NMR spectrometer usage at the BioNMR facility ETH Zürich

Similar documents
Instruction for Operating the Bruker Avance III 800 MHz NMR Spectrometers in UTMB

Step by step procedure for NMR data acquisition

User manual Bruker DPX200 NMR spectrometer

400 MHz spectrometer user manual

NMR Basics. Lecture 2

Student Name: Date Completed: Supervisor:

Your first NMR measurement

HMBC 17. Goto. Introduction AVANCE User s Guide Bruker 185

NMR Spectrometer Operation: xwinnmr

H Micro-Imaging. Tuning and Matching. i. Open any 1H data set and type wobb.

NMR Hardware 06/06/2017. Outline. Instrumentation: Magnet. Increasing magnetic field increases Sensitivity, by power of 3/2 Dispersion, linearly

Chapter 1. 1 The NMR Spectrometer. 1.1 Components of an NMR Spectrometer The Magnet

1D Transient NOE on the Bruker DRX-500 and DRX-600

6.S02 MRI Lab Acquire MR signals. 2.1 Free Induction decay (FID)

Recording EPR Spectra using ER 4102ST Resonator

(N)MR Imaging. Lab Course Script. FMP PhD Autumn School. Location: C81, MRI Lab B0.03 (basement) Instructor: Leif Schröder. Date: November 3rd, 2010

Instructions for 1 H-, 13 C-, 19 F-, and 31 P-Spectra on the Varian Mercury-Vx-300

If the magnetic field is larger, more energy is required to excite a given nucleus.

TopSpin Guide Book. Basic NMR Experiments User Manual. Innovation with Integrity. Version 002 NMR

7 Equipments. Spectrometers

2015 Spin echoes and projection imaging

Open acqi window if the button has been lost. autolocking routine, alock= y for autolocking, alock= n for typical manual locking

Implementing ultrafast 2D NMR experiments on a Bruker Avance Spectrometer

The Agilent OneNMR Probe

Lab 8 6.S02 Spring 2013 MRI Projection Imaging

An NMR Caveman s Guide to Quickly Acquiring Spectroscopic Data By Brian Sparling

Recording EPR spectra using the Loop Gap Resonator (LGR)

Gradients. Effects of B0 gradients on transverse magnetisation Similar to figure 10 of Sattler review Progr. NMR 34 (1999), 93

MRI SYSTEM COMPONENTS Module One

PINMRF. Varian 300 MHz NMR Spectrometers User Guide for Advanced 1D and Basic 2D NMR Experiments

Guide to Varian Spectrometers running VnmrJ

Two Dimensional Homonuclear Correlation Spectroscopy

EKA Laboratory Muon Lifetime Experiment Instructions. October 2006

KJM Version 1.0. Topspin 3.5 Windows 7 Topspin 1.3 Windows XP

EPR spectrometer & components

Eclipse+ NMR Training Guide

ScanArray Overview. Principle of Operation. Instrument Components

Principios Básicos de RMN en sólidos destinado a usuarios. Gustavo Monti. Fa.M.A.F. Universidad Nacional de Córdoba Argentina

In a typical biological sample the concentration of the solute is 1 mm or less. In many situations,

SUPPORTING INFORMATION

Instruction manual for T3DS software. Tool for THz Time-Domain Spectroscopy. Release 4.0

Two Dimensional Homonuclear Correlation Spectroscopy

Instructions for the Experiment

NMR Spectrometer Hardware

RENISHAW INVIA RAMAN SPECTROMETER

Magnettech GmbH Louis-Bleriot-Straße 5 D Berlin Germany

Module 1 Experimental Procedures. Experiment 1. What s in the Spectrometer?

