45EMD Portable Calibrator User Manual

Similar documents
41P Portable Calibrator User Manual

201AP Charge Amplifier User Manual

DC200A Displacement Clipper User Manual

200B Clipper Module User Manual

34134A AC/DC DMM Current Probe. User s Guide. Publication number April 2009

Model 7000 Low Noise Differential Preamplifier

INSTRUCTION MANUAL LKG 601 Electrical Safety Analyzer

Signal Conditioning Amplifier

Model 9305 Fast Preamplifier Operating and Service Manual

Glass Electrode Meter

CEC VIBRATION TRANSMITTER

2001A. 200KHz Function Generator Instruction Manual. 99 Washington Street Melrose, MA Phone Toll Free

OPERATION & SERVICE MANUAL FOR FC 110 AC POWER SOURCE

TRANSDUCER IN-LINE AMPLIFIER

SC200 Series. Signal Conditioner MNX10020, REV H 10/17/ Connection Technology Center, Inc Rae Boulevard Victor, NY (585)

INSTRUCTION MANUAL LKG

Radio Remote(s) (Installation Manual)

Series 500. Owner s Manual. Analog Transmitters by Data Industrial. Data Industrial. Data Industrial 2/95 PN 72806

INSTRUMENTS, INC. Model 2960AX Disciplined Quartz Frequency Standard 2960AX. Section Page Contents

BC145 SIGNAL ISOLATOR BOARD

Model 9302 Amplifier-Discriminator Operating and Service Manual

LINCO MEASUREMENT MODEL CP-2B MASTER METER PROVER COUNTER INSTRUCTION MANUAL

TRANSDUCER IN-LINE AMPLIFIER

A53106 SERIES DC-TO-DC CONVERTER

Model Hz to 10MHz Precision Phasemeter. Operating Manual

ELECTROSURGICAL UNIT ANALYZER

Signal Conditioning Amplifier

Model MV106J/MV116J. ±10nVdc to ±11Vdc Precision DC Voltage Standard Source. Operating Manual

Model 113 Scintillation Preamplifier Operating and Service Manual

Model 3210C. 100 Ampere AC Current Standard. Operating Manual

INSTRUCTION MANUAL FOR CELL SIMULATOR MODEL 2410

SI-125 Power Amplifier Manual 6205 Kestrel Road; Mississauga, Ontario; Canada; L5T 2A1 November 2016, Rev 0.5

Model 863 Quad Timing Filter Amplifier Operating and Service Manual

User s Guide CL309A Milliamp Loop Calibrator. Shop online at omega.com SM

Advanced Test Equipment Rentals ATEC (2832)

VT1586A Rack Mount Terminal Panel Installation and User s Manual

Wilcom MODEL T136BSBZW CIRCUIT TEST SET. Operating Instructions

Wilcom MODEL T136BGMZW CIRCUIT TEST SET. Operating Instructions

AMP-12 OPERATOR S MANUAL

Model 34A. 3Hz to 2MHz 2-Channel Butterworth/Bessel HP, LP, BP, BR Plug-In Filter Card for Model 3905/3916 Chassis.

Model 1791 VHF Radio User's Manual

A Channel Amplifier

An American Control Electronics Brand PCM4 SERIES USER MANUAL PCM4.

PXIe-1209 Dual 100 MHz Pulse Generator

Worldwide Manufacturer of Gas Detection Solutions. Transmitter EC 23. Operation Manual

SX204. Headphone Amplifier. Symetrix. Owner s Manual

Model TS. Transducer Simulator User s Guide. Sensotec Sensors

P5100A & P5150 High Voltage Probes Performance Verification and Adjustments

ECG PATIENT SIMULATOR FIVE LEAD

80i-110s AC/DC CURRENT PROBE

Model 8140VT VERSATAP INSTALLATION AND OPERATION MANUAL

PB-500. Analog Circuits Trainer Instruction Manual

PULSE DISTRIBUTION AMPLIFIER OPERATING MANUAL

Series APD4059. For Load Cells/Pressure Transducer Transmitters. Field Rangeable, with Calibration Resistor. Standard Features.

