Optimisation of Image Acquisition Bordeaux 16th November J.S. McGhie W.B. Vletter R. Frowijn No disclosures

Similar documents
Lesson 06: Pulse-echo Imaging and Display Modes. These lessons contain 26 slides plus 15 multiple-choice questions.

Physics of Ultrasound Ultrasound Imaging and Artifacts รศ.นพ.เดโช จ กราพาน ชก ล สาขาหท ยว ทยา, ภาคว ชาอาย รศาสตร คณะแพทยศาสตร ศ ร ราชพยาบาล

The Physics of Echo. The Physics of Echo. The Physics of Echo Is there pericardial calcification? 9/30/13

Lesson 06: Pulse-echo Imaging and Display Modes. This lesson contains 22 slides plus 15 multiple-choice questions.

4 Working With Scan Modes

Ultrasound Beamforming and Image Formation. Jeremy J. Dahl

Ultrasound & Artifacts

Artifacts. Artifacts. Causes. Imaging assumptions. Common terms used to describe US images. Common terms used to describe US images

Ultrasound Imaging Ultr Michael Dadd 2007

The physics of ultrasound. Dr Graeme Taylor Guy s & St Thomas NHS Trust

12/26/2017. Alberto Ardon M.D.

Lesson 12: Doppler Principles. This lesson contains 50 slides plus 26 multiple-choice questions.

SONOGRAPHIC PHYSICS, INSTRUMENTATION & DOPPLER REVIEW Part 3

Answer: TGC is needed to amplify echoes from deeper structures so that they appear as bright as similar structures located at more shallow depths.

Ultrasound Physics. History: Ultrasound 2/13/2019. Ultrasound

Image Optimization: The Sonographer s Responsibility. Prepared by Cathy Daniels, EdD, RTR, RDMS, RDCS, RVT

Nuove tecnologie per ecografia ad ultrasuoni: da 2D a 4D

Principles of Ultrasound Imaging Image Optimization

Lesson 02: Sound Wave Production. This lesson contains 24 slides plus 11 multiple-choice questions.

Physics of ultrasound

Introduction to Ultrasound Physics

S S S2 Operation Manual

3. Ultrasound Imaging(2)

Key Physics and Doppler Principles

Ultrasound physical principles in today s technology

Architecture of Quality Imaging Mary K. Henne, MS, CNMT, RDMS, RVT Ultrasound Education Specialist GE Healthcare

Physics of Ultrasound & Doppler. Sang Jae Rhee. MD., PhD. Division of Cardiovascular Medicine Wonkwang University Hospital

Doppler in Obstetrics: book by K Nicolaides, G Rizzo, K Hecher. Chapter on Doppler ultrasound: principles and practice by Colin Deane

Pass Ultrasound Physics Exam

Breast Ultrasound QA Phantom Recommended by Japan Association of Breast and Thyroid Sonology

Ultrasound Bioinstrumentation. Topic 2 (lecture 3) Beamforming

Lecture 19. Ultrasound Imaging

Introduction to Medical Engineering (Medical Imaging) Ultrasound Imaging. Ho Kyung Kim Pusan National University

Ihor TROTS, Andrzej NOWICKI, Marcin LEWANDOWSKI

Quick Reference Guide

Echo Artifacts: The Cause and Solution

Photomultiplier Tube

Design Your Performance

DC-6. Diagnostic Ultrasound System

CHAPTER 1 INTRODUCTION

The Middle East Distributor for AMBISEA Technology Corp. Electro-Medical Product Line

An Overview Algorithm to Minimise Side Lobes for 2D Circular Phased Array

SODAR- sonic detecting and ranging

Introduction. Parametric Imaging. The Ultrasound Research Interface: A New Tool for Biomedical Investigations

Diagnostic Ultrasound System. Operation Note

Sonic Distance Sensors

Chapter 4. Pulse Echo Imaging. where: d = distance v = velocity t = time

Medical Imaging (EL582/BE620/GA4426)

Attenuation and velocity of ultrasound in solid state materials (transmission)

Ultrasonic Linear Array Medical Imaging System

EVIS EUS ENDOSCOPIC ULTRASOUND CENTER EU-ME2 Dedicated ultrasound processor with versatile functionality.

Multi-Element Synthetic Transmit Aperture Method in Medical Ultrasound Imaging Ihor Trots, Yuriy Tasinkevych, Andrzej Nowicki and Marcin Lewandowski

Doppler Ultrasound. Amanda Watson.