CRL MASS SPECTROMETRY FACILITY INSTRUMENT USER MANUAL MALDI MICRO. Operating Instructions. Basement Spec Lab:

Program. short intro the hardware locking and shimming 1D proton setup presaturation spectra

Pulsed-Field Gradient NMR Diffusion Measurement Protocol

8 COSY. 8.1 Introduction. 8.2 Magnitude COSY

Fundamentals of NMR MRI Workshop Dayananda Sagar Institutes 29 May 2014 K.V.RAMANATHAN NMR Research Centre Indian Institute of Science Bangalore

Laboratory Experiments for Nuclear Magnetic Resonance Spectroscopy May 6, 2004

10. Phase Cycling and Pulsed Field Gradients Introduction to Phase Cycling - Quadrature images

Fast Methods for Small Molecules

SpinCore RadioProcessor LabVIEW Extensions

Two Dimensional Heteronuclear Correlation Spectroscopy

Ph 3455 The Franck-Hertz Experiment

PANalytical X pert Pro Gazing Incidence X-ray Reflectivity User Manual (Version: )

Gentec-EO USA. T-RAD-USB Users Manual. T-Rad-USB Operating Instructions /15/2010 Page 1 of 24

The University of Texas at Austin Electron Paramagnetic Resonance Facility. Bruker EMX+ CW EPR Spectrometer

Signal Generators for Anritsu RF and Microwave Handheld Instruments

Ultrahigh-resolution Total Correlation NMR Spectroscopy

NMR FACILITY NEWSLETTER

Exercise at the Spectrometer

Working With Drawing Views-I

Introduction to Oscilloscopes Instructor s Guide

Alibre Design Tutorial: Loft, Extrude, & Revolve Cut Loft-Tube-1

SigCal32 User s Guide Version 3.0

We recommend downloading the latest core installer for our software from our website. This can be found at:

1. Preliminary sample preparation

Introduction to Simulink Assignment Companion Document

PULSED/CW NUCLEAR MAGNETIC RESONANCE

The oscilloscope and RC filters

Nuclear Magnetic Resonance Spectrometer (600MHz)

SC5306B 1 MHz to 3.9 GHz RF Downconverter Core Module. Datasheet SignalCore, Inc.

Evaluation Chapter by CADArtifex

EENG-201 Experiment # 4: Function Generator, Oscilloscope

KJM D-SELECTIVE NMR Experiments on the AVI-600 and AVII-600. Version 1.0. Topspin 3.5 Windows 7 Topspin 1.3 Windows XP

JSM 6060 LV SCANNING ELECTRON MICROSCOPE STANDARD OPERATING PROCEDURES

Horiba LabRAM ARAMIS Raman Spectrometer Revision /28/2016 Page 1 of 11. Horiba Jobin-Yvon LabRAM Aramis - Raman Spectrometer

WITec Alpha 300R Quick Operation Summary October 2018

Horiba Jobin-Yvon LabRam Raman Confocal Microscope (GERB 120)

PHY3902 PHY3904. Nuclear magnetic resonance Laboratory Protocol

PANalytical X pert Pro High Resolution Specular and Rocking Curve Scans User Manual (Version: )

IX Feb Operation Guide. Sequence Creation and Control Software SD011-PCR-LE. Wavy for PCR-LE. Ver. 5.5x

SPECIFICATION FOR A 7.0 TESLA/400MM ROOM TEMPERATURE BORE MAGNET SYSTEM

Introduction to Autodesk Inventor for F1 in Schools (Australian Version)

Mach 5 100,000 PPS Energy Meter Operating Instructions

Eddy current flaw detector «Eddycon C»

OPTICAL BACKSCATTER REFLECTOMETER TM (Model OBR 5T-50)

PRORADAR X1PRO USER MANUAL

ENSC327 Communication Systems Fall 2011 Assignment #1 Due Wednesday, Sept. 28, 4:00 pm

5008 Dual Synthesizer Configuration Manager User s Guide (admin Version) Version valontechnology.com

PRELIMINARY SPECIFICATIONS MRBR 7.0 TESLA / 210MM ACTIVELY SHIELDED CRYO-COOLED MAGNET SYSTEM

Brightness and Contrast Control Reference Guide

Experiment 2: Electronic Enhancement of S/N and Boxcar Filtering

MFJ-208 VHF SWR Analyzer

Operating Rausch ScanCam within POSM.