PARALLEL MULTI-AMP KIT for 7200 Series AMPLIFIERS INSTRUCTION SHEET

VIM Series. 90 & 60 W, Efficient, CV Class 2 LED Drivers ORDERING INFORMATION

Model 439 Digital Current Integrator Operating Manual

AC/DC Current Oscilloscope Probe Model SL261

FRYCORDER Power Cord Burn-in Generator

Operating Manual. Model Channel. DC Signal Conditioner Amplifier

Copyright 2014 by Minarik Drives

Active CCTV Receiver Balun Installation Guide

TECHNICAL MANUAL. SERIES AP5103 DIN-Rail DC Strain Gage Conditioner ISO 9001/AS9100

P5100A & P5150 High Voltage Probes Performance Verification and Adjustments

MMSZ5221BT1 Series. Zener Voltage Regulators. 500 mw SOD 123 Surface Mount

VGM Series. 100 & 60 W, Efficient, CV Class 2 LED Drivers for Signage Applications ORDERING INFORMATION

Harris IRT Enterprises Multi-Channel Digital Resistance Tester Model XR

Model 427A Delay Amplifier Operating and Service Manual

MODEL UBP-10 UNIVERSAL IN-LINE TRANSDUCER AMPLIFIER BI-POLAR SUPPLY, 0-10 VOLT OUTPUT

Advanced Test Equipment Rentals ATEC (2832)

Model 416A Gate and Delay Generator Operating and Service Manual

M302RM OPERATING MANUAL

AM-510 Commercial / Residential Multimeter. AM-510-EUR Digital Multimeter. Users Manual

CSM-S USER S MANUAL TRIGGER DISTRIBUTION MODULE Release April 7, VXI Technology, Inc.

MODEL 3810/2 Line Impedance Stabilization Network

B MTS Systems Corp., Model Function Generator

Classic Series Public Address Amplifiers C10 & C20 Models

DL102 Counter Loop Amplifier

Pen Multimeter. Model

Model 4402B. Ultra-Pure Sinewave Oscillator 1Hz to 110kHz Typical Distortion of % Serial No. Operating Manual

Type N Coaxial Calibration Kit

Meriam Model M ma Loop Calibrator Operating Instructions

WRM-10 TM TRANSFORMER WINDING RESISTANCE METER

MODEL FS-4 INSTRUCTION MANUAL R.L. DRAKE COMPANY, MIAMISBURG, OHIO, U.S.A.

HydroLynx Systems, Inc.

IDEAL INDUSTRIES, INC. TECHNICAL MANUAL MODEL:

Safety. This symbol, adjacent to a terminal, indicates that, under normal use, hazardous voltages may be present.

SDI SPECTRADYNAMICS, INC. HIGH PERFORMANCE DISTRIBUTION AMPLIFIER OPERATING MANUAL

HTA125A/250A. Power Amplifiers. Installation & Use Manual

Model 533 Dual Sum and Invert Amplifier Operating and Service Manual

User s Guide. MultiView Series Digital MultiMeters Models: MV110 MV120 MV130

Green T-DFN Part Number Compliance Case Packaging LBS10-13 Commercial T-DFN ,000/Tape & Reel

AL8805EV1 EVALUATION BOARD USER GUIDE

ECG PATIENT SIMULATOR SIX LEAD

NCL30000LED2GEVB/D Vac up to 15 Watt Dimmable LED Driver Demo Board Operation EVALUATION BOARD MANUAL

Four-Channel Differential AC Amplifier

Installation & User Guide. For Powering Distributed Audio Systems A45-X2 TWO CHANNEL AMPLIFIER

R-X SERIES. Decade Resistor

Aluminum Electrolytic Capacitors Power Ultra Miniature Snap-In

OPERATION MANUAL LDM-1000

Transcription:

Trig-Tek 45EMD Portable Calibrator User Manual Publication No. 980958 Rev. B Astronics Test Systems Inc. 4 Goodyear, Irvine, CA 92618 Tel: (800) 722-2528, (949) 859-8999; Fax: (949) 859-7139 atsinfo@astronics.com atssales@astronics.com atshelpdesk@astronics.com http://www.astronicstestsystems.com Copyright 2011 by Astronics Test Systems Inc. Printed in the United States of America. All rights reserved. This book or parts thereof may not be reproduced in any form without written permission of the publisher.

