Amray 3600 FESEM. Standard Operating Procedure. v2.2 modified by Bryan Cord. General Notes...3. Sample Loading...5. System Loading...

Similar documents
Standard Operating Procedure for the Amray 1810 Scanning Electron Microscope Version: 29 NOVEMBER 2014

JEOL 6500 User Manual

Using the Hitachi 3400-N VP-SEM

1. Preliminary sample preparation

Title: Amray 1830 SEM#2 Semiconductor & Microsystems Fabrication Laboratory Revision: D Rev Date: 03/18/2016

RAITH e-line OPERATING INSTRUCTIONS

JEOL 6700 User Manual 05/18/2009

Scanning Electron Microscope FEI INSPECT F50. Step by step operation manual

Basic Operating Instructions for Strata Dual Beam 235 FIB/SEM

SEM Training Notebook

Figure 1 The Raith 150 TWO

SEM Training Notebook

Check that the pneumatic hose is disconnected!!!! (unless your using the BSE detector, of course)

1.1. Log on to the TUMI system (you cannot proceed further until this is done).

SEM OPERATION IN LOW VACUUM MODE

Instructions for Tecnai a brief start up manual

FE-SEM SU-8020 Operating manual (Preliminary version)

JSM 6060 LV SCANNING ELECTRON MICROSCOPE STANDARD OPERATING PROCEDURES

The ideal K-12 science microscope solution. User Guide. for use with the Nova5000

Dickinson College Department of Geology

Basic Users Manual for Tecnai-F20 TEM

Installation guide. Activate. Install your Broadband. Install your Phone. Install your TV. 1 min. 30 mins

SOP for Hitachi S-2150 Scanning Electron Microscope For review purposes only

OPERATION OF THE HITACHI S-450 SCANNING ELECTRON MICROSCOPE. by Doug Bray Department of Biological Sciences University of Lethbridge

FEI Quanta 200 ESEM Basic instructions

Operating Checklist for using the Scanning Electron. Microscope, JEOL JSM 6400.

Operating Checklist for using the Scanning Electron Microscope, JEOL JSM 6400.

FEI Tecnai G 2 F20 Operating Procedures

Horiba LabRAM ARAMIS Raman Spectrometer Revision /28/2016 Page 1 of 11. Horiba Jobin-Yvon LabRAM Aramis - Raman Spectrometer

Model SU3500 Scanning Electron Microscope

PANalytical X pert Pro Gazing Incidence X-ray Reflectivity User Manual (Version: )

Contents STARTUP MICROSCOPE CONTROLS CAMERA CONTROLS SOFTWARE CONTROLS EXPOSURE AND CONTRAST MONOCHROME IMAGE HANDLING

Installation guide. Activate. Install your TV. Uninstall. 1 min 10 mins. 30 mins

1.3. Before loading the holder into the TEM, make sure the X tilt is set to zero and the goniometer locked in place (this will make loading easier).

This document assumes the user is already familiar with basic operation of the instrument in TEM mode and use of the Microscope Control interface.

LEO 912 TEM Short Manual. Prepared/copyrighted by RH Berg Danforth Plant Science Center

General information. If you see the instrument turned off, notify MIC personnel. MIC personnel will help you insert your samples into the instrument.

1.2. Make sure the viewing screen is covered (exposure to liquid N 2 may cause it to crack).

Bruker Dimension Icon AFM Quick User s Guide

Nanoscale Fabrication & Characterization Facility. Raith e-line EBL Users Guide (updated:aug 2 nd, 2017)

This procedure assumes the user is already familiar with basic operation of the SEM and the MiraTC interface.

Standard Operating Procedure of Atomic Force Microscope (Anasys afm+)

Procedures for Performing Cryoelectron Microscopy on the FEI Sphera Microscope

2. Raise HT to 200kVby following the procedure explained in 1.6.

Vinyl Cutter Instruction Manual

CNC Using the FlexiCam CNC and HMI Software. Guldbergsgade 29N, P0 E: T:

SYNGUIDER USER'S MANUAL

Operating the Hitachi 7100 Transmission Electron Microscope Electron Microscopy Core, University of Utah

Game Making Workshop on Scratch

ENSC 470/894 Lab 3 Version 6.0 (Nov. 19, 2015)

DISCO DICING SAW SOP. April 2014 INTRODUCTION

UNIVERSITY OF WATERLOO Physics 360/460 Experiment #2 ATOMIC FORCE MICROSCOPY

Full-screen mode Popup controls. Overview of the microscope user interface, TEM User Interface and TIA on the left and EDS on the right

CHAPTER1: QUICK START...3 CAMERA INSTALLATION... 3 SOFTWARE AND DRIVER INSTALLATION... 3 START TCAPTURE...4 TCAPTURE PARAMETER SETTINGS... 5 CHAPTER2:

Optika ISview. Image acquisition and processing software. Instruction Manual

05/20/14 1. Philips CM200T. Standby Condition

Protective Equipment Nitrile gloves for handling sample holder and safety glasses for filling liquid nitrogen dewar.

