M517-E077A O P E R A T I O N G U I D E

Similar documents
YSF - 300/DAR i

SHIMADZU GENERAL RADIOGRAPHIC SYSTEM

DIGITAL TABLE C506-E049E

SONIALVISION G4 Multi-purpose Digital R/F System C506-E075

SONIALVISION G4 Multi-purpose Digital R/F System C506-E075

SONIALVISION G4 Multi-purpose Digital R/F System

C501-E029C. RADspeed. GENERAL RADIOGRAPHIC SYSTEM Automatic

C506-E064. Full digital system. Printed in Japan A-NS

ECOVIEW 9 / ECOVIEW 9 PLUS Digital Radiographic System

IDEXX-PACS * 4.0. Imaging Software. Quick Reference Guide

ECOVIEW 9 / ECOVIEW 9 PLUS Digital Radiographic System

RADspeed Pro GENERAL RADIOGRAPHIC SYSTEM Automatic C501-E027B

Remote Controlled R/F System C506-E056D

User Manual Veterinary

PLD5600A High Frequency Digital Gastrointestinal &DR System(630mA)

SHIMADZU CORPORATION KYOTO JAPAN

RADspeed Pro DIGITAL RADIOGRAPHIC SYSTEM. IBD-RADProAeroDR

Digital Radiography X-Ray System. X Twin with X Mobil Roesys GmbH [rshsmi] 1/11

4.0 How to Turn On the Selenia Dimensions

MobileDaRt Evolution MX8 Version

MXHF-1500RF is controlled by Digital key panel console that displays KV, ma and mas with APR menu programmed.

Progeny Imaging. User Guide V x and Higher. Part Number: ECN: P1808 REV. F

OPERATION MANUAL MIMAKI ENGINEERING CO., LTD.

Truly flexible to meet your clinical needs

CIRCLEX 0.3/0.8P324&0.6/1.2P324DK-85

Operation Manual. Canon CXDI-1 System Digital Radiography

Precision Performance Power

ODYSSEY HF SERIES. ULTRA High Frequency X-Ray Technology. Precision... Performance... Power. Innovations in Digital Imaging.

RADspeed Pro GENERAL RADIOGRAPHIC SYSTEM Automatic C501-E027C

R/F. Comparison of Long View Radiography Systems. 1. Introduction. 2. Methods of Long View Radiography

Progeny Imaging Veterinary

DOCUMENT SCANNER INSTRUCTIONS. Space. Backup. Count Only. New File. Scanner. Feeding Option Manual Auto Semi-Auto

BCA 9SK. 60 Hz on request Absorbed current on Stationary Vac and Vac in fluoro mode

Happy Link Software INSTRUCTION MANUAL

ZView ver.2 DR Console Software. User Manual. (for HUMAN) Manual Version : 2.1 Revision : 002

Initials. Policy #:

SPECTRALIS Training Guide

User Tool Operation Manual

China Resources Wandong Medical Equipment Co., Ltd. High Frequency 50kW Digital RF System - HF51-5

BCA 9RK. 60 Hz on request Absorbed current on Stationary Vac and Vac in fluoro mode

RICOH Stereo Camera Software R-Stereo-GigE-Calibration

AeroDR X60. Motorized X-ray System

China Resources Wandong Medical Equipment Co., Ltd. High Frequency 50kW, 150kV Radiography System - HF50-R

Get more from your images with Symphony Image Processing

DIAGNOSTIC RADIOLOGY TECHNOLOGIST COMPENTENCY CHECKLIST. Name date assessment/orientation by date

Diagnostic Ultrasound System. Operation Note

Copyright 2014 SOTA Imaging. All rights reserved. The CLIOSOFT software includes the following parts copyrighted by other parties:

Chapter5 Camera Settings and Other Functions

see thoroughly, insure safety digital C-arm system Remote Compact Fully digital

Get more from your images with Symphony Image Processing

Imaging Features Available in HTML5. it just makes sense

Introduction. Introduction

MOBILEART eco C504-E009G

DR General Radiography System. DR Series

siemens.com/luminos-fusion Luminos Fusion The 2-in-1 system that fits your needs and fits your budget

