SEM OPERATION IN LOW VACUUM MODE

Similar documents
JEOL 6500 User Manual

Using the Hitachi 3400-N VP-SEM

JEOL 6700 User Manual 05/18/2009

Dickinson College Department of Geology

JSM 6060 LV SCANNING ELECTRON MICROSCOPE STANDARD OPERATING PROCEDURES

Scanning Electron Microscope FEI INSPECT F50. Step by step operation manual

OPERATION OF THE HITACHI S-450 SCANNING ELECTRON MICROSCOPE. by Doug Bray Department of Biological Sciences University of Lethbridge

Standard Operating Procedure for the Amray 1810 Scanning Electron Microscope Version: 29 NOVEMBER 2014

Title: Amray 1830 SEM#2 Semiconductor & Microsystems Fabrication Laboratory Revision: D Rev Date: 03/18/2016

Operating Checklist for using the Scanning Electron Microscope, JEOL JSM 6400.

Check that the pneumatic hose is disconnected!!!! (unless your using the BSE detector, of course)

Operating Checklist for using the Scanning Electron. Microscope, JEOL JSM 6400.

FE-SEM SU-8020 Operating manual (Preliminary version)

1.3. Before loading the holder into the TEM, make sure the X tilt is set to zero and the goniometer locked in place (this will make loading easier).

User Operation of JEOL 1200 EX II

JEOL 5800LV SCANNING ELECTRON MICROSCOPE OPERATOR'S MANUAL

SEM Training Notebook

Operating the Hitachi 7100 Transmission Electron Microscope Electron Microscopy Core, University of Utah

2. Raise HT to 200kVby following the procedure explained in 1.6.

SEM Training Notebook

Please follow these instructions for use of the Philips CM100 TEM. Adopted from website below.

General information. If you see the instrument turned off, notify MIC personnel. MIC personnel will help you insert your samples into the instrument.

1. Preliminary sample preparation

Basic Operating Instructions for Strata Dual Beam 235 FIB/SEM

1.2. Make sure the viewing screen is covered (exposure to liquid N 2 may cause it to crack).

Basic Users Manual for Tecnai-F20 TEM

Instructions for Tecnai a brief start up manual

Model SU3500 Scanning Electron Microscope

JEOL JEM-1400 Transmission Electron Microscope Operating Instructions

RAITH e-line OPERATING INSTRUCTIONS

1.1. Log on to the TUMI system (you cannot proceed further until this is done).

Jeol JEM Responsible personell: Endy ( ) Online booking is compulsory!

Amray 3600 FESEM. Standard Operating Procedure. v2.2 modified by Bryan Cord. General Notes...3. Sample Loading...5. System Loading...

2 How to operate the microscope/obtain an image

MSE 460 TEM Lab 2: Basic Alignment and Operation of Microscope

Name EET 1131 Lab #2 Oscilloscope and Multisim

Full-screen mode Popup controls. Overview of the microscope user interface, TEM User Interface and TIA on the left and EDS on the right

1. Specimen Holder Removal, Loading, and Insertion

05/20/14 1. Philips CM200T. Standby Condition

STANDARD OPERATING PROCEDURE: JEOL TEM-2100

COMPACT MANUAL FOR GI USERS OF THE JEM 1400 FLASH BEGINNERS (For internal use only) Gray means additional information at the end of this mini-manual

JEOL 5DII. Operation introduction. By Serge Charlebois

Standard Operating Procedure Hitachi UHR CFE SU8230 SEM

SOP for Hitachi S-2150 Scanning Electron Microscope For review purposes only

How to use the Jeol 1010 TEM of GI (Liesbeth own GI version)

Tinker Tuesday Project - Drinking Glasses

CAPTURING IMAGES ON THE HIGH-MAGNIFICATION MICROSCOPE

CM20 USER GUIDE. Duncan Alexander, CIME 2010

JEOL JEM 2010 TRAINING TRANSMISSION ELECTRON MICROSCOPE USER MANUAL

Zygo Optical Interferometer SOP

Protective Equipment Nitrile gloves for handling sample holder and safety glasses for filling liquid nitrogen dewar.

