MAMMOGRAPHY - HIGH LEVEL TROUBLESHOOTING

Similar documents
THE ART OF THE IMAGE: IDENTIFICATION AND REMEDIATION OF IMAGE ARTIFACTS IN MAMMOGRAPHY

Quality Control for Stereotactic Breast Biopsy. Robert J. Pizzutiello, Jr., F.A.C.M.P. Upstate Medical Physics, Inc

Acceptance Testing of a Digital Breast Tomosynthesis Unit

Practical Aspects of Medical Physics Surveys of Mammography Equipment and Facilities

Mammography is a radiographic procedure specially designed for detecting breast pathology Approximately 1 woman in 8 will develop breast cancer over

STEREOTACTIC BREAST BIOPSY EQUIPMENT SURVEYS

Surveying and QC of Stereotactic Breast Biopsy Units for ACR Accreditation

Quality Control of Full Field Digital Mammography Units

Breast Tomosynthesis. Bob Liu, Ph.D. Department of Radiology Massachusetts General Hospital And Harvard Medical School

8/2/2017. Radiologist Responsibilities. Radiologist Responsibilities. Medical Physicist Mammography Equipment Evaluation and Annual Survey

Introduction. Digital Mammography QA: Comparing the Manufacturers Recommendations. What is QC and why is it important? Review & compare QC tests

Aspire HD. Program Manual. 2nd Edition - October 2012

7/20/2014. Outline. Outline. Disclosures. Learning Objectives. SBB: Practical Aspects of ACR Accreditation, QC and ACR On Site Surveys

Nuclear Associates

Investigation of the line-pair pattern method for evaluating mammographic focal spot performance

Ansur TNT Users Manual. Plug-In

GE Healthcare. Performa. High-performance breast imaging

MILADY. Product Data. Page 1 of 8

NJDEP Medical Physicist s Radiographic QC Survey Registration Number:

Facility, Unit and Test Equipment Data

LECTURE 1 The Radiographic Image

COMPUTED RADIOGRAPHY CHAPTER 4 EFFECTIVE USE OF CR

I. PERFORMANCE OF X-RAY PRODUCTION COMPONENTS FLUOROSCOPIC ACCEPTANCE TESTING: TEST PROCEDURES & PERFORMANCE CRITERIA

Mammography: Physics of Imaging

4/19/2016. Quality Control Activities for the RadiologicTechnologist. Objectives. 3D Tomosynthesis QC differences

Digital Breast Tomosynthesis

RAD 150 RADIOLOGIC EXPOSURE TECHNIQUE II

Overview of Safety Code 35

KODAK MIN-R S Film / 4906

Digital Imaging started in the 1972 with Digital subtraction angiography Clinical digital imaging was employed from the 1980 ~ 37 years ago Amount of

Collimation Assessment Using GAFCHROMIC XR-M2

Compliance Guidance for RADIOGRAPHIC QUALITY CONTROL (5 th Edition)

FFDM in the Field: Physicist's Role in the QC of Mammography Laser Printers May Carl R. Keener, Ph.D., DABMP, DABR

Exposure System Selection

KODAK DIRECTVIEW CR Mammography Feature User s Guide

Safelight Fog does what to contrast and density on film?

Patient-Assisted Compression Impact on Image Quality and Workflow

Image Quality. HTC Grid High Transmission Cellular Grid provides higher contrast images

Exposure Indices and Target Values in Radiography: What Are They and How Can You Use Them?

New spectral benefi ts, proven low dose

RADIOGRAPHIC EXPOSURE

Beam-Restricting Devices

FFDM -FCRm QC Requirements- What You REALLY Need to Know

GE Healthcare. Senographe 2000D Full-field digital mammography system

Image Quality. HTC Grid High Transmission Cellular Grid provides higher contrast images

Multiple Choice Identify the letter of the choice that best completes the statement or answers the question.

