HybridStage - Automated, large sample-area mapping made easy

Similar documents
Prepare Sample 3.1. Place Sample in Stage. Replace Probe (optional) Align Laser 3.2. Probe Approach 3.3. Optimize Feedback 3.4. Scan Sample 3.

Manufacturing Metrology Team

Using Autofocus in NIS-Elements

Outline: Introduction: What is SPM, history STM AFM Image treatment Advanced SPM techniques Applications in semiconductor research and industry

Standard Operating Procedure

Ionscope SICM. About Ionscope. Scanning Ion Conductance Microscopy. Ionscope A brand of Openiolabs Limited

NanoWizard 3 NanoScience AFM

contents TABLE OF The SECOM platform Applications - sections Applications - whole cells Features Integrated workflow Automated overlay

ATOMIC FORCE MICROSCOPY

Park NX20 The leading nano metrology tool for failure analysis and large sample research.

Indian Institute of Technology Bombay

Nanosurf easyscan 2 FlexAFM

Measurement of Microscopic Three-dimensional Profiles with High Accuracy and Simple Operation

Cutting-edge Atomic Force Microscopy techniques for large and multiple samples

SPM Software Release / JPK Instruments AG - all rights reserved

Nanosurf Nanite. Automated AFM for Industry & Research.

Opterra. Multipoint Scanning Confocal Microscope. Innovation with Integrity. Cell-Friendly, High-Speed, High-Resolution Imaging

Nanomechanical Mapping of a High Curvature Polymer Brush Grafted

Camera Overview. Digital Microscope Cameras for Material Science: Clear Images, Precise Analysis. Digital Cameras for Microscopy

Microscopic Structures

START-UP PROCEDURE 1 THE MICROSCOPE STAND 3 OBJECTIVES 5 STARTING WITH LAS (SOFTWARE) AND SETTING UP THE MICROSCOPE STAND 7

ASM Webinar Digital Microscopy for Materials Science

Nanonics Systems are the Only SPMs that Allow for On-line Integration with Standard MicroRaman Geometries

MFP-3D-BIO. Asylum Research. The only full-capability AFM on an inverted optical microscope

AFM of High-Profile Surfaces

Atomic Force Microscopy (Bruker MultiMode Nanoscope IIIA)

Spatial intensity distribution analysis Matlab user guide

Standard Operating Procedure of Atomic Force Microscope (Anasys afm+)

Leica SP8 TCS Users Manual

NanoFocus Inc. Next Generation Scanning Probe Technology. Tel : Fax:

Practical work no. 3: Confocal Live Cell Microscopy

UNIVERSITY OF WATERLOO Physics 360/460 Experiment #2 ATOMIC FORCE MICROSCOPY

Camera Overview. Digital Microscope Cameras for Material Science: Clear Images, Precise Analysis. Digital Cameras for Microscopy

Nature Protocols: doi: /nprot Supplementary Figure 1. Schematic diagram of Kőhler illumination.

JPK NanoWizard II. User Manual. SPM Software Release 3.3a 04 / JPK Instruments AG all rights reserved

Camera Overview. Digital Microscope Cameras for Material Science: Clear Images, Precise Analysis. Digital Cameras for Microscopy

Chemical Imaging. Whiskbroom Imaging. Staring Imaging. Pushbroom Imaging. Whiskbroom. Staring. Pushbroom

Axio Zoom.V16 The Fluorescence Zoom Microscope for Large Fields

ScanArray Overview. Principle of Operation. Instrument Components

Accessories Handbook. for NanoWizard, ForceRobot and CellHesion. systems. valid from Unlimited possibilities.

Pixel shift in fluorescence microscopy

Guide to Confocal 5. Starting session

Indentation Cantilevers

The NanomechPro Toolkit: Accurate Tools for Measuring Nanoscale Mechanical Properties for Diverse Materials

Flex-Bio. Nanosurf. Versatile Research AFM System for Life Science

Fast Laser Raman Microscope RAMAN

nanovea.com PROFILOMETERS 3D Non Contact Metrology

attosnom I: Topography and Force Images NANOSCOPY APPLICATION NOTE M06 RELATED PRODUCTS G

Proudly serving laboratories worldwide since 1979 SPECIFICATIONS

Flex-Axiom. Nanosurf. The Most Versatile AFM System for Materials Research

SOP for Micro Mechanical Analyzer

Park NX-Hivac The world s most accurate and easy to use high vacuum AFM for failure analysis.

