Agilent How To Accurately Evaluate Low ESR, High Q RF Chip Devices. Application Note

Similar documents
Agilent 4285A Precision LCR Meter

Agilent Introduction to the Fixture Simulator Function of the ENA Series RF Network Analyzers: Network De-embedding/Embedding and Balanced Measurement

Agilent AN Balanced Circuit Measurement with an Impedance Analyzer/LCR Meter/Network Analyzer Application Note

Agilent 8491A/B, 8493A/B/C, 11581A, 11582A and 11583C Coaxial Attenuators dc to 26.5 GHz

System Cabling Errors and DC Voltage Measurement Errors in Digital Multimeters

Agilent E9300 Power Sensors E-Series Technical Overview

Agilent PNA Microwave Network Analyzers

Agilent 86030A 50 GHz Lightwave Component Analyzer Product Overview

6 Tips for Successful Logic Analyzer Probing

Agilent 87415A, 87400A Microwave Amplifiers

Agilent 8761A/B Microwave Switches

Grundlagen der Impedanzmessung

Time-Domain Response of Agilent InfiniiMax Probes and Series Infiniium Oscilloscopes

Agilent Technologies Impedance Measurement Handbook December 2003

Agilent PNA Series RF Network Analyzers

Agilent 8703B Lightwave Component Analyzer Technical Specifications. 50 MHz to GHz modulation bandwidth

Phase Noise Measurement Personality for the Agilent ESA-E Series Spectrum Analyzers

Product Note E5100A-2

Agilent E8460A 256-Channel Reed Relay Multiplexer

Keysight Technologies Power of Impedance Analyzer

Agilent 8644A-1 Phase noise test with the Agilent 8644A and 8665A Signal Generators Product Note

Characterizing High-Speed Oscilloscope Distortion A comparison of Agilent and Tektronix high-speed, real-time oscilloscopes

Agilent N1911A/N1912A P-Series Power Meters and N1921A/N1922A Wideband Power Sensors. Data sheet

Agilent 4284A/4285A Precision LCR Meter Family

Agilent PSA Series Spectrum Analyzers Noise Figure Measurements Personality

Agilent E8247/E8257C PSG CW and Analog Signal Generators

Models 885 & 886 LCR METER OPERATING MANUAL

Agilent 4287A RF LCR Meter 1 MHz - 3 GHz. Technical Overview

Agilent E4438C ESG Vector Signal Generator Differential I/Q outputs. Product Note


Agilent E8267C/E8257C/E8247C PSG

Agilent 83440B/C/D High-Speed Lightwave Converters

Agilent Accessories Selection Guide For Impedance Measurements. December 2008

Agilent PSA Series Spectrum Analyzers Self-Guided Demonstration for Phase Noise Measurements

MEMS On-wafer Evaluation in Mass Production Testing At the Earliest Stage is the Key to Lowering Costs

Agilent 4070 Series Accurate Capacitance Characterization at the Wafer Level

PCI Express Receiver Design Validation Test with the Agilent 81134A Pulse Pattern Generator/ 81250A ParBERT. Product Note

Agilent 970-Series Handheld Multimeters Data Sheet

Agilent 8902A Measuring Receiver Product Note

Agilent 81600B All-band Tunable Laser Source Technical Specifications December 2002

Agilent Combining Network and Spectrum Analysis and IBASIC to Improve Device Characterization and Test Time

Agilent PN 4395/96-1 How to Measure Noise Accurately Using the Agilent Combination Analyzers

Accessories Selection Guide For Impedance Measurements. April 2005

Agilent dc Electronic Loads Models N3300A-N3307A

Agilent Equalization Techniques and OFDM Troubleshooting for Wireless LANs

Keysight E5061B-3L3/3L4/3L5 LF-RF Network Analyzer with Option 005 Impedance Analysis Function

Agilent PNA Microwave Network Analyzers

Keysight Technologies Accurate Capacitance Characterization at the Wafer Level

Agilent Highly Accurate Amplifier ACLR and ACPR Testing with the Agilent N5182A MXG Vector Signal Generator. Application Note

Agilent On-wafer Balanced Component Measurement using the ENA RF Network Analyzer with the Cascade Microtech Probing System. Product Note E5070/71-3

Advanced Test Equipment Rentals ATEC (2832)

Agilent AN Balanced Circuit Measurement with an Impedance Analyzer/LCR Meter/Network Analyzer Application Note

Agilent 4294A Precision Impedance Analyzer, 40 Hz to 110 MHz. Configuration Guide

Agilent 81980/ 81940A, Agilent 81989/ 81949A, Agilent 81944A Compact Tunable Laser Sources

Antenna Matching Within an Enclosure Part II: Practical Techniques and Guidelines

Agilent 8902A Measuring Receiver

Advanced Test Equipment Rentals ATEC (2832)

AN-1098 APPLICATION NOTE

Agilent N5250A PNA Millimeter-Wave Network Analyzer 10 MHz to 110 GHz

Jitter Analysis Techniques Using an Agilent Infiniium Oscilloscope

Keysight Technologies Using a Network and Impedance Analyzer to Evaluate MHz RFID Tags and Readers/Writers

Obtaining Flat Test Port Power with the Agilent 8360 s User Flatness Correction Feature. Product Note

