X-ray light valve (XLV): a novel detectors technology for digital mammography*

Similar documents
Radiology Physics Lectures: Digital Radiography. Digital Radiography. D. J. Hall, Ph.D. x20893

Amorphous Selenium Direct Radiography for Industrial Imaging

LCD DISPLAY TECHNOLOGY. Digital Images and Pixels

CR Basics and FAQ. Overview. Historical Perspective

Introduction. Chapter 16 Diagnostic Radiology. Primary radiological image. Primary radiological image

Unit thickness. Unit area. σ = NΔX = ΔI / I 0

Lecture 15. Lecture 15

Performance of Image Intensifiers in Radiographic Systems

AUTOMATED AND QUANTITATIVE METHOD FOR QUALITY ASSURANCE OF DIGITAL RADIOGRAPHY IMAGING SYSTEMS

Investigation of Effective DQE (edqe) parameters for a flat panel detector

Chromatic X-Ray imaging with a fine pitch CdTe sensor coupled to a large area photon counting pixel ASIC

Radiographic sensitivity improved by optimized high resolution X -ray detector design.

radiography detector

10/3/2012. Study Harder

AN AMORPHOUS SELENIUM LIQUID CRYSTAL LIGHT VALVE FOR X-RAY IMAGING

10/26/2015. Study Harder

INTERNATIONAL STANDARD

Digital Detector Array Image Quality for Various GOS Scintillators

Properties of a Detector

DRX Plus Detectors: Going from Good to Great

DIGITAL RADIOGRAPHY. Digital radiography is a film-less technology used to record radiographic images.

Examination, TEN1, in courses SK2500/SK2501, Physics of Biomedical Microscopy,

DIGITAL IMAGE PROCESSING IN X-RAY IMAGING

X-RAY MEDICAL EQUIPMENT

HIGH SPEED, HIGH RESOLUTION AND LOW COST DIGITAL RADIOGRAPHY

Tunable wideband infrared detector array for global space awareness

CMOS X-RAY DETECTORS. for Industrial Non-Destructive Testing COMMITTED TO PEOPLE. DRIVEN BY INNOVATION.

TDI Imaging: An Efficient AOI and AXI Tool

X-ray detectors in healthcare and their applications

Measurement of the Modulation Transfer Function (MTF) of a camera lens. Laboratoire d Enseignement Expérimental (LEnsE)

Digital Radiography : Flat Panel

Do you have any other questions? Please call us at (Toll Free) or , or

Mammography Solution. AMULET Innovality. The new leader in the AMULET series. Tomosynthesis, 3D mammography and biopsy are all available.

SPECTRAL SCANNER. Recycling

Ultra-high resolution 14,400 pixel trilinear color image sensor

BASICS OF FLUOROSCOPY

Multi-spectral acoustical imaging

Using molded chalcogenide glass technology to reduce cost in a compact wide-angle thermal imaging lens

Mammography is a radiographic procedure specially designed for detecting breast pathology Approximately 1 woman in 8 will develop breast cancer over

Image sensor combining the best of different worlds

PERFORMANCE CHARACTERIZATION OF AMORPHOUS SILICON DIGITAL DETECTOR ARRAYS FOR GAMMA RADIOGRAPHY

LCD DISPLAY TECHNOLOGY

SECTION I - CHAPTER 1 DIGITAL RADIOGRAPHY: AN OVERVIEW OF THE TEXT. Exam Content Specifications 8/22/2012 RADT 3463 COMPUTERIZED IMAGING

DALLA LUCE VISIBILE AI RAGGI X: NUOVI RIVELATORI DI IMMAGINI PER RAGGI X A DISCRIMINAZIONE IN ENERGIA ED APPLICAZIONI

Gas scintillation Glass GEM detector for high-resolution X-ray imaging and CT

Image Display and Perception

CXDI-70C WIRELESS SPECIFICATIONS

WHITE PAPER MINIATURIZED HYPERSPECTRAL CAMERA FOR THE INFRARED MOLECULAR FINGERPRINT REGION

Shad-o-Box HS Product Family

Lamb Wave Ultrasonic Stylus

New spectral benefi ts, proven low dose

Digital Imaging Considerations Computed Radiography

brief history of photography foveon X3 imager technology description

Dental Radiography. One of the problems of dental radiography is having different dimensions than normal.

