DIGITAL IMAGE PROCESSING IN X-RAY IMAGING

Similar documents
10/26/2015. Study Harder

Amorphous Selenium Direct Radiography for Industrial Imaging

Introduction. Chapter 16 Diagnostic Radiology. Primary radiological image. Primary radiological image

10/3/2012. Study Harder

BASICS OF FLUOROSCOPY

PERFORMANCE CHARACTERIZATION OF AMORPHOUS SILICON DIGITAL DETECTOR ARRAYS FOR GAMMA RADIOGRAPHY

Moving from film to digital: A study of digital x-ray benefits, challenges and best practices

CR Basics and FAQ. Overview. Historical Perspective

Radiology Physics Lectures: Digital Radiography. Digital Radiography. D. J. Hall, Ph.D. x20893

Acquisition, Processing and Display

Setting up digital imaging department!

Unit thickness. Unit area. σ = NΔX = ΔI / I 0

Current technology in digital image production (CR/DR and other modalities) Jaroonroj Wongnil 25 Mar 2016

Digital radiography: Practical advantages of Digital Radiography. Practical Advantages in image quality

SECTION I - CHAPTER 1 DIGITAL RADIOGRAPHY: AN OVERVIEW OF THE TEXT. Exam Content Specifications 8/22/2012 RADT 3463 COMPUTERIZED IMAGING

Examination of Pipe Welds by Image Plate Based Computed Radiography System

X-ray Tube and Generator Basic principles and construction

Seminar 8. Radiology S8 1

Digital Imaging Considerations Computed Radiography

X-RAY FLUOROSCOPY IMAGING SYSTEMS. Dr Slavik Tabakov. Luminescence: Dept. Medical Eng. & Physics King s College London

The Evaluation of Collimator Alignment of Diagnostic X-ray Tube Using Computed Radiography System

Fluoroscopy - Chapter 9

DIGITAL RADIOGRAPHY. Digital radiography is a film-less technology used to record radiographic images.

X-ray light valve (XLV): a novel detectors technology for digital mammography*

RADIOGRAPHIC EXPOSURE

Essentials of Digital Imaging

STATUS AND PROSPECTS OF DIGITAL DETECTOR TECHNOLOGY FOR CR AND DR Ulrich Neitzel Philips Medical Systems, Röntgenstrasse 24, D Hamburg, Germany

SYLLABUS. TITLE: Equipment Operation I. DEPARTMENT: Radiologic Technology

Basis of Computed Radiography & PACS

X-Ray Medical Imaging and Pixel detectors

Features and Weaknesses of Phantoms for CR/DR System Testing

X-ray Imaging. PHYS Lecture. Carlos Vinhais. Departamento de Física Instituto Superior de Engenharia do Porto

10/15/2012 SECTION III - CHAPTER 6 DIGITAL FLUOROSCOPY RADT 3463 COMPUTERIZED IMAGING

SECTION I - CHAPTER 2 DIGITAL IMAGING PROCESSING CONCEPTS

Do you have any other questions? Please call us at (Toll Free) or , or

Mammography: Physics of Imaging

The importance of radiation quality for optimisation in radiology

Veterinary Science Preparatory Training for the Veterinary Assistant. Floron C. Faries, Jr., DVM, MS

X-RAY IMAGING EE 472 F2017. Prof. Yasser Mostafa Kadah

radiography detector

Exposure Indices and Target Values in Radiography: What Are They and How Can You Use Them?

Optimization of Digital Mammography Resolution Using Magnification Technique in Computed Radiography 1

Joint ICTP/IAEA Advanced School on Dosimetry in Diagnostic Radiology and its Clinical Implementation May 2009

Digital Detector Array Image Quality for Various GOS Scintillators

Investigation of the line-pair pattern method for evaluating mammographic focal spot performance

Comparison of computed radiography and filmõscreen combination using a contrast-detail phantom

COMPUTED RADIOGRAPHY CHAPTER 4 EFFECTIVE USE OF CR

ISO INTERNATIONAL STANDARD

While digital techniques have the potential to reduce patient doses, they also have the potential to significantly increase them.