SPECIFICATIONS FOR A 4.7 TESLA/310MM BORE ACTIVELY SHIELDED MAGNET SYSTEM

Transcription:

NMR spectrometer usage at the BioNMR facility ETH Zürich Accounts 2 Safety precautions: strong magnetic fields 2 Parts of an NMR spectrometer 3 NMR data storage 3 Start topspin software 3 Initial steps for performing NMR measurements 4 The sample 4 Insert sample 4 Set Temperature 6 Locking 6 Tuning and matching 6 Shimming 7 1H pulse calibration 7 1H, 13C and 15N pulse calibration using HCN90 8 Temperature calibration 8 Basic experiments 9 Interactive optimization of acquisition parameters 10 Optimize receiver gain (rg) 10 Optimize water suppression (presaturation) 11 Calibration of water flip-back pulses 12 Chemical shift referencing of spectra 12 1 bionmr platform ETH Zürich, December 2018

Accounts Logins on the NMR workstations are independent of ETH accounts on our institute & group computers. You will need to obtain an account from the NMR administrator before working independently at the spectrometer. There are also student accounts available for training, block course & semester projects. You may work on your supervisors account with his/her consent once you have received basic training. Safety precautions: strong magnetic fields Keep magnetic materials (like iron tools or gas bottles) away from the magnets. Magnetic items may fly towards the magnet and hurt someone, or lead to a quench of the cryogenic magnet. Do not bring credit cards or electronic devices like cell phones close to the magnet. The memory may be erased by the strong magnetic fields. In case of pregnancy stay outside of 0.5 mt line. Do not bump into the magnets or other NMR equipment. The 900 (lab HPK B7/9) and 750 MHz (lab HPM A54) spectrometers have a large stray field, almost up to the spectrometer computer consoles. For the 500, 600, 600b and 700 MHz the 0.5 mt (=5 Gauss) line is the boundary of the red circles marked on the floor of the NMR labs. These spectrometers are located in lab HPK B1 (see floor plan). 2 bionmr platform ETH Zürich, December 2018

Parts of an NMR spectrometer 1. Superconducting magnet and cryostat The coil of the superconducting magnet is cooled with liquid helium (4.2 K). 2. Console The console contains the electronics (amplifier, pulse generator, shim control, digitizer, etc.) 3. Preamplifier The weak signal from the probehead is amplified before it is transmitted to the console. 4. Probehead The probehead holds the sample and contains the sending and receiving coils for the individual nuclei (2H, 1H, 13C, 15N, 31P, 19F). 5. Spinner The NMR tube is inserted into the spinner and the spinner with the tube into the magnet. 6. Cryoplatform Cooling of the cryo probehead is provided by the cryoplatform. 7. BCU The sample temperature is controller with the BCU. 8. Workstation The spectrometer is controlled by the workstation computer with the program topspin. NMR data storage Data are stored locally on the spectrometer computers in /opt/topspin/data/username/nmr. The raw nmr data (ser files) are deleted automatically from the spectrometers computers after two weeks. Copy the data to a safe place immediately after completing measurements. If any file is accidentally deleted or changed you can retrieve the old versions from a rolling backup (up to 2 weeks). Data are also backed up at the central ETH backup for six months. Details on these backups and how to retrieve data are on our NMR Database pages. Start topspin software 1. Log in to the computer which is controlling the spectrometer Each workstation screen is labeled with the spectrometer name. Be careful not to confuse the six and the six2! 2. Start topspin by clicking on the topspin icon. Topspin is the software to control the spectrometer, process and analyze the data. The current topspin version is topspin 3 bionmr platform ETH Zürich, December 2018

3.2.. Initial steps for performing NMR measurements The sample There must be a sample in the spectrometer at all times. If there is no lock signal when you start or no sample appears when you eject, contact an NMR manager. This could be a problem with a sample being stuck in the bore. Putting in a second sample can cause a major problem and may even require warming up the cryoprobe or loss of your sample. Samples must be in standard or shigemi NMR tubes (3 mm or 5 mm) and correctly positioned in the spinner to prevent damage to the probehead and sample. Samples must be cleaned before inserting into the spinner using a chemwipe. No sticky or loose parafilm may be present. If the sample is loosely held by the spinner try a different one. Do not use the heavy white ceramic spinners. At each spectrometer workstation desk there is a standard sample in a fused 5 mm standard NMR tube with a blue D2O/H2O label containing 10% D 2 O in H 2 O. The sample should be put in the spectrometer & locked after each user completes his/her session. 4 bionmr platform ETH Zürich, December 2018

