ANALYSIS OF 3RD OCTAVE BAND GROUND MOTIONS TRANSMISSION IN SYNCHROTRON RADIATION FACILITY SOLARIS Daniel Ziemianski, Marek Kozien

Similar documents
Herwig Schopper CERN 1211 Geneva 23, Switzerland. Introduction

VIBRATION MEASUREMENTS IN THE KEKB TUNNEL. Mika Masuzawa, Yasunobu Ohsawa, Ryuhei Sugahara and Hiroshi Yamaoka. KEK, OHO 1-1 Tsukuba, Japan

Coherent Synchrotron Radiation in the ANKA Storage Ring

INDUSTRIAL VIBRATION SENSOR SELECTION MADE EASY

Vibration studies of a superconducting accelerating

Response spectrum Time history Power Spectral Density, PSD

New Long Stroke Vibration Shaker Design using Linear Motor Technology

RF System Models and Longitudinal Beam Dynamics

Tutorial on Design of RF system for Indus Accelerator. Maherdra Lad Head, Radio Frequency Systems Division RRCAT, Indore

ITk silicon strips detector test beam at DESY

The impedance budget of the CERN Proton Synchrotron (PS)

FLASH at DESY. FLASH. Free-Electron Laser in Hamburg. The first soft X-ray FEL operating two undulator beamlines simultaneously

VIBRATING WIRE SENSORS FOR BEAM INSTRUMENTATION Suren Arutunian

Demonstration of exponential growth and saturation at VUV wavelengths at the TESLA Test Facility Free-Electron Laser. P. Castro for the TTF-FEL team

3 General layout of the XFEL Facility

Commissioning of National Synchrotron Light Source-II (NSLS-II) Fast Orbit Feedback System

HIGH POWER COUPLER FOR THE TESLA TEST FACILITY

SURVEY AND ALIGNMENT FOR THE SWISS LIGHT SOURCE

GROUND MOTION IN THE INTERACTION. ensured that the final focus quadrupoles on both. rms amplitudes higher than some fraction of the

BEPCII-THE SECOND PHASE CONSTRUCTION OF BEIJING ELECTRON POSITRON COLLIDER

System Integration of the TPS. J.R. Chen NSRRC, Hsinchu

MEASUREMENT OF BEAM LOSSES USING OPTICAL FIBRES AT THE AUSTRALIAN SYNCHROTRON

An Overview of MAX IV Insertion Devices & Magnetic Measurement System. Hamed Tarawneh On behalf of Insertion Devices Team

THz Pump Beam for LCLS. Henrik Loos. LCLS Hard X-Ray Upgrade Workshop July 29-31, 2009

Kistler portable triaxial Force Plate

Circumference 187 m (bending radius = 8.66 m)

SPEAR 3 - THE FIRST YEAR OF OPERATION*

Dytran Instruments, Inc. 1

Analysis of Electrical Noise in Piezoelectric Sensors

Emilia Cruz. September 21, 2015

Experiences of the QSBPM System on MAX II

ATA s Nanoradian-Class Rotational Sensors. 10 November 2009

DEVELOPMENT OF CAPACITIVE LINEAR-CUT BEAM POSITION MONITOR FOR HEAVY-ION SYNCHROTRON OF KHIMA PROJECT

Piezoelectric accelerometer design. Piezoelectric transducers Quartz and piezoceramics Mechanical design Charge amplification Design trade-offs

Bioimaging of cells and tissues using accelerator-based sources

CEBAF Overview June 4, 2010

Principles of Vibration Measurement and Analysis. Dr. Colin Novak, P.Eng July 29, 2015

Embedded Surface Mount Triaxial Accelerometer

Short-Pulse X-ray at the Advanced Photon Source Overview

EFFECTS OF ACCELEROMETER MOUNTING METHODS ON QUALITY OF MEASURED FRF S

A high resolution bunch arrival time monitor system for FLASH / XFEL

A Synchrotron Phase Detector for the Fermilab Booster

Nonintercepting Diagnostics for Transverse Beam Properties: from Rings to ERLs

The VIRGO suspensions

Superconducting RF System. Heung-Sik Kang

Third Harmonic Superconducting passive cavities in ELETTRA and SLS

VEPP-2000 COLLIDER CONTROL SYSTEM*

A STUDY OF BEAM BREAKUP IN 12 GeV UPGRADE WITH DOUBLE BEND ACHROMAT ARC OPTICS Ilkyoung Shin and Byung C. Yunn JLAB-TN November 1, 2008

Operation of a Single Pass, Bunch-by-bunch x-ray Beam Size Monitor for the CESR Test Accelerator Research Program. October 3, 2012

Examination of Microphonic Effects in SRF Cavities

THE CRYOGENIC SYSTEM OF TESLA

Using Higher Order Modes in the Superconducting TESLA Cavities for Diagnostics at DESY

