The newer Fluke 199C recording scope meters are GREAT instruments if used properly.

Similar documents
Test 1: What is the effect on cattle drinking water from an insulated water trough if an electric fence is within inches of the water tank?

Evaluating Electrical Events on the Dairy Farm

Scopes for Stray Voltage. February

Doing It Right. Phasor Labs. March voice fax

Measuring Stray Voltage. Steady state

Mr. Pulsartm. Specialized Function Generator

AE Agricultural Customer Services Play-by-Play Tekscope Manual

I. Introduction to Animal Sensitivity and Response

Oscilloscope Applications MREC Rural Energy Conference 2012 Prepared by Paul Ortmann, P.E. /

2 : AC signals, the signal generator and the Oscilloscope

I. Introduction to Animal Sensitivity and Response

Lab 04. Basic Oscilloscope Operation Capacitor Time Constant. 1. The Oscilloscope.

3.2 Measuring Frequency Response Of Low-Pass Filter :

Notes on OR Data Math Function

ECE65 Introduction to the Function Generator and the Oscilloscope Created by: Eldridge Alcantara (Spring 2007)

Reference Sources. Prelab. Proakis chapter 7.4.1, equations to as attached

The University of Jordan Mechatronics Engineering Department Electronics Lab.( ) Experiment 1: Lab Equipment Familiarization

The Oscilloscope. Vision is the art of seeing things invisible. J. Swift ( ) OBJECTIVE To learn to operate a digital oscilloscope.

ET 304A Laboratory Tutorial-Circuitmaker For Transient and Frequency Analysis

Oscilloscope and Function Generators

Making sense of electrical signals

2 Oscilloscope Familiarization

Sept 13 Pre-lab due Sept 12; Lab memo due Sept 19 at the START of lab time, 1:10pm

Waveform Generators and Oscilloscopes. Lab 6

Fluke199XRAY. Users Manual. Medical ScopeMeter

Getting Started. MSO/DPO Series Oscilloscopes. Basic Concepts

EKA Laboratory Muon Lifetime Experiment Instructions. October 2006

Experiment 5 The Oscilloscope

CHAPTER 6. Motor Driver

Making sense of electrical signals

P a g e 1 ST985. TDR Cable Analyzer Instruction Manual. Analog Arts Inc.

Introduction to Basic Laboratory Instruments

Experiment 1.A. Working with Lab Equipment. ECEN 2270 Electronics Design Laboratory 1

Notes on Experiment #1

Computer Networks. Practice Set I. Dr. Hussein Al-Bahadili

EECS 318 Electronics Lab Laboratory #2 Electronic Test Equipment

Fluke MDA-510 and MDA-550 Motor Drive Analyzer

ECE3204 D2015 Lab 1. See suggested breadboard configuration on following page!

Probe Considerations for Low Voltage Measurements such as Ripple

When you have completed this exercise, you will be able to determine the frequency response of an

Laboratory 3 (drawn from lab text by Alciatore)

Revised: Summer 2010

Reliable Power Meters Multipoint Power Recorder

Combinational logic: Breadboard adders

Complimentary Reference Material

Lecture Fundamentals of Data and signals

Virtual Lab 1: Introduction to Instrumentation

Models 296 and 295 combine sophisticated

Lab 0: Introduction to basic laboratory instruments. Revised by Dan Hoang & Tai-Chang Chen 03/30/2009

ME 365 EXPERIMENT 1 FAMILIARIZATION WITH COMMONLY USED INSTRUMENTATION

PQ Monitoring Standards

Introduction to basic laboratory instruments

Experiment 2 Effects of Filtering

Understanding The HA2500's Dynamic METER Tests

Laboratory Exercise 6 THE OSCILLOSCOPE

SAMPLE: EXPERIMENT 2 Series RLC Circuit / Bode Plot

EC310 Security Exercise 20

Experiment 2: Transients and Oscillations in RLC Circuits

Measurement Procedure & Test Equipment Used

LAB #7: Digital Signal Processing

Analog Arts SG985 SG884 SG834 SG814 Product Specifications [1]

