Potential Risks of MRI in Device Patients

Similar documents
MRI SYSTEM COMPONENTS Module One

MRI with the VNS Therapy System October 2017

(N)MR Imaging. Lab Course Script. FMP PhD Autumn School. Location: C81, MRI Lab B0.03 (basement) Instructor: Leif Schröder. Date: November 3rd, 2010

High Field MRI: Technology, Applications, Safety, and Limitations

Human Subjects in fmri Research

Medical Imaging. X-rays, CT/CAT scans, Ultrasound, Magnetic Resonance Imaging

Background (~EE369B)

Magnetic Resonance Imaging and Radio Frequency. Part 1. Produced on behalf of Mid Sussex Amateur Radio Society by M5BTB

MAGNETIC RESONANCE IMAGING

Radio Frequency Field

Acoustic noise reduction of MRI systems by means of magnetic shielding

Cardiac MR. Dr John Ridgway. Leeds Teaching Hospitals NHS Trust, UK

M R I Physics Course. Jerry Allison Ph.D., Chris Wright B.S., Tom Lavin B.S., Nathan Yanasak Ph.D. Department of Radiology Medical College of Georgia

Analysis of magnetic and electromagnetic field emissions produced by a MRI device

Cook Medical Peripheral Intervention MRI Information

2015 Spin echoes and projection imaging

BME 3113, Dept. of BME Lecture on Introduction to Biosignal Processing

Fundamentals of Pacing Therapy

Cerclage Passer. For minimally invasive application of cerclage cables.

SQUID - Superconducting QUantum Interference Device. Introduction History Operation Applications

Magnetic Resonance. Intera 1.5T Release 2.5 series Achieva 1.5T / 3.0T / XR Release 2.5 Panorama HFO Release 2.5 series. Technical Description

Hardware. MRI System. MRI system Multicoil Microstrip. Part1

1 Introduction. 2 The basic principles of NMR

Fast Field-Cycling Magnetic Resonance Imaging (FFC-MRI)

HAZARDS OF NON-IONIZING RADIOFREQUENCY (RF) RADIATION

MRI RF-Coils. Innovation with Integrity. Highest sensitivity for your preclinical MRI and MRS applications. Preclinical Imaging

MR Advance Techniques. Flow Phenomena. Class II

Technical Description. English. Release 5. Achieva

Field Simulation Software to Improve Magnetic Resonance Imaging

Azure XT SR MRI SureScan W1SR01

HCS 1.5. The countersinkable compression screw.

Magnetic Resonance Imaging Principles, Methods, and Techniques

MRI Metal Artifact Reduction

3T Unlimited. ipat on MAGNETOM Allegra The Importance of ipat at 3T. medical

MRI Systems and Coil Technology

Application Note # 5438

MRI Systems and Coil Technology

Evaluation of MWT Materials Accusorb MRI Shield. Dr. E. Kanal, Department of Radiology, UPMC. Wednesday, April 21, 2010

Receive Arrays and Circuitry

The physics of ultrasound. Dr Graeme Taylor Guy s & St Thomas NHS Trust

Magnetic Resonance Imaging

HOW TO CHOOSE AN ANTENNA RANGE CONFIGURATION

In an unmagnetized piece of iron, the atoms are arranged in domains. In each domain the atoms are aligned, but the domains themselves are random.

High frequency electomagnetic field irradiation. Andrea Contin

MARP. MR Accreditation Program Quality Control Beyond Just the Scans and Measurements July 2005

6 Appendix SCR Recording During FMRI Acquisition

Correspondence should be addressed to Takahiko Yamamoto,

Area Network Applications] Notice: This document has been prepared to assist the IEEE P It is

HETERONUCLEAR IMAGING. Topics to be Discussed:

RF Amplifiers and RF Heating for Medical Applications FTF-IND-F0690

Investigation of the Factors Responsible for Burns During MRI

int Laboratory of Integrative Neuroimaging Technology

MRI SAFETY CONSIDERATIONS AN INCIDENT STAGLIN CENTER FOR COGNITIVE NEUROSCIENCE AN INCIDENT

This is a preview - click here to buy the full publication. Exposure assessment methods for wireless power transfer systems

COMPUTED TOMOGRAPHY 1

2 Hardware for Magnetic Resonance Imaging

5. Transducers Definition and General Concept of Transducer Classification of Transducers

Analysis of spatial dependence of acoustic noise transfer function in magnetic resonance imaging

Figure 1 The switched beam forming network.