LCOS Devices for AR Applications

MTF and NPS of single-shot dual-energy sandwich detectors

Examination of Pipe Welds by Image Plate Based Computed Radiography System

ISO INTERNATIONAL STANDARD

Film Replacement in Radiographic Weld Inspection The New ISO Standard

IBEX MATERIALS DETECTION TECHNOLOGY

FRAUNHOFER AND FRESNEL DIFFRACTION IN ONE DIMENSION

Acquisition, Processing and Display

Phase Contrast Imaging with X-ray tube

Fundamentals of CMOS Image Sensors

Introduction. Chapter 1

C a t p h a n. T h e P h a n t o m L a b o r a t o r y. Ordering Information

Setting up digital imaging department!

PROCEEDINGS OF SPIE. Measurement of the modulation transfer function (MTF) of a camera lens

REAL-TIME X-RAY IMAGE PROCESSING; TECHNIQUES FOR SENSITIVITY

High Speed Hyperspectral Chemical Imaging

CPSC 4040/6040 Computer Graphics Images. Joshua Levine

True%Analog%Non-Mechanical%Beam%Steering%Using%Liquid%Crystal% Waveguide%Techniques%

SYLLABUS. TITLE: Equipment Operation I. DEPARTMENT: Radiologic Technology

Experiences of users in Digital Radiography

High Performance Amorphous Silicon Image Sensor for X-ray Diagnostic Medical Imaging Applications

Designing an MR compatible Time of Flight PET Detector Floris Jansen, PhD, Chief Engineer GE Healthcare

Compatible with Windows 8/7/XP, and Linux; Universal programming interfaces for easy custom programming.

New Detectors for X-Ray Metal Thickness Measuring

Features and Weaknesses of Phantoms for CR/DR System Testing

Strategies to improve the signal and noise performance of active matrix, flat-panel imagers for diagnostic x-ray applications

Hardware for High Energy Applications 30 October 2009

the need for an intensifier

Development of Solid-State Detector for X-ray Computed Tomography

Light guide with internal mirror array for LCD backlight

CHAPTER 9 POSITION SENSITIVE PHOTOMULTIPLIER TUBES

Medical Imaging. X-rays, CT/CAT scans, Ultrasound, Magnetic Resonance Imaging

General Hospital Catalog

Photons and solid state detection

Project Staff: Feng Zhang, Prof. Jianfeng Dai (Lanzhou Univ. of Tech.), Prof. Todd Hasting (Univ. Kentucky), Prof. Henry I. Smith

Applications of a Pixellated Detection System to Digital Mammography

DIGITAL IMAGING. Handbook of. Wiley VOL 1: IMAGE CAPTURE AND STORAGE. Editor-in- Chief

2013 LMIC Imaging Workshop. Sidney L. Shaw Technical Director. - Light and the Image - Detectors - Signal and Noise

MODULAR ADAPTIVE OPTICS TESTBED FOR THE NPOI

Minimizes reflection losses from UV-IR; Optional AR coatings & wedge windows are available.

PIXIS-XO: 1024B 1024 x 1024 imaging array 13 x 13 µm pixels

Moving from film to digital: A study of digital x-ray benefits, challenges and best practices

Unit 2 Semiconductor Devices. Lecture_2.5 Opto-Electronic Devices

BACKSIDE ILLUMINATED CMOS-TDI LINE SCANNER FOR SPACE APPLICATIONS

The new CMOS Tracking Camera used at the Zimmerwald Observatory

Seminar 8. Radiology S8 1

Transcription:

X-ray light valve (XLV): a novel detectors technology for digital mammography* Sorin Marcovici, Vlad Sukhovatkin, Peter Oakham XLV Diagnostics Inc., Thunder Bay, ON P7A 7T1, Canada ABSTRACT A novel method, based on X-ray Light Valve (XLV) technology, is proposed for making good image quality yet inexpensive flat panel detectors for digital mammography. The digital mammography markets, particularly in the developing countries, demand quality machines at substantially lower prices than the ones available today. Continuous pressure is applied on x-ray detectors' manufacturers to reduce the flat panel detectors prices. XLV presents a unique opportunity to achieve the needed price - performance characteristics for direct conversion, x-ray detectors. The XLV based detectors combine the proven, superior, spatial resolution of a-se with the simplicity and low cost of liquid crystals and optical scanning. The x-ray quanta absorbed by a 200 µm a-se produce electron hole pairs that move under an electric field to the top and bottom of a-se layer. This 2D charge distribution creates at the interface with the liquid crystals a continuous (analog) charge image corresponding to the impinging radiation's information. Under the influence of local electrical charges next to them, the liquid crystals twist proportionally to the charges and vary their light reflectivity. A scanning light source illuminates the liquid crystals while an associated, pixilated photo-detector, having a 42 µm pixel size, captures the light reflected by the liquid crystals and converts it in16 bit words that are transmitted to the machine for image processing and display. The paper will describe a novel XLV, 25 cm x 30 cm, flat panel detector structure and its underlying physics as well as its preliminary performance measured on several engineering prototypes. In particular, the paper will present the results of measuring XLV detectors' DQE, MTF, dynamic range, low contrast resolution and dynamic behavior. Finally, the paper will introduce the new, low cost, XLV detector based, digital mammography machine under development at XLV Diagnostics Inc. OBJECTIVES In many developing countries the number of breast cancers cases is increasing rather fast but their early detection is lagging due to absence of national screening programs. One of the reasons is that the user price of existing digital mammography machines is prohibitively high for the developing world. To address the large demand for low price, digital mammography machines, XLV Diagnostics Inc. is developing a novel, inexpensive, electro-optics, x-ray detectors technology based on which good quality, digital mammography machine, suitable for the developing world that can be profitably sold for about a third of the price of the machines available today. This objective is achievable because the underlying physics of electro-optics detectors allows for separating the image acquisition from image digitization and eliminates the need for using 2D active matrix array and peripheral electronics which are the most expensive components presently used within direct conversion detectors. *SPIE Conference on Medical Imaging 2014, San Diego, CA; Physics of Medical Imaging, Volume 9033 Part 2

DETECTOR OPERATION The XLV detector is a direct type detector that employs a 200 µm a-se layer evaporated on a very thin piece of glass as a radiation absorption and detection medium. The impinging x-ray quanta are stopped by the a-se layer and create electron-hole pairs which move under a strong electric field of 10V/µm towards the two opposing surfaces of the a-se layer. A very thin metallic electrode covers the quanta impinging surface at the top of the a-se layer. The bottom surface of the a-se layer rests on a piece of thin glass. At the a-se thin glass interface, the arriving charges create a 2-dimentional continuous charge distribution that is an analog mapping the volumetric attenuation information carried by the arriving quanta. This 2D charge distribution modulates local optical characteristics of a continuous thin liquid crystals layer positioned on the other side of the thin glass. The corresponding twisting of the liquid crystals molecules is monotonic in relation to the proximal charges and, when exposed to light, they reflect the light proportionally to their local twist. A calibrated light source illuminates the liquid crystals using simple focusing lens and filters and the reflected light is measured and digitized by an associated linear optical scanner. The optical scanner speed and pixel size are selectable. Digitized images are streamed to a computer outside the detector using standard, high speed, digital data interface. By maintaining the image information in analog domain until is read-out by the optical scanner, no additional noise or, equivalently speaking, secondary quantum sink, is introduced. Moreover, the read-out light s intensity can be increased, if desired, to increase the overall gain. The electrical field is applied across a-se and liquid crystals between to electrodes: the top electrode made of Chromium and the bottom electrode made of ITO. The liquid crystals layer has to be sandwiched between its two alignment layers and the ITO electrode is positioned below the bottom alignment layer. During x-ray quanta capture, the electric field is set at 10 V/µm and during analog image readout it is set at approximately 0.5 V/ µm. The analog, charge image stored at the thin glass interface with a-se can be readout repeatedly because the readout is non-destructive. The readout light source is LED based operating at 630 nm wavelength and collimated with very simple and inexpensive focusing lenses. The light reflected by the liquid crystals is read out using custom made opto-electronics mounted on the linear scanner together with a high speed, 16 bit analog-to-digital converter. The spatial resolution of the linear scanning assembly is 42 µm however, by a varying the scanning speed and electronics parameters, the pixel size can be selected dynamically for every scan. This feature allows, for example, having a fast scan at lower resolution followed by a high resolution scan taken at the nominal speed. Due to the fact the analog charge image is not digitized until the reflected

