Introduction. Sam R. Kottamasu Lawrence R. Kuhns

Similar documents
Comparison of computed radiography and filmõscreen combination using a contrast-detail phantom

Digital Radiography of the Musculoskeletal System: The Optimal Image

Investigation of the line-pair pattern method for evaluating mammographic focal spot performance

Digital Imaging Considerations Computed Radiography

The Evaluation of Collimator Alignment of Diagnostic X-ray Tube Using Computed Radiography System

Beam-Restricting Devices

Introduction. Chapter 16 Diagnostic Radiology. Primary radiological image. Primary radiological image

X-ray Imaging. PHYS Lecture. Carlos Vinhais. Departamento de Física Instituto Superior de Engenharia do Porto

3/31/2011. Objectives. Emory University. Historical Development. Historical Development. Historical Development

COMPUTED RADIOGRAPHY CHAPTER 4 EFFECTIVE USE OF CR

Dose Reduction and Image Preservation After the Introduction of a 0.1 mm Cu Filter into the LODOX Statscan unit above 110 kvp

Digital radiography (DR) post processing techniques for pediatric radiology

Teaching Digital Radiography and Fluoroscopic Radiation Protection

RADIOGRAPHIC EXPOSURE

RAD 150 RADIOLOGIC EXPOSURE TECHNIQUE II

X-RAYS - NO UNAUTHORISED ENTRY

Optimization of Digital Mammography Resolution Using Magnification Technique in Computed Radiography 1

Acquisition, Processing and Display

Exposure Indices and Target Values in Radiography: What Are They and How Can You Use Them?

Multiple Choice Identify the letter of the choice that best completes the statement or answers the question.

CR Basics and FAQ. Overview. Historical Perspective

Artefacts found in computed radiography

radiography detector

X-RAY IMAGING EE 472 F2017. Prof. Yasser Mostafa Kadah

Evaluation of a quality control phantom for digital chest radiography

Studies on reduction of exposure dose using digital scattered X-ray removal processing

Studies on reduction of exposure dose using digital scattered X-ray removal processing

Mammography: Physics of Imaging

Published text: Institute of Cancer Research Repository Please direct all s to:

Quality assurance: a comparison study of radiographic exposure for neonatal chest radiographs at 4 academic hospitals

A Practical Overview of the Clinical and Operational Impact of Computed Radiography(CR) Implementations. Shirley Weddle, RT(R)(M), CIIP, BBA

LECTURE 1 The Radiographic Image

DIGITAL IMAGE PROCESSING IN X-RAY IMAGING

I. PERFORMANCE OF X-RAY PRODUCTION COMPONENTS FLUOROSCOPIC ACCEPTANCE TESTING: TEST PROCEDURES & PERFORMANCE CRITERIA

Visualization of sources of scattered radiation from x-ray equipment used for interventional radiology

The effect of compensating filter on image quality in lateral projection of thoraco lumbar radiography

Ludlum Medical Physics

Breast Tomosynthesis. Bob Liu, Ph.D. Department of Radiology Massachusetts General Hospital And Harvard Medical School

Ask EuroSafe Imaging Tips & Tricks. Paediatric Imaging Working Group. Dose Management in Digital Radiography

New Exposure Indicators for Digital Radiography Simplified for Radiologists and Technologists

Attikon, Rimini 1, , Athens, Greece , Athens, Greece , Athens, Greece

Mammography is a radiographic procedure specially designed for detecting breast pathology Approximately 1 woman in 8 will develop breast cancer over

While digital techniques have the potential to reduce patient doses, they also have the potential to significantly increase them.

Nuclear Associates , , , , , ,

Acceptance Testing of a Digital Breast Tomosynthesis Unit

10/3/2012. Study Harder

Features and Weaknesses of Phantoms for CR/DR System Testing

Digital Imaging started in the 1972 with Digital subtraction angiography Clinical digital imaging was employed from the 1980 ~ 37 years ago Amount of

10/26/2015. Study Harder

X-Ray-Based Medical Imaging and Resolution

Film Replacement in Radiographic Weld Inspection The New ISO Standard

Nuclear Associates

Veterinary Science Preparatory Training for the Veterinary Assistant. Floron C. Faries, Jr., DVM, MS

Appropriate Inspection Distance of Digital X-Ray Imaging Equipment for Diagnosis

Iranian Journal of Medical Physics

- KiloVoltage. Technique 101: Getting Back to Basics

Overview. Professor Roentgen was a Physicist!!! The Physics of Radiation Oncology X-ray Imaging