REQUIREMENTS FOR LICENCE HOLDERS WITH RESPECT TO QUALITY CONTROL TESTS FOR DIAGNOSTIC X-RAY IMAGING SYSTEMS

X-ray Imaging. PHYS Lecture. Carlos Vinhais. Departamento de Física Instituto Superior de Engenharia do Porto

Breast Imaging Basics: Module 10 Digital Mammography

Digital Mammography Quality Control for the Mammographic Technologist

Imaging Technique Optimization of Tungsten Anode FFDM System

Multiple Choice Identify the letter of the choice that best completes the statement or answers the question.

DIGITAL RADIOGRAPHY ARTIFACTS

History of digital imaging

Breast Imaging Basics: Module 3 Producing Quality Images

Nuclear Associates

Y11-DR Digital Radiography (DR) Image Quality

Published text: Institute of Cancer Research Repository Please direct all s to:

diagnostic examination

DIAGNOSTIC ACCREDITATION PROGRAM. Radiology and CT Quality Control Procedures Workbook

X-RAY IMAGING EE 472 F2017. Prof. Yasser Mostafa Kadah

PRACTICE GUIDELINE FOR DETERMINANTS OF IMAGE QUALITY IN DIGITAL MAMMOGRAPHY

Introduction of Computed Radiography in Two Mammography Services: Image Quality and Dose Analysis

1. Carlton, Richard R., and Arlene M. Adler. Principles of Radiographic Imaging: An Art and a Science, 5th edition (2013).

TECHNICAL DATA. GIOTTO IMAGE SDL/W is pre-arranged for Full Field Digital Biopsy examination with the patient in prone position.

Teaching Digital Radiography and Fluoroscopic Radiation Protection

TESTING FLAT-PANEL IMAGING SYSTEMS: What the Medical Physicist Needs to Know. JAMES A. TOMLINSON, M.S., D.A.B.R. Diagnostic Radiological Physicist

Nuclear Associates

Quality control for digital mammography: Part II recommendations from the ACRIN DMIST trial

Predicted image quality of a CMOS APS X-ray detector across a range of mammographic beam qualities

Test Equipment for Radiology and CT Quality Control Contents

Introduction. Chapter 16 Diagnostic Radiology. Primary radiological image. Primary radiological image

Enhanced Functionality of High-Speed Image Processing Engine SUREengine PRO. Sharpness (spatial resolution) Graininess (noise intensity)

Learning Objectives: What s my motivation? (unknown screen actor) Workshop Overview

CHAPTER 6 QC Test For Fluoroscopic Equipment. Prepared by:- Kamarul Amin bin Abu Bakar School of Medical Imaging KLMUC

Photomultiplier Tube

NEXT Protocol for the Survey of Pediatric Chest Radiography

DISC QC/QA Program for Digital Imaging Systems using the DR Radchex Plus Meter

of sufficient quality and quantity

GE Healthcare. Essential for life. Senographe Essential Full-Field Digital Mammography system

Digital radiography: Practical advantages of Digital Radiography. Practical Advantages in image quality

Maximizing clinical outcomes

Image Display and Perception

Third Edition DECEMBER Collaborating institutes: QARAD, BE LUCK, BE ARCADES, FR LRCB, NL EMIFMA, BE NCCPM, UK. Contributors:

X-RAYS - NO UNAUTHORISED ENTRY

Ludlum Medical Physics

Digital radiography (DR) post processing techniques for pediatric radiology

Features and Weaknesses of Phantoms for CR/DR System Testing

Dosepix Detector as kvp-meter in Radiology and Mammography: First steps

Overview. Professor Roentgen was a Physicist!!! The Physics of Radiation Oncology X-ray Imaging

RaySafe X2. Effortless measurements of X-ray

TOPICS: CT Protocol Optimization over the Range of Patient Age & Size and for Different CT Scanner Types: Recommendations & Misconceptions

NATIONWIDE EVALUATION OF X-RAY TRENDS (NEXT) TABULATION AND GRAPHICAL SUMMARY OF THE 1999 DENTAL RADIOGRAPHY SURVEY

Digital Mammography Quality Control for the Mammographic Physicist

Mammography Solution. AMULET Innovality. The new leader in the AMULET series. Tomosynthesis, 3D mammography and biopsy are all available.