light is measured by the analog-to-digital converter on the linear scanner, the gain of the XLV detector can be increased by illuminating the liquid crystals with higher intensity light. To measure the electro-optic parameters of the sub-system formed by a-se and liquid crystals, we built a custom test station where the reflected light is measure using a very high resolution CCD. Figure2 shows the experimental set-up and Figure 3 puts in evidence that the MTF of the optical test station is much higher than the desired MTF to be measured. Figure 2 Figure 3

The DQE and the MTF of XLV based detectors prototypes were measured in accordance with the Determination of the detective quantum efficiency Detector used in Mammography - IEC 62220-1-2 Ed.1, 2007 and shown encouraging results. For readers convenience, the XLV s MTF characteristic curve is shown below in comparison with top-of-theline digital mammography machines in commercial use today. MTF Figure 4 The results demonstrate that the XLV prototype detector s MTF is equal to the MTF of the top-of-the line detectors for digital mammography and provides for the required special resolution necessary for an effective breast cancer screening machine.

Preliminary DQE evaluation also indicates that the performance of XLV based digital mammography detectors will be similar to the typical one for 200µm a-se detectors. DQE spacial frequency [mm¹] Figure 5 XLV detector s dynamic range is determined mainly by the response of the liquid crystals and the optical scanner. To cover the desired dynamic range for digital mammography, the selected liquid crystals must provide an electro-optic characteristic extending beyond fourteen bits and the associated optical scanner ADC to operate with 16 bits resolution and minimum pixel size is 42 µm.

XLV advantages Ø There is no front-end noise due to data lines capacitance because there are no data lines Ø The readout is nondestructive so the charge image can be read multiple times before erasure Ø The readout can be dynamically set to different speeds and pixel sizes Ø Without adding noise, the readout light intensity can be varied to increase output signal XLV based digital mammography machine The inherent simplicity of the XLV detectors technology provides for the opportunity to design a very robust and inexpensive digital mammography system by tightly integrating the detector with the compressing stand. Minimum control and calibration circuitry required for detector s operation allows that the majority of functions be executed in a very efficient and economic manner by the electronics resources already resident within the compression stand. Moreover, the compression stand itself can be engineered in a modular way conducive to rapid, remote testing and easy serviceability. The XLV digital mammography machines will be first equipped with a 25 cm x 30 cm detector to be followed, at a later date, by an 18 cm x 24 cm detector. All image processing and image presentation functions will be similar to those of mammography machines already in use. Summary XLV was proven to be a cost effective technology for producing good quality, large area, x-ray detectors for digital radiography. In particular, the work described in this paper demonstrated XLV suitability for medical imaging applications demanding very high spatial resolution like the one for digital mammography. Both XLV detectors and XLV digital mammography machines open a new avenue for developing good quality yet economic digital radiography machines. Acknowledgement The authors want to acknowledge Prof. John Rowlands seminal contributions to the development of XLV scientific theory and its underlying basic technologies. Patents US5847499 - Apparatus for generating multiple x-ray images of an object from a single x-ray exposure US6052432 Method of generating multiple images of an object from a single x-ray exposure US7687792 X-ray light valve based digital radiographic imaging system Contact Sorin Marcovici, XLV Diagnostics Inc., 290 Munro St., Suite 2311, Thunder Bay, ON P7A 7T1, Canada, sorin@xlvdiagnostics.com