EECS40 RLC Lab guide

LAB 7: THE OSCILLOSCOPE

A Simple Notch Type Harmonic Distortion Analyzer

Introduction to basic laboratory instruments

332:223 Principles of Electrical Engineering I Laboratory Experiment #2 Title: Function Generators and Oscilloscopes Suggested Equipment:

ScopeMeter Test Tool CSV files available for FlukeView software and Microsoft Excel Application Note

Lab 3: AC Low pass filters (version 1.3)

Fluke 192/196/199. ScopeMeter. Users Manual

ECE 2201 PRELAB 6 BJT COMMON EMITTER (CE) AMPLIFIER

University Tunku Abdul Rahman LABORATORY REPORT 1

Agilent 33220A Function Generator Tutorial

Physics 262. Lab #1: Lock-In Amplifier. John Yamrick

Analog Arts SF900 SF650 SF610 Product Specifications

54645D. Mixed Signal Oscilloscope

ENGR 1110: Introduction to Engineering Lab 7 Pulse Width Modulation (PWM)

Data Sheet. Digital Storage Oscilloscope. Features & Benefits. Applications. Ease-of-Use Feature DSO5202BMT DSO5102BMT DSO5062BMT

Switched Mode Power Supply Measurements

On-Line Students Analog Discovery 2: Arbitrary Waveform Generator (AWG). Two channel oscilloscope

EE 201 Function / Arbitrary Waveform Generator and Oscilloscope Tutorial

User Manual Series. Digital Storage Oscilloscope 6810, 6806, March Copyright Protek Test & Measurement 2005 All Rights Reserved

Advanced Lab LAB 6: Signal Acquisition & Spectrum Analysis Using VirtualBench DSA Equipment: Objectives:

Cornerstone Electronics Technology and Robotics Week 21 Electricity & Electronics Section 10.5, Oscilloscope

With any other power quality analyzer you re just wasting energy.

Why Modern Servicing Requires Complete Waveform & Circuit Analyzing!

EE 2274 RC and Op Amp Circuit Completed Prior to Coming to Lab. Prelab Part I: RC Circuit

U1604A Handheld Oscilloscopes, 40 MHz

On-Line Students Analog Discovery 2: Arbitrary Waveform Generator (AWG). Two channel oscilloscope

EENG-201 Experiment # 4: Function Generator, Oscilloscope

Oscilloscope Operation. Visualizing Signals and Making Measurements

Pulsed Measurement Capability of Copper Mountain Technologies VNAs

EECE208 INTRO To ELECTRICAL ENG LAB. LAB 2. Instrumentation

Pinpoint power quality problems faster

Department of Electrical and Computer Engineering. Laboratory Experiment 1. Function Generator and Oscilloscope

LMG500 Getting started. Get in touch

INTRODUCTION TO ENGINEERING AND LABORATORY EXPERIENCE Spring, 2015

Divide. MHz models) waveform record

Fluke 125. Getting Started. Industrial ScopeMeter

2520 Pulsed Laser Diode Test System

Test No. 2. Advanced Scope Measurements. History. University of Applied Sciences Hamburg. Last chance!! EEL2 No 2

Transcription:

CowContacttm Volume 2004 Issue I March 2004 www.phasorlabs.com If you use FLUKE ScopeMeters and FlukeView logging software..see CowContact Volume 2G. If you use the Fluke 199C read this FLUKE ScopeMeters and FlukeView logging software provide a powerful means of recording electrical events. The newer Fluke 199C recording scope meters are GREAT instruments if used properly. I discovered that the FLUKE ScopeMeter and FlukeView logging software can, under certain settings, report electrical events that did not exist at the measurement points. The engineers at FLUKE we very helpful and confirmed that this situation could exist. See CowContact Volume 2G. The Fluke 199C has internal logging features that are great features. I have not found any concerns with the Fluke 199C, but this unit when used for measuring transients or impulses must be used properly. CCVolume 4I 1 3/18/04