1.1.Clinical Artefacts

X-ray Tube and Generator Basic principles and construction

Assessment of Beam Alignment, Collimation and Half Value Layer of Some Selected X-Ray Machines in Plateau State, Nigeria

Optimization of Digital Mammography Resolution Using Magnification Technique in Computed Radiography 1

SPRINGFIELD TECHNICAL COMMUNITY COLLEGE ACADEMIC AFFAIRS

Transcription:

MAMMOGRAPHY - HIGH LEVEL TROUBLESHOOTING Maynard High New York Medical College SS2001-M.High 1

Objectives: Review MQSA and ACR annual QC tests as opportunities for troubleshooting before a significant image quality problem develops. Review MQSA equipment evaluations, and the role they play in preventing the need for troubleshooting at a later date. Present a logical decision tree approach to troubleshooting specific image quality problems. SS2001-M.High 2

The Medical Physicist CAN & SHOULD be an important resource for improving image quality of: Phantom images & Clinical images SS2001-M.High 3

Medical Physicist s Services: Compliance or Quality? Growing tendency toward compliance easier and quicker to meet minimum standards than to improve on them the MQSA inspector accepts it Both ACR and MQSA encourage the physicist to go beyond compliance checks, and to make recommendations for quality improvement. SS2001-M.High 4

Medical Physicist s Services: Compliance or Quality? During annual testing, make suggestions that may lead to improved image quality. this may prevent more intensive troubleshooting in the future Example: You see the maximum compression force is set to 30#. Although compliant, you should advise increasing the max to 45#. SS2001-M.High 5

Physicist Level Performance Evaluations & Troubleshooting: ANNUAL quality control compliance testing MQSA required testing ACR recommended testing Equipment evaluations New, repaired, relocated equipment MQSA required SS2001-M.High 6

Physicist Level Performance Evaluations & Troubleshooting: Troubleshooting (specific problems) Improving contrast, detail, noise, artifacts, etc Goes beyond performance testing of x-ray units screen-film appropriateness film processing evaluation technique factors proper use of AEC viewing conditions SS2001-M.High 7

Medical Physicist s Services in Mammography: ANNUAL quality control compliance testing Equipment evaluations Troubleshooting SS2001-M.High 8

MQSA Required Annual Physicist QC Survey: The 12 equipment tests in Table 1 testing methods not generally specified see Guidance (www.fda.gov/cdrh/mammography) Evaluation of technologist QC testing A written summary report format of report not specified to include recommendations physicist feels may improve the quality of mammography. SS2001-M.High 9

ACR Recommended Annual Physicist QC Tests: The 11 equipment tests in Table 2 testing methods detailed in 1999 ACR Mammography Quality Control Manual tests, action limits, scope of testing are very similar to MQSA, but there are several differences which Medical Physicist must consider, and use profesional judgement to develop a comprehensive performance evaluation. SS2001-M.High 10

ACR Recommended Annual Physicist QC Tests: Evaluation of technologist QC testing A written report including a summary model report formats are in the ACR manual to include recommendations physicist feels may improve the quality of mammography. SS2001-M.High 11

Resources: SS2001-M.High 12

Medical Physicist s Professional Judgment: Tailoring the Tests Guidance documents can never set the scope of testing and limits for all mammography units because: many different existing designs new designs with new features State Department of Health requirements Professional judgment & knowledge of unit must be used to tailor the tests to the unit SS2001-M.High 13

