FIRST Newsletter March 2013, Issue 20. Elemental Distribution Analysis of a Meteorite Sample from the Rochechouart Structure with the µ-xrf M4 TORNADO

Similar documents
Lab Report XRF 441 Elemental distribution analysis on geological samples with the M4 TORNADO

Bruker Nano. M4 tornado. High performance micro-xrf spectrometer. think forward

FAST ELEMENTAL MAPPING WITH MICRO-XRF

M4 TORNADO PLUS. Innovation with Integrity. Super Light Element Micro-XRF Spectrometer. Micro-XRF

M6 JETSTREAM. Innovation with Integrity. Large Area Micro X-ray Fluorescence Spectrometer. Micro-XRF

Processing of MA(or µ)-xrf Data with the M6 software

BRUKER ADVANCED X-RAY SOLUTIONS. SPECTROMETRY SOLUTIONS ARTAX mxrf SPECTROMETER

ARTAX. Innovation with Integrity. Portable Micro-XRF Spectrometer. Micro-XRF

Advancing EDS Analysis in the SEM Quantitative XRF. International Microscopy Congress, September 5 th, Outline

QUANTAX FlatQUAD. Innovation with Integrity. EDS for SEM with the XFlash FlatQUAD EDS

ANALYTICAL MICRO X-RAY FLUORESCENCE SPECTROMETER

S1 TITAN Alloy LE Calibrations (P/N: )

USING A CHARGE-COUPLED DEVICE (CCD) TO GATHER X-RAY FLUORESCENCE (XRF)AND X-RAY DIFFRACTION (XRD) INFORMATION SIMULTANEOUSLY

Quantax 100 Low-Cost EDS System. Innovation with Integrity

The SS6000 Gold Mate Series For analyzing all precious metals and other elements from Mg to U

Film Replacement in Radiographic Weld Inspection The New ISO Standard

MICRO XRF OF LIGHT ELEMENTS USING A POLYCAPILLARY LENS AND AN ULTRA THIN WINDOW SILICON DRIFT DETECTOR INSIDE A VACUUM CHAMBER

WIDE ANGLE GEOMETRY EDXRF SPECTROMETERS WITH SECONDARY TARGET AND DIRECT EXCITATION MODES

QUANTAX. Innovation with Integrity. EDS with Slim-line Technology for SEM and TEM EDS

QUANTAX. think forward. EDS Analysis for SEM and TEM MICROANALYSIS

QUANTAX EBSD with OPTIMUS TKD Transmission Kikuchi Diffraction Under Optimum Conditions

LYNXEYE XE. Innovation with Integrity. High-Resolution Energy-Dispersive Detector for 0D, 1D, and 2D Diffraction XRD

LYNXEYE XE-T. < 380 ev. Innovation with Integrity. Energy. Resolution. High-Resolution Position Sensitive Detector with Superb Energy Resolution XRD

Today s Outline - January 25, C. Segre (IIT) PHYS Spring 2018 January 25, / 26

Material analysis by infrared mapping: A case study using a multilayer

Primary reference standards for precious metal analysis (gold alloys)

High collection efficiency MCPs for photon counting detectors

XRF Applications on Native American Collections

X-Ray Spectroscopy with a CCD Detector. Application Note

IDENTIFICATION OF PAINTING MATERIALS USED FOR MURAL PAINTINGS BY IMAGE ANALYSIS AND XRF

Essential Parameters for the Visibility of IQIs and Small Indications in Digital Radiography

MICROANALYSIS WITH A POLYCAPILLARY IN A VACUUM CHAMBER

Amorphous Selenium Direct Radiography for Industrial Imaging

X-Supreme8000. A powerful innovative XRF analytical solution combining performance with flexibility

Rigaku Innovative Technologies Europe (RITE) Presented by: Dr.Peter Oberta

Application Note #548 AcuityXR Technology Significantly Enhances Lateral Resolution of White-Light Optical Profilers

K-edge subtraction X-ray imaging with a pixellated spectroscopic detector

Structural, optical, and electrical properties of phasecontrolled cesium lead iodide nanowires

New Aspects of Coin Measurement in Quality Inspection

Scanning Electron Microscopy

Digital Radiography for the Inspection of Small Defects

Nature Methods: doi: /nmeth Supplementary Figure 1. Resolution of lysozyme microcrystals collected by continuous rotation.