THANK YOU FOR PURCHASING THIS ASTRONICS TEST SYSTEMS PRODUCT For this product, or any other Astronics Test Systems product that incorporates software drivers, you may access our web site to verify and/or download the latest driver versions. The web address for driver downloads is: http://www.astronicstestsystems.com/support/downloads If you have any questions about software driver downloads or our privacy policy, please contact us at: atsinfo@astronics.com WARRANTY STATEMENT All Astronics Test Systems products are designed to exacting standards and manufactured in full compliance to our AS9100 Quality Management System processes. This warranty does not apply to defects resulting from any modification(s) of any product or part without Astronics Test Systems express written consent, or misuse of any product or part. The warranty also does not apply to fuses, software, non-rechargeable batteries, damage from battery leakage, or problems arising from normal wear, such as mechanical relay life, or failure to follow instructions. This warranty is in lieu of all other warranties, expressed or implied, including any implied warranty of merchantability or fitness for a particular use. The remedies provided herein are buyer s sole and exclusive remedies. For the specific terms of your standard warranty, contact Customer Support. Please have the following information available to facilitate service. 1. Product serial number 2. Product model number 3. Your company and contact information You may contact Customer Support by: E-Mail: atshelpdesk@astronics.com Telephone: +1 800 722 3262 (USA) Fax: +1 949 859 7139 (USA)

RETURN OF PRODUCT Authorization is required from Astronics Test Systems before you send us your product or sub-assembly for service or calibration. Call or contact Customer Support at 1-800-722-3262 or 1-949-859-8999 or via fax at 1-949-859-7139. We can also be reached at: atshelpdesk@astronics.com. If the original packing material is unavailable, ship the product or sub-assembly in an ESD shielding bag and use appropriate packing materials to surround and protect the product. PROPRIETARY NOTICE This document and the technical data herein disclosed, are proprietary to Astronics Test Systems, and shall not, without express written permission of Astronics Test Systems, be used in whole or in part to solicit quotations from a competitive source or used for manufacture by anyone other than Astronics Test Systems. The information herein has been developed at private expense, and may only be used for operation and maintenance reference purposes or for purposes of engineering evaluation and incorporation into technical specifications and other documents which specify procurement of products from Astronics Test Systems. TRADEMARKS AND SERVICE MARKS All trademarks and service marks used in this document are the property of their respective owners. Racal Instruments, Talon Instruments, Trig-Tek, ActivATE, Adapt-A-Switch, N-GEN, and PAWS are trademarks of Astronics Test Systems in the United States. DISCLAIMER Buyer acknowledges and agrees that it is responsible for the operation of the goods purchased and should ensure that they are used properly and in accordance with this document and any other instructions provided by Seller. Astronics Test Systems products are not specifically designed, manufactured or intended to be used as parts, assemblies or components in planning, construction, maintenance or operation of a nuclear facility, or in life support or safety critical applications in which the failure of the Astronics Test Systems product could create a situation where personal injury or death could occur. Should Buyer purchase Astronics Test Systems product for such unintended application, Buyer shall indemnify and hold Astronics Test Systems, its officers, employees, subsidiaries, affiliates and distributors harmless against all claims arising out of a claim for personal injury or death associated with such unintended use.

