Investigate in magnetic micro and nano structures by Magnetic Force Microscopy (MFM)

Similar documents
Basic methods in imaging of micro and nano structures with atomic force microscopy (AFM)

- Near Field Scanning Optical Microscopy - Electrostatic Force Microscopy - Magnetic Force Microscopy

Standard Operating Procedure of Atomic Force Microscope (Anasys afm+)

UNIVERSITY OF WATERLOO Physics 360/460 Experiment #2 ATOMIC FORCE MICROSCOPY

Atomic resolution of the graphite surface by STM

Optical Microscope. Active anti-vibration table. Mechanical Head. Computer and Software. Acoustic/Electrical Shield Enclosure

Lecture 20: Optical Tools for MEMS Imaging

Equipment for Attenuation and velocity of ultrasound in solid state materials (transmission), experimental set-up

ATOMIC FORCE MICROSCOPY

Attenuation and velocity of ultrasound in solid state materials (transmission)

INDIAN INSTITUTE OF TECHNOLOGY BOMBAY

Measurement of Microscopic Three-dimensional Profiles with High Accuracy and Simple Operation

Outline: Introduction: What is SPM, history STM AFM Image treatment Advanced SPM techniques Applications in semiconductor research and industry

Standard Operating Procedure

Akiyama-Probe (A-Probe) guide

Nanosurf Nanite. Automated AFM for Industry & Research.

A scanning tunneling microscopy based potentiometry technique and its application to the local sensing of the spin Hall effect

Cutting-edge Atomic Force Microscopy techniques for large and multiple samples

Akiyama-Probe (A-Probe) guide

Nanosurf easyscan 2 FlexAFM

Prepare Sample 3.1. Place Sample in Stage. Replace Probe (optional) Align Laser 3.2. Probe Approach 3.3. Optimize Feedback 3.4. Scan Sample 3.

Comparison of resolution specifications for micro- and nanometer measurement techniques

MAGNETOSCOP Measurement of magnetic field strengths in the range 0.1 nanotesla to 1 millitesla

Unit-25 Scanning Tunneling Microscope (STM)

SUPPLEMENTARY INFORMATION

Bruker Dimension Icon AFM Quick User s Guide

Electric polarization properties of single bacteria measured with electrostatic force microscopy

Keysight Technologies Scanning Microwave Microscope Mode. Application Note

Indentation Cantilevers

SENSOR+TEST Conference SENSOR 2009 Proceedings II

Basic methods in imaging of micro and nano structures with atomic force microscopy (AFM) Related Topics. Principle. Equipment

Microscopic Structures

Kirchhoff s laws, induction law, Maxwell equations, current, voltage, resistance, parallel connection, series connection, potentiometer

Keysight Technologies Using Non-Contact AFM to Image Liquid Topographies. Application Note

State of the Art Room Temperature Scanning Hall Probe Microscopy using High Performance micro-hall Probes

Profile Measurement of Resist Surface Using Multi-Array-Probe System

ATOMIC FORCE MICROSCOPY

Nanonics Systems are the Only SPMs that Allow for On-line Integration with Standard MicroRaman Geometries

; A=4π(2m) 1/2 /h. exp (Fowler Nordheim Eq.) 2 const

Advanced Nanoscale Metrology with AFM

Atomic Force Microscopy (Bruker MultiMode Nanoscope IIIA)

Bruker Dimension Icon AFM Quick User s Guide

Lateral Force: F L = k L * x

Magnetic induction with Cobra3

attosnom I: Topography and Force Images NANOSCOPY APPLICATION NOTE M06 RELATED PRODUCTS G

Alignment of the camera

University of MN, Minnesota Nano Center Standard Operating Procedure

Magnetic induction with Cobra3

IMAGING P-N JUNCTIONS BY SCANNING NEAR-FIELD OPTICAL, ATOMIC FORCE AND ELECTRICAL CONTRAST MICROSCOPY. G. Tallarida Laboratorio MDM-INFM

