Calibration of KAP meters

Similar documents
Joint ICTP/IAEA Advanced School on Dosimetry in Diagnostic Radiology and its Clinical Implementation May 2009

SPECIFICATION. Kilovoltage X-ray calibration system for protection and diagnostic level dosimetry. Prepared by

2012 :15th SESSION of ESMP

Preliminary studies of a new monitor ionization chamber

Dose Reduction and Image Preservation After the Introduction of a 0.1 mm Cu Filter into the LODOX Statscan unit above 110 kvp

Technical aspects on DAP calibration and CT calibration

MEASURING CONDITIONS AND UNCERTAINTIES FOR THE COMPARISON AND CALIBRATION OF NATIONAL DOSIMETRIC STANDARDS AT THE BIPM *

DETERMINATION OF INHERENT AND ADDITIONAL FILTRATION IN ORDER TO ESTABLISH RADIATION QUALITIES ACCORDING TO IEC 61267

I. Introduction.

Test Equipment for Radiology and CT Quality Control Contents

Introduction. Chapter 16 Diagnostic Radiology. Primary radiological image. Primary radiological image

Key words: fluoroscopy, dose-area-product, kerma-area-product, calibration of KAP meters, patient exposure

Exposure Indices and Target Values in Radiography: What Are They and How Can You Use Them?

Response characteristics of a tandem ionization chamber in standard X-ray beams

Cylindrical Ion Chambers Victoreen Model 550 Series

Update on NIST X-Ray Air-Kerma Standards and Calibrations

Half value layer and AEC receptor dose compliance survey in Estonia

熊本大学学術リポジトリ. Kumamoto University Repositor

ORAMED S MEASUREMENT AND SIMULATION CAMPAIGN FOR EXREMITY AND EYE LENS DOSES OF MEDICAL STAFF INVOLVED IN INTERVENTIONAL RADIOLOGY AND CARDIOLOGY

Estimation of signal transfer property for wireless digital detector in different measurement schemes

Beam-Restricting Devices

MC SIMULATION OF SCATTER INTENSITIES IN A CONE-BEAM CT SYSTEM EMPLOYING A 450 kv X-RAY TUBE

DOSE EVALUATION IN UROGRAPHY RADIOLOGICAL PROCEDURES

Electronic Brachytherapy Sources. Thomas W. Rusch

Cylindrical Ion Chambers

PD233: Design of Biomedical Devices and Systems

QC Testing for Computed Tomography (CT) Scanner

The importance of radiation quality for optimisation in radiology

Effect of patient support pads on image quality and dose in fluoroscopy a)

DETECTORS. Including Codes of Practice IONIZING RADIATION

New Detectors for X-Ray Metal Thickness Measuring

The Current State of EPID-Based Linear Accelerator Quality Assurance. Disclosures. Purpose of this First Talk 8/3/2017

X-rays in medical diagnostics

STEREOTACTIC BREAST BIOPSY EQUIPMENT SURVEYS

Y11-DR Digital Radiography (DR) Image Quality

INTERNATIONAL STANDARD

ABSORBED DOSE DISTRIBUTIONS USING THE ISODENSITOMETRIC METHOD FOR EXPOSURES WITH FILTER EMPLOYED FOR MAMMOGRAPHIES

Determination of the tube voltage from clinic mammographic system using two types of detectors

Title: A COMPARISON OF Cs-137 AND X-RAY SOURCES AS CALIBRATION REFERENCES FOR THERMOLUMINESCENT DOSIMETER CHIPS

8/2/2017. Radiologist Responsibilities. Radiologist Responsibilities. Medical Physicist Mammography Equipment Evaluation and Annual Survey

IONIZING RADIATION 2018 / 2019 DETECTORS. Including Codes of Practice

Ludlum Medical Physics

X-RAYS - NO UNAUTHORISED ENTRY

X-RAY IMAGING EE 472 F2017. Prof. Yasser Mostafa Kadah

Disclosures. Outline 7/31/2017. Current Implementation Status of IEC Standard : Exposure Index (EI) for Digital Radiography

X-rays. X-rays are produced when electrons are accelerated and collide with a target. X-rays are sometimes characterized by the generating voltage

Effect of pressure, temperature and humidity in air on photon fluence and air kerma values at low photon energies

Dosepix Detector as kvp-meter in Radiology and Mammography: First steps

Ask EuroSafe Imaging Tips & Tricks. Paediatric Imaging Working Group. Dose Management in Digital Radiography

Minnesota Rules, Chapter 4732 X-ray Revision

Derivation of total filtration thickness for diagnostic x-ray source assembly

Seminar 8. Radiology S8 1

MASINZA, ALWYN STANSLAUS. BSc (Hons.), PGDSc

Update on Fluoroscopy Physics AAPM MO-A-210A-1 Stephen Balter, Ph.D.