KM-4800w. Copy/Scan Operation Manual

The ideal K-12 science microscope solution. User Guide. for use with the Nova5000

Product data Remote Controlled Table

SPRINGFIELD TECHNICAL COMMUNITY COLLEGE ACADEMIC AFFAIRS

COST EFFECTIVE FLAT PANEL DIGITAL RADIOGRAPHY UPGRADE SOLUTIONS

SMX-1000 Plus SMX-1000L Plus

See what you need to see, and see it anywhere.

GENERATOR SETUP NAVIGATOR STM INSTALL V1 R2 28/04/2014. NAVIGATOR STM for generators. YOURDEVICE\_ODELlib\_Driver\Ftdi directory

Ansur TNT Users Manual. Plug-In

Thanks to Autocheck function, it is possible to perform a complete check-up of the robot thanks to a stepby-step

ScanGear CS-U 5.6. for CanoScan FB1210U Color Scanner. User s Guide

TABLE OF CONTENTS 1 SAF INTRODUCTION SAFETY REQUIREMENT LABELING

Table of Contents. Sound-Eklin rev052511

Figure 1 The Raith 150 TWO

PANalytical X pert Pro Gazing Incidence X-ray Reflectivity User Manual (Version: )

Tour 1008 User Manual

Best choice in all DR. Jumong M. SG HealthCare Jumong M Digital radiography specifications

Philips XPER FD10C R7.0.4

AxioVision 4.5 Brightfield Image Capture Procedure

Digital Portable Overhead Document Camera LV-1010

VISION V FLOOR MOUNTED ТUBE SYSTEM

Basic Operating Instructions for Strata Dual Beam 235 FIB/SEM

HP Photosmart R740 series Digital Camera. User Guide

Technical Guide for Radio-Controlled Advanced Wireless Lighting

KODAK Dental Imaging Software. Quick Start Guide

CS-200. PORTABLE TRAFFIC LIGHT CONTROLLER (Software 1.05) OPERATION AND SERVICE MANUAL

ISCapture User Guide. advanced CCD imaging. Opticstar

Nikon Firmware Update for Coolpix 950 Version 1.3

NAOMI SYSTEM User Guide

HP Designjet HD Scanner and T1200 HD Multifunction Printer

Brightness and Contrast Control Reference Guide

ENG. XYZscan User Manual P 1

ACCU-GOLD QUICK START MANUAL

Table 1. Placing the Sensor in the Sensor Cradle. Step Instruction Illustration

C-750 Ultra Zoom MANUAL FOR DIRECT PRINTING. Manual outlining new features added to the camera with firmware update.

Multi Drop Bus 5-Tube Coin Changer. Series. Service Manual B D A C E SAT

Zeiss AxioImager.Z2 Brightfield Protocol

INSTRUCTION MANUAL IP REMOTE CONTROL SOFTWARE RS-BA1

SARG: The Graphical User Interface Manual

MEASUREMENT CAMERA USER GUIDE

Enhanced Functionality of High-Speed Image Processing Engine SUREengine PRO. Sharpness (spatial resolution) Graininess (noise intensity)

Portable Digital X-ray

Towertrack 2 Optical Verification System

8000 SERIES PRECISION MULTIMETER VERIFICATION AND ADJUSTMENT GUIDE

Scorpion Antennas Controller. Instruction Manual. Firmware V11.8 November Please Read This Manual Completely Before Operating The Controller

Transcription:

O P E R A T I O N G U I D E M517-E077A

This operation guide is intended to compliment each "Operation Manual" of Sonialvision Safi re II / Safi re 17 and concisely describes its basic operations. Prior to using this guide, make sure to carefully read all "Operation Manual" booklets of Sonialvision Safi re II / Safi re 17 and fully understand them. In case of Sonialvision Safi re II, Digital Radiography System DAR-8000f Operation Manual (M517-E060) X-ray High Voltage Generator UD150B-40 Operation Manual (M501-E05) Remote-Controlled X-ray Diagnostic Table ZS-100I/IR Operation Manual (M506-E034) In case of Sonialvision Safi re 17, Digital Radiography System DAR-8000f Operation Manual (M517-E060) X-ray High Voltage Generator D150BC-40, GSC-00L Operation Manual (M501-E093) Remote-Controlled X-ray Diagnostic Table ZS-100I/IR Operation Manual (M506-E034) Notation The following notation is used in this operation guide. Notation Meaning Denotes the contents of the operation 1,, 3 Denotes the contents of the operation step by step Denotes the consequences of the operation Caution Denotes a special notice of the operation Note Denotes the information which helps operation Denotes points and notices of the operation

O P E R A T I O N G U I D E System Overview... 4 Configuration... 4 Diagnostic table console... 5 X-ray high voltage generator console... 6 Digital image processor main screen... 6 Operation flow chart... 7 System Startup and Shutdown... 8 Startup... 8 Shutdown... 9 FPD calibration... 10 Enter Study... 13 Defining usual study... 13 Defining patient information in an emergency... 14 Choosing types of study... 15 Fluoroscopy/Radiography... 16 Setting radiography condition... 16 Radiography... 17 SPOT radiography/serial radiography 17 DSA radiography (Option) 18 Digital tomography (Option) 19 Process Images... 0 Printing images... 0 Sending by DICOM function... 3 Close Study... 4 Closing active study... 4 Troubleshooting... 5 Emergency stop/recovery... 5 Error messages... 6 Actions after power failure... 30 1 3 4 5 6 7 M517-E077A

1 System Overview 1 System Overview Configuration Diagnostic table Control cabinet (Flat panel detector) Control cabinet (X-ray high voltage generator) Control cabinet (Diagnostic table) Automatic collimator X-ray tube unit Subsystem control cabinet (Flat panel detector) FPD (Flat panel detector) Image monitor Microphone Speaker unit X-ray high voltage generator console Hand switch Exposure switch Diagnostic table console Foot switch Mouse Keyboard Selector (attached mouse, monitor and keyboard) PCU (FPD control) Control cabinet (Digital image processor) 4 M517-E077A

1 System Overview Diagnostic table console L1 - L4 Selects the magnifi cation size of FPD. Table to the vertical position Table to the horizontal position R1 - R4 Selects the APR. Display panel Exposure indicator Exposure switch Oblique projection center switch Reverse the image vertically Reverse the image horizontally Stop switch SID selector switches Adjusts the contrast of the image. F1 - F5 Sets up the Subdivisional Acquisition for SPOT radiography. Sets up the tomography parameter for tomography (Option). Fluoroscopy selection switch Adjusts the brightness of image. Joystick Table/imaging unit control lever M517-E077A 5

1 System Overview X-ray high voltage generator console ma/mas up/down button (Adjust in small increments) Focus button kv up/down button (Adjust in small increments) kv shuttle (Adjust in large increments) ma/mas shuttle (Adjust in large increments) 1 3 ma/mas switch button sec up/down button (Adjust in small increments) sec shuttle (Adjust in large increments) AEC button Power ON button Power OFF button +5 0-5 Hand switch Radiography program button Touch panel Home button Log button Setup button Density up/down button X-ray radiography button Radiography preparation button Digital image processor main screen Patient information Standard control panel Standard control panel/thumbnail display button Expanded control panel 6 M517-E077A

1 System Overview Operation flow chart Start up system P.8 Calibrate FPD P.10 Start study P.13 Choose type of study P.15 Set Fluo/Rad condition P.16 Fluoroscopy/Radiography P.17 Repeat as required. Print image P.0 Send by DICOM P.3 Close study P.4 Shut down system P.9 M517-E077A 7

