COCIR SELF-REGULATORY INITIATIVE FOR MEDICAL IMAGING EQUIPMENT COMPUTED TOMOGRAPHY MEASUREMENT OF ENERGY CONSUMPTION

Similar documents
diagnostic examination

The RSM shall be extensively used in field & laboratory for verification of the accuracy of all types of single-phase energy meters.

Meter Testing Equipments, Laboratory Planning and set up

Overview of Safety Code 35

The VARIAN 250 MeV Superconducting Compact Proton Cyclotron

TOPICS: CT Protocol Optimization over the Range of Patient Age & Size and for Different CT Scanner Types: Recommendations & Misconceptions

Iterative Reconstruction in Image Space. Answers for life.

Slide 1. Slide 2. Slide 3 ACR CT Accreditation. Multi-Slice CT Artifacts and Quality Control. What are the rules or recommendations for CT QC?

QC Testing for Computed Tomography (CT) Scanner

Diffraction-enhanced X-ray Imaging (DEXI) Medical Solutions. More information using less radiation

2D, 3D CT Intervention, and CT Fluoroscopy

Maximum Performance, Minimum Space

Detector technology in simultaneous spectral imaging

Research Support. Dual-Source CT: What is it and How Do I Test it? Cynthia H. McCollough, Ph.D.

Automated dose control in multi-slice CT. Nicholas Keat Formerly ImPACT, St George's Hospital, London

NEMA XR X-ray Equipment for Interventional Procedures User Quality Control Mode

Nuclear Associates , , CT Head and Body Dose Phantom

Specification for a Fully Automatic Meter Test Equipment for testing Three-phase/Single/Transformer operated/ reference Standard energy meters.

Computed Tomography. The Fundamentals of... THE FUNDAMENTALS OF... Jason H. Launders, MSc. Current Technology

GE AMX 4+ Portable X-Ray

Moving Test - MT3000

SYLLABUS. 1. Identification of Subject:

CR Basics and FAQ. Overview. Historical Perspective

Diagnostic X-Ray Shielding

Ludlum Medical Physics

QC by the MPE in Belgium

X-RAYS - NO UNAUTHORISED ENTRY

ETSI TR V1.1.1 ( )

Digital Imaging and Communications in Medicine (DICOM) Supplement 188: Multi-energy CT Images

XRC X-Ray Calibration System

NeuViz 16 Computed Tomography. Elevating routine imaging for exceptional results

HISTORY. CT Physics with an Emphasis on Application in Thoracic and Cardiac Imaging SUNDAY. Shawn D. Teague, MD

Thermionic x-ray. Alternative technologies. Electron Field Emission. CNT Based Field Emission X-Ray Source

Wide-Detector CT for TAVR Planning:

ISPFILMQATM STATE-OF-THE-ART RADIOTHERAPY VERIFICATION SOFTWARE. Supports all major radiotherapy technologies! FilmQA TM

Digital Imaging and Communications in Medicine (DICOM) Supplement 188: Multi-energy CT Images

Connection Impact Assessment Application

Nancy G. Leveson and Clark S. Turner, An Investigation of the Therac-25 Accidents. Computer 26(7), pp , Jul Presented by Dror Feitelson

RAY COMPACT ENERGY METER MECHANICAL

RTU560. Multimeter 560CVD03. Power Measuring Display. Characteristics. Application. Data Sheet Multimeter 560CVD03

SOMATOM Esprit A Bundle of Energy

abc MHRA Philips Mx8000 IDT CT scanner technical evaluation September 2004 Best choice best practice nww.medical-devices.nhs.

SAFIRE. Sinogram Affirmed Iterative Reconstruction. Answers for life.