DEVELOPMENT OF A WAVELENGTH DISPERSIVE X-RAY FLUORESCENCE SPECTROMETER USING A MULTI-CAPILLARY X-RAY LENS FOR X-RAY DETECTION

Fast Laser Raman Microscope RAMAN

The Minting of Platinum Roubles

Report on BLP Spectroscopy Experiments Conducted on October 6, 2017: M. Nansteel

DATA REPOSITORY METHODS. Field counting of K-feldspar megacrysts

Fastest high definition Raman imaging. Fastest Laser Raman Microscope RAMAN

Georgia O'Keeffe. THE Alfred Stieglitz COLLECTION OBJECT RESEARCH Palladium print Alfred Stieglitz Collection. AIC accession number: 1949.

CT parameter studies for porous metal samples. Sören R. Lindemann Daimler AG Werk Untertürkheim

Fast Laser Raman Microscope RAMAN

DESIGN AND MEASUREMENT WITH A NEW PORTABLE X-RAY CAMERA FOR FULL-FIELD FLUORESCENCE IMAGING

The Maia Detector System: A New Way to do Scanning Probe Fluorescence Imaging

FISCHERSCOPE X-RAY Product Line X-Ray Fluorescence Measuring Instruments for the Measurement of Coating Thickness and Material Analysis

746A27 Remote Sensing and GIS. Multi spectral, thermal and hyper spectral sensing and usage

Sensor Fusion Enables Comprehensive Analysis of Laser Processing in Additive Manufacturing

Development of X-ray Tool For Critical- Dimension Metrology

State-of-the-art thin film X-ray optics for synchrotrons and FEL sources. Frank Hertlein Incoatec GmbH Geesthacht, Germany

Computed Radiography

New Detectors for X-Ray Metal Thickness Measuring

GUNSHOT RESIDUE INVESTIGATIONS USING TXRF

SOP: EDAX Eagle III Microspot XRF

SIMULTANEOUS XRD/XRF WITH LOW-POWER X-RAY TUBES

3 Analytical report of glass beads from Hoa Diem site, Khanh Hoa, Viet Nam.

DEVELOPMENT CENTER X-RAY TECHNOLOGY EZRT XEYE. Radiation proof X-ray detectors for industrial applications

METHOD FOR CALIBRATING THE IMAGE FROM A MIXEL CAMERA BASED SOLELY ON THE ACQUIRED HYPERSPECTRAL DATA

Extending Acoustic Microscopy for Comprehensive Failure Analysis Applications

X-RAY OPTICS FOR TWO-DIMENSIONAL DIFFRACTION

Manufacturing Metrology Team

COLOUR INSPECTION, INFRARED AND UV

Multifluorescence The Crosstalk Problem and Its Solution

APPLICATION OF THE DIGITAL RADIOGRAPHY IN WELD INSPECTION OF GAS AND OIL PIPELINES

SUPPLEMENTAL MATERIAL

Product Information Version 1.1. ZEISS Xradia 410 Versa Submicron X-ray Imaging: Bridge the Gap in Lab-based Microscopy

IQI-Sensitivity and Applications of Flat Panel Detectors and X-Ray Image Intensifiers A Comparison

-_.-~ Sample. HIGH SENSITIVITY TYPE TOTAL REFLECTION X-RAY SPECTROMETER SYSTEM Wafer Surface Analysis System --

Zaidi Embong and Husin Wagiran Physics Department, University Of Technology Malaysia, P.O Box 791, 80990, Johor Baharu

Portable XRF Data Under International Reporting Codes. Dennis Arne Principal Consultant Geochemistry

TEP Optimization of the CT scan quality. Related topics Number of projections, stop angle, exposure time, image statistics, binning

Multi-Lateral Shearing Interferometry: Principle and Application on X-ray Laboratory Sources

