MIF ZEISS LSM510 CONFOCAL USER PROTOCOL

Similar documents
MIF ZEISS VIOLET CONFOCAL ZEN 2009 PROTOCOL

Operating Instructions for Zeiss LSM 510

LSM 510 Training Notes

LSM 510 Meta Training Notes

Zeiss LSM 510 Confocor III Training Notes. Center for Cell Analysis & Modeling

LSM 710 Confocal Microscope Standard Operation Protocol

Zeiss LSM 780 Protocol

Zeiss LSM 880 Protocol

LSM 780 Confocal Microscope Standard Operation Protocol

ZEISS LSM510META confocal manual

REMEMBER: You have 5GB of disk space on this microscope. Check before you start if you have room for your experiment. If not delete your old data.

LSM 800 Confocal Microscope Standard Operation Protocol

Nikon Eclipse Ti A1-A Confocal Operating Manual. Start-up. Microscope

Quick Guide. LSM 5 MP, LSM 510 and LSM 510 META. Laser Scanning Microscopes. We make it visible. M i c r o s c o p y f r o m C a r l Z e i s s

Quick Guide for Zeiss 710 Laser Scanning Confocal MGH Cancer Center

CONFOCAL MICROSCOPE (Zeiss LSM 510 META v4.2)

Zeiss 780 Training Notes

Contents. Introduction

Microscopy from Carl Zeiss

Zeiss 880 Training Notes Zen 2.3

Quick Guide for Zeiss 710 Laser Scanning Confocal MGH Cancer Center

Things to check before start-up.

TRAINING MANUAL. Olympus FV1000

Practical work no. 3: Confocal Live Cell Microscopy

Microscope Confocal LSM510 META

Leica SP8 TCS Users Manual

Title: Leica SP5 Confocal User Manual

Guide to Confocal 5. Starting session

OPERATING INSTRUCTIONS

Leica SPEII confocal microscope. Short Manual

Internal Medicine Imaging Core Emory University Department of Medicine

Training Guide for Carl Zeiss LSM 5 LIVE Confocal Microscope

DIC Imaging using Laser Scanning Microscopes (LSMs) on Axio Imager Stands

Training Guide for Carl Zeiss LSM 510 META Confocal Microscope

Zeiss Axio Imager.A1 manual

Topics. - How to calibrate the LSM scanner. - How to clean the microscope. - How to adjust the pinhole alignment. - How to adjust the Collimator

Operating Checklist for using the Laser Scanning Confocal Microscope. Leica TCS SP5.

Leica Sp5 II Confocal User Guide

Zeiss AxioImager.Z2 Brightfield Protocol

Leica SP8 TCS Users Manual

Zeiss LSM 510 Multiphoton Confocal Microscope

Bi/BE 227 Winter Assignment #3. Adding the third dimension: 3D Confocal Imaging

Supplemental Method Information Zeiss LSM710

Quick Start Guide. Leica SP5 X

3 Choose the Channels button and set the Channel Settings. Set the Pinhole to 1 Airy unit.

Operation Guide for the Leica SP2 Confocal Microscope Bio-Imaging Facility Hunter College October 2009

DIC Imaging using Laser Scanning Microscopes (LSM) on Inverted Stands

SHORT INSTRUCTIONS FOR OPERATING LSM1/2 (Zeiss LSM510) AT CIAN Version 1.4, September 2014

b. Turn the power switch and key to on position for blue laser.

ZEISS LSM 710 CONFOCAL MICROSCOPE USER MANUAL

Nikon A1Rsi Confocal Start-Up Sequence

Training Guide for Carl Zeiss LSM 7 MP Multiphoton Microscope

Zeiss LSM 510 Multiphoton Confocal Microscope

Contents STARTUP MICROSCOPE CONTROLS CAMERA CONTROLS SOFTWARE CONTROLS EXPOSURE AND CONTRAST MONOCHROME IMAGE HANDLING

Everest System / Slidebook Operating Procedures

Cell Biology and Bioimaging Core

Training Guide for Carl Zeiss LSM 880 with AiryScan FAST

ZEISS LSM 710 NLO Multiphoton microscope Manual/Quick guide

Leica SP8 Resonant Confocal. Quick-Start Guide

Training Guide for Leica SP8 Confocal/Multiphoton Microscope

Confocal imaging on the Leica TCS SP8. 1) Turn the system on. 2) Use TCS user account. 3) Start LAS X software:

Nikon SIM-E & A1-R System

Zeiss LSM880 Operating Instructions. UTMB Optical Microscopy Core Jan. 16, 2018

1 Set up the confocal light path for imaging a green dye (Alexa488-EGFP). For example, the

LEICA TCS SP5 AOBS TANDEM USER MANUAL

ScanArray Overview. Principle of Operation. Instrument Components

Nikon Eclipse Ti2-E Widefield/Spinning Disk Confocal Microscope Standard Operation Protocol

RENISHAW INVIA RAMAN SPECTROMETER

Simplified Instructions: Olympus Widefield Microscope S1230

Using the Nikon TE2000 Inverted Microscope

Swept-Field User Guide

KEYENCE VKX LASER-SCANNING CONFOCAL MICROSCOPE Standard Operating Procedures (updated Oct 2017)

Zeiss LSM 510 META Guide

Renishaw InVia Raman microscope

Olympus Fluoview 1000S Spectral Confocal Microscope Introduction to the NRI-MCDB Microscopy Facility Spectral Confocal Microscope

TRAINING MANUAL. Multiphoton Microscopy LSM 510 META-NLO

Zeiss Deconvolution Microscope: A Quick Guide

Title: Nikon A1R Confocal User Manual

Supplemental Figure 1: Histogram of 63x Objective Lens z axis Calculated Resolutions. Results from the MetroloJ z axis fits for 5 beads from each

Leica DMi8A Quick Guide

Olympus Digital Microscope Camera (DP70) checklist

The Zeiss AiryScan System, Confocal Four.

Overview. About other software. Administrator password. 58. UltraVIEW VoX Getting Started Guide

User Guide to the IBIF Leica TCS SP8 MP Confocal Microscope

Usermanual for Leica SP8 confocal

Nikon A1R. Multi-Photon & Laser Scanning Confocal Microscope. Kyle Marchuk Adam Fries Jordan Briscoe Kaitlin Corbin. April 2017.

Confocal 510 Tutorial

Adobe Photoshop PS2, Part 3

Leica TCS SP8 Quick Start Guide

Using Adobe Photoshop to enhance the image quality. Assistant course web site:

Confocal Raman Microscopy (WITec Alpha 300R)

Brief manual how to start and close the Leica sp2 Confocal. (TCS SP2 AOBS system mounted on a DM IRE2)

Leica TCS SP8 Quick Start Guide

Nikon AZ100. Laser Scanning Macro Confocal Microscope. Jordan Briscoe Adam Fries Kyle Marchuk Kaitlin Corbin. May 2017.

Training Guide for Carl Zeiss AxioZoom V16 Stereo Microscope

(Quantitative Imaging for) Colocalisation Analysis

Leica TCS SL Confocal Training. Neuroscience Imaging Core Staff. Core Director. Facility Manager

BX-61: Brightfield Instruction /Continue to scroll for Fluorescent Instuctions

User Guide for TWAIN / DirectX interface for GRYPHAX USB 3.0 cameras

Use of the HSW5 Spinning Disk Confocal Microscope Updated last May 25, 2010 OK

Transcription:

MIF ZEISS LSM510 CONFOCAL USER PROTOCOL START-UP Turn on the Mercury Bulb Power Supply (if needed). Power-on the Control Box. Turn on the computer. Open the LSM 510 software. Choose Scan New Images and Start Expert Mode. Open the Laser window, and turn on all lasers. Set the argon laser output to 40%. TOGGLING BETWEEN VISUAL MODE AND LSM MODE You control the objective and the reflector wavelength using the touch panel on the right of the microscope. Please note that the 63X objective is a WATER lens. Do not put oil on the 63X objective! In order to change between the Visual and the LSM mode, use the slider rods on the lower right of the microscope. The button on the LSM menu bar will not work on this system. Side Port When the top slider is out, and the bottom slider is in, the microscope is LSM in LSM mode. (Reflector must be on Slider Rods Open ) TV When the top slider is in and the bottom slider is out, the microscope VIS is in Visual mode. In this mode you can look through the eyepieces and change the wavelength of light using the buttons on the touch panel. All of the objectives are parfocal, so you can now use the objective buttons to switch between objectives. Remember to clean your slide if you move from an oil lens back to a dry lens! LSM510 Confocal User Protocol SOP (12-12) Page 1

