Phase Imaging Using Focused Polycapillary Optics

Similar documents
Phase Contrast Imaging with X-ray tube

Introduction. Chapter 16 Diagnostic Radiology. Primary radiological image. Primary radiological image

ANALYTICAL MICRO X-RAY FLUORESCENCE SPECTROMETER

MC SIMULATION OF SCATTER INTENSITIES IN A CONE-BEAM CT SYSTEM EMPLOYING A 450 kv X-RAY TUBE

Acceptance Testing of a Digital Breast Tomosynthesis Unit

10.2 Images Formed by Lenses SUMMARY. Refraction in Lenses. Section 10.1 Questions

DALLA LUCE VISIBILE AI RAGGI X: NUOVI RIVELATORI DI IMMAGINI PER RAGGI X A DISCRIMINAZIONE IN ENERGIA ED APPLICAZIONI

SECTION I - CHAPTER 2 DIGITAL IMAGING PROCESSING CONCEPTS

X-ray Imaging. PHYS Lecture. Carlos Vinhais. Departamento de Física Instituto Superior de Engenharia do Porto

Mammography is a radiographic procedure specially designed for detecting breast pathology Approximately 1 woman in 8 will develop breast cancer over

MINIATURE X-RAY SOURCES AND THE EFFECTS OF SPOT SIZE ON SYSTEM PERFORMANCE

An Activity in Computed Tomography

This experiment is under development and thus we appreciate any and all comments as we design an interesting and achievable set of goals.

On spatial resolution

INTRODUCTION THIN LENSES. Introduction. given by the paraxial refraction equation derived last lecture: Thin lenses (19.1) = 1. Double-lens systems

Confocal Imaging Through Scattering Media with a Volume Holographic Filter

Mammography: Physics of Imaging

Exposure schedule for multiplexing holograms in photopolymer films

LECTURE 1 The Radiographic Image

IBEX TECHNOLOGY APPLIED TO DIGITAL RADIOGRAPHY

I. PERFORMANCE OF X-RAY PRODUCTION COMPONENTS FLUOROSCOPIC ACCEPTANCE TESTING: TEST PROCEDURES & PERFORMANCE CRITERIA

X-RAY IMAGING EE 472 F2017. Prof. Yasser Mostafa Kadah

GEOMETRICAL OPTICS Practical 1. Part I. BASIC ELEMENTS AND METHODS FOR CHARACTERIZATION OF OPTICAL SYSTEMS

Lenses. A lens is any glass, plastic or transparent refractive medium with two opposite faces, and at least one of the faces must be curved.

MICRO XRF OF LIGHT ELEMENTS USING A POLYCAPILLARY LENS AND AN ULTRA THIN WINDOW SILICON DRIFT DETECTOR INSIDE A VACUUM CHAMBER

PD233: Design of Biomedical Devices and Systems

FRAUNHOFER AND FRESNEL DIFFRACTION IN ONE DIMENSION

Investigation of the line-pair pattern method for evaluating mammographic focal spot performance

Seminar 8. Radiology S8 1

Comparison of FRD (Focal Ratio Degradation) for Optical Fibres with Different Core Sizes By Neil Barrie

Beam collimation with polycapillary x-ray optics for high contrast high resolution monochromatic imaging

Multiple Choice Identify the letter of the choice that best completes the statement or answers the question.

Comparison of edge enhancements by phase contrast imaging and post-processing with unsharp masking or Laplacian filtering

Amorphous Selenium Direct Radiography for Industrial Imaging

Physics 3340 Spring Fourier Optics

Optical Coherence: Recreation of the Experiment of Thompson and Wolf

Microspot x-ray focusing using a short focal-length compound refractive lenses

Lenses. Images. Difference between Real and Virtual Images

AQA P3 Topic 1. Medical applications of Physics

X rays X-ray properties Denser material = more absorption = looks lighter on the x-ray photo X-rays CT Scans circle cross-sectional images Tumours

Radiology Physics Lectures: Digital Radiography. Digital Radiography. D. J. Hall, Ph.D. x20893

Chapter 18 Optical Elements

AP Physics Problems -- Waves and Light

Applying of refractive beam shapers of circular symmetry to generate non-circular shapes of homogenized laser beams

Multi-Lateral Shearing Interferometry: Principle and Application on X-ray Laboratory Sources

Laser Beam Analysis Using Image Processing

research papers First results from a macromolecular crystallography system with a polycapillary collimating optic and a microfocus X-ray generator