Imaging Carbon Nanotubes Magdalena Preciado López, David Zahora, Monica Plisch

Supplemental Figure 1: Histogram of 63x Objective Lens z axis Calculated Resolutions. Results from the MetroloJ z axis fits for 5 beads from each

Lecture 20: Optical Tools for MEMS Imaging

EXC500p-- PATHOLOGY MICROSCOPE. EXC500hd -- HD DIGITAL PATHOLOGY MICROSCOPE. EXC500r -- RESEARCH MICROSCOPE EXC500-LABORATORY SCOPE

Fast, high-contrast imaging of animal development with scanned light sheet based structured-illumination microscopy

Accessories Handbook 2010 for NanoWizard, ForceRobot and CellHesion

DIC Imaging using Laser Scanning Microscopes (LSM) on Inverted Stands

Atomic Force Microscopes

SHORT GUIDE TO LASER MICRODISSECTION USING THE PALM COMBI SYSTEM

Life Science Instrumentation. New Generation. Light Sheet Fluorescence Microscope. Alph

Bringing Answers to the Surface

Advanced Nanoscale Metrology with AFM

Fast Laser Raman Microscope RAMAN

Basic methods in imaging of micro and nano structures with atomic force microscopy (AFM)

1. Editorial. N 9 June Content

Application Note #548 AcuityXR Technology Significantly Enhances Lateral Resolution of White-Light Optical Profilers

By: Louise Brown, PhD, Advanced Engineered Materials Group, National Physical Laboratory.

MAKE SURE YOUR SLIDES ARE CLEAN (TOP & BOTTOM) BEFORE LOADING DO NOT LOAD SLIDES DURING SOFTWARE INITIALIZATION

Widefield-NikonEclipseTE200-ORCA Nikon Eclipse TE200 Inverted Microscope with Hamamatsu 1394 Orca-ER Cooled CCD Camera and Micromanager Software

WITec Alpha 300R Quick Operation Summary October 2018

Measurement of Surface Profile and Layer Cross-section with Wide Field of View and High Precision

Scanning Ion Conductance Microscope ICnano

Leica DMi8A Quick Guide

Characterization Microscope Nikon LV150

Instructions for easyscan Atomic Force Microscope

A New Profile Measurement Method for Thin Film Surface

ProScan DC Linear Servo Stage Technology

Using the Nikon TE2000 Inverted Microscope

Nikon. King s College London. Imaging Centre. N-SIM guide NIKON IMAGING KING S COLLEGE LONDON

Opterra II Multipoint Scanning Confocal Microscope. Innovation with Integrity

Training Guide for Carl Zeiss AxioZoom V16 Stereo Microscope

NSOM (SNOM) Overview

LSM 780 Confocal Microscope Standard Operation Protocol

attocube systems Probe Stations for Extreme Environments CRYOGENIC PROBE STATION fundamentals principles of cryogenic probe stations

Development of a High-speed Super-resolution Confocal Scanner

EPIFLUORESCENCE &/OR BRIGHTFIELD MICROSCOPY

Investigate in magnetic micro and nano structures by Magnetic Force Microscopy (MFM)

Before you start, make sure that you have a properly calibrated system to obtain high-quality images.

3. are adherent cells (ie. cells in suspension are too far away from the coverslip)

Nature Methods: doi: /nmeth Supplementary Figure 1. sospim principle and representation of the sospim beam-steering unit.

Fastest high definition Raman imaging. Fastest Laser Raman Microscope RAMAN

Cell Biology and Bioimaging Core

Supplementary Information. Stochastic Optical Reconstruction Microscopy Imaging of Microtubule Arrays in Intact Arabidopsis thaliana Seedling Roots

Study of shear force as a distance regulation mechanism for scanning near-field optical microscopy

ezafm OPERATING MANUAL

FLUORESCENCE MICROSCOPY. Matyas Molnar and Dirk Pacholsky

LOW TEMPERATURE STM/AFM

Comparing FCS and FRAP as methodologies for calculating diffusion

Transcription:

HybridStage - Automated, large sample-area mapping made easy Motivation Crucial parameters that affect cell adhesion, morphogenesis, cell differentiation and cancer invasion include the molecular interactions between cells and their extracellular matrix environment, their 3D topography and the corresponding mechanical properties. [1-3]. AFM (atomic force microscopy) is an advanced multiparametric imaging technique which delivers 3D profiles of the surfaces of molecules and cells in the nm-range. It also enables the characterization of nanomechanical properties (adhesion, elasticity etc.) and the visualization of structural changes taking place at the molecular level. Technical solution The JPK BioAFM Center has developed a highly versatile solution: HybridStage. It is equipped with three scanner units that can move either the tip or the sample. A standard NanoWizard AFM head has an xyz tip scanner with a scan range of 100 x 100 x 15 µm 3. The HybridStage houses an xyz sample scanner with an optional scan range of either 100 x 100 x 100 µm 3, 200 x 200 x 200 µm 3 or 300 x 300 x 300 µm 3, which can be individually selected depending on the individual application. It also features a motorized unit for large sample movements (20 x 20 mm 2 ). The wide choice of scanners, together with optical tiling and multi-region AFM probing (figures 6, 7), enables multi-parametric characterization of soft samples over a large area, and provides additional optical data sets on inverted and upright microscopes respectively (figures 1, 2). Fig. 1: JPK HybridStage integrated into a NanoWizard system on an inverted Nikon microscope. The mapping of tissue samples and thicker multi-cellular layers as well as the nanomechanical characterization of the extracellular matrix and its embedded cells requires a (semi)automated AFM approach for a millimeter range in x,y with nm resolution. Large variations in the topography of tissue samples, combined with fluidity and stickiness of membrane and extracellular proteins, require a larger flexibility in the z-axis of the AFM [3]. Fig. 2: JPK HybridStage on a Zeiss upright AxioZoom V.16 fluorescence macroscope with an Andor EMCCD camera for force mapping experiments on non-transparent tissue samples. page 1/5

Single Cell Force Spectroscopy and membrane tethers The most frequent activities in the field of Single Cell Force Spectroscopy (SCFS) are the study of the adhesion of individual cells on (bio)materials, extracellular matrix proteins and the characterization of cell-cell interactions [4]. Due to the effect of membrane tethering, the pulling length required to separate cells from a substrate frequently exceeds several 10 ths of microns [5]. The flexible, modular design of the HybridStage, with 3D sample scanners that can scan in the range of 100 x 100 x 100 µm 3, 200 x 200 x 200 µm 3 or 300 x 300 x 300 µm 3 can meet this demand. Optical focus control with JPK SlimFocus The consequence of a long-ranged z-movement for the tip and sample scanner is that either the cantilever or the sample move out of the optical focus. Therefore, we have implemented the JPK SlimFocus (Figure 4) to synchronize the objective lens motion with the current z- scanner (figure 5). The JPK SlimFocus control software allows manual adjustment of the z-position using a slider, and the setting of two marker positions, e. g. for the tip/sample. During AFM experiments, the optical focus motion can be synchronized with the z-scanner currently being used up to 150 µm. Figure 3 illustrates a detachment curve of a fibronectin functionalized cantilever from a Vero cell. After reaching the cell surface at a piezo height of approx. 60 µm, separation is completed at a final piezo height of 105 µm. The membrane tube formations, with force plateaus of 1-15 µm in length, are the main reason for a total separation distance of about 45 µm in this example. Fig. 4: JPK SlimFocus unit equipped with an objective lens mounted on a microscope turret. The sample stage is above the lens. Fig. 3: Force distance curve of a Vero cell (insert right: with an optical microscope image of Vero cells layer) in contact with a fibronectin coated tipless cantilever (50 µg/ml for 30 min; TL-2 from NanoWorld). Force settings: 3D piezo scanner 100 x 100 x 100 µm 3, 1 nn, z-speed 10 µm/s, contact 10 sec. The graph of the force curve (lower right) is the zoom of data in the black box. Fig. 5: SlimFocus software takes into account the height corrections needed if the objective lens and sample are surrounded by different media (liquid/air) [6]. page 2/5