Nanovie. Scanning Tunnelling Microscope

Controller Design for Z Axis Movement of STM Using SPM Control Software

Instructions for easyscan Atomic Force Microscope

MEMS for RF, Micro Optics and Scanning Probe Nanotechnology Applications

Specifying and Measuring Nanometer Surface Properties. Alson E. Hatheway

A Project Report Submitted to the Faculty of the Graduate School of the University of Minnesota By

Figure for the aim4np Report

Indian Institute of Technology Bombay

Nanoscale Material Characterization with Differential Interferometric Atomic Force Microscopy

Distinguishing Between Mechanical and Electrostatic. Interaction in Single-Pass Multifrequency Electrostatic Force

RHK Technology. Application Note: Kelvin Probe Force Microscopy with the RHK R9. ω mod allows to fully nullify any contact potential difference

Park NX20 The leading nano metrology tool for failure analysis and large sample research.

Fabrication of Probes for High Resolution Optical Microscopy

ezafm OPERATING MANUAL

Characterisation of the Montana Instruments Cryostation C2 for low temperature Magneto-Optical Kerr Effect measurements using the NanoMOKE 3

Manufacturing Metrology Team

Optotop. 3D Topography. Roughness (Ra opt, Rq opt, and Rz opt) Height Distribution. Porosity Distribution. Effective Contact Area

SPM The Industry s Performance Leader High Resolution Closed-loop System Fast, Easy Tip & Sample Exchange Versatility and Value Powerful Research

Electrical Properties of Chicken Herpes Virus Based on Impedance Analysis using Atomic Force Microscopy

Study of shear force as a distance regulation mechanism for scanning near-field optical microscopy

Long-distance propagation of short-wavelength spin waves. Liu et al.

The Most Accurate Atomic Force Microscope. Park NX20 The leading nano metrology tool for failure analysis and large sample research.

Imaging Carbon Nanotubes Magdalena Preciado López, David Zahora, Monica Plisch

NanoFocus Inc. Next Generation Scanning Probe Technology. Tel : Fax:

Alejandro Mendez, Ph.D. President & CEO Mendezized Metals Corporation

Physics Faculty Publications and Presentations

Transmission Electron Microscopy 9. The Instrument. Outline

High Resolution Near-Field Measurements of Microwave

A New Profile Measurement Method for Thin Film Surface

Park NX-Hivac: Phase-lock Loop for Frequency Modulation Non-Contact AFM

Standing waves in the microwave range

The World s Most Accurate AFM System. Park NX-3DM Innovation and Efficiency for 3D Metrology.

Constant Frequency / Lock-In (AM-AFM) Constant Excitation (FM-AFM) Constant Amplitude (FM-AFM)

Proposal. Design of a Scanning Tunneling Microscope

NSOM (SNOM) Overview

nanovea.com PROFILOMETERS 3D Non Contact Metrology

PH880 Topics in Physics

Park NX-Hivac The world s most accurate and easy to use high vacuum AFM for failure analysis.

Self-navigation of STM tip toward a micron sized sample

Spotlight 150 and 200 FT-IR Microscopy Systems

Enabling Areal Density Growth

Nanomechanical Mapping of a High Curvature Polymer Brush Grafted

Nanosurf easyscan 2 Your Modular Scanning Probe Microscopy System

attocfm I for Surface Quality Inspection NANOSCOPY APPLICATION NOTE M01 RELATED PRODUCTS G

Analytical analysis of modulated signal in apertureless scanning near-field optical microscopy C. H. Chuang and Y. L. Lo *

AFM of High-Profile Surfaces

SOLVER NANO Scanning Probe Microscope

Mapping piezoelectric response in nanomaterials using a dedicated non-destructive scanning probe technique. Supporting Information

Name: Lab Partner: Section: The purpose of this lab is to study induction. Faraday s law of induction and Lenz s law will be explored. B = B A (8.