RAY MC COMPACT ENERGY METER MECHANICAL

FemtoFAB. Femtosecond laser micromachining system. tel fax Konstitucijos ave. 23C LT Vilnius, Lithuania

Translating Protocols Between Scanner Manufacturer and Model

Our promise your guarantee

ECOVIEW 9 / ECOVIEW 9 PLUS Digital Radiographic System

160-slice CT SCANNER / New Standard for the Future

Power transformers. Shunt reactors Proven history for future success

ENCORE 300 SERIES INSTALLATION AND OPERATING INSTRUCTIONS

Distributed Generation Application Form (Generation of Greater than 20 kw to 15 MW)

Wind Power Facility Technical Requirements CHANGE HISTORY

4.1. Accurate: The information is a true reflection of the original observation.

Welcome at MR Achieva 3.0T TX. About the system. Index. 1.1 Introduction

Embedded Generation Connection Application Form

COMPUTED TOMOGRAPHY 1

SPECIFICATION. Kilovoltage X-ray calibration system for protection and diagnostic level dosimetry. Prepared by

PREMIER INSTALLATION AND OPERATING INSTRUCTIONS

Facilities Operations Meter Reading Guide Version 1.0 October 27, 2017

Medical Devices cyber risks and threats

USER S MANUAL. This manual must be considered an integral part of the projector. The user must read this manual before using the projector

Best choice in all DR. Jumong M. SG HealthCare Jumong M Digital radiography specifications

RAY MC COMPACT ENERGY METER MECHANICAL

ECOVIEW 9 / ECOVIEW 9 PLUS Digital Radiographic System

A.V.R. R250. Installation and maintenance R250 0V E+ E- VOLT STAB FREQ. & L.A.M. CONFIG. REQUEST A QUOTE

HITACHI Proton Therapy System with Spot Scanning

Data. microcat +SPECT

Philips XPER FD10C R7.0.4

New spectral benefi ts, proven low dose

Multi-Access Biplane Lab

China Resources Wandong Medical Equipment Co., Ltd. High Frequency 50kW, 150kV Radiography System - HF50-R

This is a preview - click here to buy the full publication

Engineering Recommendation M30 Issue Standard Electricity Network Operator Electricity Smart Meter Configurations

HITEK POWER XR KW X-RAY POWER SUPPLY MODULE

ETSI EN V1.1.1 ( )

Technical Documentation of (EU) No 617/2013

T h e P h a n t o m L a b o r a t o r y

Energy Meters for DIN Rail Mounting Electric energy meter WS0101, WS0102,WS1102 WS0301, WS0302,WS1302

SONOMETER TM 1100 Ultrasonic compact energy meter

Quantitation of clinical feedback on image quality differences between two CT scanner models

Technical Documentation of (EU) No 617/2013

inemi Statement of Work (SOW) Medical TIG Qualification Methods for Portable Medical Products

Digital Radiography X-Ray System. X Twin with X Mobil Roesys GmbH [rshsmi] 1/11

Improvement of CT image quality with iterative reconstruction idose4

INFOCAL 8 Energy calculator

Dose Reduction and Image Preservation After the Introduction of a 0.1 mm Cu Filter into the LODOX Statscan unit above 110 kvp

The Hux Connect Portal

ETSI EN V1.1.1 ( )

RAY MECHANICAL COMPACT METER. 4 MID approved class 2 from ND 15 mm qp 0,6 m3/h up to ND 40 mm qp 10 m3/h

Authors: Cabral, Ricardo 1 ; Carvoeiras, Pedro 2 ; Fatana, João 2, ; Alves, Rita 1. 1 Centro Hospitalar Lisboa Norte - Hospital de Santa Maria; 2

SONOMETER TM 1100 Ultrasonic compact energy meter

SULTANATE OF OMAN METERING AND DATA EXCHANGE CODE

RAIL MOUNTED NETWORK METER TYPE N27D USER S MANUAL

12/21/2016. Siemens Medical Systems Research Agreement Philips Healthcare Research Agreement AAN and ASN Committees