Designing an MR compatible Time of Flight PET Detector Floris Jansen, PhD, Chief Engineer GE Healthcare

SJM MRI Activator. Handheld Device. User's Manual. Model EX4000

H 2 O and fat imaging

Introduction to MR Hardware. RF Coils C M L C T. = g * B 0. Rotating magnetization produces alternating magnetic field

Si-Hall ICs. 1 Application Notes for Differential Hall IC s

New developments in near-field acoustic holography

Variable Angle LCP Mesh Plate 2.4/2.7. Part of the Variable Angle LCP Forefoot/Midfoot System 2.4/2.7.

Health Issues. Introduction. Ionizing vs. Non-Ionizing Radiation. Health Issues 18.1

Microwave Cancer Therapy

Pulse Sequence Design and Image Procedures

EnSite Precision Cardiac Mapping System ACCURACY OF THE ENSITE NAVX NAVIGATION AND VISUALIZATION TECHNOLOGY, SENSOR ENABLED

Weber State University Radiologic Technology 4603

ABSOLUTE MAGNETIC ROTARY ENCODER ANALOG

RAD 229: MRI Signals and Sequences

Electromagnetic Field Exposure Feature of a High Resonant Wireless Power Transfer System in Each Mode

Product Safety and RF Energy Exposure Booklet for Portable Two-Way Radios

Magnetic Resonance Research Facility (MRRF) Resources

Silicon Hall ICs. 6.6 Application Notes: Differential Hall IC TLE U

LUXONDES. See the electromagnetic waves. Product 2018 / 19

Regulatory Guidance and Safety Standards

Product Safety and RF Energy Exposure Booklet for Unication Two-Way Portable Radios

Technical Report

Physics of RFID. Pawel Waszczur McMaster RFID Applications Lab McMaster University

Electronic Speed Controls and RC Motors

2.4 mm Cannulated Screw. An integral part of the Synthes Cannulated Screw System (CSS).

The SENSE Ghost: Field-of-View Restrictions for SENSE Imaging

Electromagnetic compatibility - sensibility of implantable

The promise of high-field MRI. High Field MRI Technology, Applications, Safety, and Limitations. High-field Scanners

SIGNA Pioneer: Ultra High Efficiency Gradient System Advancing the gradient technology curve

Spiral MRI on a 9.4T Vertical-bore Superconducting Magnet Using Unshielded and Self-shielded Gradient Coils

Radiofrequency: Principles & Devices

磁振影像學 MRI 磁振假影與磁振安全 磁振假影. 本週課程內容 Hardware-related Artifacts 盧家鋒助理教授 磁振假影 磁振安全

EE225E/BIOE265 Spring 2011 Principles of MRI. Assignment 6 Solutions. (y 0 + vt) dt. 2 y 0T + 3 )

MagPro coils. Versatility in Magnetic Stimulation. For Clinical and Research use

Lab 8 6.S02 Spring 2013 MRI Projection Imaging

IEEE Electromagnetic Compatibility Standards (Active & Archive) Collection: VuSpec

TECHNICAL PAPER. Smarter Sensors reduce costs for Motion Control Integrators. David Edeal. Introduction

A Closer Look. LATITUDE NXT Alerts SUMMARY. Alerts. Red Alerts

CHARACTERISTICS, DOSIMETRY & MEASUREMENT OF EMF

Chapter 21. Alternating Current Circuits and Electromagnetic Waves

LCP Pilon Plate 2.7/3.5

Transcription:

Outline Potential Risks of MRI in Device Patients Redha Boubertakh r.boubertakh@qmul.ac.uk MRI and cardiac implantable electronic devices (CIED) Components of an MRI scanner MRI implant and device safety classification Risks associated with scanning CIEDs Main magnetic field Radiofrequency waves Time-varying magnetic field gradients Summary & Conclusion Barts Health NHS Trust Queen Mary University MRI and Cardiac Implantable Electronic Devices Number of patients fitted with cardiac implantable electronic devices (CIED) is growing Large percentage will require an MRI scan over their lifetime In the past, CIEDs have been considered an absolute contraindication to MRI MRI and Cardiac Implantable Electronic Devices However, recent studies have shown that, if strict screening, safety and monitoring procedures are followed, legacy non MR Conditional CIEDs can also be scanned with minimal risks to the patient New England Journal of Medicine, 207 New England Journal of Medicine, 207 Heart Rhythm, 207 Heart, 20 General MRI Safety Risks Components of an MRI Scanner Source Static main magnetic field B0 Time-varying magnetic field gradients Radiofrequency (RF) pulses Transmit coil Body or T/R local coil Perturbed net spin magnetisation Magnetic fields ~ 0,000-00 000 stronger than the earth s Always on! Linear magnetic field gradient are used to encode the MR signal Receive coil Body or local coil Safety risks Static Magnetic field B 0 field (T,.T, 3T, ) - Projectile or missile effect - Displacement and torque effects - Device disruption - Bioeffects Radiofrequency (RF) pulses B field Amplitude~ μt, MHz frequency range - Tissue heating, Burns - Medical device heating and disruption - Interference with equipment (monitoring systems) and devices Time varying Gradient Magnetic Field G x,y,z gradients. Strength (mt/m) and slew rate (mt/m/ms) scanner dependent - Peripheral nerve stimulation (PNS) - Acoustic noise - Interference with equipments and devices Cryogens Liquid Helium at -269 oc (4 K) - Burns - Asphyxia - Hypothermia Gadolinium Based Contrast Agents -Nephrogenic Systemic Fibrosis (NFS) Affect CIEDs

