STEM alignment procedures

Similar documents
1.1. In regular TEM imaging mode, find a region of interest and set it at eucentric height.

2. Raise HT to 200kVby following the procedure explained in 1.6.

This document assumes the user is already familiar with basic operation of the instrument in TEM mode and use of the Microscope Control interface.

Full-screen mode Popup controls. Overview of the microscope user interface, TEM User Interface and TIA on the left and EDS on the right

MSE 460 TEM Lab 2: Basic Alignment and Operation of Microscope

1.3. Before loading the holder into the TEM, make sure the X tilt is set to zero and the goniometer locked in place (this will make loading easier).

Indiana University JEM-3200FS

MSE 595T Transmission Electron Microscopy. Laboratory III TEM Imaging - I

1.2. Make sure the viewing screen is covered (exposure to liquid N 2 may cause it to crack).

NANO 703-Notes. Chapter 9-The Instrument

STANDARD OPERATING PROCEDURE: JEOL TEM-2100

JEOL 6500 User Manual

1. Specimen Holder Removal, Loading, and Insertion

INSTRUCTIONS JEM-2010F FIELD-EMISSION TRANSMISSION ELECTRON MICROSCOPE WITH STEM CAPABILITY

MSE 460 TEM Lab 4: Bright/Dark Field Imaging Operation

This document assumes the user is already familiar with basic operation of the instrument in TEM mode and use of the digital camera.

Chapter 2 Instrumentation for Analytical Electron Microscopy Lecture 7. Chapter 2 CHEM Fall L. Ma

Please follow these instructions for use of the Philips CM100 TEM. Adopted from website below.

FEI Tecnai G 2 F20 Operating Procedures

Transmission Electron Microscopy 9. The Instrument. Outline

JEOL JEM 2010 TRAINING TRANSMISSION ELECTRON MICROSCOPE USER MANUAL

Standard Operating Procedure for the Amray 1810 Scanning Electron Microscope Version: 29 NOVEMBER 2014

Instructions for Tecnai a brief start up manual

Development of JEM-2800 High Throughput Electron Microscope

Operating the Hitachi 7100 Transmission Electron Microscope Electron Microscopy Core, University of Utah

How to use the Jeol 1010 TEM of GI (Liesbeth own GI version)

CM20 USER GUIDE. Duncan Alexander, CIME 2010

JEOL JEM-1400 Transmission Electron Microscope Operating Instructions

A few concepts in TEM and STEM explained

JEOL 6700 User Manual 05/18/2009

JEM-F200. Multi-purpose Electron Microscope. Scientific / Metrology Instruments Multi-purpose Electron Microscope

JEOL 2010 FasTEM & DigitalMicrograph User's Guide

User Operation of JEOL 1200 EX II

05/20/14 1. Philips CM200T. Standby Condition

OPERATION OF THE HITACHI S-450 SCANNING ELECTRON MICROSCOPE. by Doug Bray Department of Biological Sciences University of Lethbridge

Operating Checklist for using the Scanning Electron. Microscope, JEOL JSM 6400.

TEM theory Basic optics, image formation and key elements

LEO 912 TEM Short Manual. Prepared/copyrighted by RH Berg Danforth Plant Science Center

Operating Checklist for using the Scanning Electron Microscope, JEOL JSM 6400.

Introduction to Transmission Electron Microscopy (Physical Sciences)

2 How to operate the microscope/obtain an image

Basic Users Manual for Tecnai-F20 TEM

Introduction of New Products

Check that the pneumatic hose is disconnected!!!! (unless your using the BSE detector, of course)

General information. If you see the instrument turned off, notify MIC personnel. MIC personnel will help you insert your samples into the instrument.

Introduction to Electron Microscopy

The user should already be familiar with operation of the instrument in STEM mode, use of the Microscope Control interface, and TIA.

Tecnai on-line help manual --

Jeol JEM Responsible personell: Endy ( ) Online booking is compulsory!

FEI Falcon Direct Electron Detector. Best Practice Document

Procedures for Performing Cryoelectron Microscopy on the FEI Sphera Microscope

Nanotechnology in Consumer Products

Section 1: TEM parts and functions... 2

Chapter 4 Imaging Lecture 17

JSM 6060 LV SCANNING ELECTRON MICROSCOPE STANDARD OPERATING PROCEDURES

NanoSpective, Inc Progress Drive Suite 137 Orlando, Florida

COMPACT MANUAL FOR GI USERS OF THE JEM 1400 FLASH BEGINNERS (For internal use only) Gray means additional information at the end of this mini-manual

Recent results from the JEOL JEM-3000F FEGTEM in Oxford

FE-SEM SU-8020 Operating manual (Preliminary version)

