ETP ELECTRON MULTIPLIERS

Similar documents
ETP ELECTRON MULTIPLIERS

ETP ELECTRON MULTIPLIERS Care and Handling

Electron Multipliers. Introduction to ETP Electron Multipliers Selection by Technique: LC-MS 242

A New Class of Robust Sub-nanosecond TOF Detectors with High Dynamic Range

Removing Dead Time Variations Due to Detector Pulse Width in Ion Counting Systems

Nuclear Associates

Industrial X-ray Tube KL

Adhesives & Solubility

Cleaning & Maintenance Advice Textured Glass.

User instructions Compound laboratory microscope

CHEMSET Reo 502 EF TECHNICAL DATA SHEET. Extra Heavy Duty Pure Epoxy Anchoring Adhesive

User instructions Compound laboratory microscope

Basic Principles and Operation of Transformer

PRODUCT GUIDE: CATALYZED PRIMERS. Version:

HIGH PERFORMANCE FILTERS FOR COMPRESSED AIR & GAS

Klaran WD Series UVC LEDs

CORTEC Technical Support VCI Emitters - Common Questions:

MK33-W Capacitive Humidity Sensor For oil measurement applications

By using the system correctly you will obtain very high quality repairs.

USER MANUAL. Maxwell Technologies Ultracapacitor Energy Storage Modules. Models: BMOD0058 E016 B02 BMOD0006 E160 B02. Document

Rotating Anode X-ray Tube

1. CONSTRUCTION InGaN on SiC blue surface mount LEDs with colorless epoxy. 74mW Forward Current F. 20mA Peak Forward Current

NOVALYNX CORPORATION MODEL 110-WS-16TH TEMPERATURE AND RELATIVE HUMIDITY SENSOR INSTRUCTION MANUAL

MANUAL TNS Series Nut Splitter

TRANSDUCER IN-LINE AMPLIFIER

Klaran WATER DISINFECTION

V349 (CLF5043) Halide Free No Clean Core Wire Fine Wire Applications

742A Series Resistance Standards

Technical Product Information

Power LED I TECHNICAL DATA

CUPRON NEXT TECHNICAL DATA SHEET

Rotating Anode X-Ray Tube Housing Assembly. General Data. IEC Classification... Class I

Hollow Cathode Lamps

Standard Practices for Producing Films of Uniform Thickness of Paint, Varnish, and Related Products on Test Panels 1

Application Note. Soldering Guidelines for Surface Mount Filters. 1. Introduction. 2. General

PolyCarb has good resistance to many chemicals. Some chemicals may harm the PolyCarb sheets. Contact us when in doubt about any chemical.

RD-90 Developer-Replenisher F-Fix Fixer-Replenisher

Nuclear Associates

OPERATION MANUAL. Model 1010 TE Survey Meter. August Health Physics Instruments 330 South Kellogg Ave, Suite D Goleta, CA 93117

High Breakdown Voltage, Fully Depleted Series Large Active Area Photodiodes

LC/MS/MS. Page Header. triple quadrupole mass spectrometer.

Application Note. Soldering Guidelines for SMPS Multilayer Ceramic Capacitor Assemblies

Specific Lighting Product Data Sheet LTPL-C035BH450 Spec No.: DS Effective Date: 12/03/2016 LITE-ON DCC RELEASE

1/4 Sheet Palm Sander

Anatomix Plus X-Ray System

LUDLUM MODEL 182 RADON FLASK COUNTER. June 2011 Serial Number PR and Succeeding Serial Numbers

Product Bulletin

3M Scotchkote System Application Guide. High Performance Coating System for Degraded Plastisol Cladding Surfaces.

Rotating Anode X-ray Tube Assembly

Screen Printing on Rowmark Material

Next, know your supply voltage. It should be a few volts above the LED forward voltage for reliable, stable LED operation

TRANSDUCER IN-LINE AMPLIFIER

Frame Finishing Instructions

WATERFLUX 3000 Quick Start

Heiland electronic GmbH TD / TD1 / TD2. B&W-Densitometers. USERS MANUAL Version 5

Ox-RAC-08 Ribbon Angled Fiber Cleaver User Manual

INSTRUCTION MANUAL. Force Transducer Output Tube Repair Kit

TECHNICAL ADVICE. CARE & MAINTENANCE Colour fading

USER MANUAL. Maxwell Technologies BOOSTCAP Energy Storage Modules. User Manual for 15V Modules: 20 F, 23 F, 53 F, 58 F 15 Volts DC

General information 1 / 17

Anatomix RP X-Ray System

INSTRUCTION MANUAL INF Fax: (503)

SML-D1 series EXCELED TM

Solar Collector. (Evacuated Tube) Installation Guide, Operation & Maintenance Manual. Working towards a cleaner future

Rotating Anode X-Ray Tube Housing Assembly. General Data. IEC Classification... Class I Type B

ADHESIVE SWITCHABLE FILM

The Exclamation Point alerts you to particular cautions, hazards and information.

