Introduction to Scanning Electron Microscopy

Similar documents
Functions of the SEM subsystems

ELECTRON MICROSCOPY. 13:10 16:00, Oct. 6, 2008 Institute of Physics, Academia Sinica. Tung Hsu

SCANNING ELECTRON MICROSCOPY AND X-RAY MICROANALYSIS

Scanning electron microscope

MODULE I SCANNING ELECTRON MICROSCOPE (SEM)

ELECTRON MICROSCOPY. 14:10 17:00, Apr. 3, 2007 Department of Physics, National Taiwan University. Tung Hsu

S200 Course LECTURE 1 TEM

Scanning electron microscope

Scanning Electron Microscopy. EMSE-515 F. Ernst

Scanning Electron Microscopy Basics and Applications

Numerical analysis to verifying the performance of condenser magnetic lens in the scanning electron microscope.

Transmission Electron Microscopy 9. The Instrument. Outline

OPERATION OF THE HITACHI S-450 SCANNING ELECTRON MICROSCOPE. by Doug Bray Department of Biological Sciences University of Lethbridge

(Refer Slide Time: 00:10)

NANO 703-Notes. Chapter 9-The Instrument

SCIENTIFIC INSTRUMENT NEWS. Introduction. Design of the FlexSEM 1000

ELECTRON MICROSCOPY. 09:10 12:00, Oct. 27, 2006 Institute of Physics, Academia Sinica. Tung Hsu

Mohammed A. Hussein *

2.Components of an electron microscope. a) vacuum systems, b) electron guns, c) electron optics, d) detectors. Marco Cantoni 021/

Introduction of New Products

ELECTRON MICROSCOPY AN OVERVIEW

Dickinson College Department of Geology

Introduction: Why electrons?

Chapter 2 Instrumentation for Analytical Electron Microscopy Lecture 7. Chapter 2 CHEM Fall L. Ma

NanoSpective, Inc Progress Drive Suite 137 Orlando, Florida

Basics of Light Microscopy and Metallography

p q p f f f q f p q f NANO 703-Notes Chapter 5-Magnification and Electron Sources

Scanning Electron Microscopy

Electron Sources, Optics and Detectors

Topics 3b,c Electron Microscopy

Transmission electron Microscopy

Scanning Electron Microscopy SEM. Warren Straszheim, PhD MARL, 23 Town Engineering

SOP for Hitachi S-2150 Scanning Electron Microscope For review purposes only

Fabrication of Probes for High Resolution Optical Microscopy

Oct. 30th- Nov. 1st, 2017

Standard Operating Procedure for the Amray 1810 Scanning Electron Microscope Version: 29 NOVEMBER 2014

SECONDARY ELECTRON DETECTION

ZEISS EVO SOP. May 2017 ELECTRON OPTICS

Nanotechnology and material science Lecture V

Introduction to Electron Microscopy

Chapter 1. Basic Electron Optics (Lecture 2)

Title: Amray 1830 SEM#2 Semiconductor & Microsystems Fabrication Laboratory Revision: D Rev Date: 03/18/2016

2.Components of an electron microscope. a) vacuum systems, b) electron guns, c) electron optics, d) detectors. Marco Cantoni, 021/

Oct. 30th- Nov. 1st, 2017

VISUAL PHYSICS ONLINE DEPTH STUDY: ELECTRON MICROSCOPES

Low Voltage Electron Microscope

JEOL 6500 User Manual

--> Buy True-PDF --> Auto-delivered in 0~10 minutes. JY/T

Design and fabrication of a scanning electron microscope using a finite element analysis for electron optical system

Microscopy techniques for biomaterials. Engenharia Biomédica. Patrícia Almeida Carvalho

Plasmonic Nanoparticles

Schottky Emission VP FE-SEM

PROCEEDINGS OF A SYMPOSIUM HELD AT THE CAVENDISH LABORATORY, CAMBRIDGE, Edited by

Tecnai on-line help manual --

The light microscope

Introduction to Transmission Electron Microscopy (Physical Sciences)