Evaluating the Fluke 199C Scope Meter for impulse recording on a Dairy Farm Recorder Button Options: 1. Trend Plot plots a graph of selected Scope measurements over time. 2. Scope Record - records scope waveforms in deep memory 3. Meter Trend Plot plots a graph of selected Meter measurements over time. Using Trend Plot plots a graph of selected Scope measurements over time. Test #1 Channel A setup: (Push button with A ) Input A = ON-OFF set to ON if desired Coupling = DC-AC set to DC for most situations. Probe A = Voltage/Current/Temp set to Voltage for most situations. Attenuation = 1:1 if scope probe is not used or 10:1 if scope 10:1 probe is used. Input A Options = Polarity set to Normal for most situations. Input A Options = Bandwidth set to Full for most situations. Recorder Setup: (Push button with RECORDER ) Recorder = Run-Stop set on Stop. Options = Select Time of Day or From Start Set View All or Normal as desired. Recorder = Run-Stop set on Run to begin recording. Do not Exit Recorder until you are finished. Here is a sample that recorded for about 20 hours. Note that the entire recording period is shown and voltages were recorded, but you cannot tell how long each impulse lasted. CCVolume 4I 2 3/18/04

If you select the ZOOM button you can expand the chart recording as shown below: The best you can expand the recording is to 20 second per division resolution. The impulses recorded would appear to have lasted about 5 seconds, but in reality they only lasted about 10 milliseconds. The scope has a limited amount of internal memory that must be shared over the entire recording period. The longer the period is the less resolution is available to determine the impulse duration. The TREND PLOT mode is acceptable to record the magnitude of voltage impulses, but it is not adequate to determine the duration of the impulse. CCVolume 4I 3 3/18/04

Using Scope Record - records successive scope waveform sweeps in deep memory Test #2 Channel A setup: (Push button with A ) Input A = ON-OFF set to ON if desired Coupling = DC-AC set to DC for most situations. Probe A = Voltage/Current/Temp set to Voltage for most situations. Attenuation = 1:1 if scope probe is not used or 10:1 if scope 10:1 probe is used. Input A Options = Polarity set to Normal for most situations. Input A Options = Bandwidth set to Full for most situations. Range = set volts per division to the expected range you hope to record. This may be a trial and error process for cow contact impulses. Recorder Setup: (Push button with RECORDER ) Recorder = Run-Stop set on Stop. Options = Select Time of Day or From Start Display Glitches = Set Glitch On to view all short duration impulses. Set 10 khz to filter out short duration impulses. Mode = Set Single Sweep if you want to record from the time you push the RUN setting until the memory is full. Mode = Set Continuous if you want to continuously fill memory, keeping the last data recorded when you STOP. Mode = Set Ext is probably not a mode you will use. Set View All or Normal as desired. View All lets you see what is being recorded. Horizontal Sweep Time setup: Push TIME rocker switch to S or ns to select the desired sweep time from a slow sweep of 4 hours per division giving a total record time of 48 hours to a fast sweep of 500 milliseconds per division giving a total record time of 6 seconds. Recorder = Run-Stop set on Run to begin recording. Do not Exit Recorder until you are finished. CCVolume 4I 4 3/18/04

Below I am showing a test with the horizontal time sweep at 500 microseconds per division providing a total record time of 6 seconds. Three impulses have been recorded. Using the ZOOM control one of the impulses has been expanded to show duration of 300 microseconds. But, the impulse sent to the scope was only 2 microseconds long. Even at the fast sweep speed the scope will not be able to tell the duration of the voltage event. The problem is that at a cow contact point and any other point on the farm there are numerous extremely short duration and high magnitude voltage impulses that will trick the scope meter into believing there is a cow contact concern, when the impulses being recorded would never affect a dairy cow. CCVolume 4I 5 3/18/04

Below I am showing a test with the horizontal time sweep at 4 hours per division providing a total record time of 48 hours. Three impulses have been recorded. The recording was stopped after 20 hours and 19 minutes. The large impulse in the center is actually three impulses. Here I have zoomed in on the large impulse and it is really three impulses. The first impulse is created when the Mr. Pulsar test set was first connected to the scope input, the next two test pulses are individual pulses of 2 microsecond duration each. CCVolume 4I 6 3/18/04