Multiple Selectable ma Stations: All tests affected by ma should be performed at each clinically used ma station kvp radiation output rate system resolution small focus is always a different ma station SS2001-M.High 14

Multiple Filters (Mo( Mo, Rh,, Al): All tests affected by filtration should be performed for each clinically used filter HVL average glandular dose phantom image quality AEC performance artifact evaluation SS2001-M.High 15

Multiple Targets (Mo( Mo, Rh,, W): All tests affected by anode selection should be performed for each clinically used target HVL average glandular dose phantom image quality AEC performance artifact evaluation SS2001-M.High 16

Multiple Auto-kV, Auto-filter Algorithms: Selectable maximum exposure time (0.8, 1.6s) Selectable dose (STD, CNT, DOSE) HVL average glandular dose phantom image quality AEC performance Test pulses may affect dosemeters SS2001-M.High 17

Multiple AEC Sensors: Some units use separate sensors for small and large Bucky s Must test thickness/kvp tracking for each SS2001-M.High 18

Multiple AEC Sensors: Some units use a matrix of fixed detectors which can be used to find the densest portion of the breast Must test each detector independently in some way SS2001-M.High 19

Multiple Film-Screen Combos: Some facilities use a faster speed screen for mag and/or large Bucky uniformity of screen speed average glandular dose phantom image quality AEC performance system resolution SS2001-M.High 20

Medical Physicist s Professional Judgment: Tailoring the Limits Action limits can often be tightened kvp accuracy & reproducibility AEC performance can be made better than +/- 0.15 OD for 2-8 cm on some units Action limits can be tailored at acceptance testing. SS2001-M.High 21

The Annual Medical Physicist s Survey Report: Requirements Communication of results with recommendations is just as important as performing the tests. Needed on-site for annual MQSA inspections Needed to be submitted with applications for accreditation. SS2001-M.High 22

The Medical Physicist s Report: Useful for - Information resource for technologist Image quality and dose of Rh filter, AEC mode Information resource for service engineer. Details of calibration deficiencies Information resource for troubleshooting Performance baseline to which future measurements can be compared SS2001-M.High 23

The Medical Physicist s Report: Useful only if - There is sufficient information in report. Adequate description of test conditions kvp, ma, filter, target SID, magnification, receptor size phantom type and thickness AEC settings, compression force screen, film, processing conditions Test results for complete range of machine settings and modes of operation clinically used SS2001-M.High 24

The Medical Physicist s Report: Useful only if - Contains detailed recommendations for corrective action. And contains: recommendations that will improve image quality, including [those] concerning image receptors, technique factors, processing, viewing conditions and technologist QC. (ACR QC Manual) SS2001-M.High 25

Medical Physicist s Services in Mammography: ANNUAL quality control compliance testing Equipment evaluations Troubleshooting SS2001-M.High 26

Equipment Evaluations: Requirements Before use on patients, a qualified medical physicist must perform an MQSA equipment evaluation on : newly installed mammography equipment (x-ray unit or processor) disassembled and reassembled equipment equipment which has had a major component changed or repaired. SS2001-M.High 27

Equipment Evaluations: New Equipment Acceptance Test ACR Mammography QC Manual: It is assumed that mammography equipment will have been subjected to more extensive acceptance testing or a thorough performance evaluation prior to initiation of [annual] QC testing. SS2001-M.High 28

Equipment Evaluations: New Equipment Acceptance Test MQSA Compliance Guidance The equipment evaluation is more extensive than the survey. It may be regarded as an acceptance test for equipment and an annual survey alone is not sufficient to meet this requirement. SS2001-M.High 29

Equipment Evaluations: Disassembled, Reassembled Relocated, disassembled and reassembled equipment is to be evaluated as newly installed. A complete acceptance test needed SS2001-M.High 30

Equipment Evaluations: Major Repairs New x-ray tube AEC component replacement Collimator replacement Beam filter replacement Processor reassembly Generator replacement or re-calibration are considered major repairs. SS2001-M.High 31

