Nikon A1R. Multi-Photon & Laser Scanning Confocal Microscope. Kyle Marchuk Adam Fries Jordan Briscoe Kaitlin Corbin. April 2017.

Similar documents
Nikon AZ100. Laser Scanning Macro Confocal Microscope. Jordan Briscoe Adam Fries Kyle Marchuk Kaitlin Corbin. May 2017.

Leica SP8 TCS Users Manual

Nikon C1si Spectral Laser Scanning Confocal Microscope. User Guide

TRAINING MANUAL. Olympus FV1000

Things to check before start-up.

Training Guide for Leica SP8 Confocal/Multiphoton Microscope

LSM 710 Confocal Microscope Standard Operation Protocol

LEICA TCS SP5 AOBS TANDEM USER MANUAL

Confocal imaging on the Leica TCS SP8. 1) Turn the system on. 2) Use TCS user account. 3) Start LAS X software:

Nikon Eclipse Ti A1-A Confocal Operating Manual. Start-up. Microscope

Operating Instructions for Zeiss LSM 510

Leica Sp5 II Confocal User Guide

Training Guide for Carl Zeiss LSM 510 META Confocal Microscope

LSM 780 Confocal Microscope Standard Operation Protocol

Nikon SIM-E & A1-R System

Quick Start Guide. Leica SP5 X

Microscopy from Carl Zeiss

Zeiss LSM 510 Confocor III Training Notes. Center for Cell Analysis & Modeling

Operating Checklist for using the Laser Scanning Confocal Microscope. Leica TCS SP5.

Supplemental Method Information Zeiss LSM710

Nikon A1Rsi Confocal Start-Up Sequence

LSM 800 Confocal Microscope Standard Operation Protocol

Leica SP8 TCS Users Manual

Training Guide for Carl Zeiss LSM 5 LIVE Confocal Microscope

ZEISS LSM510META confocal manual

ZEISS LSM 710 CONFOCAL MICROSCOPE USER MANUAL

Practical work no. 3: Confocal Live Cell Microscopy

Zeiss LSM 510 Multiphoton Confocal Microscope

Quick Guide for Zeiss 710 Laser Scanning Confocal MGH Cancer Center

Guide to Confocal 5. Starting session

Cell Biology and Bioimaging Core

Quick Guide for Zeiss 710 Laser Scanning Confocal MGH Cancer Center

Horiba LabRAM ARAMIS Raman Spectrometer Revision /28/2016 Page 1 of 11. Horiba Jobin-Yvon LabRAM Aramis - Raman Spectrometer

User Guide to the IBIF Leica TCS SP8 MP Confocal Microscope

Title: Nikon A1R Confocal User Manual

1 Set up the confocal light path for imaging a green dye (Alexa488-EGFP). For example, the

Training Guide for Carl Zeiss LSM 7 MP Multiphoton Microscope

Title: Leica SP5 Confocal User Manual

LSM 510 Meta Training Notes

NIS-Elements C (For CONFOCAL MICROSCOPE A1) Instructions (Ver. 4.40)

Zeiss 780 Training Notes

Operation Guide for the Leica SP2 Confocal Microscope Bio-Imaging Facility Hunter College October 2009

Quick Guide. LSM 5 MP, LSM 510 and LSM 510 META. Laser Scanning Microscopes. We make it visible. M i c r o s c o p y f r o m C a r l Z e i s s

LSM 510 Training Notes

Zeiss LSM 510 Multiphoton Confocal Microscope

MIF ZEISS VIOLET CONFOCAL ZEN 2009 PROTOCOL

Nikon. King s College London. Imaging Centre. N-SIM guide NIKON IMAGING KING S COLLEGE LONDON

b. Turn the power switch and key to on position for blue laser.

Leica SP8 Resonant Confocal. Quick-Start Guide

Zeiss LSM 780 Protocol

Zeiss Axio Imager.A1 manual

Zeiss LSM 880 Protocol

Use of the HSW5 Spinning Disk Confocal Microscope Updated last May 25, 2010 OK

Contents STARTUP MICROSCOPE CONTROLS CAMERA CONTROLS SOFTWARE CONTROLS EXPOSURE AND CONTRAST MONOCHROME IMAGE HANDLING

ZEISS LSM 710 NLO Multiphoton microscope Manual/Quick guide

OPERATING INSTRUCTIONS

Confocal Application Notes Vol. 5 July 2010

Training Guide for Carl Zeiss LSM 880 with AiryScan FAST

Zeiss Axiovert 135 Fluorescence Microscope Quick Guide / Operations Manual (v. 1.0 February 09)