Using the Hitachi 3400-N VP-SEM

Model SU3500 Scanning Electron Microscope

ELECTRON MICROSCOPY. 13:10 16:00, Oct. 6, 2008 Institute of Physics, Academia Sinica. Tung Hsu

STEM Spectrum Imaging Tutorial

Scanning electron microscope

Basic Operating Instructions for Strata Dual Beam 235 FIB/SEM

SEM OPERATION IN LOW VACUUM MODE

Operating F20/F30 with SerialEM

Title: Amray 1830 SEM#2 Semiconductor & Microsystems Fabrication Laboratory Revision: D Rev Date: 03/18/2016

Scanning Electron Microscope FEI INSPECT F50. Step by step operation manual

2014 HTD-E with options

Dickinson College Department of Geology

Tecnai T12 Operating Procedures

Introduction: Why electrons?

Transmission electron Microscopy

Progress in aberration-corrected scanning transmission electron microscopy

ECEN 4606, UNDERGRADUATE OPTICS LAB

Horiba Jobin-Yvon LabRam Raman Confocal Microscope (GERB 120)

1. Preliminary sample preparation

WITec Alpha 300R Quick Operation Summary October 2018

Tutorial on Linear Image Simulations of Phase-Contrast and Incoherent Imaging by convolutions

Titan on-line help manual -- Working with a FEG

Appreciating the very little things: Status and future prospects of TEM at NUANCE

JEOL 5DII. Operation introduction. By Serge Charlebois

PANalytical X pert Pro Gazing Incidence X-ray Reflectivity User Manual (Version: )

SEM Training Notebook

Characterization Microscope Nikon LV150

ELECTRON MICROSCOPY. 14:10 17:00, Apr. 3, 2007 Department of Physics, National Taiwan University. Tung Hsu

CAPTURING IMAGES ON THE HIGH-MAGNIFICATION MICROSCOPE

SEM Training Notebook

Manual for BMS E1 eplan series, compound microscope

BX-61: Brightfield Instruction /Continue to scroll for Fluorescent Instuctions

ELECTRON MICROSCOPY. 09:10 12:00, Oct. 27, 2006 Institute of Physics, Academia Sinica. Tung Hsu

Scanning Electron Microscope. Instructions for Use

Protective Equipment Nitrile gloves for handling sample holder and safety glasses for filling liquid nitrogen dewar.

R I T. Title: Wyko RST Plus. Semiconductor & Microsystems Fabrication Laboratory Revision: A Rev Date: 05/23/06 1 SCOPE 2 REFERENCE DOCUMENTS

ISO INTERNATIONAL STANDARD

ML7520 ML7530 DIOPTER ADJUSTMENT RING BINOCULAR BODY, INCLINED 30. (a) Field Iris Control Lever. (c) Filter Slots EYEPIECES, KHW10X

Spotlight 150 and 200 FT-IR Microscopy Systems

Nikon AZ100. Laser Scanning Macro Confocal Microscope. Jordan Briscoe Adam Fries Kyle Marchuk Kaitlin Corbin. May 2017.

Transcription:

STEM alignment procedures Step 1. ASID alignment mode 1. Write down STD for TEM, and then open the ASID control window from dialogue. Also, start Simple imager viewer program on the Desktop. 2. Click on ASID from the ASID control window, and then select alignment. If scan starts, click on spot in the Simple Image Viewer. Choose a desired spot size for example spot M. 3. Turn the Brightness knob fully clockwise, and then, adjust OBJ focus in order to form a caustic spot on the screen, as shown below. 4. Center, the caustic spot with bright tilt/def x, y (often called beam tilt), as shown below. 5. Set the OBJ focus to TEM STD focus (turn the OBJ focus knob until you match OBJ lens voltage with TEM STD focus), and then adjust Brightness knob, to form a bright spot, as shown below. Bring the bright spot to the center with Beam shift x, y. Center the caustic with Beam tilt Center the bright spot with Beam shift 6. Repeat 1-5 until both caustic/bright spot are centered on the screen.

Step 2. ASID MAG mode 1. Switch from Alignment to MAG, and wait until mode change is completed. Press STD focus again to make sure that you are at STD focus. 2. Stop scanning by clicking on in the Simple Image Viewer. 3. Find amorphous material on you sample, and then obtain Ronchigram by adjusting OL or sample height. Center the Ronchigram with PLA of DEF/STIG x/y. PLA can be activated from the left control panel or PL in alignment window under Maintenance. CL stig PL alignment deflector Shift x/y Left control panel Right control panel Def/stig x/y 4. Go magnification at least higher that 250k. Higher than 1M is recommended. Press Cond stig on the left control panel, in order to correct condenser astigmatism. Make the Ronchigram as round as possible with Def/Stig x/y. See below. Overfocus Underfocus Astigmatic Corrected

