X-rays. X-rays are produced when electrons are accelerated and collide with a target. X-rays are sometimes characterized by the generating voltage

Similar documents
Mammography is a radiographic procedure specially designed for detecting breast pathology Approximately 1 woman in 8 will develop breast cancer over

X-RAY IMAGING EE 472 F2017. Prof. Yasser Mostafa Kadah

X-ray Tube and Generator Basic principles and construction

X-ray Tube and Generator Basic principles and construction

- KiloVoltage. Technique 101: Getting Back to Basics

X-rays in medical diagnostics

Technical data CAMARGUE CS-VH50/300. VARIABLE Height Bucky Table With Ceiling Suspension

Exposure System Selection

DENTAL RADIOGRAPHY KAMARUL AMIN BIN ABU BAKAR

1. Carlton, Richard R., and Arlene M. Adler. Principles of Radiographic Imaging: An Art and a Science, 5th edition (2013).

RADIOGRAPHIC EXPOSURE

SPRINGFIELD TECHNICAL COMMUNITY COLLEGE ACADEMIC AFFAIRS

Introduction. Chapter 16 Diagnostic Radiology. Primary radiological image. Primary radiological image

Partial Replication of Storms/Scanlan Glow Discharge Radiation

Mammography: Physics of Imaging

PD233: Design of Biomedical Devices and Systems

Beam-Restricting Devices

Photons interaction with matter

Explain what is meant by a photon and state one of its main properties [2]

Maltase cross tube. D. Senthilkumar P a g e 1

MILADY. Product Data. Page 1 of 8

LECTURE 1 The Radiographic Image

80 Physics Essentials Workbook Stage 2 Physics

X-RAYS - NO UNAUTHORISED ENTRY

Seminar 8. Radiology S8 1

Maximizing clinical outcomes

P R E S E N T E D B Y. K A M A R U L A M I N A B D U L L A H Dip. MED. IMG., BSc. MED. IMG. (UiTM)

Using Carbon Nano-Tube Field Emitters to Miniaturize X-Ray Tubes

MXHF-1500RF is controlled by Digital key panel console that displays KV, ma and mas with APR menu programmed.

1-1. GENERAL 1-2. DISCOVERY OF X-RAYS

V SALAI SELVAM, AP & HOD, ECE, Sriram Engg. College, Perumalpattu 1 MEDICAL ELECTRONICS UNIT IV

X-Rays and endoscopes

Medical Imaging. X-rays, CT/CAT scans, Ultrasound, Magnetic Resonance Imaging

New Detectors for X-Ray Metal Thickness Measuring

MaxRay Handheld X-ray Systems Operator Training Exam

Veterinary Science Preparatory Training for the Veterinary Assistant. Floron C. Faries, Jr., DVM, MS

Optimization of Energy Modulation Filter for Dual Energy CBCT Using Geant4 Monte-Carlo Simulation

Beam Production, characteristics and shaping

Development of a flat-panel x-ray source

Atomic and nuclear physics

X-ray tube with needle-like anode

Course Outline: At the completion of each chapter the student should be able to

SPECIFICATION. Kilovoltage X-ray calibration system for protection and diagnostic level dosimetry. Prepared by

Image Quality. HTC Grid High Transmission Cellular Grid provides higher contrast images

Visibility of Detail

AN ABSTRACT OF THE THESIS OF. W. Scott Helms for the degree of Master of Science in Radiation Health Physics

Rotating Anode X-ray Tube

X-ray Imaging. PHYS Lecture. Carlos Vinhais. Departamento de Física Instituto Superior de Engenharia do Porto

MC SIMULATION OF SCATTER INTENSITIES IN A CONE-BEAM CT SYSTEM EMPLOYING A 450 kv X-RAY TUBE

Beam Production, Characteristics and Shaping

ANALYTICAL MICRO X-RAY FLUORESCENCE SPECTROMETER

Multiple Choice Identify the letter of the choice that best completes the statement or answers the question.

CIRCLEX 0.3/0.8P324&0.6/1.2P324DK-85

Spectroscopy in the UV and Visible: Instrumentation. Spectroscopy in the UV and Visible: Instrumentation

Physics Laboratory Scattering of Photons from Electrons: Compton Scattering

Test Equipment for Radiology and CT Quality Control Contents

(12) Patent Application Publication (10) Pub. No.: US 2005/ A1

X-RAY. Lecture No.4. Image Characteristics:

Imaging Technique Optimization of Tungsten Anode FFDM System

I. PERFORMANCE OF X-RAY PRODUCTION COMPONENTS FLUOROSCOPIC ACCEPTANCE TESTING: TEST PROCEDURES & PERFORMANCE CRITERIA

Digital Imaging started in the 1972 with Digital subtraction angiography Clinical digital imaging was employed from the 1980 ~ 37 years ago Amount of

Investigation of the line-pair pattern method for evaluating mammographic focal spot performance

Scintillation Counters

Scanning electron microscope

Image Quality. HTC Grid High Transmission Cellular Grid provides higher contrast images

TECHNICAL DATA. GIOTTO IMAGE SDL/W is pre-arranged for Full Field Digital Biopsy examination with the patient in prone position.