Introduction of New Products

The solar cell as a diode

Transcription:

Investigate in magnetic micro and nano 5.3.85- Related Topics Magnetic Forces, Magnetic Force Microscopy (MFM), phase contrast imaging, vibration amplitude, resonance shift, force Principle Caution! - Set up your system on a very steady table - Do your experiments in a calm, vibrational free environment Magnet interaction between the tip and sample is used to image magnetic structures with Atomic Force Microscopy (AFM). The principle of magnetic force microscopy can be shown scanning an backup tape (DAT). For this a magnetic tip is used and is magnetized with an ultra-magnet along a certain direction to be sensitive on different direction of magnetic field lines. After imaging the topography in contact mode the cantilever is retracted a few 10 nm. At this distance magnetic force dominates the interaction between cantilever and tape. Scanning the same area it leads to stripe-like structures showing the magnetization of the tape. Also magnetic structures for data storage on floppy disks (ZIP or others) or hard disk down to a structure size of a few ten nanometer can be imaged and analyzed. Equipment 1 Compact Atomic Force Microscope (AFM) 097-99 1 Material analysis upgrade 09701- Additionally needed 1 PC, Windows XP or higher Double sided scotch tape 1 Strong permanent magnet (ultra magnet) 1 Small compass Other magnetic storage samples (DDS tape, ZIP disk, floppy disk, older hard disks) Figure 1: PHYWE Atomic Force Microscope www.phywe.com P25385 PHYWE Systeme GmbH & Co. KG All rights reserved 1

5.3.85- Investigate in magnetic micro and nano PHYWE Compact AFM set Fig. 2: Content of 097-99. Tasks 1. Set-up the microscope, magnetize the magnetic coated tip along a certain direction and approach the tip to the sample in phase imaging mode. Take a topography image. 2. Retract the tip a few 10 nm to do a MFM measurement for different distances. Magnetize the tip to another direction and compare and interpret the results. 3. Image the magnetic structures of different samples, e.g. floppy disk, ZIP floppy disk, and hard disk. Set-up and Procedure Task 1 and 2: The general handling and preparation of the AFM device incl. cantilever is described in the operating manual. A short introduction is also given in the experimental guide P25380 "Basic methods in micro and nano imaging of structures with AFM". The aim of Magnetic force microscopy (MFM) is to image a spatial distribution of a magnetic field. For MFM measurements magnetic cantilevers like Budget Sensors Multi75M-G are required. These are Silicon cantilever which are coated with a ferromagnetic material e.g. Iron. The cantilevers are usually already magnetized in the pointing direction of the tip. 2 PHYWE Systeme GmbH & Co. KG All rights reserved P25385

Investigate in magnetic micro and nano 5.3.85- Typical samples for MFM measurements are different digital data storage (DDS) devices like magnetic tapes, hard drive disks (HDD) or magneto-optical disks. However, basically any sample showing magnetic features can be imaged in MFM, as long as it fits inside the AFM and its surface is flat enough. Magnetic tapes can be preared by cutting a piece and mount with the shiny side pointing upwards on a spare sample holder using double-sided scotch tape. Pieces of disks can be mounted in the same way. The following examples show different samples of magnetic tape and hard drive disk samples which show magnetic features of different spatial sizes. Figure 3 1 shows example fotographs of the digital data storage devices used for the following scans. Figure 3 Fotographs of a HP DDS Tape (left), Exabyte Data Cartridge (center) and an opened HDD (right). Imaging Investigate the topography of the samples in phase imaging mode. As shown in figure 4 the two tapes show similar topographies, showing the magnetic film deposited on the tape. In contrast the sample of HDD shows lengthy features which come from its fabrication or the magnetic reading head scratching over its surface. However, no magnetic bits and bytes can be imaged in this way. Figure 4 Topography of the Exabyte 8MM Data Cartridge-tape (left), HP DDS Tape (center) and a HP 2.5 40 GB HDD. NOTE It is recommended to use a standart cantilever instead of a magnetic cantilever for topographic imaging in order not to damage the magnetic coating when imaging. 1 http://www.databazaar.com www.phywe.com P25385 PHYWE Systeme GmbH & Co. KG All rights reserved 3

