Half value layer and AEC receptor dose compliance survey in Estonia

Similar documents
Estimation of signal transfer property for wireless digital detector in different measurement schemes

Practical Medical Physics Session: TG-151 Dose Monitoring. August 5, 2013 Katie Hulme, M.S.

I. PERFORMANCE OF X-RAY PRODUCTION COMPONENTS FLUOROSCOPIC ACCEPTANCE TESTING: TEST PROCEDURES & PERFORMANCE CRITERIA

8/2/2017. Radiologist Responsibilities. Radiologist Responsibilities. Medical Physicist Mammography Equipment Evaluation and Annual Survey

Joint ICTP/IAEA Advanced School on Dosimetry in Diagnostic Radiology and its Clinical Implementation May 2009

Quality assurance: a comparison study of radiographic exposure for neonatal chest radiographs at 4 academic hospitals

Features and Weaknesses of Phantoms for CR/DR System Testing

Exposure Indices and Target Values in Radiography: What Are They and How Can You Use Them?

SPECIFICATION. Kilovoltage X-ray calibration system for protection and diagnostic level dosimetry. Prepared by

Learning Objectives: What s my motivation? (unknown screen actor) Workshop Overview

SUSPENSION CRITERIA FOR IMAGE MONITORS AND VIEWING BOXES.

STEREOTACTIC BREAST BIOPSY EQUIPMENT SURVEYS

A comparison of two methods for the determination of freein-air geometric efficiency in MDCT

Exposure System Selection

X-RAYS - NO UNAUTHORISED ENTRY

APPLICATION NOTE. RTI Electronics AB, Sweden Revision A, July kvp and time measurements with the Barracuda or Piranha on a typical dental unit

Nuclear Associates

TESTING FLAT-PANEL IMAGING SYSTEMS: What the Medical Physicist Needs to Know. JAMES A. TOMLINSON, M.S., D.A.B.R. Diagnostic Radiological Physicist

Overview of Safety Code 35

Dose Reduction and Image Preservation After the Introduction of a 0.1 mm Cu Filter into the LODOX Statscan unit above 110 kvp

REQUIREMENTS FOR LICENCE HOLDERS WITH RESPECT TO QUALITY CONTROL TESTS FOR DIAGNOSTIC X-RAY IMAGING SYSTEMS

Calibration of KAP meters

Acceptance Testing of a Digital Breast Tomosynthesis Unit

Investigation of the line-pair pattern method for evaluating mammographic focal spot performance

Unfors EDD-30 Radiation Protection in Fluoroscopy

Test Equipment for Radiology and CT Quality Control Contents

X-RAY MEDICAL EQUIPMENT

Beam-Restricting Devices

This document is a preview generated by EVS

Y11-DR Digital Radiography (DR) Image Quality

RaySafe X2. Effortless measurements of X-ray

Disclosures. Outline 7/31/2017. Current Implementation Status of IEC Standard : Exposure Index (EI) for Digital Radiography

Effect of patient support pads on image quality and dose in fluoroscopy a)

The importance of radiation quality for optimisation in radiology

Key words: fluoroscopy, dose-area-product, kerma-area-product, calibration of KAP meters, patient exposure

The effect of compensating filter on image quality in lateral projection of thoraco lumbar radiography

Ask EuroSafe Imaging Tips & Tricks. Paediatric Imaging Working Group. Dose Management in Digital Radiography

History of digital imaging

Veterinary Science Preparatory Training for the Veterinary Assistant. Floron C. Faries, Jr., DVM, MS

Ansur TNT Users Manual. Plug-In

Introduction. Chapter 16 Diagnostic Radiology. Primary radiological image. Primary radiological image

Assessment of Beam Alignment, Collimation and Half Value Layer of Some Selected X-Ray Machines in Plateau State, Nigeria

2012 :15th SESSION of ESMP

Exposure in Dental Radiology: A Comparison Between Intra-oral, Panoramic and Tomographic Examinations

Determination of the tube voltage from clinic mammographic system using two types of detectors

While digital techniques have the potential to reduce patient doses, they also have the potential to significantly increase them.