FLUORESCENT PROBE CONFIGURATION The Configuration Control window allows you to choose the correct filter sets and laser settings to visualize your specific fluorescent probes. Single Track configurations are used when only one fluorescent probe is present. Multi Track configurations are used when multiple fluorescent probes are present. In Multi Track configurations, be aware of Combination vs. Sequential. In Combination scanning, the lasers turn on and scan at the same time, and this may lead to bleed through. In Sequential scanning, each laser turns on, scans, and turns off before the next laser fires. Choose either Single or Multi Track, and click on the Config Button on the right of the Configuration Screen. A drop down box will appear. Scroll through the list to find the correct set of configurations that match your specific fluorescent probes. Click Apply, and close the drop down box. INTENSITY ADJUSTMENT Switch from the Vis setting to the LSM using the Side port sliders. Open up the Scan Window. Run a Fast XY scan to find and focus your sample. You can focus your sample by using the focus knobs on the microscope, or the Stage Control box located on the LSM browser menu. Once the sample is in focus, click on the palette button on the image window. Choose the Range Indicator option. This will show you areas of your image that are either too bright (orange), or too dark (blue). LSM510 Confocal User Protocol SOP (12-12) Page 2

To adjust the intensity levels of your image, click on the Channels tab, and click on the button of the wavelength channel you want to work on. Pinhole should be set to 1 Airy unit. If you click on the 1 button, the pinhole will automatically set to 1 Detector Gain amplifies the input signal by multiplication, so that bright features are brought closer to saturation and general image brightness is increased. The Detector Gain should be decreased until the saturated pixels are no longer orange. Detector gain should be set to between 500-800 optimally. Amplifier Offset sets the gray level of a selected background to zero and adjusts the darkest features in the image to black. To decrease the Amplifier Offset, move the slider to the right. Amplified Gain should be left at 1 SCANNING A QUALITY IMAGE In the Scan Control Window, select the Mode button at the top. There are several settings you should check: Frame size. The larger the frame size, the more pixels per inch. Therefore, as you increase the frame size, you increase the scan time. Scan speed: Normally start at 6 and adjust based upon your sample. Data depth: 8 Bit gives you 256 shades of grey; 12 Bit gives you 4026 shades of grey. Average: will take the average of the number of scans you have selected. You can use the Crop function to zoom in on your region of interest. The Crop button is located on the right menu bar of the image window. When you crop the image, the size of the box outline in Zoom, Rotation, & Offset menu will change to the size/position of your cropped area. LSM510 Confocal User Protocol SOP (12-12) Page 3

When all of your parameters are set, select Single and your image will start to scan. MODIFYING YOUR IMAGE Once you have scanned your image, you can add a scale bar, add arrows, pseudocolor the channels, or adjust brightness/contrast. The Overlay button opens a box that gives you the option to add a scale bar, add arrows, or add shapes. Click on the button of the option you want, and use the mouse to place the object where you want on the image. The Chan button allows you to change the color of each channel in your image. The Contr button allows you to change the brightness and contrast of the image. SAVING YOUR IMAGE All images are saved as.mdb. In the LSM Browser, select the File button. Go to Save As Create new MDB file. Place your folder in the Users Directory, and name it so you can keep track of which images you have taken To convert an image into a TIFF, click on Export. You can save your file in two ways. If you select Raw data single plane, all of the information about that image will be saved. If you made any changes in the contrast or brightness, it will not carry over. You are saving the Raw Image file. This is the most quantitative way to save your data If you select Save Contents of the Image Window, you will get the image that you have modified. When you are finished, clean the oil off of the objectives using the Lens Paper, NOT Kimwipes. LSM510 Confocal User Protocol SOP (12-12) Page 4

SHUT DOWN Open the Laser window and turn all lasers to OFF. Close out of the LSM510 software. You will get a warning box to leave power on for 5 minutes until the lasers cool. Click OK. While you are waiting, shut down the computer and clean the oil objectives (if they were used). Turn off the Remote Control Box. Turn off the Mercury Bulb Power Supply. Cover the microscope. LSM510 Confocal User Protocol SOP (12-12) Page 5