Performance test of the tig10 to be used for ECBR measurements with TITAN

Similar documents
ORTEC. Research Applications. Pulse-Height, Charge, or Energy Spectroscopy. Detectors. Processing Electronics

Gamma Ray Spectroscopy with NaI(Tl) and HPGe Detectors

Traditional analog QDC chain and Digital Pulse Processing [1]

Silicon Photomultiplier Evaluation Kit. Quick Start Guide. Eval Kit SiPM. KETEK GmbH. Hofer Str Munich Germany.

COMPTON SCATTERING. Purpose. Introduction. Fundamentals of Experiment

Nyquist filter FIFO. Amplifier. Impedance matching. 40 MHz sampling ADC. DACs for gain and offset FPGA. clock distribution (not yet implemented)

Physics Laboratory Scattering of Photons from Electrons: Compton Scattering

Digital Signal Processing for HPGe Detectors

SPECTROMETRIC DETECTION PROBE Model 310. Operator's manual

ORTEC. AN34 Experiment 14 Nuclear Lifetimes and the Coincidence Method. Equipment Needed from ORTEC. Equipment Required from Other Manufacturers

ORTEC Experiment 13. Gamma-Gamma Coincidence with Angular Correlation. Equipment Required

Instructions for gg Coincidence with 22 Na. Overview of the Experiment

K 223 Angular Correlation

CAEN Tools for Discovery

Purpose This experiment will use the coincidence method for time correlation to measure the lifetime in the decay scheme of 57

Energy Measurements with a Si Surface Barrier Detector and a 5.5-MeV 241 Am α Source

Real-Time Digital Signal Processors with radiation detectors produced by TechnoAP

Comparisons Between Digital Gamma-Ray Spectrometer (DSPec) and Standard Nuclear Instrumentation Methods (NIM) Systems

Testing of the NSC Electronics Module with the GSI Clover Detector

XRF Instrumentation. Introduction to spectrometer

INDEX. Firmware for DPP (Digital Pulse Processing) DPP-PSD Digital Pulse Processing for Pulse Shape Discrimination

NEEP 427 PROPORTIONAL COUNTERS. Knoll, Chapters 6 & 14 Sect. I & II

Keyser, Ronald M., Twomey, Timothy R., and Bingham, Russell D. ORTEC, 801 South Illinois Avenue, Oak Ridge, TN 37831s

Fast first practical help -- detailed instructions will follow- preliminary Experiment F80

Investigation of a Cs137 and Ba133 runs. Michael Dugger and Robert Lee

Virtual Laboratory of Nuclear Fission Virtual practicum in the framework of the project Virtual Laboratory of Nuclear Fission

18-fold segmented HPGe, prototype for GERDA PhaseII

A digital method for separation and reconstruction of pile-up events in germanium detectors. Abstract

Charge Loss Between Contacts Of CdZnTe Pixel Detectors

Electronic Instrumentation for Radiation Detection Systems

Week 11: Chap. 16b Pulse Shaping

Stage B: Allo scoperta del nucleo atomico: l esperimento di Rutherford. Stage C: Informatica e fisica sperimentale

Multichannel Analyser, Extended Version

1 Purpose of This Lab Exercise:

CAEN. Electronic Instrumentation. CAEN Silicon Photomultiplier Kit

DSA-LX. Digital Signal Analyzer. Radiation Safety. Amplified.

TRINAT Amplifier-Shaper for Silicon Detector (TASS)

GAMMA-GAMMA CORRELATION Latest Revision: August 21, 2007

Comparisons of the Portable Digital Spectrometer Systems

ORTEC ORTEC. Modular Pulse- Processing Electronics. What s in this Catalog? Who Needs this Catalog?

PHYSICS ADVANCED LABORATORY I COMPTON SCATTERING Spring 2002

Today s Outline - January 25, C. Segre (IIT) PHYS Spring 2018 January 25, / 26

EXPERIMENT 5. SCINTILLATION COUNTING AND QUENCH CORRECTION.

Gamma Spectrometer Initial Project Proposal

ORTEC Experiment 3. Gamma-Ray Spectroscopy Using NaI(Tl) Equipment Required. Purpose. Gamma Emission

nanomca datasheet I. FEATURES

Photon Count. for Brainies.