Physics Requirements Document Document Title: SCRF 1.3 GHz Cryomodule Document Number: LCLSII-4.1-PR-0146-R0 Page 1 of 7

3.0 Apparatus. 3.1 Excitation System

Lattice Design for PRISM-FFAG. A. Sato Osaka University for the PRISM working group

Proposal of test setup

Maurizio Vretenar Linac4 Project Leader EuCARD-2 Coordinator

Clarification of the Effect of High-Speed Train Induced Vibrations on a Railway Steel Box Girder Bridge Using Laser Doppler Vibrometer

Radiated EMI Recognition and Identification from PCB Configuration Using Neural Network

Preliminary study of the vibration displacement measurement by using strain gauge

MXD7210GL/HL/ML/NL. Low Cost, Low Noise ±10 g Dual Axis Accelerometer with Digital Outputs

(i) Sine sweep (ii) Sine beat (iii) Time history (iv) Continuous sine

Results of Vibration Study for LCLS-II Construction in the Research Yard 1

LHC TRANSVERSE FEEDBACK SYSTEM: FIRST RESULTS OF COMMISSIONING. V.M. Zhabitsky XXI Russian Particle Accelerator Conference

Piezoelectric multilayer triaxial accelerometer

Acceleration of High-Intensity Protons in the J-PARC Synchrotrons. KEK/J-PARC M. Yoshii

CONSIDERATIONS FOR ACCELEROMETER MOUNTING ON MOTORS

GENERAL PURPOSE PIEZOELECTRIC LOAD CELL

Proton Induced Thermal Stress Wave Measurements in. Solid Targets

Generic noise criterion curves for sensitive equipment

Angular Drift of CrystalTech (1064nm, 80MHz) AOMs due to Thermal Transients. Alex Piggott

FFT 1 /n octave analysis wavelet

Electron cloud effects, codes & simulations. K. Ohmi (KEK) ICAP12, Aug, 2012 Rostock

Activities on Beam Orbit Stabilization at BESSY II

Recent ground vibration measurements at CERN (Surface and underground)

S.M. Lidia, G. Bazouin, P.A. Seidl Accelerator and Fusion Research Division Lawrence Berkeley National Laboratory Berkeley, CA USA

VIBRATION SENSOR WITH TWO-WIRE INTERFACE AND BIAS USED FOR MEASURE TEMPERATURE

NINTH INTERNATIONAL CONGRESS ON SOUND AND VIBRATION, ICSV9 ACTIVE VIBRATION ISOLATION OF DIESEL ENGINES IN SHIPS

A Prototype Wire Position Monitoring System

Initial Beam Phasing of the SRF Cavities in LCLS-II

Strategy for the engineering integration of the ESS accelerator

PERFORMANCE OF THE CMS ECAL LASER MONITORING SOURCE IN THE TEST BEAM

Grounding for EMC at the European XFEL

Introduction to LIVM Accelerometers

Effects of Intensity and Position Modulation On Switched Electrode Electronics Beam Position Monitor Systems at Jefferson Lab*

Control Servo Design for Inverted Pendulum

Dynamic Angle Estimation

AN5E Application Note

Earthquake Resistance Test Specifications for Communications Equipment

Undulator K-Parameter Measurements at LCLS

Magnet Stability (and Reproducibility).

Active Stabilization of a Mechanical Structure

A SIMPLE METHOD TO COMPARE THE SENSITIVITY OF DIFFERENT AE SENSORS FOR TANK FLOOR TESTING

12 GeV Upgrade Project DESIGN SOLUTIONS DOCUMENT. Upgrade Hall A

Anthony Chu. Basic Accelerometer types There are two classes of accelerometer in general: AC-response DC-response

New Tracking Gantry-Synchrotron Idea. G H Rees, ASTeC, RAL, U.K,

LCLS-II TN Vibration measurements across the SLAC site

Beam Diagnostics, Low Level RF and Feedback for Room Temperature FELs. Josef Frisch Pohang, March 14, 2011

Design of the magnets for the MAX IV project. Martin Johansson, Beam Dynamics meets Magnets-II workshop, Bad Zurzach, Dec.