Swept-Field User Guide

SHORT INSTRUCTIONS FOR OPERATING LSM1/2 (Zeiss LSM510) AT CIAN Version 1.4, September 2014

Leica SPEII confocal microscope. Short Manual

Renishaw InVia Raman microscope

Supplemental Figure 1: Histogram of 63x Objective Lens z axis Calculated Resolutions. Results from the MetroloJ z axis fits for 5 beads from each

MIF ZEISS LSM510 CONFOCAL USER PROTOCOL

EPIFLUORESCENCE &/OR BRIGHTFIELD MICROSCOPY

Widefield 1. Switching on

Leica TCS SP8 Quick Start Guide

Contents. Introduction

Zeiss 880 Training Notes Zen 2.3

Diskovery Spinning Disk Guide

Olympus Fluoview 1000S Spectral Confocal Microscope Introduction to the NRI-MCDB Microscopy Facility Spectral Confocal Microscope

Leica TCS SP8 Quick Start Guide

The Zeiss AiryScan System, Confocal Four.

Why and How? Daniel Gitler Dept. of Physiology Ben-Gurion University of the Negev. Microscopy course, Michmoret Dec 2005

Widefield-NikonEclipseTE200-ORCA Nikon Eclipse TE200 Inverted Microscope with Hamamatsu 1394 Orca-ER Cooled CCD Camera and Micromanager Software

1 Co Localization and Working flow with the lsm700

Topics. - How to calibrate the LSM scanner. - How to clean the microscope. - How to adjust the pinhole alignment. - How to adjust the Collimator

Brief manual how to start and close the Leica sp2 Confocal. (TCS SP2 AOBS system mounted on a DM IRE2)

ZEN 2012 SP5 black edition Hotfix 12

Thor. Nikon Inverted Structured Illumination (N-SIM) & A1 Laser Scanning Confocal Microscope User Guide. v. 1.0 (12/2015)

Microscope Confocal LSM510 META

Imaging Beyond the Basics: Optimizing Settings on the Leica SP8 Confocal

CONFOCAL MICROSCOPE (Zeiss LSM 510 META v4.2)

Nikon Eclipse Ti2-E Widefield/Spinning Disk Confocal Microscope Standard Operation Protocol

Zeiss LSM880 Operating Instructions. UTMB Optical Microscopy Core Jan. 16, 2018

BX-61: Brightfield Instruction /Continue to scroll for Fluorescent Instuctions

3 Choose the Channels button and set the Channel Settings. Set the Pinhole to 1 Airy unit.

Instructions for the Leica SP5 II laser scanning confocal microscope

Zeiss AxioImager.Z2 Brightfield Protocol

Leica TCS SL Confocal Training. Neuroscience Imaging Core Staff. Core Director. Facility Manager

Internal Medicine Imaging Core Emory University Department of Medicine

Nasmyth Ultraview Vox User Protocol

Nikon E800 Microscope. Operating Instructions

START-UP PROCEDURE 1 THE MICROSCOPE STAND 3 OBJECTIVES 5 STARTING WITH LAS (SOFTWARE) AND SETTING UP THE MICROSCOPE STAND 7

Image-Pro Plus 7.0 Product Note

Olympus Digital Microscope Camera (DP70) checklist

Nikon TE300 Eclipse Wide-Field Microscope

Usermanual for Leica SP8 confocal

MAKE SURE YOUR SLIDES ARE CLEAN (TOP & BOTTOM) BEFORE LOADING DO NOT LOAD SLIDES DURING SOFTWARE INITIALIZATION

Transcription:

Nikon A1R Multi-Photon & Laser Scanning Confocal Microscope Kyle Marchuk Adam Fries Jordan Briscoe Kaitlin Corbin April 2017 Contents 1 Introduction 2 2 Start-Up 2 3 Imaging 4 3.1 Sample Alignment........................................... 4 3.2 OC Panel............................................... 6 3.3 A1plus Compact GUI......................................... 7 3.4 LUTs.................................................. 8 3.5 ND Acquistion............................................. 9 3.6 Other Useful Features......................................... 10 4 Two-Photon Imaging 12 5 Shutdown 13 6 Trouble Shooting 14 1

