Quick Start MMRC Nicolet 6700 Revised 10/24/2018

Similar documents
Seized Drugs Operational Guidelines for the Thermo FTIR Comparative and Analytical Division

Guide to Operating the Bruker FT-IR Microscope I. Basic Spectrum Acquisition with Vertex 70

Telephone: NICOLET ( ) or

Development and Applications of a Sample Compartment FTIR Microscope

The operation manual of spotlight 300 IR microscope

Measurement Method of High Absorbance (Low Transmittance) Samples by UH4150 INTRODUCTION

RENISHAW INVIA RAMAN SPECTROMETER

Spotlight 150 and 200 FT-IR Microscopy Systems

The FTNIR Myths... Misinformation or Truth

OPT3: Operating Procedure for Horiba Jobin Yvon LabRam Aramis Raman/PL System See LabSpec_6_2 General User Quick Start Guide on the computer desktop

Spectrum GX systems FT-IR SPECTROSCOPY.

UV-Vis-NIR Spectrophotometer Quick Start Guide

FTIR Spectrophotometer

Modular Fourier Transform Infra-Red Spectrometer (FT-IR) Solution

IBIL setup operation manual for SynerJY software version

WITec Alpha 300R Quick Operation Summary October 2018

Infrared wire grid polarizers: metrology, modeling, and laser damage threshold

Chemistry 524--"Hour Exam"--Keiderling Mar. 19, pm SES

Automated Double Aperture Accessory

Kit for building your own THz Time-Domain Spectrometer

Horiba Jobin-Yvon LabRam Raman Confocal Microscope (GERB 120)

Single-Slit Diffraction. = m, (Eq. 1)

PANalytical X pert Pro Gazing Incidence X-ray Reflectivity User Manual (Version: )

Confocal Raman Microscopy (WITec Alpha 300R)

Things to check before start-up.

Instruction manual for T3DS software. Tool for THz Time-Domain Spectroscopy. Release 4.0

The IVIS Spectroscopy Laboratory (v1.5)

Horiba LabRAM ARAMIS Raman Spectrometer Revision /28/2016 Page 1 of 11. Horiba Jobin-Yvon LabRAM Aramis - Raman Spectrometer

Bruker Dimension Icon AFM Quick User s Guide

Ocean Optics R-2000 Raman Spectrometer Setup and Operating Instructions Arlen Viste and Deanna Donohoue April 2000 Update 2003, DEW

SOP-P051. Scanning of Optical Filters With USB2000. Objective: To determine the spectral transmittance properties of an optical filter.

Microscopy from Carl Zeiss

Supplementary Materials

Installing the detector filter assembly (G )

TriVista. Universal Raman Solution

771 Series LASER SPECTRUM ANALYZER. The Power of Precision in Spectral Analysis. It's Our Business to be Exact! bristol-inst.com

Advance Concrete. Tutorial

Renishaw InVia Raman microscope

Installation and User Guide. FlexIR TM NIR Fiber Optic Accessory

3.003 Lab 3 Part A. Measurement of Speed of Light

Lab 12 Microwave Optics.

Agilent Cary 7000 Universal Measurement Spectrophotometer (UMS)

Chapter 2 Fundamentals of FT-IR

Welcome to SPDL/ PRL s Solid Edge Tutorial.

BEAMAGE-3.0 KEY FEATURES BEAM DIAGNOSTICS AVAILABLE MODELS MAIN FUNCTIONS SEE ALSO ACCESSORIES. CMOS Beam Profiling Cameras

Experiment 19. Microwave Optics 1

EDUCATIONAL SPECTROPHOTOMETER ACCESSORY KIT AND EDUCATIONAL SPECTROPHOTOMETER SYSTEM

ERS KEY FEATURES BEAM DIAGNOSTICS MAIN FUNCTIONS AVAILABLE MODEL. CMOS Beam Profiling Camera. 1 USB 3.0 for the Fastest Transfer Rates

Educational Spectrophotometer Accessory Kit and System OS-8537 and OS-8539

Dario Cabib, Amir Gil, Moshe Lavi. Edinburgh April 11, 2011

BEAMAGE KEY FEATURES AVAILABLE MODELS. CMOS Beam Profiling Cameras

Physics 476LW. Advanced Physics Laboratory - Microwave Optics

TRAINING MANUAL. Olympus FV1000

TRAINING MANUAL. Multiphoton Microscopy LSM 510 META-NLO

Lab 5: Brewster s Angle and Polarization. I. Brewster s angle

Bruker Dimension Icon AFM Quick User s Guide

Laboratory 1: Motion in One Dimension

Thermo Scientific SPECTRONIC 200 Education

SUPPLEMENTAL MATERIAL

Applications of Steady-state Multichannel Spectroscopy in the Visible and NIR Spectral Region

