ABB ph/orp analyzer and probe package

Similar documents
ABB conductivity analyzer and probe package

AX460, AX466 and AX416 Single and dual input analyzers for ph/redox (ORP)

ABB MEASUREMENT & ANALYTICS DATA SHEET. AX460, AX466 and AX416 Single and dual input analyzers for ph/redox (ORP)

Analyze IT Single and Dual Input Analyzers for ph/redox (ORP)

Analyze IT Single and Dual Input Analyzers for ph/redox (ORP)

Conductivity Analyzer

ABB MEASUREMENT & ANALYTICS DATA SHEET. AX410, AX411, AX416, AX450 and AX455 Single and dual input analyzers for low level conductivity

Single and Dual Input Analyzers for Low-level Conductivity

ph or ORP Transmitter

Multi-electrode Conductivity Analyzers

AC200/AX400 BROCHURE JUNHO 2013

Model 50 Ground Loop ph Sensor Model 55 Solution Ground ph Sensor

LST400 Ultrasonic level / open channel flow transmitter

GRUNDFOS DATA BOOKLET DIP. Measuring amplifiers and controllers, preassembled systems. Up to 3 parameters (Cl 2, ClO 2, O 3, ph, ORP)

WATER MADE EASY MARINE ENERGY MUNICIPAL INDUSTRIAL

Metal Finishing. W600 Series Controllers. Summary of Key Benefits

General Specifications

General Specifications

OPTISENS PH 8390 Technical Datasheet

Memosens CPS96D. Technical Information

MJK Expert 700 / 800 / 900 / 1100 / 2100 MJK Expert 1400 / 3400

ABB MEASUREMENT & ANALYTICS DATA SHEET. Ultrasonic level LST400 Open channel flow transmitter

OPTISENS ORP 8590 Technical Datasheet

Conex DIA-1, -2, -2Q. Measuring amplifiers and controllers, preassembled systems. One or two parameters (Cl 2, ClO 2, O 3, H 2 O 2, PAA, ph, ORP)

General Specifications

Instruction manual SMART 96-2 / WG-2 ph

Key Number I II III IV V _ - - _ - - _ - _, _,

Technical Information Memosens CPS96D

Corporate Profile. needs and product demands. CON 08X /03 Rev 0

Condumax CLS21 and CLS21D

Maintenance: Check the measurement signal: regularly required once a month Calibration: typically 6 to 12 months after installation

Technical Information Orbipore CPS91D and CPS91

GRUNDFOS ALLDOS DATA BOOKLET DIP. Measuring amplifiers and controllers Preassembled systems 50/60 Hz

For reliable and continuous measurement of water quality insist on Aqua SMARTPro Transmitters

Ceramax CPS341D. Technical Information

Model Q40CT. System Features. Toroidal Conductivity Monitor...

Temperature Transmitters - Head Mounted

Signet 9950 Dual Channel Transmitter

Process ph/orp Analysis Systems

Analytical Transmitters

OPTISENS PH ph sensor. Technical Datasheet

Signet 9950 Dual Channel Transmitter

Temperature Transmitters - DIN Rail Mounted

EXPERT-LINE. KOBO-pH. Transmitter for ph-value and ORP Model APM-X

1000TR. Instructions

Technical Information Orbipac CPF81D/CPF82D and CPF81/CPF82

Orbipac CPF81/CPF82 and CPF81D/CPF82D

Kotron Sentinel Model 804 RF Level Transmitter

OPTISENS PH 8100 Technical Datasheet

CSI 9330 Vibration Transmitter

GRUNDFOS ALLDOS DATA BOOKLET. Conex DIS-PR, DIS-D. Measuring amplifiers and controllers Preassembled systems 50/60 Hz

9330 Vibration Transmitter

Universal Controller

C1901 Single Pen Circular Chart Recorder. C1901 a rugged, reliable recorder for all single channel recording applications

New Technology for Monitoring Cycle Chemistry. David M. Gray McIlvaine Hot Topic Hour February 21, 2013

Conductivity Transmitter

Condumax CLS15 and CLS15D

ph/orp Transmitter SIGNET

MIT510/2, MIT520/2 and MIT1020/2

ph / ORP Measurement mypro CPM 431

Condumax W CLS15 and CLS15D

Screw-in resistance thermometers with plug connector

Remote Controlled FOUNDATION Fieldbus Two-Wire ph/orp Transmitter

Flow Measurement SITRANS F M. Flow sensor MAG /49. Overview

OPTITEMP TT 20 Technical Datasheet

COMPACT-LINE. KOBO-pH Transmitter for ph-value Model APM-Z

General Specifications

The transducer component converts the measured signal and displays the actual value.