Easy Ultrasonic Phased Array Inspection of Corrosion - Resistant Alloys and Dissimilar Weld Materials

Performing ultrasound probe quality assurance assessments: A How-to Guide

Explain what is meant by a photon and state one of its main properties [2]

Endoscopic Ultrasonography System

Equipment for Attenuation and velocity of ultrasound in solid state materials (transmission), experimental set-up

DC-6 Expert. Diagnostic Ultrasound System

Basic Radar Definitions Introduction p. 1 Basic relations p. 1 The radar equation p. 4 Transmitter power p. 9 Other forms of radar equation p.

COMPUTER PHANTOMS FOR SIMULATING ULTRASOUND B-MODE AND CFM IMAGES

Interaction of Sound and. logarithms. Logarithms continued. Decibels (db) Decibels (db) continued. Interaction of Sound and Media continued

ENDOSCOPIC ULTRASOUND SYSTEMS

The Script of ZST + Presentation. MIS Upstream Marketing Team [ 日期 ]

Medical Imaging. X-rays, CT/CAT scans, Ultrasound, Magnetic Resonance Imaging

UGEO H60. Performance in Style. Features

inter.noise 2000 The 29th International Congress and Exhibition on Noise Control Engineering August 2000, Nice, FRANCE

Fig. 1

M5 Diagnostic Ultrasound System

INTRODUCTION TO RADAR SIGNAL PROCESSING

Multi-spectral acoustical imaging

FREQUENTLY ASKED QUESTIONS

ADAPTIVE CORRECTION FOR ACOUSTIC IMAGING IN DIFFICULT MATERIALS

An acousto-electromagnetic sensor for locating land mines

Ground Penetrating Radar

Phased Array Inspection of Coarse Grain Welds (Austenitic, CRA, etc)

ENDOSCOPIC ULTRASOUND SYSTEMS

MAKING TRANSIENT ANTENNA MEASUREMENTS

Standard Guide for Evaluating Performance Characteristics of Phased-Array Ultrasonic Testing Instruments and Systems 1

Ques on (2): [18 Marks] a) Draw the atrial synchronous Pacemaker block diagram and explain its operation. Benha University June 2013

Set No.1. Code No: R

Simulation of Ultrasonic Testing of Rail Wheel Face using Phased Array and DDF technique

Physics in Modern Medicine Fall 2010

Real Time Deconvolution of In-Vivo Ultrasound Images

ACOUSTIC MICRO IMAGING ANALYSIS METHODS FOR 3D PACKAGES

REAL-TIME B-SCAN ULTRASONIC IMAGING USING A DIGITAL PHASED. Robert Dunki-Jacobs and Lewis Thomas General Electric Company Schenectady, New York, 12301

SU-1 EG-580UT EG-580UR ENDOSCOPIC ULTRASONOGRAPHY SYSTEM

Evaluation of automatic time gain compensated in-vivo ultrasound sequences

Biomedical. Measurement and Design ELEC4623/ELEC9734. Electrical Safety and Performance Standards

Optimization of Axial Resolution in Ultrasound Elastography

PHYSICALLY, the speed of sound in human tissue limits


Velocity Estimation in muscular Tissue by Ultrasound

MULTI-FREQUENCY ULTRASOUND IMAGING: PHANTOM STUDY

Further Developments in Ultrasonic Phased Array Inspection of Aging Aircraft

INSPECTION OF THERMAL BARRIERS OF PRIMARY PUMPS WITH PHASED ARRAY PROBE AND PIEZOCOMPOSITE TECHNOLOGY

We are IntechOpen, the world s leading publisher of Open Access books Built by scientists, for scientists. International authors and editors

MR Advance Techniques. Flow Phenomena. Class II

IMAGING OF DEFECTS IN CONCRETE COMPONENTS WITH NON-CONTACT ULTRASONIC TESTING W. Hillger, DLR and Ing. Büro Dr. Hillger, Braunschweig, Germany

BEAM DISTORTION IN DOPPLER ULTRASOUND FLOW TEST RIGS: MEASUREMENT USING A STRING PHANTOM

Transcription:

Optimisation of Image Acquisition Bordeaux 16th November 2016 J.S. McGhie W.B. Vletter R. Frowijn No disclosures

Image optimisation: The Echo machine It looks difficult to drive an echo machine!!