MWPC Gas Gain with Argon-CO 2 80:20 Gas Mixture

Physics 4BL: Electricity and Magnetism Lab manual. UCLA Department of Physics and Astronomy

Zaidi Embong and Husin Wagiran Physics Department, University Of Technology Malaysia, P.O Box 791, 80990, Johor Baharu

ABOVETABLE X-RAY SOURCE FLUOROSCOPIC AND SPOT-FILM SYSTEMS

OPTI510R: Photonics. Khanh Kieu College of Optical Sciences, University of Arizona Meinel building R.626

PRACTICAL CONSIDERATIONS AND EFFECTS OF METALLIC SCREEN FLUORESCENCE AND BACKSCATTER CONTROL IN GAMMA COMPUTED RADIOGRAPHY

X-ray Imaging (XRI) Lectures on Medical Biophysics. Dept. Biophysics, Medical Faculty, Masaryk University in Brno. X-ray Imaging (XRI)

2014 CINDE Toronto Portable X-ray & X-Ray Production

CR Basics and FAQ. Overview. Historical Perspective

Version 1.0. TechnicVR. Student Guide

Principle of X-Ray Systems

Rotating Anode X-Ray Tube Housing Assembly. General Data

Ludlum Medical Physics

Scanning electron microscope

Terahertz Radiation of a Low-inductance Discharge in Vacuum with Laser-plasma Initiation

Tomosynthesis. Energy Subtraction. Long View Imaging

New spectral benefi ts, proven low dose

Human Retina. Sharp Spot: Fovea Blind Spot: Optic Nerve

Dose Reduction and Image Preservation After the Introduction of a 0.1 mm Cu Filter into the LODOX Statscan unit above 110 kvp

Digital Radiography System

Effect of pressure, temperature and humidity in air on photon fluence and air kerma values at low photon energies

Ph 3324 The Scintillation Detector and Gamma Ray Spectroscopy

SCANNING ELECTRON MICROSCOPY AND X-RAY MICROANALYSIS

Study of increased radiation when an x-ray tube is placed in a strong magnetic field

Nuclear medicine imaging has been an integral component

Diffraction-enhanced X-ray Imaging (DEXI) Medical Solutions. More information using less radiation

Rotating Anode X-Ray Tube Housing Assembly. General Data. IEC Classification... Class I

Overview. Professor Roentgen was a Physicist!!! The Physics of Radiation Oncology X-ray Imaging

MODULE I SCANNING ELECTRON MICROSCOPE (SEM)

I. Introduction.

Radiology. Radiograph: Is the image of an object made with use of X- ray instead of light.

Spectrophotometer. An instrument used to make absorbance, transmittance or emission measurements is known as a spectrophotometer :

KENYA NATIONAL EXAMINATION COUNCIL REVISION MOCK EXAMS 2016 TOP NATIONAL SCHOOLS

Components of Optical Instruments

Transcription:

X-rays Ouch! 1

X-rays X-rays are produced when electrons are accelerated and collide with a target Bremsstrahlung x-rays Characteristic x-rays X-rays are sometimes characterized by the generating voltage 0.1-20 kv soft x-rays 20-120 kv diagnostic x-rays 120-300 kv orthovoltage x-rays 300 kv 1 MV intermediate energy x-rays > 1MV megavoltage x-rays 2

Bremmstrahlung Bremsstrahlung x-rays occur when electrons are (de)accelerated in the Coulomb field of a nucleus 3

Bremsstrahlung 4

Bremsstrahlung The power radiated from an accelerating charge is given by Larmor s equation 2 2 P = 2 3 e c a 3 In the case of an electron in the Coulomb field of a nucleus F Ze a = = k ~ m r 2 m 2 Z 5

Bremsstrahlung The probability of bremsstrahlung goes as Z 2, hence high Z targets are more effective than low Z The energy of the x-rays varies from zero to the maximum kinetic energy of the electron (x-ray tube kvp) The energy spectrum from a thick target goes as 1/E but inherent (1mm Al eq) plus additional (few mm Al) filtration removes the lower energy x-rays Here I am referring to diagnostic x-rays 6

Bremsstrahlung The unfiltered energy spectrum is approximately given by Kramer s law which was an early application of quantum mechanics I ( ) ( ) E = KZ T E γ e γ 7

Bremsstrahlung 8

Characteristic x-rays After excitation, ions with a vacancy in their inner shell can de-excite Radiatively through x-ray fluorescence Non-radiatively through the emission of Auger electrons 9

Characteristic X-rays Thus an x-ray spectrum will also show characteristic x-rays arising from L to K and M to K transitions after ionization of a K electron Usually transitions to higher shells absorbed by the filtration or are not x-rays 10

Characteristic X-rays The probability of K shell fluorescence increases with Z 11

Characteristic X-rays 12

Characteristic X-rays Sometimes the characteristic x-rays are emphasized using the same material for target and filter Characteristic x-rays from molybdenum are effective in maximizing contrast in mammography 13