Secondary Electron Detector

Scanning Electron MICROSCOPES. SEM-20/30, TableTop SEM-20. The Wise SEM for Your Application

attocfm I for Surface Quality Inspection NANOSCOPY APPLICATION NOTE M01 RELATED PRODUCTS G

ELECTRON MICROSCOPY. 14:10 17:00, Mar. 8, :10 17:00, Mar. 15, 2018 P101, Institute of Physics, Academia Sinica. Tung Hsu

Procedures for Performing Cryoelectron Microscopy on the FEI Sphera Microscope

2. Raise HT to 200kVby following the procedure explained in 1.6.

Module 4B7: VLSI Design, Technology, and CAD. Scanning Electron Microscopical Examination of CMOS Integrated Circuit

General information. If you see the instrument turned off, notify MIC personnel. MIC personnel will help you insert your samples into the instrument.

Operating the Hitachi 7100 Transmission Electron Microscope Electron Microscopy Core, University of Utah

How to choose a Scanning Electron Microscope (SEM)

SEM Training Notebook

1.1. In regular TEM imaging mode, find a region of interest and set it at eucentric height.

Observing Microorganisms through a Microscope LIGHT MICROSCOPY: This type of microscope uses visible light to observe specimens. Compound Light Micros

Scanning Electron Microscope. Instructions for Use

1. Preliminary sample preparation

Microscope. Dr. Leena Barhate Department of Microbiology M.J.College, Jalgaon

Development of JEM-2800 High Throughput Electron Microscope

Transmissions Electron Microscopy (TEM)

Operating Checklist for using the Scanning Electron Microscope, JEOL JSM 6400.

!"#$$%$&'()*"+,-$'.%",-/"-0*'-$'#'12%0'

Low Voltage Electron Microscope

LVEM 25. Low Voltage Electron Mictoscope. fast compact powerful

Full-screen mode Popup controls. Overview of the microscope user interface, TEM User Interface and TIA on the left and EDS on the right

1.3. Before loading the holder into the TEM, make sure the X tilt is set to zero and the goniometer locked in place (this will make loading easier).

Scanning Electron Microscope FEI INSPECT F50. Step by step operation manual

Operating Checklist for using the Scanning Electron. Microscope, JEOL JSM 6400.

Bringing Answers to the Surface

A Portable Scanning Electron Microscope Column Design Based on the Use of Permanent Magnets

Experiment 6: Franck Hertz Experiment v1.3

MSE 460 TEM Lab 2: Basic Alignment and Operation of Microscope

Low Voltage Electron Microscope. Nanoscale from your benchtop LVEM5. Delong America

MCR Scanning Electron Microscopy Laboratory Portfolio

2 How to operate the microscope/obtain an image

Lecture 20: Optical Tools for MEMS Imaging

LEO 912 TEM Short Manual. Prepared/copyrighted by RH Berg Danforth Plant Science Center

TEM theory Basic optics, image formation and key elements

Titan on-line help manual -- Working with a FEG

FEI Quanta 200 ESEM Basic instructions

Using the Hitachi 3400-N VP-SEM

SEM OPERATION IN LOW VACUUM MODE

1.1. Log on to the TUMI system (you cannot proceed further until this is done).

JEOL 6700 User Manual 05/18/2009

Model SU3500 Scanning Electron Microscope

ELECTRON OPTICS. Prof. John G. King Dr. John W. Coleman Dr. Edward H. Jacobsen. Graduate Students. Steven R. Jost Norman D. Punsky

Transcription:

Introduction to Scanning Electron Microscopy By: Brandon Cheney Ant s Leg Integrated Circuit Nano-composite This document was created as part of a Senior Project in the Materials Engineering Department at San Jose State University. It is intended to provide an introduction scanning electron microscopy and techniques for better imaging. References are provided at the end of the paper for those who wish to study the subject more thoroughly.