NIM INDEX. Attenuators. ADCs (Peak Sensing) Discriminators. Translators Analog Pulse Processors Amplifiers (Fast) Amplifiers (Spectroscopy)

SPECTROMETRIC CHARACTERISTIC IMPROVEMENT OF CdTe DETECTORS*

nanomca 80 MHz HIGH PERFORMANCE, LOW POWER DIGITAL MCA Model Numbers: NM0530 and NM0530Z

Application Note. Monitoring the Release of Radioactive Noble Gases Through the Stack of a Nuclear Power Plant (NPP): Stack Monitor System

Multianode Photo Multiplier Tubes as Photo Detectors for Ring Imaging Cherenkov Detectors

Charge-Sensing Particle Detector PN 2-CB-CDB-PCB

Amptek Silicon Drift Diode (SDD) at High Count Rates

RF Generators. Requirements:

Triplet polarimeter update

Partial Replication of Storms/Scanlan Glow Discharge Radiation

Dual 500ns ADC User Manual

LUDLUM MODEL MODEL AND MODEL GAMMA SCINTILLATORS. June 2017

User s Manual for Integrator Short Pulse ISP16 10JUN2016

A user-friendly fully digital TDPAC-spectrometer

Implementation of A Nanosecond Time-resolved APD Detector System for NRS Experiment in HEPS-TF

Bryn Mawr College Department of Physics Undergraduate Teaching Laboratories Alpha Particle Spectroscopy


Summary. Introduction

AN-DPP-003 Rev A2: Using the DP5 with HPGe USING THE DP5 WITH GERMANIUM DETECTORS

GRETINA. Electronics. Auxiliary Detector Workshop. Sergio Zimmermann LBNL. Auxiliary Detectors Workshop. January 28, 2006

Physics 342 Laboratory. Scattering of Photons from Free Electrons: Compton Scattering

Photon Counters SR430 5 ns multichannel scaler/averager

Studies of a Bulk Micromegas using the Cornell/Purdue TPC

Electrical Test of HP 0.5-µm Test Chip for Front-end Electronics for GLAST Tracker

Diamond sensors as beam conditions monitors in CMS and LHC

ORTEC. Time-to-Amplitude Converters and Time Calibrator. Choosing the Right TAC. Timing with TACs

Performance of 8-stage Multianode Photomultipliers

Radiation Detection Instrumentation

50 MHz Voltage-to-Frequency Converter

High collection efficiency MCPs for photon counting detectors

X-rays. X-rays are produced when electrons are accelerated and collide with a target. X-rays are sometimes characterized by the generating voltage

PARTICLE DETECTORS (V): ELECTRONICS

nanomca-ii-sp datasheet

IDM-200-P. High Purity Germanium (HPGe) Interchangeable Detector Module. Intelligent, Modular Solutions for Nuclear Security Monitoring Systems

Ph 3324 The Scintillation Detector and Gamma Ray Spectroscopy

Comparisons of the DSPEC and DSPEC Plus Spectrometer Systems

--- preliminary Experiment F80

Citation X-Ray Spectrometry (2011), 40(4): 2. Right final form at

EE-4022 Experiment 3 Frequency Modulation (FM)

Characterisation of the VELO High Voltage System

Commissioning Status and Results of ATLAS Level1 Endcap Muon Trigger System. Yasuyuki Okumura. Nagoya TWEPP 2008

A Modular Readout System For A Small Liquid Argon TPC Carl Bromberg, Dan Edmunds Michigan State University

Digital coincidence acquisition applied to portable β liquid scintillation counting device

nanodpp datasheet I. FEATURES

Digital trigger system for the RED-100 detector based on the unit in VME standard

P. Branchini (INFN Roma 3) Involved Group: INFN-LNF G. Felici, INFN-NA A. Aloisio, INFN-Roma1 V. Bocci, INFN-Roma3

Positron Emission Tomography

NIM. ADCs (Peak Sensing) Analog Pulse Processors Amplifiers (Fast) Amplifiers (Spectroscopy) Attenuators Coincidence/Logic/Trigger Units

A Prototype Amplifier-Discriminator Chip for the GLAST Silicon-Strip Tracker

Operation of a Single Pass, Bunch-by-bunch x-ray Beam Size Monitor for the CESR Test Accelerator Research Program. October 3, 2012

SILICON DRIFT DETECTORS (SDDs) [1] with integrated. Preliminary Results on Compton Electrons in Silicon Drift Detector

PX4 Frequently Asked Questions (FAQ)

AMPTEK INC. 14 DeAngelo Drive, Bedford MA U.S.A FAX:

Transcription:

Performance test of the tig1 to be used for ECBR measurements with TITAN T. Brunner, S. Ettenauer, A. Gallant January 17, 21 1 Introduction At the TITAN experiment there will be several detectors installed to observe radiation that is being emitted by ions stored inside the EBIT. Currently, a thick crystal HPGe (Ge), a thin crystal low energy Ge (LeGe) and several planar Si (PIPS) detectors are installed at the experiment. These detectors are read out by various preamplifier and thus provide different output signals and ranges (see table 1). The requirements for the data acquisition system are to provide an energy resolution comparable to analog systems and the time stamp of the event. To read out these signals we investigated the performance of a tig1 Detector Preamp Output polarity Output range Typ. range Ge RC - -5V -1 V? LeGe TRP + -5 V -5 V PIPS 23BT + -5 V -.2 V Mesytec (+/)- -±5 V -±.2 V Table 1: List of detectors and preamplifier currently installed at TITAN. 14bit sampling ADC system. Herefore, efficiency, resolution and performance were compared with an ORTEC DSPEC system, a digital waveform processing ADC. For direct comparison we installed the sources in front of the detectors and took data with both systems, DSPEC and tig1. This reduces any effects due to different source placement. 2 Performance of the DSPEC To get energy spectra from the LeGe we used a 133 Ba source due to its several energy lines between 6 kev and 35 kev. In order to obtain spectra with the Ge detector we used a 6 Co source in combination with a 57 Co source. This provides lines in the regions of 12 kev and 13 kev. The spectra were fitted with Radware, a software package for interactive graphical analysis of γ-ray 1

