Disclosures. Outline 7/31/2017. Current Implementation Status of IEC Standard : Exposure Index (EI) for Digital Radiography

Similar documents
Exposure Indices and Target Values in Radiography: What Are They and How Can You Use Them?

Practical Medical Physics Session: TG-151 Dose Monitoring. August 5, 2013 Katie Hulme, M.S.

Digital radiography (DR) post processing techniques for pediatric radiology

Estimation of signal transfer property for wireless digital detector in different measurement schemes

Y11-DR Digital Radiography (DR) Image Quality

Ask EuroSafe Imaging Tips & Tricks. Paediatric Imaging Working Group. Dose Management in Digital Radiography

COMPUTED RADIOGRAPHY CHAPTER 4 EFFECTIVE USE OF CR

I. PERFORMANCE OF X-RAY PRODUCTION COMPONENTS FLUOROSCOPIC ACCEPTANCE TESTING: TEST PROCEDURES & PERFORMANCE CRITERIA

3/31/2011. Objectives. Emory University. Historical Development. Historical Development. Historical Development

Features and Weaknesses of Phantoms for CR/DR System Testing

An Exposure Indicator for Digital Radiography. Report of AAPM Task Group 116

Acquisition, Processing and Display

Learning Objectives: What s my motivation? (unknown screen actor) Workshop Overview

of sufficient quality and quantity

2217 US Highway 70 East Garner, NC Main: Fax:

SPECIFICATION. Kilovoltage X-ray calibration system for protection and diagnostic level dosimetry. Prepared by

Key words: fluoroscopy, dose-area-product, kerma-area-product, calibration of KAP meters, patient exposure

New Exposure Indicators for Digital Radiography Simplified for Radiologists and Technologists

Quality assurance: a comparison study of radiographic exposure for neonatal chest radiographs at 4 academic hospitals

Outline. Digital Radiography. Understanding Digital Modalities: Image Quality and Dose. Image Quality. Dose Control

- KiloVoltage. Technique 101: Getting Back to Basics

Breast Tomosynthesis. Bob Liu, Ph.D. Department of Radiology Massachusetts General Hospital And Harvard Medical School

NEMA XR X-ray Equipment for Interventional Procedures User Quality Control Mode

A study of exposure index value fluctuations in computed radiography and direct digital radiography using multiple manufacturers

Teaching Digital Radiography and Fluoroscopic Radiation Protection

Digital Imaging Considerations Computed Radiography

DISC QC/QA Program for Digital Imaging Systems using the DR Radchex Plus Meter

Mammography: Physics of Imaging

Joint ICTP/IAEA Advanced School on Dosimetry in Diagnostic Radiology and its Clinical Implementation May 2009

Acceptance Testing of a Digital Breast Tomosynthesis Unit

Half value layer and AEC receptor dose compliance survey in Estonia

The effect of compensating filter on image quality in lateral projection of thoraco lumbar radiography

Evaluation of cassette-based digital radiography detectors using standardized image quality metrics: AAPM TG-150 Draft Image Detector Tests

Update on Fluoroscopy Physics AAPM MO-A-210A-1 Stephen Balter, Ph.D.

Digital Imaging started in the 1972 with Digital subtraction angiography Clinical digital imaging was employed from the 1980 ~ 37 years ago Amount of

History of digital imaging

1. Patient size AEC. Large Patient High ma. Small Patient Low ma

Unfors EDD-30 Radiation Protection in Fluoroscopy

CR Basics and FAQ. Overview. Historical Perspective

STEREOTACTIC BREAST BIOPSY EQUIPMENT SURVEYS

Automated dose control in multi-slice CT. Nicholas Keat Formerly ImPACT, St George's Hospital, London

GE AMX 4+ Portable X-Ray

Calibration of KAP meters

A Practical Overview of the Clinical and Operational Impact of Computed Radiography(CR) Implementations. Shirley Weddle, RT(R)(M), CIIP, BBA

Learning Objectives. Outline. Getting Started with CR. Converting the Radiology Department from Film-Screen to Digital: Making the Transition