Zooming in on a 2 microsecond pulse as far as possible shows the following screen: It would appear to some reading the scope screen that the pulse duration was about 10 seconds but it was only 2 microseconds. So be careful, in this test the impulse appears to be 5,000,000 times greater in duration than it really is! Does this mean the Fluke Scope recorded in error? NO. When the recording mode was set for a horizontal sweep of 4 hours per division the scope had a maximum record time of 48 hours (48*60*60=172,800 seconds). The deep memory in the scope has 27,000 memory points so the recording interval per memory point is (172,800 / 27000 = 6.4 seconds. The oscilloscope can detect impulses as short as 50 nanoseconds so it had no problem detecting and recording a 2 microsecond long impulse. The above scope screen shows that in the 6.4 second time period an impulse did occur, but it cannot tell you exactly how long the impulse lasted. This is very important in stray voltage work. Another problem that can occur is if a 2 microsecond impulse occurs every 7 seconds, the scope chart may falsely show a continuous voltage for an even longer period of time. CCVolume 4I 7 3/18/04

Using Scope Record - records successive scope waveform sweeps in deep memory Test #3 This test has the same scope settings as Test #2. Below is a test where the scope is recording at 4 hours per division for a total record time of 48 hours. For a 60 second period I applied one 2 microsecond pulse every five seconds. The first screen seems to indicate an impulse of 60 seconds duration: Next I zoomed in to maximum resolution and the scope still indicates an impulse of 60 second duration: The test the impulse is reported to be 30,000,000 times greater in duration than it really is! CCVolume 4I 8 3/18/04

Why is the false reporting of short duration impulses so important? On the farm there are numerous sources of extremely short duration impulses. These impulses come from on-farm equipment that is turning on and turning off. I have found that equipment turning off creates greater impulses than equipment turning on. When a light is switched or a vacuum pump motor stops, there is a high magnitude impulse of very short duration that is created on the farm electrical system. This impulse can be observed at cow contact points and usually has a magnitude from 1 volt peak to as high as 70 volts peak. Of course the voltage does not last long enough for a cow to feel the voltage, but modern high-speed instruments such as the Fluke 199C can detect and record the events. It is extremely important when measuring voltages at cow contact points to measure both the magnitude of the voltage and the duration of the voltage. Fortunately the Fluke 199C and other scopes of its class provide the ability to do a perfect job of recording data if the unit is used properly. Below is a screen of a Fluke 199C properly recording a voltage impulse at a cow contact point. The vertical voltage scale is 1 volt per division and the horizontal sweep speed is 5 microseconds per division. The impulse has a peak voltage of approximately 0.6 volts and duration of approximately 4 microseconds. As you can see even the 5% most sensitive cows would not detect the impulse. CCVolume 4I 9 3/18/04

To determine if the cow felt the impulse I refer to a cow sensitivity chart that I have prepared using data and assistance from the University of Wisconsin. This chart is the summation of many different experiments and field measurements by many different researchers. 10,000 V Behavioral response for 5% of the most sensitive cows using sine waves from muzzle to hooves exposure V o l t s - Z e r o 1,000 V 100 V Typ. Fencer 1 Cycle, Biphasic 1 Cycle, Monophasic Multiple Cycle t o 5 10 V p e a k - 5 0 0 2 1 V (1,000 mv) 0.1 V Higher Frequency Lower Frequency USDA LOC PSCW LOC o h m s 0.01 V (10 mv) 0.001 V (1 mv) 1000 on your AM radio dial 300 Hz (5th Harmonic) 1 millisecond 60 Hz 0.1 1 10 100 1,000 10,000 Phase Duration (microseconds) = Time between voltage zero crossings Reprinted with permission C. Forster 7/5/00, 6/2003 Used properly, the ScopeMeter and FlukeView logging software can be a powerful survey tool for stray voltage. It is important that the system be used properly. Please see the liability disclaimer at the end of Volume 2001 Issue B. Chuck Forster, P.E. cforster@phasorlabs.com Fluke products can be viewed at www.fluke.com Mr. Pulsar can be viewed at www.phasorlabs.com CCVolume 4I 10 3/18/04