Basic P-XRD instructions for Operating the Instrument

Part 1: Standing Waves - Measuring Wavelengths

NIR SPECTROSCOPY Instruments

Getting Started. Spectra Acquisition Tutorial

ENSC 470/894 Lab 3 Version 6.0 (Nov. 19, 2015)

NCSL International 2995 Wilderness Place, Suite 107 Boulder, Colorado Office: (303) Fax: (303)

Shimadzu RF-5301 Fluorimeter operation guide for students

Thermo Scientific SPECTRONIC 200 Visible Spectrophotometer. The perfect. teaching instrument

Unique Scattering Measurements Using the Agilent Universal Measurement Accessory (UMA)

Swept Wavelength Testing:

Instructions XRD. 1 Choose your setup , Sami Suihkonen. General issues

PHYS 3153 Methods of Experimental Physics II O2. Applications of Interferometry

INTRODUCTION TO DATA STUDIO

Chem 524 Lecture notes (Sect. 7) update 2011

Experiment P55: Light Intensity vs. Position (Light Sensor, Motion Sensor)

Borders. Use QuiltMotion Touch for BERNINA to create a variety of border styles. Quick Borders. Borders using Simple Block Designs

RAITH e-line OPERATING INSTRUCTIONS

Lab 1B LabVIEW Filter Signal

Measuring optical filters

Information & Instructions

PHONE: +1 (519) FAX: +1 (519)

Microwave Optics. Department of Physics & Astronomy Texas Christian University, Fort Worth, TX. January 16, 2014

Using the Hitachi 3400-N VP-SEM

Fourier Transform Infrared Spectrophotometer. IRAffinity-1 C103-E076D

Conversational CAM Manual

Modeling Basic Mechanical Components #1 Tie-Wrap Clip

11Beamage-3. CMOS Beam Profiling Cameras

Technical Procedure for Infrared Spectroscopy

Operating Checklist for using the Scanning Electron. Microscope, JEOL JSM 6400.

PANalytical X pert Pro High Resolution Specular and Rocking Curve Scans User Manual (Version: )

Training Guide for Carl Zeiss LSM 510 META Confocal Microscope

Introduction to Autodesk Inventor for F1 in Schools (Australian Version)

University of Wisconsin Chemistry 524 Spectroscopic Components *

The quantitative relationship between distance, time and speed

Optical Spectrum Analyzers

MICROWAVE OPTICS. Instruction Manual and Experiment Guide for the PASCO scientific Model WA-9314B G

Content Spectrophotometers

FTIR. Far - Infrared Teraher t z. Tests. Instruments. Terahertz Spectrometer

FPPO 1000 Fiber Laser Pumped Optical Parametric Oscillator: FPPO 1000 Product Manual

PM-IRRAS Accessory. Hardware Manual

Transcription:

Do Not Remove Quick Start MMRC Nicolet 6700 Revised 10/24/2018 1. Fill out the logbook before you begin. 2. Before you start the Nicolet should be on with purge gas flowing. 3. Check that the purge gas is N2 and that it has a flow rate of 30 ft 3 /h (14 l/s). The purge gas flow meter is behind the FTIR. Note that you can purge the system at a higher flow rate but at higher flow rates the performance of the spectrometer is not as good. 4. If you want to use the MCT or Si detector you need to check that they are in the beam path (if you cannot tell you can see what is installed when you run Omnic). See table below for ranges of the detectors and beamsplitters. 5. For the MCT detector you need Liq N2 for cooling. You add the LN2 using the small white funnel near the instrument and opening the little door above the detector. Add the LN2 slowly at first because the rapid boil off tends to not let much LN2 into the detector. It will take about 5 10 minutes to fill it the first time and the detector holds ` 750ml of LN2. Allow the detector to cool ~20 minutes Detector Ranges/cm -1 Bean Splitter Ranges/cm -1 DTGS MCT Si KBr Quartz 10,000 420 10,000 370 25,000-9,000 (400-1100 nm) 7,800 370 28,000 4,000 (350-2,500 nm) 6. If you have been trained and need to change the beam splitter and or the detectors see the directions for that below. 7. If you need to install an accessory do that now. 8. Start the OMIC software. It will offer an Experiment which if you know what you want you can choose. Otherwise cancel the experiment window. 9. You should now see the window shown in Figure 1. 10. Choose the icon Expt Set to setup your experiment. 11. The experiment setup window will open with 6 tabs on top. a. Choose the Bench tab shown in Figure 2. i. You need to choose Figure 1 Omnic main screen your detector, beamsplitter and source. The drop down menus for detector and source will show what is already in the spectrometer. If you need to change any of them see instructions below.