Criteria for Optical Systems: Optical Path Difference How do we determine the quality of a lens system? Several criteria used in optical design

Assembly and Experimental Characterization of Fiber Collimators for Low Loss Coupling

The Evaluation of Collimator Alignment of Diagnostic X-ray Tube Using Computed Radiography System

CHAPTER 9 POSITION SENSITIVE PHOTOMULTIPLIER TUBES

Construction of a Talbot Interferometer for phase-contrast imaging

Experiment 1: Fraunhofer Diffraction of Light by a Single Slit

COMPUTED RADIOGRAPHY CHAPTER 4 EFFECTIVE USE OF CR

An Activity in Computed Tomography

Bias errors in PIV: the pixel locking effect revisited.

IBEX MATERIALS DETECTION TECHNOLOGY

Characterization of a High-Energy X-ray Compound Refractive Lens

Laser Scanning for Surface Analysis of Transparent Samples - An Experimental Feasibility Study

APPLICATIONS FOR TELECENTRIC LIGHTING

LOS 1 LASER OPTICS SET

Acquisition, Processing and Display

Polycapillary optic source combinations for protein crystallography

Collimation Assessment Using GAFCHROMIC XR-M2

2017 West Coast Educators Conference Orlando. Projection Geometry. 1. Review hierarchy of image qualities (amplified version):

STUDENT REVIEW QUESTION SET K CR/DR CONTENT AREA

TSBB09 Image Sensors 2018-HT2. Image Formation Part 1

IMAGE SENSOR SOLUTIONS. KAC-96-1/5" Lens Kit. KODAK KAC-96-1/5" Lens Kit. for use with the KODAK CMOS Image Sensors. November 2004 Revision 2

X-ray phase-contrast imaging

Be aware that there is no universal notation for the various quantities.

FIRST INDIRECT X-RAY IMAGING TESTS WITH AN 88-mm DIAMETER SINGLE CRYSTAL

Thermal tuning of volume Bragg gratings for high power spectral beam combining

Transmission- and side-detection configurations in ultrasound-modulated optical tomography of thick biological tissues

ADVANCED OPTICS LAB -ECEN Basic Skills Lab

Optimization of Digital Mammography Resolution Using Magnification Technique in Computed Radiography 1

X-rays. X-rays are produced when electrons are accelerated and collide with a target. X-rays are sometimes characterized by the generating voltage

Quality control of Gamma Camera. By Dr/ Ibrahim Elsayed Saad 242 NMT

INTRODUCTION TO FLEXIBLE BRONCHOSCOPY. Fluoroscopy Synopsis HENRI G COLT MD SECOND EDITION THE BRONCHOSCOPY EDUCATION PROJECT SERIES

Unit thickness. Unit area. σ = NΔX = ΔI / I 0

MICROANALYSIS WITH A POLYCAPILLARY IN A VACUUM CHAMBER

Applications of Optics

Imaging Technique Optimization of Tungsten Anode FFDM System

Talbot- Lau interferometry with a non- binary phase grating for non-destructive testing

Small-bore hollow waveguides for delivery of 3-mm laser radiation

EE119 Introduction to Optical Engineering Spring 2002 Final Exam. Name:

ADVANCED OPTICS LAB -ECEN 5606

CR Basics and FAQ. Overview. Historical Perspective

Εισαγωγική στην Οπτική Απεικόνιση

Beam-Restricting Devices

Introduction. Strand F Unit 3: Optics. Learning Objectives. Introduction. At the end of this unit you should be able to;

Published text: Institute of Cancer Research Repository Please direct all s to:

Examination, TEN1, in courses SK2500/SK2501, Physics of Biomedical Microscopy,

Q-switched resonantly diode-pumped Er:YAG laser

Film Replacement in Radiographic Weld Inspection The New ISO Standard

Observational Astronomy

LENSES. INEL 6088 Computer Vision

Gas scintillation Glass GEM detector for high-resolution X-ray imaging and CT

EXPRIMENT 3 COUPLING FIBERS TO SEMICONDUCTOR SOURCES

HIGH RESOLUTION COMPUTERIZED TOMOGRAPHY SYSTEM USING AN IMAGING PLATE

Visibility of Detail

Transcription:

Phase Imaging Using Focused Polycapillary Optics Sajid Bashir, Sajjad Tahir, Jonathan C. Petruccelli, C.A. MacDonald Dept. of Physics, University at Albany, Albany, New York Abstract Contrast in conventional imaging of soft tissues is often limited due to the very similar attenuation of tissues to be distinguished. Phase contrast techniques can enable discrimination of tissues with similar attenuation. A major limitation to the widespread adoption of phase-contrast techniques is that for tabletop sources the required degree of coherence generally requires a small (1 to 5 µm) source. In this work, a polycapillary optic was employed to create a small virtual source from a large spot rotating anode. Phase contrast images obtained with two optics and several pinholes have been analyzed and preliminary results obtained for quantitative phase measurements. Keywords: Phase imaging, propagation-based, edge enhancement, radiography, polycapillary optics, virtual source. 1. Introduction Since the discovery of X rays by Roentgen, absorption-based images utilize the attenuation properties of the object being imaged to generate a two dimensional map. Such types of images produce high contrast for objects where the electron density differences are substantial, e.g., bone and flesh, but result in poor contrast for soft tissue imaging where the tissues exhibit similar attenuation, e.g., glandular tissue and infiltrating ductal carcinoma in breast imaging. 1 However, X rays can accumulate significant differential phase delay even in weakly absorbing materials. If the change of phase of X rays can be imaged, tissues which would normally be indistinguishable under attenuation contrast can be distinguished. Phase contrast imaging renders different phase delays as intensity variation in the detector plane. The amount of attenuation and phase delay of a given material can be described by its complex-valued index of refraction, where β and δ are real-valued parameters describing the absorption and phase delay of X rays, respectively. For X-ray energies of 1 to 1 KeV, δ of tissues is about 1-6 to 1-8 and is about 1-9 to 1 11, so the phase delay term is approximately 1 times greater than the absorption term. Because δ remains relatively large at high energies where absorption is low, phase imaging may give the possibility of dose reduction at these energies. Several phase contrast methods 2 have been investigated to visualize phase delay. The proposed techniques all require a fairly spatially coherent source, which for typical medical X-ray sources requires a small source spot. Aside from the requirement on source coherence, many techniques utilize optical elements or gratings of fine pitch to obtain phase contrast images. Here, we investigate the simplest to implement X-ray phase contrast techniques: propagation based phase imaging, in which the physics of X-ray propagation through free space generates an image with phase contrast. In the current study, this method has been implemented to image several test objects including small diameter rods and insects. khokhar.sajid@gmail.com, sbashir@albany.edu (1) Advances in X-Ray/EUV Optics and Components IX, edited by Christian Morawe, Ali M. Khounsary, Shunji Goto, Proc. of SPIE Vol. 927, 927X 214 SPIE CCC code: 277-786X/14/$18 doi: 1.1117/12.263229 Proc. of SPIE Vol. 927 927X-1

In order to generate a source with the required spatial coherence, polycapillary optics were used to focus the beam from a polychromatic rotating anode Mo X-ray source through a pinhole. Polycapillary optics are a collection of small hollow glass tubes through which X rays are guided along the length of the optics by total external reflection and the exiting beam is convergent to create a small intense source of X rays. 3 The differences in phase delay induced by the objects, transformed into an intensity pattern on the detector plane can be quantified using the transport of intensity equation. In the near-field region and for weakly attenuating objects, the intensity is described by 4 ) (2) where is the magnification, is the distance of object from the X-ray source, R 2 the distance of detector from the object, I o is the incident intensity, T is the object transmission, ) is the phase delay introduced by the object. is the de-focusing distance and 2. Experiment Set Up A 5 KW Molybdenum (Mo) rotating anode X-ray source with a source size of 3 µm was used. Three types of detectors were also used for different purposes: for alignment a solid state CdZe Amptek XR-1 CR X-ray detector; and for imaging a Fuji Image plate M1419 with an active area of 1 cm 12 cm and 5 µm resolution or a phosphor coated CCD camera with an active area of 3 cm 4 cm and a pixel size of 22 µm. A 1.6 mm polyethylene rod was imaged to test edge enhancement. pinhole detector Figure 1. Schematic diagram of the experimental set-up. The optic was translated horizontally and vertically at each point along z-axis to align it with the x-ray source. The input and output focal lengths of optic were 81 mm and 5.5 mm as given in Table 1. In this geometry, the polycapillary optic collected radiation from a divergent source and focused onto a small point. Table 1. Characteristics of Polycapillary Optic Manufacturer X-ray Optical System Inc. ID Number 69 Input Focal Length (mm) 81 Output Focal Length (mm) 5.5 Channel Diameter (µm) 9.4 Proc. of SPIE Vol. 927 927X-2