System startup and shutdown System startup and shutdown Startup X-ray high voltage generator console 1 Verify Image monitor of digital image processor and peripherals are turned OFF. Press will be illuminated after approx. seconds. Image monitor, peripherals, and control cabinet are turned ON. Stand-by indicator 1 3 +5 0-5 Digital image processor While the system is running the power up self-testing routine, a splash screen appears. Startup is completed when the following screen appears. Main image area 8 M517-E077A

System startup and shutdown Shutdown 1 Close Digital image processor active study 1 Click on the Standard control panel. A confirmation dialog appears. Click [OK]. The study is closed. Close the application program 1 Click on the Standard control panel. A confirmation dialog appears. Click [Yes]. The system will be automatically shut down and the control cabinet will be turned OFF. 3 Turn X-ray high voltage generator console OFF the X-ray high voltage generator Press OFF. ON lamp is turned off, and then OFF lamp is turned ON. 1 3 +5 0-5 Caution After pressing the OFF button on the X-ray high voltage generator console, do not press the ON button for at least 10 seconds. If you do, the system may not operate properly. M517-E077A 9

System startup and shutdown FPD calibration This operation maintains the quality of fl uoroscopy/radiography images. Be sure to perform the FPD calibration once a day after system startup. Verify that no substance is located on the diagnostic table during FPD calibration. Approx. 15 minutes is required to complete the FPD calibration. Auto FPD calibration on system startup Perform when the system starts up. 1 Start up the system After system startup, an FPD calibration dialog appears. A on diagnostic table console blinks, and the message appears. Press and hold the [SET] on the diagnostic table console The diagnostic table moves to calibration position. It sounds beep when the diagnostic table arrives in calibration position. Press [F5 : SET] to move Table to calibration starting position. Press [F5 : SET] to move Table to calibration starting position. An FPD Calibration start dialog appears. [SET] 10 M517-E077A

System startup and shutdown 3 Click [Start] An FPD Calibration starts. [Close] button appears when FPD calibration is complete. 4 Click [Close] Caution Make sure there are no personnel in the study room before performing FPD calibration procedure as the diagnostic table exposes X-ray during the procedure. M517-E077A 11

System startup and shutdown FPD Calibration from Right-click menu Perform while the system is running. Perform FPD calibration after the study is fi nished because the operation cannot be performed during study. 1 Right-click on the main image area The right-click menu appears. Select [FPD Calibration] The FPD Calibration dialog appears. The following procedures are same as auto FPD calibration in system startup steps. 1 M517-E077A

3 Enter study Defining usual study 3 E n t e r s t u d y Enter the patient information to start study. 1 Open Patient List window 1 Click on the Standard control panel. A Patient List window appears. Click [New]. A Patient Information window appears. Enter patient information Patient Name Patient ID Patient DOB Fields indicated by an asterisk ( ) are required. 3 Click [Open] The Patient Information window is closed, and then the study starts. A patient s information appears on the upper left corner of main image area. M517-E077A 13

3 E n t e r s t u d y Defining patient information in an emergency The system automatically sets a tentative data such as a patient s name and patient ID when the time is pressing or the patient s information is unknown during emergency medical service. 1 Open Patient List window 1 Click on the Standard control panel. A Patient List window appears. Click [Emergency]. The Patient List window is closed, and then the study starts. The patient s information appears on upper left of the main image area. The following patient information is automatically defined. Patient Name (Last Name): ELNMyyyymmdd Current year/month/date is set. Patient Name (First Name): EFNMhhmmss Current hour/minute/second is set. Patient ID: EPIDyyyymmdd-hhmmss Current year/month/datehour/minute/second is set. ELNM0060501 EFNM1134 EPID0060501-1134 OTHER Sex: OTHER The tentative data must be corrected after the study. 14 M517-E077A

3 E n t e r s t u d y Choosing types of study 1 Choose [Cell Title] according to types of study The top 4 APRs according to type of study appear. Diagnostic table console display panel The top 4 APRs registered by digital image processor appear. M517-E077A 15

