Electrochemical Impedance Spectroscopy

Compliance Voltage How Much is Enough?

Interface 5000 Potentiostat/Galvanostat/Zero-Resistance Ammeter

Potentiostat/Galvanostat/Zero Resistance Ammeter

POT/GAL 15 V 10 A and POT/GAL 30 V 2 A. Electrochemical Impedance Potentiostat Galvanostat Test Interface for Alpha-A Analyzer

173 Electrochemical Impedance Spectroscopy Goals Experimental Apparatus Background Electrochemical impedance spectroscopy

Universal Dummy Cell 3. Operator's Manual

Universal Dummy Cell 2. Operator's Manual

Reference 3000 Potentiostat/Galvanostat/ZRA

Potentiostat / Galvanostat / Impedance Analyzer

Experiment 1: Instrument Familiarization (8/28/06)

Experiment 1: Instrument Familiarization

Potentiostat / Galvanostat / Impedance Analyzer

multi-channel Potentiostat / Galvanostat / Impedance Analyzer Rev

Calsytech, # 38 North Mada Street Nandambakkam, Chennai, Tamil Nadu. Discipline Electro-Technical Calibration Issue Date

ESA400 Electrochemical Signal Analyzer

Internal Model of X2Y Chip Technology

The Benchmark for Electrochemical Research Instrumentation

Potentiostat / Galvanostat / Impedance Analyzer

EIS Measurement of a Very Low Impedance Lithium Ion Ba ery

PeakForce SECM with Bio-Logic SP-300 Potentiostat

potentiostat/galvanostat/impedance analyser

Passive Probe Ground Lead Effects

Zahner 3/2011. femto-farad Probe

TT-SI MHz Active Differential Probe

Defining High Performance Electrochemistry...

SENSOR AND MEASUREMENT EXPERIMENTS

potentiostat/galvanostat/impedance analyser

Ten problems to solve before 6 December, 2010

EIS measurements on Li-ion batteries EC-Lab software parameters adjustment

Lithium-ion batteries are emerging as the de facto. battery.

ET1210: Module 5 Inductance and Resonance

Tender Document. National Centre of Excellence in Analytical Chemistry, ISSUED BY: University of Sindh, Jamshoro.

Methods for Reducing Emissions from Switching Power Circuits. A. McDowell, C. Zhu and T. Hubing

Suitability of the INPHAZE impedance analyzer for Bioimpedance

EE2210 Laboratory Project 1 Fall 2013 Function Generator and Oscilloscope

USER MANUAL. DC 13kHz High Voltage Amplifier WMA V to +175V output. Stable with all capacitive loads, generates no overshoot

University of Pennsylvania Department of Electrical and Systems Engineering ESE319

WHALETEQ. ESU Neutral Electrodes Impedance Tester. Model: HFPA150. User Manual

Setting up a Multi sine impedance measurement

High Power Monolithic OPERATIONAL AMPLIFIER

EE 230 Lab Lab nf C 2. A. Low-Q low-pass active filters. (a) 10 k! Figure 1. (a) First-order low-pass. (b) Second-order low-pass.

Homework Assignment 12

Laboratory 4: Amplification, Impedance, and Frequency Response

AP Physics C. Alternating Current. Chapter Problems. Sources of Alternating EMF

Opamp stability using non-invasive methods

ECE 3155 Experiment I AC Circuits and Bode Plots Rev. lpt jan 2013

Experiment 8 Frequency Response

High Power Monolithic OPERATIONAL AMPLIFIER

Background (What Do Line and Load Transients Tell Us about a Power Supply?)

Proper Termination of Digital Incremental Encoder Signals

BENCH METER Model>9803. Wavecom Instruments

Lab 3: AC Low pass filters (version 1.3)

PCI4 Potentiostat/Galvanostat/ZRA. Operator's Manual

When you have completed this exercise, you will be able to determine the frequency response of an

ECE 2274 Lab 2. Your calculator will have a setting that will automatically generate the correct format.

