The physics of ultrasound. Dr Graeme Taylor Guy s & St Thomas NHS Trust

Similar documents
12/26/2017. Alberto Ardon M.D.

The Physics of Echo. The Physics of Echo. The Physics of Echo Is there pericardial calcification? 9/30/13

Physics of Ultrasound Ultrasound Imaging and Artifacts รศ.นพ.เดโช จ กราพาน ชก ล สาขาหท ยว ทยา, ภาคว ชาอาย รศาสตร คณะแพทยศาสตร ศ ร ราชพยาบาล

Ultrasound Imaging Ultr Michael Dadd 2007

Interaction of Sound and. logarithms. Logarithms continued. Decibels (db) Decibels (db) continued. Interaction of Sound and Media continued

Explain what is meant by a photon and state one of its main properties [2]

Ultrasound Physics. History: Ultrasound 2/13/2019. Ultrasound

Artifacts. Artifacts. Causes. Imaging assumptions. Common terms used to describe US images. Common terms used to describe US images

Ultrasound Beamforming and Image Formation. Jeremy J. Dahl

Lesson 02: Sound Wave Production. This lesson contains 24 slides plus 11 multiple-choice questions.

Lesson 06: Pulse-echo Imaging and Display Modes. These lessons contain 26 slides plus 15 multiple-choice questions.

Ultrasound Bioinstrumentation. Topic 2 (lecture 3) Beamforming

Optimisation of Image Acquisition Bordeaux 16th November J.S. McGhie W.B. Vletter R. Frowijn No disclosures

Ultrasound & Artifacts

Introduction to Ultrasound Physics

Medical Imaging (EL582/BE620/GA4426)

Chapter 4. Pulse Echo Imaging. where: d = distance v = velocity t = time

Fig. 1

INTRODUCTION. Have applications for imaging, detection and navigation.

Ultrasonic Linear Array Medical Imaging System

Medical Imaging. X-rays, CT/CAT scans, Ultrasound, Magnetic Resonance Imaging

Doppler Ultrasound. Amanda Watson.

Photomultiplier Tube

Doppler in Obstetrics: book by K Nicolaides, G Rizzo, K Hecher. Chapter on Doppler ultrasound: principles and practice by Colin Deane

3. Ultrasound Imaging(2)

GraspIT Questions AQA GCSE Physics Waves

Sonic Distance Sensors

Chapter Introduction. Chapter Wrap-Up. and the Eye

Physics of ultrasound

Lesson 06: Pulse-echo Imaging and Display Modes. This lesson contains 22 slides plus 15 multiple-choice questions.

Section 1: Sound. Sound and Light Section 1

REAL-TIME B-SCAN ULTRASONIC IMAGING USING A DIGITAL PHASED. Robert Dunki-Jacobs and Lewis Thomas General Electric Company Schenectady, New York, 12301

used to diagnose and treat medical conditions. State the precautions necessary when X ray machines and CT scanners are used.

PRODUCT 4.06 IMAGE MANAGEMENT

Physics in Modern Medicine Fall 2010

4 Working With Scan Modes

Answer: TGC is needed to amplify echoes from deeper structures so that they appear as bright as similar structures located at more shallow depths.

(A) 2f (B) 2 f (C) f ( D) 2 (E) 2

Introduction to Medical Engineering (Medical Imaging) Ultrasound Imaging. Ho Kyung Kim Pusan National University

Lesson 12: Doppler Principles. This lesson contains 50 slides plus 26 multiple-choice questions.