Base Station Installation and Maintenance

Agilent EPM Series Power Meters

Agilent LCR Meters, Impedance Analyzers and Test Fixtures

Agilent 4294A Precision Impedance Analyzer 40 Hz to 110 MHz Technical Overview

Using an MSO to Debug a PIC18-Based Mixed-Signal Design

Agilent E8267C PSG Vector Signal Generator

Agilent 81662A DFB Laser Agilent 81663A DFB Laser Agilent Fabry-Perot Lasers

Advanced Test Equipment Rentals ATEC (2832)

Agilent AN Applying Error Correction to Network Analyzer Measurements

Agilent PSA Series Spectrum Analyzers Noise Figure Measurements Personality

Improving TDR/TDT Measurements Using Normalization Application Note

Agilent 4-Port PNA-L Network Analyzers

Keysight Technologies Impedance Measurements

How to Drive the Agilent Technologies Microwave Matrix and Transfer Switch via the E8483A Microwave Switch/Step Attenuator Driver.

Agilent 83711B and 83712B Synthesized CW Generators

Picking the Optimal Oscilloscope for Serial Data Signal Integrity Validation and Debug

Circuit Characterization with the Agilent 8714 VNA

Lecture 4. Maximum Transfer of Power. The Purpose of Matching. Lecture 4 RF Amplifier Design. Johan Wernehag Electrical and Information Technology

CHQ SERIES. Surface Mount Chip Capacitors: Ultra High Frequency

Single Device Solution for High Speed Testing and Frequency Sweeping IMPEDANCE ANALYZER IM3570

Keysight Technologies Performing Impedance Analysis with the E5061B ENA Vector Network Analyzer. Application Note

Advanced Test Equipment Rentals ATEC (2832)

Agilent Technologies 8114A 100 V/2 A Programmable Pulse Generator

Agilent 8766/7/8/9K Microwave Single-Pole Multi-Throw Switches

Exercises for the Antenna Matching Course

Lecture 4 RF Amplifier Design. Johan Wernehag, EIT. Johan Wernehag Electrical and Information Technology

expanding the possibilities

Agilent Upgrade Guide for the 8510 Vector Network Analyzer Product Note

Migrating 4195A to E5061B LF-RF Network Analyzer. April 2010 Agilent Technologies

Radio Frequency Electronics

Agilent U2000 Series USB Power Sensors. Data Sheet


There is a twenty db improvement in the reflection measurements when the port match errors are removed.

Highly Efficient Resonant Wireless Power Transfer with Active MEMS Impedance Matching

UWB Antenna Measurements with the 20 GHz E5071C ENA Network Analyzer

Agilent Technologies 3000 Series Oscilloscopes

Internal Model of X2Y Chip Technology

Keysight Technologies MEMS On-wafer Evaluation in Mass Production

Transcription:

Agilent How To Accurately Evaluate Low ESR, High Q RF Chip Devices Application Note 1369-6

Contents The Changing Requirements of RF Component Testing............ 3 Measurement challenges................................... 3 Advantages of the RF I-V Method.............................. 5 Calibration and Compensation in the RF Range.................. 7 Conventional calibration methodology........................ 7 Advanced calibration methodology........................... 8 High-Q measurement...................................... 9 Fixture compensation...................................... 10 Residual impedance error................................... 10 Electrical length error...................................... 10 Measurement Correlation and Repeatability..................... 11 Residual parameters at high frequencies...................... 11 Variance in residual parameter values........................ 12 A difference in contact condition............................ 12 A difference in open/short compensation conditions............ 13 Electromagnetic coupling with a conductor near the DUT........ 14 Defining short-bar residual inductance........................ 15 Measurement repeatability................................. 17 Summary................................................... 18 Appendix: Special Calibration Method for High-Q Measurement.... 19 Outline of the special calibration............................. 19 Details of the low loss capacitor (LCC) calibration.............. 20 Phase measurement error factors............................ 22 Related Literature........................................... 23 2

The Changing Requirements of RF Component Testing Several trends in wireless communications are changing the requirements for RF component testing. Today s hottest new applications including 3G UMTS (W-CDMA), Bluetooth, and wireless LAN applications are being offered at the 2 3 GHz frequency bands. Design engineers therefore must evaluate the RF components used in wireless communication devices at these higher operating frequencies, under realistic conditions, to ensure the quality and capability of their final products. While the operating frequencies of wireless devices are moving higher, device size is shrinking and so is the size of the device s RF components. For example, chip inductors as small as 0603 (0.6 x 0.3 mm or 0.2 x 0.1 inch) are being used today in mobile phones and PDAs. Smaller, higher frequency devices require higher performing RF circuits and more efficient battery operation. To meet increasingly stringent design requirements, design engineers use passive components such as chip inductors and capacitors with low equivalent series resistance (ESR) and high quality factor (high Q). Because component selection is so important in determining the final performance of a product, component parameters must be characterized as completely and accurately as possible. Proper measurement techniques, the right test instrument, and fixtures that can accommodate the minute size of today s RF chip devices are all necessary to enable thorough, accurate RF characterization and testing. This application note discusses how these needs are met with a new impedance measurement solution. Measurement challenges As the operating frequency of an RF component increases, so the selfresonant frequency (SRF) gets higher. Furthermore, impedance at the SRF becomes large. To characterize the component, we must be able to measure across a broad impedance range. The common practice of using a vector network analyzer and the reflection coefficient measurement technique gives very accurate results at 50 or 75 ohms. However, it does not cover the higher and lower impedance areas. This situation is illustrated in Figure 1. Impedance range [Ohm] 10 0K 10 K 1K 10 0 10 1 10 0m 10 m Inductor measurements How can we measure this area? Reflection coefficient method (VNA) 10 K 10 0K 1M 10 M 100M 1G 10 G Frequency range [Hz] Figure 1. The reflection coefficient method does not adequately measure the impedance range of a modern chip inductor. 3