ii. You should also set the wavelength range. Remember if there are absorptions outside the range you choose they will be folded into the range you are observing. Best if any region excluded has no light intensity. Figure 2 Bench tab in Experimental setup window. iii. The Gain, optical velocity and aperture interact. 1. Optical Velocity: This is the speed of the moving mirror in the interferometer. The velocity can be set from 0.18 to 6.3 in relative units. The slower the velocity the long it takes to collect a spectrum; however, as you increase the velocity you decrease the intensity of the spectrum. So sometimes with a slower velocity you can take fewer scans. a. DTGS detector scan rate should be 0.6329 if uncooled with 4 cm -1 resolution. b. MCT can scan as fast as 1.89 at 4 cm -1 2. Aperture: This determines the area of the opening to the light beam. Doubling the aperture doubles the area and should also roughly double the amount of light hitting the detector. However, a larger aperture will increase the absolute signal but also decrease the resolution. The instrument will tell you if the Detector Aperture DTGS 63 MCT, or Si 32 aperture is too large for the selected resolution. 3. Gain: The gain needs to be set so that there is sufficient signal to measure.

a. The autogain setting will measure the signal just before the spectrometer starts to collect a spectrum and set it accordingly. This takes only a small amount of time. b. Otherwise one can set the gain by putting you sample in the spectrometer and setting the gain so that the peak to peak maximum of the interferogram is between 3 and 10. b. Choose the Collect tab and you will see the window show in Figure 3 i. Set the No. of scans, the Resolution, and the check boxes you want. Figure 3 Bench tab of Experimental Setup Window When you change the resolution, it interacts with the settings on the Bench tab. ii. The Final Format can be Interferogram, SingleBeam, %T, Abs, etc. If you choose SingleBeam you will need to ratio your reference and sample spectra to get %T after taking the spectra (see 15 d ii Processing below). If you choose %T or Abs you need to take a reference scan before you take a sample scan and the instrument automatically references the two. iii. Correction: This is for the use of accessories like ATR, reflectance or other. For normal transmission spectra none is ok iv. If you automatically save the spectra make sure the path is the one to your folder. (You can set the path under the menu Edit:options.) v. It is always good practice to add an Experimental description that tells the sample you are running. vi. Finally, you can save your setup but since everyone s setup is saved in the same folder you need to add you name or initials at the start of the file and a descriptive title like BSB_Abs1000. c. Diagnostic tab: You can align the bench in this table ( need to set the gain to 1 to do an align).

d. Advanced tab: Can set Zero filling (none but I like level 1), Apodization (Happ Genzel), Phase correction (Mertz), sample spacing (2 or IR, 1 for NIR), filters (based on velocity) and single sided (not check) this gives a double side interferogram. Note that when you reprocess (see 15 d ii below) that these same parameters need to be set. 12. Once all the experimental parameters are set close the Experimental Setup window. 13. Now on the main Omnic window choose collect sample Col Smp icon if you are in single beam mode or Collect background Col Bkg if you are in Transmission or Abs mode. In the latter case after it collects the background put in your sample and collect the sample spectrum. As if scans you will see the spectrum in the Figure 4 Spectra Window plotting area of the window. Hint: if you choose %T or Abs you will see an actual spectrum; however, if you choose single beam you will see the single beam intensity with doe not show much spectral information. 14. You can save your spectra in a folder with your name. All spectra folders have the path C:\my_documents\omnic\spectra\XXXX. Please do not save spectra to the desktop or in other people s folders. 15. Spectra Window: Figure 4 a. The red spectrum is active. There can me multiple spectra shown at one time. You can delete or save the active spectrum. Select a different spectrum by clicking on it. Select multiple by shift clicking. b. The View finder at the bottom right of window can change be used to change the wavenumber/wavelength limits of plot c. Tools: Lower left of window i. Right 4 icon lets you set x and y scale ii. Yellow arrow tool at left lets you select, expand or contract, move spectra iii. Other tools let you choose region of spectrum, create x,y cursor, measure peak height, measure peak area, and annotate. 16. Options for other menus: a. File: Save selected spectra, Open spectra, save and open configuration file. b. Edit: editing annotations, toolbar, or menus, Options i. Options: set path and many many other options. Note if you change the default options please save your configuration with your name and do not make it the default configuration.