4 3 2 0 2 3 4 4 3 2 0 2 3 4 Z Axis (m) 4 3 2 0 2 3 4 4 3 2 0 2 3 4 Z Axis (m).8.6.4.2 0.8 0.6 0.4 0.2 0. T. T.9 T T 0.2 0 0.2 0.4 0.6 0.8.2.4.6.8 2 Z Axis (m) Side view Patient table MRI Safety Terminology for Implants and Devices MRI Safety Terminology for Implants and Devices Any device/implant falls into one of these three categories No CIED is MR Safe An MR Conditional device can be safely scanned if conditions defined by the manufacturer are met The development of MR Conditional CIEDs (2008) has made MRI scans safe to use Condition of use include: The type of device Generator + Leads type Programming modes and parameters, Time since implantation, The MRI environment Type of magnet, maximum value of field strength Spatial magnetic field gradient Time varying magnetic field gradients Type of imaging sequence, induced heating Imaged body part Type of imaging coils used, Main Magnetic Field Potential Risks of CIEDs in the MRI Environment The static magnetic field B 0 (mainly.t or 3T on clinical scanners) is one of the main sources of danger Strong attractive force exerted on ferromagnetic objects Projectile effect Additionally, a device can experience a torque (rotational forces) to align it with the direction of B 0 Associated Risks: Device motion, vibration and displacement Magnetic field B0 Static Field Characteristics Main Magnetic Field Attractive forces depend on the ferromagnetic content of the device High ferromagnetic content higher risk An important parameter is the spatial magnetic field gradient Rate of change of the static magnetic field within and around the magnet Siemens scanner Vertical Axis (m) 200 mt 40 mt 20 mt 0 mt mt 3 mt mt 0. mt Horizontal Axis (m) 200 mt 40 mt 20 mt 0 mt mt 3 mt mt 0. mt Vertical (m) Attractive forces are proportional to the spatial gradient of the static field Force highest at the bore entrance However, torque is proportional to the static field strength Torque largest at the centre of the magnet bore Both forces depend on the ferromagnetic/paramagnetic content of the device MR conditional devices are designed with reduced ferromagnetic content Stainless steel and titanium alloys used 0. T B0 field - Side view B0 field - Top view Spatial field gradient - Side view 2

Main Magnetic Field Main Magnetic Field Conditions Even for non MR Conditional pacemakers, these risks are less of an issue for device post 2000 due to low ferromagnetic content Subcutaneous tissue fibrosis around the device Luechinger, R et al, PACE 200 207 HRS expert consensus statement, Heart Rhythm Reduction in device size throughout the years has sensibly diminished potential risks St. Jude Medical Field strength and spatial magnetic field gradient are specified in the conditions of use of MR Conditional devices Example: Horizontal cylindrical bore magnet, clinical MRI systems with a static magnetic field of. Tesla (T) must be used Maximum spatial magnetic field gradient of 70 G/cm Imaging Exclusion Zones Reed Switch Malfunction An MR Conditional CIED does not always allow all body parts to be scanned MRI conditions usually also include possible Exclusion Zones If an exclusion zone of an MR Conditional device is imaged Device is used off-label Follow non MR Conditional safety protocols Strong magnetic fields can result in device malfunction For non MR Conditional devices, reed switches have been shown to malfunction A reed switch is used to program devices A small magnetic field is used to change device mode to asynchronous pacing Reed Switch Malfunction Time-Varying Magnetic Field Gradients Switch behaviour has been shown to be unpredictable depending on orientation Unexpected switch opening or closure Scheidegger, M et al, PACE 2002 Electronic solid-state Hall effect sensors have improved reliability Behave in a predictable manner MRI sequences use magnetic gradients to spatially encode the imaged object Important scanner parameters: Slew rate (up to 200 T/m/s) Maximum gradient strength (such as 4 mt/m) Max strength RF pulses Slice gradient Phase encoding gradient Frequency encoding gradient 40 o TR TR TE Rise time 3