data. This program is fitting a function consisting of a Gaussian peak, a step function to account for Coulomb scattered photons, a background and a skew function to account for photons that got scattered out of the crystal. The FWHM of spectra fitted with Radware agree with the results obtained with Maestro, Ortec s program to read out DSPECs. Quoted uncertainties are only uncertainties of the fit. 2.1 Ge detector resolution The resolutions of the Ge detector for several peaks, read out with the DSPEC, are listed in table 2. Analyzed with ORTEC s program Maestro, the FWHM of the peaks were determined to be.99 kev @ 122 kev,.96 kev @ 136 kev, 1.75 kev @ 1173 kev and 1.82 kev @ 1333 kev. Position [kev] PosErr [kev] FWHM [kev] FWHM error [kev] 14.246.15.887.9 122.59.2.984.1 136.485.2.978.6 1173.285.2 1.76.4 1332.522.2 1.796.5 Table 2: FWHM of the Ge detector at different energies ( 6 Co and 57 Co) achieved with the DSPEC and fitted with Radware. 2.2 LeGe detector resolution To determine the resolution of the LeGe detector a 133 Ba source was used. The spectrum was fitted in Origin with a Gaussian and a step function and the results are listed in table 3. The fitted resolution achieved with Radware are listed in table 4. Due to the small crystal, the number of photons scattered out of the crystal is rather small and the results of both programs agree very well. The fitted peak positions plotted as function of gamma energies results in a straight line for both results of Origin and Radware. The specrum taken with DSPEC and LeGe detector is linear in the range between 5 kev and 4 kev (see Figure 1). 2

Position [kev] FWHM [kev] 6.7 ±.3.162 ±.6 14.711 ±.3.28 ±.6 25.359 ±.3.282 ±.6 35.231 ±..314 ±.1 36.98 ±.1.334 ±.2 53.415 ±.1.358 ±.2 79.86 ±.5.441 ±.3 81.237 ±..45 ±.1 122.294 ±.2.579 ±.3 136.697 ±.11.615 ±.16 16.826 ±.12.685 ±.17 276.564 ±.3 1.1 ±.6 32.984 ±.3 1.49 ±.2 356.127 ±.3 1.123 ±.2 Table 3: FWHM of the LeGe detector at different energies ( 133 Ba and 57 Co) achieved with the DSPEC and fitted with Origin. RW Position [kev] RW FWHM [kev] 53.163 ±.1.347 ±.4 79.64 ±.1.435 ±.2 81.19 ±.1.441 ±.1 276.57 ±.8.919 ±.2 33.28 ±.11.983 ±.28 356.24 ±.7 1.147 ±.1 Table 4: FWHM of the LeGe detector at different energies ( 133 Ba and 57 Co) achieved with the DSPEC and fitted with Radware. 3 Performance of the tig1 During the tests with the tig1 several parameters were optimized to achieve the best energy resolution. Besides, we also investigated the contribution of electric noise to the spectrum. The tig1 card was set to 4 V mode. 3.1 tig1 energy resolution Several parameters need to be set for the use of the tig1. The parameters that mainly determine the energy calculation of a sampled wave form are L and M.An ODB script was modified in order to scan through many different L and M parameters on the tig1 card. The script currently resides in the directory " /online/tig1" on titan4. 3

5 3 35 35 (a) y =.2916(399) +.9995(2) x Origin Fitted Peak Position [kev] 3 1 5 Radware Fitted Peak Position [kev] 35 3 5 (b) y = -.3422(57) + 1.72(2) x 5 1 3 35 NNDC Peak Energy [kev] NNDC Peak Energy [kev] Figure 1: Fitted peak position versus gamma energy of LeGe detector and DSPEC, analyzed with Origin and Radware. 3.1.1 Ge detector energy resolution It was found that the L parameter had the largest effect on the resolution while the M parameter had little to no effect. The lack of dependence of the resolution on the M parameter is surprising given that it is a major component in correcting for the decay of the pulse from the pre-amp. (a) Scan for the L parameter for given M values for the 1.33 MeV peak of 6 Co. The best resolution obtained is 2.2-2.3 kev. (b) Scan for the M parameter for given M values for the 1.33MeV peak of 6 Co. For a given L parameter there is no significant variation with M. Figure 2: FWHM of the 1.33 MeV peak for different scans of L and M. One of the highest energy resolutions was obtained for the parameters L=6 and M=5. The data file containing this spectrum is called run155.mid, the reference DSPEC file is called Ge-co6-co57.Chn. Both spectra were fitted with Radware to exclude biasing due to different fitting functions and routines. The results are shown in table 5. The values in table 5 show that the energy resolution for tig1 and DSPEC are comparable for high energy photons with the tig1 being about 2% worse. For lower energy photons the energy resolution achieved with the tig1 is about 1.5 times worse than the one obtained with the DSPEC. To check the reliability of fit and the efficiency of the data acquisition a quotient of the different peak areas was calculated ( Peak1 Peak2 [tig1] and Peak1 Peak2 [DSPEC]) (see 4