c. View: Allows you to control how the spectrum is displayed: i. Full scale draws all the spectra in the plot area so that each one will have a peak that goes full scale. The only spectra that relates to the Y axis shown is the one selected. In this mode the amplitudes on various spectra cannot be compared. ii. Common Scale draws the spectra show that they are all on the same scale. Be careful is the spectra have fundamentally difference Y axes such as one with %T and another with Abs. d. Process: Allows you to post-process a displayed spectrum: i. Abs, %T, etc. converts from one to the other. ii. Reprocess: Allows one to convert a spectrum to other formats and correct it using a background file (see details in Reprocessing Single Beam Spectra section below). iii. Baseline correction: Allows one to manually select a baseline and subtract it from a selected spectrum. iv. Subtract: Allows one to subtract two displayed spectra. Note that the two spectra must be simultaneously selected by holding down the control key before selecting Subtract. 17. When you are done make sure the instrument is purging with air. You can quit Omnic. Reprocessing SingleBeam Spectra: 1) SingleBeam spectra can be converted to other formats such as %T, Abs, etc. and simultaneously corrected with a background file using the Reprocess built-in function in the OMIC software. a. Note that a saved background spectrum is needed in order to use Reprocess. 2) The Reprocess window can be accessed by selecting the Process tab at the top of the main window (see Figure 1) and then selecting Reprocess among the available option. 3) In the Reprocess window (see Figure 5) select an option for the following parameters: a. Resolution: Input the resolution used for collecting the selected single beam spectrum. b. Apodization: Happ Genzel is recommended. c. Phase correction: Mertz is recommended. d. Zero filling: none can be selected but Bruce likes Figure 5 Reprocess Window level 1. e. Final format: Interferogram, SingleBeam, %Transmittance, Absorbance, Kubelka- Munk, Photoacoustic, %Reflectance, Log(1/R) or Raman spectrum. f. Correction: None is recommended.

g. Background file: click on browse to select the corresponding background single beam file. Note that the background and single beam spectrum should have been collected with the same resolution, limits, source, detector, etc. 4) Once all parameters have been selected click on OK to convert and correct the selected SingleBeam spectrum. FTIR Hints: 1) DTGS detector scan rate should be 0.6329 if uncooled. 2) MCT can scan as fast as 1.8988, 3) Gain of detector should be set so the peak to peak reading for the interferogram is ~ 10 V p to p 4) In most cases you should scan forward and back and 2 sided 5) Quick checks of Instrument: a) DTGS detector. With nothing in beam path, using the DTGS detector, scan rate of 0.4747, aperture of 100%, and gain of 1, you should see an interferogram peak of about +5 and min of -3 or a max to min of about 8V. b) MCT detector. With nothing in beam path, using the MCT detector cool (i.e. LN2 in detector), scan rate of 1.8988, aperture of 18%, and gain of 1, you should see an interferogram peak of about +7.4 and min of -6 or a max to min of about 13.4V. c) ATR accessory. Run background with nothing in the beam. Using the same conditions (mirror velocity, aperture, detector, etc.) run a sample spectrum with only the ATR in the beam and no sample. The throughput should be 30% for the diamond Smart ATR at 1000 cm -1, and Seagull at 2000 cm -1. For the Horizon and Praying Mantis it should be between 24 and 40% at 2000 cm -1. Changing the Detectors: The detectors that we have and their ranges are shown below. Detector Range cm -1 (nm) Sensitivity MCT 11,700-420 (1 10 3-2 10 4 ) D* > 5 x 10 9 cm Hz 1 2 W -1 DTGS 12,500-370 (1 10 3-3 10 4 ) D* > 2 x 10 8 cm Hz 1 2 W -1 Si diode 25,000-9,000 (400 1,100) NEP < 10-14 W Hz -1 2 1) Turn off the power to the Nicolet FTIR (switch on power supply brick). 2) Open the top of the Nicolet, by first loosening three screws on the bottom front of instrument, and prop open. 3) Use a screwdriver to loosen the detector you want to remove. 4) Put the new detector into the place of the removed detector and press down on the base to make sure the electrical plug makes contact. 5) Screw down the base of the detector. 6) Close the top of the Nicolet 7) Turn on the Nicolet. 8) Check that Omnic sees the new detector.

Changing Beamsplitter The beamsplitters have the ranges shown at right. 1) Unlock and open the beamsplitter door (top right-hand side). 2) Rotate the locking handle counter clockwise 180º on the beamsplitter holder. Beamsplitter Range cm -1 (nm) KBr 7,800-370 (1,300 30,000) Quartz 27,000-4,000 (370 2,500) 3) Remove the beamsplitter in the instrument and place it in the storage area. 4) Remove the other beamsplitter from the storage area place it into the instrument. 5) Make sure that the beamsplitter is seated in position and rotate locking handle clockwise 180º. 6) The label on the handle on the beamsplitter should face forward. 7) Close and lock beamsplitter door. 8) Start the Omnic software and go to experiment setup. 9) Go to bench and make sure the correct beamsplitter, detector, and light source are chosen. 10) Go to diagnostics in the experimental setup screen. 11) If you are working in the NIR or vis see instructions on web for NIR or vis. 12) Make sure you see the interferogram (if you do not try to reduce or open the aperture to see if the interferogram appears also check the settings of the beamsplitter, source and detector). 13) If you cannot see an interferogram see below. 14) After you see the interferogram Press align. 15) When alignment completes you are ready.