Image Optimization: The Sonographer s Responsibility. Prepared by Cathy Daniels, EdD, RTR, RDMS, RDCS, RVT

Breast Ultrasound QA Phantom Recommended by Japan Association of Breast and Thyroid Sonology

Attenuation and velocity of ultrasound in solid state materials (transmission)

Principles of Ultrasound Imaging Image Optimization

Biomedical. Measurement and Design ELEC4623/ELEC9734. Electrical Safety and Performance Standards

Ultrasound physical principles in today s technology

Key Physics and Doppler Principles

COMPUTED TOMOGRAPHY 1

Radar Imaging Wavelengths

CHAPTER 1 INTRODUCTION

Equipment for Attenuation and velocity of ultrasound in solid state materials (transmission), experimental set-up

Architecture of Quality Imaging Mary K. Henne, MS, CNMT, RDMS, RVT Ultrasound Education Specialist GE Healthcare

Nuove tecnologie per ecografia ad ultrasuoni: da 2D a 4D

Lecture 19. Ultrasound Imaging

Project = An Adventure : Wireless Networks. Lecture 4: More Physical Layer. What is an Antenna? Outline. Page 1

COMPUTER PHANTOMS FOR SIMULATING ULTRASOUND B-MODE AND CFM IMAGES

Harmonic Source Wavefront Correction for Ultrasound Imaging

Multi-spectral acoustical imaging

P6 Quick Revision Questions

Fiberoptic and Waveguide Sensors

Light and Applications of Optics

SODAR- sonic detecting and ranging

An Overview Algorithm to Minimise Side Lobes for 2D Circular Phased Array

Retrospective Transmit Beamformation. Whitepaper. ACUSON SC2000 Volume Imaging Ultrasound System. Answers for life.

Physics B Waves and Sound Name: AP Review. Show your work:

4/18/2012. Supplement T3. 3 Exam Questions, 3 Groups. Amateur Radio Technician Class

4.6 Waves Waves in air, fluids and solids Transverse and longitudinal waves

Spectral Distance Amplitude Control for Ultrasonic Inspection of Composite Components

Beamforming in ultrasound

4.6.1 Waves in air, fluids and solids Transverse and longitudinal waves Properties of waves

27/11/2013' OCEANOGRAPHIC APPLICATIONS. Acoustic Current Meters

Range Sensing strategies

AQA P3 Topic 1. Medical applications of Physics

X rays X-ray properties Denser material = more absorption = looks lighter on the x-ray photo X-rays CT Scans circle cross-sectional images Tumours

Physics of Ultrasound & Doppler. Sang Jae Rhee. MD., PhD. Division of Cardiovascular Medicine Wonkwang University Hospital

Ultrasound Physics and Instrumentation, 5e Chapter 7: Level 1 Quiz Answers. 1) Which of the following is stated in the introduction of the chapter?

The Human Brain and Senses: Memory

CHAPTER ONE SOUND BASICS. Nitec in Digital Audio & Video Production Institute of Technical Education, College West

Multi-Element Synthetic Transmit Aperture Method in Medical Ultrasound Imaging Ihor Trots, Yuriy Tasinkevych, Andrzej Nowicki and Marcin Lewandowski

ATS 351 Lecture 9 Radar

SONOGRAPHIC PHYSICS, INSTRUMENTATION & DOPPLER REVIEW Part 3

Underwater Acoustics Research

An Activity in Computed Tomography

Module 5: Experimental Modal Analysis for SHM Lecture 36: Laser doppler vibrometry. The Lecture Contains: Laser Doppler Vibrometry

Introduction. Chapter 16 Diagnostic Radiology. Primary radiological image. Primary radiological image

Introduction. Parametric Imaging. The Ultrasound Research Interface: A New Tool for Biomedical Investigations

Episode 323: Diffraction

BEAM DISTORTION IN DOPPLER ULTRASOUND FLOW TEST RIGS: MEASUREMENT USING A STRING PHANTOM

Antennas and Propagation

Light sources can be natural or artificial (man-made)

Life Science Chapter 2 Study Guide

Phased Array Velocity Sensor Operational Advantages and Data Analysis

LlIGHT REVIEW PART 2 DOWNLOAD, PRINT and submit for 100 points

ACOUSTIC MICRO IMAGING ANALYSIS METHODS FOR 3D PACKAGES

An Activity in Computed Tomography

Wave Review Questions Updated

Ques on (2): [18 Marks] a) Draw the atrial synchronous Pacemaker block diagram and explain its operation. Benha University June 2013