ECE 2274 Lab 2 (Network Theorems)

A potentiostat is an electronic instrument that controls the voltage between two electrodes

AC Magnitude and Phase

Lecture 16 Date: Frequency Response (Contd.)

Lab 9: Operational amplifiers II (version 1.5)

PIEZO FILM LAB AMPLIFIER

When input, output and feedback voltages are all symmetric bipolar signals with respect to ground, no biasing is required.

SCOPE OF ACCREDITATION TO ISO 17025:2005 & ANSI/NCSL Z

APPLICATION NOTE 33 Battery Cell Electrochemical Impedance Spectroscopy N4L PSM3750 Impedance Analyzer + BATT470m Current Shunt

Multi-function Gain-Phase Analyzer (Frequency Response Analyzer) Model 2505

Application Note 5525

Testing & Calibration Lab, 204, Diamond Industrial Estate No. 2, Ketki Pada Road, (Near Dahisar Toll Naka), Dahisar (East), Mumbai, Maharashtra

Combined EIS- and Spectro-Electrochemical Absorbance Measurement Experiment. Practical Course 2 C.-A. Schiller

Dual, Current Feedback Low Power Op Amp AD812


SDM3045X Digital Multimeter. DataSheet

Lab 1: Basic RL and RC DC Circuits

Low Cost, General Purpose High Speed JFET Amplifier AD825

MAX2720/MAX2721 Evaluation Kits

Wilcom MODEL T136BGMZW CIRCUIT TEST SET. Operating Instructions

MECH 1100 Quiz 4 Practice

Wilcom MODEL T136BSBZW CIRCUIT TEST SET. Operating Instructions

SDM3055 Digital Multimeter. DataSheet

TETRIS 1000 High Impedance Active Probe. Instruction Manual

The Tuned Circuit. Aim of the experiment. Circuit. Equipment and components. Display of a decaying oscillation. Dependence of L, C and R.

CH Instruments, Inc. Electrochemical Instrumentation

Technical Services Department, BHEL, Piplani, Bhopal, Madhya Pradesh. Discipline Electro Technical Calibration Issue Date

Electrochemical Impedance Spectroscopy

SDM3045X Digital Multimeter. DataSheet

Single Supply, Rail to Rail Low Power FET-Input Op Amp AD820

I-V, C-V and AC Impedance Techniques and Characterizations of Photovoltaic Cells

PHYS 3322 Modern Laboratory Methods I AC R, RC, and RL Circuits

Technology in Balance

PowerAmp Design. PowerAmp Design PAD541 COMPACT POWER OP AMP

MIC915. Features. General Description. Applications. Ordering Information. Pin Configuration. Pin Description. Dual 135MHz Low-Power Op Amp

supplies and frequency converters. BumbleBee is also very effective in fast transient measurements

Testing Power Sources for Stability

E84 Lab 3: Transistor

USER. manual. Falco Systems WMA-100. High Voltage Amplifier DC - 500kHz

Understanding, measuring, and reducing output noise in DC/DC switching regulators

PHYS 235: Homework Problems

Precision OPERATIONAL AMPLIFIER

Transcription:

Quick Check of EIS System Performance Introduction The maximum frequency is an important specification for an instrument used to perform Electrochemical Impedance Spectroscopy (EIS). The majority of EIS manufacturers specify a maximum frequency of 1 MHz. Some manufacturer s specifications, however, misrepresent true system performance. Many EIS instruments specified to measure impedance at 1 MHz can apply a 1 MHz signal and measure a response -- but have huge errors when they do so. This document is a protocol to quickly assess the performance of any EIS instrument. You can use it as an easy and effective comparison between systems. You should test before you buy! The test routine involves measuring five EIS spectra. Three spectra are recorded using resistors with ideal EIS response. Two other spectra probe the limits of an EIS system. One spectrum is recorded in potentiostatic mode on a cell that approximates infinite impedance. The second spectrum is recorded in galvanostatic mode on a cell with close to zero impedance. Gamry Instruments will be happy to provide three ± 1% accurate resistors and a piece of braided copper for the near-zero impedance test. Simply call or email us and ask for the EIS Test Resistors. All high quality EIS systems should allow 4-terminal connections to the cell. In this common scheme, four leads are connected to the sample being tested. Two leads apply current and two leads measure voltage. Impedance is calculated by dividing the measured voltage by the current. EIS spectra cannot be measured accurately without a 4- terminal connection. What do you do if your system cannot make 4-terminal connections? Connect the Reference lead to the Counter electrode lead in any convenient manner. Don t expect high quality results. This document contains detailed cable positioning information. Improper connections to a resistor will add capacitive or inductive impedance to the ideal resistive impedance. If you follow the cable positioning shown here, the resistor spectra should have less than 2 of phase shift up to 1 MHz. 1 kω Resistor The 1 kω resistor is marked with 3 brown and 2 black colored bands: (brown, black, black, brown, brown) Hooking up to the 1 kω resistor is easy -- see the figure below. Connect the Working and Working Sense leads to one side of the resistor and the Reference and Counter leads to the other side. Keep the leads about 4 cm apart to minimize capacitive coupling between the leads. Gamry s cell cable color codes are shown in the table below. Color Lead Function Green Working Electrode Current Flow Blue Working Sense Voltage Sense White Reference Electrode Voltage Sense Red Counter Electrode Current Flow Measure the resistor s EIS spectrum using 20 mv rms AC and zero DC voltage. Test frequency should extend from 0.2 Hz to the system s maximum frequency.

100 Ω Resistor The 100 Ω resistor is marked with 3 black and 2 brown colored bands: (brown, black, black, black, brown) Lead placement is critical for a non-inductive spectrum of the 100 Ω resistor. The Reference and Working Sense leads should be twisted together and attached very close to the resistor s body. The Counter and Working leads are also twisted together. They approach the resistor from the other side as shown below. Measure the resistor s EIS spectrum using 20 mv rms AC and zero DC voltage. Test frequency should extend from 0.2 Hz to the system s maximum specified frequency. 10 kω Resistor The 10 kω resistor is marked with colored bands: (brown, black, black, red, brown) This is the only resistor with a red band. Connect the Working and Working Sense leads to one side of the resistor and the Reference and Counter leads to the other side. Keep the leads far apart to minimize capacitance across the resistor. We have measured 0.25 pf capacitance between alligator clips 10 cm apart -- enough to cause a 0.9 phase error at 1 MHz. The figure below shows a very low capacitance connection. Record the potentiostatic EIS spectrum of this resistor with zero DC and 20 mvrms AC voltage. The frequency range should extend from 0.2 Hz to the system s maximum specified frequency. Open Lead Test The ideal cell for this test has infinite impedance. A large air-gap between the cell leads approximates this ideal. Connect the Working and Working Sense leads together and place them inside a Faraday Cage. The Cage must be connected to the potentiostat s ground. Connect the Reference electrode and Counter leads together. With a Gamry EIS system, these leads can be left outside of the Faraday Cage. NOTE: With other systems, the Counter and Reference electrode leads outside the Faraday Cage may cause excessive noise. If this is true, move these leads inside the Cage. The separation between the pairs of leads should be at least 20 cm. Record the potentiostatic EIS spectrum of this open lead configuration. We recommend testing with 50 mv rms AC amplitude and zero DC voltage. The frequency range should extend from about 25 mhz to the system s maximum frequency. Shorted Lead Test The ideal cell for this test is zero impedance. The cell leads shorted together approach this ideal. In shorted lead tests the placement of the cell leads and the way the leads are connected is critical. NOTE: Clipping the cell leads to a piece of braided copper gives you a good low impedance connection. Clipping the alligator clips to each other always gave us more than 500 µω of contact resistance. Twist the Reference and Working Sense leads together and clip them to one side of the braided copper. Twist the Counter and Working leads together and clip them to the opposite side of braid. See the figure below for the lead placement.