FOR YOUR SAFETY Before undertaking any troubleshooting, maintenance or exploratory procedure, read carefully the WARNINGS and CAUTION notices. This equipment contains voltage hazardous to human life and safety, and is capable of inflicting personal injury. If this instrument is to be powered from the AC line (mains) through an autotransformer, ensure the common connector is connected to the neutral (earth pole) of the power supply. Before operating the unit, ensure the conductor (green wire) is connected to the ground (earth) conductor of the power outlet. Do not use a two-conductor extension cord or a three-prong/two-prong adapter. This will defeat the protective feature of the third conductor in the power cord. Maintenance and calibration procedures sometimes call for operation of the unit with power applied and protective covers removed. Read the procedures and heed warnings to avoid live circuit points. Before operating this instrument: 1. Ensure the proper fuse is in place for the power source to operate. 2. Ensure all other devices connected to or in proximity to this instrument are properly grounded or connected to the protective third-wire earth ground. If the instrument: - fails to operate satisfactorily - shows visible damage - has been stored under unfavorable conditions - has sustained stress Do not operate until performance is checked by qualified personnel.

Publication No. 980958 Rev. B 45EMD User Manual Table of Contents Chapter 1... 1-1 Introduction... 1-1 Features... 1-2 Description... 1-2 Specifications... 1-3 Frequency... 1-3 Amplitude... 1-3 Power... 1-3 Dimensions... 1-3 Chapter 2... 2-1 Operation... 2-1 Rechargeable Battery... 2-1 FREQUENCY and X1-X10 Switches... 2-1 AMPLITUDE and X1-X10 Switches... 2-2 UNITS mv/pc Switch... 2-3 OFFSET Controls... 2-3 Chapter 3... 3-1 Performance Test... 3-1 Test Equipment... 3-1 Switch Settings... 3-1 Performance Test... 3-1 Chapter 4... 4-1 Calibration Procedure... 4-1 Test Equipment... 4-1 Adjustment Potentiometers... 4-1 Calibration Procedure... 4-1 Astronics Test Systems i

45EMD User Manual Publication No. 980958 Rev. B List of Figures Figure 1-1, 45EMD Portable Calibrator... 1-1 Figure 2-1, 45EMD Control Switches... 2-2 Figure 4-1, Potentiometer Adjustment Locations... 4-2 ii Astronics Test Systems

Publication No. 980958 Rev. B 45EMD User Manual DOCUMENT CHANGE HISTORY Revision Date Description of Change A 02/02/2011 Document Control release B 01/28/2013 Revised per ECN01650 updating AC accuracy spec in Chapter 1 and Performance Test in Chapter 3. Astronics Test Systems iii

45EMD User Manual Publication No. 980958 Rev. B This page was left intentionally blank. iv Astronics Test Systems

Publication No. 980958 Rev. B 45EMD User Manual Chapter 1 Introduction The 45EMD Portable Calibrator (Figure 1-1) is designed to provide precision signals for field calibration of measurement instrumentation. This hand-held unit generates the required outputs to simulate a variety of transducers including temperature, pressure and vibration, with laboratory accuracy. It is powered by a rechargeable NICAD battery and operates for approximately 6 continuous hours on a single charge. The 45EMD s output frequency is crystal controlled and accurate to.05%. Frequency is selectable from 10 to 999.9 Hz with 0.1 Hz resolution or 100 to 9999 Hz with 1.0 Hz resolution. A four-digit thumb switch and an X1 or X10 switch allow mv or pc level to be set from 0.1 to 999.9 with a 0.1 resolution or from 1.0 to 9999 with 1.0 resolution. Selectable output units include DC, PK, PK-PK or RMS. The mv output signal is on a BNC and the pc on a microdot connector. A differential pc output is provided on a three-pin connector. A DC offset voltage with selectable polarity can be superimposed on the AC outputs. Also a constant synchronous square wave output is provided. A carrying case and battery charger are included. Figure 1-1, 45EMD Portable Calibrator Astronics Test Systems Introduction 1-1