Thyristor power controller B DI1002 V1.0 Manual/Operating instructions

DIGITAL MOBILE CXR THE NIGERIA FIELD EXPERIENCE NIGERIA TEAM UN CC, ADDIS ABABA 5-8, 2010

General Hospital Catalog

Energy Management Three-phase energy meter with output modules Type EM4-DIN

IHE Radiology Technical Framework Supplement. Stereotactic Mammography Image (SMI) Trial Implementation

Transcription:

COCIR SELF-REGULATORY INITIATIVE FOR MEDICAL IMAGING EQUIPMENT COMPUTED TOMOGRAPHY MEASUREMENT OF ENERGY CONSUMPTION Revision: 1 Date: June 2015 Approved: June 2015

TABLE OF CONTENT 1. INTRODUCTION... 3 2. SCOPE... 3 3. DEFINITIONS... 4 4. SYSTEM POWER MODES... 5 5. USE MODES OVERVIEW... 5 6. EXAMINATION DISTRIBUTION AND SCAN PROTOCOLS... 6 7. RESOURCES... 6 8. UNIT UNDER TEST (UUT)... 6 9. POWER MEASUREMENT DEVICE... 7 10. MEASUREMENT OF POWER... 8 10.1. TRANSITION OFF-TO-IDLE AND IDLE POWER MEASUREMENT... 8 10.2. TRANSITION IDLE-TO-OFF MEASUREMENT... 9 10.3. SCAN MODE ENERGY MEASUREMENT... 9 10.4. TRANSITION IDLE-TO-LOWPOWER MEASUREMENT... 9 10.5. TRANSITION LOWPOWER-TO-IDLE MEASUREMENT... 9 11. ENERGY CONSUMPTION CALCULATION... 10 12. COCIR CT DATA COLLECTION SPREAD SHEET... 10 2 / 10

1. INTRODUCTION The Energy-related Products (Ecodesign) Directive, 2009/125/CE, enables the European Commission (EC) to set Ecodesign requirements through new regulations for any group of products which uses energy or is related to energy consumption. In 2007, Medical Devices were identified as a Priority A product group by the European Commission for future regulation. COCIR Companies presented in 2009 a Self-regulatory Initiative for Medical Imaging Equipment, driven by the Ecodesign Steering Committee, committing to improve the environmental performances of their products. As the Steering Committee has done for magnetic resonance, a working group was formed to determine a measurement methodology and ecodesign targets for computed tomography (CT) from manufacturers of these systems. The outcome of the group was an agreed upon procedure for measuring typical energy consumption of CT equipment. This procedure defines specific states of system operation and instructions for determining a set of scanning protocols to be analysed. 2. SCOPE This methodology can be used to measure all CT systems. Equipment and accessories beyond a basic CT product and not required for a basic scan, or customer-provided equipment, such as optional reconstruction hardware, patient vital signs accessories, facility-provided cooling equipment and hardware for advanced medical applications are outside the scope of this procedure. It is not intended for use on interventional CT equipment. The methodology is not suited for the measurement of technologies combining CT with other imaging modalities, such as CT/PET. 3 / 10

3. DEFINITIONS 12h day-time: time from 7 am to 7 pm. 12h night-time: time from 7 pm to 7 am. Energy: The capacity to do work. In this document, the unit of energy is kilowatthours (kw h). Power: The rate at which energy is generated or consumed. In this document, the unit of power is kilowatts (kw). Procedure Type: An examination is a collection of scans for an individual patient. Procedure Type refers to an exam for a specific anatomy or type of exam. (e.g. Abdomen or Vascular). Off: The system is shut down, AC mains off, according to the user manual. The system consumes no energy. Low-power: The system functions into the minimum energy consumption state that the user can select according to the user manual. Idle mode: A state of the system when fully powered but no scan has been prescribed. This mode does NOT include x-ray tube rotor or gantry rotation. Scan mode: A state of the system between individual scans and during scans (e.g. during patient handling, examination planning, contrast agent injection and active scanning with x-ray generation). This mode includes tube rotor rotation, gantry rotation and generation of image, and any possible idle between scans. 4 / 10