Equipment Evaluations: Scope for Major Repairs Tests would include all those affected by the component repaired or replaced. SS2001-M.High 32

Equipment Evaluations: Example - New AEC sensor AEC testing Dose determination Phantom tests SS2001-M.High 33

Equipment Evaluations: Rebuilt/Replaced Processor Sensitometric testing Phantom tests Artifact evaluation Dose determination AEC testing Verification of proper processing solutions SS2001-M.High 34

Equipment Repairs: Physicist Involvement Medical physicist evaluates in person Medical physicist provides oversight Medical physicist involvement optional Should be discussion between facility and medical physicist SS2001-M.High 35

The Mammography Quality Standards Act Final Regulations Document #4 May 23, 2001 www.fda.gov/cdrh/mammography SS2001-M.High 36

From CDRH Guidance Document #4 Table: Medical Physicist Involvement in Equipment Adjustments, Changes, or Repairs For any adjustment, change, or repair not listed in the table below, or if the facility is unsure as to the full extent of the adjustment, change, or repair, the facility should consult their medical physicist to determine the proper extent of his or her involvement in evaluating the item. Item Major Repair Medical Physicist Involvement Automatic Exposure Control AEC Replacement Y MP conducts evaluation in person Thickness compensation internal* MP conducts evaluation Y adjustment in person AEC sensor replacement Y MP conducts evaluation in person AEC circuit board replacement Y MP conducts evaluation in person Density control - internal* adjustment N MP oversight Bucky (New to Facility) Replacement AEC also replaced Y MP conducts evaluation in person AEC not replaced N MP oversight SS2001-M.High 37

Medical Physicist s Services in Mammography: ANNUAL quality control compliance testing Equipment evaluations Troubleshooting SS2001-M.High 38

Manufacturer s Troubleshooting Resources: Film processor service company Film manufacturer representative Mammography unit service engineer Mammography unit applications person Should be advised where appropriate SS2001-M.High 39

Manufacturer s Troubleshooting Resources may not resolve issue - Poorly trained service person Narrow focus and expertise of individual Management decisions resulting in poorly compatible system components film/screen/chemistry/processor Physicist can analyze entire imaging system to determine where improvement can be made. SS2001-M.High 40

Troubleshooting by the Medical Physicist Involvement of medical physicist in solving specific problems concerning image quality Will evaluate entire imaging chain Film processing Film-Screen combination X-ray unit Technologist technique SS2001-M.High 41

Triggering Events for Physicist s Troubleshooting: #1 Failure of Technologist s QC test processor sensitometry phantom image quality Physicist must be knowledgeable in performing technologist s tests and in all variables which affect test results. SS2001-M.High 42

Triggering Events for Physicist s Troubleshooting: #2 Patient s current mammograms perceived to be not as good as previous images. Physicist must separate reality from impression establish which image quality parameter degraded establish possible cause of degraded parameter make recommendations for improvement SS2001-M.High 43

Triggering Events for Physicist s Troubleshooting: #3 Facility fails its application for accreditation because of image quality This risk can be reduced if physicist reviews all images before submission. Of course physicist must be familiar with the image quality criteria used by the ACR for both phantom and clinical images SS2001-M.High 44

ACR Image Quality Criteria for Clinical Images: ACR Mammography QC Manual Accreditation Application Package Accreditation Results Report returned to facility Literature (Bassett et al. 2000) SS2001-M.High 45

Clinical Image Failures : 1997* Positioning 20% Exposure 15% Compression 14% Sharpness 13% Contrast 13% Artifacts 11% Labeling 8% Noise 5% SS2001-M.High *Bassett, et al;radiology 2000;215:698-702 46

SS2001-M.High *Bassett, et al; Radiology 2000; 215: 698-702 Inadequate pectoralis on MLO Sagging breast on MLO Poor visualization post. tissue on MLO Skin folds Poor visualization post. tissue on CC Post. Nipple line on CC <1cm of MLO Positioning Problems : 1997* 20 % 22% 14% 14% 12% 12% 10% 47