Better results can be obtained with the Agilent E4991A RF impedance/ material analyzer, which uses a proprietary RF I-V technique for highly accurate impedance measurement to 3 GHz. Unlike the reflection coefficient method, RF I-V directly measures current and voltage (I-V) through the device under test, achieving an impedance measurement accuracy of ±0.8% over a wide impedance and frequency range. The area of coverage is shown in Figure 2. Impedance range (Ohm) 100K 10K 1K 100 10 1 100m 10m RF I-V method 10K 100K 1M 10 M 100 M 1G 10G Frequency range (Hz) Reflectioncoefficient method Figure 2. The RF I-V method used by the Agilent E4991A RF impedance analyzer extends the possibility of accurate impedance measurement across the RF range. The E4991A is calibrated using a low-loss capacitor standard for improved phase accuracy. As a result, it can measure components with small capacitance and inductance values over the frequency range, and offers unique advantages in characterizing high Q devices and evaluating temperature characteristics. A variety of test fixtures for different components and materials are available for the E4991A, with options for material measurement, temperature characteristic evaluation, and on-wafer measurement. 4

Advantages of the RF I-V Method To test today s higher frequency RF chip devices, the RF I-V method employs a simple operating principle, illustrated in Figure 3. A test signal is applied from the signal source to the device under test (DUT). A vector voltage meter measures the voltage (V 1 ) across the DUT. At the same time, current (I) flowing through the DUT is picked up by a balun transformer, converted into a voltage, and measured by a second vector voltage meter (V 2 ). The impedance of the DUT (Z) is then obtained from the values of V 1, V 2, and the resistance (R). R V 2 V 1 DUT I V 2 = I x R V 1 Z= = I V 1 R V 2 Figure 3. The RF I-V technique is based on a simple operating principle. The simple operating principle of the RF I-V method enables measurement over a wider frequency range than was previously possible using a single instrument. The E4991A, for example, has a frequency range of 1 MHz to 3 GHz. The analyzer s basic measurement accuracy of 0.8% is a dramatic improvement over the accuracy achieved using the more traditional reflection-coefficient technique and a vector network analyzer. Another advantage provided by the RF I-V method is improved measurement stability after calibration. Generally, changes in environmental temperature significantly influence the measurement of vector impedance at high frequencies. In a network analyzer, the reference and test channels of the voltage ratio detector (VRD) independently measure the input signals, as shown in Figure 4. Tracking errors, which occur when the relative gain and phase of the two channels vary after calibration is performed, are often the result of temperature variance. Because tracking errors cause the impedance measurement values to drift, these errors must be eliminated to improve measurement stability. 5

In the VRD section of the E4991A, also shown in Figure 4, the input signals V v and V i are multiplexed by the input switch. Each signal is measured alternately with two-channel VRD circuits in a measurement cycle. As a result of the vector voltage ratio calculation, tracking errors are canceled, which makes it possible to obtain stable impedance measurement results. Network analyzer Vector ratio detector Ref Impedance analyzer (RF I-V) Vector ratio detector V v Ch. 1 Zx Test V R V A Zx V i Ch. 2 V 1 V 2 Figure 4. Block diagrams of the vector ratio detector in a network analyzer and the E4991A. An additional advantage of the E4991A and RF I-V method is in the evaluation of temperature characteristics. The E4991A is not limited by the temperature dependencies that are evident in measurements using a network analyzer. For example, if we use a modern vector network analyzer and typical reflection coefficient (Γ) and S-parameter (S 11 and S 12 ) configurations to measure ESR, we see changes in the test results illustrated in Figure 5 at 100 MHz and 1 GHz for temperature settings of 18 C, 23 C, and 28 C. (The temperature range is satisfied with instrument specifications). In the same figure, however, we see that measurements made using the RF I-V method are stable across the temperature range, regardless of the frequency and the DUT s impedance. Capacitance Capacitance (pf) 10.25 100 M Hz DUT: 10pF capacitor 10.2 S2 1 10.15 S1 1 10.1 10.05 10 9.95 RF I-V 9.9 13 18 23 28 33 Temperature (deg C) Capacitance Capacitance (pf) 16 15 14 13 12 RF I-V S1 1 S2 1 1 G Hz 11 13 18 23 28 33 Temperature (deg C) 1. 5 0. 5-0.5-1 1 0-1.5 E4991A : RF I-V ESR DUT: 10pF capacitor ESR (Ω) ESR (Ω) S1 1 RF I -V S2 1 100 M H z -2 13 18 23 28 33 Temperature (deg C) 1. 6 1. 4 1. 2 1 0. 8 0. 6 E4991A : RF I-V ESR S1 1 RF I -V 1 G Hz S2 1 0. 4 13 18 23 28 33 Temperature (deg C) DUT: 10pF capacitor DUT: 10pF capacitor DUT: 10pF capacitor Figure 5. Temperature characteristics of a 10 pf capacitor. The measurements made using the E4991A and RF I-V method are stable across the 18 to 28 temperature range. 6