4. SYSTEM POWER MODES The operation modes are defined as Off, Low power, Idle, and Scan. The energy consumption differs between the modes and the transition between modes occurs by operator selection. The anticipated power of these modes, from high to low, is: Scan > Idle > Low power > Off The table below shows a possible state transition order, for a CT system. Initial State Transition To Method Off / Low-power Idle Operator starts system Idle Scan Operator starts a scan Scan Idle Scanning completes Idle Off / Low Power Operator selection 5. USE MODES OVERVIEW Typical daily system operation is set as follows: Time period Mode Duration 12h night-time Off or Low Power 12 hours 12h day-time Idle and Scan 12 hours The typical daily energy consumption of a CT system is the sum of the energy consumption for each of the two time periods. 5 / 10

6. EXAMINATION DISTRIBUTION AND SCAN PROTOCOLS Within each procedure type, the specific scan parameters representing each exam type have been chosen to harmonize across manufacturers to general values as follows: Table 1: Exam Protocols Scan type Head Chest Spine Abdomen Kv 120 120 120-140 120 ma 200-300 100-250 300 187-400 Rotation speed 0,5-1 0,5 1 0,5-0,8 CTDI Maximum 60 8 25 15 Scan Length (mm) 200 200 200 200 Reconstruction jobs 1 1 1 1 Tube current modulation ON ON ON ON 7. RESOURCES The following personnel are recommended: An engineer or technician familiar with the power distribution of the system and power electronics safety. An engineer or applications specialist familiar with scanner operation and the prescribing of clinical protocols. 8. UNIT UNDER TEST (UUT) System Configuration: System configuration should be recorded and configured to perform the set of specified procedures with appropriate mandatory equipment. System Voltage: The system should be installed according to the manufacturer s installation instructions at an input voltage of 380-415 VAC. Installation: The system shall be installed and calibrated according to its specification, including all system-critical items needed to perform a basic scan, e.g. gantry, x-ray generator and tube, reconstruction engine(s), required electronics such power supplies, controllers, console/computer, and patient table. Any equipment and accessories beyond the basic product offering that is not required for a basic scan (e.g. customer-provided equipment, optional Supplier equipment such as patient vital signs accessories, hardware for advanced medical applications, etc.) shall not be included in the measurement. Environmental Conditions: The measurements are to be taken at a steady-state operating temperature within manufacturer s specified ambient temperature and humidity limits. 6 / 10

Measurement: Prior to each mode s measurement, the equipment shall remain in that mode for sufficient time to allow temperature and other pertinent transient conditions to stabilize, except when measuring transient conditions (Off/Low-power to Idle and Idle to Off/Low-power). Phantoms: For the scan a 200-350mm water/polypropylene phantom shall be used, so that the spiral scan is not performed without any structure in the scan beam. 9. POWER MEASUREMENT DEVICE The measurements shall be recorded using a device capable of measuring 3-phase voltage and current, and calculating the integral of power with respect to time (energy) or a power meter able to sample average power ratings. Figure 1: Power usage of a CT scanner over 24 hours. The level of the low power mode depends on the specific CT models and technology. The power measurement device shall be installed onto the input to the AC Mains disconnect panel of the system to ensure that all energy consumption of the CT equipment is captured, including any cooling equipment provided by the CT supplier as part of the base system. 7 / 10

10. MEASUREMENT OF POWER Trial executions of testing showed a further simplification of the test procedure is possible by approximating the range of protocols by only the Abdomen scan. Setting up Scan Protocols Prepare the abdomen scan protocols provided by the CT scanner (e.g. abdomen routine, abdomen standard, ) according to the user manual using the parameters defined in section 5. If it is not possible for the CT system under test to use a certain sequence or specified value, use a sequence or value as close as possible to the sequence specified given the same diagnostic results. The power measurement of different system states, transitions etc, described in the following sections shall finally allow to calculate the complete 24h power consumption according to figure 2. Figure 2: exemplary power usage of CT scanner for calculation of daily energy consumption. The figure reports also variable idle power. 10.1. TRANSITION OFF-TO-IDLE AND IDLE POWER MEASUREMENT 1) Shut the system to Off mode according to the user manual. 2) Wait to ensure that all system elements have established a static Off power operation 3) Start the power measurement on the measurement device. 4) Then switch on the CT system to idle according to the user manual. The measurement device now records the complete transition from Off state to Idle state (energy consumption of transition Off-to-Idle) 5) Record E off-idle (Energy from Off to Idle) [kwh] 6) Record total transition time, T off-idle [h] 7) Record the power consumption in Idle state (Idle power consumption) in kw 8 / 10

