Patient-Assisted Compression Impact on Image Quality and Workflow

Similar documents
Acceptance Testing of a Digital Breast Tomosynthesis Unit

GE Healthcare. Essential for life. Senographe Essential Full-Field Digital Mammography system

Tomosynthesis and Motion

Breast Tomosynthesis. Bob Liu, Ph.D. Department of Radiology Massachusetts General Hospital And Harvard Medical School

GE Healthcare. Senographe 2000D Full-field digital mammography system

1.1.Clinical Artefacts

New spectral benefi ts, proven low dose

Digital Breast Tomosynthesis

Practical Aspects of Medical Physics Surveys of Mammography Equipment and Facilities

8/2/2017. Radiologist Responsibilities. Radiologist Responsibilities. Medical Physicist Mammography Equipment Evaluation and Annual Survey

Understanding Quantra 2.0 User Manual

Digital radiography (DR) post processing techniques for pediatric radiology

MAMMOGRAPHY - HIGH LEVEL TROUBLESHOOTING

Mammography: Physics of Imaging

Surveying and QC of Stereotactic Breast Biopsy Units for ACR Accreditation

Quality Control of Full Field Digital Mammography Units

THE ART OF THE IMAGE: IDENTIFICATION AND REMEDIATION OF IMAGE ARTIFACTS IN MAMMOGRAPHY

Collimation Assessment Using GAFCHROMIC XR-M2

STEREOTACTIC BREAST BIOPSY EQUIPMENT SURVEYS

Mammography Solution. AMULET Innovality. The new leader in the AMULET series. Tomosynthesis, 3D mammography and biopsy are all available.

4.1. Accurate: The information is a true reflection of the original observation.

Image Quality. HTC Grid High Transmission Cellular Grid provides higher contrast images

KODAK DIRECTVIEW CR Mammography Feature User s Guide

Introduction. Digital Mammography QA: Comparing the Manufacturers Recommendations. What is QC and why is it important? Review & compare QC tests

GE Healthcare. Performa. High-performance breast imaging

Improved Tomosynthesis Reconstruction using Super-resolution and Iterative Techniques

RAD 150 RADIOLOGIC EXPOSURE TECHNIQUE II

DICOM Correction Item

Image Quality. HTC Grid High Transmission Cellular Grid provides higher contrast images

Quality Control for Stereotactic Breast Biopsy. Robert J. Pizzutiello, Jr., F.A.C.M.P. Upstate Medical Physics, Inc

2012 :15th SESSION of ESMP

Influence of different iteration levels in fourth generation iterative reconstruction technique on image noise in CT examinations of the neck

2 nd generation TOMOSYNTHESIS

Open. the Digitized world. Fuji Computed Radiography

4/19/2016. Quality Control Activities for the RadiologicTechnologist. Objectives. 3D Tomosynthesis QC differences

Challenges in Mammography: Part 1, Artifacts in Digital Mammography

Breast Imaging Basics: Module 10 Digital Mammography

Aspire HD. Program Manual. 2nd Edition - October 2012

A Practical Overview of the Clinical and Operational Impact of Computed Radiography(CR) Implementations. Shirley Weddle, RT(R)(M), CIIP, BBA

Performance and care. all in one

GE Healthcare. Senographe Crystal The choice is crystal clear

Y11-DR Digital Radiography (DR) Image Quality

Image Quality in Digital Mammography: Image Acquisition

Redefining Ergonomics

Imaging Technique Optimization of Tungsten Anode FFDM System

TECHNICAL DATA. GIOTTO IMAGE SDL/W is pre-arranged for Full Field Digital Biopsy examination with the patient in prone position.