Calibration and Compensation in the RF Range Conventional calibration methodology Instrument calibration and compensation are processes that play a critical role in RF component testing. The conventional way to calibrate a network analyzer is with open, short, and load calibration. To perform this method of calibration, the open, short, and load reference terminations are connected to the test port and then each termination is measured. The calibration data is then stored in instrument memory and used to calculate and remove instrument errors. Impedance values for the reference terminations are indicated in two ways: with vector impedance coordinates and with a Smith chart, as shown in Figure 6. jx SHORT 0 ohm LOAD 50 ohm R SHORT LOAD OPEN 0 ohm Γ = 1 π 50 ohm Γ = 0 0 S Γ = 1 0 Error vector jx Note: OPEN impedance is infinite, so it is not shown in the graph. (a) Vector impedance plane (b) Smith chart Figure 6. Impedance values for reference terminations are graphed on the vector impedance plane and on the Smith chart. Although all three terminations are indispensable for calibration, the 50 Ω LOAD termination impedance is particularly important for ensuring precise calibration and has a large influence on the resulting measurement accuracy. The uncertainty of the LOAD termination impedance is represented by a circle that encloses the error vector, as we see in Figure 6a. The uncertainty of its phase angle increases with the frequency and becomes a considerable error factor, especially in measurements of high-q devices at high frequencies. 7

Advanced calibration methodology To achieve highly accurate measurement results using the RF I-V method, the E4991A employs an advanced calibration method that incorporates a low-loss capacitor (LLC) calibration standard. This calibration method is superior to a network analyzer s typical 50-ohm calibration, and it enables the E4991A to measure parameters such as high-q with great accuracy. To do this, Agilent provides a new standard, a low-loss capacitor, to define phase the absolute X axis as shown in Figure 7. The R and X axes are also defined, which allows us to accurately measure high Q. When we calibrate the E4991A using the LLC standard, phase is determined as follows: First, we assume that Z 0 phase is equivalent to 0 degrees. Then we measure the LLC phase in order to get the difference from 0 degrees. Finally, we subtract this difference from Z 0 to define the phase standard. See Figure 7. Note that the LLC calibration method is effective for measurements above 300 MHz. X Z +? Q = tanθ R SHORT OPEN LOAD Low loss (air) capacitor Z = Z 0 Phase = 90 Phase = 0? Figure 7. Instruments that use the RF I-V method of impedance measurement employ an advanced calibration technique to obtain high accuracy. 8

High-Q measurement High-Q (or low-d) measurement is currently a hot topic in RF component testing. Quality factor (Q) is defined as the ratio between reactance (X) and resistance (R) and is commonly used for inductors. The reactance is the product of 2πfL. A low-q inductor (that is, one with high R) absorbs more energy than a high-q inductor. Therefore, high Q is a desirable characteristic in mobile equipment, which must conserve battery energy. Why is it so difficult to measure a high-q or low-d component accurately? To make this measurement, the test instrument must be able to measure a large reactance and a very small resistance at the same time. If, for example, Q = 100, the resistance will be 100 times smaller than the reactance, as depicted in Figure 8. The instrument s inaccuracy, shown as a small circle at the end of measured impedance vector, makes it difficult to measure the small amount of resistance. Even a tiny fluctuation can create a large error in the Q measurement result. Overcoming this difficulty requires a highly accuracy instrument. X Error vector X X Q = R Q=100 R value is 1/100 of X value R R Figure 8. Instrument inaccuracy can add significant error to high-q measurements. If we use a conventional vector network analyzer to measure impedance at Qx=100, the measurement error will be greater than 65% and could go as high as 100%. Even if the value of Q is below 10, the measurement error will be approximately 10%. We can obtain better results using the E4991A s RF I-V method with LLC calibration. As the chart in Figure 9 shows, at Q=100 the E4991A can measure up to 3 GHz with an error of 30% or less. At Qx=50, the measurement error is 15%. Q factor accuracy (%) 120 100 80 60 40 20 Network analysis E4991A RF I-V method 0 1 M 10 M 100 M 1 G 10 G Frequency (Hz) Note: Q accuracy is compared at a Q factor of 100 at 50 Ω impedance. Figure 9. The RF I-V method with LLC calibration provides better Q measurements to 3 GHz. 9