45EMD User Manual Publication No. 980958 Rev. B Features Features include: Laboratory Accuracy 10 to 9999 Hz (XTAL) 0.1 to 9999 mv or pc Differential pc Output Square wave Output Adjustable DC Offset Carrying Case Battery Charger Description The 45EMD Portable Calibrator is a convenient signal generator and is especially useful for field testing of monitoring systems. It has the required outputs to simulate a variety of pickups, temperature pressure, vibration (acceleration, velocity or displacement) or can be used in a laboratory as a precision signal source. The unit has a rechargeable NICAD battery that provides approximately 6 hours of continuous use before recharging. The 45EMD has many features. It provides a crystal controlled frequency output with a.05% accuracy. Its frequency is set using a four-digit thumbswitch working with an X1-X10 switch thus the frequency can be set with 0.1 Hz resolution from 10 to 999.9 Hz or 1 Hz resolution from 100 Hz to 9999 Hz. A UNITS switch selects OFF, DC, PEAK, PK-PK, or RMS settings. A four-digit thumbswitch selects in conjunction with an X1-X10 switch provides 0.1 to 999.9 mv (millivolts) and pc (picocoulomb), with 0.1 mv and pc resolution or 1 to 9999 with 1 mv and pc resolution. The mv output is on a BNC and the pc output is on a microdot connector. Also a differential pc signal is available on a three-pin connector (MS3102A.105L-3S). For convenience, the unit has a constant square wave output, on a BNC at the set frequency to be used as a sync or reference signal. A DC Offset voltage can be superimposed on the AC output signals this is accomplished by a minus-off switch to allow plus or minus 0 to 12 Volts with a single turn control. Introduction 1-2 Astronics Test Systems

Publication No. 980958 Rev. B 45EMD User Manual Specifications Frequency Amplitude Range 10 to 9999 in two ranges (usable to 1 Hz) Accuracy ±0.05% Resolution 0.1 Hz X1 range 10.0-999.9 Hz 1 Hz X10 range 100-9999 Hz The mv (Millivolt) output is converted to picocoulomb at the pc (Picocoulomb) output. The Frequency is the same; the Picocoulomb level is the same as the Millivolts ±0.3%. Impedance AC Accuracy Distortion DC Accuracy mv (millivolts) less than 50 Ohms pc (picocoulomb) 1000 pf capacitor 10 to 9999 Hz 0.5% of reading ±0.3 FS (for set outputs) SQ Wave 5 Volt p-p ±0.5V Less than 1% (30 to 9999 Hz) ±0.5% of setting ±0.2% FS Power 120 or 230 VAC battery charger 9 VDC at 200 ma output Dimensions 4.75 wide x 6.5 high x 2 deep. Astronics Test Systems Introduction 1-3

45EMD User Manual Publication No. 980958 Rev. B This page was left intentionally blank. Introduction 1-4 Astronics Test Systems

Publication No. 980958 Rev. B 45EMD User Manual Chapter 2 Operation The 45EMD Portable Calibrator is contained in a high impact plastic case. It is light and very durable. It provides settings of 000.1 to 999.9 mv or pc output levels with an X10 to provide 0001 to 9999 mv or pc at the two separate output jacks. The battery used provides more than 6 hours of operation before recharge is needed. The 45EMD Portable Calibrator is a straight forward instrument with selfexplanatory controls. The operation of each control is given and information regarding the battery charging. Rechargeable Battery The 45EMD Portable Calibrator contains a six-volt rechargeable battery that should give long, trouble-free service. The battery will supply power to the 45EMD for more than 6 hours when fully charged. In order to fully recharge the battery, it is necessary to plug the charge adapter into 115 V RMS power and connect to the 45EMD Portable Calibrator. To extend the life of the battery, it should be completely discharged before recharging. The battery requires ten hours to fully charge. The 45EMD provides circuitry such that the charger may be left pluggedin indefinitely without damaging the battery. When the low battery (LOW BATT) light illuminates, it does not necessary mean the battery is fully discharged. It does mean that the battery has dropped below the voltage required to maintain the specifications for the 45EMD. The 45EMD will probably continue to operate, but the output voltages may not be within specifications to completely discharge the battery leave the unit ON until the Low Battery light goes out, then charge the battery. FREQUENCY and X1-X10 Switches The frequency is set by a four station thumbswitch to set from 0001 to 9999. Incorporated with this switch is an X1-X10 slide switch to accomplish two full scale frequencies of 999.9 Hz or 9999 Hz. The low range has a resolution of 0.1 Hz; the high range has 1 Hz resolution. Astronics Test Systems Operation 2-1