Design and Performance Characteristics of Computed Radiographic Acquisition Technologies

RAD 150 RADIOLOGIC EXPOSURE TECHNIQUE II

Computed Radiography

Radiographic sensitivity improved by optimized high resolution X -ray detector design.

Acceptance Testing of a Digital Breast Tomosynthesis Unit

Digital Image Management: the Basics

I. PERFORMANCE OF X-RAY PRODUCTION COMPONENTS FLUOROSCOPIC ACCEPTANCE TESTING: TEST PROCEDURES & PERFORMANCE CRITERIA

DIGITAL IMAGING Recognise the importance of quality assurance

Dedicated Veterinary Imaging Solutions Digital, CR and Analog Imaging Solutions for any size patient and any size budget.

Mammography is a radiographic procedure specially designed for detecting breast pathology Approximately 1 woman in 8 will develop breast cancer over

X-rays in medical diagnostics

3/31/2011. Objectives. Emory University. Historical Development. Historical Development. Historical Development

Digital Radiography : Flat Panel

Photostimulable phosphor plates (PSPs)

Overview. Professor Roentgen was a Physicist!!! The Physics of Radiation Oncology X-ray Imaging

X-ray detectors in healthcare and their applications

STUDENT REVIEW QUESTION SET K CR/DR CONTENT AREA

Scintillation Counters

DELWORKS DR MEDICAL. take the next step

Chromatic X-Ray imaging with a fine pitch CdTe sensor coupled to a large area photon counting pixel ASIC

NPTEL NPTEL ONLINE COURSE. NPTEL Online Certification Course (NOC) NPTEL. Theory and Practice of Non Destructive Testing

TESTING FLAT-PANEL IMAGING SYSTEMS: What the Medical Physicist Needs to Know. JAMES A. TOMLINSON, M.S., D.A.B.R. Diagnostic Radiological Physicist

Advances in X-Ray Scintillator Technology Roger D. Durst Bruker AXS Inc.

COMPUTED RADIOGRAPHY (CR)

Visualization of sources of scattered radiation from x-ray equipment used for interventional radiology

REAL-TIME X-RAY IMAGE PROCESSING; TECHNIQUES FOR SENSITIVITY

Gas scintillation Glass GEM detector for high-resolution X-ray imaging and CT

Dose Reduction and Image Preservation After the Introduction of a 0.1 mm Cu Filter into the LODOX Statscan unit above 110 kvp

Evaluation of a quality control phantom for digital chest radiography

Beam-Restricting Devices

Radiology. Radiograph: Is the image of an object made with use of X- ray instead of light.

Half value layer and AEC receptor dose compliance survey in Estonia

Small Animal Radiographic Techniques and Positioning COPYRIGHTED MATERIAL

Photons interaction with matter

DR _ solutions. We understand that customers don t need just products, they want. solutions

PRACTICAL CONSIDERATIONS AND EFFECTS OF METALLIC SCREEN FLUORESCENCE AND BACKSCATTER CONTROL IN GAMMA COMPUTED RADIOGRAPHY

Computed Radiography of Resistance Temperature Sensor for Indian PHWR

Studies on reduction of exposure dose using digital scattered X-ray removal processing

Studies on reduction of exposure dose using digital scattered X-ray removal processing

We are IntechOpen, the first native scientific publisher of Open Access books. International authors and editors. Our authors are among the TOP 1%

DALLA LUCE VISIBILE AI RAGGI X: NUOVI RIVELATORI DI IMMAGINI PER RAGGI X A DISCRIMINAZIONE IN ENERGIA ED APPLICAZIONI

COST EFFECTIVE FLAT PANEL DIGITAL RADIOGRAPHY UPGRADE SOLUTIONS

Estimation of signal transfer property for wireless digital detector in different measurement schemes

you can Portable DR with High Sensitivity. Less is more. CXDI-501C DIGITAL RADIOGRAPHY SYSTEM

Hardware for High Energy Applications 30 October 2009

Essentials of Digital Imaging

Image Quality Artifacts in Digital Imaging

Real Time Linear Array Imaging. Brian Caccamise

RADIOGRAPHY TERMS TO KNOW SELF STUDY DENTALELLE TUTORING

TECHNICAL DATA. GIOTTO IMAGE SDL/W is pre-arranged for Full Field Digital Biopsy examination with the patient in prone position.