Instructional Resources/Materials: Light vocabulary cards printed (class set) Enough for each student (See card sort below)

Pass Ultrasound Physics Exam

Transcription:

The physics of ultrasound Dr Graeme Taylor Guy s & St Thomas NHS Trust

Physics & Instrumentation Modern ultrasound equipment is continually evolving This talk will cover the basics

What will be covered? The basic properties of ultrasound Generating and detecting it The common imaging modalities A, B & M mode Spectral Doppler Colour flow & Power Doppler Safety issues

What is ultrasound? Just a normal sound wave but at a higher frequency than we can hear (>1 MHz) It is a pressure wave which travels through a medium In medical imaging, longtitudinal waves are commonly used

What are its major properties? It travels through a medium with a characteristic wave velocity It reflects off interfaces between media of different acoustic impedance It refracts (bends) when passing obliquely between media of different sound velocity It scatters in media containing small objects It is absorbed by most media

Sound velocities Air about 300 m/s Water about 1450 m/s Soft tissue about 1560 m/s Fat, muscle, blood all have slightly different velocities Bone about 3000 m/s It takes about 6 μsec for sound to travel through 1 cm of soft tissue

Sound reflection When sound travels across a junction between two media, some bounces back and some travels onward This depends on the product of the density ρ and sound velocity c of each medium ( its impedance = ρc) As the impedances differ more, so more is reflected and less transmitted. No sound travels across air-tissue boundary

Sound refraction Unlike X rays sound does not always travel in a straight line though imagers assume that is what s happening When passing at an oblique angle between media with different velocity, the sound can change direction (like light) This can distort the ultrasound image Important when using it to guide interventions!

Scattering Scatter (sound bouncing off in all directions) occurs when a medium contains lots of small (<1 wavelength) items of a different impedance. This is what gives tissues their greyness on an ultrasound image (Wavelength at 1 MHz = 1.5mm)

Absorption This is where energy is lost to the tissues It means deeper structures return weaker echoes Soft tissue absorption is approximately proportional to frequency Bone has high absorption, nearly proportional to the square of frequency To image deep structures use lower frequency

Absorption the consequences All imagers need to compensate for absorption with depth Imaging close to the surface can use higher frequencies (up to 10 MHz) Large folk need to be imaged at lower frequency Imaging through bone (skull) is only practicable at 2 MHz or less Resolution Depth compromise

Imagers some available modes B Scan Colour flow mapping Spectral Doppler

Imagers not forgetting M mode B Scan M mode Used to visualise the movement of moving reflecting objects

The basis of imaging - echo ranging The fundamental principle is that of Directional pulse-echo location The distance to a sound reflector is proportional to the delay in the echo Depth α delay time = distance / sound velocity

Instrumentation A Scan To make a successful A (amplitude) Scan tracing one needs: a short ultrasound pulse a narrow ultrasound beam A scan Echo Amplitude Time

Instrumentation B Scan To make a successful B (brightness) Scan image one also needs: a way to steer/move beam to cover a two dimensional area All scanners assume some fixed sound velocity is it correct for your patient

What frequency to use? Shorter sound pulses have better resolution but higher frequency content So the shortest pulse is limited by the absorption of sound in tissue To obtain echoes from deep structures one needs to use low frequency pulses 1-4 cm deep use 7-12 MHz 10-20 cm deep use 2-5 MHz (Guide depth resolution ~ 2 wavelengths)

How big a transducer? The beam width is defined by the shape of the active transducer A small transducer has a narrow beam close up, but widens out & weakens quickly A trade off between resolution and penetration

Instrumentation Swept gain Swept gain or depth gain compensation must be used to compensate for quieter distant echoes Echo amplitude Depth This is inbuilt in all scanners but assumes some average attenuation which may be wrong Gain increased with time after transmission

Practical Transducers The all electronic transducer is the basis of the modern ultrasound B scanner No moving parts Array of small elements Wide frequency range Electronically controlled

Instrumentation - transducers The all electronic transducer is a large array of small elements Each is individually wired to the scanner