Fixture compensation In addition to performing calibration to remove instrument inaccuracies, we must compensate for the electrical characteristics of the test fixture and cable. To get the real characteristics of the DUT, two sources of error that exist between the calibration plane and the DUT must be removed: residual impedance and electrical length. See Figure 10. DUT Test fixture - Residual impedance error Cable - Electrical length error Calibration plane Figure 10. Two kinds of error residual impedance and electrical length can be introduced between the DUT and the calibration plane. Residual impedance error Small amounts of impedance, inductance, conductance, and capacitance exist in a test fixture, and these can lead to an impedance measurement error. With the E4991A, we can remove the impedance error by using open/short compensation at the test fixture port. Electrical length error Electrical length error is caused by a phase shift of the test signal at a 50-ohm electric line between the calibration plane and the DUT. It can cause a phase measurement error when we measure complex impedance parameters. We can compensate for this electrical length error by using the port extension feature of the E4991A. Using compensation values of Yo and Zs and the measured value of Zm, we can calculate the real impedance of the DUT (Zdut) with the equation in Figure 11. Zs Zm Y0 Zdut Zdut = Zm - Zs 1 - (Z m - Zs)Yo Figure 11. The DUT impedance can be obtained with proper compensation. 10

Measurement Correlation and Repeatability It is possible to obtain different measurement results for the same device when the same instrument and test fixture are used. There are many possible causes of such measurement discrepancies, as well as residuals. Typical factors that contribute to the measurement discrepancies in RF impedance measurements are described as follows. Residual parameters at high frequencies It is important to consider the effects that residual parameters have on measurement results at high frequencies. Assume, for example, that a residual inductance of 0.1 nh and a stray capacitance of 0.1 pf exist around the measurement terminals of the test fixture. The effects of these small residuals differ depending on the frequency, as depicted graphically in Figure 12. 100 k 10 k 1 pf 0.1 pf 0.01 pf Impedance (Ω) 1 k 100 10 1 0.1 1 nh 0.1 nh 0.01 nh 0.01 1 M 10 M 100 M 1 G 10 G Frequency (Hz) Figure 12. Relationships between the residual parameter values and the typical impedance measurement range using the RF I-V method. In the low-frequency range, the residual parameter values are much smaller than the values of normally measured devices. This is because the capacitors and inductors, which are designed for use in low-frequency electronic equipment, possess large values compared to small residuals. Those devices used for higher-frequency circuits and equipment, however, have lower values. In the frequency range above 100 MHz, the majority of DUTs are low-value devices; that is, their values are in the low nh and low pf, approaching the values of the residuals. Consequently, the residual parameters have a greater effect on higherfrequency measurements, and they become a primary factor in measurement error. The accuracy of the measurement results after compensation depends on whether the open/short measurements have been properly performed. To obtain measurement results with good correlation and repeatability, the compensation must be performed each time under identical conditions. Any differences in the compensation method will result in differences in the measured values, leading to problems in correlating the measurement results. 11

Variance in residual parameter values The effective residual impedance and stray capacitance will vary depending on the position at which the DUT is connected to the measurement terminals. Residual inductance increases when the DUT is connected to the tips of the terminals rather than the bottom. Stray capacitance also varies with the DUT s relative position. See Figure 13. DUT L1 C1 To Test port DUT C2 L2 L2 > L1 C2 < C1 Test fixture Calibration plane Test port Figure 13. Difference in residual parameter values due to DUT positioning. A difference in contact condition Changes in the orientation of the DUT to the terminals are another sources of measurement discrepancy. When the DUT makes contact evenly across the measurement terminals, the distance of the current flow between the contact points is minimized, thus providing the lowest impedance measurement value. If the DUT tilts or slants, however, the distance of current flow increases, yielding an additional inductance between the contact points. See Figure 14. Residual resistance also varies in this scenario, producing changes in the measured D, Q, and R values. Such positioning errors affect the measurement of low-value inductors and degrade the repeatability of measured values. Measurement terminal Ground terminal Measurement terminal DUT Straight contact (provides minimum distance of current flow) Ground terminal Slant or tilt contact (yields extra distance of current flow) Figure 14. Altering the contact conditions changes the measured values of inductance and residual resistance. 12

A difference in open/short compensation conditions Improper open/short measurements similarly decrease the accuracy of compensated measurement results. If the open/short measurement conditions vary, again we get inconsistent measurement values. In particular, each short device has an inherent impedance (or inductance) value that, if not properly defined, can introduce error. The effective impedance of the short device will vary depending on how the DUT makes contact with the measurement terminals. As we see in Figure 14, the contact points for a bottom electrode test fixture differ from those for a parallel electrode fixture. Contact points Shorting chip Contact points Bottom electrode fixture Parallel electrode test fixture Figure 15. Different contact points can affect the results of open/short measurements. Furthermore, if the short device is not straight that is, if it curves slightly the measured impedance will vary depending on which side of the device is facing up. These effects are usually small, but they should be taken into consideration, especially when we make very low inductance measurements, typically below 10 nh. 13

Electromagnetic coupling with a conductor near the DUT Electromagnetic coupling between a DUT and a metallic object near the DUT varies with mutual distance and causes variance in measured values. Leakage flux generated around an inductive DUT induces an eddy current in a closely located metallic object. The eddy current suppresses the flux, decreasing the measured inductance and Q factor values. The distance of the metallic object from the DUT is a factor of the eddy current strength, as shown in Figure 16a. Because test fixtures contain metallic objects, they can be important causes of measurement discrepancies. Open-flux-path inductors usually gene rate leakage flux with directivity. As a result, measured values will vary depending on the direction of the DUT. Differences in the eddy current due to leakage flux directivity are illustrated in Figure 16b, c, and d. If a parasitic capacitance exists between the DUT and an external conductor, it is difficult to remove the effect on the measurement because the guarding technique will be invalid. Thus, the DUT should be separated from the conductor with enough distance to minimize measurement errors. Leakage flux Eddy current Metallic object (a) Metallic object near the leakage flux of inductor will cause an eddy current effect. This effect increases as the distance of the metallic object from the inductor decreases. (b) Inductor with leakage flux directivity. Turning the inductor by 90 degrees reduces the eddy current as shown in figure (c). (c) Metallic object in parallel with the leakage flux causes less eddy current. (d) Inductor with less leakage flux directivity. The eddy current effect is almost independent from the direction of inductor's side faces. Figure 16. Eddy current effects and the magnetic flux directivity of a DUT are important test considerations. 14