Nuclear Associates

2017 West Coast Educators Conference Orlando. Projection Geometry. 1. Review hierarchy of image qualities (amplified version):

SECTION I - CHAPTER 1 DIGITAL RADIOGRAPHY: AN OVERVIEW OF THE TEXT. Exam Content Specifications 8/22/2012 RADT 3463 COMPUTERIZED IMAGING

NATIONWIDE EVALUATION OF X-RAY TRENDS (NEXT) SUMMARY OF 2003 FLUOROSCOPY SURVEY

TORNIER BLUEPRINT. 3D Planning + PSI SCAN PROTOCOL

Wide-Detector CT for TAVR Planning:

DISC QC/QA Program for Digital Imaging Systems using the DR Radchex Plus Meter

ACR AAPM SIIM Practice Guideline for Determinants of Image Quality in Digital Mammography

FFDM in the Field: Physicist's Role in the QC of Mammography Laser Printers May Carl R. Keener, Ph.D., DABMP, DABR

PRACTICE GUIDELINE FOR DETERMINANTS OF IMAGE QUALITY IN DIGITAL MAMMOGRAPHY

Maximizing clinical outcomes

Image Display and Perception

TOPICS: CT Protocol Optimization over the Range of Patient Age & Size and for Different CT Scanner Types: Recommendations & Misconceptions

Luminos drf Max. Taking 2-in-1 to the MAX in radiography and fluoroscopy. siemens.com/luminos-drf-max

Studies on reduction of exposure dose using digital scattered X-ray removal processing

Studies on reduction of exposure dose using digital scattered X-ray removal processing

LECTURE 1 The Radiographic Image

Threshold contrast visibility of micro calcifications in digital mammography

Quality control for digital mammography: Part II recommendations from the ACRIN DMIST trial

NATIONWIDE EVALUATION OF X-RAY TRENDS (NEXT) TABULATION AND GRAPHICAL SUMMARY OF 2001 SURVEY OF ADULT CHEST RADIOGRAPHY

12/21/2016. Siemens Medical Systems Research Agreement Philips Healthcare Research Agreement AAN and ASN Committees

NEMA XR X-ray Equipment for Interventional Procedures User Quality Control Mode

7/20/2014. Outline. Outline. Disclosures. Learning Objectives. SBB: Practical Aspects of ACR Accreditation, QC and ACR On Site Surveys

4.0 How to Turn On the Selenia Dimensions

Published text: Institute of Cancer Research Repository Please direct all s to:

Predicted image quality of a CMOS APS X-ray detector across a range of mammographic beam qualities

GE Healthcare. Senographe Crystal The choice is crystal clear

Exposure Indices and Target Values in Radiography: What Are They and How Can You Use Them?

Mammography is a radiographic procedure specially designed for detecting breast pathology Approximately 1 woman in 8 will develop breast cancer over

IHE Radiology Technical Framework Supplement. Stereotactic Mammography Image (SMI) Trial Implementation

Minnesota Rules, Chapter 4732 X-ray Revision

Facility, Unit and Test Equipment Data

Full-field digital mammography with grid-less acquisition and software-based scatter correction: Investigation of dose saving and image quality

Digital Mammography Quality Control for the Mammographic Technologist

1. Carlton, Richard R., and Arlene M. Adler. Principles of Radiographic Imaging: An Art and a Science, 5th edition (2013).

DR _ solutions. We understand that customers don t need just products, they want. solutions

Advanced digital image processing for clinical excellence in fluoroscopy

Exposure System Selection

Digital Quality Control Phantom for X-ray Mammography Image

DELWORKS DR MEDICAL. take the next step

siemens.com/luminos-fusion Luminos Fusion The 2-in-1 system that fits your needs and fits your budget

IBEX TECHNOLOGY APPLIED TO DIGITAL RADIOGRAPHY

REQUIREMENTS FOR LICENCE HOLDERS WITH RESPECT TO QUALITY CONTROL TESTS FOR DIAGNOSTIC X-RAY IMAGING SYSTEMS

Get more from your images with Symphony Image Processing

Simultaneous Multi-Slice (Slice Accelerated) Diffusion EPI

Overview of Safety Code 35

Balancing Bandwidth and Bytes: Managing storage and transmission across a datacast network

FFDM -FCRm QC Requirements- What You REALLY Need to Know

Ansur TNT Users Manual. Plug-In

Digital Mammography Quality Control for the Mammographic Physicist

Practical Medical Physics Session: TG-151 Dose Monitoring. August 5, 2013 Katie Hulme, M.S.