Some things never change they just look better 1975 2016 Current machines control much of the image and signal processing

Image optimisation: Echocardiographic machine capable of: - 2D imaging, M-mode - pulsed & continuous wave Doppler - colour flow Doppler mapping - electrocardiogram - analysis software - optional: tissue doppler analysis, 3D echocardiography. Machine operator: - fully trained,dedicated,sonographer,physicians - continuing education credits

Image optimisation: Knobology Echo machine controls can be grouped according to application: - Imaging Controls - Spectral Doppler Controls - Colour Flow Mapping Controls - Display, Measurement and Recording - Set Up Differences exist in manufactures with respect to how much OPERATOR control is allowed.

Image optimisation: The Echo machine Monitor: Adjust the contrast and brightness so that both the weakest and strongest gray levels are present on the screen System preset: Standard for your lab to produce uniform appereance of studies. A MUST in colour Doppler maps ie33

Image optimisation Gain iscan Depth Focus TGC

Image optimisation

Image optimisation: Technical skill Patient position - left (right) lateral decubitus position supine position Selection of the transducer Higher frequency: Better resolution less penetration (pediatric) Lower frequency: Poorer resolution better penetration (adult) Transducer position - Minor adjustment of transducer position can provide a better acoustic window. - Consider contact between transducer tip and skin Focus

Image optimisation: Selection of the transducer Fundamental S12-4 Best spatial resolution Most impaired penetration 2 nd Harm. 1.3 2.6Mhz Impaired spatial resolution Most possible penetration The worse the acoustic window the lower the frequency

Image optimisation: Technical skill Patient position - left (right) lateral decubitus position supine position Selection of the transducer Higher frequency: Better resolution less penetration (pediatric) Lower frequency: Poorer resolution better penetration (adult) Transducer position - Consider contact between transducer tip and skin - Minor adjustment of transducer position can provide a better acoustic window. Focus

Image optimisation: Transducer position

Image optimisation: Transducer position

Image optimisation: Focus To optimize resolution at a specific distance Structures proximal to the focus level are better visualized

Image optimisation: Gain settings Gain setting Overall gain Adjusts the amplitude of the received signals over the total length of the ultrasound beam. Time-gain compensation Allows differential adjustments along this length to compensate for the effect of attenuation Compression The amplitude range of the reflected signal is compressed into a range of values from white to black

Image optimisation: Gain setting Correct gain setting Overall gain Time gain compensation (TGC)

Image optimisation: Compression Gainsetting Overall gain Adjusts the amplitude of the received signals over the total length of the ultrasound beam. Time-gain compensation Allows differential adjustments along this length to compensate for the effect of attenuation Compression The amplitude range of the reflected signal is compressed into a range of values from white to black

Image optimisation: Compression Compress 30 Gain 62% Compress 40 Gain 62% Compress 50 Gain 62% Compress 60 Gain 62% To provide an image with a gradation of gray levels the number of levels of gray can be adjusted by the compress / dynamic range setting Default setting 50 55 (ie 33)

Image optimisation: Tissue Harmonic Imaging Tissue Harmonic Imaging In tissue harmonic imaging the harmonic frequency energy is generated as the ultrasonic wave propagates through the tissue By processing the received signals the second harmonic is filtered out and displayed Framerate Number of images per second depends on the number of scan lines and adjusted depth (routine setting 40 frames per second)

Image optimisation: Fundamental 2 nd Harmonic Fundamental (FR 75Hz) Gain 44% 2 nd Harm. 1.7 3.4Mhz (FR 64Hz) Gain 62% Harmonic imaging improves signal-to-noise-ratio. Better lateral resolution poorer axial resolution

Image optimisation: Fundamental 2 nd Harmonic Important The type of processing required to filter out the received harmonic signal does make structures within the heart appear slightly thicker in harmonic as compared with fundamental frequency

Image optimisation: 2 nd Harmonic - XRES

Image optimisation: Framerate Tissue Harmonic Imaging In tissue harmonic imaging the harmonic frequency energy is generated as the ultrasonic wave propagates through the tissue By processing the received signals the second harmonic is filtered out and displayed Framerate Number of images per second depends on the adjusted depth, sector angle and number of scan lines. (routine setting 40 frames per second)

Image optimisation: Framerate Begin with sufficient depth to see beyond the heart Reduce depth till area of interest fills the screen Depth 19cm = FR 50Hz Depth 13cm = FR 55Hz

Image optimisation: Framerate Decrease sector angle Decrease line density Depth 13cm = FR 72Hz (speckle tracking) Depth 13cm = FR 88Hz