Posterior nipple line on CC not within 1 cm of that on MLO 6.5 cm 8.0 cm SS2001-M.High 48

Mammography Troubleshooting Categories: Film processing Film-Screen combination X-ray unit Technologist technique SS2001-M.High 49

Troubleshooting the Film Processing: General principles Film must be processed as recommended by film manufacturer (or equivalently) physicist can compare processing at another site using recommended processing. Sensitometry does not guarantee proper processing, only consistency. SS2001-M.High 50

Troubleshooting the Film Processing: General principles Look for changes film type, emulsion number chemistry type, supplier, mixer staffing, hours of operation, films/day PM schedule, PM personnel Film processing must be evaluated before troubleshooting the x-ray unit SS2001-M.High 51

Troubleshooting Resources: SS2001-M.High 52

Troubleshooting Resources: SS2001-M.High 53

Troubleshooting Resources: SS2001-M.High 54

Mammography Troubleshooting Categories: Film processing Film-Screen combination X-ray unit Technologist technique SS2001-M.High 55

Troubleshooting the Film-Screen Combination: High system speed Analysis of 31,000 phantom images submitted to ACR showed: High failure for masses & fibers (30-40%) associated with doses lower than 0.75 mgy (Haus, Yaffe, Feig, et al. 2000) Low doses should trigger a careful evaluation of image noise. SS2001-M.High 56

Troubleshooting the Film-Screen Combination: Low System Speed May result in blur due to long exposure times May result in low contrast if kvp is raised to achieve proper exposure times Dose and risk may be needlessly high. SS2001-M.High 57

Average Glandular Dose: An important troubleshooting tool It should be within an appropriate range SS2001-M.High 58

Mammography Troubleshooting Categories: Film processing Film-Screen combination X-ray unit Technologist technique SS2001-M.High 59

Troubleshooting the Mammographic X-Ray Unit: Physicist may feel this is an easy category Make a few measurements Compare with last report Voila! The offending parameter pops out. Things are seldom so simple SS2001-M.High 60

Example: Troubleshooting AEC because of Low OD on Images Seldom fruitful to have service change AEC calibration without determining reason for low OD Need a troubleshooting algorithm SS2001-M.High 61

Troubleshooting Low OD: A possible algorithm First, get the facts: Under what conditions were images light? All images, parts of images, dense, fatty, thick, thin, high kvp, Rh filter, large Bucky only, etc Was there any correlation with x-ray unit, cassette #, technologist, radiologist, time of day, day of week, PM schedule, etc SS2001-M.High 62

Areas of film with OD < 1.0 are UNDER-EXPOSED EXPOSED OD = 1.6 OD = 0.4 SS2001-M.High 63

Troubleshooting Low OD: A possible algorithm Next, check film and film processing: Nothing can be learned about AEC if processing is not in control Any film, chemistry, processing changes? Review sensitometry records Is processing in control today so AEC testing will be meaningful? SS2001-M.High 64

Troubleshooting Low OD: A possible algorithm Next, check phantom image QC records: Has mas been stable indicating stability of AEC? Review phantom OD plot for stability If there are changes, are they correlated with changes in film sensitometry? If needed, go back to original images. SS2001-M.High 65

Troubleshooting Low OD: A possible algorithm Next, review technologist technique: Check Technique chart Query all technologists about AEC modes used, filters chosen, sensor positioning Compression adequate to spread tissue? Physicist needs working knowledge of the various modes available on machine. SS2001-M.High 66

Troubleshooting Low OD: A possible algorithm Review inputs used by AEC: Thickness (position of compression paddle) Compression force kvp ma, exposure time Attenuation measurements (test pulse) Film type selected Dose or time mode chosen SS2001-M.High 67