SmartRAD. Advanced Digital Radiography System

Transcription:

DIGITAL IMAGE PROCESSING IN X-RAY IMAGING Shalini Kumari 1, Bachan Prasad 2,Aliya Nasim 3 Department of Electronics And Communication Engineering R.V.S College of Engineering & Technology, Jamshedpur, Jharkhand. India ABSTRACT: In this paper, an application of digital image processing and analysis technique has been discussed, which can be useful in healthcare domain to predict some major diseases for human being. This contribution discusses a selection of today's techniques and future concepts for digital x-ray imaging in medicine. This application is an image processing system, which works on the basis of a new unified radiography/fluoroscopy solid-state detector concept. Advantages of digital imaging include the possibility to archive and transmit images in digital information systems as well as to digitally process pictures before display, include the possibility to archive and transmit images in digital information systems as well as to digitally process pictures before display, we examine the real time acquisition of dynamic x-ray images (x-ray fluoroscopy). Here, particular attention is paid to the implications of introducing charge-coupled device cameras. We then present a new unified radiography/fluoroscopy solid-state detector concept. As digital image quality is predominantly determined by the relation of signal and noise, aspects of signal transfer, noise, and noise-related quality measures like detective quantum efficiency feature prominently in our discussions. Finally, we describe a digital image processing algorithm for the reduction of noise in images acquired with low x-ray dose. Keywords: Image acquisition, Image enhancement, Image sharpening, Image restoration [1] INTRODUCTION In this paper, we discuss selected current topics of digital image acquisition and processing in medicine, focusing on x-ray projection imaging [1]. If you just look at this name - digital image processing; you find that there are 3 terms. First one is processing, then image, then digital. So, a digital image processing means processing of images which are digital in nature by a digital computer. Image processing is a method to perform some operations on an image, in order to get an enhanced image or to extract some useful information from it. It is a type of signal processing in which input is an image and output may be image or characteristics/features associated with that image. Here, we find that digital image processing techniques is motivated by 2 major applications. The first application is improvement of pictorial information for human perception. So, this means that whatever image you get, we want to enhance the quality of the image so that the image will have a better look and it will be much better when you look at the image [2]. A key feature of digital imaging is the inherent separation of image acquisition and display. Also, digital images can be stored and transmitted Shalini Kumari, Bachan Prasad and Aliya Nasim 1

DIGITAL IMAGE PROCESSING IN X-RAY IMAGING via picture archiving and communication systems (PACS),' and be presented on different output devices, like film printers or cathode ray tube (CRT) monitors (softcopy viewing) as shown in Figure 1 The separation of image acquisition and display in a digital system is illustrated by comparing analog and digital acquisition of single high resolution projection images (x-ray radiography). The principle of the imaging setup is sketched in Figure 2. X radiation passes through the patient before exposing a detector [1]. Widely used for image detection are analog screen/film combinations as shown in Fig.1, which consist of a film sheet sandwiched between thin phosphor intensifying screens. The phosphor screens convert the incoming x radiation into visible light blackening the film, which, after developing, is examined by viewing on a light box. Figure: 1. Principle of conventional x-ray image detection by a screen/film combination. The light-sensitive film is sandwiched between two phosphor intensifying screens which convert the incoming x-radiation into visible light. Well-established digital alternatives include storage phosphor systems (SPS), 2-4 also known as computed radiography (CR) systems, and a selenium detector based digital chest radiography system [(DCS) [5, 6]. Shalini Kumari, Bachan Prasad and Aliya Nasim 2