DRX Plus Detectors: Going from Good to Great

Tailoring automatic exposure control toward constant detectability in digital mammography

Quality Control of Full Field Digital Mammography Units

Aspire HD. Program Manual. 2nd Edition - October 2012

Appropriate Inspection Distance of Digital X-Ray Imaging Equipment for Diagnosis

Fully Automatic X-ray meter kvp, dose, rate, HVL and time Extremely easy to use. Unfors ThinX RAD for Radiographic Quality Control

Diagnostic Imaging Specialists Corporation 163 St.Malo Street St.Malo, MB Canada R0A1T0

RAD 150 RADIOLOGIC EXPOSURE TECHNIQUE II

SYLLABUS. TITLE: Equipment Operation I. DEPARTMENT: Radiologic Technology

Overview of Safety Code 35

2012 :15th SESSION of ESMP

Introduction. Digital Mammography QA: Comparing the Manufacturers Recommendations. What is QC and why is it important? Review & compare QC tests

Practical Aspects of Medical Physics Surveys of Mammography Equipment and Facilities

Dosepix Detector as kvp-meter in Radiology and Mammography: First steps

Nuclear Associates EZ CR-DIN Phantoms

Minnesota Rules, Chapter 4732 X-ray Revision

DIAGNOSTIC ACCREDITATION PROGRAM. Radiology and CT Quality Control Procedures Workbook

DRAFT Technical evaluation of Philips Microdose SI digital mammography system

Clinical Experiences with a Patient Skin Dose Monitoring and Tracking Program

Nathan Childress, Ph.D., DABR

Test Equipment for Radiology and CT Quality Control Contents

An Introduction to TG-142 Imaging QA Using Standard Imaging Products. Mark Wiesmeyer, PhD, DABR Technical Product Manager Standard Imaging, Inc.

Collimation Assessment Using GAFCHROMIC XR-M2

Surveying and QC of Stereotactic Breast Biopsy Units for ACR Accreditation

Digital Detector Array Image Quality for Various GOS Scintillators

Enhanced Functionality of High-Speed Image Processing Engine SUREengine PRO. Sharpness (spatial resolution) Graininess (noise intensity)

Outline ASRT Changes Impact on current curriculum Potential new courses WECM Changes Last update Resources and needs

The Evaluation of Collimator Alignment of Diagnostic X-ray Tube Using Computed Radiography System

DOSELAB TOMOTHERAPY TG-148 QA QUICK GUIDE TG-148 RECOMMENDED TESTS 1. V.B.1.C. - Y-JAW DIVERGENCE/BEAM CENTERING

New spectral benefi ts, proven low dose

ELECTRONIC CONTROL CONCEPTS 160 Partition Street Saugerties, NY or local phone

Investigation of the line-pair pattern method for evaluating mammographic focal spot performance

INTERNATIONAL STANDARD

Overview. Professor Roentgen was a Physicist!!! The Physics of Radiation Oncology X-ray Imaging

X-RAY IMAGING EE 472 F2017. Prof. Yasser Mostafa Kadah

X-RAYS - NO UNAUTHORISED ENTRY

Beam-Restricting Devices

Protocol for the Quality Control of the Physical and Technical Aspects of Digital Breast Tomosynthesis Systems

Exposure System Selection

AUTOMATED AND QUANTITATIVE METHOD FOR QUALITY ASSURANCE OF DIGITAL RADIOGRAPHY IMAGING SYSTEMS

A Comprehensive Review of Image Production

DigiMam Conformance Statement for DICOM V3.0

ESTABLISHING A QUALITY ASSURANCE ROUTINE FOR DIGITAL IMAGING

CT radiation profile width measurement using CR imaging plate raw data

Quality Control for Stereotactic Breast Biopsy. Robert J. Pizzutiello, Jr., F.A.C.M.P. Upstate Medical Physics, Inc

Research Support. Dual-Source CT: What is it and How Do I Test it? Cynthia H. McCollough, Ph.D.