Troubleshooting Low OD: A possible algorithm Now expose some phantoms: Simulate patient exposure by using normal compression force with normally used AEC mode. Process films as technologists do emulsion down lengthwise on right side of feed tray SS2001-M.High 68

Film must be fed properly MinR 2000 Emulsion side down MinR 2000 Emulsion side up OD=1.51 OD=1.41 SS2001-M.High 69

Troubleshooting Low OD: A possible algorithm Analyze the data: What is the proper OD? Phantom > 1.4, Patient >1.0 in densest portion But, Depends on type of film viewbox luminance radiologist SS2001-M.High 70

Troubleshooting Low OD: A possible algorithm Finally review all information and make recommendations: There may be more than one cause. Areas of possible improvement unrelated to the light films may be uncovered. Make recommendations concerning these SS2001-M.High 71

LESSON LEARNED #1 Troubleshooting in mammography generally involves several problem sources. They all need to be considered. SS2001-M.High 72

LESSON LEARNED #2 The medical physicist must be knowledgeable about mammography equipment design and how it is used clinically. SS2001-M.High 73

LESSON LEARNED #3 The medical physicist must develop a logical and efficient troubleshooting algorithm. SS2001-M.High 74

Troubleshooting Decision Trees: Identify the particular image quality problem. Make a list of the possible causes or contributing factors. Create a decision tree algorithm to test causes. SS2001-M.High 75

Sensitometry Density and/or Density Difference Change: Film emulsion # (use reserved film) Developer temp Immersion time Replenish rates Contaminated dev. Improperly mixed developer. Change in developer type/brand Oxidized developer Expired/improperly stored film SS2001-M.High 76

Sensitometry Base + Fog Increase Contaminated developer Light leak Expired/improperly stored film Improper safelight SS2001-M.High 77

Phantom OD or Contrast Change Film emulsion # (use clinical film) kvp calibration or setting See all items for sensitometry AEC calibration or setting Target/filter setting Cassette/screen changed Position of phantom & AEC sensor SS2001-M.High 78

Phantom mas Change AEC calibration or setting Cassette/screen changed kvp calibration or setting Position of phantom & AEC sensor Target/filter setting SS2001-M.High 79

Phantom Score Change See all items for sensitometry See all items for Phantom OD or Contrast change See items for low contrast, blur and noise Excessive artifacts SS2001-M.High 80

Low Contrast on Patient Images Most common cause generalized underexposure, or underpenetration of dense portions of breast OD = 0.4 AEC sensor not under densest portion of breast. SS2001-M.High 81

Exposure Problems : 1997* 15% 46% 40% Generalized under-exposure Inadequate penetration of dense areas 9% Generalized over- exposure SS2001-M.High *Basset,et al;radiology 2000;215:698-702 82

Contrast Problems : 1997* 89% 13% 10% Inadequate Contrast Excessive Contrast SS2001-M.High *Bassett,et al;radiology 2000;215:698-702 83

Low Contrast on Patient Images See all items for sensitometry See all items for Phantom OD or Contrast change Film processing does not meet manufacturer s recommendations High fog level SS2001-M.High 84

Low Contrast Case Study #1 Chemistry changed to non film manufacturer by department manager, but not caught by sensitometry or phantom tests because mistakes were made in plotting. Non-optimal optimal film processing. SS2001-M.High 85

Low Contrast on Patient Images kvp calibration or setting (may be affected by AEC mode) AEC calibration or setting leading to low OD (may affect kvp or target/filter) Target/filter setting (may be affected by AEC mode) Inadequate compression SS2001-M.High 86

Compression Force is often operator adjustable SS2001-M.High 87

Deflection Should be < 1 cm 6.5 cm 5.5 cm SS2001-M.High 88

Low Contrast Case Study #2 Compression force miscalibrated,, indicated 20 dn, measured <25 # Resulted in 9 cm compressed thickness (compared with 5 cm previous year) and AOP changed filter to Rh. SS2001-M.High 89