A Guide to Radiochromic Film Dosimetry with EBT2 and EBT3

Wide beam CT dosimetry. Elly Castellano

Assessment of Beam Alignment, Collimation and Half Value Layer of Some Selected X-Ray Machines in Plateau State, Nigeria

X-ray Imaging. PHYS Lecture. Carlos Vinhais. Departamento de Física Instituto Superior de Engenharia do Porto

A Generalized Strategy for 3D Dose Verification of IMRT/VMAT Using EPID-measured Transit Images

Introduction of a Single Chip TLD System for Patient Dosimetry

Studies on reduction of exposure dose using digital scattered X-ray removal processing

Mammography is a radiographic procedure specially designed for detecting breast pathology Approximately 1 woman in 8 will develop breast cancer over

Studies on reduction of exposure dose using digital scattered X-ray removal processing

QA Considerations. QA for LGK Perfexion : : Follow NRC licensing guidelines (10( CFR ) Leksell Gamma Knife Perfexion

IBEX TECHNOLOGY APPLIED TO DIGITAL RADIOGRAPHY

Digital Imaging started in the 1972 with Digital subtraction angiography Clinical digital imaging was employed from the 1980 ~ 37 years ago Amount of

1-1. GENERAL 1-2. DISCOVERY OF X-RAYS

X-ray Tube and Generator Basic principles and construction

Summer Student project report

Radiographic sensitivity improved by optimized high resolution X -ray detector design.

Use of Back Scattered Ionizing Radiation for Measurement of Thickness of the Catalytic Agent Active Material

Report on the study and optimisation of chest and lumbar spine X-ray imaging

Master of Science Thesis. SIMIND Based Pinhole Imaging

Your worldwide source of grids. What a difference a grid makes

Medical Imaging: A Look inside. Medical Imaging. Medical Imaging. Visible Human Project

QUALITY CONTROL TESTS IN SOME DIAGNOSTIC X-RAY UNITS IN BANGLADESH

Optimization of Energy Modulation Filter for Dual Energy CBCT Using Geant4 Monte-Carlo Simulation

CR Basics and FAQ. Overview. Historical Perspective

Received 3 December 2016; revised 17 May 2017; editorial decision 21 June 2017; accepted 4 July 2017

China Resources Wandong Medical Equipment Co., Ltd. High Frequency 50kW, 150kV Radiography System - HF50-R

Effect of slit scan imaging techniques on image quality on radiotherapy electronic portal imaging

I. PERFORMANCE OF X-RAY PRODUCTION COMPONENTS FLUOROSCOPIC ACCEPTANCE TESTING: TEST PROCEDURES & PERFORMANCE CRITERIA

Maximizing clinical outcomes

Renato Padovani, University Hospital, Udine, Italy

Evaluation of X-Ray Beam Quality Based on Measurements and Estimations Using SpekCalc and Ipem78 Models

The effect of compensating filter on image quality in lateral projection of thoraco lumbar radiography

Breast Tomosynthesis. Bob Liu, Ph.D. Department of Radiology Massachusetts General Hospital And Harvard Medical School

ISO INTERNATIONAL STANDARD. Non-destructive testing of welds Radiographic testing Part 1: X- and gamma-ray techniques with film

NOT FOR DISTRIBUTION JINST_083P_0914 v1

When small things matter. Small Field Dosimetry Application Guide

Teaching Digital Radiography and Fluoroscopic Radiation Protection

Features and Weaknesses of Phantoms for CR/DR System Testing

Monte Carlo study on a new concept of a scanning photon beam system for IMRT

Exposure System Selection

Clinical Experiences with a Patient Skin Dose Monitoring and Tracking Program

Development of Personal Dosimeter Using Electronic Dose Conversion Method

RADIOGRAPHIC EXPOSURE

CyberKnife Iris Beam QA using Fluence Divergence

Predicted image quality of a CMOS APS X-ray detector across a range of mammographic beam qualities

Technical data CAMARGUE CS-VH50/300. VARIABLE Height Bucky Table With Ceiling Suspension

Transcription:

Calibration of KAP meters Alexandr Malusek! Division of Radiological Sciences Department of Medical and Health Sciences Linköping University! 2014-04-15 1

Outline 1. KAP meter construction 2. Air kerma-area product 3. Calibration methods (NRPB, tandem, IAEA) 4. IAEA s NK based formalism 5. Calibration corrections 6. Suggestions for improvement 2

Introduction A KAP meter is a plane-parallel ionization chamber calibrated to measure the air kerma-area product. Old calibration methods (NRPB) cannot be used as films are no longer used in clinics New calibration methods (IAEA) have been developed but issues with accuracy still exist. Work on improvements continues 3