Get more from your images with Symphony Image Processing

Architecture of Quality Imaging Mary K. Henne, MS, CNMT, RDMS, RVT Ultrasound Education Specialist GE Healthcare

Test Equipment for Radiology and CT Quality Control Contents

Transcription:

Patient-Assisted Compression Impact on Image Quality and Workflow Senographe Pristina In 2017, GE Healthcare s Senographe Pristina ( Pristina ) was approved by the FDA using the standard technologist-controlled () compression mode. GE Healthcare sought to improve the features on the Pristina to include an optional patient-assisted () compression mode. The patient-assisted feature enables the patient to personally refine breast compression using a hand-held remote control after the compression has been initiated by the technologist, which is required to ensure proper breast positioning. To extend the compression modes available on Pristina, GE Healthcare conducted a single-blind prospective randomized research study that compared the acceptability of overall image quality in two-view (CC and MLO) breast images acquired using and compression modes. Clinical breast images were obtained from an Ethics Committee-approved clinical case collection study, where two-view (CC and MLO), unilateral and compression image sets were independently evaluated by two (2) MQSAqualified readers who were blinded to the clinical care and histories of the patients, and the compression mode used during image acquisition. The results of the study clearly demonstrated that compared to images where compression was applied solely by the technologist, patient-assisted compression produces images of similar quality. Image Acqusition and Procedural Data Collection Consenting female patients presenting for screening exams with digital mammography were enrolled in this study and had one breast ( Breast of Interest ) imaged with both and compression modes in standard CC and MLO views; the other breast was imaged with only, per standard of care. Patients were randomly placed into two groups: a) the breast of interest (BoI) imaged first or b) BoI imaged second, so that approximately half the patients were assigned into each group. For each patient s BoI, compression and image acquisition were conducted first, consisting of both CC and MLO views. No imaging was administered using compression mode. Following compression image acquisition, compression and standard of care imaging were conducted on the same BoI. Two (2) study-specific image sets of the BoI were collected for each subject. Image sets consisted of CC and MLO views obtained using compression and CC and MLO views obtained using compression. Study staff collected and documented the following workflow and procedural data on case report forms (CRF) for each subject: size of paddle used for imaging, amount of time required to educate subject on compression mode, laterality and order of imaging (first or second) for the BoI, start times for and compressions, technologist intervention(s) during compression, and incidence and reason for any imaging. Compression force, breast thickness, entrance skin air kerma (ESAK), and acquisition parameters were acquired from the DICOM headers and recorded on CRFs by the study staff. (ESAK is determined on the mammography system and is used to calculate Average Glandular Dose.) Image Attribute Reviews Evaluable two-view (CC and MLO), unilateral and compression image sets were independently evaluated by two (2) MQSAqualified readers who were blinded to the