Examination, TEN1, in courses SK2500/SK2501, Physics of Biomedical Microscopy,

Scientific Analysis of the mural paintings of Aruch, Kobayr and Akhtalà

X-ray Tube and Generator Basic principles and construction

Applications Information

Thermal management and thermal properties of high-brightness diode lasers

Damage-free failure/defect analysis in electronics and semiconductor industries using micro-atr FTIR imaging

Image interpretation and analysis

Copyright -International Centre for Diffraction Data 2010 ISSN

Supplementary Figures

CHAPTER-V SUMMARY AND CONCLUSIONS

Electronic Nose Sensing in Agriculture

Fluorescence X-ray Spectrometer System ZSX Series

Introduction. Chapter 16 Diagnostic Radiology. Primary radiological image. Primary radiological image

High resolution ion mobility-mass spectrometry for separation and identification of isomeric lipids

Science on the Fly. Preview. Autonomous Science for Rover Traverse. David Wettergreen The Robotics Institute Carnegie Mellon University

Mod. 2 p. 1. Prof. Dr. Christoph Kleinn Institut für Waldinventur und Waldwachstum Arbeitsbereich Fernerkundung und Waldinventur

Add CLUE to your SEM. High-efficiency CL signal-collection. Designed for your SEM and application. Maintains original SEM functionality

Dosepix Detector as kvp-meter in Radiology and Mammography: First steps

Applications of Steady-state Multichannel Spectroscopy in the Visible and NIR Spectral Region

Moving from film to digital: A study of digital x-ray benefits, challenges and best practices

Transcription:

FIRST Newsletter March 2013, Issue 20 Elemental Distribution Analysis of a Meteorite Sample from the Rochechouart Structure with the µ-xrf M4 TORNADO By Dr. Roald Tagle, Ulrich Waldschlager, Dr. Michael Haschke, Bruker Nano GmbH, Berlin, Germany. Michael Beauchaine, Bruker AXS Inc., Madison, WI, USA Geological samples are inhomogeneous and the distribution analysis of their different compounds is important for understanding geological processes, exploration of mineral resources, origin, and other applications. Micro X-ray fluorescence (µ-xrf) is well suited for these purposes. It offers great spatial resolution combined with small spot size and high sensitivity for element concentrations. This report describes the elemental examination of a meteorite sample from the Rochechouart structure in France. This meteorite impact structure was dated and revealed an age of 214 +/- 8 million years. The Rochechouart is the first impact structure on earth in which traces of the impacting asteroid were found. Previously they had only been found by Apollo samples from the moon. The traces of the projectile were partially vaporized, melted or highly fragmented during the impact event and mixed with the molten and shocked rocks from the target of the Massif Central. The presence of a projectile in an impact melt rock can be recognized by the enrichment in certain elements more abundant in meteorites than in rock form the earth s crust. These elements are mainly platinum group elements (PGE = Os, Ru, Ir, Pd, Pt, and Rh), but also other siderophile elements such as Ni. Whereas the found PGE are in the low ppb concentration range, the concentration of Ni is in the medium ppm range and therefore detectable with X-ray fluorescence. The analyzed sample shown in Figure 1 was a roughly polished slab. The multi-element imaging of the M4 TORNADO with a detection limit in the low ppm range for many elements allows a better understanding of the components involved in the formation and the hydrothermal evolution of the impact melt. Figure 1: The highlighted sample area was measured with the M4 TORNADO. Instrument The measurements were performed with the Bruker M4 TORNADO. This instrument is characterized by the following features: Large vacuum chamber for samples up to 200 x 300 x 150 mm³ Fast X-Y-Z stage with TurboSpeed for fast mapping and distribution analysis on-the-fly