Image optimisation: Framerate X5-1 matrix transducer Matrix array technology utilizes 2400 fully-sampled elements for 360-degree focusing and steering xplane

Image optimisation: Framerate Frame rate 74Hz Frame rate 37Hz xplane lateral tilt +30 (+45 ) -30 (-45 )

Image optimisation: Framerate ROI: LV (VR 33HZ) ROI: LA (VR 51HZ)

Color Doppler Flow Imaging Gain Setting Adjusts the degree of amplification of received Doppler signals To optimize the flowsignal the gain setting is just below the level of random background noise Framerate + Velocity range Sector depth Sector width Line density Pulse repetition frequency (PRF)

Color Doppler Flow Imaging: 2D gain Reduce 2D gain! Color flow data is not displayed on the top of structures (including noise due to excessive gain)

Color Doppler Flow Imaging: Color gain setting Radom noise Correct gain setting Too low gain = diminished sensitivity (large jets appear smaller)

Color Doppler Flow Imaging: FR sector size Frame rate 8 Hz Frame rate 21 Hz Increased sector width requires more scan lines resulting in slower frame rate So keep color area to a minimum

Color Doppler Flow Imaging: FR sector size Frame rate 22Hz xplane lateral tilt Frame rate 11Hz +30 (+45 ) -30 (-45 )

Color Doppler Flow Imaging: FR sector depth FR 17 Hz FR 17 Hz FR 29 Hz Same transmit-receiving time No difference in frame rates Less transmit-receiving time Higher frame rate

Color Doppler Flow Imaging: Velocity range VR 91 cm/s VR 61 cm/s VR 30 cm/s Default setting velocity scale 60 cm/s

Color Doppler Flow Imaging: Velocity range(tcpc) Low PRF setting use a low wall filter, therefore low velocity flow is visible

Spectral Doppler modes Continuous Wave (CW) No upper velocity limit but No depth discrimination Pulsed Wave (PW) Upper velocity limited by Aliasing but Does provide spatial location

Pulsed Wave Doppler: Sample volume position

Standard settings pulsed/ continuous wave Doppler Filter setting Baseline shift High pass filters eliminate low frequency Doppler shifts Use Baseline shift to maximise Velocity Scale Velocity range Gain setting Optimize for precise velocity measurements Too high gain setting overrate velocity measurements

Continuous Wave Doppler: Gain setting Vmax 3.2 m/s Max PG 41 mmhg Mean PG 24 mmhg VTI 83 cm Vmax 3.5 m/s Max PG 50 mmhg Mean PG 29 mmhg VTI 94 cm

Pulsed Wave Doppler: Gain setting CWD Aortic valve PWD LVOT correct gain setting higher gain setting PG 82 mmhg VTI 96.1 cm (LVOT 17 mm) VTI 36.5 cm AVA 0.86 cm 2 VTI 44.3 cm AVA 1.04 cm 2

Ultrasound Image Artifacts Definition: False, multiple or misleading information introduced by the imaging system or by interaction of ultrasound with the adjacent tissue Can be falsely interpreted as real pathology May obscure pathology Important to understand and appreciate

Ultrasound Image Artifacts Do not believe everything you see

Ultrasound Image Artifacts Acoustic enhancement Acoustic shadowing Wide beam artifact Side lobe artifact Reverberation artifact Gain artifact Contact artifact

Image Artifacts: Reverberation This causes evenly spaced lines at increasing depths Sound reflects back and forth between the surface of the probe and a strong reflector close to the surface Bron: A.Pally

Image Artifacts: Reverberation TTE TTE

Image Artifacts: Side Lobe The probe cannot produce a pulse that travels purely in one direction

Image Artifacts: Acoustic shadowing Occurs distal to any highly reflective or highly attenuating surface Failure to visualize the source of a shadow is usually caused by the object being outside the plane of the ultrasound beam Bron: A.Pally

Image Artifacts: Refraction X Sound is refracted as it passes from one medium to another. Thus the direction in which it travels changes, when the angle is < 90 this can lead to: subtle miss placement of structures degeneration of image quality Bron: A.Pally

Image Artifacts: Refraction Ghost Image A dramatic example of refraction. A structure is represented twice or more, side by side

Image optimisation: Take Home Message Important - use all modalities of the echocardiographic techniques Settings have to be - optimized for each individual patient - adjusted with respect to the target lesions that have to be analysed Artifacts - change the transducer position - change the angulation of the transducer WRONG SETTINGS! Important findings will not be documented and can also be overlooked

THANK YOU