Position tig1 FWHM DSPEC FWHM FWHM [kev] [kev] [kev] ( tig1 DSPEC ) 122.6 1.456.983 1.48 136.47 1.494.977 1.53 1173.2 2.39 1.74 1.2 1332.5 2.147 1.794 1.2 Table 5: FWHM of the Ge detector at different energies ( 6 Co and 57 Co) achieved with the tig1. Also shown is the resolution achieved with the DSPEC and the fraction fig1/dspec. Both spectra, tig1 and DSPEC, were fitted with Radware. table 6). The error value arises from the uncertainty of the fit calculated by Quotient tig1 DSPEC Peak1/Peak2 1332/1173.848(8).885(3) 1332/136 2.679(45) 2.85(6) 1332/122.332(3).343(1) 136/122.124(2).12(1) Table 6: Quotient of Ge peak areas recorded with tig1 and DSPEC, fitted with Radware. Radware. The quotients agree reasonably well with each other. In an earlier test the detection efficiency of the system was investigated. Two spectra with Ge and the same source ( 57 Co) were taken with DSPEC and tig1. The spectra were normalized to the duration of the data acquisition. Afterwards, the spectra were integrated over three different regions and the peaks at 122 kev and 136 kev (see Figure 7). These integrated counts were then used Integral DSPEC Integral tig1 to form the quotient of DSPEC and tig1 ( ). As values in table 8 show, both systems are equally efficient, the quotient is 1. The spectra were normalized on run time. The settings for the tig 1 were L=28 and M=15. 3.2 LeGe detector energy resolution The same scan that was performed with Ge detector and tig1 was also done with LeGe detector and tig1. The best energy resolution was achieved for similar parameters L and M as with the Ge detector (L=6 and M=5, run155.mid). Both spectra were fitted with Radware to exclude influences due to different fitting functions and routines. The results are listed in table 9. If the spectra are fitted with either root or Origin, the FWHM agree within error. For lower energies the resolution of the tig1 is 1.5 to 2 times worse than that obtained with the DSPEC. But for higher energies, the resolution is comparable in both cases with the DSPEC being slightly better. 5

1 In te g ra l R e g io n I In te g ra l o f 1 2 2 k e V p e a k In te g ra l o f 1 3 6 k e V p e a k [5 /6 /2 9 1 9 : 5 "/G ra p h 1 " (2 4 5 4 9 5 7 )] L in e a r R e g re s s io n fo r C a lib ra tio n _ C : Y = A + B * X P a ra m e te r V a lu e E rro r ------------------------------------------------------------ A.9 7 4 9 2.2 1 7 9 8 B.7 6 7 4 6 2.3 7 7 6 E -5 ------------------------------------------------------------ Region or peak Quotient DSPEC tig1 C o u n ts [3 6 s liv e tim e ] 1 1 1 In te g ra l R e g io n II In te g ra l R e g io n III 122 kev.97(1) 136 kev.97(1) I 1.(1) II 1.2(1) III 1.2(1) 1 2 4 6 8 1 1 2 1 4 1 6 C h a n n e l Table 7: Regions that were integrated to compare the acquisition efficiency of DSPEC and tig1. Table 8: Quotient of integrated regions recorded with DSPEC and tig1. The tig1 spectrum was integrated with root, the DSPEC spectrum was integrated using Origin. Position tig1 FWHM DSPEC FWHM FWHM [kev] [kev] [kev] ( tig1 DSPEC ) 53.681(1).346(4) 1.97 79.6.787(1).435(2) 1.81 81.79(2).441(1) 1.79 276 1.23(6).919(2) 1.11 33 1.121(4).982(28) 1.14 356 1.18(4) 1.147(1).97 Table 9: FWHM of the LeGe detector at different energies ( 133 Ba) achieved with the tig1. Also shown is the resolution achieved with the DSPEC and the fraction fig1/dspec. Both spectra, for tig1 and DSPEC, were fitted with Radware. Huge deviations arise when the quotient is formed with peak intensities relative to each other. Here, the values calculated for tig1 and DSPEC vary by up to a factor 4 as shown in table 1. Assuming that the DSPEC records spectra more Quotient tig1 DSPEC Peak1/Peak2 33/276 2.2(21).945(16) 81/33 18.8(11) 4.8(5) 33/53.11(1).42(6) 81/53 19.876(5) 17.138(117) Table 1: Quotient of LeGe peak areas recorded with tig1 and DSPEC, fitted with Radware. 6