Operation manual. Model 24 Infrasound Sensor

Dominant wavelength (nm) or CCT (K) Royal Blue 455nm 465nm 65 mw. Blue 465nm 475nm 4.5 lm. Green 520nm 535nm 18 lm. Amber 585nm 595nm 8.

(

Student Sheet. P003S: Making Paint with Minerals

Roline L1 Series. Humidity - Temperature Transmitters INSTRUCTION MANUAL

HYGROCHIP 1/6 DIGITAL HUMIDITY SENSOR HYT-271

M1 Series. Humidity - Temperature Transmitter INSTRUCTION MANUAL

The Exclamation Point alerts you to particular cautions, hazards and information.

ALMD-CY3G-YZx02-DS100

A GUIDE TO BETTER SPRAYING

EVERLIGHT ELECTRONICS CO.,LTD.

Common Coating Problems

2K weather resistance clear coat A850

Stationary Anode X-Ray Tube. General Data

3M Industrial & Transportation Business. Windows and Doors Design and Assembly Solutions. Windows. and Doors. Solutions

Operating Manual 6 Industrial Bench Grinder ATBG280/

Reference only P.1 /14

Gluing windows with SABA Glasstack 760

ENS31 Automatic Isolation Unit Product Description

EVERLIGHT ELECTRONICS CO., LTD.

SF 6 Characteristics and Re-use Concept

Infrared Gun. Perfect For. Easy To Use. Features. Get Professional Results Every Time! Battery Installation. Model: IN1022

Circular Drying Time Recorder

I P I OUT R L. Applications. Standards. N 52.D ; Page 1/8

D D. Wood Doors Catalogue

Elcometer Muller Laboratory Grinder

Mylar Speaker Specification. Series FMS. Model Number: FMS15R03N08XTHN. Version 1.0

INSTALLATION INSTRUCTIONS FOR #101NS & #101QR FIXED BASE

5-Wire Analog Resistive Touch Screen Specification Guide

Base model features TTL buffer compatible modulation input level and 24/28Vdc supply.

Instruction Manual. For Type E Plasma Tube Electrodes

POTTING BENCH. SAFETY EQUIPMENT Gloves Goggles Dust mask

ML7520 ML7530 DIOPTER ADJUSTMENT RING BINOCULAR BODY, INCLINED 30. (a) Field Iris Control Lever. (c) Filter Slots EYEPIECES, KHW10X

Transcription:

ETP ELECTRON MULTIPLIERS ] Care and Handling Electron Multipliers division of

Maintenance, Storage, and Handling of ETP Electron Multipliers ETP Electron Multipliers from SGE are being used in a growing number of analytical applications, including mass spectrometry, surface science, and general physics. They offer the higher performance required by modern instrumentation and have a proven reputation for reliability, ruggedness, and performance. This booklet is intended as a guide to the maintenance, storage, and handling of ETP Electron Multipliers. Following the basic procedures described herein will enable you to obtain the optimum life and performance from your ETP Electron Multiplier. i

CONTENTS: Title Pg Storage of ETP Electron Multipliers 1 Care and Handling of ETP Electron Multipliers 2 Cleaning ETP Electron Multipliers 3 Installing and Operating ETP Electron Multipliers 5 Obtaining Optimum Lifetime from ETP Electron Multipliers 6 ii

Storage of ETP Electron Multipliers The critical dynode surfaces of an ETP Electron Multiplier consist of a specially formulated and applied aluminum oxide material which is very rugged and completely stable, even when repeatedly exposed to room atmosphere. The original package as supplied by ETP is designed for long term storage. The multiplier is enclosed in two sealed plastic bags. The outer bag contains a silica gel package to minimize adsorption of moisture which could otherwise lead to excessive outgassing upon installation. In its original packaging, the shelf-life of an ETP Electron Multiplier is guaranteed for up to two years from the date of shipment! If it is necessary to store the multiplier outside its original packaging, it should be kept in a clean, dust-free, low-humidity environment, such as a glass desiccator. 1

Care and Handling of ETP Electron Multipliers Standard high-vacuum techniques are recommended when handling ETP Electron Multipliers in order to minimize the possibility of contamination. Clean, powder-free gloves or finger cots should be worn to prevent transfer of skin oils to critical or mounting surfaces of the multiplier. All tools, mountings, and equipment should be thoroughly cleaned prior to coming in contact with the multiplier. Care should also be taken to minimize exposure of the multiplier to dust or other airborne contamination. Dust particles within the detector may result in increased background noise during operation. Exposure of the multiplier to high humidity should be avoided in order to minimize outgassing during initial operation. If the multiplier exhibits excessive noise due to adsorbed moisture, this condition may be rectified by either leaving the multiplier in high vacuum for about 24 hours, or by using a simple vacuum bake procedure. 2