Transducers Each element is a simple piezo-electric transducer electrical sound electrical They are not individually directional They must work in unison

Instrumentation - transducers The transducer is made directional by a beam former an electronic process which synchronously controls the action of many of the independent elements simultaneously To make a simple forward looking directional beam - just use plenty of elements together

Instrumentation linear array To model a rectilinear scanner just repeat pulse-echo cycle with different sets of active elements Linear array

Instrumentation phased array To model a sector transducer, one only needs to add short time delays between firing elements and also on receiving ehoes. sector This effectively bends the beam Repeating the pulse-echo sequence with differing delays gives a sector scan

Focal zones It is possible to narrow the beam width at a certain depth for better resolution This is done by adding symmetrical time delays to inner elements in the group selected when transmitting and receiving Greater delay difference = closer focus

Multiple focal zones Can optimise beam width of pulse for a given depth range (the focal zone) To improve resolution at various depths, make separate images, each with different focal zone placement, then cut-&-paste This reduces the frame rate

Real-time compounding If one can CHANGE DIRECTION of the transducers line-of-sight, one can interrogate tissue from different directions & combine to give better image Removes speckle improves echoes from non-aligned specular reflectors

Image showing advantage of compounding Conventional image of breast (Speckle & shadow) With compounding SonoCT

transducer bandwidth A transducer s bandwidth refers to the range of frequencies which it can transmit & receive Older transducers typically had limited bandwidths This limited their use to a fixed depth range and resolution F

wide bandwidth Wideband transducers have been developed, along with scanners which control the range of frequencies transmitted and received at any one time Vary receive bandwidth with depth of echo One transducer - various centre frequencies Harmonic imaging - Send at low frequency, receive at 1 st harmonic

Modern image possibilities

Spectral Doppler Basic B Scan techniques do not work well with blood, it appears dark and it travels too fast Movement sensitive techniques are needed Spectral Doppler Colour Flow Mapping Power Doppler Christian Andreas Doppler

Spectral Doppler Red cells scatter back a small proportion of incident ultrasound MOVING cells will reflect ultrasound with a DOPPLER SHIFT in FREQUENCY This small shift is proportional to the red cell velocity θ

Spectral Doppler If the transmit frequency is in the low MegaHertz range, then the Doppler-shift frequency for blood flow is in the audio range Basic B scan equipment can be adapted, and it is helped if the transducer has a wide bandwidth and the beam can be steered to get a good Doppler angle Usually Pulsed Doppler is provided

Duplex - Spectral Doppler Spectral Doppler techniques use similar pulses, beam shaping & steering methods to Bscan The basic difference is that the pulse-echo process is repeated over and over, but for just one line-of-site.

Duplex Spectral Doppler These two techniques are combined with a method for displaying the sound spectrum - a sonogram Note it is common to show the vertical axis in cm/sec though it really is only known in khz!

Duplex - Colour flow The data is colour coded according to frequency (velocity) and superimposed on the B scan image Be aware of direction of flow detection and aliasing

Duplex - Power Doppler Velocity and direction can sometimes confuse, and one can choose Power Doppler mode This displays the areas where flow is detected but with the brightness proportional to the POWER of the movement detected. Good for slow flow and when vessels hard to see on B scan

Some safety issues Thermal & mechanical effects Significant deleterious bio-effects on either patients or operators of diagnostic ultrasound procedures have not been reported in literature ALARP (as low as reasonably practicable) MI mechanical index (likelihood of cavitation) TI thermal index (1.0 = 1ºC temperature rise)

Some safety issues Use BMUS GUIDELINES FOR THE SAFE USE OF DIAGNOSTIC ULTRASOUND EQUIPMENT Eg: avoid - an embryo less than eight weeks after conception; - the head, brain or spine of any foetus or neonate; - an eye (in a subject of any age).

Imaging summary B SCAN pulse echo, short pulses, displays anatomy M MODE displays mechanical motion SPECTRAL DOPPLER displays flow signal at one point COLOUR FLOW displays flow over image area