KODAK DIRECTVIEW CR Mammography Feature User s Guide

Nuclear Associates

TESTING FLAT-PANEL IMAGING SYSTEMS: What the Medical Physicist Needs to Know. JAMES A. TOMLINSON, M.S., D.A.B.R. Diagnostic Radiological Physicist

Comparison of computed radiography and filmõscreen combination using a contrast-detail phantom

Assessment of Beam Alignment, Collimation and Half Value Layer of Some Selected X-Ray Machines in Plateau State, Nigeria

A comparative study of several digital flat panel X-ray units: patients doses and image quality in chest radiography

Protocol for the Quality Control of the Physical and Technical Aspects of Digital Breast Tomosynthesis Systems

Determination of the tube voltage from clinic mammographic system using two types of detectors

Mammography is a radiographic procedure specially designed for detecting breast pathology Approximately 1 woman in 8 will develop breast cancer over

Transcription:

Current Implementation Status of IEC Standard 62494-1: Exposure Index (EI) for Digital Radiography July 31, 2017 Ryan Fisher, PhD, DABR Katie Hulme, MS, DABR None Disclosures Outline Review of IEC Standard 62494-1 Vendor-specifics of verifying compliance with IEC Standard Vendor Implementation of IEC Standard Determination of: Value of Interest Implications for calculated Exposure Index Deviation Index and Target Exposure Index Values 1

TG-116 - An Exposure Indicator for Digital Radiography Appendix C (p42) One true index to unite them all A unified EXPOSURE INDEX for all digital radiography systems is needed to simplify its usage, e.g. for the establishment of exposure guidelines, particularly when systems of different manufacturers are used within the same department. This standard defines such a concept of the EXPOSURE INDEX. IEC 62494-1 (page 5) Exposure Index The exposure index (EI) shall be related to the value of interest (V) according to the formula: EI = c 0 g(v) c 0 = 100 μgy -1 g(v) is the equipment-specific inverse calibration function 2

IEC 62494-1 ORIGINAL DATA The following corrections shall be made to the RAW DATA as in normal clinical use: Bad or defective pixels Flat-field corrections Corrections for geometric distortion IEC 62494-1 Determine the ORIGINAL DATA Image Segmentation Histogram Based Other IEC 62494-1 Determine the Image Segmentation Histogram Based Other The determination of the RELEVANT IMAGE REGION should be done by methods that identify the attenuated regions of the beam that are relevant to the diagnostic purpose of the acquired image. NOTE 2 While it is understood that the selection of the RELEVANT IMAGE REGION is an important step in the generation of the EXPOSURE INDEX and that a single unified method may be desirable, it is not feasible at this time. Future version of the standard may address this issue. IEC 62494-1 (page 9) 3

IEC 62494-1 Determine the V Mean Median Mode Trimmed mean Trimean Other Measure of Central Tendency IEC 62494-1 Determine the V Exposure Calculate the Index EI = c 0 g(v) Exposure Index Calibration The EI is to be calibrated such that under calibration conditions: EI = c 0 K CAL c 0 = 100 μgy -1 K CAL is the image receptor air kerma in μgy under the calibration conditions 4

Calibration Conditions Fixed beam quality (~RQA5): HVL of 6.8 ± 0.3 mm Al Added filtrations of either: 21 mm Aluminum 0.5 mm Copper and 2mm Aluminum X-ray tube voltage 66kVp 74 kvp Adjust tube voltage to obtain target HVL Calibration Conditions Homogenous irradiation of the effective image receptor area Measurement of image receptor air kerma should be made freein-air (no backscatter) Calibration Conditions Homogenous irradiation of the effective image receptor area using FIXED BEAM QUALITY 5

Calibration Conditions Determine the Central 10% of image area Calibration Conditions Determine the V CAL Mean Median Mode Trimmed mean Trimean Other Measure of Central Tendency Calibration Conditions Determine the V CAL Calculate K CAL K CAL = g(v CAL ) EI = c 0 K CAL 6