Flow Measurement SITRANS F M

Single Input ph/redox (ORP) Analyzer

FAST SAMPLING CONVERTER

OEM pressure transmitter for general industrial applications Models O-10 (T), O-10 (5)

Type F-20-3A Sanitary Pressure Transmitter NEMA 4X with Integral Junction Box

Type PLC Fitting. Valve for Continuous control. Diaphragm valve. Output

OEM radar transmitter, for aggressive media level measurement

SITRANS F flowmeters SITRANS F M. Flow sensor MAG /48 Siemens FI Overview

COMPACT-LINE. KOBO-RP Transmitter for ORP Model ARM-Z

Additel 875 Series Dry Well Calibrators

Measuring and Controlling Unit Multronic OC

COMPACT-LINE. Transmitter for ph-value

phix Compact 4.10 PH, TEMPERATURE AND REDOX MEASUREMENT DATASHEET EN 4.10 PHIX DATASHEET 1812

Signet ph/orp Electrodes General Purpose

OPTIFLUX 5000 Technical Datasheet

Condumax CLS15 and CLS15D

VARIABLE FREQUENCY DRIVE SPECIFICATION

FC Series Signal Conditioners

Features. Picture. Specifications. Pressure Transmitter 4-20mA 2-wire MPG Type: MPGxxxxxxxx for level measurement Order No.: see Table 1 / Page 4

Flow Level Temperature Pressure Valves Analyzers Accessories TekValSys

TEK-TEMP 2100A. Explosion-Proof Temperature Transmitter. TEMPERATURE. Technology Solutions

General Specifications

Digital conductivity transmitter

Electromagnetic Flowmeter Hygienic/Sanitary

JUMO ecotrans Lf 03 Microprocessor Transmitter / Switching Device for conductivity or resistivity and temperature

Signet 9900 Transmitter

Multi-Electrode Mounting (used for ph, Orp and Specific Ion) Honeywell Proprietary

Miniature resistance thermometer Model TR33, thread-mounted

Condumax CLS21 and CLS21D

P5201 Universal Programmable Transmitters with Galvanic Isolation

SMARTPAT PH 1590 Technical Datasheet

S8000 Series ph/orp Electrode Platform. Self Cleaning Flat Surface ph/orp Electrodes LIQUID ANALYTICAL SOLUTIONS THE NEXT GENERATION

Transcription:

Water Technologies & Solutions fact sheet ABB ph/orp analyzer and probe package Series AX466 dual input analyzer description and use AX466 analyzers incorporate the latest technology to provide highly reliable, flexible, feature-packed devices that satisfy a diverse range of process monitoring and control applications. AX466 is used with glass, enamel and antimony ph electrodes and metal Redox sensors, providing measurements with exceptional accuracy and performance. reduced maintenance through in-line cleaning Many installed ph and ORP sensors benefit from regular cleaning. When carried out manually, this can be laborious and costly so is often ignored. Advanced AX466 ORP analyzers make it possible to automate sensor cleaning by using a programmable wash-timer contact to activate a cleaning agent valve or pump. Pulsed water-wash and jet-wash chemical clean actions are standard, user-selectable functions. The frequency and duration of the wash can be tuned to meet the requirements of the application. current output proportional control In this case, the control signal is provided by the 4 to 20mA output of the AX466 analyzers rather than using the alarm/control contacts. The magnitude of the current output changes in proportion to the deviation from set point. This method is used generally with motorized valves or when controlling the speed of dosing pumps with a 4 to 20mA proportional input. specifications - ORP Input resistance Reference 1 x 1013 ohms Range 1200 to +1200mV Minimum span 100mV Temperature compensation modes Automatic or manual Nernstian compensation Range 10 to 200 C (14 to 392 F) Process solution compensation with configurable coefficient Range 10 to 200 C (14 to 392 F) adjustable 0.05 to +0.02%/ C ( 0.02 to +0.009%/ F) Temperature sensor Programmable Pt100 Find a contact near you by visiting www.suezwatertechnologies.com and clicking on Contact Us. *Trademark of SUEZ; may be registered in one or more countries. 2017 SUEZ. All rights reserved. AM-FScfABB_pH_ORP_Package_EN.docx Jul-12

Slope Between 40 and 105% (low limit user-configurable) Calibration with auto-stability checking Automatic 1 or 2 point calibration selectable from: ABB DIN Merck NIST US Tech 2 x user-defined buffer tables for manual entry, 2-point calibration or single-point process calibration specifications - ph Input resistance Glass >1 x 1013 ohms Reference 1 x 1013 ohms Range 2 to 16pH Minimum span Any 2pH span Resolution 0.01pH Accuracy 0.01pH Temperature compensation modes Automatic or manual Nernstian compensation Range 10 to 200 C (14 to 392 F) Process solution compensation with configurable coefficient Range 10 to 200 C (14 to 392 F) Adjustable 0.05 to +0.02%/ C ( 0.02 to +0.009%/ F) Temperature sensor Programmable Pt100 Calibration Ranges 0 to 14pH Slope Between 40 and 105% (low limit user-configurable) number of relays Three supplied as standard Relay contacts Single-pole changeover Rating 5 A, 115/230 V AC, 5 A DC Insulation 2 kv RMS contacts to earth/ground analog outputs Number of current outputs (fully isolated) Two supplied as standard Output range 0-10 ma, 0-20 ma, or 4 20 ma analog output programmable to any value between 0 and 22 ma to indicate system failure Accuracy ±0.25% FSD, ±0.5% of reading (whichever is the greater) Resolution 0.1% at 10 ma, 0.05% at 20 ma Maximum load resistance 750 ohms at 20 ma power supply Voltage requirements 100 to 240 V AC 50/60 Hz Page 2 AM-FScfABB_pH_ORP_Package_EN.docx