Excessive Blur on Patient Images Marginal focal spot performance Inadequate compression (at least 25#, but as much as possible) Long exposure times (kvp may be too low) time should be between 1-2 sec. Poor film-screen contact SS2001-M.High 90

Compression Problems : 1997* 59% 14% 23% 9% Poor separation of parenchymal tissues Patient motion Non uniform exposure levels SS2001-M.High *Bassett, et al; Radiology 2000; 215: 698-702 91

Excessive Noise on Patient Images Film-screen combination too fast Correlates with low Average Glandular Dose Film processing does not meet manufacturer s recommendations (this is a special concern in case of non-dedicated processor) SS2001-M.High 92

Noise Problems : 1997* 74% 5% 26% Visually Striking Mottle Pattern Limited Visualization of Detail because of Noise SS2001-M.High *Bassett, et al; Radiology 2000; 215: 698-702 93

Excessive Artifacts on Patient Images X-ray unit filter Screen artifacts Film handling Darkroom dust Improperly installed or vented processor Dirty/worn or misaligned rollers Nightly cleaning of cross-over rollers Not following manufacturer s recommendations for processing SS2001-M.High 94

Artifact Problems : 1997* 11% 37% 29% 12% 9% 12% Dirt or Lint Scratches or Pickoff Grid Roller Marks Other SS2001-M.High *Bassett, et al; Radiology 2000; 215: 698-702 95

Mammography Troubleshooting Categories: Film processing Film-Screen combination X-ray unit Technologist technique SS2001-M.High 96

Technologist skill affects image quality: Responsible for positioning & compression Controls OD by AEC sensor positioning Controls contrast by kvp and target/filter selection Controls motion blur (exposure time) by kvp selection SS2001-M.High 97

Compression is one of most important image quality factors Separates structures within breast Reduces thickness of breast more uniform OD less motion shorter exposure time reduced geometric blur SS2001-M.High 98

Physicist can & should evaluate compression Is compression force properly set and calibrated? Physicist should perform test personally and compare with technologist s records Does compression force hold for length of time it takes to complete patient exposure? SS2001-M.High 99

Physicist can & should evaluate compression Is compression mode used properly by technologist? Some units have programmable, multi-step compression modes, that if not understood, can result in incomplete compression. SS2001-M.High 100

Physicist can & should evaluate compression Is the technologist using adequate force? Many units print compression force on the film or on a sticker The physicist should review patient films to see that at least 25# of force is being applied. SS2001-M.High 101

kvp controls both contrast and exposure time RULE of THUMB Maximize contrast by selecting the lowest kvp consistent with an exposure time between 1 and 2 sec to reduce motion blurring. SS2001-M.High 102

kvp controls both contrast and exposure time Is the technologist using a kvp which properly balances contrast and time? Review auto-kvp AEC modes used with technologists and explain their influence on kvp Review exposure times on films Consult technique charts SS2001-M.High 103

Technique charts are valuable Is technique chart conspicuously posted? If not, images may not be consistent among all technologists. Is technique chart current? Chart may not have been changed to reflect changes in screen/film/chemistry SS2001-M.High 104

Technique charts are valuable Do all technologists follow chart? Need to query all technologists. Are recommended settings appropriate? Do AEC phantom tests result in an acceptable exposure time and kvp? SS2001-M.High 105

Mammography Image Quality Troubleshooting Requires good acceptance and periodic QC testing data Requires analysis of entire imaging chain Film processing Film-Screen combination X-ray unit Technologist technique SS2001-M.High 106

Mammography Image Quality Troubleshooting The medical physicist needs to expand his/her knowledge and expertise beyond x- ray unit testing to be a more valuable resource for improving image quality. SS2001-M.High 107

My recommended solution to film processing problems: * * May not be FDA approved SS2001-M.High 108