1 Introduction The Nikon A1R microscope consists of two illumination options; four confocal laser lines (408 nm, 488 nm, 561 nm, 638 nm) and a Mai Tai DeepSee IR laser (740-1040 nm). For the confocal laser lines there are 4 PMT detectors (blue, green, red, far red) that can reach video rate and faster. For multi-photon imaging there is a high sensitivity non-descanned detector (NDD) for deep tissue imaging. The system utilizes an Nikon base in the upright geometry, an automated Prior XY stage, and a PIFOC objective piezo for scanning. Additionally, there is a perfusion system with an in-line and stage heater and an anesthesia setup making the microscope suitable for slide, slice, explant, and intravital imaging. Currently, there are 3 long working distance objectives consisting of a 16x/0.8 NA Plan Apo water immersion, 20x/1.0 NA Plan Apo water immersion, and 25x/1.1 NA Plan Apo water immersion. Also, there are higher magnification objectives consisting of a 40x/1.3 NA Plan Fluor Oil immersion, 60x/1.2 NA Plan Apo water immersion, and a 100x/1.4 NA Plan Apo oil immersion. Please sign up for time using the MyCores microscope scheduler. Reservations can be made up to 4 weeks in advance for owner s consortium and beginning on Friday for the following week for non-owners. 2 Start-Up 1. Check the Mai Tai laser (Figure 1). (a) The power supply should always be on. DO NOT turn it off. (b) The key should always be turned to ON. (c) If the screen on the lower box indicates coolant needs to be added, add a small volume. If you have not done this before, please contact a BIDC employee and we will show you how. Note: Even if you are only using the diode lasers to image, the Mai Tai needs to be on and talking to the computer for the software to run. Figure 1: Power and Key always ON. 2. Make sure the large stage is completely down (Figure 2). 3. Turn on the microscope peripherals (Figure 3). (a) Turn on the Prior XY stage (1). (b) Turn on the z-axis piezo (2). (c) Turn on the microscope power supply (3) and wait until initialization is complete. (d) Turn on the microscope control box (4). 2

Figure 2: Stage Down on Startup. 4. Turn on the arc lamp for visualization (Figure 3). Note: The lamp must be on for at least 10 minutes before shutting it off. (a) Turn on the power switch located on the back of the box. (b) Press local for lamp control. (c) Press 1 to start the lamp. Note: If your acquisition will run for hours, please turn off the lamp to conserve bulb lifetime. Figure 3: Peripherals and Lamp. 5. Turn the keys on confocal lasers from Standby to ON (Figure 4). Note: Only turn on the lasers you will be using. 6. Launch the NIS-Elements software from the desktop. 7. Clear previous user s presets: (a) Clear the Z Intensity Correction tab in the lower left of the GUI. (b) Reset the Z range under the ND Acquistion Z tab. (c) Clear all XY stage positions under the ND Acqusition XY tab. 3

8. Install objective. (a) Check out the objective from the BIDC office (filling out the form on the cork board). (b) Screw the objective into the z-axis piezo. Note: If you have not handled an objective before, please have a BIDC employee show you an appropriate technique. Figure 4: Turn keys from Standby to On. 3 Imaging 3.1 Sample Alignment 1. Place your sample on stage. (a) For open solution samples, first put down the plastic drop tray. Note: Cut-out must be in the upper-right corner to avoid collision with the micro-encoder. (b) Add the black metal piece with the circular hole for slides. (c) Place your slide and add the correct immersion fluid for the objective. (d) If you are using the heated stage and/or the perfusion pump, have a BIDC employee show you the proper setup. 2. Move sample under the objective. (a) Carefully raise the large stage by hand. Pro Tip: Work your hands under the stage by the adjustable legs instead of pulling up. This will reduce the risk a hitting the objective with your sample. (b) In combination with the XY stage, move your sample so that the objective is touching your immersion liquid. (c) Tighten the knobs so the stage does not drift downwards. (d) Using the coarse adjustment, lower the objective until it is just above your sample. 3. Align sample by eye. (a) Change the microscope s light path using the knob on the right side of the microscope stage to BINO. 4