In order to align the optic for maximum transmission, it was first placed as close to the source as possible and translated iteratively in the horizontal and then in the vertical direction to achieve the best alignment. The optic was moved away from the source and same procedure was repeated. At each distance, the full width at half maximum (FWHM) of the scan along the x and y axes was fit to a Gaussian distribution ( ) (3) using Origin 6. software. At the input focal point, the width was minimum with a value of.34 ±.2 mm, as shown in Figure 2. The variation of beam width with the distance is shown in Figure 3. The minimum expected value of the beam width w calc at the input focal point is approximately 5 ) (4) where s is the source size of 3 µm, f in is the input focal length and is critical angle for total reflection, which for the borosilicate glass is For an X-ray energy of 2 KeV, the critical angle is 1.5 mrad. The expected beam width is.35 mm, which is in good agreement with the measured value. (5) Intensity 14 12 1 8 6 4 2-1. -.5..5 1. displacement from beam axis(mm) data Gaussian fit An image plate was used to estimate the output spot size. The image plate (Fuji M1419) was placed close to the optic focal spot and the resulting image was then read in film reader (Fuji Films BAS-18) using Image Gauge software. The smallest resolution of 5 µm was selected during reading of the image plate. Figure 4 shows the magnified view of the spot at a distance of 5.5 mm from the optic. The profile across the focal spot size drawn in Figure 4 gave a fit Gaussian distribution with a width of 114 ± 5 µm at the output focal point. The variation of spot size with the distance from the optic is shown in Figure 5..44.42 -E.4.38 =.36 L_.34.32 Figure 2. Graph resulted from the scan of optic at the input focal point. The red line is the Gaussian fit to data. The width is.34 ±.2 mm..3 6 7 8 9 1 Distance from source (mm) Figure 3. Alignment of optic with the x-ray source. At the input focal point, the beam width is minimum. Proc. of SPIE Vol. 927 927X-3

Data Gaussian fit 14 12 1 8 intensity 6 4 2-2 -.2..2.4.6.8 1. 1.2 1.4 1.6 1.8 Profile range (mm) Figure 4. Magnified view of focal spot at the output focal point. The profile across the spot was drawn and the resulting graph was used to estimate the FWHM. The expected output focal spot size of the optic is where c is the channel size of a single fiber, f out is the output focal length of the optic. For the optic used here, expected value of spot size is 12.3 µm. The difference between the measured and expected focal spot size may be due to an error in manufacturing of the optic. For short focal length optics it can be difficult to cut the optic at the position for which the channels are directed toward a single point. This optic was actually designed to be used with the short focal length side facing the source, to enable efficient source capture, in which case it was not important that all of channels be precisely aligned. To make the spot size smaller, 25-1 µm diameter pinholes were placed at the output focal point of the optic and carefully aligned to maximize the output. After aligning the optics and pinhole with the X-ray source, the output beam had high enough spatial coherence for phase imaging and was used to form phase enhanced images of rods and insects. (6) 3. Images of Rod 3.1. 1 µm Pinhole 3.1.1. Image Plate A cylindrical 1.6 mm diameter polyethylene rod was imaged using a 1 µm diameter pinhole. The rod was placed 17 cm from the pinhole ( ) and the image plate was positioned at a distance of 64 cm from the rod ( ). The magnification Optic Spot Size (µm) 35 3 25 2 15 1 5 5 1 15 Image plate distance from spot (mm) was 4.7. (7) Figure 5. The source size decreases with distance. It has minimum value at the output focal point. Proc. of SPIE Vol. 927 927X-4

The rod image presents a dark region with the body of the rod due to attenuation and a white band at the rod s edge which is edge enhancement due to the phase imparted by the rod. For quantative analysis, a profile of the rod was drawn perpendicular to the length of the rod and plotted against distance, as shown in Figure 7. This line was created by summing 5 neighboring pixels. Both the intensity dip due to attenuation and the intensity peak due to phase are evident in this plot. The presence of noise in the image diminishes the strength of the phased enhanced signal. The conspicuity of the edge enhancement 6 can be described by the edge-enhancement to noise ratio (EE/N) ratio (8) where P is the peak intensity, is the average intensity over a region of background labeled B in Figure 7, and is the standard deviation of intensity over region B. The edge-enhancement to noise ratio was.91. Figure 6. Image of 1.6 mm diameter rod taken with a 1 µm pinhole. The white lines show edge enhancement. The image in Figure 6 was acquired at 3 kvp and 3 ma for two and a half minutes. The product of current and exposure time is higher than the normally used for soft tissue imaging in clinical settings, 7 however, the intensity loss was before the object and would not create additional patient dose. In addition, an optic which is designed for this application would produce a small spot without a pinhole 5 and thus would have higher intensity and shorter exposure time. Normalized Intensity 1.2 1.8.6.4.2 B P 2 4 6 8 Length of profile drawn normal to rod (mm) Figure 7. Profile of the 1.6 mm diameter rod taken with a 1 µm pinhole. The small peaks on both sides of the valley are due to edge enhancement. Proc. of SPIE Vol. 927 927X-5