4 Fluoroscopy/Radiography 4 Fluoroscopy/Radiography Setting radiography condition 1 Select APR Diagnostic table console Press any key of R1 - R4 on the console to select APR. R1 - R4 Digital image processor The APRs selected on the diagnostic table console appear on the Standard control panel. X-ray high voltage generator console The radiography program associated with APR is displayed. Radiography item APR Cell title Select the magnification size Press any key of L1 - L4 on the console to select FPD magnification size. L1 - L4 16 M517-E077A

4 Fluoroscopy/Radiography Radiography Set the radiography condition prior to radiography as necessary. SPOT radiography/serial radiography 1 Verify Patient information Types of study (APR cell title and radiography item) Select 3 Fluoroscopy 4 Expose 1 Select APR for SPOT or SERIAL radiography. SPOT radiography Select Subdivisional Acquisition format from [F1] - [F4]. The selected Subdivisional Acquisition format is displayed inverted. 1 Verify that the fluoroscopy selection switch illuminates. The X-ray is not exposed when the fluoroscopy selection switch is not illuminated even if the foot switch is pressed. Press the fluoroscopy selection switch to light. Press down the foot switch. The exposure indicator illuminates. The fluoroscopy image appears on the main image area of digital image processor. Press exposure switch. SERIAL radiography F1 F F3 F4 Confirm the [IBS] and [Pulse Fluoroscopy] are selected on touch panel of X-ray high voltage generator console ([IBS] lights up when ON). X-ray exposure terminates after images are acquired within preset time. Acquired images are serially displayed (Auto Replay: ON) or the last image appears (Auto Replay: OFF) after exposure. Releasing the exposure switch stops the image acquisition during exposure. IBS Note Pulse Fluoroscopy M517-E077A 17

4 Fluoroscopy/Radiography DSA radiography (Option) 1 Verify Patient information Type of study (APR cell title and radiography item) Select Select APR for DSA radiography. 3 Fluoroscopy 4 Prepare 5 Expose 1 Confirm that the fluoroscopy selection switch illuminates. The X-ray is not exposed when the fluoroscopy selection switch is not illuminated even if the foot switch is pressed. Press the fluoroscopy selection switch to light. Press down the foot switch. The exposure indicator illuminates. The fluoroscopy image appears on the main image area of digital image processor. injector Press and hold down the hand switch. Mask images followed by live images are acquired (automatically). The DSA image appears on the main image area of digital image processor. X-ray exposure terminates after images are acquired within preset time. Releasing the exposure switch stops the image acquisition during exposure. 18 M517-E077A

4 Fluoroscopy/Radiography Digital tomography (Option) 1 Verify Patient information Types of study must be selected. Select 3 Set 4 Move 5 Exposure Select APR for tomography. up parameters 1 To select a parameter, press [F: Select]. To change the value, press [F3: Value] or [F4: Value]. The exposure time is automatically determined when the diagnostic table moves to the starting point of tomography. diagnostic table to starting point Press [F5: SET] until the key is highlighted. Press [F1: Cancel] for canceling the exposure operation. Layer height Pitch Number of exposure (serial mode) TEST RUN F1 F F3 F4 F5 Press and hold down the hand switch. The tomography image appears on the main image area after each exposure. Hold down the hand switch until the table stops. When the hand switch is released during the process, the images for which the exposures have been completed are stored on the hard disk. Returning angle of X-ray tube unit Press the oblique projection center switch. M517-E077A 19

5 Process images 5 Process Images Printing images Do one of the following to print images. Click [Print Image] : Confi rm the actual image displayed on the main image area and print the image. Select from right-click menu: Confi rm the actual image on the main image area and select/deselect to print the image. Select from Patient List: Entire patient fi le or specifi c study can be printed. Print by [Print Image] 1 Click A mark "S" appears in the lower right corner on the main image area. The images will be printed when the number of frames reaches the quantity of film format (up to 1 frames). S Click Print image without sending specific quantity The Print Layout menu appears. 3 Select [Eject Film] The images will be printed. Note Printed images are saved in the "Print ShotSave" folder. 0 M517-E077A