Defining short-bar residual inductance Surface mount device (SMD) test fixtures for the E4991A are furnished with short bars that can be used for various-sized devices. Optional short bars are available for standard industrial sizes (EIA, IEC, EIAJ). The 16197A comes furnished with only the industrial standard sizes. Because the residual impedances of the short bars differ according to short-bar size, Agilent recommends using the same sized short bar whenever compensation is performed. Two short bars of different sizes were used to make the 2 nh SMD inductor measurements in Figure 15. There is a discrepancy of approximately 200 ph between the measurements that could have been prevented by using short bars of the same size. The residual impedance of the short bar is defined not simply by the existing residuals in the short bar itself. It also depends on how the short bar is placed in the surrounding conditions. Factors that can affect residual impedance include the permittivity, thickness, micro-stripline, and ground conditions of the printed circuit board, as well as other environmental factors. Unless variations in the surrounding conditions are minimized, it is very difficult to simulate the residuals of the short bar. When we measure extremely small inductance values, there are two methods of defining the residual inductance of the short bar. One definition is to let short equal 0 H. With this method, the measurement result is the relative value of the SMD inductor's impedance to the short bar's impedance. The short bar's residual inductance based on its size and shape is not estimated. Traditionally, this has been the primary method of compensating the test fixture's residual inductance. The second definition is to let SHORT equal xh. With this method, the measurement result is the absolute value of the SMD inductor's impedance. The short bar's residual inductance based on its size and shape is estimated and is used as a reference value. Both measurement results are correct, and the difference in results is due only to the difference in the definition methods. However, we may still have a problem in obtaining a good correlation. For example, if we use the same short bar to measure a 10 nh inductor, and the short bar's residual inductance is defined as 0 nh in one instance and 0.4 nh in another, the measurement result will differ by 4%. 15

Changing the short bar s residual parameters has a greater effect on higher-frequency measurements. In Figure 17, different reference residual inductance values are used to measure a 100 pf capacitor. We see that the resonant frequency shifts by more than 200 MHz. If short equals 0H, resulting in the short bar s residual equaling 0.4 nh, the SRF is calculated to be 762 MHz a higher value than the 550 MHz resonant frequency obtained when SHORT equals 0.4 nh. SHORT=0.4nH 100 pf 0.837 nh The SRF shifted by 200 MHz by just changing the short bar s reference residual inductance value. SHORT=0nH 100 pf 0.437nH(0.837 nh-0.4nh) SHORT=0H SHORT=0.4nH Figure 17. Using different reference residual inductance values results in different SRFs. Although the short bar s residual inductance can be defined as short = 0H or short = xh, and either definition will give correct results, we have to use the same method of definition as well as the same measurement conditions (test fixture, compensation method, frequency and test signal) when we correlate measurement results, especially those of inductors with low values. Agilent generally chooses to use short = 0 H, but for the 16196A/B/C family of test fixtures that are used with the E4991A, the short bar s residual inductance values have been provided in the form of theoretical values, so either method of definition can be used and residuals can be calculated without error. 16

Measurement repeatability Table 1 compares the major Agilent RF SMD test fixtures. Short repeatability, which becomes critical when we measure low nh inductors, is smallest with the 16196x and greatest with the 16191A/2A/3A fixtures. Proportional error, which becomes critical when we measure impedance at high frequencies, is again smallest with the 16196x and greatest with the 16191A/2A fixtures. Table 1. To maintain data correlation, we need to use a test fixture that has high repeatability and low proportional error. Key Items 16196x 16197A 16191A/2A Test Frequency Range Dc to 3GHz Dc to 3GHz Dc to 2GHz Short Repeatability [mω] 30+125 x f [GHz] 30+ 150 x f [GHz] 30+250 x f [GHz] Open Repeatability [µs] 5+40 x f [GHz] 2+30 x f [GHz] 2+30 x f [GHz] Proportional Error [%] 1.0 x f 2 [GHz] 1.2 x f 2 [GHz] 1.5 x f 2 [GHz] Table 2 summarizes the importance of calibration and compensation methodologies that are used to measure low-esr/high-q components. Table 2. This table illustrates the main features of each calibration and compensation method. Theory Calibration Cable correction Eliminates instrument system errors Defines the "Calibration Plane using the CAL standard Low Loss C Calibration improves high Q measurement in RF Eliminates the effects of cable error Extends the "Calibration Plane" to the end of the cable Compensation OPEN/SHORT Compensation OPEN/SHORT/LOAD Compensation Eliminates the effects of error sources existing between the "Calibration Plane" and the DUT Eliminates the effects of simple fixture residuals Note: Use the same short bar size and definition value to keep the data correlation Eliminates the effects of complicated fixture residuals Note: Use the same short bar size and definition value to keep the data correlation 17