5.3.85- Investigate in magnetic micro and nano Figure 5 (left) Forces in MFM: Force gradient of a point dipole. (right) MFM imaging in dynamic mode Theoretical Background When performing MFM measurements the tip shouldn't come into mechanical contact with the surface as the influence of the mechanical contact is usually stronger than the magnetic effects. Therefore during MFM the tip is scanned at a defined distance above the sample. The magnetic force which is measured in MFM can be approximated by the force gradient of a point dipole seen by the magnetic tip (see figure 5 (left)), where is the unit vector normal to the cantilever plane and the magnetic force When the AFM is operated in dynamic mode, e. g. when the tip is vibrated in -direction (vertically to the sample) during the scanning, the influence of the magnetic force in this direction is dominant and thus determines the contrast the MFM image (see figure 5 (right)). The magnetic field influences the spring constant of the cantilever probe in the following way: where is the original spring constant. Therefore a force gradient pointing away from the tip decreases the spring constant and a force gradient pointing towards the tip increases the spring constant. Like in dynamic mode, this can be related to a decrease in height of the topography for the lower spring constant and an increase in height for a higher spring constant. The effect of the magnetic force on the amplitude and phase signal are also comparable to the dynamic mode topography imaging. For the amplitude the change of the spring constant induces a shift of the resonance peak and the difference in the amplitude is thereby related to the magnitude of the magnetic field (see figure 6 (left)). The same shift is also observed in the phase signal (see figure 6 (right)). The advantage of the phase signal over the amplitude signal is its local linearity and the thus simpler and linear relation between the force and the phase shift: where is the -factor of the cantilever probe. For this reason, we will use phase images for imaging the magnetic samples structures. 4 PHYWE Systeme GmbH & Co. KG All rights reserved P25385

Investigate in magnetic micro and nano 5.3.85- Figure 6 Amplitude (left) and phase shift (right): The magnetic tip sample interaction influences the spring constant of the cantulever probe and provokes a shift in resonance frequency. The resonance shift can be detected in the amplitude signal (left) and phase signal (right). Technical Implementation The most important change compared to the standard dynamic mode is that the probe needs to be scanned at a defined distance above the surface. This is realized by performing measurements in constant height mode as shown in figure 7. In this mode, the feedback loop is switched off. Figure 7 Linear scan at a given height over the sample surface Preparing the Measurement 1. Mount the MFM Probe on the AFM. Usually, new MFM cantilevers are magnetised. However, you might neew need to magnetise the tip with a strong permanent magnet as shown in figure Magnetisation of the Tip (figure 8). 2. Determine the north-pole of the permanent magnet (for example using a compass). 3. Mark the north-pole, in order to be able to magnetise the tip in a reproducible direction. 4. Bring the pole of the magnet to within 1-2 mm distance of the tip. 5. Move the magnet away from the tip in a direction perpendicular to the cantilever surface. 6. Select cantilever type in the software. If the present cantilever type is not there create a new cantilever type according to the specifications of the manufacturer. 7. If not done yet, select the phase contrast imaging mode. www.phywe.com P25385 PHYWE Systeme GmbH & Co. KG All rights reserved 5

5.3.85- Investigate in magnetic micro and nano Figure 8 Magnetization of the tip Approaching the Surface Set the free vibration amplitude to 2mV. Clear the check box Auto. start imaging in the SPM Parameter -> Approach panel as shown in figure Auto Start (figure 9). A direct start could harm the tip. Approach the surface. Figure 9 Auto start imaging CAUTION Only change the amplitude calibration value when the tip is witdrawn far from the sample. Otherwise you could damage the tip. Calibrating the Vibration Amplitude The calibration value of the vibration amplitude depends on the cantilever type, because of changes in the length and resonance frequency of the cantilever. As it is important to know the vibration amplitude it needs to be calibrated. This needs to be done only when a cantilever was changed. Use the following procedure to calibrate the vibration amplitude: Measure a vibration amplitude vs. distance curve in spectroscopy mode as shown in figure Amplitude Calibration (figure 10). Determine the slope with the measure length tool and calculate the calibration value. Withdraw the tip from the surface again. Set the free vibration amplitude to a value in order of the magnetic feature size you intend to measure. The feature size can be determined by using the Measure Length tool. An example is given in Task 3. Starting the measurement Ensure that the check box Auto. start scanning is still cleared. Approach the surface. 6 PHYWE Systeme GmbH & Co. KG All rights reserved P25385