3.2. 25 µm Diameter Pinhole 3.2.1. Image Plate The 1.6 mm diameter polyethylene rod was then imaged using a 25 µm diameter pinhole shown in Figure 8. The rod was placed at a distance of 25 cm from the output focal point of the optic ( ) and the image plate was placed approximately 8-81 cm from the rod ( ) giving a magnification of 4.2. The distance was increased to increase the beam coherence. The rod was exposed for 6 minutes with voltage of 3 kvp and a current of 4 ma. The exposure was increased to accommodate the loss of intensity in using the 25 µm diameter pinhole. As with the 1 micron pinhole, the captured image shows attenuation through the rod s body with edge enhancement due to phase at the rod s edges. A five pixel wide profile was drawn normal to the rod in both images and a background division was performed to remove the effect of background intensity variation due to the optic structure in the image. The normalized resulting profile is shown in Figure 9. The peaks on both sides of the valley show edge enhancement. The edge-enhancement to noise ratio was 2.56. Figure 8. Image of a 1.6 mm rod taken with a 25 µm pinhole. The white line along the length of rod shows edge enhancement. Normalized Intensity 1.2 1.8.6.4.2 2 4 6 8 Length of profile drawn normal to rod Figure 9. Profile of the image of figure 8. Edge enhancement is evident from the two peaks. Background division was applied to remove nonuniformity in the beam. 3.2.2. Charged Coupled Device (CCD) Camera The previous rod images were taken with the image plate that had a pixel size of 5 µm. A higher resolution digital detector (CCD camera) with a pixel size of 22 µm was used to test whether the edge enhancement would be improved. Flat-field and dark field images were acquired with no rod in place and were applied to the rod images. The rod-to-pinhole distance was increased to 4 cm, and an image of the rod was taken with the object placed at a distance of 4 cm from source and a detector to rod distance of 67 cm, for a magnification of 2.6. As the exposure per frame was restricted to 3 seconds, average of 3 frames was taken to increase the exposure time to 15 minutes. A bright line along the length of the rod could be seen on both sides of the rod as shown in Figure 1. Figure 1. A 1.6 mm diameter rod image taken with CCD camera after flat field correction. Two bright lines along the length of rod are clearly seen. Proc. of SPIE Vol. 927 927X-6

Additional images were taken at varying object-to-detector distances. The effect of increased magnification on edge enhancement is shown in Figure 11.The edge enhancement increases approximately linearly with increasing distance of detector to object as expected from the work of other investigators 6 using a small spot microfocus source with no optic. The virtual source provided by a polycapillary optic and pinhole performed as a microfocus source. 51 7 6 4 3 c 1 m 1 2 3 4 5 6 7 Distance of detector from object (cm) Figure 11. The effect of distance R 2 on edge enhancement. The edge enhancement to noise ratio increases with distance. Insect Images Insect images were taken to demonstrate the feasibility of phase contrast in biological tissues. A conventional absorption image, shown in Figure 12, was also acquired with the insect at a distance of R 1 = 4 cm from the source, with the image plate placed just behind the object at a distance of R 2 = 2 cm. The phase image shown in Figure 13 was taken at a distance of 67 cm from the insect keeping R 1 = 4 cm. The phase image in Figure 13 shows not only the internal structures (due to absorption) but also shows clearly defined boundaries evidencing edge enhancement. Some of the improvement is due to the smaller effective pixel size, because of which there is visible edge-enhancement. Figure 12. Absorption image taken with the detector placed just behind the insect. Figure 13. Edge-enhanced image of the insect with the detector was placed at 67 cm from the insect. The edges of internal structures are clearly visible even though the lower resolution detector was employed. Proc. of SPIE Vol. 927 927X-7