LEO 912 TEM Short Manual. Prepared/copyrighted by RH Berg Danforth Plant Science Center

FEI Quanta 200 ESEM Basic instructions

ZEISS EVO SOP. May 2017 ELECTRON OPTICS

MSE 595T Transmission Electron Microscopy. Laboratory III TEM Imaging - I

Procedures for Performing Cryoelectron Microscopy on the FEI Sphera Microscope

MEASUREMENT CAMERA USER GUIDE

Operating Procedures for MICROCT1 Nikon XTH 225 ST

STEM alignment procedures

These aren t just cameras

Operation Guide. Hitachi S-3400N. Variable Pressure Scanning Electron Microscope. with. Deben Peltier Coolstage

Tinker Tuesday Project - Fabric Engraving

SAMUEL ROBERTS NOBLE ELECTRON MICROSCOPY LABORATORY. Operating Procedures for the Zeiss 9 S-2. Transmission Electron Microscope

SOP: EDAX Eagle III Microspot XRF

Tinker Tuesday Project - Stainless Steel Tumblers

Strata DB235 FESEM FIB

JEOL 2010 FasTEM & DigitalMicrograph User's Guide

Figure 1 The Raith 150 TWO

Workflow for Betterlight Imaging

CATHODOLUMINESCENCE IMAGING UNIVERSITY OF ST ANDREWS

This document assumes the user is already familiar with basic operation of the instrument in TEM mode and use of the Microscope Control interface.

Section 1: TEM parts and functions... 2

English PRO-642. Advanced Features: On-Screen Display

Functions of the SEM subsystems

LLS - Introduction to Equipment

Gardall Locks Operating Instructions

WITec Alpha 300R Quick Operation Summary October 2018

Photo Editing in Mac and ipad and iphone

Tinker Tuesday Project - Pumpkin Carving

Characterization Microscope Nikon LV150

Contents STARTUP MICROSCOPE CONTROLS CAMERA CONTROLS SOFTWARE CONTROLS EXPOSURE AND CONTRAST MONOCHROME IMAGE HANDLING

Nanoscale Fabrication & Characterization Facility. Raith e-line EBL Users Guide (updated:aug 2 nd, 2017)

Zeiss AxioImager.Z2 Brightfield Protocol

Bruker Dimension Icon AFM Quick User s Guide

NOTES ON PRINTING 08/06

CNC Using the FlexiCam CNC and HMI Software. Guldbergsgade 29N, P0 E: T:

Olympus Digital Microscope Camera (DP70) checklist

DISCO DICING SAW SOP. April 2014 INTRODUCTION

12. Creating a Product Mockup in Perspective

INTRODUCTION We believe that every laboratory working in the field of nanotechnology needs an SEM, therefore we would like to introduce to you our IEM

SOP: Polyvar Met Light Microscope

Introduction to Scanning Electron Microscopy

S S S2 Operation Manual

FEI Tecnai G 2 F20 Operating Procedures

BX-61: Brightfield Instruction /Continue to scroll for Fluorescent Instuctions

What is an EQ? Subtract Hz to fix a problem Add Hz to cover up / hide a problem

Horiba LabRAM ARAMIS Raman Spectrometer Revision /28/2016 Page 1 of 11. Horiba Jobin-Yvon LabRAM Aramis - Raman Spectrometer

Olympus LEXT OLS 4000 Confocal Laser Microscope

ML7520 ML7530 DIOPTER ADJUSTMENT RING BINOCULAR BODY, INCLINED 30. (a) Field Iris Control Lever. (c) Filter Slots EYEPIECES, KHW10X

Operation Guide for the Leica SP2 Confocal Microscope Bio-Imaging Facility Hunter College October 2009

Transcription:

SEM OPERATION IN LOW VACUUM MODE Instructions for JEOL 5800 LV The EVAC light of the SEM specimen chamber should be already lit when you approach the SEM & the SEM will have been left in the high vacuum or SEI mode. Sample display reading: HT READY SEI 20KV X 300,000 WD35 X = + 0.00 Y = +0.00 R= +0.00 Z= +35.0 T = +0.0 WD & Z depend position of sample holder during removal of sample from chamber. To load sample: 1. Press VENT and hold for about 1-2 seconds to bring the specimen chamber to atmosphere. Vent button lights up, Evac light goes off. 2. Wait for Beep. 3. UNLOCK specimen chamber door. 4. OPEN specimen chamber door. 5. INSERT specimen holder containing sample into specimen chamber using the tongs. Keep fingers off of specimen holder & specimen use gloves. 6. CHANGE the magnification from x300,000 (see display panel) to a lower mag by pressing the instant magnification INST MAG button. Magnification will instantly change from 300,000x to approximately 100x. You can further lower the magnification if you want using the magnification knob to ~60-80x. Lowest possible mag depends on your working distance (WD; see display panel). 7. Use the JOYSTICK to POSITION the sample holder in the specimen chamber. Make sure sample is not too close to detector! Position highest point of sample approximately 10-15 mm from the detector. To center sample, use joystick in X/Y mode (X/Y button is lit). Move it left/right. (If Magnification is too high, X/Y movement is nill. To raise/lower sample, use joystick in T/Z mode (T/Z button is lit, and small T/Z button flashes). Move it up/down (towards left monitor or towards you). Note: Stage moves the fastest at the lowest possible magnification. IMPORTANT! Press X/Y button when sample holder is in position. (X/Y button is now lit). 8. CLOSE & LOCK specimen chamber door. * Write down the WD and position of your sample for future reference. 1