reliable, looking at these values rises a question on the reliability of the read out of the LeGe detector with the tig1. One explanation might be the considerable electrical noise of the tig1 for lower channels. For the Ge the noise contribution was investigated and is presented in chapter 5.4. It was found that below 4 kev the spectrum of the Ge detector strongly depends on the trigger threshold set in the ODB. This limit of 4 kev for the Ge detector corresponds to a limit of about 3 kev for the LeGe detector and thus might explain, why the count rate with the tig1 is about four times the count rate of the DSPEC. Constant noise should not affect the peak area, so it can t be the only explanation why the quotients in table 1 vary that strongly. Due to the transistor reset preamplifier the height of an X-ray event is in the 1 mev range. This results in a FWHM of about 3 channels for the tig1. In comparison with the DSPEC, this is a reduction of a factor six. The FWHM of the DSPEC is about 18 channels (see Figure 3). 1 D S P E C L e G e n o rm a liz e d to tig 1 1 8 s tig 1 ru n tim e 1 C o u n ts [a.u.] 1 1 1 1 2 3 E n e rg y [k e V ] Figure 3: The difference of tig1 and DSPEC in resolution for the low energy region. The FWHM in channels of the tig1 (red line) is about a factor six less than of the DSPEC (black line). Peaks that can be resolved with the DSPEC are not resolvable by the tig1. 4 Trigger on Transistor Reset For the low energy X-ray detector (LeGe) we are using a transistor reset preamplifier. Figure 4 displays examples for transistor resets (yellow signal) together with the inhibit signal (blue). The inhibit signal is a TTL signal of 1 µs width and starts coincidentally with the transistor reset. In some cases the transistor reset is done in steps (see Fig. 4b). The steps are separated in time by about 55 µs each of which is accompanied by the 1 µs inhibit signal. The tig1 triggers on these transistor resets: This is most likely due to the fact that in most 7

(a) (b) Figure 4: LeGe signal during a transistor reset (Yellow) and the inhibit signal (Blue). cases the signal peaks before the transistor resets. 4.1 Transistor Reset inhibit signal To tag triggers on these transistor resets we send the inhibit signal to another channel of the tig1 card. Ideally, we could then exclude those LeGe events whose timestamp are within a certain range of the timestamp of an inhibit signal. Initially, we sent the inhibit signal straight into the tig1 channel. However, the number of events registered by the tig1 in this channel did not agree with those recorded by a separate scaler. Due to the buffer size of 16 events, the difference between the number of events of the scaler and in the tig1 midas data file should equal to +15. However, we observed more counts with the tig1 than on the scaler. We thus concluded that the tig1 cannot handle the 1 µs TTL signal properly. The problem could be avoided by feeding the inhibit signal to a Gate and Delay Generator with a 1 µs output signal. Figure 5 shows the waveforms stored by the tig1 for the TTL inhibit signal and the LeGe signal with the transistor reset. Both channels were individually triggered. For all observed cases, the event builder correctly assigned the same event number for both channels (in the case of Fig. 5 it is event #15). Thus, in principle the LeGe events could be sorted and those events corresponding to transistor resets could be excluded. Also, for transistor resets the tig1 has so far always reported a vanishing energy and will, thus, not be considered as a regular event. But, due to the 8

1.8 x 14 event #115 ch E= TS=572976556 ns 1.6 1.4 1.2 1 18 event #115 ch9 E= TS=572976556 ns 16 14 12 1 8.8.6 inhibit signal 6 4 2 LeGe signal.4 5 1 15 2 25 t [1ch = 1 ns] 5 1 15 2 25 t [1ch = 1 ns] Figure 5: Tig1 waveforms of a transistor reset: TTL inhibit signal on the left and LeGe signal on the right. TTL events 4 events of unknown origin Figure 6: Tig1 charge spectrum for a channel with the TTL inhibit signal. Most TTL signals appear around 88, however there are also four out of 312 notunderstood events on the right. problem of Re-triggering (see 5), it will be essential to correctly identify events due to transistor resets in case the Re-triggering problem cannot be fixed. Note also that in Figure 5 the tig1 reports zero charge (E = ) for the TTL signal. This is due to an incorrect setting of the latency parameter. With the correct settings for the latency parameter, the tig1 assigns a charge value to a 1 µs TTL signal which is different to zero. A test run of 1 hour has been performed and all TTL events were assigned about the same charge value. However, in other individual cases different charge values have been observed for a TTL signal fed into the tig1 (see Figure 6). Their origin is currently not understood. 9