45EMD User Manual Publication No. 980958 Rev. B Frequency and X1-X10 Switches Amplitude and X1-X10 Switches Units mv/pc Switch Offset Controls Figure 2-1, 45EMD Control Switches AMPLITUDE and X1-X10 Switches The amplitude is set using a four-digit thumbwheel switch (Figure 2-1) providing settings from 000.0 to 999.9 mv, or pc amplitude for the units selected by the UNITS switch. The 000.1 to 999.9 Amplitude setting can be multiplied by an X1- X10 Switch. Refer to Figure 2-1. When switched to X1 the amplitude is 000.0 to 999.9 mv and pc. In the X10 position, amplitudes of 0000 to 9999 mv and pc can be set. Operation 2-2 Astronics Test Systems

Publication No. 980958 Rev. B 45EMD User Manual UNITS mv/pc Switch This UNITS mv/pc switch (Figure 2-1) has five positions. The first is DC. When the switch is in this position, only DC voltages appear at the mv (millivolt) output. The output level of the DC is determined by the AMPLITUDE switch, and can provide outputs of 000.1 to 999.9 millivolts or 0001 to 9999 millivolts DC in two ranges by placing the X1-X10 Switch to X10. The +/ Switch selects either a positive or negative DC at the millivolt output. The next position of the switch is PEAK. When in this position the AMPLITUDE is calibrated for AC millivolts peak at the mv (millivolt) output. The output at the BNC connector labeled pc (picocoulomb) is a charge output equivalent to the applied voltage through a precision capacitor. The precision 0.3% capacitor is 1000 pf. Hence, if 100 mv PEAK is set in on the front panel switch, a voltage of 100 mv is applied through the 1000 pf capacitor providing an equivalent charge of 100 pc peak at the BNC pc output. An MS3106-105L-3S connector also provides a differential picocoulomb output. The third position of the switch is AC PK-PK. When in this position the mv (millivolt) and pc (picocoulomb) outputs are calibrated in PK-PK units. While in the PK-PK position the mv output can be offset 0 to 12 V DC adjustable by the DC SET control. When no DC is needed be sure the DC OFFSET Switch is in the OFF position. OFFSET Controls The OFFSET Controls (Figure 2-1) are the X1-OFF-X10 Switch, the single turn OFFSET control and the +/ SWITCH. Either 0-1.2V DC in the X1 position of the OFFSET Switch or 0-12V DC in the X10 position can be set by the OFFSET control. The polarity of the offsets is set by the +/ SWITCH. Astronics Test Systems Operation 2-3

45EMD User Manual Publication No. 980958 Rev. B This page was left intentionally blank. Operation 2-4 Astronics Test Systems

Publication No. 980958 Rev. B 45EMD User Manual Chapter 3 Performance Test The following performance test procedure is a method of testing the 45EMD Portable Calibrator unit to verify that it is performing within the manufacturer s specifications. The unit uses integrated circuits and very stable parts and should not require calibration more often than every twelve months. In the event that a reading is out of tolerance, refer to Chapter 4, Calibration Procedures. Test Equipment Note: Equivalent equipment can be used. AC/DC voltmeter Keithley 191 Frequency Counter Leader LDC-822 Switch Settings Set the FREQUENCY switch to 100.0 Hz X1. Set AMPLITUDE switch to indicate 0000. Set UNITS switch to DC. Set the AMPLITUDE X1, X10 switch to X10. Turn the OFFSET knob to the full counterclockwise position. The offset X1-OFF-X10 to switch to OFF. Place the +/ Switch to the (+) position. Performance Test 1. Connect the DC voltmeter to the mv OUTPUT (BNC) jack. 2. Observe an indication of 0 ±20.0 mv on the DC voltmeter. 3. Set the AMPLITUDE switch to 9999. 4. Observe an indication of 9999 ±70 mv on the DC voltmeter. 5. Set the AMPLITUDE X1-X10 switch to X1. 6. Set the AMPLITUDE switch to 100.0. 7. Observe an indication of 100 ±2.5 mv on the DC voltmeter. Astronics Test Systems Performance Test 3-1