5 Process images Print from Right-click menu 1 Right-click on the main image area The Right-click menu appears. Select [Print Select] A mark "H" appears in the lower right corner of the main image area. The mark "H" appears when the image to be printed is selected. To delete the mark "H", select [Print Select] from Right-click menu again. H 3 Select [Print Start] The image will be printed. M517-E077A 1

5 Process images Print from Patient List 1 Click on the Standard control panel The Patient List window appears. Select patient file or study Patient Either the patient file or specific study can be selected from Patient List window. Study 3 Click either [Print Patient] or [Print Study] Print Study Print Patient The Configure Print Settings dialog appears. Clicking [Print Patient], all images in the selected patient file are printed. Clicking [Print Study], all images in the selected study are printed. 4 Change or set the configurations 5 Click [OK] All images are printed. M517-E077A

5 Process images Sending by DICOM function The images including study can be sent to CD-R or DICOM network (Option) in DICOM format in the following ways. CD-R Export to DICOM CD from Patient List DICOM network (Option) Manual DICOM Send Auto DICOM Send DICOM Send from Patient List DICOM Send from Thumbnail This clause describes only Manual DICOM Send. For more DICOM exporting, refer to "DAR-8000f Operation Manual". Manual DICOM Send (Option) For manual DICOM send, open the patient fi le including image and allow that image to be displayed on the main image area. 1 Click on the Standard control panel The DICOM menu appears. Choose 3 Select 4 Click 5 Click [Send Study] or [Send ShotSave] Clicking [Send Study] sends the entire image in selected study. Clicking [Send ShotSave] sends images in ShotSave folder. The Archive Destinations dialog appears. the destinations [OK] The Archive Destinations dialog is closed and a confirmation window appears. [OK] The image or patient files are sent. The patient file is closed. M517-E077A 3

6 C l o s e S t u d y 6 Close Study Closing active study 1 Click on the Standard control panel A confirmation dialog appears. Click [OK] The patient file is closed. 4 M517-E077A

7 Troubleshooting Emergency stop/recovery 7 Troubleshooting Emergency stop the diagnostic table operation Press the stop switch on either the diagnostic table console or diagnostic table. Diagnostic table console Diagnostic table Recovery from emergency stop To recover from emergency stop, be sure to turn the stop switch clockwise. Caution Ensure to pause 10 seconds or more between halt and recovery. M517-E077A 5

7 Troubleshooting Error messages X-ray high voltage generator console When the system detects a fault, an error message appears on X-ray high voltage generator console. Touch the key to close the message screen, and then take action described below for the error message. Messages related to the X-ray high voltage generator (fault indicator) on the console illuminates and one of the following error messages appears. The system cannot expose while these messages are displayed. Message Meaning Procedure Radio Over Current Radiography tube current exceeded set value + 00 ma. Measured kv OVER Starter ERR Starter is not working Measured tube voltage exceeded permitted range. Starter is in abnormal situation or condition. Starter is not working when fl uoroscopy. If this message appears repeatedly, contact Shimadzu service representative. I.F. OVER Abnormal fi lament heating current. Line Voltage OVER Supply voltage exceeded permitted range. Charge Volt ERR Abnormal charging voltage of primary smoothing capacitor. Power Down Abnormal control circuit supply voltage. Contact Shimadzu service representative. H. V. T Not Connected Abnormal connection to high-voltage transformer. Arcing Trouble Repeated arcing occurred. 6 M517-E077A