Insert sample Magnets without SampleCase (nine, seven) and SampleJet in manual mode (sechs) 1. Eject sample: ej (H 2 O/D 2 O sample should be ejected if you just started your session) 2. Remove the previous sample from the magnet. 3. Insert your sample tube into a spinner, and adjust the insertion depth by centering the liquid around the active volume of the probe using the plexiglas device. The sample insertion depth should maximally be 21 mm from the center. 3. Insert sample on top of magnet, and type ij Magnets with SampleCase (six, five, sept) 1. Put your sample with spinner into a free holder in the SampleCase carousel. 2. Type sx holder_nr to insert the sample from holder_nr into the magnet. 3. Holder 23 and 24 contain the standard samples for temperature calibration and sucrose to be inserted in the magnet in case there is no sample from the user. 4. To eject the sample type ej. SampleJet (sechs) Please ask Simon or Alvar if you like to use the SampleJet. The SampleJet robot needs special caps for the tubes. Create new experiment Create a data directory for your experiments. This is done by copying a directory from one of your earlier experiments or from another user or by creating a brand new experiment by entering edc and loading a parameter set using rpar. To copy an experiment directory you first have to load that data set into topspin: navigate to the directory on the left panel. Right click on the directory and select display The experiments you want to copy is now the currently active dataset in topspin. In order to copy the currently active dataset directory type edc. The following window will appear: 5 bionmr platform ETH Zürich, December 2018

To create a copy of the current dataset in your folder you need to change the username (in the example above accept ) to your username. In case the DIR window is not visible, click on Options. You may also want to change the name of the directory so that you can easily identify the type of sample, experiment and date later. A good naming convention is for example DATE_SAMPLE_EXPTYPE, for instance 20180103_Ubiquitin_bbassignment where DATE is in YYYYMMDD format. Set Temperature Type edte or double-click on the temperature display icon at the bottom of the topspin window. Allowed temperature range: 0 C to 55 C. Allow the temperature to equilibrate during 3 5 minutes (3 min for 3 mm tubes, 5 min for 5 mm tubes) The temperature should be calibrated before each experimental session. See section Temperature calibration. 6 bionmr platform ETH Zürich, December 2018

Locking Lock the sample by typing lock and selecting the lock solvent. The purpose of the lock is to keep the magnetic field constant by monitoring a reference signal, which is the deuterium signal of the lock substance, and adjusting the magnetic field to compensate for external (like moving metal objects) or internal (drift) perturbations. Type lock to open a window to select the lock solvent. Select the solvent for your sample (usually 10 % D 2 O / 90 % H 2 O for biological samples). The spectrometer now reads the parameters for the corresponding lock solvent, lock power, gain, frequency and searches for the lock signal within a given range. Wait until lock finished appears. Tuning and matching Start tuning and matching by typing atmm. A new window will open and show the wobble-curve of the channel with the lowest frequency. E. g. if you loaded the parameter set zgpr.eth it has three channels defined channel 1: H1 channel 2: C13 channel 3: N15 The tuning and matching curve of N15 is displayed first. The Larmor frequency of N15 is shown by a vertical red line. You can adjust TUNE and MATCH using the < and > buttons for small steps, << and >> for medium steps, and <<< and >>> for large steps. When done Save, Clicking switch nucleus will change to C13 which has the next higher frequency. Repeat procedure. Click switch nucleus again & finally adjust H1 and save. You can also do this entire procedure automatically by entering the command atma. It says atma finished once atma is completed. Shimming Shim the sample by typing topshim gui. Start shimming by clicking on Start. The message topsim complete indicates that shimming is finished. The quality of the shim can be assessed from the report tab. The parameters envelope width and B 0 inhomogeneity are supposed to be below 0.2 and 0.5 Hz, respectively. Shimming is required to create a homogeneous magnetic field. An in-homogeneous magnetic field result in broad and/or asymmetric lineshapes. 1 H pulse calibration Calibrate the pulse length for the proton 90 deg pulse using the water resonance at 4.7 ppm. 1. Change to your data directory. 7 bionmr platform ETH Zürich, December 2018

