X-RAY FLUOROSCOPY IMAGING SYSTEMS. Dr Slavik Tabakov. Luminescence: Dept. Medical Eng. & Physics King s College London

Similar documents
X-ray Tube and Generator Basic principles and construction

BASICS OF FLUOROSCOPY

Introduction. Chapter 16 Diagnostic Radiology. Primary radiological image. Primary radiological image

Fluoroscopy - Chapter 9

10/15/2012 SECTION III - CHAPTER 6 DIGITAL FLUOROSCOPY RADT 3463 COMPUTERIZED IMAGING

Joint ICTP/IAEA Advanced School on Dosimetry in Diagnostic Radiology and its Clinical Implementation May 2009

I. PERFORMANCE OF X-RAY PRODUCTION COMPONENTS FLUOROSCOPIC ACCEPTANCE TESTING: TEST PROCEDURES & PERFORMANCE CRITERIA

TESTING FLAT-PANEL IMAGING SYSTEMS: What the Medical Physicist Needs to Know. JAMES A. TOMLINSON, M.S., D.A.B.R. Diagnostic Radiological Physicist

DIGITAL IMAGE PROCESSING IN X-RAY IMAGING

10/26/2015. Study Harder

Beam-Restricting Devices

LECTURE 1 The Radiographic Image

10/3/2012. Study Harder

Amorphous Selenium Direct Radiography for Industrial Imaging

X-ray Imaging. PHYS Lecture. Carlos Vinhais. Departamento de Física Instituto Superior de Engenharia do Porto

Seminar 8. Radiology S8 1

X-RAYS - NO UNAUTHORISED ENTRY

Basis of Computed Radiography & PACS

Do you have any other questions? Please call us at (Toll Free) or , or

Dose Reduction and Image Preservation After the Introduction of a 0.1 mm Cu Filter into the LODOX Statscan unit above 110 kvp

Multiple Choice Identify the letter of the choice that best completes the statement or answers the question.

Photons interaction with matter

X-ray Tube and Generator Basic principles and construction

AN ABSTRACT OF THE THESIS OF. W. Scott Helms for the degree of Master of Science in Radiation Health Physics

X-Ray Medical Imaging and Pixel detectors

Basis of Computed Radiography & PACS

Distributed by. Ecotron Anyview Series. The Intelligent C-Arm System

Nuclear Associates

Acquisition, Processing and Display

Radiology. Radiograph: Is the image of an object made with use of X- ray instead of light.

Radiographic sensitivity improved by optimized high resolution X -ray detector design.

Exposure System Selection

siemens.com/luminos-fusion Luminos Fusion The 2-in-1 system that fits your needs and fits your budget

SPRINGFIELD TECHNICAL COMMUNITY COLLEGE ACADEMIC AFFAIRS

CHAPTER 10. DIGITAL SUBTRACTION ANGIOGRAPHY (DSA) 10.1 Introduction

CHAPTER 6 QC Test For Fluoroscopic Equipment. Prepared by:- Kamarul Amin bin Abu Bakar School of Medical Imaging KLMUC

Radiology Physics Lectures: Digital Radiography. Digital Radiography. D. J. Hall, Ph.D. x20893

Nuclear Associates

SPRINGFIELD TECHNICAL COMMUNITY COLLEGE ACADEMIC AFFAIRS

X-rays. X-rays are produced when electrons are accelerated and collide with a target. X-rays are sometimes characterized by the generating voltage

Photomultiplier Tube

Mammography: Physics of Imaging

1-1. GENERAL 1-2. DISCOVERY OF X-RAYS

X-rays in medical diagnostics

Fluoroscopy & Tomography. Rhodes

Version 1.0. TechnicVR. Student Guide

Unit thickness. Unit area. σ = NΔX = ΔI / I 0

Mammography is a radiographic procedure specially designed for detecting breast pathology Approximately 1 woman in 8 will develop breast cancer over

X-RAY IMAGING EE 472 F2017. Prof. Yasser Mostafa Kadah

Essentials of Digital Imaging

ProX Intraoral X-ray. PLANMECA is proud to introduce a new intraoral X-ray unit to its comprehensive collection of imaging products- the ProX.

NDE SOLUTIONS RADIOGRAPHY COURSE OUTLINE

Standard Guide for Radioscopy 1

X-RAY. Lecture No.4. Image Characteristics:

1. Carlton, Richard R., and Arlene M. Adler. Principles of Radiographic Imaging: An Art and a Science, 5th edition (2013).