clinical care and histories of the patients, and the compression mode used during image acquisition. A third reader provided adjudication when there was disagreement between the two (2) readers evaluations of overall clinical image quality. Images were displayed on mammography review workstations per the randomization scheme. Each reader evaluated each image set and documented acceptability ( acceptable or unacceptable ) of overall image quality and image characteristics on CRFs. Readers indicated whether a acquisition would be required in standard practice, and if yes, which view (CC, MLO, or both) would require reimaging. Readers also specified the breast density for each image set. Study Patients A total of 30 patients completed the study, with ages ranging from 1 to 83 years old. Of the 30 patients, 16 completed the study from the BoI first group and 1 completed from the BoI second group. A total of 60 image sets (30 and 30 image sets) were acquired from the 30 completed patients, thus representing the analysis set for all endpoint analyses, unless otherwise specified in the following sections. Overall Image Quality All 60 image sets (30 and 30 image sets) were evaluated for acceptability of overall image quality by two (2) readers. Both readers evaluated all image sets from all patients. Due to disagreement between Readers 1 and 2, Reader 3 provided adjudication for four () image sets and five (5) image sets. Including Reader 3, there were a total of 129 image evaluations included in the primary endpoint analysis of overall image quality (Table 1). Overall, 100% of image sets evaluated for both and compression were found to be of acceptable image quality by the readers (Table 1). When compared to image sets, 100% of the image sets were of equal or better acceptability for image quality ( Table 2). Table 1 Acceptability of overall image quality, both overall and by reader Reader 1 Reader 2 Final, adjudicated Acceptability Overall Image Quality [%(n/n)] [95% CI] [%(n/n)] [95% CI] Unacceptable 0 0 86.7% (26/30) [69.3%,96.2%] Unacceptable 13.3% (/30) [3.8%,30.7%] 83.3% (25/30) [65.3%,9.%] 16.7% (5/30) [5.6%,3.7%] Unacceptable 0 0 Table 2 Patient-assisted () versus technologist-controlled () compression, overall image quality Acceptability of Overall Image Quality ( vs ) equal or better than %(n/n) 95% CI Reader 1 Yes 100.0% (30/30) Reader 2 Final, adjudicated Yes 96.7% (29/30) [82.8%,99.9%] No 3.3% (1/30) [0.1%,17.2%] Yes 100.0% (30/30) Repeat Image Acquisition Results Repeat imaging for and image sets were indicated by technologists during image acquisitions and by readers during image evaluations. Technologists indicated only one (1) image set that would have undergone imaging due to positioning and specified as clipping

bottom of breast. For, however, four () Technologist-indicated Reader-indicated acquisitions acquisitions (Reader 2 only) Reason for Positioning image sets were ed by the technologists due to positioning (n=) and other reasons (n=3), which were specified "opposite BoI skin fold. There were no s indicated due to compression ( Table 3). Technologist-indicated acquisitions Reason for Positioning 1 Reader-indicated acquisitions (Reader 2 only) View for Both (CC & MLO) Reader 2 was the only reader with unacceptable overall image quality evaluations, and therefore, was the only reader who was prompted to indicate if imaging would be requested, and if yes, which view. There were nine (9) image sets that Reader 2 evaluated as unacceptable, but only eight (8) image sets were indicated for imaging four () and four () image sets. CC, MLO, and both (CC and MLO) views were indicated for image sets, whereas MLO and both were indicated for the image sets ( Table 3). Table 3 Summary of imaging results Image Attribute Review Results Readers 1 and 2 evaluated all and image 1 1 Compression - - CC 1 0 Other, specify 1 3 MLO 2 3 1 (3.3%) 1 View for Both (CC & MLO) 1 1 Compression - - CC 1 0 Other, specify 1 3 MLO 2 3 1 (3.3%) sets for a pre-defined set of image attributes Attribute Assessed Breast positioning, assessing coverage of the breast on craniocaudal and mediolateral oblique view, separately Exposure, assessing visualization of the adipose and fibroglandular tissues, and visualization of breast tissue underlying the pectoralis muscle, separately Breast compression, assessing overlapping breast structures, uniformity of exposure of fibroglandular tissues, adequacy of penetration of thicker portions of the breast, exposure of thinner areas, and motion unsharpness Image contrast for differentiation of subtle tissue density differences Sharpness, assessing the edges of fine linear structures, tissue borders, and benign calcifications Tissue visibility at the skin line Noise, i.e., noise obscuring breast structures or suggestive of structures not actually present Artifacts due to image processing, detector failure and other factors external to the breast on hard-copy and soft-copy displays Image Attribute Review Final, Adjudicated (% ) (Table ). When there was disagreement between the readers for overall image quality, Reader 3 provided adjudication. The final adjudicated results showed 100% of the image sets were acceptable for all eight (8) image attributes, as determined by the (% ) 100.0% 96.7%