Time-varying Magnetic Field Gradients Effects of EMI Rapidly switching gradients can induce Electromagnetic interference (EMI) Electrical currents and voltages can be induced in conduction wires Device leads Generator Interferences could be interpreted by a CIED as a real or missing heart rhythm signal (oversensing, undersensing). EMIs can lead to: Therapy inhibition A pacemaker may be withhold pacing Pacing dependent patients Innapropriate shocks An ICD may interpret an interference as requiring an unnecessary shock Effects of EMI Effects of EMI High-frequency noise and noise response ventricular pacing are noted on the bipolar ventricular channel from a single-chamber pacemaker during MRI Figure. Artefact mimicking narrow complex tachycardia EMI can cause a power-on-reset The device programming parameters go back to the factory default Pacing could be inhibited Activation of antitachycardia therapy Device battery status and longevity affected Battery drain may also occur Beinart R et al, Circulation 203 Gimbel J et al, Annals of Noninvasive Electrocardiology 200 Preventing EMI Radiofrequency (RF) Pulses Electromagnetic interference can be reduced by filtering the ECG signal In MR Conditional devices, generators and circuitry are shielded to minimize the effects of interferences Better protection of the power supply Intracardiac ECG signals in a canine model inside an MR scanner Very short RF pulses are used to disturb/tip the net spin magnetization within a tissue RF pulse frequency is matched to the imaged nuclei ( H) at a given static magnetic field ~ 64 MHz at.t ~ 28 MHz at 3T The body will absorb some of this energy Resistive heating Heating generated by a sequence is measured by the Specific Absorption Rate (SAR) in Watt/kg Nazarian S et al, Circulation 2008 4

SAR MODE for CIEDs Leads Heating Effects Leads can act as antennae and concentrate RF energy Lead lengths (40 60 cm) comparable to MRI RF wavelengths Effect strongly associated to ratio of lead length to RF wavelength and presence of loops and lead geometry Normal Mode ( 2 W/kg) for all CIEDs MHRA, Safety Guidelines for Magnetic Resonance Imaging Equipment in Clinical Use, March 20 High electrical currents can be induced heating Resistive effect Leads Heating Effects Leads Heating Effects Potential risk of thermal injury by ohmic loss in myocardial tissue around the tip Generated currents may lead to: Myocardial stimulation Temporary or permanent changes in impedance and thresholds Device malfunction and damage Abandoned and fractured leads, broken lead tips and lead loop configurations may increase heating effects Epicardial lead tips not cooled by blood flow Figure. For pacemaker leads 20 60 cm in length the RF-induced pacemaker lead tip heating is shown for pacemaker-attached and abandoned capped pacemaker leads. Pacemaker leads that are abandoned and capped heat significantly more than pacemaker-attached leads for clinically relevant lead lengths (40 60 cm). Langman D et al, PACE 20 Reducing Lead Heating Effects Summary of Potential Risks Limit the RF power used during imaging Reduce SAR level Changes in lead design: Improvements to lead inner- and outer-coils structure Lead tip coating with polarization resistant material Use of heat-dissipating filters Force, Torque Patient discomfort, surrounding tissue damage, device malfunction Vibration Patient discomfort, device malfunction Induced currents/voltages Induced VT, arrhythmia, pacing inhibition Heating Tissue damage, impedance and threshold changes, loss of sensing and/or pacing capture Device malfunction Device reset, mode changes, loss of therapy, patient shocks Static Magnetic field B0 Time-varying gradients RF pulses

Scanning Conditions Unknown. Could be MR unsafe. Conclusion Conclusion Scanning of MR Conditional cardiac devices is safe if specified conditions are followed Relatively straightforward Can be done at any MRI general imaging unit No reason to deny a patient an MRI examination Strong evidence from registries and clinical studies show that non MR Conditional CIEDs can be scanned safely if strict clinical and scanning protocols are observed Specialized centres There still exists a confusion about the used terminology and determining scanning conditions Spatial static field gradient not always easy to check MR safety labelling MR? Non MR Conditional? 6