Characteristic X-rays Mo target, filter, and result 14

Directionality For MeV electrons, bremsstrahlung x- rays are preferentially emitted in the electron s direction For kev electrons, bremsstrahlung x- rays are emitted at larger angles Characteristic x-rays are emitted isotropically since there is no angular correlation between the incident electron that causes the ionization and the fluorescent photon 15

X-ray Tube A simplified x-ray tube (Coolidge type) shows the idea behind most x-ray tubes today 16

X-ray Tube In addition to bremsstrahlung and characteristic x-ray production, electrons also loose energy through collisions Collision losses dominate in this energy region radiation loss EZ ( E in MeV) collision loss 820 For 100 kev electrons in W radiation loss collision loss 0.1 74 820 0.009 Thus >99% of the electron energy goes into heating the target rather than x-rays Removing heat from the anode in a vacuum is an issue = 17

X-ray Tube Efficiency of x-ray production depends on the tube voltage and the target material W (Z=74) in this example P P deposited radiated = VI = 0.9 10 Efficiency ε = P P 9 ZV radiated deposited 2 I = 0.9 10 9 ZV kvp (V) 50 200 6000 Heat (%) 99.7 99 65 X- rays (%) 0.3 1 35 18

X-ray Tube X-ray tubes 19

X-ray Tube More detail 20

X-ray Tube Housing for shielding (Pb) and cooling (oil) 21

X-ray Tube More detail 22

X-ray Tube The main parts of the x-ray tube are Cathode/filament Typical electron current is 0.1-1.0 A for short exposures (< 100 ms) Anode/target Glass/metal envelope Accelerating voltage Typical voltage is 20-150 kvp 23

Cathode consists of Cathode Low R tungsten wire for thermionic emission Tungsten has a high melting point (3370C) and minimum deposit on the glass tube Tube current is controlled by varying the filament current which is a few amps A focusing cup Uses electric field lines to focus the electrons Typically there are two filaments Long one: higher current, lower resolution Large focal spot Short one: lower current, higher resolution Small focal spot 24

Cathode Dual focus filament is common 25

Anode Usually made of tungsten in copper because of high Z and high melting point Molybdenum and rhodium used for soft tissue imaging Large rotating surface for heat distribution and radiative heat loss Rotation of 3k-10k revolutions/minute Resides in a vacuum (~10-6 torr) Thermally decoupled from motor to avoid overheating of the shaft Target is at an tilted angle with respect to axis Bremsstrahlung is emitted at ~ right angles for low energy electrons Determines focal spot size 26

Anode 27

Anode 28

Anode The heating of the anode limits the voltage, current, and exposure time An exposure rating chart gives these limits 29

Anode Power = V x I (watts) Energy = Power x time = V x I x s (joules) HU (Heating Unit) ~ J Damaged anodes 30

Anode The angle determines the projected focal spot The smaller the angle the better the resolution Typically 7-20 degrees θ Angle θ Angle Incident electron beam width Actual focal spot size Incident electron beam width Apparent focal spot size Actual focal spot size Increased apparent focal spot size Film Film 31

X-rays The energy of the photons depends on the electron energy (kvp) and the target atomic number Z The number of photons depends on the the electron energy (kvp), Z, and the beam current (ma) A typical number / area is ~ 10 13 / m 2 About 1% will hit the film ~ 10 11 / m 2 Absorption and detection efficiency will further reduce this number 32

Automatic Exposure Control X Ray tube Collimator Beam Air Soft tissue Bone Patient Table Grid AEC detectors Cassette AEC detectors can ionization chambers or solidstate detectors 33

Automatic Exposure Control Most modern x-rays machines are equipped with automatic exposure control also called a phototime The AEC sets the technical parameters of the machine (kv, ma, time, ) in order to avoid repeated exposures AEC is used to keep the radiographic quality (film density) equal on all patients AEC detectors can be ionization chambers or solid state detectors 34

Grid To reduce the number of secondary scattered photons making it to the film, a grid between the patient and film is used 35

Details Grid Grid bars are usually lead whereas the grid openings are usually made of aluminum or carbon Grid thickness is typically 3 mm Grid ratio is H/W and 10/1 is typical Grid frequency of 60 lines / cm is typical B/W/H on the figure might be 0.045, 0.120, 1.20 in mm The Bucky factor is the entrance exposure w/wo the grid while achieving the same film density 4 is average 36

Accelerating Voltage The potential difference between cathode and anode must be generated by 60 Hz 220V AC power High voltages are produced using a transformer 37

Accelerating Voltage Electrons are accelerated when the filament is at a negative potential with respect to the target Diode circuits can be used to provide rectification (AC to DC voltage) Three phase power (6 pulse or 12 pulse) can be used to reduce ripple Constant potential operation can be achieved by using constant potential (voltage regulations) or high frequency x- ray generators 38

Half-wave Rectifier Not very efficient 39

Full-wave Bridge Rectifier This circuit allows the entire input waveform to be used 40

Accelerating Voltage kv ripple (%) 100% 13% 4% Single phase single pulse Single phase 2-pulse Three phase 6-pulse Three phase 12-pulse Line voltage 0.01 s 0.02 s 41