Digital radiography: Practical advantages of Digital Radiography. Practical Advantages in image quality

SPRINGFIELD TECHNICAL COMMUNITY COLLEGE ACADEMIC AFFAIRS

Image Quality Artifacts in Digital Imaging

Moving from film to digital: A study of digital x-ray benefits, challenges and best practices

HIGH RESOLUTION COMPUTERIZED TOMOGRAPHY SYSTEM USING AN IMAGING PLATE

Effect of patient support pads on image quality and dose in fluoroscopy a)

CONTRASTING VIEWS DIGITAL VS CONVENTIONAL RADIOGRAPHY

Amorphous Selenium Direct Radiography for Industrial Imaging

Effect of Backscattered Radiation on X-Ray Image Contrast

Visibility of Detail

Radiology. Radiograph: Is the image of an object made with use of X- ray instead of light.

Half value layer and AEC receptor dose compliance survey in Estonia

Digital Image Management: the Basics

STEREOTACTIC BREAST BIOPSY EQUIPMENT SURVEYS

Effects of Plastic Wedges on Whole Foot Radiograph in Anteroposterior and Oblique Positions

Quality Control for Stereotactic Breast Biopsy. Robert J. Pizzutiello, Jr., F.A.C.M.P. Upstate Medical Physics, Inc

Nuclear Associates

Do you have any other questions? Please call us at (Toll Free) or , or

Radiographic Techniques, Contrast, and Noise in X-Ray Imaging

Your worldwide source of grids. What a difference a grid makes

Radiology Physics Lectures: Digital Radiography. Digital Radiography. D. J. Hall, Ph.D. x20893

Learning Objectives: What s my motivation? (unknown screen actor) Workshop Overview

7/24/2014. Image Quality for the Radiation Oncology Physicist: Review of the Fundamentals and Implementation. Disclosures. Outline

Exposure System Selection

DIGITAL RADIOGRAPHY ARTIFACTS

STUDENT REVIEW QUESTION SET K CR/DR CONTENT AREA

2017 West Coast Educators Conference Orlando. Projection Geometry. 1. Review hierarchy of image qualities (amplified version):

Essentials of Digital Imaging

Nuclear Associates

Nuclear Associates

History of digital imaging

2217 US Highway 70 East Garner, NC Main: Fax:

of sufficient quality and quantity

Maximizing clinical outcomes

The importance of radiation quality for optimisation in radiology

A study of exposure index value fluctuations in computed radiography and direct digital radiography using multiple manufacturers

NDE SOLUTIONS RADIOGRAPHY COURSE OUTLINE

Joint ICTP/IAEA Advanced School on Dosimetry in Diagnostic Radiology and its Clinical Implementation May 2009

Quality Control of Full Field Digital Mammography Units

Fabrício Sampaio Péres Kury Federal University of Rio de Janeiro Medical School

1. Carlton, Richard R., and Arlene M. Adler. Principles of Radiographic Imaging: An Art and a Science, 5th edition (2013).

Some operation methods show in the catalog reguire optional eguipment

Luminos RF Classic. Where value meets performance.

Transcription:

Pediatr Radiol (1997) 27: 119 123 Springer-Verlag 1997 Sam R. Kottamasu Lawrence R. Kuhns Musculoskeletal computed radiography in children: scatter reduction and improvement in bony trabecular sharpness using air gap placement of the imaging plate Received: 29 December 1995 Accepted: 27 March 1996 S.R. Kottamasu () ) Department of Pediatric Imaging, Children s Hospital of Michigan, Wayne State University, 3901 Beaubien Blvd., Detroit MI 48201-2196, USA L. R. Kuhns Department of Radiology, Mott Children s Hospital, University of Michigan, Ann Arbor, Michigan, USA Abstract The effect of various air gaps on computed radiographic musculoskeletal images was investigated using a knee phantom. Scatter to primary radiation ratios were measured using the beam stop method at air gaps ranging from 0 to 30 in. (0 762-mm). Bony trabecular sharpness, line pair resolution, quantum mottle and visualization of low-contrast beads in the soft tissues were evaluated. A significant reduction of scatter to primary radiation ratio, from a value of almost 1 at table top to about 0.4 at 10-in. (254- mm) air gap and about 0.2 at 25-in. (635-mm) air gap placement of the computed radiography (CR) imaging plate, was obtained. A progressive improvement in bony trabecular sharpness and line pair resolution, compared with the table top and Bucky images was observed on 10-in. (254-mm) through 25-in. (635-mm) air gap images. Sharpness of the bony trabeculae and line pair resolution were best on the 25-in. (635-mm) air gap images. The skin entrance radiation dose does not have to be increased for air gap digital radiography. The radiographic noise or quantum mottle is highest on the Bucky image, higher on air gap images and minimal on the table top images, despite a high scatter to primary radiation ratio at the table top. The lower quantum mottle on the table top images allowed for maximal visualization of low contrast densities in the soft tissues. Air gap radiography further improves musculoskeletal computed imaging by reducing the scatter to primary radiation ratio without an increase in the skin entrance dose. For significant reduction of the scatter to primary radiation ratio and best evaluation of line pair spatial resolution and bony trabeculae, a 25-in. (635-mm) air gap with digital radiography would be optimal. For evaluation of low contrast densities in the soft tissues, table top placement would be the technique of choice. Introduction We have used digital radiography for musculoskeletal imaging in the Pediatric Orthopedic Clinic at Children s Hospital of Michigan for 4 years and there has been good acceptance of the images by clinicians. During this time we have attempted to optimize the processing parameters of the Fuji Digital Radiography system (Fuji Medical Systems USA, Stamford, Conn.) for each age group and body part, using table top and Bucky (moving grid) techniques. We decided that the next step in improving the image quality would be to investigate the effect of various air gaps on scatter reduction, sharpness of bony trabeculae and line pair resolution, using a knee phantom to simulate an average skeletal part in children.

120 Fig. 1 Scatter to primary radiation ratio (S/P ratio) at various air gaps below the knee phantom are displayed. S/P ratio decreased from a value of almost 1 at table top to 0.2 at 25 in. (635-mm) Table 1 Line pair resolution per mm at various positions of the phosphor plate Receptor placement Mean Standard deviation Table top 2.59 0.27 Bucky 2.79 0.44 Air gap, 5 in. (127 mm) 2.50 0.32 Air gap, 10 in. (254 mm) 2.78 0.17 Air gap, 15 in. (381 mm) 3.09 0.39 Air gap, 20 in. (508 mm) 3.20 0.33 Air gap, 25 in. (635 mm) 3.30 0.20 Air gap, 30 in. (762 mm) 3.20 0.33 Materials and methods A 0.3-mm focal spot radiographic tube was placed 40 in. (1016 mm) above the table top. A standard knee phantom with a line pair phantom (Nuclear Associates, New York) taped to the under surface of the knee and acrylic beads of varying sizes from 1 to 5 mm in diameter taped over the soft tissue portion of the phantom was used. Fuji Standard III Imaging receptors were used and all images were processed on an AC-1 plus image processor using the same processing parameters (density, contrast, frequency filtering and edge enhancement). The images were obtained using Bucky placement, table top placement of the receptor, and placement of the receptor at 5, 10, 15, 20, 25, and 30 in. (127, 254, 381, 508, 635, and 762 mm) below the table top after moving the table top aside. There was no modification of radiographic equipment. A table top technique of 65 kvp and 3.2 mas was used for all images, including those obtained on the Bucky apparatus. The scatter to primary radiation ratio at various air gaps below the knee phantom was measured. These measurements were obtained using film densitometry and the beam stop method [1] by M. Goodsitt (Department of Radiology, University of Michigan). Film density measurements were translated into radiation exposure values using the characteristic curve (also called the H and D curve, after Hunter and Driffield, who first used such a curve) [2]. It was assumed that exposure levels in shadows of the beam stops represented only scattered radiation, while the levels immediately surrounding the beam stop represented scattered radiation plus unscattered primary radiation [1]. Two board-certified radiologists and a board-certified orthopedic surgeon independently evaluated all images without knowing the experimental conditions of the study. Each physician assessed sharpness of the bony trabeculae on each image on a 1 5 scale, with 5 being the best. They evaluated line pair resolution on all images using a 12 magnification loop. They then evaluated quantum mottle on each image on a 1 5 scale with 5 representing the greatest mottle. They next evaluated visualization of the low-contrast acrylic beads in the soft tissue portion of the phantom, with 1 being least and 5 being most visible. Results The scatter to primary radiation ratio decreased from a value of almost 1 at table top to about 0.4 at a 10-in. (254-mm) air gap and about 0.2 at 25 in. (635-mm) (Fig. 1). Line pair resolution per millimeter on Bucky and 10-inch (254-mm) air gap images was similar (Table 1). A progressive improvement in bony trabecular sharpness and line pair resolution, compared with the table top and Bucky images was noted on the air gap images from 10 through 25 in. (254 635-mm) (Figs. 2, 3, Table 1). Sharpness of the bony trabeculae and line pair resolution were best on the 25-in. (635-mm) air gap images. The quantum mottle was highest on the Bucky image, higher on air gap images and least with the table top technique (Fig. 4). Visualization of the low-contrast beads in the soft tissue area was best on the table top images (Fig. 5). The 25-in. (635-mm) air gap images were next best for detection of the low-contrast beads. A combination of visualization of beads in the soft tissues and trabecular sharpness was best achieved on the 25- in. (635-mm) air gap images (Fig. 6).