7 Troubleshooting Messages related to the radiography conditions (caution indicator) on the console illuminates and one of the following error messages appears. The system cannot expose while these messages are displayed. Message Meaning Procedure mas OVER mas too small Emission OVER Generator Load OVER mas/time OVER mas value exceeded set value by 800 mas. mas value was less than 0.5 mas. Or, tube current was below minimum value for mas setting method. Emission characteristics out of range. Equipment ratings exceeded. mas value too large and radiography time exceeded 10 sec. Or, radiography time exceeded permitted range. Change the set value. Increase radiography tube voltage or reduce radiography tube current. Decrease radiography tube voltage or radiography tube current. Reduce the set value of mas or sec. control range OVER Setting out of range for mas setting method. Reduce the set value. Messages related to the radiography conditions One of the following error messages appears. The system cannot expose while these messages are displayed. Message Meaning Procedure HU OVER Predicted Value predicted to go out of range if the heat units increase from the current heat unit value according to the set radiography. Change the radiography conditions or cease operation until heat unit value reduces. HU Full Stored Thermal OVER Heat units reached the permitted limit. X-ray tube unit temperature exceeded the permitted limit. Cease operation until heat unit value reduces. M517-E077A 7

7 Troubleshooting Error messages at power ON Message Meaning Procedure Panel Battery WARNING Cabinet Battery WARNING Panel Battery EMPTY Cabinet Battery EMPTY The data recording battery in the operation panel must be replaced soon. The data recording battery in the control cabinet must be replaced soon. The data recording battery in the operation panel is empty. The data recording battery in the control cabinet is empty. Contact Shimadzu service representative. X-ray switch ERR The radiography button remains ON. Fluo switch ERR The fl uoro foot switch remains ON. Messages related to Communication (Option) (caution indicator) on the console illuminates and one of the following error messages appears. The system cannot expose while these messages are displayed. Message Meaning Procedure Communication ERR Received radiographic conditions cannot be set for the equipment. Change the setting at the external instrument. Other messages (caution indicator) on the console illuminates and one of the following error messages appears. The system cannot expose while these messages are displayed. Message Meaning Procedure Door/Interlock AEC OVER The fl uoroscopy table is not prepared The examination room door is open. AEC failed and AEC backup was used during AEC radiography. The fl uoroscopy table is not prepared due to incorrect position of X-ray tube. Close the door. Increase radiography tube voltage or extend the radiography time. Confi rm the diagnostic table confi guration. 8 M517-E077A

7 Troubleshooting Diagnostic table console Message Position Detect Error XXX Position Error XXX Motor Error XXX Meaning Position detect error in operating parts Potentiometer break detect Motor error Procedure Reset by pushing STOP switch, or re-startup the system. If the system is not restored, contact Shimadzu service representative. M517-E077A 9

7 Troubleshooting Actions after power failure In case of power failure due to lightning, UPS (Uninterruptible Power Supply) contained in control cabinet detects the power failure, and then automatically shuts down the FPD (PCU). Without automatic startup for FPD (PCU) after power failure recovery (backup by UPS), perform manual startup. 1 Turn ON the system "Startup" P.8 Turn 3 Shut 4 Verify ON the power switch of PCU Application software "Xcat" for FPD control starts up automatically. Approx. 5 minutes is required to complete the startup. down the system Turn ON the system again the FPD power control condition Click [Tool], and then click [Power ]. A HV Power Control dialog appears. Verify that the check boxes of [Synchronize with the host system] and [Enable timer control] are filled. Note Battery backup time by UPS is approx. 3 minutes. It sounds beep from control cabinet while the power is being backed up by UPS. If the power fails because backup time for UPS expires, call a Shimadzu service representative. 30 M517-E077A

TOKYO OFFICE 3, Kanda-Nishikicho 1-chome, Chiyoda-ku, Tokyo 101-8448, Japan Phone: 81(3)319-5641 FAX: 81(3)319-5710 Cable Add.: :SHIMADZU TOKYO Overseas Telex No.: 03-391 (SHMDT J) KYOTO OFFICE 1, Nishinokyo-Kuwabaracho, Nakagyo-ku, Kyoto 604-8511, Japan. Cable Add.: SHIMADZU KYOTO Overseas Telex No.: 054-166 (SHMDS J) SHIMADZU EUROPA GmbH Albert-Hahn-Strasse 6-10, D-4769 Duisburg, F.R. Germany Phone: 49(03)7687-0 Fax: 49(03)7666-5