Laboratory Test of Vibration of Micro/Nano Satellite for Environment Test Standardization

Transcription:

ANALYSIS OF 3RD OCTAVE BAND GROUND MOTIONS TRANSMISSION IN SYNCHROTRON RADIATION FACILITY SOLARIS Daniel Ziemianski, Marek Kozien Cracow University of Technology, Institute of Applied Mechanics, al. Jana Pawla II 37, 31-864 Cracow, PL email: dziemianski@pk.edu.pl Marcin Nowak National Synchrotron Radiation Centre SOLARIS, ul. Czerwone Maki 98, 30-392 Cracow, PL, The purpose of the measurement is to determine the levels of Ground Motion (GM) in experimental hall in synchrotron radiation facility SOLARIS. Ground motion in various areas, such as experimental hall, and a Linear Accelerator (LINAC) tunnel, are presented. Machines with high accuracy that are sensitive to ground vibrations are generally designed using crude assumptions on the dynamic properties of the floor where they are placed. This also holds for the prediction of machine accuracy in the design phase, based on expected or measured ground vibration spectra. It is well known that floor vibrations are one of the main disturbance sources for high precision equipment. Especially for equipment in the synchrotron radiation facility, like SOLARIS, but also in the other particle physics laboratory, the impact of floor vibrations on the machine accuracy is generally determining the dynamic architecture of the system. It is said that the peak around 0.3[Hz] in the GM spectra is caused by ocean waves. This phenomenon is shown by comparing GM spectra for the concrete floor in an experimental hall of SOLARIS with wave height of the ocean. Also 3rd Octave Band Analysis it was presented and corresponding Vibration Criteria were presented and discussed. Keywords: Octave Band, Ground Motions, Synchrotron 1. Introduction The first polish third generation light source Solaris is based on the linear accelerator and the storage ring connected with dog-leg transfer line. The linac provides the beam with maximum 600 [MeV] energy which can be injected to the storage ring and ramped up to the nominal energy 1.5 [GeV]. A storage ring layout consists of 12 novel highly integrated Double Bend Achromat (DBA) magnets designed by MAX-IV Laboratory in Sweden. One year period of the machine assembling has started in May 2014. Then the commissioning phase started and the process of finetuning for reaching the designed parameters is still ongoing. Most important parameters of this 1.5 [GeV] storagering are presented in Table 1. A detailed description of the machine and the layout can be found in [1 4]. Recent optimization of linear injector and storage ring allowed to reach 600 ma of injected beam at 525 [MeV] energy stored in the ring and 400 [ma] beam ramped to the nominal 1.5 [GeV] energy. Machine optics, was corrected closely to the designed values. Sufficiently stable beam along with good reproducibility of beam parameters from injection to injection allowed to start the commissioning of the first beam line UARPES. 1

Table 1: Solaris Storage Ring Parameters Parameter Value Energy 1.5 GeV Beam current 500 ma Circumference 96 m Number of bending magnets 12 Main RF frequency 99.931 MHz Number of bunches 32 Horizontal emittance (bare lattice) 6 nm rad Tune Q x, Q y 11.22, 3.15 Natural chromaticity ξ x, ξ y -22.96, -17.4 Corrected chromaticity ξ x, ξ y +1, +1 Beam size (straight section) σ x, σ y 184 μm, 13 μm Beam size (dipole) σ x, σ y 44 μm, 30 μm Total lifetime 13 h 2. Instrumentation Piezoelectric sensors use Newton s Second Law and the piezoelectric effect to measure the acceleration of a mass. A piezoelectric accelerometer contains a seismic mass mounted so that the force applied to the mass by movement of the housing squeezes or stretches a natural quartz crystal or manmade piezoelectric ceramic measuring element. The pressure on the measuring element produces an electrical charge within the material that is proportional to the force applied the piezoelectric effect. This force, in turn, is proportional to acceleration. IEPE Integrated Electronics Piezoelectric) sensors have built-in signal conditioning circuits having low impedance output electronics, compatible with a two-wire constant current supply providing a DC voltage bias. IEPE sensors are very popular in numerous industrial applications except those which special requirements such as static (zero [Hz]) sensing, high temperature applications, or process control applications. Four mono-axial seismic accelerometers were used to measure the ground motion, with a sensitivity of about 1000 [mv/m/s2].the mass of each transducer is 635 [g] (PCB 393831). The usable frequency range of a sensor is from about 0.1 [Hz] up to 1/3 to 1/2 of the natural resonance frequency (700 [Hz]). In order to increase the frequency, either the stiffness should be increased or the mass decreased. Since decreasing mass also decreases sensitivity, increasing stiffness is preferable. In most cases, a combination of the two approaches is required. The accelerometers were glued on the structure with wax. The natural frequency of an unmounted sensor is different from that of a mounted sensor, because the mounted sensor has a stiffness determined by the stiffness of the structure in which it is mounted. Attaching the sensor to a structure can lower the natural frequency by a substantial amount. Connectors and cables are often the weak link in an instrumentation chain. The triboelectric effect is the generation of an error signal in a charge output sensor whenever its attached cable is physically moved. The only effective way to eliminate this error signal is to clamp the cable as close to the accelerometer as possible. The cables were securely fastened to the structure to minimize cable whip as well as the connector strain. 2 ICSV24, London, 23-27 July 2017