121 Fig. 2 Trabecular detail. Scores for each observer and total scores for each imaging technique are shown. The lowest score was seen using the table-top technique, where the scatter to primary ratio is high. The highest scores were achieved at air gaps of 25 in. (635-mm) and 30 in. (762-mm), where the scatter was minimized Fig. 3 Line pair resolution. The line pair resolution for each observer and the total line pair resolution for all three observers are displayed for each imaging technique. The line pair resolution was maximized using a 25-in. (635-mm) air gap Statistical analysis The inter-observer variability was evaluated using weighted Kappa analysis. The average Kappa value for all three paired observers was 0.73 for trabecular sharpness, 0.05 for beads in soft tissues, and 0.52 for quantum mottle. Values of weighted Kappa above 0.75 indicate excellent agreement beyond chance, whereas values below 0.40 indicate poor agreement [3]. Thus, trabecular sharpness alone was evaluated by our observers with fairly minimal inter-observer variability. Discussion Computed radiography (CR) is ideal for pediatric musculoskeletal imaging. Storage phosphor CR is a digitized imaging system, which uses conventional radiographic equipment to expose an image on a photostimulable imaging plate (IP) instead of a film screen combination. Scatter reaching the IP is significantly less when a grid is employed. However, a grid absorbs about 40% of the total radiation emerging from the patient, which makes it necessary to increase the skin entrance dose [4]. The present project demonstrates that air gap radiography further improves musculoskeletal computed imaging by reducing the scatter to primary radiation ratio without an increase in the skin entrance dose. An air gap between the patient and the imaging plate reduces scatter reaching the receptor and also geometrically magnifies the object. Scatter to primary ratio is reduced significantly from a value of almost 1 at table top to about 0.4 at 10-in. (254-mm) air gap and 0.2 at 25-in. (635-mm) air gap placement of the IP (Fig. 1). Sharpness of the bony trabeculae and line pair resolution at 10-in. (254-mm) air gap are better than the table top images. For best evaluation of the sharpness of bony trabeculae and line pair resolution, a 25-in. (635-mm) air gap would be optimal (Figs.2, 3, 6). This represents only 1.6 times

122 Fig. 4 Quantum mottle. The score of each observer for noise and the total score for all three observers is displayed for each imaging technique. The noise was greatest for the Buckey technique because of reduction of both scatter and primary radiation reaching the receptor. The noise was lowest for the table-top technique, where scatter radiation reaching the receptor was maximized Fig. 5 Low-contrast objects (beads) in soft tissue. The scores of each observer and the total score for all three observers are displayed for each imaging technique. The visualization of the beads was best accomplished using the tabletop technique, where the total photon flux (scatter and primary radiation) was maximized geometric magnification. Standard reduction of a 14 in. 17 in. (356-mm 432-mm) image to a 10 in. 12 in. (254-mm 305-mm) hard copy by CR, therefore produces a near life-size image of the object, so that many body parts (elbow, knee, hand, wrist, shoulder) can be imaged using air gap placement of the IP. The quantum mottle is inversely proportional to the total number of photons reaching the imaging plate. When the same exposure factors are used, the quantum mottle is highest on the Bucky image, higher on air gap images and minimal using the table top, non-grid technique (Fig. 4). When the air gap is used, more of the scattered photons do not reach the receptor and the total number of photons incident on the IP decreases resulting in greater quantum mottle. The lower quantum mottle allows maximal visualization of the low contrast densities in the soft tissues (Fig. 5) despite a high scatter to primary radiation ratio on the table top images (Fig. 1). Thus for visualization of low-contrast foreign bodies in the soft tissues, table top placement of the receptor would be the chosen technique. For the best overall imaging technique for soft tissues and trabecular detail, a 25-in. (635-mm) air gap would be chosen (Fig.6). Based on the above experiments, we have demonstrated that the same table top non-grid radiographic exposure technique can be used for air gap imaging with CR. The skin entrance dose does not have to be increased for air gap placement of the IP due to the linear response over a very wide dynamic range of the CR system [5, 6]. Murphey et al. [7] reviewed digitized radiographs of the original radiographs in 42 patients with non-displaced or minimally displaced fractures and found that all fractures were seen on digitized images with 2.88 LP/mm or greater resolution. Following a review of 122 digitized images of plain radiographs selected to represent a wide variety of musculoskeletal lesions, Wegryn [8] reported that 2.5 LP/mm was adequate to visualize all the abnormalities seen on plain radiographs.