readers. One-hundred percent (100%) of image sets were deemed acceptable for all attributes, except for the first attribute (see Table ). Table Summary of final, adjudicated image attribute reviews Imaging Duration Results Prior to imaging, the study staff documented the amount of time it took to educate each subject on use of the compression mode. It took 1-2 minutes for 70% of patients (n=21) and 3-5 minutes for the other 30%. image acquisition time was also collected for each image set. Image acquisitions using and compression modes lasted an average of 3.9 and 3. minutes, respectively (Table 5). Four () image sets were excluded from the analysis because the patients required imaging per standard of care; imaging was not allowed for compressions. One subject s and image sets were also excluded from the analysis due to a protocol deviation associated with the order of and imaging on the BoI. In total, one (1) image set and five (5) image sets were excluded from this analysis. Table 5 Summary of imaging durations by compression type Imaging Duration (minutes) Mean (SD) 3.9 (1.0) 3. (1.0) Median.0 3.0 Range (3.0-6.0) (2.0-5.0) Table 6 Summary of procedural characteristics Procedural Characteristics The mean compression force of compressions for CC and MLO views were 9.1 and 8.2 dan, respectively, versus 8.3 and 8.7 dan, respectively, for compressions. Breast thicknesses for and compressions were 52.0 and 53.2 mm, respectively, for CC views and 5.6 and 53.7 mm, respectively, for MLO views. Mean ESAK for compression was.5 mgy for CC views and.9 mgy for MLO views. ESAK for compression was.7 mgy for CC views and.8 mgy for MLO views ( Table 6). Acquisition parameters for each compression mode are presented in Table 6. Conclusions Parameters Compression Force (dan) Patient-Assisted () Compression CC MLO Technologist-Controlled () CC Compression MLO Mean (SD) 9.1 (3.0) 8.2 (2.7) 8.3 (2.0) 8.7 (2.2) Median 8.0 8.0 8.0 9.0 Range (.0-15.0) (3.0-1.0) (.0-11.0) (.0-12.0) Breast Thickness (mm) Mean (SD) 52.0 (1.3) 5.6 (16.1) 53.2 (1.0) 53.7 (15.1) Median 50.9 53.7 51.1 52.9 Range (29.-96.0) (29.8-98.7) (29.5-95.7) (29.-93.5) Entrance Skin Air Kerma (ESAK, mgy) Mean (SD).5 (2.).9 (2.6).7 (2.).8 (2.) Median 3.7.1 3.8 3.8 Range (2.6-12.5) (2.5-13.5) (2.6-12.7) (2.5-12.6) Acquisition Parameters (kvp) Mean (SD) 33.3 (0.5) 33.3 (0.5) 33.3 (0.) 33.3 (0.5) Median 33.0 33.0 33.0 33.0 Range (33.0-3.0) (33.0-3.0) (33.0-3.0) (33.0-3.0) Acquisition Parameters (mas) Mean (SD) 35.7 (16.3) 38.5 (18.0) 36.9 (16.9) 37.6 (17.0) Median 30.0 32.0 30.5 31.0 Range (22.0-88.0) (22.0-95.0) (23.0-91.0) (22.0-91.0) A total of 60 evaluable (n = 30) and (n = 30) two-view, unilateral image sets were collected from 30 distinct female patients using the and compression modes of Senographe

Pristina in a clinical setting. Mammography procedures were performed by qualified personnel, and independent image evaluations were conducted by two (2) MQSAqualified readers, with adjudication performed by a third MQSA-qualified reader in cases of disagreement about overall image quality. The final overall clinical image quality was acceptable for 100% of the compression image sets evaluated in this study, and 100% of image sets were of equal or better image quality when compared to the image sets. Furthermore, 100% of the image sets were deemed acceptable for all image attributes. These cumulative results substantiate that compared to images where compression was applied solely by the technologist, patientassisted compression produces images of similar quality. In terms of workflow, the study demonstrated that patient education took 1-2 minutes in most cases, and the image acquisition time was minimally impacted by use of C. There was only one (1) instance of technologist intervention with C in this study, and subgroup analyses of patients with their BOI imaged first or second did not reveal significant differences between the two groups. These results confirm that the device is easily operable by patients, when guided by their technologist, and the education received from the technologist at the start of the exam and the guidance received during device operation were sufficient to achieve an adequate level of compression.