Figure:2 Principle of x-ray projection radiography (view from above, 1: x-ray tube, 2:x-ray beam, 3: patient, 4:detector) In CR systems, the image receptor is a photostimulable phosphor plate, which absorbs and stores a significant portion of the incoming x-ray energy by trapping electrons and holes in elevated energy states. The stored energy pattern can be read out by scanning the plate with a laser beam. The emitted luminescence is detected by a photomultiplier and subsequently digitized. Common plate sizes are 35X35 cm 2 sampled by a 1760X 1760 matrix, 24x30sampled by a 7576X1976 matrix, and for high resolutions 18x24 cm 2 sampled by a 1770x2370 matrix. The resulting Nyquist frequencies are between 2.5 and 5 lp/mm. An example CR image is given in Figure 3. The detector of a DCS consists of an amorphous selenium layer evaporated onto a cylindrical aluminum drum. Exposure of the drum to x radiation generates an electrostatic charge image, which is read out by electrometer sensors. Maximum size of the sampled image matrix is 2166x2448 pixels, with a Nyquist frequency of 2.7 lp/mm[3]. Figure: 3. A Portion of size 700x1846 pixels from a radiograph of a foot (dorso-plantar) acquired by a CR system. Shalini Kumari, Bachan Prasad and Aliya Nasim 3

DIGITAL IMAGE PROCESSING IN X-RAY IMAGING In analog as well as in digital systems, the acquired radiographs are degraded by non ideal system properties. These include limitations of contrast and resolution, and are described for instance by the modulation transfer function (MTF). Other undesired effects are spatially varying detector sensitivity and unwanted offsets. Additional degradations can be introduced by accidental over- or underexposure. Unlike screen/film systems, however, digital systems enable the compensation of such known degradations by suitable processing like gain and offset correction and MTF restoration [2]. Furthermore, the problem of over- or underexposures is virtually eliminated by the wide latitude of the SPS and DCS image receptors (about four orders of magnitude) and the possibility to digitally adjust the displayed intensity range. Finally, methods like "unsharp masking" and "harmonization" can be employed to enhance with respect to diagnostically less important information, and to optimize image presentation on the selected output device.'-'" Figure 2.1 shows the result of applying such enhancement techniques to the radiograph in Figure 3[4] In the next section we review x-ray image quality measures needed later in the paper. We then consider digital real time dynamic x-ray imaging, known as x-ray fluoroscopy. Here, we pay particular attention to differences in noise behavior of electronic camera tubes and solid-state charge-coupled device (CCD) cameras. This is followed by a discussion of a new flat solid state x-ray sensor for both digital x-ray fluoroscopy and high resolution radiography. Finally, in, we describe a recently developed quantum noise reduction filter [5]. [2] X-RAY IMAGE DETECTION An x-ray tube generates x-radiation by accelerating electrons in an electric field towards a tungsten anode. On hitting the anode, about l% of the electrons generate x-ray quanta, which leave the tube through an x-ray transparent window. The x-ray beam consists of a discrete number of x-ray quanta of varying energy, with the maximum energy being limited by the applied tube voltage. Typical values for the tube voltage range between 60 and 150 kv. The energy distribution of the x-ray quanta determines the beam quality. A thin aluminum plate about 3 mm thick, which absorbs low-energy x-ray quanta unable to pass through the patient, is integrated directly into tube window[1]. These Digital x-ray imaging quanta would only add to the absorbed patient dose without contributing to the imaging process. In the following, the thus reduced range of energies is approximated by a single, average energy, i.e., we assume mono energetic x radiation. For tube voltages of 150 and 60 kev, these average energies are about 63 and 38 kv, respectively [3] Owing to the discrete nature of x radiation only a limited, potentially small number of x-ray quanta contributes to the imaging process at each pixel. For instance, in x-ray fluoroscopy the typical x-ray dose for an image is about 10 ngy at a beam quality of 60 kev. This results in a quantum flow q 0 of roughly q 0=300 quanta/mm 2. X-ray quantum noise is caused by random fluctuations of the quantum flow, which obey a Poisson distribution. Therefore, the standard deviation o of quantum noise is proportional to (q 0) 1/2 [5]. Shalini Kumari, Bachan Prasad and Aliya Nasim 4