1.8 x 14 event #11 ch E= TS=691336928 ns 2 x 14 event #11 ch9 E= TS=691336928 ns 1.6 1.4 1.2 1 1.5 1 ch9.8.6 ch.4 5 1 15 2 25 t [1ch = 1 ns].5 5 1 15 2 25 t [1ch = 1 ns] 41 event #12 ch9 E=8796 TS=691337568 ns 475 event #13 ch E=1524 TS=691992744 ns 48 46 ch9 47 44 465 42 4 3998 46 455 ch 3996 5 1 15 2 25 t [1ch = 1 ns] 45 5 1 15 2 25 t [1ch = 1 ns] Figure 7: Example for waveforms of re-triggered events. Event #15 is the trigger on the transistor reset. Event #12 and #13 are re-triggered events. 5 Re-triggering We observed that the tig1 sometimes triggers twice on the same event. First it triggers on the correct event, a second trigger appears a bit later. 5.1 Re-trigger on transistor reset Figure 7 shows an example of a trigger on a transistor reset where ch is sampling the waveform of the TTL inhibit signal and ch9 is recording the LeGe signal. In ch9 a second event is being recorded 6.4 µs after the correct transistor reset event, while in ch the second event is recorded about 6.6 ms later. Latter waveform could be a real signal of not understood origin, while the first one is white noise and should not be triggered on. Note that both re-triggered events are stored with a non-zero charge value. The charge values for re-triggered events vary from case to case and will thus add to the background unless they can be correctly identified and removed from the spectra. Re-triggered events have been observed in several scenarios which can be classified as listed in Table 11. 1

Type energy signal inhibit signal I re-trigger? yes yes delay 6.3-6.5 µs 6.5 ms charge non-zero, no systematic around a fixed value waveform noise as in Fig. 6 occurrence 5 out of 826 events II re-trigger? yes yes delay 6.3-6.5 µs differs from case to case, max 6 ms charge non-zero, no systematic small values including zero waveform noise noise occurrence 13 out of 826 events III re-trigger? no yes delay no systematic known charge same as II waveform noise occurrence 1 out of 369 events IV re-trigger? no yes delay 6.5 ms charge same fixed value as in I waveform as in Fig. 6 occurrence 1 out of 369 events V re-trigger? yes no delay 6.2 (once 6.7) µs charge non-zero, no systematic waveform no signal occurrence 8 out of 826 events Table 11: Types of re-triggered events after a transistor reset 11

trigger threshold ratio of events comment with zero charge [%] 5 >.1 3 2 2 34 1 75 probably includes already real noise events, too Table 12: Trigger threshold and re-triggered events: The trigger threshold for the PIPS detector was varied. For lower thresholds the amount of re-triggered events increased. A re-triggered event is indicated by its vanishing charge. It was checked by looking on the waveforms that this is an appropriate indicator. 5.2 Re-trigger on regular events For Germanium (LeGe and Ge) and PIPS detectors we observed re-triggered events also on regular events. The waveforms of these re-triggered events contains no information, i.e. it is noise. More detailed investigations were done for the PIPS detectors. For these detectors, the number of the re-triggered events is trigger threshold dependent. For typical settings the delay between real signal and re-trigger was 6.1 µs in every case. All observed re-triggered events were stored by the tig1 with a charge value of zero. A possible explanation could be found in the waveforms of the detectors. As shown in Fig. 8, all detectors carry some overshoot or ringing on the signal, which the tig1 might have troubles with. Together with Chris Pearson we looked on the waveform processing for individual events on the tig1 card. Ideally, the detector signal and its processed waveforms should look like as it is sketched in Fig. 9a. The width of the pulse in the clipped waveform is fixed to 35 ns. However, we have observed processed waveforms as sketched in Fig. 9b. The width of the clipped waveform was about 1.3 µs and the overshoot of the detector signal could potentially lead to a second trigger depending on the trigger threshold. In this case, the L parameter was set to L = 4.8 µs and the re-triggered event would thus identical to the 6.1 µs which we normally observe for the delay of re-triggered events in the PIPS detector. However, this explanation failed to explain the 6.4 µs delay of the re-triggered events for the transistor reset in the LeGe detector for which the L parameter was set to 2.8 µs. General remark: All tests including the inhibit signal and the re-triggering have been performed before a change in the tig1 frontend in October 29. It is, thus, advisable to investigate these issues again. 5.3 Threshold limit for Ge detector We investigated the behavior of the tig1 for different Trigger Threshold level. Spectra were taken for different trigger threshold values and normalized to 12