Optical Tiling and Multi-Region Imaging If the x,y piezo scan range and the optical observation window (typically 100 µm) are much smaller than the sample size (in the range of mm) to be investigated with the AFM, a conflict arises. In order to enlarge the microscopic optical overview, the motorized sample scanner executes a repetitive pattern, acquiring the corresponding optical images of the sample which are then tiled and displayed in the SPM software (Figure 6). When tiling starts, the file chooser opens an automatically named sub-folder. The calibration file and the image file are saved in the sub-folder and are easily accessible for data processing. microscope can be seen in Figure 7. First, DirectOverlay was performed to correlate the cantilever tip position with the corresponding optical image. Second, optical tiling with 4 x 4 images on 500 x 500 µm 2 was carried out (see figure 6) to increase the region of interest. Third, a multiscan range was selected. For this specific example, 4 x 4 QI scans, with an individual map size of 50 µm, were performed (frames are labeled in magenta). In Figure 7, 13 of the 16 maps have already been executed. The green frame on position 14 indicates the current map position. Using the advanced force oscilloscope, multiple data viewers can be activated, e. g. height (of the sample) at force setpoint, adhesion (between cantilever and sample), slope (as an indication of sample stiffness) and reference force height(s) at a defined value of the force setpoint. In the specific example shown in Figure 7, two different data viewer channels are displayed (height at 100% of the setpoint and reference height at 80% of the setpoint). To obtain a better contrast of cellular structures, the reference height is indicated as a pixel difference image. Fig. 6: Screenshot of the NanoWizard control software. Right: Live camera image with Vero cells in liquid on an inverted microscope with optical phase contrast. The shadow of the retracted cantilever is visible. Left: The first optical image is calibrated in x,y with piezo accuracy using the AFM scanner (scan range marked in green, here 100 x 100 µm 2 ). A Tiling Tool button allows the user draw a tiling rectangle on the surface. The software automatically considers the number of optical images required after selecting the tiling area (here 4 x4 images for 500 x 500 µm 2 ). For multi-region AFM imaging, the user will be asked whether they wish to examine more than one measurement position. They can then select points of interest before a measurement starts, which will then be displayed in the data viewer with tiled optical images. When the measurement begins, the positions are processed sequentially. Different types of measurement definitions are available: single points, lines and rectangular regions. The multi-scan option is available for most measurement modes like force mapping and spectroscopy, AFM imaging, QI TM and multiple DirectOverlay TM snapshots. An example of multi-scan QI imaging of mammalian Vero cells on an inverted optical Fig. 7: Screenshot of the NanoWizard HybridStage control software. Adherent Vero cells in phase contrast optical microscope, imaged in PBS buffer, under 37 C temperature control in a JPK PetriDishHeater. Top: Optical tiling over 500 x 500 µm 2 consists of 4 x 4 optical images. The selected multi-scan region is indicated by the magenta colored frame: 4 x 4 maps with 50 x 50 µm 2 size. Top left: Height channel maps. Top right: Pixel difference map of the same region. Bottom: Advanced force oscilloscope and a force curve for a specific index position. QI mode settings: Force 0.23 nn, z- length 300 nm, 256 x 256 pixels. A PFQNM-LC-A cantilever with a tip height of approx. 19 µm and blunt tip (radius 70 nm) was used. page 3/5