2. rpar zg.eth : load the data set called zg.eth. 3. pl1db -13 (-13 db, the exact value depends on the spectrometer) : Set the power level for the high power proton pulse. 4. p1 1 us : Set the pulse width to a small value (1 μs) and acquire a spectrum to set the phases of the water signal correctly. 5. ns 1 : number of scans 6. ds 0 : dummy scans 7. rg 1 : receiver gain set low to avoid receiver overflow because of the intense water signal 8. zg : zero go 9. ft : fourier transform 10.apk : automated phase correction. Observe the intense water signal at around 4.7 ppm and make sure it is a symmetric gaussian curve. 11.Set the carrier at the center of the water signal at about 4.7 ppm using the cursor by clicking on the icon. Move the cursor to the center of the peak and click the left mouse button, select O1. The o1 values is now stored for the current parameter set. For the actual calibration, the 360 deg pulse duration is determined and divided by four to give the 90 deg pulse. Determining the 360 deg pulse has two advantages: once the p1 is close to 360 the water magnetization is returned to the positive z axis and therefore only a short equilibration delay is needed. Furthermore, it is visually more straightforward to identify a zero transition, like in the figure above, than a maximum in a sine curve. Start with p1 = 4 * p90. p90 is an estimate to the 90 deg pulse width. The value from the logbook is a good starting point. Acquire spectrum like described in 8, 9, 10. If the p1 value is too high, this will result into a positive water signal and vice versa. For a perfect 360 deg pulse, the spectrum will look almost like a flat line with some wiggles at the position of the water signal. Adjust p1 accordingly and repeat steps 8, 9, 10. 8 bionmr platform ETH Zürich, December 2018

1 H, 13 C and 15 N pulse calibration using HCN90 Type HCN90. HCN90 will automatically determine the 90 deg pulses for 1 H, 13 C and 15 N. It takes a few minutes for HCN90 to finish. Note that your sample needs to contain reasonable concentrations (> 50 μm) of 13 C and/or 15 N labeled molecules (protein or RNA) to work. Unless you are using the 900 MHz, you can use the automatically determined 1 H pulse from HCN90. HCN90 does not give good values for 1 H on the 900 MHz. Write down the pulses and power levels in the logbook. Type xxx90 to read calibrated parameters into current data-set. Type geteth to calculate and set decoupling power, shaped pulse power, and recommended maximum TD values for the current spectral widths. If you change spectral widths, you must repeat geteth to obtain maximum TD values. Type DOTALL in the command line to open documentation on the DOTALL parameter sets, which includes and more detailed explanation of geteth. Temperature calibration The sample temperature is regulated with a stream of cooled or heated air which constantly passes the sample in the probehead. The temperature which is displayed in topspin show the temperature of the airstream which can deviate from the actual sample temperature. The actual temperature within the NMR tube is calculated from the temperature dependent chemical shifts of the methanol signals. 1. Insert fused 5 mm NMR tube with yellow label MeOD 99.8% 2H sample. 2. Lock on deuterated methanol by typing lock and selecting the solvent MeOD (methanol-d4). 3. Observe the lock signal. Open the large lock-signal window by double-clicking on the lock icon at the bottom panel of the topspin window or by typing lockdisp in the topspin command line. 4. Open the BSMS panel by clicking on its icon or by typing bsmsdisp. 5. Adjust the height of the lock signal by changing the lock gain (in the range between 100 to 115 ) and power (depending on the solvent: for 10% D 2 O around -18 db). Ideally, the lock signal (green and red moving lines) is at the same height as the top horizontal grid line. 6. Automatically adjust lock parameters by typing loopadj in topspin command line. 7. Perform tuning and matching ( atmm ). 8. Read the parameter set zg.eth ( rpar zg.eth ). 9. Make sure the following parameters are set properly for the zg.eth experiment: ns 1 (number of scans) ds 0 (dummy scans) 9 bionmr platform ETH Zürich, December 2018