To start LV mode: 9. Press LV Button located on front left panel of SEM. Display will read 9.9.9 Pa. 10. Press AUTO button. 11. Press EVAC button to evacuate the specimen chamber. LV panel will flash to the number previously set. Suggested LV setting for a wet sample is 150 Pa; for a dry sample try 20 Pa. To change LV setting: turn controller knob to the right (clockwise) to turn the pressure down (i.e., 50 Pa to 7 Pa), and to the left (counter clockwise) to increase pressure in specimen chamber (i.e., 10 Pa to 150 Pa). It takes a while for the pressure to change - be patient! You can move on to the next step before specimen chamber reaches desired vacuum, but remember to listen for beep before saturating the filament. LV panel stops flashing when it reaches the desired vacuum that you set. 12. LOOK at display panel under left monitor. It will change from VENT to PRE-EVAC to EVAC to HT READY. Sample display panel reading: HT READY BE/C 20KV X 200 WD35 X = + 0.00 Y = +0.00 R= +0.00 Z= +35.0 T = +0.0 WD & Z depend sample height in sample holder & focus setting. 13. TURN UP the brightness & contrast on the two monitors. Don t turn them up all the way. 14. If the display panel shows a different accelerating voltage than you want, use the computer mouse to choose MENU (if not already in this mode). Click on ACCV to choose the accelerating voltage you wish. When finished, a right mouse click or escape will get rid of this submenu. Remember: a high accelerating voltage (30kV) will penetrate samples deeper and damage them more quickly than a low one (10kV). 15. In the Menu mode click on BEIW (left side of screen). Make sure the submenu reads: Polarity POSI When activated, box is yellow. Image COMP Compo Brt CRS Shadow ON Choose a middle shadow level (click mouse on middle position). 2

To saturate the filament: 16. First verify the filament knob is at the seven o clock position (i.e., off). Press SPOT SIZE button (on extreme right of keyboard). The button will light up green when activated. CRS, SIZE and FINE will light up orange, and SIZE will flash on & off. This button may have been on TONE. 17. Check the number of chevrons (>>) visible in the display panel which may look as follows: HT READY BE/C 20KV x 300,000 WD34 SPOT SIZE >>>>>>>> >>>>>>>> > -- -- -- -- -- -- 18. Press the HT button (extreme right of keyboard). This button will light up orange when activated. The filament light to its immediate left will light up and flash continuously. 19. Press DMAG button on top left keyboard. The button will light up green when activated. 20. Press LSP1 button (to right of DMAG button). The button will light up green when activated, and a line appears on the left monitor. 21. Make sure that at least 12 chevrons are lit (ballpark setting: anywhere in the middle set of chevrons). Change spot size by rotating the knobs for CRS and FINE. Spot size is dependent on the chosen accelerating voltage you are working with. Go to TONE. 22. SLOWLY saturate the filament by slowly turning the filament knob clockwise from a 7 o clock position to approximately 2 o clock. Notice the filament light stops flashing. What you need to look for in the left monitor as you rotate the filament knob so that you know the filament is saturated: Observe the horizontal line on the left monitor. As you slowly saturate the filament, the line should become wavy (noisy). When the panel display reads 12 o clock, the line should be very wavy at this point. If not press TONE button (is lit green), and rotate knob controlling contrast (CONT) & knob controlling brightness (BRT) until the line on the left monitor is wavy. Don t let the line to go to the top of the screen & become flat - adjust using BRT & CONT knobs. As you continue to slowly turn the filament knob, the wavy line should: 1) increase in amplitude, then 2) decrease amplitude, then 3) increase again. At the point of the second increase the filament is saturated. IMPORTANT! Do not turn the filament knob past 2 o clock. Look at the panel display which should show L.C. and a number (in µa) instead of HT ready. If µa doesn t increase with knob rotation you have saturated the filament!!! Back off. A sample display panel in LV mode with filament saturated. Record start time on filament. Close canopy lid. L.C BE/C 83µA 20KV X 200 WD35 X = + 0.00 Y = +0.00 R= +0.00 Z= +35.0 T = +0.0 3