INTERNATIONAL STANDARD

ISO INTERNATIONAL STANDARD

A study of exposure index value fluctuations in computed radiography and direct digital radiography using multiple manufacturers

I. Introduction.

Studies on reduction of exposure dose using digital scattered X-ray removal processing

NEMA XR X-ray Equipment for Interventional Procedures User Quality Control Mode

Studies on reduction of exposure dose using digital scattered X-ray removal processing

X-RAY IMAGING EE 472 F2017. Prof. Yasser Mostafa Kadah

DISC QC/QA Program for Digital Imaging Systems using the DR Radchex Plus Meter

RAD 150 RADIOLOGIC EXPOSURE TECHNIQUE II

QUALITY CONTROL TESTS IN SOME DIAGNOSTIC X-RAY UNITS IN BANGLADESH

Measurement of table feed speed in modern CT

RADIOGRAPHIC EXPOSURE

Appropriate Inspection Distance of Digital X-Ray Imaging Equipment for Diagnosis

Practical Aspects of Medical Physics Surveys of Mammography Equipment and Facilities

Radiographic viewing conditions at Johannesburg Hospital

QC in Diagnostic Radiology. Main steps for a QC survey in Diagnostic Radiology

of sufficient quality and quantity

ABOVETABLE X-RAY SOURCE FLUOROSCOPIC AND SPOT-FILM SYSTEMS

New Exposure Indicators for Digital Radiography Simplified for Radiologists and Technologists

Mammography: Physics of Imaging

Fully Automatic X-ray meter kvp, dose, rate, HVL and time Extremely easy to use. Unfors ThinX RAD for Radiographic Quality Control

2217 US Highway 70 East Garner, NC Main: Fax:

Report on the study and optimisation of chest and lumbar spine X-ray imaging

Outline ASRT Changes Impact on current curriculum Potential new courses WECM Changes Last update Resources and needs

QC Testing for Computed Tomography (CT) Scanner

Title: A COMPARISON OF Cs-137 AND X-RAY SOURCES AS CALIBRATION REFERENCES FOR THERMOLUMINESCENT DOSIMETER CHIPS

X-ray Tube and Generator Basic principles and construction

Predicted image quality of a CMOS APS X-ray detector across a range of mammographic beam qualities

INTERNATIONAL STANDARD

Digital radiography (DR) post processing techniques for pediatric radiology

COMPUTED RADIOGRAPHY CHAPTER 4 EFFECTIVE USE OF CR

ISO Photography Intra-oral dental radiographic film and film packets Manufacturer specifications

Quality Control for Stereotactic Breast Biopsy. Robert J. Pizzutiello, Jr., F.A.C.M.P. Upstate Medical Physics, Inc

DRAFT Technical evaluation of Philips Microdose SI digital mammography system

Version 1.0. TechnicVR. Student Guide

Teaching Digital Radiography and Fluoroscopic Radiation Protection

Comparison of computed radiography and filmõscreen combination using a contrast-detail phantom

Surveying and QC of Stereotactic Breast Biopsy Units for ACR Accreditation

Introduction of a Single Chip TLD System for Patient Dosimetry

Calculati tions - BIR meth thod and NCRP 147

Outline. Digital Radiography. Understanding Digital Modalities: Image Quality and Dose. Image Quality. Dose Control

DIGITAL IMAGE PROCESSING IN X-RAY IMAGING

Interventional Radiological Equipment selection and installation

SYLLABUS. TITLE: Equipment Operation I. DEPARTMENT: Radiologic Technology

Attikon, Rimini 1, , Athens, Greece , Athens, Greece , Athens, Greece

Aspire HD. Program Manual. 2nd Edition - October 2012

DETERMINATION OF INHERENT AND ADDITIONAL FILTRATION IN ORDER TO ESTABLISH RADIATION QUALITIES ACCORDING TO IEC 61267

ELECTRONIC CONTROL CONCEPTS 160 Partition Street Saugerties, NY or local phone

A comparative study of several digital flat panel X-ray units: patients doses and image quality in chest radiography

Nuclear Associates

USER MANUAL. RaySafe X2. RaySafe X2 Solo

DIAGNOSTIC ACCREDITATION PROGRAM. Radiology and CT Quality Control Procedures Workbook

Derivation of total filtration thickness for diagnostic x-ray source assembly

Transcription:

Half value layer and AEC receptor dose compliance survey in Estonia K. Kepler, A. Vladimirov Training Centre of Medical Physics, University of Tartu Testing Centre of the University of Tartu, Estonia E-mail: kalle.kepler@ut.ee 9th International Conference & Workshop Medical Physics in the Baltic States, Kaunas, 13-15 October 2011

Introduction CEC Radiation Protection 91 (1997) Criteria of acceptability of radiological (including radiotherapy) and nuclear medicine installations CEC Radiation Protection 162 Radiation criteria for acceptability of medical radiological equipment used in diagnostic radiology, nuclear medicine and radiotherapy Suspension levels (type A, B, C and D) quantitative indicators of performance which must be met Draft for public consultation 2009 Final draft - Dublin workshop, Sept. 2011 http://www.neyhqarc.nhs.uk/rp162/

Introduction During the last decade practically all radiology departments of Estonia have successfully moved from film-screen based technology to filmless digital technology - computed radiography (CR) and digital radiography (DR). Now the majority (about 90%) of the radiographic systems are CR systems, but the number of DR systems is increasing. In the same time many AEC systems of the radiographic equipment have remained adjusted as for film-screen systems sensitivity and energy response. The question is, if the AEC needs readjustment for being used with CR system. One of the significant modifications in the new draft of Radiation Protection 162 (RP162) includes specification of acceptable half value layer (HVL) values for diagnostic X-ray equipment at different tube potentials. The new criteria are based on the international standard IEC 60601-1-3:2008. In the previous Radiation Protection 91 (RP91) only criteria for total filtration (>2.5 mm Al), was given. In practice, total filtration is not directly measurable as HVL.

Minimum permissible HVL in X-ray equipment by IEC 60601-1-3 X-ray tube voltage, kv IEC (1994) HVL, mm Al IEC (2008) HVL, mm Al 50 1,5 1,8 60 1,8 2,2 70 2,1 2,5 80 2,3 2,9 90 2,5 3,2 100 2,7 3,6 110 3,0 3,9 120 3,2 4,3 130 3,5 4,7 140 3,8 5,0 150 4,1 5,4

Materials and methods Half value layer (HVL) was estimated by Barracuda MPD detector (RTI, Sweden) using narrow beam geometry with aluminium filter thicknesses of 1, 2, 2.5, 3 and 4 mm for 226 general radiographic X-ray tube assemblies at 80 kvp and for 53 dental X-ray tube assemblies at 70 kvp. Receptor entrance dose set by AEC was estimated for 54 AEC systems in routine QC tests during last two years in order to get distribution of AEC setup curves at different tube potentials. The indicated tube current - exposure time product value (mas) was used to calculate receptor entrance dose taking into account the measured tube output, HVL and estimated scattering for 15 cm PMMA phantom. The selection included CR systems manufactured by Agfa (31), Fujifilm (19) and Kodak (1) and DR systems manufactured by Trixell, GE and DRTech.

Measurement geometries for receptor dose by IPEM IPEM. Measurement of the Performance Characteristics of Diagnostic X-Ray Systems: Digital Imaging Systems. IPEM Report Number 32, Part VII (2010)

Results: HVL in general radiography IEC, 1994 Only one unit of the tested 226 X-ray tube assemblies evaluated by the IEC 60601-1- 3:1994 (HVL>2.3 mm Al at 80 kv) was subject to readjustment. 14% of the tested X-ray tube assemblies evaluated by the IEC 60601-1-3:2008 and draft RP162 criteria (HVL>2.9 mm Al at 80 kv) were subject to readjustment.

Results: HVL in dental radiography Only one unit of the tested 53 dental X-ray tube assemblies evaluated by the IEC 60601-1-3:1994 and RP91 criteria (HVL>2.1 mm Al at 70 kv or total filtration >1.5 mm Al) was subject to readjustment. IEC 60601-1-3:2008 sets the minimum HVL at 70 kv to 2.5 mm Al which have made necessary to troubleshoot 59% of the tested dental X-ray tube assemblies. Draft RP162 does not state any criteria for HVL in dental radiography.

Results: AEC & receptor entrance dose In studies of 54 AEC units the estimated receptor entrance dose varied, depending on AEC calibration: at 60 kv from about 3 to 16 Gy, third quartile 6.4 Gy at 80 kv from about 2 to 9 Gy, third quartile 4.5 Gy It is obvious that AEC of some CR systems have remained calibrated by film-screen sensitivity at lower energies (60 kv) some receptor doses are located at about 10-12 Gy.