Accuracy of KAP meters Large uncertainties in PKA values measured by KAP meters in clinics are common owing to the strong energy dependence of these chambers. Manufacturers typically guarantee the accuracy of 25% (k=2) required by IEC specifications. In diagnostic radiology, IAEA s [3] and ICRU s [4] specifications require accuracy better than 7% (k=2 ). This holds true for PKA measurements too. 4

KAP meters sensitive area ~ 14 cm x 14 cm thickness ~ 1.6 cm conductive coating ~ 10 nm transparent to visible light 1.5 mm 5.9 mm 1.0 mm 5.9 mm 1.5 mm conductive coating outer electrode air cavity inner electrode air cavity outer electrode 5

X-ray tubes with collimator housing C-arm x-ray units 6

Air kerma-area product, PKA x ray tube Z P KA = A K air da reference plane Α Kair beam axis x ray tube Α PKA does not depend on the position of the reference plane if photons in the beam are neither scattered nor absorbed K air 1/d 2 A d 2 A d is the distance from focus 7

Photon scatter and absorption x ray tube KAP meter patient plane some photons registered by the KAP meter may not reach the patient plane owing to scatter or absorption some scattered photons may increase PKA at the patient plane Φ 1 Φ 1 Φ 2 > Scatter increases fluence of particles 8

Calibration methods National Radiological Protection Board (NRPB) [1] Tandem calibration (developed at STUK) [2] International Atomic Energy Agency (IAEA) [3] modification of NRPB s method for screen-film systems and computed radiography systems tandem calibration in IAEA s geometry laboratory calibration of the reference KAP meter 9

NRPB calibration method focal spot Features:! collimator KAP meter Kair is measured with an ionization chamber at the beam axis beam axis A is determined as the area within 50% of the maximum optical density optical density d m K air ionization chamber film PKA = Kair A Problems:! Kair measured at one point only d m /2 Films are no longer used 10

Tandem calibration method focal spot collimator clinical KAP meter Features:! Reference KAP meter measures PKA for incident radiation dcr ~ 30 cm to 40 cm d cr Attenuation in air is neglected Problems:! reference KAP meter A difference between clinical and standards laboratory beam qualities may result in systematic error A KAP meter holder is needed 11

IAEA calibration methods 20 cm 10 cm 95 cm 100 cm x ray tube KAP meter diagnostic detector styrofoam couch laboratory calibration 4 6 cm Features! K air is measured with detector at the beam axis Beam size 10 cm x 10 cm at the position of the detector The nominal area is determined as the area contained within 50% of the maximum optical density / pixel value P KA = K air A Alternatively: P KA is measured using a reference KAP meter Problems:! K air is measured in one point only in the beam-area method 12

The effect of distance 30.0 cm 30.0 cm 100 cm r x ray tube clinical KAP meter P KA,r P KA,p (r) Method:! Monte Carlo simulations of PKA using MCNP Ring detectors were used to score Kair X-ray spectra for 40, 80 and 140 kv filtered with 5 mm Al PKAp(r) PKAr 1.000 0.995 0.990 0.985 0.980 0.975 0.970 40 kv 80 kv 140 kv 5 10 15 20 25 30 radius / cm From [7] Cylindrical KAP meter based on VacuTec 70157! Results:! Beam attenuation in air Larger beam radius resulted in lower relative difference 13

Stray radiation Relative difference in PKA between the two configurations was less than 3% for considered tube voltages. Attenuation in air was the main factor. Stray radiation may be undetected in the IAEA calibration method. or? KAP meter stray radiation 14

NK formalism of IAEA s TRS-457 dosimetric quantity calibration coefficient z} { K =(M Q0 {z} z } { Y M 0 {z} ) N K,Q0 i k i {z} dosimeter reading, reference conditions dosimeter reading, no beam correction factors 15

Correction for temperature and pressure temperature of air in o C reference pressure P0 = 101.3 kpa k TP = 273.2+T 273.2+T 0 P0 P reference temperature T0 = 20 o C pressure of air in kpa 16

Correction for humidity Reference value of relative humidity of air is 50% No correction is needed in the range 30% - 80% (TRS-457) 17

Correction for radiation quality unknown calibration coefficient c.c. obtained from standards laboratory K Q = M Q z } { N K,Q = M Q N K,Q N K,Q0 z } { N K,Q0 = M Q N K,Q0 k Q,Q0 {z } Thus N K,Q = N K,Q0 k Q,Q0 beam quality correction factor 18