Effective excitation of fluorescence through high brilliance X-ray tube together with X-ray optics for concentration of the tube radiation to spot sizes down to 25 µm Detection of fluorescence radiation with silicon drift detectors (SDD) with high count rate capability Quantification with standardless models using full pattern fitting The measurement was performed under the following conditions: Excitation: 50 kv, 600 µa Analyzed area: approx. 48 x 25 mm Number of pixels: 600 x 313 Acquisition time per pixel: 5 ms and 200 ms, respectively Total measurement time: 20 min and 10.5 h, respectively Total Measurement Time and HyperMap The total time for the measurement is not only determined by the acquisition time per pixel but also by the time required for stage movement. The M4 TORNADO performs the measurement on-thefly, which means that the stage moves constantly and does not stop at every pixel. This saves a large amount of time and makes the total measurement time almost identical to the sum of the pixel times. The M4 TORNADO uses Bruker s HyperMap technology in which a complete spectrum is saved for every pixel of the map. This offers a lot of advantages, in particular for the data evaluation. Distribution of Single Elements The availability of data for every pixel permits the calculation of the distribution of single elements. This provides additional information on the Figure 2: Results of a distribution analysis with an acquisition time per pixel of 5 ms (top) and 200 ms (bottom). composition of the sample. The distribution of components of the mineral is shown in Figure 2. The resolution of the element distribution images is influenced by the measurement time. This can be concluded by comparing the distribution images of the same element at two different measurement times. Figure 3 shows the element distribution of Ni as a trace element, which is interesting for determining the base of the impact crater. The differences in spatial resolution at both measurement times can be clearly observed. The image resulting from the analysis with the longer measurement time shows clear element distribution with sharp structures. As for the analysis at the short measurement time, the Ni image is more blurred and unclear. Element maps for each element present are displayed in Figure 4.

Figure 3: Distribution of Ni for 5 ms (left) and 200 ms (right) acquisition time per pixel. Figure 4: Single element distribution images of a series of elements calculated from the HyperMap at 200 ms acquisition time per pixel.

Sum spectra of complete mapping HyperMap also offers the possibility to accumulate a spectrum for every pixel and displays the sum spectrum of the complete mapping. Figure 5 shows the sum spectra of both mapping in a log scale. The intensity of one of the spectra is significantly higher (factor 40), due to the longer measurement time. However, other differences between the spectra are negligible, since the measurement time was long enough in both cases. Single point and accumulated spectra The acquisition of a spectrum shows the result of a single pixel at 5 ms measurement time. This allows for the identification of main components, but no further evaluation. In Figure 6 the green spectrum is the result of measuring the same pixel for 200 ms. It has better statistics and allows for the identification of minor elements. Finally, the blue spectrum results from the measurement of a large area with similar composition. It shows good statistics and traces even at a measurement time of only 5 ms and allows for quantification. Trace analysis HyperMap also offers the function Maximum Pixel Spectrum, which generates a spectrum out of the highest channel content for every pixel of the complete map. Figure 7 compares the spectrum with the sum spectrum of the entire map. The Maximum Pixel Spectrum provides more information on trace elements, which can be used for the calculation of their distribution. This is valid for Ni, Cu, and Zr. Figure 5: Sum spectrum of both mappings (log scale).

Figure 6: Selected area of the mineral for accumulation of the point and area spectrum. The image on the right shows the spectra of a single point for 5 ms (red) and 200 ms (green) measurement time (absolute counts). The blue one is the accumulated spectrum of the highlighted sample area for 5 ms measurement time (counts per second). Figure 7: Comparison of the sum spectrum (red) with the Maximum Pixel Spectrum (blue).

Conclusion The analysis of a geological sample with the M4 TORNADO offers a wide variety of information on the structure and composition of the sample. This includes the complete distribution of several elements and the identification and analysis of single grains. The high count rate capability and the excellent detector performance of the M4 TORNADO permit reliable data collection within a very short measurement time. The TurboSpeed stage allows fast mapping and measurement onthe-fly. Even acquisition times in the low millisecond range can deliver first impressions of the structure of the sample. Longer measurement times ensure better spatial resolution due to better statistics and the enhanced contrast between pixels. Longer measurement times also offer the possibility to examine traces. 2013 Bruker Corporation. All rights reserved. Bruker AXS Inc. Madison, WI, USA +1 800 234 9729 info@bruker-axs.com www.bruker.com