The Working Sense and Reference leads are the inside pair of connections. Don t allow the clip leads to touch except at the braid connections. recorded with a poor EIS system. This EIS system claims operation at 1 MHz. At 1 MHz, the errors are greater than 40% and 30. This is unacceptable. NOTE: Shorted lead measurements must be made using galvanostatic cell control. Potentiostatic mode does not work. If your EIS system does not offer galvanostatic mode skip this test! Record the galvanostatic EIS spectrum of the shorted lead configuration using an AC current that is about ½ of the instrument s rated current. Use zero DC current. The frequency range should extend from 0.1 Hz to about 20 khz. Resistor Results The Bode Plot below shows EIS spectra of the three resistors measured with a Gamry Reference 600 Potentiostat. The circles are magnitude and the crosses are phase. Open Lead Results The open lead impedance represents the highest impedance that an EIS system will measure. All spectra measured on real-world samples consist of the sample s impedance in parallel with the open lead impedance. When you compare open lead spectra recorded on different EIS systems, higher impedance is better. The magnitude of an ideal resistor s impedance is independent of frequency and the phase is zero. In the spectra above, the magnitude variation is less than ± 1% and the phase angle is less than ± 2 at all frequencies. The following figure shows spectra recorded on EIS systems from different vendors. The cell was 10 kω in both tests. Again, the circles are magnitude and the crosses are phase. The spectrum in blue was measured using a Gamry Instruments Reference 600. The red spectrum was The Bode Plot above is an open lead spectrum measured using a Gamry Reference 600 Potentiostat. Over a wide frequency range, this looks like the spectrum of a small capacitor. A capacitor model was fit to the region between 0.1 Hz and 100 khz. The calculated capacitance was 0.16 pf ± 0.0004 pf.

This system s impedance at low frequency exceeds 10 TΩ (1 x 1013 Ω). Many EIS systems have low frequency open lead impedance below 100 GΩ. This will severely limit their ability to measure high impedances. Shorted Lead Results Shorted lead impedance is the lowest impedance that an EIS system can measure. All spectra measured on real-world samples consist of the sample s impedance in series with ideal shorted lead impedance. When you compare shorted lead spectra recorded on different systems, lower impedance is better. The Bode Plot below is a shorted lead spectrum measured using a Gamry Instruments EIS system with a Reference 600 Potentiostat. The cell leads were clipped to a piece of copper braid. The AC current was 300 ma rms. The 3.6 mm resistive path seems reasonable. NOTE: This shorted lead spectrum was measured with the standard 60 cm cell cable provided with every Reference 600 Potentiostat. Significantly lower inductance will be measured using Gamry s special Low Z cable for the Reference 600. The shorted lead test illustrates some of the experimental difficulty in measuring very small impedance. You must carefully separate the current paths and sense paths in your 4-terminal connection. You cannot ignore the resistance of metal connections. Shorted lead curves measured on other EIS systems will probably look similar. The resistance and inductance values are likely to be higher than those from the above spectrum. If you use shorted lead spectra to compare two EIS systems, make sure that you are comparing the two systems and not the quality of their shorted lead connections. Conclusions This test protocol can quickly check the function of an EIS system. It will quickly reveal marginal EIS systems, which typically exhibit very poor frequency response and unsatisfactory open lead and shorted lead spectra. These tests also demonstrate the ability of an EIS system to measure very small or very large impedances. Knowledge of these limitations is very important. It can help you distinguish instrument response from chemistry. This spectrum fits well to a series RL model with a resistance of 30 µω and inductance of 42 nh. The resistance value is interesting. The copper braid used to make the shorted connection has a measured resistance of 8.4 µω per mm. 30 µω corresponds to a path of 3.6 mm through the braid. Quick Check of EIS System Performance Technical Note, Rev. 1.0 12/9/2010 Copyright 2009 Gamry Instruments, Inc. 734 Louis Drive Warminster PA 18974 Tel. 215 682 9330 Fax 215 682 9331 www.gamry.com info@gamry.com