Suggestions for improvement Problem 1: Large energy dependence of KAP meters Calibration coefficients (RQR) Calibration coefficients (RQA) 2000 NKA (Gym 2 C 1 ) 1800 1600 1400 1200 1000 10 nm 15 nm 20 nm KAP 1 KAP 2 KAP 3 NKA (Gym 2 C 1 ) 1800 1600 1400 1200 1000 10 nm 15 nm 20 nm KAP 1 KAP 2 40 60 80 100 120 140 40 60 80 100 120 140 U (kv) U (kv) Figure: Calibration coefficient as a function of tube voltage for RQR and RQA beam qualities. Measured (markers) and simulated (lines) values. 19

Suggestions for improvement Problem 2: Beam qualities at clinics differ from beam qualities at standards laboratories.!! As a consequence, transfers of calibration coefficients from the standards laboratory to clinics are associated with uncertainties. 20

Our approach Calibration coefficients of built-in KAP meters should be beam-quality specific. This is a task for manufacturers. Before it happens, hospital physicists can determine factors correcting PKA values reported by the built-in KAP meters. To do so, three beam qualities have to be considered: Q0: reference beam quality at standards laboratory Q1: reference beam quality at the clinic Q: any diagnostic beam quality at the clinic 21

Our approach P KA = NP ref KA,QMQ ref N ref P KA,Q 1 N P KA,Q ref NP ref KA,Q M 0 Q ref 1 P KA,Q 0 = N ref P KA,Q N ref P KA = kq,q ref k ref 1 Q 1,Q N ref 0 P KA,Q M ref 0 Q where the beam quality correction factors are: k ref Q,Q 1 k ref Q 1,Q 0 Clinical reference beam quality Q1 to diagnostic beam quality Q. E.g. 70 kv and 0.1 mm Cu to 140 kv and 0.3 mm Cu. Standards laboratory reference beam quality Q0 to clinical reference beam quality Q1. E.g. RQR 5 to 70 kv and 0.1 mm Cu. 22

Our approach NPKA, Q NPKA, Q1 1.2 1.1 1.0 0.0 mm Cu, 10 nm 0.0 mm Cu, 15 nm 0.0 mm Cu, 20 nm 0.1 mm Cu, 10 nm 0.1 mm Cu, 15 nm 0.1 mm Cu, 20 nm 0.3 mm Cu, 10 nm 0.3 mm Cu, 15 nm 0.3 mm Cu, 20 nm 0.0 mm Cu, measured 0.1 mm Cu, measured 0.3 mm Cu, measured 60 80 100 120 140 U (kv) Figure: Simulated and measured values of the beam quality correction factor kq,q1. (Q1: 70 kv, 0.1 mm Cu, Q: 0.0, 0.1, and 0.3 mm Cu, any U). 23

Our approach Table: Simulated values of the beam quality correction factor kq1,q0 as a function of the KAP meter s conductive coating thickness. Thickness (nm) k 10 0.913 15 0.892 20 0.878 24

Our approach Result for an x-ray stand at the clinic: Manufacturer s calibration was in error by up to 30%. 25

Primary references [1] NRPB. National Protocol for Patient Dose Measurements in Diagnostic Radiology. Chilton: National Radiological Protection Board, 1992. [2] Toroi, P, T Komppa, and A Kosunen. A Tandem Calibration Method for Kerma area Product Meters. Physics in Medicine and Biology 53 (September 21, 2008): 4941 58. [3] International Atomic Energy Agency. Dosimetry in Diagnostic Radiology: An International Code of Practice. Vienna: International Atomic Energy Agency, 2007. 26

Secondary references [4] ICRU. Patient Dosimetry for X Rays Used in Medical Imaging (Report 74). International Comission on Radiation Units & Measurements, 2005. [5] Toroi, P, T Komppa, A Kosunen, and M Tapiovaara. Effects of Radiation Quality on the Calibration of Kerma-Area Product Meters in X-Ray Beams. Physics in Medicine and Biology 53 (September 21, 2008): 5207 21. [6] Malusek, A, J P Larsson, and G Alm Carlsson. Monte Carlo Study of the Dependence of the KAP-Meter Calibration Coefficient on Beam Aperture, X- Ray Tube Voltage and Reference Plane. Physics in Medicine and Biology 52 (February 21, 2007): 1157 70. [7] Malusek, Alexandr, and Gudrun Alm Carlsson. Analysis of the Tandem Calibration Method for Kerma Area Product Meters via Monte Carlo Simulations. In Standards, Applications and Quality Assurance in Medical Radiation Dosimetry (IDOS), 1:129 36. Vienna: IAEA, 2011. [8] Larsson, J P, J Persliden, and G Alm Carlsson. Ionization Chambers for Measuring Air Kerma Integrated over Beam Area. Deviations in Calibration Values Using Simplified Calibration Methods. Physics in Medicine and Biology 43 (March 1, 1998): 599 607. 27