1. Introduction The SEM instrument is made up of two main components, the electronic console and the electron column. The electronic console provides control knobs and switches that allow for instrument adjustments such as filament current, accelerating voltage, focus, magnification, brightness and contrast. The FEI Quanta 200 is a state of the art electron microscope that uses a computer system in conjunction with the electronic console making it unnecessary to have bulky console that houses control knobs, CRTs, and an image capture device. All of the primary controls are accessed through the computer system using the mouse and keyboard. The user need only be familiar with the GUI or software that controls the instrument rather than control knobs and switches typically found on older style scanning electron microscopes. The image that is produced by the SEM is usually viewed on CRTs located on the electronic console but, instead with FEI the image can be seen on the computer monitor. Images that are captured can be saved in digital format or printed directly. 2. Electron Column The electron column is where the electron beam is generated under vacuum, focused to a small diameter, and scanned across the surface of a specimen by electromagnetic deflection coils. The lower portion of the column is called the specimen chamber. The secondary electron detector is located above the sample stage inside the specimen chamber. Specimens are mounted and secured onto the stage which is controlled by a goniometer. The manual stage controls are found on the front side of the specimen chamber and allow for x-y-z movement, 360 rotation and 90 tilt however only the tilt cannot be controlled through the computer system thus there is no need to use all of the manual controls manipulate the orientation of the sample inside the sample chamber. Below is a diagram of the electron column and a description of each of the components of the electron column.

Figure 2. Scanning electron microscope column [1]. 1) Electron gun: Located at the top of the column where free electrons are generated by thermionic emission from a tungsten filament at ~2700K. The filament is inside the Wehnelt which controls the number of electrons leaving the gun. Electrons are primarily accelerated toward an anode that is adjustable from 200V to 30 kv (1kV=1000V). 2) Condenser Lenses: After the beam passes the anode it is influenced by two condenser lenses that cause the beam to converge and pass through a focal point. What occurs is that the electron beam is essentially focused down to 1000 times its original size. In conjunction with the selected accelerating voltage the condenser lenses are primarily responsible for determining the intensity of the electron beam when it strikes the specimen [1].

3) Apertures: Depending on the microscope one or more apertures may be found in the electron column. The function of these apertures is to reduce and exclude extraneous electrons in the lenses. The final lens aperture located below the scanning coils determines the diameter or spot size of the beam at the specimen. The spot size on the specimen will in part determine the resolution and depth of field. Decreasing the spot size will allow for an increase in resolution and depth of field with a loss of brightness [1]. 4) Scanning System: Images are formed by rastering the electron beam across the specimen using deflection coils inside the objective lens. The stigmator or astigmatism corrector is located in the objective lens and uses a magnetic field in order to reduce aberrations of the electron beam. The electron beam should have a circular cross section when it strikes the specimen however it is usually elliptical thus the stigmator acts to control this problem [1,4]. 5) Specimen Chamber: At the lower portion of the column the specimen stage and controls are located. The secondary electrons from the specimen are attracted to the detector by a positive charge. 3. Vacuum System The ability for a SEM to provide a controlled electron beam requires that the electronic column be under vacuum at a pressure of at least 5x10-5 Torr. A high vacuum pressure is required for a variety of reasons. First, the current that passes through the filament causes the filament to reach temperatures around 2700K [2]. A hot tungsten filament will oxidize and burn out in the presence of air at atmospheric pressure. Secondly, the ability of the column optics to operate properly requires a fairly clean, dust-free environment. Third, air particles and dust inside the column can interfere and block the electrons before the ever reach the specimen in the sample chamber [1]. In order to provide adequate vacuum pressure inside the column, a vacuum system consisting of two or more pumps is typically present.