Strata DB235 FESEM FIB

ZEISS EVO SOP. May 2017 ELECTRON OPTICS

GXCapture 8.1 Instruction Manual

Stitching MetroPro Application

Fusion M2/32/40 focus, home and table calibration

ACCU-GOLD QUICK START MANUAL

AgilEye Manual Version 2.0 February 28, 2007

SOP-P051. Scanning of Optical Filters With USB2000. Objective: To determine the spectral transmittance properties of an optical filter.

Please follow these instructions for use of the Philips CM100 TEM. Adopted from website below.

JEOL 5DII. Operation introduction. By Serge Charlebois

R I T. Title: Wyko RST Plus. Semiconductor & Microsystems Fabrication Laboratory Revision: A Rev Date: 05/23/06 1 SCOPE 2 REFERENCE DOCUMENTS

Start Here. Installing your Microtek ScanMaker 9800XL Plus PC:

2020 DRAWBOT INSTALLATION AND USE. Robert Ashford Henry Arnold 4-H OABB

reference guide reference guide reference guide

OverDrive for Kindle, Kindle Paperwhite, Kindle Voyage, and Kindle Oasis (not Kindle Fire and Fire Tablet) Contents

User Operation of JEOL 1200 EX II

Horiba Jobin-Yvon LabRam Raman Confocal Microscope (GERB 120)

SWIFT SERIES M2252DGL MICROSCOPE

Leica DMi8A Quick Guide

ADD A REALISTIC WATER REFLECTION

Digital Portable Overhead Document Camera LV-1010

Sketch-Up Guide for Woodworkers

Bruker Dimension Icon AFM Quick User s Guide

Guide to AmScope SZM Series Dissecting Scope for Sample Imaging with MU Series Camera

Zygo Optical Interferometer SOP

Welcome 1. Precaution

Scratch for Beginners Workbook

CCD User s Guide SBIG ST7E CCD camera and Macintosh ibook control computer with Meade flip mirror assembly mounted on LX200

PANalytical X pert Pro High Resolution Specular and Rocking Curve Scans User Manual (Version: )

The operation manual of spotlight 300 IR microscope

Using the AmScope Microscope Cameras

Recording EPR spectra using the Loop Gap Resonator (LGR)

LumaSpec 800S User Manual

Visioneer OneTouch Scanner. Installation Guide FOR WINDOWS

STEM Spectrum Imaging Tutorial

so you want to get to know Onsrud... Onsrud1 : machine set up

CHROMACAL User Guide (v 1.1) User Guide

QUICK-START FOR UNIVERSAL VLS 4.6 LASER! FRESH 21 SEPTEMBER 2017

MSE 460 TEM Lab 2: Basic Alignment and Operation of Microscope

Installing Your New Creature From The Black Lagoon Tail Light DMD Panel MOD

ALTURA EDS. Rev. 0915

Service Manual for XLE/XLT Series Laser Engravers

WELCOME WHAT S IN THE BOX

Transcription:

Amray 3600 FESEM Standard Operating Procedure v2.2 modified 5.13.13 by Bryan Cord Contents General Notes...3 Sample Loading...5 System Loading...8 Imaging...12 Saving Data...16 System Unloading...18 Troubleshooting...20 1

2

General Notes This guide is intended as a quick reference for doing basic imaging on the Amray 3600. It assumes you re familiar with the system and is NOT a substitute for training. Contact Bryan Cord (bcord@umn.edu) for training if you need it. The system has a lot of advanced capabilities that aren t covered here. Fortunately, it also has a comprehensive help file. The help menu will give you a detailed explanation of every element in each of the different control windows. There s also a manual (linked on the desktop) that goes over everything in a more how-to format. Due to low-level incompatibility between Windows 7 and the 35-year-old SEM control hardware, moving the mouse from the Windows PC screen to the SEM display screen will cause the mouse to double across both screens. The mouse pointer on the PC screen is just a ghost and won t affect anything; move the pointer off the SEM display to get the Windows pointer back. Good SEM imaging takes practice, especially on this system. If you re not getting the resolution you feel like you should be getting, experiment with different settings or ask for help. 3