Summary The RF I-V method of impedance measurement, implemented in the E4991A impedance analyzer, provides accurate and stable results across a wide RF frequency range. An advanced method of calibration using a low loss capacitor enables highly accurate measurement of high Q. Test fixture compensation should be made using one short-bar size and the same definition of short each time. Also, keep in mind that the position of the connection and the orientation of the short bar in relation to the test fixture terminals will influence the impedance measurement results. Ensuring consistency in the contact environment is important in maintaining data correlation. 18

Appendix: Special Calibration Method for High Q Measurement Agilent has developed a special, easy-to-perform method of calibrating the E4991A RF impedance/material analyzer that provides the impedance phase accuracy needed for high-q device measurements. This calibration technique is described below. Outline of the special calibration Although a measurement instrument may be stable, high phase accuracy is still required to measure Q factors accurately. For instance, if we want to measure the Q factor with 10% uncertainty for a DUT whose Q value is almost 100, the uncertainty for phase-scaling must be smaller than 1E-3. The phase accuracy of the instrument is determined almost entirely by the uncertainty of the 50 Ω load standard used in the OSL calibration. One method of improving the accuracy of the phase measurement is to use a phase-calibrated load standard. However, we cannot be sure that the phase uncertainty of the calibrated 50 Ω load will be smaller than 1E-3 at high frequencies (such as 1 GHz). An additional technique is to use, in addition to the normal open-short- load standards, a low loss capacitor as the second load (Load 2), keeping the dissipation factor (D) below 1E-3 at 1 GHz. In this case the uncertainty of the measured phase is decreased from the phase uncertainty of the 50 Ω load (Load 1) to the uncertainty of D of the low loss capacitor (Load 2) for almost all DUT impedances. 19

Details of the low loss capacitor (LCC) calibration We want a calibration method that will reduce the error in phase measurement in spite of the phase error associated with the 50 Ω load. Consider a case in which we have a 50 Ω load standard whose impedance phase is unknown, but whose impedance magnitude is known. We add another load (Load 2), and this time the impedance magnitude is unknown and the impedance phase is known. Load 2 is a low loss capacitor. The number of unknown circuit parameters is still three at most. However, two more unknowns related to standards have been added. There are now eight real unknown parameters: Circuit parameter K1 (two real parameters) Circuit parameter K2 (two real parameters) Circuit parameter K3 (two real parameters) Impedance phase θls1 of the 50 Ω LOAD (one real parameter) Impedance magnitude Zabs_ls2 of the low loss capacitor LOAD2 (one real parameter) We solve this problem analytically. For the simplest case in which both the open and the short standard are ideal, the circuit parameters are found as follows. K1 = A x Z ls x R 0 K2 = Z sm / R 0 K3 = Y sm x R 0 where: R 0 = characteristic impedance A = (1 - Zlm i *Yom)/(Zlm i - Zsm) Yom = measured admittance for OPEN STD Zsm = measured impedance for SHORT STD Zlm i = measured impedance for LOAD STD (i = 1:LOAD1, i = 2:LOAD2) Zls i = true impedance for LOAD STD (i = 1:LOAD1, i = 2:LOAD2) Zls 1 = Zabs_ls1*EXP(j*θls1) Zls 2 = Zabs_ls2*EXP(j*θls2) θls1 = θ2 - θ1 + θls2 Zabs_ls2 = A1/A2*Zabs_ls1 Zabs_ls1 = impedance magnitude for LOAD1 (50 Ω): known θls2: impedance phase for LOAD2 (low loss capacitor): known θ1 = arg((1-zlm 1 *Yom)/(Zlm 1 - Zsm)) θ2 = arg((1-zlm 2 *Yom)/(Zlm 2 - Zsm)) A1 = (1-Zlm 1 *Yom)/(Zlm 1 - Zsm) A2 = (1-Zlm 2 *Yom)/(Zlm 2 - Zsm) 20

In actual cases, these circuit parameters are expressed by far more complicated equations. Therefore, we adopted a simpler procedure consisting of two steps: Step 1 Regard the impedance of the 50 Ω LOAD as Zls1 = 50 + j*0 (that is, the phase of LOAD1 is zero). Find the circuit parameters K1, K2, and K3 by normal OSL calibration using the LOAD value (Zls1). Execute correction for LOAD2 and get the corrected impedance (Zcorr2). Calculate the phase difference ( θ) between the phase of Zcorr2 and the true phase of LOAD2. Step 2 Modify the impedance of LOAD1 to Zls1 whose phase is θ and whose impedance magnitude is still 50 Ω. Calculate the circuit parameters again by normal OSL calibration using modified LOAD impedance Zls1. Although this method gives an approximation, performing just these two steps provides enough accuracy for our purposes. 21