Figure: 4 Enhanced version of Figure 3. First, middle and high spatial frequencies were amplified relative to very low ones in order to make such details better visible (harmonization). In a second stage, the image was given a sharper appearance by additional amplification of high spatial frequencies by unsharp masking.. A Portion of size 700x1846 pixels from a radiograph of a foot (dorso-plantar) acquired by a CR system. [3] DIGITAL X-RAY FLUOROSCOPY X-ray fluoroscopy is a real time dynamic x-ray imaging modality which allows a physician to monitor on-line clinical procedures like catheterization or injection of contrast agents. Shalini Kumari, Bachan Prasad and Aliya Nasim 5

DIGITAL IMAGE PROCESSING IN X-RAY IMAGING Figure: 5 E Sketch of fluoroscopy system [1 : movable C arm, 2 : x-ray tube, 3 : x-ray beam, 4 : patient, 5 : operating table, 6 : detection front end,7 : video signal fed to processing unit and monitor (not shown) enhanced version An x-ray fluoroscopy system is sketched in Figure 5, a movable C-shaped arm bearing the x-ray tube and the image detection "front end" is mounted close to the operating table. The position of the C arm can be adjusted arbitrarily during the clinical procedure. The detected dynamic images are displayed on a CRT monitor placed near the operating table, hence providing the physician with immediate visual feedback [4]. [4] FLUOROSCOPY IMAGE DETECTION Today's detection front ends consist of an x-ray image intensifier. (XRII) coupled by a tandem lens to a TV camera, which is followed by an A/D converter. The XRII is a vacuum tube containing an entrance screen attached directly to a photocathode, an electron optics, and a phosphor screen output window. Images are detected by a fluorescent caesium iodide (CsI) layer on the entrance screen, which converts the incoming x-ray quanta into visible photons, which in turn reach the photocathode. The CsI screen is a layer approximately 400µm thick and evaporated onto an aluminum substrate. The absorption of theses creensis about 60%-70%. In addition, CsI is grown in a needle-like structure such that the individual needles act as optical guides for the generated light photons. This prevents undesired lateral propagation within the CsI layer, and thus ensures a relatively good screen MTF[3]. Shalini Kumari, Bachan Prasad and Aliya Nasim 6

[5] CONCLUSION This article described selected topics of medical x-ray image acquisition and processing by digital techniques, some of which are already well established, while others are presently emerging. By first comparing digital radiography systems to analog ones, it was shown that a key advantage of digital imaging lies in the inherent separation of image acquisition and display media, which enables one to digitally restore and enhance acquired images before they are displayed. It turned out that the amount of restoration and enhancement which can be applied is fundamentally limited by noise. ACKNOWLEDGEMENT We would like to thank Prof. Shalini Kumari and Prof. Rakesh Kumar in helping us during the research work for this paper. Under their guidance it was possible for us to complete our research work. REFERENCES [1] D. Meyer-Ebrecht, "The filmless radiology departrnent-a challenge for the introduction of image processing into medical routine work," in 4th IEE International Conference on Image Processing and its Applications, Maastricht, NL (April 1992). [2] Prof. Biswas P.K, "Digital Image Processing," in Indian Institute of Technology, Kharagpur. [3] M. Ishida, "Image Processing," in Computed Radiography, Y. Tateno, T. Iinum4 and M. Takano, Eds., pp. 25-30, Springer-Verlag, Berlin (1987). [4] U. Neitzel, I. Maaclq and S. Günther-Kohfahl, "Image quality of a digital chest radiography system based on a selenium detector," Med. Phys. 2r(4), 509-516 (1994). [5] W. Hillen, S. Rupp, U. Schiebel, and T. Zaengel, "Imaging performance of a selenium-based detector for high-resolution radiography," in Medical Imaging III: Image Formation, Proc. SPIE 1090, 296-305 (1989). [6] G. Spekowius, H. Boerner, W. Eckenbach, P" Quadflieg, and G. J. Laurenssen, "Simulation of the imaging performance of x-ray image intensifier/tv camera chains," in Medical Imaging 1995, Proc. SPIE u3\ t2-23 /1995). Shalini Kumari, Bachan Prasad and Aliya Nasim 7