(b) Select how you would like to visualize your sample through the eyepiece by selecting GFP Eyes, DAPI Eyes, etc in the OCPanel. Pro Tip: Use your brightest or most concentrated dye for initial alignment (c) Looking through the eyepiece, turn the microscope focus towards you (thumbs up) until your sample comes into focus. 4. Switch to imaging with software. (a) Once you have found your sample with the eyepieces, change the light path to F with the knob at the side of the microscope. (b) In the OC Panel turn off the lamp illumination by selecting Close Lamp Shutter. (c) In the software select the proper objective under the OC Panel (at the top right of the screen). This affects the metadata for your acquisition (stitching, pinhole size, etc.) (d) On the left side of the screen, click the red Remove Interlock button to open the light path. 5

There are 4 main components of the software s Graphical User Interface. Many of the settings and selections will work in cooperation with others. The following sections of this user guide will attempt to explain how to use the most popular buttons and settings. 3.2 OC Panel This panel is mostly only used during the start of your imaging session (Figure 5). Visualization: Selects the filter used for finding your sample by eye. Press Close Lamp Shutter to turn off lamp illumination. Confocal: Saved settings for different imaging techniques. Default Confocal Settings is a good place to start for a new sample. Shut Down is pressed before closing the software. Objectives: Select the objective used for imaging. This is important for stage stitching, calculated parameters, and metadata. Special Apps: Saved settings for particular imaging conditions. Not useful to the average user. Figure 5: OC Panel 6

3.3 A1plus Compact GUI Within this tab many of the options for laser and PMT control are found (Figure 6). Scan: Runs a live preview. Capture: Takes a single image. Galvano vs Resonant: This switches between scan settings within the microscope. Resonant scans at a continuous frequency, while Galvano will dwell at each pixel position. Resonant is fast and is useful for alignment. Galvano is slower, but will produce nicer images. uni- vs bi-drectional scanning Sets the PMTs to collect light while the y mirror is scanned in only in one direction or both. Bi-directional is faster, while uni- is more precise. Control by: Determines how the user sets the frame rate for acquistion, whether by the pixel dwell time, or by setting a frame per second. Size: The size, in pixels, of your image. Some sizes are not accessible due to the imaging speed set above. If you want to access larger image areas, change the frame rate to a longer acquisition setting. Normal vs line averaging vs line integration: These settings allow the user to scan lines multiple times and either average or integrate the signal. Sample acquisition will take longer depending on the amount of averaging or integration chosen. The Normal button clears those settings. Ch Series: The dropdown allows the user to choose the channels and the order in which they are fired. This is a useful feature to limit emission bleedthrough. The more passes selected, the slower the acquisition. Unselecting the Ch Series button fires all channels at once. Pinhole: Changes the physical size of the pinhole for light collection. Pressing the 1.2 AU sets the size to 1.2 Airy Units based on your selected objective. This is a good starting point for most experiments. Opening the pinhole increases light throughput, but descreases resolution. The dropdown menu chooses which wavelength (laser illumination) the pinhole is calculated for. DU4 vs IR NDD: This selects which detectors are used. DO NOT select IR NDD if the lights are on or you re using the visible diode lasers. This can cause permanent harm to the detectors. Use IR Laser: Select this option when doing 2P-imaging (see below). Detectors and Lasers: The checkboxes select which PMTs to use in imaging. HV is the gain setting (typical 80-110). Offset is a current bias on the PMTs (keep at 0). The pushbuttons turn on or off the emission of the laser. The slider and input box is the power of the laser in percentage (typical 1-20). We do not recommend going below 1-2% as the laser intensities will become unstable. 7

Figure 6: A1plusCompactGUI 3.4 LUTs Look up tables (LUTs) control how the intensity is displayed on the screen. They always control the active image (whether during or post acquisition). They do not change how the data is collected (Figure 7). Autoscale (All): The computer gives a best guess on the best way to view the data. This is a good button to press, as you can then adjust the min and max sliders to dial in your display. Dropdown: Selects which LUTs are displayed. Saturation Display: Highlights which pixels are being saturated with the current illumination settings. It s best to have this on while setting up acquisition parameters. 8