4

Sample Loading 1. For small pieces, mount your samples on one of the provided stubs using carbon tape discs. Please do not tape samples directly to the block! a. To make life easier, you may want to put some gold nanoparticles near the part of the sample you want to image to make it easier to locate in the SEM. Use the syringe in the toolbox to deposit ~5 µl of particles from the bottle on your sample. Make sure you put the syringe back in its protective tube when finished. b. If you re using nanoparticles, make sure you give the drop time to dry before pumping down the airlock! Even sucking up a small amount of water will shorten the lifetime of the airlock pump. 2. Place the stem of your sample stub in one of the holes on the raised block in the center of the chuck. Use the small allen wrench near the load lock to tighten the set screw that holds it in place. 5

3. For full wafers, remove the piece-part block from the chuck using the larger allen wrench in the toolbox and put it in a safe place. Clamp your wafer into the chuck using the spring clips. Be sure to replace the piece-part block when finished! 4. To view samples at an angle (for cross-sections and the like), use the provided angle chuck. Mount it on the wafer holder using the same two screws as the piece-part chuck. The arrows scribed on top of the block should point toward the SEM chamber when the chuck is attached properly. Attach your sample to either the 45 or 85 degree tilt using the copper clips. If the clips are in the way feel free to take some of them off while you re working as long as you put them back on when you finish. 6

7

System Loading 1. Slide the chuck, with your sample mounted on it, into the load lock per the picture below. 2. Close the load lock door and press the Load button on the nearby panel a. You can also use the Sample load command in the Vac menu on the SEM Control software panel to do this. If you re getting errors about stage referencing, this way usually works better. 3. If the SEM software isn t already on, start it using the Start SEM icon on the desktop. If the PC is logged out for some reason, use the username 3600 without a password to log 8

into it. 4. Wait a few minutes while the system evacuates the load lock and transfers the chuck into the chamber. When you see the gate valve door close on the camera display screen, the sample is loaded. 5. Load the correct holder profile for the holder with the File/Open command in the Stage Control window. If the piece-part block is attached, use 4in_stub.stg ; use 4inch.stg if you re imaging a full wafer and angle.stg if you re using the angle chuck. The holder profile is what keeps the stage from crashing into the final lens, so don t forget to do this! 6. Double-click on the stage map in SEM Control to move the stage to the approximate location of your sample. If using the piece-part stub, the crosshairs represent the different mounting sites along the stub. 7. Set the z-axis to the approximate working distance you d like to use (6-8mm is good) by dragging the line in the Z window of the Stage Control window up and down. Make sure you set the working distance in the SEM Control window too (the FL setting) so the system knows where to focus. 9

8. Set the operating voltage in the SEM Control window (the KV box) to the desired value. 5000V is fine for most imaging, but when imaging surfaces or thin features it may be useful to go to 1-2 kv. 9. Set the probe current by adjusting the C1 value in the SEM Control window. A higher (larger negative) current will give you better signal to noise but increase charging and contamination. -12 is a good starting point with this setting; you can go to -15 or higher if the image is too noisy, or -10 or even lower if charging is a major issue. 10. Open the column valve, either by pressing the column isolation valve button on the column controls or by double-clicking the blank box in the lower-right corner of the SEM Control window. 10

11

Imaging 1. To change magnification, use the six buttons on the lower-left corner of the console. You can also type new magnifications in SEM control or use the labeled function keys to get to pre-set magnification levels. Moving around can be accomplished with the joystick, the stage control software, or the beam shift knob. 2. Brightness and contrast can be adjusted using the knobs of the same name on the console. The auto video setting will automatically optimize brightness and contrast. 3. Focus can be changed using the large knob under the joystick on the console. Note that the autofocus/auto-stigmation capabilities don t really work well and are not recommended. 4. The easiest way to calibrate your imaging settings is to zoom out to the lowest mag and find an easily-visible garbage feature like a scratch, dust speck, etc. Try not to use the thing you re actually imaging to calibrate your settings, as charging and/or contamination could ruin it before you get a chance to actually image it. a. Moving around can be done with the joystick, the SEM display screen (right-click to change the mouse icon into a target, then double-click the target to bring the double-click site to the screen center), the Stage Control window, or the X and Y beam shift knobs. b. This is much easier to do with a fast scan speed. Turn off slow scan on the console and adjust the RS value to 1 or 2 in the SEM Control window (click on RS, then right-click to lower the value or left-click to raise it) for best results. 12