10.2. TRANSITION IDLE-TO-OFF MEASUREMENT 1) Allow the system to stabilize in Idle mode then start the power measurement on the measurement device. 2) Shut the system to Off mode according to the user manual (no energy consumption). 3) The measurement device now records the complete transition from Idle state to Off state (energy consumption of transition Idle-to-Off) 4) Record E idle-off (Energy from Idle to Off) [kwh] 5) Record total transition time, T idle-off [h] 10.3. SCAN MODE ENERGY MEASUREMENT 1) Begin kwh datalogging at time T=0. 2) Prescribe a patient and begin an abdomen exam according to Table 1, Exam Protocols. 3) Allow the exam to complete normally. 4) Let the system run until time T=35 minutes. 5) Record the total energy in kilowatt-hours. 6) Record the scan energy E A in kwh. 10.4. TRANSITION IDLE-TO-LOWPOWER MEASUREMENT 1) Allow the system to stabilize in Idle mode, then start the power measurement on the measurement device. 2) Switch the system to LowPower mode according to the user manual. 6) After a static LowPower state is reached, stop the power measurement on the power measurement device. The measurement device now records the complete transition from Idle state to LowPower state (energy consumption of transition Idle-to-LowPower) 7) Record E idle-lowpower [kwh] 8) Record total transition time, T idle-lowpower [h] 9) Record the power consumption in Idle state (Idle power consumption) in kw 10.5. TRANSITION LOWPOWER-TO-IDLE MEASUREMENT 1) Allow the system to stabilize in LowPower mode, and then start the power measurement on the measurement device. 2) Switch the system to Idle mode according to the user manual. 3) After a static Idle state is reached, stop the power measurement on the power measurement device. The measurement device now records the complete transition from LowPower state to Idle state (energy consumption of transition LowPower-to-Idle) 4) Record E lowpower-idle [kwh] 5) Record total transition time, T lowpower-idle [h] 9 / 10

11. ENERGY CONSUMPTION CALCULATION The 24h energy consumption according to figure 2 has to be calculated out of the measured power and energy consumption of the different system states and transitions, with the following formula for the 3 scenarios: Scenario Off: The CT scanner is in Off mode for 12h during night time E tot = E off + P idle x 20mins + 20 x E A Scenario Idle: The CT scanner is in Idle mode for 12h during night time E tot = E idle (12h) + P idle x 20mins + 20 x E A Scenario Low : The CT scanner is in LowPower mode during 12h night time E tot = E lowpower + P idle x 20mins + 20 x E A Where: E off : E idle-off + E off-idle E idle : Energy in Idle mode over 12h night time E lowpower : E idle-lowpower + P lowpower * (12h - T idle-low T low-idle ) + E lowpower-idle P idle : Power in Idle mode 12. COCIR CT DATA COLLECTION SPREAD SHEET The data obtained according to the present methodology have to be filled in the appropriate template that can be downloaded from the COCIR website www.cocir.org for the calculation of the energy consumption in the different scenarios. 10 / 10

Self-regulatory Initiative for medical imaging equipment Copyright 2011 COCIR All Rights Reserved. All material in this document is, unless otherwise stated, the property of COCIR. Copyright and other intellectual property laws protect these materials. Reproduction or retransmission of the materials, in whole or in part, in any manner, without the prior written consent of the copyright holder, is an infringement of copyright law.