Phase measurement error factors We considered the following error factors during phase measurement in the LLC calibration: 1) Uncertainty for impedance magnitude of LOAD1 2) Impedance phase of LOAD1 3) Impedance magnitude of LOAD2 4) Uncertainty for impedance phase of LOAD2 5) Uncertainty for admittance magnitude of OPEN Notice that Factor 2 and Factor 3 do not cause errors if we use the analytical solution. With computer simulations the phase error is shown to behave as follows: 1) Sensitivity of phase measurement error due to the uncertainty of impedance magnitude for LOAD1 is small. 2) Sensitivity of phase measurement error due to the impedance phase of LOAD1 is small. 3) Sensitivity of phase measurement error due to the impedance magnitude of LOAD2 is small. 4) Uncertainty for the impedance phase of LOAD2 directly affects the phase measurement error. 5) In the case of reactive DUT sensitivity of the phase measurement error due to the uncertainty for admittance magnitude of OPEN ( Yopen ) is reduced to R 0 * Yopen *(Copen/Cls2). In the case of resistive DUTs, the sensitivity is the same as in the normal OSL calibration. where: R 0 = characteristic impedance = 50 Ω, Copen = capacitance of OPEN STD, and Cls2 = capacitance of LOAD2 (low loss capacitor). Based on the discussion above, the phase measurement error when we use the low- loss- capacitor calibration method is determined mainly by Uncertainty for the impedance phase of LOAD2 Uncertainty for the admittance magnitude of OPEN Now we evaluate two other items. The D factor for the capacitor (3 pf) used in the calibration can be small because its dimensions are small and the space between the inner and outer conductors is filled almost completely with air. The D value is verified as 500E-6 at 1 GHz from the residual resistance measurement at the series resonant frequency. The D factor is proportional to the frequency, due to the skin effect. By using zero as the D value for the capacitor during calibration, a phase measurement error of 500E-6 is realized at 1 GHz. On the other hand, the uncertainty for the open capacitance is ±5 ff at most. This gives us a phase-measurement error less than ± 100E-6 at 1 GHz. In all, a phase measurement uncertainty of 500E-6 is realized by using the low loss capacitor calibration. 22

For More Information www.agilent.com/find/impedance Related Literature Literature Number 5952-1530E Agilent LCR Meters, Impedance Analyzers, and Test Fixtures Literature Number 5950-3000 Impedance Measurement Handbook Literature Number 5965-4792E Accessories Selection Guide For Impedance Measurements Literature Number 5980-1233E Agilent E4991A RF Impedance/Material Analyzer Technical Specifications Literature Number 5980-1234E Agilent E4991A RF Impedance/Material Analyzer Literature Number 5988-0728EN Advanced impedance measurement capability of the RF I-V method compared to the network analysis method AN 1369-2 23

Agilent Technologies Test and Measurement Support, Services, and Assistance Agilent Technologies aims to maximize the value you receive, while minimizing your risk and problems. We strive to ensure that you get the test and measurement capabilities you paid for and obtain the support you need. Our extensive support resources and services can help you choose the right Agilent products for your applications and apply them successfully. Every instrument and system we sell has a global warranty. Support is available for at least five years beyond the production life of the product. Two concepts underlie Agilent s overall support policy: Our Promise and Your Advantage. Our Promise Our Promise means your Agilent test and measurement equipment will meet its advertised performance and functionality. When you are choosing new equipment, we will help you with product information, including realistic performance specifications and practical recommendations from experienced test engineers. When you use Agilent equipment, we can verify that it works properly, help with product operation, and provide basic measurement assistance for the use of specified capabilities, at no extra cost upon request. Many self-help tools are available. Your Advantage Your Advantage means that Agilent offers a wide range of additional expert test and measurement services, which you can purchase according to your unique technical and business needs. Solve problems efficiently and gain a competitive edge by contracting with us for calibration, extra-cost upgrades, out-of-warranty repairs, and onsite education and training, as well as design, system integration, project management, and other professional engineering services. Experienced Agilent engineers and technicians worldwide can help you maximize your productivity, optimize the return on investment of your Agilent instruments and systems, and obtain dependable measurement accuracy for the life of those products. Agilent Email Updates www.agilent.com/find/emailupdates Get the latest information on the products and applications you select. Agilent T&M Software and Connectivity Agilent s Test and Measurement software and connectivity products, solutions and developer network allows you to take time out of connecting your instruments to your computer with tools based on PC standards, so you can focus on your tasks, not on your connections. Visit www.agilent.com/find/connectivity for more information. By internet, phone, or fax, get assistance with all your test & measurement needs Phone or Fax United States: (tel) 800 452 4844 Canada: (tel) 877 894 4414 (fax) 905 282 6495 China: (tel) 800 810 0189 (fax) 800 820 2816 Europe: (tel) (31 20) 547 2323 (fax) (31 20) 547 2390 Japan: (tel) (81) 426 56 7832 (fax) (81) 426 56 7840 Korea: (tel) (82 2) 2004 5004 (fax) (82 2) 2004 5115 Latin America: (tel) (305) 269 7500 (fax) (305) 269 7599 Taiwan: (tel) 0800 047 866 (fax) 0800 286 331 Other Asia Pacific Countries: (tel) (65) 6375 8100 (fax) (65) 6836 0252 Email: tm_asia@agilent.com Online Assistance: www.agilent.com/find/assist Product specifications and descriptions in this document subject to change without notice. Agilent Technologies, Inc. 2003, 2004 Printed in USA January 24, 2004 5989-0258EN