5. Once you ve found a feature to use for calibration, zoom in a bit and twist the focus knob to bring it into focus. Check the working distance (FL) setting in the SEM Control window and make sure it s approximately what you were looking for; drag the z-axis up or down accordingly (in the Stage Control window) if it isn t and refocus. 6. With the feature in focus, degauss the detector using the Degauss button in the SEM Control window. This will cause the image to jump out of focus. Refocus the image and degauss again. The detector is fully de-magnetized when degaussing doesn t affect the focus. Degaussing is critical if you re doing quantitative imaging, as a magnetized detector can cause spatial distortions in your image. 7. Zoom in on your calibration feature until you have problems getting it into focus. At this point, you ll need to align the aperture. Press the wobble button on the console (or use the equivalent software setting/hotkey). You should see the image wobbling in the x and/or y axes. Use the X and Y aperture alignment knobs on the column (see below) to adjust the aperture until the feature stops wobbling and just breathes in and out of focus in one place. Turn off the wobble when finished. 13

a. If adjusting the apertures is causing the screen to go black, the beam is out of alignment. Stop the focus wobble and twist the x and y axes of the beam alignment knob until the image is as bright as possible, then continue. b. Try not to lean on the column while adjusting the aperture knobs, as this will make aligning them even harder. 8. With the aperture aligned, continue to zoom in and focus until you re again having problems focusing on a feature. Now focus as much as possible and adjust the x and y stigmators until the image comes fully into focus. a. There are coarse and fine stigmator knobs on the console. For best results, move both fine stigmator knobs to the 12 o clock position and use the coarse stigmator knobs to get as close as possible to the optimum setting. Then use the fine knobs to fine-tune, as well as compensate for any stigmation drift that occurs while imaging. 14

b. Continue to refocus while adjusting stigmation. c. Using a reduced scan field (pressing the reduced field button on the console) will let you use a much smaller image to adjust stigmation. Since this image refreshes much more quickly than a full-screen image, you ll get more immediate visual feedback that can make your life much easier. You can adjust the sized and position of the reduced field using the mouse on the display monitor. 9. Aperture adjustment, focus, and stigmation are an iterative process. Continue to zoom in and adjust all three settings (using the focus wobble for aperture alignment) until you have a clear image at the desired magnification. 15

Saving Data 1. Once you have a good image with optimized settings, you ll want to save a high-quality scan of it. The best way to do this is to use a slow scan ( Slow Scan button on the console) setting between 1 and 4 (click SS then left/right click to raise and lower the scan speed in SEM Control). Press Freeze during the scan to freeze the scan at the end of the frame. a. It should be mentioned that the Photo button, which was originally connected to an old-timey film camera, doesn t do anything anymore. 2. You can add annotations if you want (see appendix) at this point. For measurement, you can simply draw a line/box on the image with the mouse on the SEM display monitor. To add extra text, use the Alpha menu in the SEM Control window. 3. To save images, use the Image Control window. You can change the resolution of the saved images here, as well as click or unclick the G icon to either save or not save (respectively) annotations and the data zone at the bottom of the image. a. If you re planning to publish an image, it s a good idea to save a high-res version without annotations and a low-res version with annotations. That way you ll know all the dimensions, but you can add your own (better-looking) decorations to the high-res image later. 4. Once you ve got your save settings the way you want them, click the fourth icon from the left (or File/Save) to save your image. Because this is an old SEM, you re restricted to 8- character filenames. Live with it. a. Try to save in the c:\userdata directory, after creating a subdirectory with your name on it, just for the sake of organization. 5. You can upload saved images to your UMN Netfiles locker for download later. Open Google Chrome (on the desktop) and log in using your X.500 username. a. PLEASE DO NOT USE USB STICKS TO TRANSFER FILES! USB sticks are major virus-transmission vectors on Windows machines. Netfiles is fast and 16

convenient. You can access it from any PC at http://netfiles.umn.edu b. Anything left on the hard drive is subject to deletion at any time, so make sure anything important is backed up somewhere else. 17

Unloading 1. Close the column valve with either the button on the column or the control in the corner of the SEM Control window 2. Unload your sample by pressing the Unload button on the load lock or doing Vac/Unload in the SEM Control menu. The system will pump the airlock, unload your sample, and vent automatically. When the load lock button stops flashing, your sample can be removed. 3. Remove your sample from the stage and clean off any carbon tape or other residue you may have left in the process 4. The load lock stays vented when the system isn t actively loading and unloading, so make sure the door is closed to minimize contamination. 18