8 1.255 x 14 432 7 1.25 43 6 5 4 3 2 LeGe 1.245 1.24 1.235 1.23 PIPS #75855 428 426 424 422 42 small PIPS 1 1 2 t [1ch = 1 ns] 1.225 1 2 t [1ch = 1 ns] 418 1 2 t [1ch = 1 ns] Figure 8: Examples for signals of several detectors. (a) (b) threshold re-trigger? Figure 9: tig1 waveform processing: The clipped waveform leads to an event trigger. Its height is proportional to the detector s signal height. (a) ideal settings (b) observed for small PIPS detector 13

their exact runtime afterwards. Above 4 kev, the spectra are identical, but below 4 kev, the spectra vary strongly depending on the threshold chosen. Depending on the threshold, the intensity of the 14 kev peak of 57 Co changes significant. This suggests that the trigger threshold is not well defined but smeared out. The problem is that it is impossible to reduce the threshold further because the card would trigger on too many noise events. It becomes impossible to determine the peak area of any peak below 4 kev, because the spectrum is depending on the parameters set for the tig1 (see Figure 1). Figure 1: Dependency of the spectrum for different thresholds. The thresholds used for these spectra were 8, 9, 1 and 11. To confirm that this behavior is tig1 related and not an issue of the detector or the positioning of the source, a comparison spectrum with the DSPEC was taken. This spectrum clearly resolves the 14 kev line and results in an energy resolution, that is better by a factor of two. In Figure 11 the spectra of tig1 and DSPEC are presented together with the DSPEC spectrum plotted as dashed black line, whereas the tig1 spectra are plotted similar to Figure 1. More details on the settings of tig1 and DSPEC are listed in the EBIT elog, Message ID 11. All these tests were performed with a firmware of the tig1 that uses only one sampled value to trigger. During these tests the tig1 was shielded against electro-magnetic noise by copper sheets. 5.4 Noise In order to determine the electric noise of the tig1, the pre trigger value was changed to 6 in order to start sampling 6 µs before the actual rise of the signal (see Figure 12). Each event was inspected visually to make sure that in the region of interest there is indeed only noise. Then, the noise channels of each event were Fourier transformed individually and the results added up. During these tests we had some issues with the tig1. For wave forms larger than 6 the tig1 would become unstable. Sometimes non-consistent behavior 14

Figure 11: Comparing threshold of tig1 and DSPEC. The DSPEC has a better defined trigger level and the peak at 14 kev can clearly be resolved. The tig1 has a soft trigger Threshold and doesn t resolve the 14 kev in a reasonable way. The spectra are normalized by their run time. Figure 12: Sampled waveform of a triggered event. The first 6 channel were used for a fast furrier transformation. in the count rate was observable. Occasionally, the count rate without a source was higher than with a source. Additionally, we could find several triggered events, that were just triggered by noise, i.e. without an energy signal. In order to reduce the noise shielding was added to the tig1. On both sides of the card copper sheets were installed inside the VXI crate. The front of the crate was completely closed with copper sheets and the top of the crate was 15

covered with a aluminum grid. The grid was chosen to allow air flow through the crate for cooling. All the sheets were screwed together and connected with a grounding copper cable. This grounding cable was then connected to the ground of the rack. The FFT frequency spectra are shown in Figure 13 and Figure 14. Especially for frequencies higher than 4 MHz, shielding with copper Figure 13: FFT frequency spectrum of the Ge detector BGND signal. Figure 14: FFT frequency spectrum of the Ge detector BGND signal. plates resulted in good noise reduction. For lower frequencies noise reduction due to shielding is very limited. Significant is the peak at 25 MHz that is present in all the spectra. According to Jean Pierre Martin, this peak originates from the sampling process. It should vanish if the K parameter is set to a multiple of 4. For these runs the K value was set to 2, so the noise at this frequency has to be of different origin. The full report of these tests is listed in the EBIT elog at message ID 94 and 95. 5.5 Coincidence test Since the time stamp of an event is recorded, it was possible to create coincidences. This feature wasn t investigated in detail during the tests described above. So far, only coincidences of the same trigger event were used. During TI- TAN s ECBR run with 124,126 Cs it could be demonstrated that the coincidence condition works. Figure 15 shows the spectra of Ge detector on the left and Si (beta) detector on the right. The first row shows spectra without coincidence condition, while the second row exhibits the cleaned spectra with coincidence condition. In the last row the spectrum of the time difference between β + and γ event is illustrated. 16

he[] GE ADC he[] Entries 632477 Mean 212.2 RMS 183.7 PIPS2 ADC 9 he[1] Entries 1154443 Mean 37.4 RMS 69.54 4 1 8 7 3 1 6 5 2 1 4 1 3 2 1 run73.root 5 1 15 2 25 3 35 1 run73.root 2 4 6 8 1 Ge(PIPS2) Coincidence 5 hcoin[] Entries 377 Mean 216.9 RMS 15.5 PIPS2(Ge) Coincidence 35 hcoin[1] Entries 377 Mean 371.5 RMS 7.28 3 4 25 3 2 2 15 1 1 5 run73.root 2 4 6 8 1 12 14 run73.root 5 1 15 2 25 3 35 t_ge - t_pips2 18 htdiff Entries 377 Mean 4223 RMS 465.7 16 14 12 1 8 6 4 2 3 35 4 45 5 run73.root Figure 15: Coincidence spectra taken during the 126 Cs beam time. Gamma spectra of the Ge detector are plotted on the left, while beta spectra are plotted on the right. The first row shows the complete spectrum while the second row consists of spectra cleaned under the coincidence condition. The timing spectrum in the last row illustrated the timing between beta and gamma event. 6 PIPS detectors Before the PIPS detectors 1 were installed in the EBIT, their performance was tested with an analog system and with the tig1. Due to a noisy environment (caused by the vibrations of the vacuum pumps) the tests were performed in the TIGRESS detector lab. An unsealed 27 Bi source (with an activity of 37 kbq measured on Sep. 15, 24) was placed about 2.5 cm above a PIPS detector and β + and conversion electron spectra were taken. The vacuum in the chamber was around 2 mtorr. The analog system consisted of a Canberra Amplifier 226 with a shaping time of 4 µs (coarse gain 1, fine gain 213) and a MCA. Figure 16 shows a spectrum taken with the analog system. Afterwards, a spectrum was taken with the tig1 (K = 4, L = 48). Table 13 compares the energy resolution achieved with both systems. The tig1 s resolution is almost 5 % worse than the one of the analog system. 1 Note that PIPS2 is NOT the small PIPS! 17