5. Click on Anode to activate the anode wobbler, then select Gun deflector. Adjust gun tilt with def/stig x/y so that the beam contracts and expands concentrically with respect to the center of the Ronchigram. Gun deflector Anode wobbler 6. Mark the center of the Ronchigram (bring the center to the screen center. Move to an area of interest. If your crystal is tilted, you see that the center of the Kikuchi pattern does not match the center of the Ronchigram. Go large over focus. Tilt your sample so that the Kikuchi pattern center matches the center of the Ronchigram, as shown be low. : center of the Ronchigram 7. Now adjust your focus. As you approach the optimum defocus, lattice fringes of your sample will become visible. Remember how the Ronchigram looks like in focus. Insert CL aperture #3 or #4 to the center of the Ronchigram. 8. Make sure that there is not crystal tilt. If there is a tilt, carefully tilt your sample to exact zone axis in order to achieve atomic resolution. 9. Now the microscope is ready to perform high-resolution STEM imaging.

Overfocused Si Ronchigram Focused Si Ronchigram Underfocused Si Ronchigram Slightly Tilted On Axis

How to obtain STEM ADF images. JEOL BF JEOL HAADF 1. For JEOL HAADF detector, first, set camera length to 60~120cm, and then select STEI-DF+ click in. As soon as In is checked, the upper ADF is inserted (you hear the sound). 2. Start scanning by clicking on. Adjust OL focus until the image is focused. Adjust CL stig if necessary. 3. Adjust Brightness and Contrast in ASID control panel as you see STEM image in Simple Image Viewer. 1. For JEOL BF detector, first, set camera length to 100~120cm, and then select STEI-BF+ click In. As soon as In is checked, the upper BF is inserted (you hear the sound). 2. Start scanning by clicking on. Adjust OL focus until the image is focused. Adjust PL alignment coil to maximize brightness of your STEM image (centering your beam to the BF detector). 3. Adjust Brightness and Contrast in ASID control panel as you see STEM image in Simple Image Viewer.

How to obtain STEM BF images. After the STEM alignment described above. For JEOL BF detector, set camera length to 40cm or larger, you can also insert HC aperture #3 or #4 to the center of the Ronchigram, if you would like to make the collection angle smaller. Insert JEOL BF detector manually, and put screen cover on the viewing chamber. Select STEI-BF and Click on in the Simple Image Viewer. control window, and then, the microscope starts scanning. JEOL BF Gatan BF 4. Focus the image by OBJ fine. If necessary, adjust CL stig. For Gatan BF detector, set camera length to 8cm or larger, you can also insert HC aperture #3 or #4 to the center of the Ronchigram, if you would like to make the collection angle smaller. Insert Gatan BF detector by pushing the BF button on the Gatan detector controller, and lift the screen by pressing F1 on the right control panel. Put screen cover on the viewing chamber. Select Ext in the ASID control window, and then, click on. First, focus with OL, and then center the beam with PLA, by maximizing the contrast in the image. If necessary, correct astigmatism.

Step 1. Things to do in TEM Alternative STEM alignment procedures Perform standard TEM alignment with CL aperture #3 or #2. Find a distinctive feature in your sample and bring it to center of your screen. Focus and take an image for reference. Click on ASID from the ASID control panel under Dialogue, and then select MAG. Choose a desired spot size for example spot M. Set camera length to 60~120cm, and then select STEI-UDF+ click in. As soon as In is checked, the upper ADF detector is inserted (you hear the sound). JEOL HAADF 1. Go to Simple Image Viewer. 2. Adjust Brightness and Contrast in ASID control panel as you see STEM image in Simple Image Viewer. 1. Find the feature you found in TEM. If necessary, lower your magnification, and then focus.

TEM image of feature STEM image After centering 1. Press Bright Tilt on the left panel control, and then use shift x and y to bring the feature to the center of scanning image. 2. Select super rapid scan SR in Simple Image Viewer. 3. Turn on HT Wobbler, and make the wobbling concentric with Def/Stig x/y (while Bright Tilt is still pressed). CL stig Shift x/y Left control panel Right control panel Def/stig x/y 1. Click on in Simple Image Viewer to stop scan, and then look at the Ronchigram with CL aperture still inserted. After HT centering After aperture centering

1. Center the CL aperture with respect to the Ronchigram center using Def/stig x,y with Bright Tilt pressed, as shown in the previous page. If necessary, correct astigmatism with Def/stig x,y after pressing Cond Stig. 2. If it is difficult to correct astigmatism with CL aperture inserted, mark the center of your aperture, and then retract CL aperture. Correct astigmatism, making Ronchigram round, and then re-insert CL aperture to exactly where it was before. Overfocus Underfocus Astigmatic Corrected 1. Now you can move to step 5 in original alignment