3. Results While spectral transforms like FFT provide constant absolute bandwidth (i.e. the bins are equally spaced in terms of Hz), 1/n octave analysis provides an analysis bandwidth which is proportional to the centre of each band throughout the analysed frequency range. This is called constant relative bandwidth. 3.1 Vibration criteria for work with sensitive equipment The VC Curves Specification define the vibration criteria which need to be met for certain types of sensitive equipment to be used in the specified area. The comparison of the data acquired from the measurement and the curves is typically done for the RMS value of velocity integrated over the 1/3 octave bands within the frequency range from 1 to 80 [Hz]. The highest VRMS over this range and calculated this way, determines the actual VC range for the tested area. Table 1 shows the velocity values which need to be met for certain vibration criteria. The same information is shown on the graph above. The ground vibration measurement is performed for all three directions (2 horizontal, 1 vertical) and the VC specification needs to be used for each one of them. Typically for high precision equipment, vertical V RMS values should be below the VC-F curve, while the horizontal ones below VC-G. 3.2 Speed RMS [μm/s per 1/3 octave band] In the 1/n octave analysis, the signal to be analysed is split into partial signals by a digital filter bank before the vibration level is determined. The filter bank consists of several filters connected in parallel, each with a bandwidth of 1/n octave. An octave filter is a filter whose upper cut-off frequency is twice the lower cut-off frequency, whereas 3rd octave filters further subdivide each octave band into three parts and so on. This means that octave filters or 1/n octave filters don't have a constant absolute bandwidth, but a constant relative bandwidth, i.e. the frequency bands are equidistant on a logarithmic frequency scale. Figure 1: Speed RMS [μm/s per 1/3 octave band] - vertical axis ICSV24, London, 23-27 July 2017 3

Figure 2: 3.2 Speed RMS [μm/s per 1/3 octave band] - horizontal axis NS Figure 3: 3.2 Speed RMS [μm/s per 1/3 octave band] - horizontal axis EW 4 ICSV24, London, 23-27 July 2017

3.3 Evaluate the power spectral density The first analysis was to evaluate the power spectral density for each direction of sensors and event. The ground motions at several locations measured in are plotted in (Figure 4). These locations are divided into three groups. The first group (red lines) is the synchrotron bunker, the second group (green lines) is the experimental hall and the third group (black lines) is the tunnel linac. SOLARIS Figure 4: Power Spectrum Density Integrated is used to sum up the total vibration of the PSD. It gives the RMS (Root Mean Square) value of the total vibration The integrated RMS ground motion in vertical direction ( the average of the measurements) are show and compared with the results of similar research centres in the world. The (Table 2.) shows the results at 1 [Hz]. RMS Integraded values Table 2: Integrated RMS value at 1 [Hz] @1 Hz [nm] SOLARIS 4,8 CLEX experiment 13,3 AEGIS experiment 13,0 PSI particle accelerator 11,8 CMS experiment 6,8 CesrTA particle accelerator 3,8 LSS4 Caver-SPS Accelerator 4,1 TT1 tunnel (ISR) 2,3 LHC tunnel (DCUM584) 1,9 ICSV24, London, 23-27 July 2017 5

4. Conclusions The ground motion spectrum can be separated in two main regions. The low frequency region below about 2 [Hz] which is dominated by natural motions sources like ocean waves and earth tides, and the high frequency region above 2 [Hz] which is dominated by cultural motion sources like traffic, industry, transformers, pumps and people walking. The power spectral density graphs above clearly shows the distinction between these two regions. In the low frequency region, all locations have a similar ground motion dominated by well correlated natural motions. In higher frequencies the ground motion is dominated by the local cultural noise, and the highest ground motions occur at locations with high cultural activity. From the vibration measurements performed in the laboratory SOLARIS and the following 3rd Octave Band Analysis in can be seen that vibration levels in the specified locations are below the VC-G curves at all times in which the measurements were performed and within the whole relevant frequency range (1-80 [Hz]). This in turn means that the specified location is adequate for the purpose of operating a high precision equipment such as a parts of detector experimental lines in question. REFERENCES 1 MAX IV Detailed Design Report, http://www.maxlab.lu.se/maxlab/max4/index.html. 2 M. Eriksson, et al., The MAX IV Synchrotron Light Source,, conference IPAC2011, San Sebastián, Spain. 3 S. C. Leemann, Particle Accelerator Conference, New York, USA, (2011). 4 Ady M., Hermann M., Kersevan R., Vandoni G., Ziemianski D., Leak propagation dynamics for the HIE-ISOLDE superconducting linac, IPAC2014: proceedings of the 5th International Particle Accelerator Conference, June 15-20, 2014, Dresden, Germany. 6 ICSV24, London, 23-27 July 2017