123 We know from our work compared with that of previous investigators that the line pair resolution achieved using an air gap of 10 in. (254-mm) or greater is more than adequate for routine diagnostic musculoskeletal imaging [6 12]. To date, we have performed air gap radiography in a few patients in the Pediatric Orthopedic Clinic for evaluation of the appendicular skeleton with good initial results. Modification of the radiographic tables used for acquiring CR images to include an adjustableheight cassette holder underneath a transparent table top for visualization of the field of collimation on the cassette would facilitate easy air gap placement of the IP for routine clinical use of this technique. In conclusion, computed radiography is an alternative to film-screen radiography in the evaluation of the musculoskeletal system in children. A significant reduction in the scatter to primary radiation ratio, improvement in bony trabecular sharpness and line pair resolution can be achieved using air gap placement with CR without an increase in the skin entrance dose compared with the table-top technique. a Fig. 6 a Table top and b 25-inch (635-mm) air gap images of the knee with a line pair phantom. A marked improvement in bony trabecular sharpness and line pair resolution is noted on the 25- inch (635-mm) air gap image compared with the table top image b Acknowledgements We thank Maureen A. Humphries for technical assistance, Drs. Thomas Slovis, Carl Stanitski, and Joseph Reed for acting as independent observers in this study, Mitch Goodsitt, PhD for measurement of scatter to primary radiation and Anthony Schork, PhD for performing the statistical analysis. Statistical support was provided by the University of Michigan Multipurpose Arthritis Center, NIH grant P60-AR220557-12. References 1. Sorenson JA, Nelson JA, Niklason LT, Jacobsen SC (1980) Rotating disk device for slit radiography of the chest. Radiology 134: 227 231 2. Michel M, Ter-Pogossian (1969) The physical aspects of diagnostic radiology, Harper & Row, New York 3. Landes JR, Koch GG (1974) The measurement of observer agreement for categorical data. Biometrics 33: 159 174 4. Tucker DM, Souto M, Barnes GT (1993) Scatter in computed radiography. Radiology 188: 271 274 5. Kuhns LR, Kottamasu SR (1995) Pediatric air-gap chest digital imaging: an experimental study. Pediatr Radiol 25: S99 S201 6. Neitzel U (1992) Grids or air gaps for scatter reduction in digital radiography: a model calculation. Med Phys 19: 475 481 7. Murphey MD, Bramble JM, Cook LT, Martin NL, Dwyer SJ (1990) Nondisplaced fractures: spatial resolution requirements for detection with digital skeletal imaging. Radiology 174: 865 870 8. Wegryn SA (1990) Comparison of digital and conventional musculoskeletal radiography: an observer performance study. Radiology 175: 223 228 9. Murphey MD (1989) Digital skeletal radiography: spatial resolution requirements for detection of subperiosteal resorption. AJR 152: 541 546 10. Richmond BJ, Powers C, Piraino DW, Freed H, Meziane MA, Hale JC, Schluchter MD, Schils J, Gragg LA (1992) Diagnostic efficacy of digitized images vs. plain films: a study of the joints of the fingers. AJR 158: 437 441 11. Wilson AJ (1991) Photostimulable phosphor digital radiography of the extremities: diagnostic accuracy compared with conventional radiography. AJR 157: 533 538 12. Murphey MD, Quale JL, Martin NL, Bramble JM, Cook LT, Dwyer SJ III (1992) Computed radiography in musculoskeletal imaging: state of the art. AJR 158: 1071 1080