PLD5600A High Frequency Digital Gastrointestinal &DR System(630mA)

Digital Detector Array Image Quality for Various GOS Scintillators

Effect of patient support pads on image quality and dose in fluoroscopy a)

Luminos drf Max. Taking 2-in-1 to the MAX in radiography and fluoroscopy. siemens.com/luminos-drf-max

V SALAI SELVAM, AP & HOD, ECE, Sriram Engg. College, Perumalpattu 1 MEDICAL ELECTRONICS UNIT IV

COMPUTED RADIOGRAPHY CHAPTER 4 EFFECTIVE USE OF CR

QC in Diagnostic Radiology. Main steps for a QC survey in Diagnostic Radiology

Luminos RF Classic. Where value meets performance.

NPTEL NPTEL ONLINE COURSE. NPTEL Online Certification Course (NOC) NPTEL. Theory and Practice of Non Destructive Testing

X-RAY TUBE ARCING: MANIFESTATION AND DETECTION DURING QUALUTY CONTROL

CR Basics and FAQ. Overview. Historical Perspective

Film Replacement in Radiographic Weld Inspection The New ISO Standard

Contrast. Contrast: the difference in density on adjacent areas of a radiograph or other image receptor. Subjective. Long Scale (Low Contrast)

Nuclear Associates

Setting up digital imaging department!

SYLLABUS. 1. Identification of Subject:

Overview. Professor Roentgen was a Physicist!!! The Physics of Radiation Oncology X-ray Imaging

Breast Tomosynthesis. Bob Liu, Ph.D. Department of Radiology Massachusetts General Hospital And Harvard Medical School

Mobile digital radiography system for nondestructive testing of large diameter pipelines

Overview of Safety Code 35

RADIATION SAFETY DIRECTORATE

ISO INTERNATIONAL STANDARD

A Comprehensive Review of Image Production

Examination, TEN1, in courses SK2500/SK2501, Physics of Biomedical Microscopy,

Medical Imaging. X-rays, CT/CAT scans, Ultrasound, Magnetic Resonance Imaging

PROCEEDINGS OF A SYMPOSIUM HELD AT THE CAVENDISH LABORATORY, CAMBRIDGE, Edited by

Estimation of signal transfer property for wireless digital detector in different measurement schemes

COPYRIGHT 2002 by Srinivasan Vedantham ALL RIGHTS RESERVED. Use or inclusion of any portion of this document in another work intended for

Phase Imaging Using Focused Polycapillary Optics

Nuclear Associates

CPSC 4040/6040 Computer Graphics Images. Joshua Levine

BCA 9SK. 60 Hz on request Absorbed current on Stationary Vac and Vac in fluoro mode

Teaching Digital Radiography and Fluoroscopic Radiation Protection

Essentials of Digital Imaging

PRACTICAL CONSIDERATIONS AND EFFECTS OF METALLIC SCREEN FLUORESCENCE AND BACKSCATTER CONTROL IN GAMMA COMPUTED RADIOGRAPHY

BCA 9RK. 60 Hz on request Absorbed current on Stationary Vac and Vac in fluoro mode

Performance of Image Intensifiers in Radiographic Systems

PD233: Design of Biomedical Devices and Systems

Technical information. X-RAY PRODUCTS/Integrated solutions

Real Time Linear Array Imaging. Brian Caccamise

DENTAL RADIOGRAPHY KAMARUL AMIN BIN ABU BAKAR

SECTION I - CHAPTER 2 DIGITAL IMAGING PROCESSING CONCEPTS

Essential Parameters for the Visibility of IQIs and Small Indications in Digital Radiography

Test Equipment for Radiology and CT Quality Control Contents

Examination of Pipe Welds by Image Plate Based Computed Radiography System

Transcription:

X-RAY FLUOROSCOPY IMAGING SYSTEMS Dr Slavik Tabakov OBJECTIVES - Image Intensifier construction - Input window - Accelerating and focusing electrodes - Output window - Conversion factor - II characteristics - Modulation Transfer function -Digital fluoroscopy Dept. Medical Eng. & Physics King s College London E-mail: slavik.tabakov@kcl.ac.uk, slavik.tabakov@emerald2.co.uk Fluoroscopy delivers very high patient dose. This can be illustrated with an example: The electrical energy imparted to the anode during an exposure is A = C 1. U a. I a. T The X-ray tube anode efficiency is E = C 2. Z. U a Luminescence: Fluorescence - emitting narrow light spectrum (very short afterglow ~nsec) - PM detectors; II input screens (CsI:Tl) Phosphorescence - emitting broad light spectrum (light continues after radiation) - monitor screens, II output screens (ZnCdS:Ag) The old fluoroscopic screens are no longer used due to high dose and low resolution From the two equations follows that the energy produced in a single exposure will be X = C. A. E = C. Z. (U a ) 2. I a. T = (C. Z). 2. mas Radiography of the lumbar spine (with parameters 80, 30 mas): X = k. 80.80.30 = k. 192,000 Fluoroscopy - 3 minutes Barium meal (with parameters 80, 1mA) X = k. 80.80.1.3.60 = k. 1,152,000 In this example fluoroscopy delivers approx. 6 times more X-ray energy (dose)