Inverse Calibration Function g(v CAL ) is the inverse calibration function: K CAL = f 1 V CAL = g(v CAL ) The specified inverse calibration function shall have an uncertainty of less than 20% under calibration conditions Beam Quality Dependence Calibration Condition 21 mm Al CsI data acquired on GE Flashpad DR Detector GoS data acquired on Fuji FDR D Evo II DR Detector Deviation Index If target exposure index (EI T ) values are provided by the system, the deviation index (DI) shall be automatically calculated according to: DI = 10 log 10 EI EI T NOTE 1 For this purpose, the TARGET EXPOSURE INDEX values for different examinations/applications need to be available on the digital x-ray imaging system, e.g. in a data base. Such values may be established by professional societies or by the responsible organization. IEC 62494-1 (page 12) 7

Summary IEC 62494-1 The IEC Exposure index (EI) is linear with incident detector air kerma The standard explicitly defines the conditions under which the EI shall be calibrated Relationship between EI and incident detector air kerma will vary with beam quality Summary IEC 62494-1 The standard does not define the method by which the vendor: Determines the Calculates the Value of Interest Summary IEC 62494-1 The standard does not attempt to establish Target Exposure Index (EI T ) values Leaves this to professional societies and responsible organizations States that they should be available on the system for the purposes of calculating the Deviation Index (DI) Acknowledges that EI T values may depend on: Type of detector Type of examination Diagnostic question 8

Implementation Status Who has implemented IEC 62494-1? What do you need to know for verifying compliance with IEC 62494-1? Differences in vendor approaches to implementation? Who has implemented IEC 62494-1? Vendor Legacy Exposure Indicator IEC Compliant Exposure Index Agfa lgm (No longer in use) EI Canon REX EI Carestream EI* EI Fuji S EI GE DEI (No longer in use) EI Konica S EI Philips EI* EI_s Siemens EXI Clinical EXI* *Not to be confused with the IEC exposure index (EI) **The Clinical EXI is normalized to calibration conditions (Physical EXI is NOT, unless requested by site), note, however, that the Clinical EXI is determined from histogram analysis as opposed to a fixed ROI Vendor Implementation The IEC standard explicitly states that the following items shall be documented: The filter and x-ray tube voltage used for calibration The method used for the selection of the relevant image region The method used to calculate the value of interest The inverse calibration function and the range of image receptor air kerma for which the inverse calibration function can be used to calculate the image receptor air kerma from the value of interest under calibration conditions 9

Vendor Questionnaire What is the inverse calibration function and range of image receptor air kerma for which the inverse calibration function can be used to calculate the image receptor air kerma from the value of interest under calibration conditions? What are the vendor s calibration conditions for verifying accuracy of the IEC exposure index (EI)? Recommended beam quality (filter, HVL, kvp) and SID? What exam tag(s), image processing parameters, and/or tools should be used to ensure a fixed region of interest at the center of the image is used when verifying accuracy of the EI? Size of VOI used to determine EI under these conditions? What is the vendor s tolerance for accuracy of the EI? If EI is out of tolerance, what is the methodology for calibrating the EI to bring back it in tolerance (if it exists)? For clinical images, how is the relevant image region and value of interest (VOI) determined? Are there specific assumptions made that the user should be aware of that can affect the EI that is displayed? Does the vendor recommend Target Exposure Index values for various exams / anatomical regions? If so, do these come pre-programmed and can they be modified? Vendor Implementation Filter and beam quality used for calibration verification Vendor Calibration Filter Beam Quality Agfa 21 mm Al or 0.5 mm Cu+ 2 mm Al (recommended but not provided) RQA5 (HVL of 6.8± 0.3 mm Al) Canon 21 mm Al (recommended but not provided) ~RQA5 (HVL of 7.1± 0.1 mm Al Carestream 21 mm Al or 0.5 mm Cu + 2 mm Al (recommended but 0.5 mm Cu & 1 mm Al provided) RQA5 (HVL of 6.8± 0.3 mm Al) Fuji 21 mm Al or 0.5 mm Cu + 2 mm Al (recommended but not provided) RQA5 (HVL of 6.8± 0.3 mm Al) GE 20 mm Al (filter provided by manufacturer) RQA5 (HVL of 6.8± 0.3 mm Al) Konica 21 mm Al or 0.5 mm Cu+ 2 mm Al (recommended but not provided) RQA5 (HVL of 6.8± 0.3 mm Al) Philips 21 mm Al (filter provided by manufacturer) RQA5 (HVL of 6.8± 0.3 mm Al) Siemens 0.6 mm Cu (insert internal filtration) ~RQA5 (HVL of 6.8± 0.3 mm Al) Vendor Implementation Method used for the selection of the relevant image region* *For Verification of EI Under Calibration Conditions Vendor Exam Tag How to Define a FIXED ROI Size of ROI Agfa System Diagnosis with Flat Field Use manual ROI tool User defined Canon Service Tag AEC (used for AEC calibration) n/a (determine by exam tag) 10cm x 10cm (center of detector) Carestream Pattern Use manual ROI tool User defined Fuji Sensitivity n/a (determine by exam tag) 10cm x 10cm (center of detector) GE n/a Enable Technical Mode 10cm x 10cm (center of detector) Konica S-Value n/a (determine by exam tag) 10cm x 10cm (center of detector) Philips n/a Use Simple Ranger Tool User defined Siemens n/a Physical EXI* According to DIN 6868 Part 58 (center of detector) *The Clinical EXI is normalized to calibration conditions (Physical EXI is NOT, unless requested by site), note, however, that the Clinical EXI is determined from histogram analysis as opposed to a fixed ROI 10