Results: AEC & receptor entrance dose In studies of 54 AEC units the estimated receptor entrance dose varied, depending on AEC calibration: at 100 kv from about 2 to 7 Gy, third quartile 4.2 Gy at 125 kv from about 2 to 7 Gy, third quartile 3.8 Gy

Results: AEC curves for CR and DR Average receptor dose for CR systems is given with standard deviation (vertical) bars compared with the receptor dose for three DR systems from different manufacturers and pre-adjustments.

Conclusions About every seventh of the tested radiographic X-ray tube assemblies and more than half of the tested dental X-ray tube assemblies need some filtration to be added in order to comply with the new IEC standard and RP162 requirements. The results correspond well with the draft RP162 recommended type C criteria for receptor entrance dose: maximum allowed dose of 10 Gy per plate while using grid. In practice some AEC used in CR systems may need adjustment at lower tube potentials. Apparently the AEC system has remained with the same adjustment as it was for screen-film systems using higher receptor doses at low potentials. Average receptor dose of the CR systems was about twice as high as for the best optimised DR systems. The draft RP162 specifies radiation criteria for the radiological equipment useful for radiation protection purposes and carrying out routine tests. Clinically optimal setup of AEC system is still an open question and depends on the examination and the detector features. Evaluation of AEC curves and distribution of receptor entrance doses enables to establish baseline for the future dose vs digital image quality optimisation methods.

Additional remarks It is not probably justified to differ suspension levels for X-ray equipment depending on the year of manufacture because of absence of such references (no on year 2006 nor 2012). EN 60601-1-3:2008 states: This European Standard supersedes EN 60601-1-3:1994. However, EN 60601-1-3:1994 remains valid until all the Parts 2 that are used in conjunction with it have been withdrawn. The older particular standards (Part 2 standards) for radiography and fluoroscopy, CT and mammo have been withdrawn already. For dental radiography the particular standards are not published yet (draft 2011) but the following statement from EN 60601-1-3:2008 applies for them: when Part 1-3 is used for appliances not covered by a Part 2, EN 60601-1-3:1994 is not to be used after 2009-09-12. Radiation quality criterion (expressed as minimum HVL) is a general radiation protection requirement and should be applicable to all X-ray diagnostic equipment in clinical use, independent on the year of manufacture. In the same time it is quite easy to add filtration, if needed, even for older types of X-ray equipment.

References IEC (International Electrotechnical Commission) (1994) Medical electrical equipment - Part 1-3: General requirements for safety - Collateral Standard: General requirements for radiation protection in diagnostic X-ray equipment. Edition 1.0. IEC 60601-1-3. IEC, Geneva, Switzerland. IEC (International Electrotechnical Commission) (2008) Medical electrical equipment - Part 1-3: General requirements for basic safety and essential performance - Collateral Standard: Radiation protection in diagnostic X-ray equipment. Edition 2.0 IEC 60601-1-3. IEC, Geneva, Switzerland. EC (European Commission) (1997) Criteria for acceptability of radiological (including radiotherapy) and nuclear medicine installations. Radiation Protection 91. EC, Luxembourg. EC (European Commission) (2009). Radiation criteria for acceptability of medical radiological equipment used in diagnostic radiology, nuclear medicine and radiotherapy. Final draft amendedv1.4-091001. EC, Luxembourg. IPEM (lnstitute of Physics and Engineering in Medicine) (2010) Measurement of the performance characteristics of diagnostic X-ray systems: digital imaging systems. IPEM Report 32, Part VII. IPEM, York, UK. Walsh, C., Gorman, D., Byrne P., Larkin, A., Dowling, A. Malone, J. F. (2008) Quality assurance of computed and digital radiography systems, Radiation Protection Dosimetry, Vol. 129, No. 1-3, 271-275. Aichinger, H., Dierker, J., Joite-Barfuss, S., Sabel, M. (2004) Radiation exposure and image quality in X-ray diagnostic radiology. Springer-Verlag, Berlin. Sandborg M., Carlsson G. A., Persliden J. and Dance D. R. Comparison of different materials for test phantoms in diagnostic radiology. Radiation Protection Dosimetry Vol. 49, No 1/3, 345-347 (1993)

Thank you!