Figure 7: LUTs 3.5 ND Acquistion The ND Acquisition tab is where most of your experimental setup will take place (Figure 8). Save to File: Check this box to automatically save your file after completion to the directory with the chosen filename. The filename will continuously update itself so no data is overwritten. It is good practice to always have this checked. Order of Experiment: Choose the order in which the data is acquired. The rule of thumb is to minimize physical movements. Channels are faster than z stacks are faster than stage movements. Time: Set up a time lapse experiment. XY: Choose different positions across your sample. Z: Used to set up scanning through your sample in the z direction. 1. Top: Assign the top of the z stack from the current objective position. 2. Bottom: Assign the bottom of the z stack from the current objective position. 3. Step: The size of the steps through the z stack. 4. Steps: The number of the steps through the z stack. 5. Step push button: Assigns the steps and step size based upon the Nyquist sampling limit calculated for that objective. Lambda: Used to set up non traditional illumination schemes. See a BIDC employee if help is needed. Large Image: Used to set up a large tiled image. Run Now: When everything has been selected. Run Now will start the acquisition using the laser and PMT settings from the A1plus Compact GUI. Run Z Corr: Runs the experiment like Run Now with the exception that it using the interpolated power and PMT settings from the Z Intensity Correction tab (See Below). 9

Figure 8: ND Acquisition 3.6 Other Useful Features Some other quality of life features can be found around the software (Figures 9,10,11). XY Mouse At the top of the imaging window press the target looking button. This will allow you to control the stage with the mouse. Z Intensity Correction Located in the lower left of the GUI. This allows you to set laser intensity and PMT gain specific to the z position. Intensity at z steps is linearly interpolated. Split Channel View Collected images can be split into their channels. Volume View A 3D view of collected data. Tile View Data is broken into z positions. Max Projection View the max intensity at all pixels through a dimension. Sliders The blue tile buttons at the bottom of your image allow you to slide through z positions and time. Scan Large Image Under the Aquire menu there is the Scan Large Image option. This is a useful menu that quickly creates a tiled view at the current z position. Once the image is collected, the user can right-click on the image and interact with it, such as setting a feature to the image center. Users must select the proper objectives. 10

Figure 9: Z Intensity Correction Figure 10: Image Interaction 11

Figure 11: Scan Large Area 4 Two-Photon Imaging Two-Photon Imaging parameters and acquisition setup works in much the same way as the visible diode lasers. 1. In the A1plus Compact GUI, check the Use IR Laser Box. This will enable the A1plus MP GUI (upper left). 2. Press the On button to enable the Mai Tai Deep See Laser. This will take a few minutes as indicated by the flashing yellow square. 3. Clicking on the yellow circle option opens the shutter. 4. Choose the wavelength to set the laser. Switching wavelengths takes a few seconds as indicated by the flashing yellow square. 5. Change the power within any of the activated PMT windows. Since there is only one laser line, relative channel intensities can only be balanced by wavelength selection and PMT gain. 6. WARNING: Only press the IR NDD option once the lights are off and the microscope cover is closed. This button selects the sensitive IR NDD detectors for two-photon imaging. Each time you need to turn on the lights, select the DU4 detectors. 12

5 Shutdown Check the scheduler. If there is no one scheduled within two hours of your experiment s completion, follow the complete shutdown procedure, otherwise only shutdown the software. 1. Software (a) In the OC panel select Shut Down. (b) Close the software. 2. Microscope (a) Turn off the microscope control box (4). 3. Devices (a) Turn off the microscope power supply (3). (b) Turn off the z-axis piezo (2). (c) Turn off the Prior XY stage (1). 4. Lamp (a) Under the local setting, press 2 to shut down the lamp. (b) Turn off power. 5. Lasers (a) Turn the lasers you used from ON to Standby. 6. Objective (a) Gently unscrew the objective from the piezo. (b) Clean the objective with water and lens paper. Note: If you have not cleaned an objective before, please have a BIDC employee show you a proper technique. (c) Return the objective to the BIDC office. 13

6 Trouble Shooting Please let us know when you run into a software or hardware problem, so we can add it to the list. 1. There is no light getting to the eyepiece. (a) Check that the arc lamp is powered and emitting. (b) Check that the light path knob is set to BINO. (c) Open the filters and shutters menu in the software. i. Check if the shutter is open. ii. Check that the correct filter cube is selected. (d) Check whether the neutral density filters on the left side of the microscope are engaged. 2. No light is getting to the detector. (a) Make sure the light path knob is set to F (not R or BINO ). (b) Click the red interlock button in the software. 3. During software startup an error pops up saying it cannot connect to a device. (a) Press the Shut Down button in the OC Panel. (b) Close the software. (c) Lower the microscope stage. (d) Check the piece of hardware and cycle the power, if needed. (e) Restart the software. 4. During acquisition your sample seems to be zoomed in. (a) Check the viewing area zoom factor. (b) Check your pixel array size (it sometime becomes bigger than the screen). (c) Check the Scan Area tab to make sure you are scanning the whole image area. 14