45EMD User Manual Publication No. 980958 Rev. B 8. Set the AMPLITUDE switch to 010.0. 9. Observe an indication of 10.0 ±2.1 mv on the DC voltmeter. 10. Set the AMPLITUDE switch to 900.0, and the UNITS switch to RMS. 11. Connect an AC voltmeter to the mv OUTPUT (BNC) jack. 12. Observe an indication of 900 ±7.5 mv RMS on the AC voltmeter. 13. Set the UNITS switch to PEAK. 14. Observe an indication of 636 ±5.0 mv RMS on the AC voltmeter. 15. Set the UNITS switch to PK-PK. 16. Observe an indication of 318 ±2.5 mv RMS on the AC voltmeter. 17. Set the Frequency Switch to 999.9. 18. Connect the Frequency Counter to the mv OUTPUTS (BNC) jack. 19. Observe an indication of 999.9 ±0.5 Hz on the Frequency Counter. 20. Set the FREQUENCY X1-X10 Switch to X10. 21. Observe an indication of 9999 ±5.0 Hz on the Frequency Counter. 22. Return the UNITS switch to RMS. 23. Observe an indication of 900 ±7.5 mv RMS on the AC voltmeter. 24. Set the AMPLITUDE X1-X10 switch to X10. 25. Observe a 9000 ±75 mv RMS on the AC voltmeter. Performance Test 3-2 Astronics Test Systems

Publication No. 980958 Rev. B 45EMD User Manual Chapter 4 Calibration Procedure The unit has been carefully calibrated using frequency counters and voltmeters that are traceable to the National Bureau of Standards. Do not make any adjustments unless properly calibrated equipment is available. Make sure the battery is fully charged. Test Equipment Note: Equivalent equipment can be used. Frequency Counter Leader LDC-822 Digital Multimeter Keithley 191 Adjustment Potentiometers There are five separate adjustments inside the 45EMD Portable Calibrator. See Figure 4-1. Remove the back cover of the 45EMD by removing the six screws found on the back of the Calibrator. Pull the cover back and lift it up and off. Place the unit in front of you with the front panel down. Looking down into the calibrator you will see the five adjustment potentiometers. The five adjustment potentiometers provide the following. R1 Peak Level R2 RMS Level R3 DC Level R4 DC Zero R5 Factory Zero Adjust (factory adjustment only) Calibration Procedure 1. Connect a DC voltmeter to the MV (BNC) OUTPUT jack. 2. Set FREQUENCY switch to 5000 X1. 3. Place the UNITS switch to DC. (The DC OFFSET Control must be completely counterclockwise), and OFFSET switch to OFF. Astronics Test Systems Calibration Procedure 4-1

45EMD User Manual Publication No. 980958 Rev. B 4. Set the AMPLITUDE switch to 0000 X1. 5. Adjust the DC Zero ADJ R4 for a 0 ±0.1 mv indication on the DC Voltmeter. 6. Set the AMPLITUDE switch to 900.0. 7. Set the DC LEVEL ADJ R3 for a 900 ±1.0 mv indication on the DC voltmeter. 8. Connect an AC voltmeter to the mv (BNC) jack. 9. Place the UNITS switch to RMS. Make sure DC OFFSET Control is in the full counterclockwise position and OFFSET Switch to OFF. 10. Set the RMS LEVEL ADJ R2 for a 900 ±1.0 mv RMS indication on the AC voltmeter. 11. Place the UNITS switch to PEAK. 12. Set the PEAK LEVEL ADJ R1 for a 636.3 ±1.0 mv RMS indication on the AC voltmeter. A high grade 0.3% precision 1000 pf capacitor is in series with the BNC output connector. The above calibrations have automatically adjusted the pc (picocoulomb) output, as these calibrated voltages are used via the capacitor to the pc (picocoulomb) output. Figure 4-1, Potentiometer Adjustment Locations Calibration Procedure 4-2 Astronics Test Systems