Multi-Scan Tissue Mapping In order to obtain a better understanding of the development and progression of cancer, analyzing changes in the mechanical properties of cells and their surrounding extracellular matrix material appears to be crucial [3]. Using the example of human cervix tissue, the benefits of the HybridStage and the mode of operation, in combination with an upright macroscope, are presented here. apparent Young s modulus of the multi-scan shown in Figure 9. Fig. 8: Zeiss upright AxioZoom V.16 (lens 0.5x) fluorescence macroscope was used for the overview transmission light image (left) and fluorescence image (right) of human cervix tumor tissue. Fluorescence staining was performed with HOECHST for the nucleus labelling. The white frames illustrate the 1000 x 1000 µm 2 AFM mapping area. The optical images were done without an AFM head to obtain better optical quality for the fluorescence image. Thanks to its long working distance, the upright macroscope allows the collection of optical images with or without the AFM head. Furthermore, replacement of the AFM head does not require a readjustment of the sample. The excellent replacement accuracy of the head with a mounted cantilever is in the range of < 2µm in x,y,z. Once the region of interest is found in the optical image, it can be used to position the AFM probe with respect to the area of interest (Figure 8). A HybridStage with a 3D piezo sample scanner of 300 x 300 x 300 µm 3 in conjunction with the motorized stage was used to perform a multi-scan map of 1000 x 1000 µm 2 with 5 x 5 single maps of 200 x 200 µm 2 and a pixel distance of 10 µm each (Figure 9). All individual maps were analyzed using the batch processing option for indentation experiments in the JPK data processing (DP) software. Alternatively, the data can be exported, analyzed and composed using a third-party software e. g. Matlab. Figure 10 summarizes the frequency distribution of the Fig. 9: Overview of apparent Young s modulus multi-scan map for 1000 x 1000 µm 2 (compare figure 8) with 5 x5 individual force maps of 200 x 200 µm 2. The single force maps were analyzed using the NanoWizard DP software. The indentation experiments were performed at 3 nn setpoint force with a Nanoworld-Cont cantilever with a 6 µm bead which was separately glued at the tip apex. When performing a multi-scan force map, over a larger area on a very rough sample with varying heights, it may e. g. i) damage the cantilever, or ii) the AFM measurements may not yield usable data. Furthermore, in very soft samples with sticky surfaces, the cantilever only separates from the sample surface after several 10 th of micrometers, resulting in noticeable deformation of the sample or removal of the membrane tube. Fig. 10: Histogram of the apparent Young s modulus of the data displayed in Fig. 9 with a unimodal frequency distribution. A typical indentation depth at 3 nn setpoint force was around 1µm 3 µm. page 4/5

Here too, the z-range must be increased in order to successfully complete the measurement. To solve this technical difficulty, the NanoWizard software can set the height of both the piezo-retract and the motor-retract. In this way, an individual map and entire scan can be optimally performed. Figure 11 shows that the tissue sample examined is very rough and has a very varied height. Even in the individual map of 200 x 200 µm 2, the height difference is approx. 80 µm. Despite this, the multi-scan could still stably perform the measurement without having to stop. Conclusion The HybridStage is a versatile tool for advanced, multiparametric AFM characterization of samples in the range of mm to nm and pn resolution, which features three scanner units that can move the tip/sample. A wide range of samples can be studied, including biomaterials, cells and microaggregates, embryos and tissues, model organisms in developmental biology (zebra-fish, C. elegans, etc) and implants. Particularly the increased z- piezo range, equipped with wide-ranged motorized sample motion, is perfect for the nanomechanical characterization of tissues required in cancer, nanomedicine or developmental biology. Fig. 11: Height profile of a single map of 200 x 200 µm 2. of the cervix tissue (map from a top right corner of the multi-scan area). HybridStage Compatibilities Optical microscope: Upright o Zeiss Axio Zoom.V16 o Leica MF205 o Olympus MVX 10 with epi-fluorescence and brightfield Optical microscope: Inverted o Zeiss AxioObserver o Nikon TE, Ti and Ti2 lines o Leica DMI 6000/DMi8 o Olympus IX71 with epi-fluorescence and brightfield The HybridStage frees experiments from the constraints of the AFM piezo range. Large-range tiling of optical images provides a clear visual overview, allowing a fast setup of optically guided experiments and direct selection of the optical features for investigation. Navigate around the sample, collect a list of interesting features for multiscan, or even map force responses over greatly extended scan ranges. The HybridStage is a modular, piezo-based sample scanner stage combined with motorized XY sample movement, giving direct access to anything you can see. The highly versatile solution for all of your sample requirements. Sample courtesy The data for tissue characterization shown in Figures 2 and 8-11 are kindly provided by Prof. J. Käs and Dr. Th. Fuhs (University Leipzig). The Vero cell samples were provided by Prof. A. Hermann and his group (Humboldt university Berlin). Literature [1] Engler AJ. et al. Cell 2006; 126:677-689 [2] Barriga EH. et al. Nature 2018; 554: 525-527 [3] Plodinec M. et al. Nature nanotechnol. 2012; 7: 757 765 [4] Alsteens D. et al. Nature reviews/materials 2017; 2/17008: 1-16 [5] Friedrichs J. et al. Nature protocol 2010; 5/5: 1353-1361 [6] Beyer H., Riesenberg H. Handbuch der Mikroskopie 1988; ISNB 3-341-00283-9 page 5/5