(90 V Min. to 264 V Max. AC) 12 to 30 V DC Power consumption 10 W Insulation Mains to earth (line to ground) 2 kv RMS environmental data Operating temperature limits 20 to 55 C ( 4 to 131 F) Storage temperature limits 25 to 75 C ( 13 to 167 F) Operating humidity limits Up to 95% RH non condensing approvals, certification and safety Safety approval UL CE Mark CE Mark Covers EMC & LV Directives (including latest version EN 61010) General safety EN61010-1 Overvoltage Class II on inputs and outputs languages configurable English French German Italian Spanish AM-FScfABB_pH_ORP_Package_EN.docx Page 3

overall dimensions Page 4 AM-FScfABB_pH_ORP_Package_EN.docx

AP201pH/ORP sensor system The AP201 ph/orp system is designed to provide high reliability and withstand the toughest environments for process monitoring and control. The rugged assembly is built to bear the rigours of weather and process. Parts in contact with the process comprise of 316 stainless steel Insertion systems tolerate temperatures up to 130 C (266 F) and pressures up to 6 bar (90 psi). The inner electrode connections are ingress-protected to IP67/NEMA 6P (exceeds NEMA 4X). AP121/6 Pt Redox (ORP) ± 2000mV 0 to 130 C (32 to 266 F) -AP121/0 General Process 0 to 14 ph 0 to +100 C (32 to 212 F) The heartbeat of the AP201 System is the AP120 Series combination electrode. With high process integrity it delivers low running costs and low replacement outlay. just-in-time maintenance One of the most vulnerable times for a ORP electrode is during reintroduction into the line. Analyzers with advanced diagnostic facilities, such as ABB's AX466 Series, can detect sensor breakages as they occur. In addition, they can detect sensors 'out-of-process' and several reference faults. These valuable facilities enable just-in-time maintenance. The AP201 holder is equipped with an integral solution earth facility. This provides high stability of measurement and enables the transmitter to supply full sensor diagnostics. ph/orp sensor series AP120 with rapid temperature response for critical process The AP121 is a series of highly process-resistant 12mm combination ORP electrodes used in the AP201 System. Some of the benefits available to users are: Exceptional resistance to poisoning and pressure IP67 (NEMA 6P) Ingress Protection High corrosion resistance and steam sterilizability Assured high accuracy of calibration and operation Rapid response enabling tighter process control Unparalleled resistance to plugging Extended life and improved performance exceptional process resistance AP121 electrodes are equipped with a rugged solid electrolyte providing effective resistance to process poisons and fluctuating pressures. This far outstrips the performance of lesser, gelled-slurry type electrodes. The sensor is also backed with an annular, PTFE junction giving excellent dirt-repellence. ORP Probe ph probe reduced maintenance through in-line cleaning 50% of ph sensors in chemical processes benefit from regular cleaning. When done manually, this can be laborious and costly so is often ignored. AM-FScfABB_pH_ORP_Package_EN.docx Page 5

The AP201 makes it possible to automate sensor cleaning, driven by advanced Analyzers such as ABB's AX466. The AP201 system earth rod also serves as a jet-spray head enabling the user to supply a periodic burst of cleaning agent across the sensor. This is delivered through two nozzles, one directed at the measuring element and the other at the reference junction. An aggressive, but little-andoften approach, keeps the sensor in prime condition, even in gypsum-laden scrubber applications. Generally, water is a good solvent but, on other occasions, use of an acid, alkali or even detergent is beneficial. Process Connections Stainless steel adapters 1 in. NPT Length of Cable 10 m (33 ft) specifications Materials Shaft and cap Polypropylene Sensor body Ryton* PPS Ground rod/spray tube 316 Stainless steel Jet-wash facility Non-return function Integral one-way valve Spray tube connection 6mm compression fitting SUEZ part numbers 3064317 Analyzer Package with probes and cables 3063661 ph Probe Assembly Only with cable 3063662 ORP Probe Assembly Only with cable 3063654 Analyzer Only 3095360 Replacement ORP electrode only 3095326 Replacement ph electrode only Maximum Sensor Temperature 130 C (266 F) PPS and steel adapters Maximum Sensor Pressure 6 bar (90 psi) @ 25ºC (77ºF) Page 6 AM-FScfABB_pH_ORP_Package_EN.docx