To get an image: *Insure that nothing or no one is touching/against the column console. Image quality is greatly affected. 23. Press the SL1 button. The line on the left monitor should disappear. If the picture is too bright or dark; in TONE mode, rotate the BRT and/or CONT knobs, so that sample is visible. An easier way to adjust is listed below. If you press SL1 again, a graphical display (called a wave form monitor) will appear overlaying the image in the left monitor. Adjust BRT & CONT so that they are centered on the graph. The sample is most likely out of focus at this time. 24. Press the COARSE button above the focus knob (right side of keyboard). This will let you quickly focus on the sample. 25. Increase magnification by 5-10 times or more (e.g., from 150x to 550x). Use coarse focus to get a sharp image. 26. When you get the best focus on the sample, press the COARSE button again (coarse button is no longer lit) this lets the focus knob adjust with fine focus. 27. Press the STIG button and use X & Y knobs to get a sharply focused image; then return to the desired low mag (e.g., 150x). When you change ACCV, recheck filament saturation and stigmate. 28. Press the SL2 button once. Observe the image - if it is acceptable, you can go to step #31 or #32. If not, press SL2 button again to obtain wave form monitor function, then use fine focus to get sample in focus, and adjust CON & BRT. 29. Press SL2 button (now in regular raster mode) and use fine focus if necessary. 30. Press DMAG button (not lit) to get a full screen view. (Optional). To move the specimen: 31. If you are in the wrong area of the specimen, you may move the specimen by using the joy stick. BEFORE USING JOYSTICK: Make sure the X/Y button on the extreme left of keyboard is lit. Make sure the display panel reads T = +0.00 (see #12). If display panel says Spot Size, press Tone button to see sample position on display panel. Moving joy stick to left makes image move left, and you can see what was out of view to the right. Moving joy stick up (towards left monitor) makes image shift up, and you can see what was below the current view. 4

To take a video print picture: 32. In SL2 mode, obtain a good image that is in focus with appropriate contrast & brightness. 33. Press SL2 button twice to see a graphical display overlaying the image in the left monitor. In TONE mode adjust BRT and CONT so that they are centered on the graph. 34. Press SL3 and wait for one full screen (raster). 35. Press the Freeze L button to freeze the good image on left monitor. (Can also freeze right image and print while continuing to work with left screen). 36. Select L button under select on panel near Polaroid camera. (or R if froze R image). 37. Press PRINT under stored image. 38. Remove video print from bottom right front of SEM. Picture tears off. To take a Polaroid picture (photo): 39. Repeat steps #32 - #36 above. 40. Use Polaroid film Type 52 at F16, Pro 400 at F16, or Type 55 at F8. 41. Verify that Polaroid camera lever is in L position. 42. Insert film in camera. Leave lever in L position. 43. Pull film out. 44. Press Stored PHOTO. Photo light is on. 45. Wait for exposure by watching orange photo light go out & you ll also hear a beep. 46. Push film back in. 47. Move lever to P position. 48. Pull film out. 49. Wait a specified number of seconds as indicated on film envelope. 50. Leave lever in P position. To take save image to computer disk: 51. Repeat steps #32 - #34. 52. Insert disk into disk drive (Disk label faces left). 53. Click on ALBUM in computer screen. 54. Click on SAVE. 55. Answer questions in submenu (filename, screen choice, TIFF/Bitmap). 56. Wait for image to load on disk. In SL3 mode, only one image fits on disk. In other modes up to 4 images can fit on disk. Icon appears on computer screen. 5

To change the sample: 57. Slowly turn down the filament knob. Press the orange HT button to turn off the high tension. The HT button should no longer be lit. Keep track of filament time. 58. Press VENT on the specimen chamber. 59. Remove old sample and add new one: SEE steps #1-8 above. Be careful of specimen positioning. 60. Press EVAC button. 61. When display panel says HT Ready you can press the HT button to turn on the high tension. 62. Resume steps #23+ above to get an image of the new sample. 6

To shut down: 63. Slowly turn FILAMENT KNOB counter clockwise back to 7 o clock. Notice orange filament light next to HT button flashes. Keep track of filament time. 64. Press HT button if still on. It s orange light should be off, as well as that of the filament. 65. Turn MAGNIFICATION KNOB all the way to 300,000x. See display panel to verify. 66. Press DMAG button to turn it off, if you haven't done so already. It should not be lit. 67. Press Freeze L or R buttons so that they are not lit. 68. Press TONE button if it isn t already activated (should be lit green). 69. Press SL3 button. Should light up green. 70. Leave computer in MENU mode. 71. Turn down brightness and contrast on two monitors. 72. Press VENT. 73. Wait for beep. 74. Unlock and open specimen chamber door. 75. Lower specimen holder using joystick in T/Z mode. Look while doing this. 76. Press X/Y button so that it is lit green. 77. Close and lock specimen chamber door. 78. Press LV button (lit up) on front left panel to turn off LV (button no longer lit). 79. Press EVAC. Notice display mode shows SEI instead of BE/C. 80. IMPORTANT! CLEAN UP the area. 81. Scope is now in SEI mode. Leave scope in this mode. A sample display panel when you are finished. HT READY SEI 20KV X 300,000 WD34 X = +19.10 Y = +10.09 R= +0.00 Z= +35.0 T = +0.0 Note: WD & Z depend on where you left sample holder when you took the sample out. Only controls lit (green) are SL3, X/Y, and TONE. 7