Investigate in magnetic micro and nano 5.3.85- Figure 10 Amplitude-distance curve with indicated measure length tool. Auto adjust the scanning slope by pressing the Adjust slope button in the SPM Parameter -> Imaging panel (figure 11). This will make sure the tip is scanned parallel to the surface and will not crash. Activate the constant height mode by checking the box Enable Constant Height Mode also found in the SPM Parameter -> Imaging panel (figure 12). Start at about and decrease the distance manually until you reach the optimum contrast, an example is given in Task 3. The scanning time per line can be set to low values of as the tip is not touching the sample and therefore will not be harmed by a fast scanning speed. Figure 11 Auto start imaging Figure 12 Enable constant heigt mode CAUTION In the constant height box a negative value means a displacement away from the surface. A positive value will crash the tip into the sample. www.phywe.com P25385 PHYWE Systeme GmbH & Co. KG All rights reserved 7

5.3.85- Investigate in magnetic micro and nano Task 3: Image the magnetic micro structures of different samples In the following the samples of magnetic tapes and a hard drive disk are imaged in MFM. Selecting the Vibration Amplitude The magnitude of the MFM (phase) signal, the phase noise and the resolution of the MFM signal depend on the vibration amplitude. When the amplitude is too small, the noise is too high, when the amplitude is too large, the resolution becomes low, and ultimately the signal will also become too small. It can be shown that the vibration amplitude should be of a similar magnitude as the smallest details you wish to see. The accurate vibration amplitude can be calculated by multiplying the previously measured value from the spectroscopy with the mean feature size. In the measurement of a magnetic tape (Exabyte 8MM Data Cartridge) shown in figure 13, the aproximate feature size is, the amplitude calibration gave, therefore the corresponding recommended vibration amplitude would be However, such large amplitudes exceed the maximum possible virbration amplitude of the cantilever. In this case the the largest possible amplitude should be used. Nevertheless, the results show a good resolution. Figure 13 MFM measurements of a Exabyte 8MM Data Cartridge-tape using different scan sizes. The mean feature size is. Selecting the Relative Tip-Position In Order to achieve the best contrast, the distance to the sample must be as small as possible. At a certain point, however the tip will touch the surface and will be harmed when the contact is to strong. A safe procedure is to start from a safe tip to sample distance (e.g. ) and reduce it step wise until the tip begins to feel the surface. Usually, this is the case at a distance well below. Immediately increase the distance until the topography features disappear from the phase signal. 8 PHYWE Systeme GmbH & Co. KG All rights reserved P25385

Investigate in magnetic micro and nano 5.3.85- Figure 14 shows the measurement of another magnetic tape (Hewlett Packard Digital Data Storage 4GB). Here we see a different magnetic stucture with a mean feature size of.. Furthermore we see, there are 2 parallel aligned tracks. By using the Measure Angle tool we can determine the tracks are tilted by with respect to each other. The reason for this is that there are each two read and write heads in DDS drives. Figure 14 MFM measurements of a Hewlett Packard Digital Data Storage 4GB-tape using different scan sizes. The mean feature size is. In figure 15 the magnetic structure of a hard drive disk (Samsung, 40 GB) is shown. The mean features size is around. Figure 15 MFM measurements of a Samsung size is. hard drive disk using different scan sizes. The mean feature The measurements show that not only the magnetic feature size is growing smaller with increasing storage capacity but also there are different structural arrangements for the respective techniques. Further- www.phywe.com P25385 PHYWE Systeme GmbH & Co. KG All rights reserved 9

5.3.85- Investigate in magnetic micro and nano more, these arrangements are not visible in topographic AFM imaging but only in MFM measurements probing the local magnetization of the data storage devices. In currently used Data storage devices the feature size usually is already very small, such that it is only possible to image the magnetic structure using high end MFM devices. 10 PHYWE Systeme GmbH & Co. KG All rights reserved P25385