4. Quantitative Phase Retrieval Quantitative phase retrieval involves using knowledge of the physics that produces phase contrast images in order to retrieve a quantitative map of phase imparted to the X-ray beam by the object. For propagation-based imaging, one attempts to invert the transport of intensity equation, Eq. (2), to recover phase from intensity measurements. Because phase and attenuation information are both present in any pixel of the image, one typically acquires two (or more) images at different propagation distances to allow unambiguous phase reconstruction. 8 Capturing multiple images for processing can be difficult as the phase features are fine structures in the images (enhanced edges) which can shift by many pixels on the detector as the distances are changed in the imaging system. Even shifts of a single pixel can produce severe artifacts in a reconstructed image. To get around this problem, there are a number of phase retrieval techniques that attempt to reconstruct phase from a single image by making strong assumptions on the relationship between attenuation and phase. In this project, the phase-attenuation duality concept was employed to obtain the phase image of the insect. According to this theory the phase image and the attenuation image can both be determined by the same map of the projected electron density. The electron density was derived from the single phase-contrast image 9 as ) ( { ) [ ( )] }), (9) where I is the intensity in the image plane, I in is the entrance intensity, is the two-dimensional Fourier transform and is it inverse, u is the spatial frequency vector in the object plane, and is the total cross-section for X- ray photon Compton scattering from a single electron derived from Klein-Nishina formula. The right image in Figure 14 is the phase at each pixel, calculated by assuming the phase is proportional to electron density. o Measured intensity PAD phase reconstruction - 2-4 -6-8 (a) (b) Figure 14. Image (a) is the edge-enhanced image of the insect and image (b) is the pure phase image of the same insect. The phase image was obtained by employing the concept of phase-attenuation duality. The left image of Figure 14 is the edge-enhanced image of Figure 13, except after background subtraction with a no object image. Additional features are clearly visible in the phase image. Proc. of SPIE Vol. 927 927X-8

Conclusion Propagation based phase imaging was successfully tested with a large spot rotating anode X-ray source aligned with polycapillary optic. The polycapillary optic accepts radiation from a large X-ray source and converges it to create a small spot. A rod of 1.6 mm diameter was imaged with different geometric configurations using an image plate and CCD camera. Edge-enhancement to noise ratios up to a value of 6.5 were obtained. Conventional absorption and phase images of an insect were acquired to determine the feasibility of phase imaging in biological tissues using this technique. A pure phase object was obtained using a phase attenuation duality approach by taking only one image. Additional optimization of the optical system is required to reduce the exposure time. Acknowledgement The authors gratefully acknowledge the National Institutes of Health (NIH) project number R1EB9715. References 1 Monnin, S. P., Quantitative Characterization of edge enhancmenet in phase contrast x-ray imaging, Med.Phys.31 1372-1383 (24). 2 Momose, A., Recent advances in x-ray phase imaging, Jpn. J. Appl.Phys.44, 6355 (21). 3 MacDonald, C. A, Focusing Polycapillary Optics and their Applications, X-Ray Optics and Instrumentation, special issue X-Ray Focusing: Techniques and Applications, Volume 21 (21), Article ID 86749, 17 pages, doi:1.1155/21/86749. 4 Gureyev, T.E., and Wilkins, S.W., On X-ray phase imaging with a point source, Vol. 15, J. Opt. Soc. Am. A 579-585 (1998). 5 MacDonald, C. A., and Gibson, W.M., Polycapillary Optics, Handbook of Optics, Third Edition Volume V: Atmospheric Optics, Modulators, Fiber Optics, X-Ray and Neutron Optics, M. Bass, C. DeCusatis, J Enoch, V. Lakshminarayanan, G. Li, C.A. MacDonald, V.N. Mahajan, E. Van Stryland, eds., McGraw-Hill (21). 6 Donnelly, E.F. Characterization of the phase-contrast radiography edge enhancement in a cabinet x-ray system. Physics in Medicine and Biology, 51, 21-3 (26). 7 American Association of Physicist in Medicine, Equipment Requirement and Quality Control for Mammography, Report of Task Group 29, (199). 8 Gureyev,T. E., et al. Refracting Rontgen s rays: Propagation-based x-ray phase contrast for biomedical imaging, J. Appl. Phys. 15, 125 (29). 9 Xizeng Wu, et al., X-ray Phase-attenuation Duality and Phase Retrieval, Vol. 3, No. 4 Optics Letters (24). Proc. of SPIE Vol. 927 927X-9