Vendor Agfa Canon Vendor Implementation What if my EI is out of tolerance? Carestream Fuji GE Konica Philips Siemens Tolerance on EI Accuracy ± 20% (per IEC) Method to Re-Calibrate EI? No No Yes Yes Yes (fixed rooms) No (mobile units) Yes (formerly No) ACCEPTANCE TESTING IS IMPORTANT No Yes* *Separate calibration factors for Clinical EXI and Physical EXI, vendor procedures only address calibration of Clinical EXI Vendor Implementation V? Relevant Image Region Value of Interest EI = c 0 g(v) 11

Value of Interest Calculation 12

Agfa Canon Carestream Fuji GE Konica Philips Siemens Vendor Implementation Method used to calculate the value of interest Vendor Determination (Clinical) Value of Interest Image Pixel Histogram Clinical features Geometric location & exposure level Noise, signal & tissue variation measurements Calculation Method Median Exposure field identification Histogram analysis Mean Exposure field identification Histogram analysis Clinical Features Trimmed Mean Exposure field identification Histogram analysis Mean Exposure field identification Histogram analysis Median Uses the same ROI as used in automatic gradation processing Split pattern recognition Exposure field identification Histogram analysis 2-point average of Reference Values (L and H) Exposure field identification Histogram analysis Trimean Exposure field identification Histogram analysis Mean Multi-Vendor Head Phantom Study Imaged head phantom on various DR panels 8 x10 10 x12 14 x17 Consistent positioning and collimation Same detector dose in all images Recorded Exposure Index values scaled and corrected for input dose & panel sensitivity Theoretically, differences in EI should be mostly due to differences in vendor approach EI (average) = 576 STD = 117 COV = 20% 70 13

70 Target EI Factory preset EI T? Vendor Vendor EI T values Implementation of DI Agfa Recommended Values Configured by Apps (editable) Yes Canon Not currently provided (editable) Yes Carestream Preprogramed (editable) Yes Fuji Preprogramed (editable) Yes GE Preprogramed (editable) Yes Konica Preprogramed (editable) Yes Philips No No Siemens Preprogramed (editable)* Yes* Conclusions IEC Standard 62494-1 is widely implemented by vendors Sometimes in tandem with legacy dose indicators Vendor documentation can be hard to come by Differences in vendor methods of implementation can lead to appreciable variation in EI values Range of more than 2.5x with head phantom for same input AK Differences in vendors likely to be exam specific as well Shows need for vendor specific Target Expose Indexes at this time Panel EI sensitivity may need to be accounted for Some vendors don t allow recalibration Variation in implementation can make quality control efforts difficult 14

15