rg 1 (receiver gain) lb 8 (line broadening) 10.Perform shimming ( topshim gui ) and click on Start. 11.Start the experiment by typing zg. 12.Process the data using the following commands: em (exponential multiply) ft (fourier transform) apk (automated phase correction) 13.Calculate the sample temperature from the processed methanol spectrum by typing calctemp (select D, which is the default) Calctemp computes the actual sample temperature using the chemical shift difference between the methanol methyl and OH proton resonances. 14.Compare the temperature calculated by calctemp with the set temperature. If the calculated and the set temperature are different then adjust the set temperature and repeat the temperature measurement (steps 10 to 13). Basic experiments 1. zg-wg3919.eth : one dimensional proton experiment using the 3-9-19 watergate for water suppression important parameters: o1, p1, pl1 (pl1db), ns, rg 2. 15Nhsqc.eth: 2D 1H15N HSQC important parameters: o1, p1, pl1, p3, pl3, ns, rg, 1 td (number of increments) Interactive optimization of acquisition parameters In the previous sections we have described how to determined all parameters required to start the experiment, like the pulse lengths and o1. However, for samples in aqueous buffer the parameters for water suppression using the presaturation method need to be adjusted interactively. Once the water suppression is optimized, the receiver gain (rg) can be set to the highest possible value. Interactive optimization of parameters is done by adjusting the parameters during acquisition. Clicking above or below the slider button moves the slider a small increment in that direction. The command gs open a dialog box and starts interactive acquisition. Note that in gs mode the acquired data is not stored on the disc. Open the experiment whose parameters you need to optimize and type gs. gs left panel Modify the parameter using the slider. 10 bionmr platform ETH Zürich, December 2018

Save - save the parameter that was changed last. Save all - save all changes. Restore - restore the parameter that was changed last. Restore all - restore all changes. Stop - stop the acquisition and leave the gs dialog box. Optimize receiver gain (rg) The receiver gain determines the amplification of the raw NMR signal in the receiver. A low receiver gain results in noisy spectra. A too high rg results in receiver overflow by the high intensity water signal. The goal is to set rg to at least 64. 1. Start gs mode by typing gs. 2. Click on the receiver tab on the left panel. A slider appears to interactively adjust the receiver gain. 3. Move the slider up or down to increase or decrease the receiver gain. Changes in the rg can be immediately observed in the intensity of the FID (free induction decay) on the right panel. 4. Too high rg values result in a intense FID which will be chopped off at the top and bottom of the FID window and the message: Warning ADC overflow will appear. Low values of rg result in a weak FID signal and noisy spectra. 5. Increase the rg to just below ACD overflow. 6. Click on Save and Stop on the left panel. The new rg value is now stored for the current data set. Low rg values are usually caused by poor water suppression. Reasons for poor water suppression are (1) Misadjusted presat power level (pl9) (2) Misadjusted proton carrier frequency (3) poor shims or a combination of the three. Optimize water suppression (presaturation) One of several possibilities for water suppression is the presaturation method. The parameters to be optimized are the power level for the presaturation (pl9 in W or pl9db in db) and the offset frequency for the proton channel (o1). The goal is is to keep the intensity of the water signal low. Typical values for pl9db are around 48 db. How to obtain the o1 value see section 1H pulse calibration (point 11). 11 bionmr platform ETH Zürich, December 2018

1. Start gs mode by typing gs. 2. Adjust the rg according to the previous section. 3. Click on Square Power on the left panel. 4. Select pl9db using the radio button. 5. Change the sensitivity for the slider to 0.1 by clicking on the sensitivity arrow down button. 6. Adjust pl9db using the slider in 0.1 db steps. Note that it takes 2 or 3 scans (about 3 seconds) for the change to become visible in the FID. In case of poor water suppression most of the FID intensity result from water. A decrease in the water signal intensity can therefore be observed by an overall intensity decrease of the FID. Decrease pl9db until the FID intensity does not decrease any further. In case decreasing pl9db does not result in decreased FID intensity try to increase pl9db. 7. Click on Save to store the optimized pl9db value for the current data set. 8. Click on the Frequency tab on the left panel. 9. Select o1 using the radio button. 10.Change the sensitivity for the slider to 0.1 by clicking on the sensitivity arrow down button. 11.Adjust o1 in 0.1 increments (increase or decrease) and observe the change in the shape of the FID. If the o1 value is off the center of the water signal the FID appears wavelike. Try to adjust o1 to obtain an FID which is symmetric around the time axis. 12. Click on Save and Stop on the left panel. 12 bionmr platform ETH Zürich, December 2018