19

Troubleshooting Problem Stage has crashed into the final lens; alarm is beeping. Stage has crashed into electron detector Stage won t move Loadlock reference error on SEM display Can t type user annotations onto the SEM display Scan Rot button won t turn off (other buttons might do this too) Solution Find the BNC cable plugged into a port with the ground symbol near it on the side of the main SEM chamber. Unplug that cable to stop the touch alarm and unlock the stage. Reference the stage with the reference command in the stage menu of the Stage Control window to drop it away from the final lens. Confirm that the lens isn t damaged by focusing on something and make a note in CORAL that a stage crash occurred. Crashing into the electron detector is more problematic than crashing into the lens, since it doesn t trigger the touch alarm. If the stage looks like it s contacting the detector (check the IR cam) IMMEDIATELY hit the STOP button in the Stage Control. Re-reference the stage if necessary (see final lens crash procedure above) and drop the z-axis to 20mm or some other large working distance. Make a note in CORAL that a detector crash occurred. The stage is probably out of reference. The lower-right corner of the Stage Control window will say UNREF in this case. Reference the stage using the reference command in the stage menu of the Stage Control window. Re-reference the load lock, either by pushing the REF button on the load lock or using the Reference Airlock command in the Vac menu of the SEM Control window. Click the mouse on the PC monitor. For some reason, the PC monitor has to be focused for user annotations to work. Push it twice quickly. You may have to do it a few times. I ll get around to cleaning the contact on it eventually. 20

Typing on the PC just causes annotation letters to appear on the SEM display. Can t get rid of unwanted annotations on SEM display Can t save images (TIFFERROR message when you try) Software appears to be locked up Go into the alpha menu in the SEM Control window and turn off user annotation. On the PC monitor, click the SEM Control window and hit ESC then 0. Once you restart scanning, that should clear all annotations. Due to a weird software bug, the system can t save images if thumbnail images are displayed in an open directory window. If you have a directory window open, either close it or switch to list or detail view. Reset it using the Stop SEM and Start SEM desktop icons. IMMEDIATELY run a stage reference ( reference command in the stage menu of the Stage Control window) after doing this or you risk crashing the stage! If the column chamber valve is open (check that first), the beam is probably mis-aligned. No image visible (black screen at all B/C levels) Image goes black during aperture alignment Scan the beam alignment knobs (near the focus control) around until you find the setting where the image is brightest (at 5kV, this should be around 12 o clock, but it could be very different at other voltages). Note that you ll probably have to do quite a bit of aperture realignment after changing this setting. If the beam is partially mis-aligned, certain aperture positions may block it completely. When it goes black, use the beam alignment controls to find a position where the image is restored. You may have to adjust the beam and aperture alignments iteratively to get everything lined up just right. 21

Check to make sure the video is working, first of all (see No image visible above). If that seems to be OK, try finding a large feature (scratch, speck of dust, nanoparticles) and get the column aligned and focused before you go looking for your features. Can t find features you re attempting to image If you re imaging very thin features, going to a lower beam voltage may be useful; try 2500V and work downward in 500V steps. You ll have to realign the column and aperture at each voltage, unfortunately. Set the high voltage back to 5kV when you re done! A useful trick for locating sparsely-placed features on a sample is to diamond-scribe a circle around the area of interest (assuming you can see it optically). Alternately, using the syringe in the toolbox to put a drop of gold nanoparticles on or near the area of interest can make locating your features much easier too. Make sure you let the drop dry before pumping down the airlock! Can t image features smaller than 20-25 nm This is likely a vibration issue. The source of the vibrations is currently unknown, but in the meantime imaging at night (when there s no construction and much less activity in the building) may give you better resolution. Alternately, use the Charfac SEMs for your ultra-high-res needs. Can t image thin features or surface detail Turn down the accelerating voltage. The system has a LEAP lens capable of going to voltages as low as 500V without distortion, so you have some headroom here. Lower voltage means more interaction between the beam and the surface, so you ll see more details and may be able to resolve thin features that were invisible at 5000V. Note that the tradeoff here is often increased charging, due to increased beam-surface interaction. No good way around that one. 22

There are a few solutions to this. The most obvious is to make sure the sample is solidly grounded (carbon tape holding it on a metal stub will usually work for conductive samples). If possible, coat the sample with a thin layer of metal (preferably Au or Au-Pd) to aid charge dissipation. Sample is charging If you re still having problems, turn down the beam current by setting C1 to a lower negative value in the SEM Control window. -12 is the default; you can go all the way to -4 if needed. You ll need to refocus/realign after each change, so go down in steps of 2-4 for your own sanity. If that still isn t helping, you can sometimes work around charging by getting your focus/stigmation settings optimized at another site, then moving to your imaging site and quickly taking a single slow scan before charging has affected it much. 23