Separate pumps are required because one pump isn t really capable of doing all the work but, in conjunction they can provide a good vacuum pressure relatively quickly and efficiently. A majority of the initial pumping is done by the action of a mechanical pump often called a roughing pump. The roughing pump operates first during the pump-down process and has excellent efficiency above 10-2 Torr. Although many mechanical pumps used in SEMs are capable of producing pressures better than 5x10-5 Torr, a very long pump down time would mostly be required. Pressures lower than 10-2 Torr are more easily acquired by the action of a turbo-molecular pump. Turbo-molecular pumps make use of a turbine that rotates at 20,000 to 50,000 rotations per minute to evacuate gas molecules and particulates found inside the column [7]. Turbo-molecular pumps are expensive and sensitive to vibrations thus it is important to remember that sudden jolts to the instrument can not only affect the beam but, the severely damage turbo pumps. The FEI Quanta 200 is design to operate in three separate vacuum modes, high vacuum (10-6 to 10-7 Torr), low vacuum (0.1 to 20 Torr) and ESEM (0.1 to 20 Torr) [6]. In the low vacuum and ESEM mode of operation the pressure inside the specimen chamber operates at a lower pressure than that inside the actual column which is always at high vacuum. The high vacuum mode is typically where most SEMs operate. In this mode the highest resolutions and magnifications can be achieved although, it is not suitable for all specimens. Generally anything that is conductive and has a high density works well in high vacuum mode. Specimens that are non-conductive or have a low density are more suitable for the low vacuum or ESEM mode. The ESEM mode of operation is necessary for wet, non-conductive samples. 4. Electron Beam-Specimen Interactions Originally microscopy was based on the use of the light microscope and could provide specimen resolutions on the order of 0.2 microns. To achieve higher resolutions, an electron source is required instead of light as the illumination source, which allows for resolutions of about 25 Angstroms. The use

of electrons not only gives better resolution but, due to the nature electron beam specimen interactions there are a variety of signals that can be used to provide information regarding characteristics at and near the surface of a specimen. In scanning electron microscopy visual inspection of the surface of a material utilizes signals of two types, secondary and backscattered electrons. Secondary and backscattered electrons are constantly being produced from the surface of the specimen while under the electron beam however they are a result of two separate types of interaction. Secondary electrons are a result of the inelastic collision and scattering of incident electrons with specimen electrons. They are generally characterized by possessing energies of less than 50 ev [1]. They are used to reveal the surface structure of a material with a resolution of ~10 nm or better [1]. Backscattered electrons are a result of an elastic collision and scattering event between incident electrons and specimen nuclei or electrons. Backscattered electrons can be generated further from the surface of the material and help to resolve topographical contrast and atomic number contrast with a resolution of >1 micron. While there are several types of signals that are generated from a specimen under an electron beam the x-ray signal is typically the only other signal that is used for scanning electron microscopy. The x-ray signal is a result of recombination interactions between free electrons and positive electron holes that are generated within the material. The x-ray signal can originate from further down into the surface of the specimen surface and allows for determination of elemental composition through EDS (energy dispersive x-ray spectroscopy) analysis of characteristic x-ray signals. Figure 3 is a diagram which displays a cross section of the volume of primary excitation illustrating zones from which signals may be detected [1].

Figure 3. Signals that result from electron beam-specimen interaction [2]. 5. Observation Techniques Scanning microscopes are becoming much easier to use these days with the advancement of electronics and introduction of new techniques. After just a short amount of training nearly anybody can acquire relatively good images. As users begin to investigate a broader range of materials and encounter more complex analysis scenarios, satisfactory images become more difficult for the inexperienced user to obtain. When the image is not clear enough or an image of a particular feature is seemingly impossible to obtain, it requires further problem solving to determine what the cause really is. This section will focus on the techniques for successful imaging.

Image Disturbances There are many types of image disturbances that occur while imaging specimens. In some cases these disturbances are attributed to defects with the instrument; however, more than likely the disturbances are due to operator s lack of experience, improper sample preparation, or external influences such as instrument vibration. Disturbances can be examined and corrected using a systematic approach based on their appearance. Table 2 below summarizes image disturbances and conditions that might cause them. Table 2. Image disturbances and their possible causes [6].