The multiplier may be baked in vacuum at up to 350 C. NOTE: Some models may not be baked above 100 C. See the installation notes supplied with each ETP Electron Multiplier for information concerning temperature limitations. For multiplier models that can be baked to 350 C in vacuum, then if baking must be done at atmosphere, temperature should be limited to 120 C. The multiplier should be heated and cooled slowly to avoid damage due to thermal shock. The rugged design of ETP Electron Multipliers reduces the risk of damage during handling; however, an electron multiplier should be considered a precision component, and all reasonable care should be used when installing or removing. Cleaning ETP Electron Multipliers In normal operation it is not recommended to claen the multiplier in any way. However if the multiplier becomes contaminated (for example as a result of vacuum system failure) then in some cases it may be possible to clean the multiplier. Caution: Never operate an electron multiplier suspected of being contaminated with pump oil until it has first been cleaned as described overleaf. Operation of a contaminated multiplier will permanently damage the critical surfaces. 3

Not all Active Film Multiplier models can be cleaned using an ultra-sonic cleaner. Check with SGE technical support to determine if your model multiplier can be cleaned. If your multiplier can be cleaned, then non-polar solvents may be used. The recommended procedure is as follows (any deviation may void the warranty): Cleaning Equipment 1. Ultra-sonic cleaner 2. Clean glass beaker (100mL or larger) 3. AR- or HPLC-grade Heptane solvent Cleaning Method * 1. Place the multiplier in the beaker and fill with Heptane until multiplier is fully immersed. 2. Place the beaker in the ultra-sonic cleaner and clean for approximately ten minutes. 3. Dispose of solvent in an appropriate manner and refill the beaker. 4. Repeat steps 1-3 as required. 5. Remove the multiplier from the beaker and allow to dry at room temperature. 6. Bake the multiplier for three hours in air at 100 C. *Check with SGE technical support before proceeding. 4

Installing and Operating ETP Electron Multipliers ETP Electron Multipliers generally require no preconditioning; however, when first installed, it is recommended that applied voltage not exceed 2200V for the first 24 hours of operation. This voltage limit is 3200V for fast Time-Of-Flight multipliers with model number between 14870 and 14899. Multiplier leads should be positioned for a minimum clearance of 3mm between the lead and any other part of the multiplier mount or vacuum chamber. The multiplier should be mounted so that there are no sharp objects in close proximity to the entrance aperture. This condition could result in noise during operation. Operating pressure should be as low as possible for best lifetime and minimum noise. Under no circumstances should the multiplier be operated at a pressure in excess of 1x10-4 mbar to avoid risk of damage due to discharge or arcing. 5

Obtaining Optimum Lifetime from ETP Electron Multipliers The gain of a correctly installed electron multiplier will fall very gradually over time, requiring the applied voltage to be periodically increased. This is a normal part of the multiplier aging process. Total operational life varies depending on application and may be affected by a number of factors: Operating Environment: Lower operating pressure and lower partial pressure of hydrocarbons in the vacuum chamber will result in increased operational life, all other conditions remaining the same. Operating Gain: Operation should always be at the lowest applied voltage consistent with good signal acquisition. In general, the lower the gain, the longer the life of the multiplier. 6

Output Current: Multiplier lifetime is inversely proportional to average output current. Excessive input signals should be avoided, especially when operating at high gain. Rapid degradation of multiplier performance may result from failure of the vacuum pumping system. This may result in severe contamination of the detector or damage due to arcing caused by the sudden increase in chamber pressure. If vacuum failure occurs, the multiplier should be cleaned before being operated again. Degradation caused by vacuum failure may be permanent and irreversible, especially if the multiplier high voltage was ON during the vacuum failure. For technical or ordering information contact your nearest ETP or SGE office or visit www.etpsci.com. 7

Electron Multipliers USA SGE, Incorporated 2007 Kramer Lane, Austin, Texas, 78758 Toll Free: (800) 945 6154 Tel: (512) 837 7190 Fax: (512) 836 9159 E-mail: usa@sge.com Australia and Pacific Region SGE International Pty. Ltd 7 Argent Place, Ringwood, VIC 3134 Tel: +61 (0) 3 9837 4200 Fax: +61 (0) 3 9874 5672 E-mail: support@sge.com Europe SGE Europe Ltd. (UK) 1 Potters Lane, Kiln Farm, Milton Keynes, MK11 3LA, UK Tel: +44 (0) 1908 568844 Fax: +44 (0) 1908 566790 E-mail: uk@sge.com Visit www.etpsci.com for technical information and on-line training on ETP Electron Multipliers Copyright 2003 SGE International Pty.Ltd. All rights reserved BR-0061-A Rev11 03/03