35 3 25 # counts 2 15 1 5 1 15 2 25 3 35 Energy [channel] Figure 16: 27 Bi spectrum taken with the PIPS detector and analog electronics. PIPS 1 (#75855) energy [kev] FWHM analog [kev] FWHM tig1 [kev] 481.69 11. 14.7 975.65 1.6 14.6 PIPS 2 energy [kev] FWHM analog [kev] FWHM tig1 [kev] 481.69 1.8 14.7 975.65 1. 14.7 Energy resolution of PIPS detectors for 27 Bi internal conversion elec- Table 13: trons. 18

7 Conclusion For TITAN s coaxial Ge detector the energy resolution of the tig1 for photons of more than 1 MeV is about 2 % worse than those of the DSPEC. In the energy region of 12 kev the resolution of the tig1 is about twice the resolution achieved with the DSPEC. Using a script to take spectra with the tig1 for different parameters resulted in optimized parameters to take gamma spectra. The quotients of peak areas were similar for tig1 and DSPEC. This gives confidence to determine branching ratios with Ge detector and tig1 for photon energies above 4 kev. For photons with less than 4 kev the spectrum is dependent of the trigger threshold. This is an issue for applications at TITAN, since it prohibits one from observing X-ray spectra in the region between 15 kev and 3 kev. Copper shielding that got installed, reduced the high frequency noise but didn t improve the situation with the soft trigger issue below 4 kev. The digitalization of a LeGe detector signal of a transistor reset preamplifier with the tig1 didn t perform well. The energy resolution of the tig1 in the region of 5 kev is twice the resolution of the DSPEC. For higher energies the resolution becomes comparable with the DSPEC being about 1% more precise. Comparing the quotients of different peak areas didn t result in any reasonable values. This might be due to electrical noise. During the tests it became clear that the important points for TITAN s future spectroscopic DAQ are: Resolution of Ge and LeGe detector Performance of the DAQ with the coaxial Ge detector in the low energy range. Are peak heights between 5 and 4 kev dependent of the trigger threshold, e.g. is there a soft trigger threshold? What happens during the reset pulse of the LeGe s preamplifier? Is it possible, to also record the reset inhibit pulse on another channel and clean up the spectrum afterwards? Can we record the signal of betas hitting a PIPS (Si surface barrier) detector? Can we create a coincidence between betas and gammas? A lot of people spent time and energy on these tests. First we would like to thank the TIGRESS group for borrowing us tig1 card and VXI crate. Thanks a lot also to Pierre Amaudruz, Scott Williams and Chris Pearson, who spent many hours trying to adopt the tig1 to our needs. Thank as well to Aaron and Stephan, who spent days in analyzing spectra and trying to figure out what s happening with the detector signals inside the tig1. 19

A tig1 parameter scan F W H M [k e V ] @ 1 2 2. 6 k e V 3. 3. 2.7 5 2.8 2.6 2.5 2.4 2.2 5 2.2 2. 2. 1.7 5 1.8 1.6 1.5 1 2 3 4 5 L P a ra m e te r 6 2 7 8 9 6 5 4 3 M P a ra m e te r F W H M [k e V ] @ 1 3 7.4 7 k e V 1 1.4 2 3 4 5 L P a ra m e te r 6 2 7 8 9 6 5 4 3 M P a ra m e te r 3.2 5 3.2 5 3. 3. 2.7 5 2.7 5 F W H M [k e V ] @ 1.1 7 M e V 2.5 F W H M [k e V ] @ 1.3 3 M e V 2.5 2.2 5 2 1. 2 3 4 5 L P a ra m e te r 6 2 7 8 9 6 5 4 3 M P a ra m e te r 2.2 5 1 2 3 4 5 L P a ra m e te r 6 2 7 8 9 6 5 4 3 M P a ra m e te r Figure 17: K-L-parameter scans for different photon energies taken with tig1 and Ge detector. 2