Effect of Accelerating Voltage The accelerating voltage on the Qunata 200 is adjustable from 200V-30 kv. Choosing the right accelerating voltage is critical for obtaining a good clear image however the most suitable voltage level depends will depend on mostly on the type of material being examined. The more conductive the material the better it will behave under higher voltages. Higher voltages (15-30kv) generally allow for high resolution at high magnifications although, this can damage the specimen very quickly if it is not highly conductive. Thus, when imaging polymers and ceramics it is more suitable to use voltages below 10 kv. The figure below details the effect of using high and low accelerating voltages. Figure 4. Effect of accelerating voltage [6].

5 kv Accelerating Voltage 25 kv Accelerating Voltage Figure 5. The above images were taken at different accelerating voltages, notice how the resolution is improved with increased accelerating voltage [6]. 30 kv Accelerating Voltage 5 kv Accelerating Voltage Figure 6. The above images were with different accelerating voltages, notice how surface structures appear clearer at lower voltages [6].

Effect of Working Distance Besides the accelerating voltage one s choice of working distance and spot size will greatly influence the image quality. As with accelerating voltage there exists a give and take situation when choosing the most suitable settings for working distance and spot size. Generally speaking a working distance of 10mm should be used and will allow for a good depth of field while maintaining good resolution. In most cases one may want to reduce the working distance to achieve better resolution especially when using lower accelerating voltages. Figure 5 below illustrates the effect of working distance. Figure 7. Effect of working distance [6]. 10 mm Working Distance 38 mm Working Distance Figure 8. The above images were taken at two different working distances, notice how the depth of field increases with working distance [6].

Effect of Spot Size Spot size basically restricts the beam current and will thereby cause for brightness and contrast compensations. Smaller spot sizes will require higher brightness and contrast levels thus there can be a limits when using a small spot size. Typically smaller spot sizes allow for higher resolution and a greater depth of field. Figure 6 below shows the effects of spot size on image quality. Figure 9. Effect of Spot Size [6]. 600 micron spot size 200 micron spot size Figure 10. The above images were taken with two different size spot sizes. Notice how much the depth of field is improved in the 200 micron spot size image [6].

6. References 1. M.T. Postek, K.S. Howard, A.H. Johnson and K.L. McMichael, Scanning Electron Microscopy: A Student s Handbook, (Ladd Research Ind., Inc. Williston, VT., 1980). 2. C.E. Lyman, D.E. Newbury, J.I. Goldstein, D.B. Williams, A.D. Romig, J.T. Armstrong, P. Echlin, C.E. Fiori, D.C. Joy, E. Lifshin and Klaus-Ruediger Peters, Scanning Electron Microscopy, X-Ray Microanalysis and Analytical Electron Microscopy: A Laboratory Workbook, (Plenum Press. New York, N.Y., 1990). 3. J.I. Goldstein, H. Yakowitz, D.E. Newbury, E. Lifshin, J.W. Colby, J.W. Colby and J.R. Coleman, Pratical Scanning Electron Microscopy: Electron and Ion Microprobe Analysis, edited by J.I. Goldstein and H. Yakowitz (Pelnum Press. New York, N.Y., 1975). 4. I.M. Watt, The Principles and Practice of Electron Microscopy, (Cambridge Univ. Press. Cambridge, England, 1985). 5. FEI. The Quanta 200 User s Operation Manual 2 nd ed. (2004). 6. JEOL. Guide to scanning electron Microscopy [Online]. Available at http://www.jeol.com/sem/docs/sem_guide/tbcontd.html (accessed 1 Feb 2005). 7. WIKIPEDIA (2005). Turbo molecular pump [Online]. Available at http://en.wikipedia.org/wiki/turbomolecular_pump (accessed 1 Feb 2005).