Basic Components of an Image Intensifier - Input window (Ti or Al) 95% transmission - Input screen: CsI (new) or ZnS (old) phosphor - Photocathode (a layer of CsSb 3 ) - Accelerating electrodes zoom (e.g. 30/23/15 cm) - Output screen (2.5 cm) - II housing (mu-metal) - Output coupling to the TV camera II Input screen: Columnar crystals of CsI which reduces dispertion (collimation); absorbs approx. 60% of X-rays Photocathode applied directly to CsI both light spectrum match very well II Accelerating electrodes II Output screen: Phosphor (ZnCdS:Ag) on glass base The accelerated e - produce multiple light photons; thin Al foil prevent return of light (veiling glare) Coupling: fibre optic or tandem optic Conversion factor ~100-1000 (cd.m -2 /μgy.s -1 ) = (output phosphor light / input screen dose rate) Total gain (out. light photons /inp. X photons )

Total gain (out. light photons /inp. X photons ) 1 X-ray photon >> 1000 light photons (input screen) >> >>50 photo e - >> 3000 light photons (output screen) in the case above the total gain is 3000 MTF of II depending on zoom (magnification) Some II Characteristics: Minification gain -D m -inp./output diam. (D inp / D out ) 2 Flux gain -F x (approx. 30-60): Out.scr. light photons / inp. ligh photons to photocath. Brightness gain -G B G B = D m x F x * Zooming increases the resolution, but requires higher dose rate!! Contrast Ratio -X-ray scatter at input window, input phosphor -Light scatter within phosphor, not-absorbed light by phosphor -Back scatter from output phosphor (to photocathode), at output window L c light intensity at centre of image (pure white) Cont. Ratio (C v )= L c /L d : ideally max/0 ; in reality approx. 30/1 L d - light intensity at centre of image (cover with Pb) II field size 40 cm (16 ) 32 cm (12.5 ) 20 cm (8 ) 15 cm (6 ) Resolution (Lp/mm) 4.0 4.2 5.5 6.0 Contr. ratio 20:1 25:1 30:1 35:1 Automatic Brightness Control System (ABS) - produces images with constant brightness by keeping constant entrance dose rate to the II The feedback C1 have two options - taking signal from D1 (dosimeter) or D2 (photometer). * II entr. dose rate is approx. 1 μgy/sec and should not exceeds 2 μgy/sec. * The maximal patient entrance skin dose should not exceed 0.01 Gy/min). Convers. Factor (cd/m / mr/s) 166 100 60 50 Distortion (pincushion %) 9 4.5 1.4 1 Dose (relative) 0.25 0.5 0.75 1 Table from: D.Dowsett, P.Kenny, E.Johnston - different types and characteristic curves of changing the /ma Graph from: E Krestel (SIEMENS)

TV camera types: Vidicon - gamma 0.7; slow response, some contrast loss (light integration), high dark current, but low noise - suitable for organs Plumbicon - gamma 1; quick response, small dark current, but high noise - suitable for cardiac examinations Modulation Transfer Function and Contrast Transfer Function 10% - cut-off frequency (lim. sp. res.) Modulation Transfer Function MTF=(recorded signal f)/(origin. signal f); also MTF(f) = FT{LSF(x)} MTF~m = sin π.u f / π.u f,where u f = f/l * (M-1)/2M,where M-magnif.; f focal spot; L period of the structure (~ to spatial frequency)

Overall II-TV system MTF = MTF 1 x MTF 2 x x MTF n Dynamic range of II -much larger than this of radiographic film (output luminance per dose unit) Resolution and Magnification of II - electronic zoom up to 4 times (lp/mm) Digital Fluoroscopy Digital subtraction and unsharp masking

Mathematical operation in DSA: Functional imaging; Logarithmic & Square Root Subtraction, etc. Functional Imaging 60 70 II contrast with different (constant ma) 90 100